MRI-PET; A new approach for multi modality imaging system

download MRI-PET;  A new approach for multi modality imaging system

If you can't read please download the document

description

MRI-PET; A new approach for multi modality imaging system. Karl Ziemons Member of Crystal Clear Collaboration. Dr. Karl Ziemons Forschungszentrum Jülich GmbH Central Institute of Electronics Tel.: +49-2461-615685 Fax: +49-2461-613990 E-Mail: [email protected]. Motivations. - PowerPoint PPT Presentation

Transcript of MRI-PET; A new approach for multi modality imaging system

  • MRI-PET; A new approach for multi modality imaging systemKarl Ziemons Member of Crystal Clear CollaborationDr. Karl ZiemonsForschungszentrum Jlich GmbHCentral Institute of Electronics

    Tel.: +49-2461-615685Fax: +49-2461-613990E-Mail: [email protected]

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    MotivationsPET & MRI are medical imaging techniques that in widespread use both for patient diagnosis and management, and in clinical research

    playing a key role in a wide range of fields from mapping of the human brain to the development of new treatments for cancer

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    MotivationsPET in-vivo information about metabolism and functionality

    (f)MRI Anatomical information with a better soft tissue contrast as CT and does not apply additional radiation dose in-vivo information about neuronal activities; the so-called blood oxygenation level dependent (BOLD) contrast has proved to be a very sensitive MRI marker

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    Complementary nature of MRI & PETHence: The Sum of PET and MRI should be excellent and even better MRI + PET minimal scatterGradients need more currentStronger coupling of RF coil

    M.Schwaiger, S.Ziegler, et al., 2005

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design ChallengesLimited space for the PET detectorNeed minimum space for patientPET detector must not use magnetic materialsCould distort MR imagePET detector must not emit in MR frequencyCould produce MR image artifacts MR-compatible PET shielding materialsCould distort MR image

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design ChallengesPET detector must not use magnetic materialsDistortions and interference lines are produced in a uniform phantom imageCourtesy by R.Grazioso, Siemens

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design ChallengesPET detector must not emit in MR frequencyRF artifacts can be seen in the RF noise measurement due to improper shielding of the PET electronics.Courtesy by R.Grazioso, Siemens

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design ChallengesMR-compatible PET shielding materialsBulk leadHot-pressed lead monoxideBGOPure lead powderRed lead oxide powder(Courtesy of D. Struhl, et al., TNS 2003)Spin-echoGradient-echoVarious materials can produce an MR signature even if it is not conductive or magnetic.

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design Challenges

    MR gradient field-eddy currentsCould produce noise in detectorCould heat detectorMR RF transmitCould produce false PET eventsMR materialsWill produce more gamma attenuation

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET/MR Design ChallengesMR Radio frequency

    degrading the energy and timing resolution of the PET system.

    Courtesy by R.Grazioso, Siemens

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    PET Insert for a Clinical MRI ScannerSimultaneous Measurement in 1.5 T

    APD based detectorMulti-slice PET images High quality MR images

    Courtesy by R.Grazioso a. M.Schmand, Siemens

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    BrainPET Project(G)-APD PET insert:trapezoid monolithic block design: LSO block with two (G)-APD in double readout technology and MR compatibleSystem peak sensitivity: > 15%Spatial resolution: < 1.3mm over the whole FOVCurrent CCC projectGoal:

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    MRTPETSimultaneous Imaging of a [F-18]-FDG Mouse HeadStep and shoot PET acquisition (12 angles, each 6 min) while MRT images were taken.Filtered Back Projection (2.5 mm Gaussian image filtering post reconstruction)70/30 Bruker Biospec system. B-GA20 gradient set. Micro Imaging Coil. FLASH MRT sequence. 1mm slice thickness. with Two Coincident APD Based LSO Block-Detectors and a 7Tesla Small Animal MRT SystemCourtesy by B.Pichler, Tbingen

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    New MR-compatible solid-state PET detectorsMagnetically insensitive detectors for MR/PETSilicon photomultipliers (SiPM)an array of geiger-mode APDs showing good timing and energy resolution but not fully developed yet (Eres:12.5%, Tres: 560 psec). (Lorenz et al., Some studies for a development of a small animal PET based on LYSO crystals and geigermode-APDs)Cadmium Zinc Telluride (CZT)has been investigated for a long time but new for PET. Some recent results show good timing (2.6ns vs BaF2). (Verger at al., New trends in gamma-ray imaging with CdZnTe/CdTe at CEA-LETI)Hybrid Photodiode (HPD)sensitive to magnetic fields but can operate normally if positioned parallel to the magnetic field. (Igor Rubashov, Apparatus for combined nuclear imaging and magnetic resonance imaging, and method thereof,)

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    MRI-PET: Which Problems to solve?to measure or calculate attenuation maps for quantitative PET image reconstructionNeed such a solution for Brain and WholeBody Imaging!!to assess motion in the FOVMotion can deteriorate MR and PET images and may cause serious artefacts! This is especially true for images derived from multiple acquisitions!MRI

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    MRI based AC: BrainPET-based Attenuation CorrectionCorresponding MR imagesMRI-based segmentation (whole brain, bone, soft-tissue, sinus arca)

    (IEEE, 2006: E. Rota Kops, et al.)CTMRIDifferentBone signal

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    Problem of Patient Motion (18F-Altanserin)[Herzog et al., JNM, 2005]Motion affectedMotion correctedShould be derived from MRI?!

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    Why MRI-PET Hybrid Imaging?Want true simultaneous data acquisition in a single device Want combined functional and morphological data acquisition at the same timeWant multi modal functional acquisitions at the same time (fMRI / MRS - PET)Want to cross-validate activations measured with PET and fMRI under the same conditions, at the same time, in the same status

    but still want quantitative PET imageMRI

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    ConclusionPET/CT was a medical revolution and a technical evolution

    MR/PET seems to be a technical revolution and a medical evolution

    Frontiers in Imaging Science - Rome 2006

    Forschungszentrum JlichCentral Institute of Electronics

    AcknowledgmentsThanks to Ralph Ladebeck and Ron Grazioso (Siemens) for MR specifications and material

    N.J.Shah and U.Pietrzyk (IME) ,T. Beyer (Essen) And all the others for images and material

    Frontiers in Imaging Science - Rome 2006

    The MR will emit RF at a specified frequency (42.58 MHz/T for hydrogen). The RF can interfere with the detector electronics.