National University of Singapore Projects Repository 2018... · In land scarce Singapore, future...
Transcript of National University of Singapore Projects Repository 2018... · In land scarce Singapore, future...
National University of Singapore
PC4199 - Honours Project in Physics
Quantifying Sources of Indoor Radon in OfficeEnvironments
Sam Koh Boon Kiat
A0140608M
Supervisor: Associate Professor Chung Keng Yeow
Acknowledgements
I am grateful to my Supervisor Associate Professor Chung Keng Yeow for giving me the opportunity to pursue
this Final Year Project, and for guiding me throughout the project. I am also thankful to the staff at Singapore
Nuclear Research and Safety Initiative (SNRSI) for their support in my project. In particular, Mr Chan Yi
Meng for assisting me in data collection for this project.
Contents
1 Introduction 4
2 Radon 5
2.1 Problem of Indoor Radon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1 Quantifying Radon Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Preliminary Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Emanation from Concrete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.1 Mass Balance Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.2 Solution of the Differential Equation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Instrumentation 12
3.1 Interior of Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3 Reliability of Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Results 16
4.1 Initial Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Hard Surface Emission Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3 Bulk Emission Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3.1 Results with Concrete Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.3.2 Results without Concrete Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4 Dose of Ionising Radiation Received . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.5 Risk of Lung Cancer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5 Mitigating Factors and Future Work 35
5.1 Use of Sealant or Paint as a Protective Barrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
ii
iii
5.2 On Indoor Spaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3 Improved Bulk Emission Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.4 Predictions using Formula . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6 Conclusion 39
Chapter 1
Introduction
The United Nations Scientific Committee on the Effects of Atomic Radiation reports that over 50% of an
individual’s annual dose to ionising radiation originates from radon. Scientists around the world have conducted
research on the effects of exposure to indoor radon and the source of radon indoors.1 Many studies have reported
that concrete is a major source of indoor radon due to the radium present in the concrete aggregate. A highly
urbanised city like Singapore, sometimes also dubbed the concrete forest, would therefore have an abundance of
indoor radon sources. Despite the health risks that Singaporeans are exposed to on a daily basis, there are not
many studies conducted particularly targeting indoor radon in Singapore.
This report is one of the initial studies conducted on indoor radon in Singapore. In chapter 2, we start with the
generic properties of radon, before discussing the source of indoor radon in Singapore. In chapter 3, we discuss
the instrumentation of the measuring instrument used to measure radon concentrations. Then, in chapter 4 we
present some results based on experiments conducted at various locations in Singapore. Next, we discuss the
possible health effects, and how the numbers obtained in this study affects the average Singaporean. Finally, we
end off in chapter 5 with possible mitigation factors that may be adopted to reduce the concentration of indoor
radon.
4
Chapter 2
Radon
Radon is the sixth heaviest noble gas. Out of the different isotopes of radon, 222Rn has the longest half life of
3.8 days. This is the main isotope that would be of concern to us as it is within the timescale for it to escape
the soil after it is formed from the decay of radium. It is a decay product of the decay chain of Uranium 238,
and since it is unstable, it undergoes further decay in the form of alpha decay.
2.1 Problem of Indoor Radon
Figure 2.1. shows the decay chain of uranium, with radon as an intermediate decay product in the decay chain.
Out of the different intermediate decay products in the decay chain, only radon is in the gaseous state. Radon
present in indoor air can undergo decay, and the daughter nucleus (218 Po) can attach itself to aerosols present.
Aerosols with progenies of radon attached can be inhaled, and be deposited in the lungs. Natural aersols, such
as dust may have a high fraction of it being deposited in the lungs.2 Radon progenies would continue undergoing
a decay while in the lungs and releases alpha particles in the lungs, and biological tissues in the lungs are in the
direct path of the alpha particle. Ordinarily, alpha particles are not penetrating in nature, and sources of alpha
particles outside the body does little damage as the skin provides a sufficient barrier to prevent direct contact
with biological tissues. However, since the alpha decay took place in the lungs, this exposes biological tissues
directly to the highly ionising alpha particle, and it is the proposed mechanism in which radon increases the risk
of lung cancer for individuals. The alpha particle carries with it a large amount of energy, which can increase the
possibility of damage to DNA. It is this highly ionising property that is exploited in Nuclear Medicine therapy,
as it becomes possible to target individual cancer cells.3
The Environmental Protection Agency (EPA) in the United States of America reported that exposure to indoor
5
6
radon is the second leading cause of lung cancer, behind smoking.4 With this statistic in mind, it is therefore
important to quantify, identify, and reduce the sources of indoor radon in a highly urbanised country like
Singapore.
Figure 2.1. Decay Chain of 238U, with 222Rn as an intermediate decay product
In USA, many live in standalone houses, with possibly a basement in their houses. The levels of radon is the
highest at the bottom of the house, where the basement sits. The source of this radon has been traced to the
surrounding soil around the basement. The radon is produced when radium present in the soil undergoes a decay
and releases radon. The radon then enters through cracks on the walls of the basement. In addition, radon is
significantly denser than air. Any source of radon indoors can effectively cause radon to be accumulated in the
lowest point of the dwelling. Hence, while the results of the effects of radon exposure can be applied to the local
context, the source of indoor radon, and possibly, mitigating factors proposed by the EPA may not be applicable
at this point.
In land scarce Singapore, future developments have been planned to be downwards into the earth.5 When
such plans become a reality in the future, the indoor spaces underground will not only have concrete in the
surroundings, but also an abundance of soil. Long-lived nuclei above radon in the decay chain also leads to the
continual source of radon in soil. When terrestrial matter is used in the building materials, this may potentially
increase the amount of radon in such indoor spaces, and hence increase the risks associated with radon exposure.
7
This heightens the importance of a study to quantify the source of indoor radon in Singapore.
Furthermore, radon is a gas that is completely undetectable without the right instrument. It is colourless,
odourless, and chemically inert. Exposure of the general population to radon is therefore difficult to ascertain.
Coupled with the fact that effects on one’s health only sets in in the long term, most would be unaware of the
risks of radon exposure.
2.1.1 Quantifying Radon Concentration
The SI unit that could be used to quantify the concentration of radon in indoor air is Bq m−3 (becquerel per
metre cubed). Other units that are used includes pCi L−1 (pico curies per litre), which is primarily used in the
United States and almost nowhere else. The curies is a historical unit named after Marie and Pierre Curie who
discovered radium in 1898. One curie is the activity from one gram of radium (37 billion decays per second).
Almost everywhere else, including the World Health Organisation (WHO), uses the metric system and measures
radon in Bq m−3. The conversion factor between the two unit systems can then be derived to be
1 pCi L−1 = 3.7× 1010 × 10−12 × 103 = 37 Bq m−3 (2.1)
The action level is the concentration of radon, above which, actions should be taken to actively reduce the
amount of indoor radon. The WHO has recommended the action level to be 100 Bq m−3.6 The EPA, however,
has fixed its action level at 148 Bq m−3, or, in the non conventional units, 4 pCi L−1. Various member states
of the WHO reported having action levels ranging from 100 Bq m−3 to 400 Bq m−3, when surveyed in the
International Radon Project in 2005.7 Most countries also responded that newer buildings have a lower action
level (more stringent requirements) compared to older buildings. Table 2.1. shows the various action levels of
some countries.
Table 2.1: Action Levels of Some Member Countries of the WHO.7
Country For Existing Buildings / Bq m−3 For New Buildings/ Bq m −3
China 1000 (Workplace) 200
Germany 100 100
UK 200 200
USA 148 148
8
2.2 Preliminary Results
A preliminary study was done to measure the level of indoor radon in typical residential and office spaces in
Singapore. The results have been verified, and the trends observed were accountable. Figure 2.2. shows the
variation of radon concentration with time in an enclosed meeting room in an office.
Figure 2.2. Concentration of radon as a function of time in an small meeting room inside
CREATE tower. Shaded out time intervals represents intervals when ventilation system was
switched off.
The preliminary results obtained shows that radon concentration indoors is not a constant, and that it fluctuates
with time. In particular, the trend shows a strong correlation with the ventilation rate of the room. Most studies
conducted have also established outdoor radon to be at very low levels, in the range of 5 - 25 Bq m−3. The graph
is explained by ventilation in the room. Radon accumulates when there is a lack of ventilation, and is removed by
the ventilation system when there is exchange of air with outdoor air, as seen in figure 2.2. Radon concentration
is highly dependent on the ventilation rate of the room. As soon as air was exchanged with outdoor air, the
amount of radon is reduced significantly. Newer, modern buildings built to be more energy efficient are more
well sealed against influx of outdoor air. This reduces cost of air conditioning for offices, and buildings where
the majority of indoor space is air conditioned. However, this also means that natural ventilation and exchanges
9
between indoor and outdoor air is reduced when the air conditioning system is not switched on.1,8–11
Based on the preliminary results, it also suggests that being in the office during office hours when the air
conditioning system is switched on is not a cause of concern. However, the same could not be said during
non-office hours, when the ventilation system is not switched on. The exchange of air is prevented, and radon
is allowed to accumulate in the office environment.
2.3 Emanation from Concrete
Concrete has been identified as the major contributor to indoor radon.12 Concrete is made up of three main
raw materials- cement, aggregate (gravel and sand), and water. The aggregates present in the concrete consists
of gravel and sand in a large proportion. Even though most indoor environments in high rise buildings are
not surrounded by gravel and sand, it is kept inside the concrete and is all around us. The radium present is
primarily responsible for radon being emanated from the surface. This motivates the attempt to describe the
rate of change of radon concentration in a particular room, due to radon emanating from concrete.
2.3.1 Mass Balance Analysis
Consider a source of radon enclosed in a vessel. The rate of change of the concentration of radon (C) may be
attributed to a several factors. These are outlined and shown in the following mass balance equation.
dC
dt=AE
V− λC − q
V(C − Cair)−DC (2.2)
where
• A is the area where radon can emanate, with units of m2,
• E is the emanation rate of radon (Bq m−2 h−1),
• V is the volume of the system, (m3)
• λ is the decay constant of radon (h−1)
• q is exchange (ventilation) rate of air between the vessel and its surroundings, which may be of the form
of leakage (m3 h−1),
• Cair is the concentration of radon in the background (Bq m−3)
• D is the back diffusion coefficient (h−1).
10
The first term represents the contribution of radon from a concrete wall with area A that is exposed for radon
emanation. The second term represents the natural decay of radon. The third term refers to radon removed via
ventilation, or leakage from the vessel. Due to the air exchange, radon present in ambient air (air outside the
vessel) may also be introduced to the vessel. This is accounted for by the fourth term. Finally, the fifth term
represents a back diffusion of radon into the concrete walls.
2.3.2 Solution of the Differential Equation
The differential equation can be rewritten to group terms linear in C together, and terms independent of C
together.
dC
dt= −
(λ+
q
V+D
)C +
AE
V+qCair
V(2.3)
Before solving the differential equation, we can make the assumption that the back diffusion coefficient is
negligible, compared to the other terms.13,14 This value has been determined by other researchers to be much
lower compared to the other terms. With this simplification, the differential equation becomes:
dC
dt= K1 +K2C (2.4)
with
K1 =EA+ qCair
V, K2 = −
(λ+
q
V
)
Using the following integrating factor
exp
(∫−K2 dt
)(2.5)
the differential equation is solved to give the following solution:
C(t) =
[EA+ qCair
V λ+ q
]+
(Cair −
EA+ qCair
V λ+ q
)exp
(−(λ+
q
V
)t)
(2.6)
Equation (2.6) is the main equation that would be used in this research. The results obtained in this project
can be fitted to a curve of the equation described above using MATLAB.
11
If we are just considering the initial rate, then the exponential may be expanded to give the following expression,
which is an expression linear in time.
C(t) ≈ Cair +
[EA+ qCair
V− Cair
(λ+
q
V
)]t (2.7)
Chapter 3
Instrumentation
The field of radon concentration measurement is a highly specialised one. There are companies that provide
radon concentration measurement instruments as part of their huge catalogue of other equipment. The instru-
ment available for this project was obtained from a company specialising in radon concentration measurement.
Durridge is a company based in the United States that manufactures the RAD7 Radon Detector and maintains
them. This is the instrument that was used for this project, and radon concentrations are consistently reported
in Bq m−3.
The RAD7 is an instrument that allows direct sample analysis. It takes in samples of air and detects the
number of alpha particles that arises from the decay of radon and its progenies. Due to the short-lived nature
of radon, it is also not practical to collect a sample of air back for measurement in a laboratory. There are also
other detectors that measures radon concentration. However, such detectors are generally used for long term
monitoring of radon (about 3 months to a year) and not suitable for the purpose of this project.
3.1 Interior of Instrument
The sample holder inside the RAD7 consists of a hemispherical charged shell with volume 0.7 L. The hemisphere
is positively charged, and pushes the positively charged alpha particles and decay products upwards to where a
detector is positioned. The frequency at which the air is sampled from the environment can be set on the RAD7
itself. The RAD7 has an internal pump which takes in air at the specified interval from the surroundings.
12
13
Figure 3.1. Interior of the RAD7.
The hemispherical shell is positively charged so that daughter nuclei produced in the decay of radon will be
repelled upwards towards the detector. Radon progenies are then stuck on the detector, and can undergo further
decay in which the alpha particles are detected. The detection process is not 100% efficient, and the manufacturer
have taken this into account during the calibration process.
The presence of moisture in the interior will adversely affect the efficiency of the detector. As such, a desiccant
is interfaced to the RAD7, in order to dry the air before it enters the chamber. If there is high humidity, it
reduces the efficiency of repulsion of decay products to the detector, which hence reduces the counts of alpha
particles detected. In such an event, the software provided by the manufacturer attempts to correct this, and the
reported radon concentration is corrected to give a higher reading. This is particularly the case over long-term
monitoring, when the desiccant can become depleted.
Just before the air is allowed to enter the RAD7, it is filtered to prevent other particulates from entering the
14
RAD7 to ensure that the counts of alpha particles in the desired energy ranges are solely contributed by the
progenies of radon. This prevents false positives from being recorded in the RAD7, and ensures that counts are
solely due to contribution from radon gas.
3.2 Detection
The RAD7 uses a solid state alpha detector that allows it to perform alpha spectroscopy within the unit itself.
This feature allows the parent nucleus to be identified, and the amount of that parent nucleus can then be
quantified. The part of the decay chain that is relevant for detection within the RAD7 is as follows.
222Rnα−−−−−−→
5.49MeV
218Poα−−−−−−→
6.00MeV
214Pbβ−−→ 214 Bi
β−−→ 214Poα−−−−−−→
7.69MeV214 Pb
As radon decays, the positively charged daughter nuclei are propelled towards the detector. As they undergo
further decay, the alpha particle arising from such decays have a 50% probability of entering the detector and
producing an electrical signal proportional to its energy. After about three hours, when 214Po reaches secular
equilibrium with 222 Rn, this count is also taken into account when computing the concentration of radon.
The end result from a given run is presented in a spectrum, in which a plot of counts of alpha particles at a
given energy is shown against the energy of alpha particles. To facilitate the use of the spectrum provided at
the end of the run, the RAD7 divides the energy axis into different channels, each one depicting a certain range
of energies of alpha particles. Each channel is labelled with the alphabets, and a sample spectrum is shown in
figure 3.2.
Figure 3.2. Sample Alpha Spectrum obtained in a test run.
15
Channel A provides the counts of alpha particles arising from the decay of 218Po, and channel C provides the
counts of alpha particles arising from the decay of 214 Po.
The counts per minute of decay products measured by the RAD7 can then be used to compute the concentration
of radon, which is pre-calibrated by the manufacturer of the RAD7.
3.3 Reliability of Results
Numerous runs were performed using the RAD7, before a second RAD7 was made available for the course of
this project. To ensure that results obtained are independent of which instrument was used, both RAD7s were
switched on simultaneously to measure the radon concentration in the same room, placed slightly less than a
metre apart. The results obtained from this test run are shown below.
Figure 3.3. Data obtained from both RAD7s. The two instruments are placed slightly less than
a metre away from each other, and are sampling the air in the same room.
From the comparison of the data obtained using both RAD7s, it tells us that the measurements obtained are
consistent with each other. The average percentage difference between the two instruments is 7 %.
Chapter 4
Results
Once we are confident the RAD7 is giving us usable data, radon concentrations at various locations, subject to
different conditions can then be measured. In this chapter, we present the initial measurements performed using
the RAD7s and some observations made. Then, we look at the measurements obtained using accessories, and
evaluate if these accessories are fit for its intended purpose.
4.1 Initial Measurements
The initial measurements performed using the RAD7 were mainly for the purpose of familiarising with the
instrument. The initial measurements were conducted at a residential location, and at an office in CREATE
tower. Radon in residential locations appear not to be a problem due to the consistently low values of radon
measured. One of the measurements taken from a residential unit is shown in Figure 4.1. The residential unit is
a HDB unit in Sengkang. During the measurement process, the room was enclosed as tightly as possible. That
is, windows were all closed, and the door was closed for majority of the time during the measurement process.
The gap between the door and the floor was slightly covered during this process.
As this was a preliminary test that was conducted, the settings on the RAD7 were different compared to the other
tests. In this case, each cycle time was 20 minutes. The sample of air is kept in the RAD7’s interior chamber
for 20 minutes before it is refreshed. Nevertheless, the results of this test shows that radon in a residential unit
is relatively low. It is mostly below the action level set by the EPA, and hence, it would not be the focus of this
project.
In this project, at least 5 other runs were conducted with the room less sealed, and the concentration of radon
measured were at or below the levels as shown in Figure 4.1.
16
17
Figure 4.1. Data obtained from initial measurements. This was conducted in a resident unit in
Sengkang.
Additionally, a preliminary study on indoor radon in residential areas in Singapore arrived at the same conclusion,
at other residential units.15 This reinforces the fact that residential radon is not a cause of concern within the
units we have sampled.
The movement of radon after it emanates from concrete walls was also investigated as a preliminary study.
Both RAD7s were placed in adjacent rooms, with the door open. This allows radon to diffuse freely across
the boundary of the two rooms. The two RAD7s were placed approximately 5 metres apart from each other
in an office in CREATE tower. The concentration gradient would drive the diffusion process. The underlying
hypothesis is that diffusion of radon across rooms would be registered, and when the concentration of radon in
one room decreases, the concentration of radon in the adjacent room would increase. The results from this run
is shown in Figure 4.2.
18
Figure 4.2. Concentration of radon placed inside and outside of a meeting room in an office.
The door joining the two adjacent room was kept open during the measurement process.
The results obtained from this run shows that for the 2 rooms system, the concentration of radon does not seem
to differ significantly, and the diffusion coefficient of radon in air is so high that on the time scales at which
measurements are taken, the radon concentration profile is at a steady state. This allows us to proceed with a
zero dimensional assumption in the mass balance analysis equation, and for even smaller systems, we will assume
the concentration of radon to be the same throughout the system.
The variation of radon concentration with time in an enclosed environment was shown earlier in section 2.2-
Preliminary Results. The measurements provides a good and clear gradient during the accumulation of radon
process. This could also be used to determine the emanation rate of concrete from the room, given the dimensions
of the room, and under certain assumptions. The assumptions that we would make are:
• Radon emanates from the 4 vertical walls. Even with furniture placed in directly in front of the wall such
as cupboards, this would not affect the flux of radon into the room.
• There are no other sources of radon in the room apart from the walls.
• Treating the meeting room as a ‘big box’, Cair is taken to be the concentration of radon in the adjacent
19
room, which was shown to be the same as the meeting room itself. This modifies the equation describing
the concentration of radon to be:
C(t) =EA
V λ+
(C(0)− EA
V λ
)exp(−λt) (4.1)
where C(0) is the initial concentration of radon.
Since in the measurement process, only the initial rate was detectable and no saturation concentration was
achieved. This approximates the solution to
C(t) = C(0) +
(C(0)− EA
V λ
)(−λ)t (4.2)
With these assumptions, the curve fitting could be performed, and the emanation rate obtained. By considering
the slopes, and obtaining the physical dimensions of the room, the emanation rates obtained over the course of
three days in the particular room of interest are 21.8 Bq m−2 h−1, 18.7 Bq m−2 h−1, and 25.0 Bq m−2 h−1,
with an average of 21.8 Bq m−2 h−1. Slight variations were expected in the determination of the emanation
rate, given that there are uncertainties in the measurements of radon concentration.
4.2 Hard Surface Emission Chamber
With the initial measurements, and the primary source of radon identified, the hard surface emission chamber
was used to measure the emission from a hard concrete surface. The hard surface emission chamber is also
manufactured by the same manufacturer. It interfaces with the RAD7, and allows for the concentration of
radon in the closed system to be measured. The results obtained was then fitted to the equations in chapter 2,
in order to obtain the empirical constants such as the emanation rate. The hard surface emission chamber was
fixed to the wall using a Caulking cord, also provided by the manufacturer. This seals the hard surface emission
chamber to the wall, and air exchange is kept to a minimum. As the hard surface emission chamber is fixed to
the wall directly, it allows for direct, on-site sample analysis. The chamber can be fixed to the concrete wall and
subsequently removed when the experiment is completed. However, this technique proved to be less than ideal,
as the emission chamber appears to be more suited for a horizontal surface, rather than a vertical wall.
Figure 4.3. shows a schematic of the closed system with the hard surface emission chamber interfaced.
20
Figure 4.3a. Schematic of Hard Surface Emission Chamber attached to a Concrete Wall.
Figure 4.3b. The Hard Surface Emission Chamber used in this project.
The hard surface emission chamber was attached to a concrete wall based in physics department in NUS. The
resulting graph of concentration of radon against time in hours is shown in Figure 4.4.
21
Figure 4.4. Concentration of radon as a function of time obtained by attaching the hard surface
emission chamber to a wall. The chamber is then interfaced to the RAD7.
From the results, it is clear that concrete is a significant contributor of indoor radon. In the absence of a large
ventilation rate to exchange air with the surrounding air, radon can accumulate to a high level, well above the
action level of any country, or the action level set by the WHO. The saturation concentration of radon achieved
in this set up is approximately 1550 Bq m−3.
As the RAD7 measured the concentration of radon in time units of hours, the following calculations will be
done in the corresponding units. For instance, the decay constant of radon will be given as ‘per hour’, with the
numerical value of 0.00756 h−1.
The parameters that needs to be filled into the mass balance equation includes the volume of the setup, the area
of radon emanation as well as the ambient concentration of radon, Cair. The volume of the system is estimated
using values provided by the manufacturer, to be 1520.1 cm3. This was computed by taking into account the
volumes of all apparatus used. The volumes of the hard surface emission chamber, the RAD7, the desiccant, as
well as tubings used in the experiment were taken into account. Next, the area in which radon can be emanated
is computed using the physical values provided by the manufacturer, verified by measurements physically on the
hard surface emission chamber, to be 0.036644 m2. Finally, the last parameter was independently measured by a
second RAD7 running parallel to the one attached to the hard surface emission chamber. The purpose of this is
to measure the concentration of radon in the ambient air, ideally right outside the hard surface emission chamber
22
simultaneous to the RAD7 interfaced to the hard surface emission chamber. To ensure that the concentration
of radon measured in the ambient air is a good representation of the air surrounding the hard surface emission
chamber, the inlet to this RAD7 was placed as close as possible to the hard surface emission chamber. Figure
4.5. shows the background obtained in this run. There are some fluctuations in the values recorded, but the
average value can be taken as the value for Cair. The average value was obtained to be 21.4 Bq m−3.
Table 4.1 summarises the parameters that would be substituted into the curve fitting equation.
Table 4.1: Parameters substituted into Curve Fitting Equation.
Parameter / (units) Value
Volume of system / m3 0.0015201
Area that radon can emanate through / m2 0.036644
Cair / Bq m−3 21.4
Figure 4.5. Background concentration of radon measured by the second RAD7.
The initial rate was first used to obtain the emanation rate. The linear equation (2.7) was used for this purpose.
The first 6 points on the graph was used to fit the linear equation. Using the initial points, the gradient was
determined to have a value of 50 Bq m−3 h−1. This gives an equation for the gradient of the graph.
Gradient =
(Cair −
EA+ qCair
V λ+ q
)(λ+
q
V
)= 50 Bq m−3 h−1 (4.3)
23
This forms one equation. To solve for the unknown parameters E and q, the steady state data has to be used.
This is the saturation concentration of radon as recorded by the RAD7. Due to fluctuations in values recorded
of radon, an average value of 1550 Bq m−3 was adopted as the saturation concentration of radon in this set up.
This forms the second equation.
C(∞) =AE + qCair
V λ+ q≈ 1550 Bq m−3 (4.4)
Solving the two equations simultaneously, the emanation rate and ventilation rate was obtained.
Table 4.2: Parameters obtained using Curve Fitting
Parameter / (units) Value
Emanation rate, E / Bq m−2 h−1 2.1
Ventilation rate, q / m3 h−1 3.8× 10−5
The ventilation rate enters into the equation for the effective decay constant of this set-up. The two terms that
contribute to decay of radon in the system are the natural decay constant, λ, as well as the ventilation of the
system. Based on the curve fit, the following shows the ratio of the decay constant to the removal of radon via
ventilation.
λ
q/V=
0.0076
0.025= 0.304 (4.5)
The ratio implies that the fraction of effective decay constant due to leakage from the system is significant.
The experimental procedure was replicated in a small room in the office of CREATE tower. The hard surface
emission chamber was attached to a concrete floor, and the concentration of radon measured as a function of
time. Ideally, the results from this test would have a lower value of leakage, as it is on a horizontal surface, which
would ensure the system is sealed more tightly. Similar to the previous run in NUS Physics office, the second
RAD7 was measuring the concentration of radon in the ambient air simultaneously. The results obtained from
this is shown in figure 4.6.
24
Figure 4.6. Graph obtained in a small room with hard surface emission chamber attached to
concrete floor. The second RAD7 was monitoring the ambient concentration of radon
simultaneously.
Here, we make the following observations on the trend of data obtained.
• The concentration of radon inside the hard surface emission chamber is consistently higher than the
concentration of radon of the ambient air. This suggests that there is some form of accumulation inside
the chamber, relative to the ambient air.
• There is a correlation in the trend of radon concentration inside the hard surface emission chamber and
the concentration in the ambient air. This suggests that there must have been significant exchange of air
between the interior of the hard surface emission chamber and the ambient air.
The second observation would lead us to the conclusion that the effective decay constant of the hard surface
emission chamber set up in this case is much larger compared to the previous experimental set-up. Due to the
larger effective decay constant, no saturation concentration was achieved. Equilibrium with the ambient air was
attained between measurement times such that each measurement records an already saturated concentration.
This is supported by the observation that there is a strong correlation between the two graphs. This means
that the effective decay constant of this system is so large that the equilibrium, or saturation concentration, is
attained in the order of less than an hour. A closer inspection of the hard surface of concrete studied revealed
that there are cracks on the surface. This would explain the large decay constant observed in this set-up.
25
This is a significant revelation because it might render future tests using the hard surface emission chamber
less valid, and extreme care must be taken to account for such defects in the concrete matrix. For other
tests conducted during this research using the hard surface emission chamber, leakage that was observed could
therefore also have been due to the presence of such inevitable defects in the concrete. Such defects may go
beyond just the surface, and might not be visible. This limits the use of the hard surface emission chamber, and
the use of a bulk emission chamber is proposed to overcome the limitation of cracks present in concrete. The
use of the bulk emission chamber is discussed in the next section.
4.3 Bulk Emission Chamber
Another possible way to measure the emanation rate from a sample of concrete is by means of a bulk emission
chamber. This is a self-made chamber made up of PVC plastic, with two holes drilled into the sides. This allows
the tubes for airflow to be connected to the RAD7, and the concentration of radon in the air of the closed system
to be measured. However, compared to the hard surface emission chamber, this is a more indirect method to
measure the emanation rate from concrete, because a portion of the concrete has to be removed and placed
inside the bulk emission chamber.
Figure 4.7a. Schematic of Bulk Emission Chamber where a cube of concrete is enclosed in it.
26
Figure 4.7b. Bulk Emission Chamber where a cube of concrete is enclosed in it.
This method is considered a destructive method. A sample of the building material is required for this purpose.
In this part, we used the following parameters for the equation. The concentration of radon in the laboratory
is lower compared to the experiment conducted using the hard surface emission chamber. The environment in
the laboratory is always controlled, and the ventilation system is switched on. This promotes the exchange of
indoor and outdoor air, so the concentration of radon is consistently lower compared to the on-site test with the
hard surface emission chamber.
Table 4.3: Parameters substituted into Curve Fitting Equation.
Parameter / (units) Value
Volume of system / m3 0.019304
Area that radon can emanate through / m2 0.06
Cair / Bq m−3 7
There were two main experiments that were performed using the bulk emission chamber. The first involved
a concrete block being placed inside the chamber, and the corresponding radon concentration is measured as
a function of time. The second experiment removes the concrete block from the chamber, and measures how
the concentration of radon will decay with time. Both experiments will be able to determine a value for the
ventilation rate, q. In an ideal situation, the q value obtained from both experiments should agree with each
other.
27
4.3.1 Results with Concrete Block
The concrete block was placed inside the bulk emission chamber, as shown in Figure 4.8. The experiment was
allowed to run for 12 days, in a controlled environment. In the laboratory, the ventilation system is always
switched on, and at no time was the ventilation system being switched off during the measurement procedure.
The data obtained from this gives a significant accumulation of radon in the chamber.
Figure 4.8. Data obtained from a concrete block sealed inside the bulk emission chamber.
By performing a curve-fitting, the equation obtained that describes this curve is
C(t) = 1553.70− 1537.49 exp (−0.009799t) (4.6)
The effective decay constant allows for q to be determined immediately. Subsequently, substitution of the
parameters as shown in table 4.4. yields the following fitted parameters.
Table 4.4: Parameters obtained from curve fitting
Parameter / (units) Value
Emanation rate, E / Bq m−2 h−1 4.88
Ventilation rate, q / m3 h−1 4.244× 10−5
Although there is a range of values for the emanation rate from literature,9,16 the values obtained in this project
are at the same order of magnitude.
28
For the bulk emission chamber, the ratio of the natural decay constant to the decay constant due to ventilation
is larger compared to the ratio obtained from the hard surface emission chamber.
λ
q/V=
0.0076
0.00220≈ 3.5 (4.7)
This implies that the natural decay of radon is more significant compared to the leakage due to ventilation.
Although these two values are at the same order of magnitude, the natural decay term is now larger compared
to the ventilation decay term- an improvement from the hard surface emission chamber. By modifying the fitted
equation such that the ventilation rate, q, becomes zero, an ideal situation can be attained. This allows us to
predict the variation of the concentration of radon in an ideal scenario where there is no leakage from the system.
This is shown in figure 4.9.
Figure 4.9. Comparison of data points obtained vs ideal scenario of no leakage.
4.3.2 Results without Concrete Block
In this experiment, the source of radon was removed by opening the lid, and physically removing the concrete
block before measurements were made. The concentration of radon was allowed to attain a steady state before
the block was removed. This gives a decay curve of radon in the bulk emission chamber, due to both the natural
decay of radon and the ventilation. The curve obtained in this part is shown in Figure 4.10.
29
Figure 4.10. Concentration of radon as a function of time in the bulk emission chamber. The
concentration of radon was allowed to reach saturation levels before the concrete block was
removed.
In this case, the equation to be used for curve fitting was modified slightly due to a differing boundary condition.
Previously, the boundary condition set by the constants of integration assumed the initial radon concentration
to be the same as the ambient concentration, Cair. This is not the case in this experiment. After adjusting for
the modified boundary condition, the following equation was used for the curve fitting instead.
C(t) =qCair
q + V λ+
[C(0)− qCair
q + V λ
]exp
(−(λ+
q
V
)t)
(4.8)
We first examine that the correct limiting cases are obtained.
• The initial concentration is C(0).
• In the long time limit, the limiting concentration is given by
qCair
q + V λ
which, in the limit of large ventilation rate, the limiting concentration is given by Cair, and in the case
where there is no ventilation, all the radon in the bulk emission chamber eventually decays to give a
30
concentration of zero.
Fitting the curve gave the following equation for the experiment conducted without the concrete block in the
bulk emission chamber.
C(t) = 381.1 exp (−0.01028t) + 1.82 (4.9)
With more data points providing the characteristic decay curve, the effective decay constant was first determined,
which allows the value of q to be determined.
Table 4.5: Parameters obtained from curve fitting
Parameter / (units) Value
Ventilation rate, q / m3 h−1 5.154× 10−5
Figure 4.11. Limiting cases together with data points shown.
By adjusting the parameters, we can verify that the decay relation is correct.
Examination of the curves in figure 4.11. we observe that in the absence of any ventilation, i.e. air exchange with
the surrounding ambient air, the decay curve will be less steep, and take a longer time to reach the concentration
of radon in the ambient air. The graph will tend to the value of zero, as all radon present in the chamber will
decay away. The converse is true when the ventilation rate is increased. When the ventilation rate is artificially
31
increased by a factor of 10, the decay slope is much steeper, and the equilibrium with the ambient air is attained
at a much faster time scale. Curves obtained by running the experiments using the same set up will be in
between these two limiting cases. In the data points shown, it can be seen that the leakage is not as significant,
as the curve obtained is closer to the case when q = 0.
4.4 Dose of Ionising Radiation Received
In this section, we compute the dose of ionising radiation received by an individual due to indoor radon. The
generic method to compute the annual dose is in the following equation.
Dose received = EF ×OF × t×D × C (4.10)
where
• EF is the equilibrium factor (assumed to be 0.4 for indoors17),
• OF is the occupation factor,
• t is the total time in a year given in hours,
• D is the dose conversion factor,
• C is the average concentration of radon.
There are two international committees committed to the study of radioisotpes and its effect on public health.
The two oragnisations are the United National Scientific Committee on the Effects of Atomic Radiation (UN-
SCEAR), and the International Commission on Radiological Protection (ICRP). Each organisation published
a value for the dose coefficient for inhaled radon. UNSCEAR published this value to be 9 nSv (Bq m−3 h)−1,
and maintained in 2016 that this value is a realistic and practical value.17 Revised in 2018, the dose coefficient
published by ICRP for inhaled radon takes the value of 6.7 nSv per Bq m−3 h.18
In our calculations, we use the higher value for calculating the dose of ionising radiation received due to indoor
radon. We start by partitioning the general population into two- one that works only during office hours, and the
other that works beyond office hours. An assumption we will be making is that ventilation systems in the office
are switched on only during office hours, and that it does not cater for workers in the office working overtime.
A realistic estimate of the number of hours of a ‘normal’ office worker is 8 hours. The average concentration of
radon in the air inhaled by normal office workers is taken to be the average value during office hours, 9 am to
32
5 pm. This takes a value of 20 Bq m−3. We first compute the dose of ionising radiation received by a normal
office worker.
The occupation factor is computed by accounting the number of hours in the office in a year. In a year with
365 days, assuming that there are 104 (2 × 52) days of the weekend, there are 261 days in the office. With 11
public holidays in a year, this reduces the number of days in the office to be 250 days. From this, the occupation
factor factor can be computed.
OF =8× 250
24× 365= 0.228 (4.11)
The corresponding annual dose of ionising radiation received by a normal office worker is therefore
Dose received = 0.4× 0.228× 8760× 9× 10−6 × 20 = 0.144 mSv (4.12)
This is the dose that a normal office worker would receive under such conditions at the workplace only, and it
is considered on the lower end of the dose received from radon world wide.
For overtime workers, the concentrations of radon in the air they inhale are higher, and also there is a higher
occupation factor. We adjust the occupation factor by making an assumption that they work for 12 hours a day,
say from 9 am to 9 pm. As the dose was received under different conditions, for a more accurate representation,
we consider the following:
Total dose = Dose received during office hours + Dose received during overtime hours (4.13)
In terms of equation 4.10, this is expressed as
Total dose = 0.4× 8760× 9× 10−6(OF (normal)× C(normal) +OF (overtime)× C(overtime)) (4.14)
And with the earlier assumptions, the occupation factor due to overtime work is
OF (overtime) =3× 250
24× 365= 0.0856 (4.15)
The average concentration of radon after the ventilation systems are switched off is higher, but as it has not
accumulated significantly, this value was measured to be 30 Bq m−3. The total dose an overtime worker received
from inhaled radon is therefore given by
33
Total dose = 0.4× 8760× 9× 10−6((0.228× 20) + (0.0856× 30)) = 0.225 mSv (4.16)
Table 4.6. Summarises this set of results for the dose received.
Table 4.6: Annual dose received by normal office workers and overtime office workers.
Annual dose due to inhaled radon / mSv
Normal office worker 0.144
Overtime office worker 0.225
For workers who return to office on a weekend, with even higher radon concentrations, the dose would be
increased further. Depending on the nature of work, the number of weekends a worker would return to office
differs greatly. We will make a reasonable assumption that an overtime worker would return to his workplace
for 5 hours per weekend for one year. The average radon concentration over a particular weekend in the office
sampled was 240 Bq m−3. The annual effective dose received by this worker due to the additional time in office
over the weekend is:
Annual dose received due time in office over weekends = 0.4× 5× 52
24× 365× 8760× 9× 10−6 × 240 = 0.225 mSv
(4.17)
4.5 Risk of Lung Cancer
Health effects of ionising radiation are well known and studied at high doses. The effect of low dose ionising
radiation of 100 mSv and below, however, is not well understood. Low dose ionising radiation can be defined
to be doses of 100 mSv and below. One such model to predict the effects of low dose ionising radiation is
known as the Linear No Threshold model (LNT model). In this model, effects of high dose ionising radiation
are extrapolated to predict the effects of low dose ionising radiation. There is debate on the validity of the LNT
model. The EPA have adopted the LNT model in their research to claim that lung cancer is the second leading
cause of lung cancer in the USA.19 Adopting this model used by the EPA, some calculations will be performed
based on the data obtained in this project.
In the report by ICRP publication 103 published in 2007, with an assumption of the LNT model, the combined
detriment due to excess cancer and heritable effects remains unchanged at around 5% per Sv.20 Translating
this into the probability of cases per million population, it corresponds to a probability of 50 cancer cases per
34
million population per milli-Sievert. Singapore has a population of 5.612 million as of 2017, and we assume that
2 million Singaporeans work in indoor offices. Based on the dose received due to inhaled radon and its progenies
by normal workers, this gives an occurrence of 58 lung cancer cases per year in the workplace.
Number of excess cases of lung cancer = 0.144 mSv× 2× 106 × 50× 10−6mSv−1 ≈ 14 (4.18)
In comparison, the Singapore Cancer Registry’s Annual Registry Report on the Trends in Cancer Incidents in
Singapore for the years 2009 to 2013 published in 2015 reported that the total number of lung cancer cases over
this 5 year period was 6612.21 On average, this amounts to 1322 cases per year.
The number calculated is solely due to exposure to indoor radon by a normal office worker, and the value used
is an internationally recognised value. In reality, the number may vary as no statistics on the actual number of
office workers in Singapore is available, and the conversion factor may be disproved by new research methodology.
Chapter 5
Mitigating Factors and Future Work
Is indoor radon therefore inevitable? Concrete, being essential in the construction of buildings, will continue to
emanate radon. Due to the presence of long-living parents of radon, concrete can continue to emanate radon for
a long time. Some possible methods to reduce the level of indoor radon are proposed.
5.1 Use of Sealant or Paint as a Protective Barrier
Figure 5.1. Measured radon concentration from hard surface emission chamber attached to
laboratory floor.
In one of the experiments to measure the emission of radon from a concrete floor coated with epoxy, the readings
obtained were not within expectations. No significant accumulation was observed. The results obtained in this
35
36
experiment is shown in figure 5.1.
The location chosen for this experiment was a laboratory in CREATE tower, that is underground. Figure 5.2.
shows the on site location of the laboratory.
Figure 5.2. Laboratory where hard surface emission chamber was fixed to the floor to measure
emanation from concrete floor.
The phenomenon of no accumulation was attributed to the presence of a sealant applied over the floor. Due to
the nature of laboratory work, the floor had to be additionally protected in the event of any chemical spills. This
layer of protection, however, acts as a diffusion barrier to radon, which explains why there was no accumulation
of radon in the hard surface emission chamber. Physically, due to this barrier, the increased diffusion constant
increases the amount of time needed for radon to diffuse through the floor. The result is that a greater proportion
of radon would decay before it is replenished with fresh radon from the concrete floor.
In indoor locations with elevated concentrations of radon, similar sealants could be used to prevent the emanation
of radon from concrete into the indoor environment, effectively preventing the build up of radon in a poorly
ventilated area.
37
For office spaces, the use of such sealants might not be feasible. It has also been proposed that wall paints may
be used to mitigate the amount of radon indoors.22 Future work includes investigating the effects of the thickness
of wall paint on the concentration of radon, and how different types of paint may affect the concentration of
radon indoors. Due to the limitation of the hard surface emission chamber as discussed earlier, it would be
more feasible to employ the use of the bulk emission chamber for this purpose. Concrete blocks or slabs may be
used, and the effect of the thickness of paint on the emanation rate of radon from the concrete can be studied
effectively. However, we also note that the best measurement would be a direct measurement of the vertical
surface of a wall on-site, compared to an indirect measurement of a sample of building material. Until the
limitations due to the defects present in concrete is effectively taken into account, the hard surface emission
chamber is not feasible to measure the emanation rate from concrete walls.
5.2 On Indoor Spaces
In order to give a better statement on the total dose one would receive from ionising radiation from radon and its
progenies, more indoor spaces should be sampled so as to obtain a better estimate on the average concentration
of radon one is exposed to. There is expected to be large variations in the concentrations of radon measured,
since radon concentration is highly dependent on the amount of radium present in the concrete. For example,
more offices could be sampled in order to obtain a more representative average value of radon concentration.
More studies on the risk association with indoor for Singaporeans should also be done. With more studies
performed, a more accurate value for the risk of lung cancer associated with indoor radon can be determined.
In this study, the value was adopted from studies performed overseas, and is one that is used internationally.
As shown in the preliminary results, areas with high ventilation rate shows lower concentrations of indoor radon.
Areas with significant build-up of radon should be ventilated actively when occupied by personnel. This will
reduce the concentration of radon present in that indoor environment. For buildings in the future, it may be
beneficial if there could be openings to allow air exchange with the outdoor air when the ventilation systems are
not switched on.
5.3 Improved Bulk Emission Chamber
In this research, there were limitations of the equipment set-up available. For instance, when allowing radon
to saturate in the bulk emission chamber and subsequently removing the concrete block, the concrete block
was removed by opening the lid of the box, and physically removing the block from the chamber. Although
radon is significantly denser than air, this process of removing the concrete block have reduced the saturation
38
concentration by more than half. In an ideal situation, a box containing the concrete block should be separately
constructed, and connected to the bulk emission chamber via tubings available. A pump may then be used
to circulate the air between both housings, so that the concentration is uniform in both. When saturation
concentration is attained, the box containing the concrete block can be detached from the bulk emission chamber
with minimal loss of radon. The tubes are connected to the bulk emission chamber via valves, which will only
open when a tube is connected. This process will be able to achieve a higher starting concentration for the
monitoring of radon concentration without a source.
Additionally, a ‘blank’ bulk emission chamber should be allowed to run so as to ascertain that there is no
contribution of radon from the bulk emission chamber. Although it is highly unlikely that the PVC box will
contribute a significant amount of radon, this should ideally be verified.
5.4 Predictions using Formula
A possible future work for this research topic would be to verify the equation that describes the concentration
of radon, scaled up to a room. Currently, the model has been assumed to work for a larger scale room. This
should be verified by actual measurements on-site. For instance, an initial measurement of a wall at the on-site
location could be done using the hard surface emission chamber, and the emanation rate of radon determined.
Subsequently, the location should be sealed and radon allowed to be accumulated. The emanation rate of radon
could then be determined once again using the whole room monitoring, and then compared to the emanation
rate obtained using the hard surface emission chamber.
Chapter 6
Conclusion
In summary, this project has employed a method to quantify sources of indoor radon. The emanation rates of
radon from concrete was measured on-site using a hard surface emission chamber, and the emanation rate of
radon from a concrete block was measured in the laboratory using a bulk emission chamber. The emanation
rate from a concrete block was measured to be 4.88 Bq m−2 h−1, and the emanation rate from two walls 2.1
Bq m−2, and 21.8 Bq m−2 h−1 respectively. The values obtained were compared to literature values, and agree
in the order of magnitude. Of the two methods, the bulk emission chamber provides greater control over the
experimental parameters, as the environment in the lab is controlled and the background concentration of radon
is generally low. However, this method is more destructive in nature. The more accurate measure is still to
measure the emanation rate on-site, although there is less control over the experimental parameters. The dose
to individuals due to indoor radon at the workplace was then computed, and the excess risk for lung cancer was
computed using values proposed by ICRP.
39
Bibliography
[1] Le, H.; Soo Dong, S.; Nguyen An, S.; Tran Huu, D. Measurement of Indoor Radon Concentration in Dalat
area. Science and Technology Development Journal 2018, 21, 71–77.
[2] Rissler, J.; Gudmundsson, A.; Nicklasson, H.; Swietlicki, E.; Wollmer, P.; Londahl, J. Deposition efficiency
of inhaled particles (15-5000 nm) related to breathing pattern and lung function: an experimental study in
healthy children and adults. Particle and fibre toxicology 2017, 14, 10.
[3] Mango, L.; Pacilio, M. Therapy with Alpha Rays. ARC J. Radiol. Med. Imaging 2016, 1, 1–3.
[4] Environmental Protection Agency, Health risk of Radon. https://www.epa.gov/radon/
health-risk-radon, Date accessed: 27 Feb 2019.
[5] Chandran, R, Digging deep: Singapore plans an underground future. 2018;
https://www.reuters.com/article/us-singapore-landrights-planning/digging-deep-singapore-plans-an-
underground-future-idUSKCN1ON01J, Date accessed: 27 Feb 2019.
[6] World Health Organisation, World Health Organization lowers recommended action level for radon; goal
is to reduce lung-cancer risk. 2009; https://www.consumerreports.org/cro/news/2009/09/world-health-
organization-lowers-recommended-action-level-for-radon-goal-is-to-reduce-lung-cancer-risk/index.html,
Date accessed: 27 Feb 2019.
[7] Zeeb, H. International Radon Project - Survey on Radon Guidelines, Programmes and Activities. World
Health Organisation 2007, 12–13.
[8] Font, L.; Baixeras, C. The RAGENA dynamic model of radon generation, entry and accumulation indoors.
Science of The Total Environment 2003, 307, 55 – 69.
[9] Chao, C. Y.; Tung, T. C.; Burnett, J. Influence of ventilation on indoor radon level. Building and Environ-
ment 1997, 32, 527 – 534.
40
41
[10] Collignan, B.; Powaga, E. Impact of ventilation systems and energy savings in a building on the mechanisms
governing the indoor radon activity concentration. Journal of Environmental Radioactivity 2019, 196, 268
– 273.
[11] Vasilyev, A. V.; Yarmoshenko, I. V.; Zhukovsky, M. V. Low air exchange rate causes high indoor radon
concentration in energy-efficient buildings. Radiation Protection Dosimetry 2015, 164, 601–605.
[12] Yarmoshenko, I.; Vasilyev, A. Effect of energy-efficient measures in building construction on indoor radon
in Russia. Radiation Protection Dosimetry 2016, 174, 419–422.
[13] Abo-Elmagd, M. Radon exhalation rates corrected for leakage and back diffusion Evaluation of radon
chambers and radon sources with application to ceramic tile. Journal of Radiation Research and Applied
Sciences 2014, 7, 390 – 398.
[14] Chao, C. Y.; Tung, T. C.; Chan, D. W.; Burnett, J. Determination of radon emanation and back diffusion
characteristics of building materials in small chamber tests. Building and Environment 1997, 32, 355 – 362.
[15] Ang, W. L. J. Measuring Radon Levels In Singapore. 2018; https://www.physics.nus.edu.sg/student/
Honours%20Projects%20Repository%202017-18/Ang%20Wee%20Lin%20Joyce.pdf, Final year project re-
port.
[16] Sahoo, B.; Sapra, B.; Gaware, J.; Kanse, S.; Mayya, Y. A model to predict radon exhalation from walls
to indoor air based on the exhalation from building material samples. Science of The Total Environment
2011, 409, 2635 – 2641.
[17] UNSCEAR, SOURCES, EFFECTS AND RISKS OF IONIZING RADIATION- Report to the General
Assembly, with Scientific Annexes; 2016.
[18] ICRP, Summary of ICRP Recommendations on Radon; 2018.
[19] John J. Cardarelli, I.; Ulsh, B. A. It Is Time to Move Beyond the Linear No-Threshold Theory for Low-Dose
Radiation Protection. Dose-Response 2018, 16, 1559325818779651.
[20] ICRP, Publication 103- The 2007 Recommendations of the International Commission on Radiological Pro-
tection; 2007.
[21] NRDO, Singapore Cancer Registry Annual Registry Report Trends in Cancer Incidence in Singapore 2009
2013 ; 2015.
[22] Ridha, A. A.; Ayyed, H. K. Using wall paints as a barrier to radon gas emission. Journal of Physical Studies
2018, 22 .