FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at www.alma-alma.fr OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume conversion [MDV399] DSM XS B 2015/04/13 Creation DSM AH Issue Date Nature of modifications Written by Approved by Relevant software 348+v2.02.XX

Transcript of MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is...

Page 1: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24

This document is available at www.alma-alma.fr

OPERATING MANUAL

MU 7038 EN C

GRAVICOMPT MANIFOLD

C 2015/09/08 Volume conversion [MDV399] DSM XS

B 2015/04/13 Creation DSM AH

Issue Date Nature of modifications Written by Approved by

Rel

evan

t sof

twar

e 3

48+

v2.0

2.X

X

Page 2: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 2/24

This document is available at www.alma-alma.fr

CONTENTS

1 GENERAL PRESENTATION AND DESCRIPTION: ...................................................................................4

2 OPERATING RECOMMENDATIONS: ...................................................................................................5

3 OPERATION MODES OF THE INDICATOR DEVICE: ...............................................................................5

4 DRIVER MODE: .................................................................................................................................6

4.1 Menu DELIVERY .......................................................................................................................6

4.1.1 Delivery in preset mode ................................................................................................................................ 7

4.1.2 Delivery in FREE mode .................................................................................................................................. 7

4.1.3 Delivery in not-counted mode ...................................................................................................................... 7

4.2 Menu PRINT ............................................................................................................................8

4.3 Menu DISPLAY .........................................................................................................................9

4.3.1 Sub-menu TOTALISER ................................................................................................................................... 9

4.3.2 Sub-menu MEMORIZATION .......................................................................................................................... 9

4.4 Menu MAINTENANCE ............................................................................................................ 10

4.5 List of alarms ......................................................................................................................... 11

5 SUPERVISOR MODE: ....................................................................................................................... 12

5.1 Menu CALIBRATION ............................................................................................................... 12

5.1.1 Sub-menu ENTER GAUGE VOLUME ............................................................................................................ 12

5.1.2 Sub-menu GAUGE FILLING .......................................................................................................................... 13

5.2 Menu PRODUCTS SETTING ..................................................................................................... 13

5.3 Menu VEHICLE ....................................................................................................................... 13

5.4 Menu SETTINGS ..................................................................................................................... 13

5.4.1 Sub-menu VOLUMES SETTINGS .................................................................................................................. 13

5.4.2 Sub-menu TIMING SETTINGS ...................................................................................................................... 14

5.4.3 Sub-menu DEFAULT SETTINGS .................................................................................................................... 15

5.5 Menu TIME ADJUSTMENT ...................................................................................................... 15

5.6 Menu LANGUAGE .................................................................................................................. 15

5.7 Menu PARAMETERS ............................................................................................................... 15

5.7.1 Sub-menu CONFIGURATION ....................................................................................................................... 15

5.7.2 Sub-menu measuring system EMA ............................................................................................................. 16

6 METROLOGICAl MODE: .................................................................................................................. 17

6.1 Menu INDICATOR REFERENCE ................................................................................................ 17

6.2 Menu CONFIGURATION.......................................................................................................... 17

6.2.1 Sub-menu COMPARTMENTS....................................................................................................................... 17

6.2.2 Sub -menu UNIT AND ACCURACY ............................................................................................................... 18

6.2.3 Sub-menu CONVERSION ............................................................................................................................. 19

6.3 Menu measuring system EMA ................................................................................................ 19

6.3.1 Sub-menu METER COEFFICIENT .................................................................................................................. 20

Page 3: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 3/24

This document is available at www.alma-alma.fr

6.3.2 Sub-menu CORRECTION ............................................................................................................................. 20

6.3.3 Sub-menu METER FLOWRATES ................................................................................................................... 20

6.3.4 Sub-menu VOLUMES .................................................................................................................................. 20

6.3.5 Sub-menu TEMPERATURE .......................................................................................................................... 21

6.3.6 Sub-menu DETECTOR .................................................................................................................................. 21

6.4 Menu DATE AND TIME ........................................................................................................... 21

ANNEXE ................................................................................................................................................. 22

RELATED DOCUMENTS ........................................................................................................................... 24

Page 4: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 4/24

This document is available at www.alma-alma.fr

1 GENERAL PRESENTATION AND DESCRIPTION:

The GRAVICOMPT MANIFOLD is an interruptible measuring system. It is intended for delivery of liquids other than water. Liquids flow from each of the compartments of a road tanker connected to the manifold.

The GRAVICOMPT MANIFOLD measuring system is fitted with: � A turbine meter � A MICROCOMPT+ electronic calculator-indicator device � A differential pressure sensor � A gas detection sensor located upstream of the turbine meter � A transfer valve which regulates flow � Air-operated gates connecting each compartment with the manifold � A non-return vent valve to ensure the manifold fills and empties properly � A printer (option). The calculator-indicator MICROCOMPT+ allows to display: � Either volume at metering conditions – Vm � Or volume converted to base conditions – Vb.

It is designed to measure volumes of liquid by gravity (pre-set or not). An option takes into account the temperature of liquid.

The calculator-indicator controls one manifold connected to up to 10 compartments with a maximum of 16 products which names are configurable

In option, it may print delivery tickets, internal totaliser, parameters, and events diary

NOTE: Information printed on the printer has no metrological value. Only the values displayed by the main indicating device providing proof.

Presentation of the MICROCOMPT+ calculator-indicator:

Sealed cup protecting access

to metrological functions

SUPERVISOR magnetic key to access configuration and calibration menu

Page 5: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 5/24

This document is available at www.alma-alma.fr

Buttons function:

The MICROCOMPT+ calculator-indicator manages measuring operation and computerizes the measuring system defaults.

2 OPERATING RECOMMENDATIONS:

Before using a GRAVICOMPT MANIFOLD measuring system, the operator must ensure that the following conditions have been met:

� The piping linking each compartment and the transfer valve must have a minimum pitching of 3%. The vehicle on which the measuring system is installed must be fitted with a device to ensure it is horizontal

� The end-of-counting probe is placed so that it can detect the vacuity of the collector on the smallest free surface.

3 OPERATION MODES OF THE INDICATOR DEVICE:

Driver mode This is the normal using mode in exploitation. Refer to DRIVER MODE

Supervisor mode To access the supervisor mode, the magnetic key must be put at the right of the MICROCOMPT display. This mode is used to set the measuring system and to access the calibration menu. Refer to SUPERVISOR MODE for setup.

Metrological mode To access the METROLOGICAL mode, the MICROCOMPT has to be unsealed. Only an authorized person can remove the seal. This mode allows setting all metrological parameters. It’s done at the putting into use of the measuring system and sometimes during metrological controls. Refer to METROLOGICAL MODE for configuration.

Page 6: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 6/24

This document is available at www.alma-alma.fr

4 DRIVER MODE:

4.1 Menu DELIVERY

A delivery includes several operations. An operation includes the following stages:

• Selection of the compartment • Selection of the product • Selection of the distribution mode: preset, free or not counted.

At the end of an operation, press MENU to start a new operation.

At the end of the delivery, choose the menu END DELIVERY? Validate with OK.

MODE�PRESET

MODE�FREE

PRODUCT�GO

PRODUCT�SP95

END DELIVERY?

DELIVERY XC1 / -- -- -- -- --

MODE�NOT COUNTED

Place the delivery ticket in the printer before validating

>DELIVERY TICKET? >ADD PAPER >PRINTING FINISHEDPRINTING...

Remove the ticket

During the measurement, some information may be displayed by pressing MENU:

Unit: m3/h or L/minINSTANT. FLOW

PRODUCT HEIGHT

471 kg/mm/min /h

PRODUCT HEIGHT

INSTANT. FLOW HF

59,3 L kg/m3

/min /h

with active optionINSTANTANEOUS TEMP.

+18,3 kg/°C/min /h

INSTANTANEOUS TEMP.

OK

STOP

DELIVERY

PRINT

DISPLAY

MAINTENANCE

MENU

Back to normal display is automatic.

DON’T PRESS STOP PUSHBUTTON.

Page 7: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 7/24

This document is available at www.alma-alma.fr

4.1.1 Delivery in preset mode

Validate the delivery mode MODE�PRESET and set the volume. Pressing OK makes the preset unloading begin. If the preset volume is lower than the authorised volume, the measurement will be invalidated at the end of the operation; it will be displayed alternately with dashes: ‘-- -- --’.

1000 LEND OPERATION

1000 LCONTINUE OPERATION

1000 LOPERATION FINISHED

MODE�PRESET 000000 LENTER PRESET

1000 LC1/ GO / START

0 LFILLING

256 LC1/ GO /PRESET

1000 LDRAIN IN PROGRESS

256 LSTOP OPERATION

1000 LDRAIN IN PROGRESS

IF OPERATION STOPPED

4.1.2 Delivery in FREE mode

Validate the distribution mode MODE�FREE. Pressing OK makes the unloading begin in order to empty the compartment. Pressing the STOP button stops the unloading.

MODE�FREE 0 LC1/ GO / START

256 LC1/ GO /FREE

2259 LDRAIN IN PROGRESS

2259 LOPERATION FINISHED

256 LSTOP OPERATION

256 LEND OPERATION

256 LCONTINUE OPERATION

256 LC1/ GO / START

256 LDRAIN IN PROGRESS

256 LOPERATION FINISHED

0 LFILLING

4.1.3 Delivery in not-counted mode

Validate the distribution mode MODE�NOT COUNTED. The unloaded volume is not verified by the measuring system (measuring container). This delivery mode is used when the end of counting probe is out of order).

MODE�NOT COUNTED ENTER VOLUME 00000 LENTER VOLUME

2300 LEND DELIVERY?

- - - - - - LEND DELIVERY?

Remember to follow any regulations in force

Page 8: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 8/24

This document is available at www.alma-alma.fr

4.2 Menu PRINT

PRINT

For printing a daily summary.

SUMMARY 19.07DATE (DD.MM)

DELIVERY TICKET >ADD PAPER >PRINTING FINISHED

The delivery ticket of the last measurement is printed when this item is chosen.

TOTALISER >ADD PAPER

For printing the general totaliser.

PARAMETERS >ADD PAPER

For printing the metrological parameters and the setting value stored in the calculator device.

EVENTS RECORDED 02.07.15DATE (DD.MM.YY)

>ADD PAPER

>ADD PAPER

PRINTING...

PRINTING...

PRINTING...

PRINTING...

PRINTING...

Page 9: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 9/24

This document is available at www.alma-alma.fr

4.3 Menu DISPLAY

4.3.1 Sub-menu TOTALISER

Visualization of the general totaliser.

00011 LINDEX TOTAL 00011548

548 LINDEX TOTAL 00011548

General totaliser displayed in liters.

4.3.2 Sub-menu MEMORIZATION

This menu is available in stand-by mode or during an intermediate stop. It allows the proofreading of all the measurement results stored by the GRAVICOMPT MANIFOLD. That can be done in two ways: LIST: Display all the measurement details recorded, from the newest to the oldest, sorted by day then by measurement number. SELECTION: Display a specific measurement by selecting the day number For each measurement, are displayed: the number and the name of the product, the measured volume, temperature and density with active options. MEMORIZATION LIST

SELECTION

000217INDICATOR REFERENCE

202DAY NUMBER

003MEASUREMENT NUMBER

000217INDICATOR REFERENCE

202DAY NUMBER

003MEASUREMENT NUMBER

001000 LCOMPARTMENT X

750.0 kg/m3

PRODUCT X/ XXX

000990 LCOMPARTMENT X

Vm

Vb

+22.9 °CMEAN TEMPERATURE

002MEASUREMENT NUMBER

000400 LCOMPARTMENT X

750.0 kg/m3

PRODUCT X/ XXX

Vm

If conversion is ON

If temperature is ON

750.0 kg/m3

PRODUCT X/ XXX

If conversion is ON

Page 10: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 10/24

This document is available at www.alma-alma.fr

4.4 Menu MAINTENANCE

EMA/ECP (XX)MAINTENANCE

842 mmPRODUCT HEIGHT

Status of the end of counting probe which can be:DRY, WET or FAULT (out of order).

Instantaneous height of product in the compartiment (mm).

+21 °CINSTANTANEOUS TEMP.

Instantaneous temperature of the product in the compartment (°C).

Page 11: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 11/24

This document is available at www.alma-alma.fr

4.5 List of alarms

Apparition of a default makes the pouring stop by closing the transfer valve. At the same time, the MICROCOMPT+ displays the associated alarm message. The operator must deal with the default and then validate the alarm.

BL

OC

KIN

GN

ON

BL

OC

KIN

G

MEMORY OVER LOADED Delivery diary is full

DISPLAY MEANING ACTION

MEMORY LOST (PILE)

MEMORY LOST

PROM DEFAULTRAM DEFAULT

EEPROM MEMORY LOST

Loss of saved memory

Delivery diary lost

Loss of software or resident integritySaved memory fault

Loss of metrological configuration

Substitution of the backup battery

Substitution of the AFSEC+ electronic card

TEMPERATURE FAULT

DISPLAY DEFAULT

Temperature determination failure

Problem with display card

TOTALISER LOST Loss of totaliser

If steady alarm, substitution of the display card

Substitution of the backup battery

Check the temperature probe status /If steady alarm, see a reparator for trouble shooting

LOW FLOW DEFAULT Low flowrate (less than 4m3/h)HIGH FLOW DEFAULT

PRINTER DEFAULTPOWER SUPPLY PROBLEM

ZERO FLOW DEFAULT

DATE AND TIME LOST

STOP OPERATION

High flowrate (greater than maximum flowrate)

No more communication with the printerPower outage during operation

Zero flow

Loss of date and time

Intentional interruption of discharge

Check the parameters / Reduce flowrate

Check the connection cable, on-off switch and fuseCheck the cause / Restore power supply

Check if the pulse transmitter is powered (red indicators)

Set date and time in supervisor mode (supervisor key)

Continue, stop or finish the operation

Check the hydraulic system (valve, strainer, nozzle…)

METERING PROBLEM

FLAP LEAK DEFAULT

Metering problem with the measuring device

Product leakage from a flap

Check if the pulse transmitter is powered (red indicators)

Check the flap

US

ER

Enter and exit the METRO mode / If steady alarm, substitution of the backup battery

Substitution of the AFSEC+ electronic cardSubstitution of the AFSEC+ electronic cardSubstitution of the AFSEC+ electronic card

RE

PA

IRE

R

If steady alarm, see a reparator for trouble shooting

DIARY DEFAULT Reset of the events diary Acknowledge the alarm, check the date in supervisor mode (supervisor key)

MANIFOLD NOT EMPTYThe manifold is not empty at the beginning of the

operationFollow the manifold release sequence

GAS DEFAULT Détection d’air en phase de grand débit

END DG DEFAULT End of pouring probe out of orderPRESSURE DEFAULT Pressure determination failure If steady alarm, see a reparator for trouble shooting

If steady alarm, see a reparator for trouble shooting

WATCHDOG DEFAULT Fault with display or power card or AFSEC+ cardSwitch on-off the MICROCOMPT+ /

If steady alarm, substitution of the faulty card

VOLUME CONVER DEFAULT Problem during volume conversion If steady alarm, substitution of the AFSEC+ electronic card

Page 12: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 12/24

This document is available at www.alma-alma.fr

5 SUPERVISOR MODE:

MENU

STOP

OK

SETTINGS

VEHICULE (XX)

CALIBRATION / GAUGE

PRODUCTS SETTINGS

TIME ADJUSTMENT

LANGUAGE (XX)

PARAMETERS

5.1 Menu CALIBRATION

CALIBRATION / GAUGE ENTER GAUGE VOLUME

GAUGE FILLING

5.1.1 Sub-menu ENTER GAUGE VOLUME

This menu allows you to check the accuracy of the measuring system by calculating the measuring device error and the new corrected coefficient. First, use the GAUGE FILLING menu to calibrate the measuring system. Enter the volume read on the gauge and validate. The following information is then displayed: � The signed error in % � The coefficient revised as a function of the error � The average flow of the delivery

Each of these values is shown step by step and by pressing the green button OK. Then the product in the manifold is released.

00598.6 LENTER VOLUME (REF)

+00.17ERROR (PERCENT)

11.4892COEFFICIENT (P/L)

ENTER GAUGE VOLUME 47.3AVERAGE FLOWRATE

kg/m3

/h

00598.6 LENTER VOLUME (REF)

+00.17ERROR (PERCENT)

11.4892COEFFICIENT (P/L)

47.3AVERAGE FLOWRATE

L kg/m3

/min /h 34 LMANIFOLD DRAINING?

34 LRELEASE PRODUCT

34 LOPERATION FINISHED

After menu GAUGE FILLING

Page 13: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 13/24

This document is available at www.alma-alma.fr

5.1.2 Sub-menu GAUGE FILLING

This menu is used for filling the gauge with keeping the manifold full of product. Use it the same way as the user mode; at the end of the operation, the manifold is not drained. At the end of the operation, the display returns to ENTER GAUGE VOLUME menu after pressing the green button OK. GAUGE FILLING MODE�PRESET 000000 L

ENTER PRESET

600 LOPERATION FINISHED

0 LCX/ SP95 / START

0 LFILLING

600 LCX/ SP95/ PRESET

PRODUCT�SP95 0 LC1 / -- -- -- -- --

5.2 Menu PRODUCTS SETTING

Definition of products. For the 5 first products, default names are proposed. To remove a product, enter blank space as name product. Maximum number of characters: 5

PRODUCTS SETTING 1CHOOSE PRODUCT

01/NAME (GO ) 01/NAME�GO

01/CORRECTION (XX) 01/CORRECTION�OFF

01/CORRECTION�ON

16CHOOSE PRODUCT

16/NAME ( ) 16/NAME�

16/CORRECTION (XX) 16/CORRECTION�OFF

16/CORRECTION�ON

5.3 Menu VEHICLE

Set the vehicle registry number on which the GRAVICOMPT MANIFOLD is installed. This number will be printed on delivery tickets, invoices…

VEHICLE (AA--000--AA) VEHICLE�AA--000--AA

5.4 Menu SETTINGS

5.4.1 Sub-menu VOLUMES SETTINGS

This menu allows setting the volume parameters: VOLUME LF�HF: Set the volume (in liters) beyond which the MICROCOMPT+ controls the high flowrate at the beginning of the measurement. Example: "0030" for 30 litres. VOLUME HF�LF: Set the volume (in liters) beyond which the MICROCOMPT+ controls the low flowrate at the end of a preset measurement when the end of counting probe is still wet.

VOLUMES SETTINGS 0 LVOLUME LF�HF

0000 LVOLUME LF�HF

10 LVOLUME HF�LF

0010 LVOLUME HF�LF

Page 14: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 14/24

This document is available at www.alma-alma.fr

5.4.2 Sub-menu TIMING SETTINGS

This menu allows setting the duration parameters: OPENING INCREMENT(S): Set the command increment duration of the API adapter opening valve (in seconds). Minimal value: 0.03 second. Maximal value: 3.999 seconds. Default value: 0.070 second (70 millisecondes). OPENING RELAX.(S): Set the relaxation duration between two API adapter opening command increments (in seconds). Maximal value: 3.999 seconds. Default value: 1 second. CLOSING INCREMENT(S): Set the command increment duration of the API adapter closing valve (in seconds). Maximal value: 3.999 seconds. Default value: 0.070 second (70 millisecondes). CLOSING RELAX.(S): Set the relaxation duration between two API adapter closing command increments (in seconds). Maximal value: 3.999 seconds. Default value: 1 second. ANTI-VORTEX STOP(S): Set the API adapter closing duration after an ANTI-VORTEX breakdown. Minimal value: 5 seconds. Maximal value: 99 seconds. Default value: 5 seconds. MANIFOLD FILLING(S): Set the manifold filling duration (in seconds). Minimal value: 20 seconds. Maximal value: 59 seconds. Default value: 30 seconds. MANIFOLD DRAINING(S): Set the manifold draining duration (in seconds). Minimal value: 20 seconds. Maximal value: 59 seconds. Default value: 30 seconds. WET PROBE: Set the maximum duration before the end of counting probe become wet (in seconds). Minimal value: 20 seconds. Maximal value: 99 seconds. Default value: 10 seconds. TIMING SETTINGS 0.070

OPENING INCREMENT(S)

0.070OPENING INCREMENT(S)

1.000OPENING RELAX.(S)

1.000OPENING RELAX.(S)

0.070CLOSING INCREMENT(S)

0.070CLOSING INCREMENT(S)

1.000CLOSING RELAX.(S)

1.000CLOSING RELAX.(S)

5ANTI-VORTEX STOP(S)

05ANTI-VORTEX STOP(S)

30MANIFOLD FILLING(S)

30MANIFOLD FILLING(S)

30MANIFOLD DRAINING(S)

30MANIFOLD DRAINING(S)

10WET PROBE (S)

10WET PROBE (S)

Page 15: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 15/24

This document is available at www.alma-alma.fr

5.4.3 Sub-menu DEFAULT SETTINGS

This menu allows setting the temperature and density default values when CONVERSION is ON.

DEFAULT SETTINGS +15.0 °CTEMPERATURE

+15.0 °CENTER TEMPERATURE

750.0 kg/m3

DENSITY

750.0 kg/m3

ENTER DENSITY

5.5 Menu TIME ADJUSTMENT

Date and time are set in METROLOGICAL mode. The hour may be adjusted (±2h) one time a day through this menu.

TIME ADJUSTMENT 14.41TIME (HH:MM)

e.g. 14.14 means 2:41 pm

5.6 Menu LANGUAGE

This menu is available if a translation catalogue has been uploaded in the MICROCOMPT+. The message “INCORRECT TRANSLATION CATALOG” appears in the event that no catalogue is uploaded.

LANGUAGE (XX) EN

FR

Choose the display language.

5.7 Menu PARAMETERS

This menu allows displaying the parameters set in METROLOGICAL MODE. See the related chapter for the meaning.

PARAMETERS 00000INDICATOR REFERENCE

CONFIGURATION

EMA

5.7.1 Sub-menu CONFIGURATION

See the METROLOGICAL MODE chapter for the meaning of each parameter.

Page 16: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 16/24

This document is available at www.alma-alma.fr

CONFIGURATION Compartments managed by the GRAVICOMPT.

N° of the first compartment.

UNIT AND ACCURACY Unit of flowrate 123.4FLOWRATE UNIT

L kg/m3

/min /h

LOW FLOW HEIGHT XX mmC X/LF HEIGHT Height thresholds below

which the unloading is done in low flow.

XCOMPARTMENT NUMBER

XFIRST COMPARTMENT

XX mmC X/LF HEIGHT

5.7.2 Sub-menu measuring system EMA

See the METROLOGICAL MODE chapter for the meaning of each parameter.

EMA

200 LMINIMUM QUANTITY

DETECTOR (XX) Type of gaz detector on the manifold: ATEX (DG3001) or non-ATEX (DG3000).

24.0 LMANIFOLD VOLUME

5.0 LFIXED VOLUME

00.0000METER COEFFICIENT

Coefficient of the meter of the measuring system (in pulses/liter).

+0CORRECTION(THOUSAND)

SETTINGS VOLUMES SETTINGS

FLOW RATES SETTINGS 8.0MINIMUM FLOWRATE

L kg/m3

/h

80.0MAXIMUM FLOWRATE

L kg/m3

/h

18.0OBJECTIVE LOW FLOW

L kg/m3

/h

Correction factor per thousand (‰) of the measuring system for a measurement with high viscosity products.

Page 17: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 17/24

This document is available at www.alma-alma.fr

6 METROLOGICAL MODE:

INDICATOR REFERENCE

DATE AND TIME

MENU

STOP

OK

CONFIGURATION

EMA

6.1 Menu INDICATOR REFERENCE

Set the GRAVICOMPT serial number (5 numeric values). 00000ENTER REFERENCE

00000INDICATOR REFERENCE

6.2 Menu CONFIGURATION

CONFIGURATION

UNIT AND ACCURACY

COMPARTMENTS

CONVERSION (XX)

6.2.1 Sub-menu COMPARTMENTS

This menu is to describe the compartments configuration. COMPARTMENT NUMBER: Set the compartments number depending on the measuring system (maximum 7 or 10). FIRST COMPARTMENT: Set the first compartment to determine which compartments will be useful. LOW FLOW HEIGHT: Set the compartment height threshold in mm. Below this threshold, the unloading will be done in low flow. The reference zero is the tapping of the differential pressure transmitter. (Example: set "00635" for a 635mm height from the tap point).

Page 18: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 18/24

This document is available at www.alma-alma.fr

LOW FLOW HEIGHT 100 mmC X/LF HEIGHT

COMPARTMENTS 0COMPARTMENT NUMBER

0FIRST COMPARTMENT

01ENTER CPT. NUMBER

00100 mmENTER LF HEIGHT

100 mmC X/LF HEIGHT

00100 mmENTER LF HEIGHT

X = COMPARTMENT NUMBER

01ENTER COMPARTMENT

6.2.2 Sub -menu UNIT AND ACCURACY

Choose the flow rate unit which is displayed and printed.

UNIT AND ACCURACY

1234CHOSSE UNIT

L kg/m3

/min /h

123.4FLOWRATE UNIT

L kg/m3

/min /h 123.4CHOOSE UNIT

L kg/m3

/min /h

Page 19: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 19/24

This document is available at www.alma-alma.fr

6.2.3 Sub-menu CONVERSION

CONVERSION (XX) CONVERSION�OFF

CONVERSION�ON

+15.0 °C

REFERENCE TEMP.

+15.0 °C

ENTER REFERENCE TEMP.

MAIN DISPLAY (XX) MAIN DISPLAY�VM

MAIN DISPLAY�VBASE

+15.0 °C

DENSITY TEMP. (REF)

+15.0 °C

DENSITY TEMP. (REF)

CONVERSION FORMULA (XX) >API54A

>API54B

>LPG

>EN14214

>ETH15

DENSITY AT XX 1CHOOSE PRODUCT

16CHOOSE PRODUCT

750.0 kg/m3

DENSITY AT 15 (1)

>ETH20

>FAME

>ETBE

DENSITY RANGE 500.0 kg/m3

MIN DENSITY

500.0 kg/m3

MIN DENSITY

999.9 kg/m3

MAX DENSITY

999.9 kg/m3

MAX DENSITY

XX=DENSITY TEMP. (REF)

Example for XX=15°C

6.3 Menu measuring system EMA

EMA

CORRECTION(THOUSAND)

METER COEFFICIENT

METER FLOWRATES

TEMPERATURE

VOLUMES

DETECTOR

Page 20: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 20/24

This document is available at www.alma-alma.fr

6.3.1 Sub-menu METER COEFFICIENT

Set the meter coefficient of the measuring system in pulse per liter (p/l).

0.0000METER COEFFICIENT

00.0000ENTER COEFFICIENT

6.3.2 Sub-menu CORRECTION

Set the correction factor per thousand (‰) of the measuring system for a measurement with high viscosity products.

+0CORRECTION(THOUSAND)

+0ENTER CORRECTION

6.3.3 Sub-menu METER FLOWRATES

MINIMUM FLOWRATE: Set the metrological minimal flowrate of the measuring system in m3/h or L/ m3 depending on the configured flow unit. MAXIMUM FLOWRATE: Set the metrological maximal flowrate of the measuring system in m3/h or L/ m3 depending on the configured flow unit. OBJECTIVE LOW FLOW: Set the objective low flow in m3/h. In low flow phases, a regulation will be done around this value with a tolerance of ±3m3/h. This value increased by 3 must be less than the maximum flowrate.

METER FLOWRATES

18.0OBJECTIVE LOW FLOW

L kg/m3

/h

008.0ENTER MINIMUM FLOW

L kg/m3

/h

080.0ENTER MAXIMUM FLOW

L kg/m3

/h

018.0ENTER OBJECTIVE LF

L kg/m3

/h

80.0MAXIMUM FLOWRATE

L kg/m3

/h

8.0MINIMUM FLOWRATE

L kg/m3

/h

6.3.4 Sub-menu VOLUMES

MINIMUM QUANTITY: Set, in liters, the minimum measured quantity of the measuring system to guaranty the measurement (authorized volume) MANIFOLD VOLUME: Set the volume of the manifold in liters (depends on the compartments number). FIXED VOLUME: Set the end of counting fixed volume of the measuring system in liters.

200 LMINIMUM QUANTITY

0200 LMINIMUM QUANTITY

VOLUMES

24.0 LMANIFOLD VOLUME

24.0 LMANIFOLD VOLUME

5.0 LFIXED VOLUME

05.0 LFIXED VOLUME

Page 21: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 21/24

This document is available at www.alma-alma.fr

6.3.5 Sub-menu TEMPERATURE

This menu os an option. It is used to calibrate the temperature into the MICROCOMPT+. Refer to FM 8510.

TEMPERATURE (XX) TEMPERATURE�OFF

TEMPERATURE�ON +99.9 °CCALIBRATE TEMP.

+99.9 °CENTER TEMPERATURE

Operating with temperature or not.

Calibration of temperature.

-20.0 °CMINIMUM TEMPERATURE

-20.0 °CENTER MINIMUM TEMP.

+50.0 °CMAXIMUM TEMPERATURE

+50.0 °CENTER MAXIMUM TEMP.

LOCKED ALARM (XX) LOCKED ALARM�OFF

LOCKED ALARM�ON

6.3.6 Sub-menu DETECTOR

Set the gas detector model for the end-of-counting probe:

3001: ATEX gas detector which status needs to be validated. The end-of-counting probe must be dry before validating the ‘dry’ status. 3000: Non ATEX gas detector.

DETECTOR (XX) DETECTOR�3001 Type of gaz detector on the manifold: ATEX (DG3001) or non-ATEX (DG3000).

DETECTOR�3000

DETECTOR�DRY

6.4 Menu DATE AND TIME

DATE AND TIME 14.41TIME (HH:MM)

Enter the day, the month and the year, and then enter hour and minutes (e.g.14.41 means 2:41 pm).

16.07.15DATE (DD.MM.YY)

Page 22: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 22/24

This document is available at www.alma-alma.fr

ANNEXE

SUMMARY: PARAMETERS: Print the daily summary Print the calculator parameters

GRAVICOMPT 348+.000 BOARD v1r8VERSION 02.02.00 OF 30.06.15VEHICULE: AA-000-AAINDICATOR: 03000 PRINTED ON THE 21/07/15 at 10h20

SUMMARYOF MEASUREMENTS OF 21/07/15DAY 202 003 MEMORISED RESULTS

**** DAILY TOTALISERS ****

GO (1) : 00009908 LSP95 (2) : 00000000 LE-10 (3) : 00000000 L SP98 (4) : 00000000 LFOD (5) : 00008066 L

TOTAL FROM 1 TO 16 : 00017974 L

************** SUMMARY **************N CPT PRODUCT TEMP VOLUME 001 3 (1) GO 22.9°C *00316 L 002 3 (1) GO 22.9°C 01000 L 003 4 (5) FOD 22.9°C 01500 L 004 4 (5) FOD 22.9°C 00020 L 005 4 (5) FOD 22.9°C 01400 L 006 3 (1) GO 22.9°C 02539 L 007 3 (1) GO 22.9°C *02200 L

* : NO GUARANTEED DELIVERY

If the Temperature option is not activated, FLOW is printed instead of the product temperature 'TEMP'.

************* PARAMETERS *****************

COMPARTMENT NUMBER : 5INDEX BEGINNING : 1LF HEIGHT (MM):C1 (1234) C2(1234) C3(1234) C4(1234)C5 (1234) HEIGHT : 470 MMFLOWRATE UNIT : M3/HCOEFFICIENT : 02.9000 P/LCORRECTION COEFF : +0 °/ooMIN FLOW: 8.0 / MAX : 80.0 M3/HOBJECTIVE LOW FLOW : 18.0 M3/HMINIMUM QUANTITY : 200 LMANIFOLD VOLUME : 24.0 LFIXED VOLUME : 5.0 LTEMPERATURE : ONGAS DETECTOR : DG3001CONVERSION : VBCONVERSION FORMULA : API54BMIN DENSI:500.0 / MAX :999.9 KG/M3PRODUCT NAME CORRECT DENSITY PROD 1 GO OFF 750.0 KG/M3 PROD 2 SP95 ON 750.0 KG/M3 PROD 3 E-10 ON 750.0 KG/M3 PROD 4 SP98 ON 750.0 KG/M3 PROD 5 FOD OFF 750.0 KG/M3VOLUMES : LOW TO HIGH FLOW : 30 L HIGH TO LOW FLOW : 10 LTIMING: OPENING INCREMENT :0.070 S OPENING RELAX. :1.000 S CLOSING INCREMENT :0.070 S CLOSING RELAX. :1.000 S WET PROBE :10 S ANTI-VORTEX STOP : 5 S MANIFOLD FILLING :30 S MANIFOLD DRAINING :30 SSTOP FLOW AT 9.8 M3/H

GRAVICOMPT 348+.000 BOARD v1r8VERSION 02.02.00 OF 30.06.15VEHICULE: AA-000-AAINDICATOR: 03000 PRINTED ON THE 21/07/15 at 10h20

Instantaneous Height

Page 23: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 23/24

This document is available at www.alma-alma.fr

TOTALISER: EVENTS RECORDED: Print the general totaliser Print the events recorded

********** TOTALISER **********

GENERAL TOTALISER: 00056638 L

GRAVICOMPT 348+.000 BOARD v1r8VERSION 02.02.00 OF 30.06.15VEHICULE: AA-000-AAINDICATOR: 03000 PRINTED ON THE 21/07/15 at 10h20

25 RECORD(S)

14:49:55 TEMPERATURE DEFAULT14:49:53 USER MODE14:30:03 SWITCH ON14:24:33 RESET APPLICATION...

09:47:15 METROLOGICAL MODE09:47:06 DATE MODIFICATION09:42:57 PARAM@10= 19509:12:36 PARAM@ 9= 108:59:02 PARAM@26= 13 08:58:57 PARAM@24= 1

GRAVICOMPT 348+.000 BOARD v1r8VERSION 02.02.00 OF 30.06.15VEHICULE: AA-000-AAINDICATOR: 03000 PRINTED ON THE 26/07/15 at 15h50EVENTS OF 21/07/15

DELIVERY ORDER :

********** DELIVERY **********

DELIVERY 001

COMPARTMENT : 1PRODUCT : GOMEASUREMENT 1 : 00400 LITERSMEASUREMENT 2 : 01000 LITERSMEASUREMENT 3 : 01000 LITERS

--------------------TOTAL CPT 1 : 02400 LITERS

IN CASE OF DISPUTE, THE MEASUREMENT RESULTS STORED BY THE MAIN INDICATING DEVICE PROVIDING PROOF

GRAVICOMPT 348+.000 BOARD v1r8VERSION 02.02.00 OF 30.06.15VEHICULE: AA-000-AAINDICATOR: 03000 PRINTED ON THE 22/07/15 at 9h42

Page 24: MU 7038 EN C€¦ · FORM DOC 011 A MU 7038 EN C GRAVICOMPT MANIFOLD Page 1/24 This document is available at OPERATING MANUAL MU 7038 EN C GRAVICOMPT MANIFOLD C 2015/09/08 Volume

FORM DOC 011 A

MU 7038 EN C GRAVICOMPT MANIFOLD Page 24/24

This document is available at www.alma-alma.fr

RELATED DOCUMENTS

GU 7038 User Guide

MV 5006 Verification Guide

FM 8000 Replacement of the backup batteries on the AFSEC and AFSEC+ electronic board

FM 8001 Diagnostic support for power supply failure

FM 8002 Diagnostic support for a display failure

FM 8003 Diagnostic support for DEB_0 or ZERO FLOW DEFAULT alarm

FM 8004 Diagnostic support for GAS or PRESENCE GAS alarm

FM 8005 Diagnostic support for METERING PROBLEM alarm

FM 8007 Diagnostic support for MEMORY LOST or DEF MEMO alarm

FM 8008 Diagnostic support for DATE alarm

FM 8010 Diagnostic support for EEPROM MEMORY LOST alarm

FM 8011 Configuration of jumpers and adjustment of metering thresholds on the AFSEC+ electronic board

FM 8510 Adjustment of a temperature chain in a MICROCOMPT+