MRI(2)

download MRI(2)

of 12

description

MRI medical imaging

Transcript of MRI(2)

  • Fundamentals of Magnetic Resonance

    Imaging

    -- Pulse Sequence & Safety

    Outline

    Image formation

    Basic Pulse Sequence

    Safety

    Image Formation

    Gradients

    A gradient is simply a deliberate change in the magnetic field

    Gradients are used in MRI to linearly modify the magnetic field from one point in space to another

    Gradients are applied along an axis (i.e. Gx

    along the x-axis, Gy

    along the y-axis, Gz

    along the z-axis)

    What happens to the frequency of precession when we turn on a gradient?

    Start, 16/2 8:18pm

  • Effect of Gradient on Rate of Precession Effect of a Gradient

    Spatial Localization

    Gradients, linear change in magnetic field, will provide additional information needed to localize signal

    Makes imaging possible/practical

    Couldnt spatially localize MRI signal instead moved subject to get each voxel

    Nobel prize awarded for this idea!

    Slice Selection

    Use magnetic gradient to modify

    frequency of the protons

    precession

    A slice will be selected with the

    protons precess with the same

    frequency as that of the RF

    pulse.

    Slice location vs slice thickness

    The slice selection gradient is

    always applied perpendicular to

    the slice plane.

    As we need to know where the signal come from

  • Just like a tuning fork!

    Slice Orientation

    Inplane Spatial Localization

    Phase encodingFrequency encoding

    Could be

    the other

    way round!

    An example

    Phase Encoding

    Apply gradient in one direction

    Protons decrease or increase their

    rate of precession

    Turn off the gradient

    All of the protons will again precess

    at the same rate

    Difference is that they will have

    different phase from each other

    1. Initial state

    2. Gradient turned on

    3. Gradient turned off

    Only the same thing will resonance

  • Before slice selection (only spin) At selected slice (precess with Larmor frequency)

    phase encoding gradient is applied phase encoding gradient is turned off

    A 2D Scenario Phase encoding in practice

    F

    r

    e

    q

    u

    e

    n

    c

    y

    e

    n

    c

    o

    d

    i

    n

    g

    Phase encoding

    Phase

    encoding

    gradient of

    signal 128

    Phase

    encoding

    gradient of

    signal 1

    Phase

    encoding

    gradient of

    signal 256

    Frequency Encoding

    Apply gradient in one direction to modify the rate at which the

    protons spin based on location of the proton

    Leave it on

    Result:

    Protons that experience a decrease in the net magnetic field

    precess slower

    Protons that experience an

    increase in the net magnetic

    field precess faster

    = x G

    Frequency offset from the center

    x position relative to center

    G frequency encoding gradientBefore frequency

    encoding

    Frequency encoding

    turned on

    Nobel Prize in Physiology or Medicine

    in 2003 for discoveries concerning magnetic resonance imaging

    Make MRI applicable to imaging human

    Paul C. Lauterbur

    USA

    Sir Peter Mansfield

    UK

    Developed a mathematical

    process to speed the image

    reading, i.e., EPI

    Introduced frequency and phase

    encoding principle in the magnetic field

    for spatial localization in MRI

  • Signal Detection Detecting Net Magnetization using Coil

    C

    Electromagnetic Induction!

    K-Space

    k-space is what is actually measured in MRI (i.e.,

    the signal from M0 is transformed into x and y

    values via k-space)

    kPE

    K-Space and MR Image

    Echo

    k-space lines

    =

    Inverse Fourier

    transform

    Mxy signal does not become 0 instantly, the spins merely diphase

    Can rephase spins to form a symmetrical MR signal -> echo

    MR data (k-space) from scanner is a line by line acquisition of echoes

    Inverse Fourier Transform of the data gives required image

    Rane SD. Texas A&M University, 2005.

  • K-Space and MR Image

    original image full k-space data

    high

    signal

    Imaging and K-Space

    Center of k-space

    Imaging and K-Space

    Everything else

    Imaging and K-Space

    Full Frequency Half Phase

  • Nobel Prize in Chemistry in 1991

    for the development of Fourier

    Transform nuclear magnetic

    resonance spectroscopy and

    the development of multi-

    dimensional NMR techniques

    Richard Robert Ernst

    Switzerland

    Nobel Prize in Chemistry in 2002

    For his development of nuclear magnetic

    resonance spectroscopy for determining

    the three-dimensional structure of

    biological macromolecules in solution"

    Honorary Professor, The Chinese

    University of Hong Kong

    Kurt Wthrich

    Switzerland

    Basic Pulse Sequences

    Timing Diagram for an Imaging Sequence

    as a function of time

    a 90o slice selective pulse

    a slice selection gradient pulse

    a phase encoding gradient pulse

    a frequency encoding gradient pulse

    a signal

    Repeat it!

    TR: Repetition Time (the time between repetitions)

    Phase encoding gradient changes: magnitude varies in equal steps between

    the maximum amplitude of the gradient and the minimum value.

    time

  • The simplest signal form generated in MRI

    The magnetization component has a non-zero component in

    the xy plane

    The precessing magnetisation will induce a corresponding

    oscillating voltage in a detection coil surrounding the sample

    Free Induction Decay (FID)

    Tissue 1 Tissue 2

    MR signal intensity

    very flexible and allow the user to acquire images in

    which either T1 or T2 (dominantly) influences the signal

    intensity displayed in the MR images.

    Spin Echo (SE)

    Gf

    G

    Gs

    TI

    TE

    TRhttp://www.youtube.com/watch?v=dFp2Z3wjrmo&list=PLCD41685D8499AAB1

    http://www.youtube.com/watch?v=vMh11VtUA5o&list=PLCD41685D8499AAB1

    For SE, relatively long imaging times

    But not for GE, due to

    flip angle is typically smaller than 90 (i.e., 20 ~ 60)

    They have no spin-echo because there is no 180 pulse

    Rephasing is done by means of gradient reversal only

    Gradient Echo (GE)

    http://www.youtube.com/watch?v=Rvpa1gqG06g&list=PLAE12114468910462

    The fastest 2D imaging sequence currently available

    Could be a SE or a GE sequence, multiple echoes are generated in one

    excitation

    Acquisition time TA for one image is 100ms and even lower

    Acquisition time TA2D =Nph*TR/ETL (Nph is inplane phase encoding steps)

    ETL is the Echo Train Length (i.e., the number of echoes per excitation)

    Used in functional MRI, diffusion and perfusion imaging

    Echo planar imaging (EPI)

    Gf

    G

    Gs

    Higher induced voltage can give a brighter signal

    TE= only can give you one piece of data

    TR =repetition time for if 3D image is needed

  • Major Pulse Sequence Parameters

    Echo Time (TE) time after 90o RF pulse until readout. Determines

    how much transverse relaxation will occur before reading one row

    of the image.

    Repetition Time (TR) Determines how much longitudinal

    relaxation will occur before constructing the next row of the image.

    In spin echo, TR is the time between two successive 90o RF

    pulses

    In gradient echo, TR is the time between centers of two small

    angle pulse

    Contrast in MRI: T1-Weighting

    500ms 3000ms

    Contrast in MRI: T2-Weighting

    180ms30ms

    50ms

    50ms30ms 180 ms

    Biological Effect of MRI and Safety

  • 1. Static Magnetic Field Bo

    2. Gradient Magnetic Fields

    3. Radiofrequency Electromagnetic Fields

    Safety Concerns Recall: Magnetic Properties of Material

    iron, nickel, or cobalt

    e.g., aneurysm clips, parts of pacemakers, shrapnel, etc.

    large positive magnetic susceptibility

    remain magnetized after an external magnetic field is removed

    oxygen and ions of various metals like Fe, Mg, and Gd

    positive magnetic susceptibility

    Contrast agent

    magnetic susceptibility between ferromagnetic and paramagnetic

    e.g., iron containing contrast agents for bowel, liver, and lymph

    node imaging.

    Contrast agent

    no intrinsic atomic magnetic moment

    small negative magnetic susceptibility

    e.g. water, copper, nitrogen, barium sulfate, and most tissues are

    diamagnetic

    The main magnetic field of a 1.5 T magnet is about 30,000

    times the strength of the earth's magnetic field.

    Ferromagnetic Objects Projectile Effect

    Effect on implants

    pacemakers is disturbed

    intracranial aneurysm clips could be ferromagnetic and experience a

    torque or twisting in a magnetic field

    aneurysm clip might experience a fatal hemorrhage

    heart valves (e.g., Star-Edwards) could be torqued in a magnetic field,

    but not necessarily

    Safety Concern 1 -- Static Magnetic FieldSafety

    The whopping strength of the magnet makes safety

    essential.

    Things fly Even big things!

    Safety Concern 1 -- Static Magnetic Field

  • SafetySafety Concern 1 -- Static Magnetic Field

    Changing magnetic field induce electrical currents in conductors.

    Effect:

    For surgical metal implant, the potential exists for electrical currents being

    induced in the metal with subsequent heating

    Very rapidly changing magnetic fields as may be achieved with EPI

    could cause nerve stimulation, which can affect motor nerves --

    muscle contraction

    Safety Concern 2 Varying Gradient Fields

    Specific Absorption Rate (SAR)

    defined as the power absorbed per mass of tissue

    Unit: W/kg

    the heating is normally insignificant

    But EPI and MRS are possible of over heating tissue

    Solution:

    Shorter ETL

    Longer TR

    Less slices

    Revise the RF sequence to reduce the energy

    Safety Concern 3 -- Radio Frequency

    Proportional to

    the square of Bo

    !

    Burn from looped cables

    Safety

    Anyone going near the magnet subjects, staff and visitors

    must be thoroughly screened:

    Subjects must have no metal in their bodies:

    pacemaker, aneurysm clips, metal implants (e.g.,

    cochlear implants), interuterine devices (IUDs)

    some dental work (fillings okay)

    Subjects must remove metal from their bodies

    jewellery, watch, piercings, coins, wallet, etc.

    any metal that may distort the field (e.g., underwire

    bra)

    Subjects must be given ear plugs (acoustic noise can reach

    120 dB)

    Guidelines to Observe

    Functional MRI not suggest

  • What the radiologist say

    http://www.radiologyinfo.org/en/photocat/ga

    llery3.cfm?image=mr-safety-

    explained.jpg&pg=sfty_mr

    What does radiologist say?Reference

    The Essential Physics of Medical Imaging by Bushberg et al.

    (Basics).

    The Basics of MRI, Joseph P. Hornak, Ph.D. (principle)

    Magnetic Resonance Imaging by Stark and Bradley, second

    edition (Artifacts, Basics, Instrumentation, Pulse Sequences).

    Safety Considerations in MR Imaging by Kanal et al. Radiology

    176:593-606, 1990 (Safety).

    Reference