MRI in Patients with Implanted Electronic Medical … Lifelong Learning...The Issue There is an...
Transcript of MRI in Patients with Implanted Electronic Medical … Lifelong Learning...The Issue There is an...
MRI in Patients with
Implanted Electronic
Medical Devices
Dr James FraserDalhousie University
CAR 2016
Objectives
1. Describe the safety concerns and technical challenges
presented by imaging patients with implanted cardiac
electronic medical devices. (CanMEDS role: Health
Advocate)
2. Explain the term MR-conditional device and its
implications.
3. Demonstrate a basic understanding of the CHR/CAR
collaborative Consensus Statement on Magnetic
Resonance Imaging with Cardiac Implantable Electronic
Devices. (CanMEDS roles: Collaborator, Communicator)
Disclosures
No financial interests with any of the companies or
products discussed
The Issue
There is an increasing number of patients with implanted
cardiac devices, many of whom will require an MRI
examination for optimal imaging assessment.
There are serious and potentially fatal risks of performing
MRI in these patients without following appropriate
precautions and protocols.
An increasing number of MR-conditional implanted
cardiac electronic devices (e.g. pacemakers and ICDs)
are becoming available within Canada.
Potential risks of MRI
with conventional pacemakers
Risks associated with MRI due to interaction between Pacemaker system and the static magnetic field, gradient magnetic fields, and radiofrequency energy.
Static magnetic field: movement, torque, dislodgment, magnetic sensor activation and unpredictable reed-switch behavior, magnetohydrodynamic effect.
Gradient magnetic fields: rapidly changing magnetic fields can induce electrical currents in pacemaker leads -> oversensing, undersensing, even life-threatening arrhythmia.
Radiofrequency Energy: Leads can act as “antennae”, concentrate radiofrequency energy, = heat and electrical currents -> reset, battery depletion, effects on sensing, pacingthresholds, lead impedances, causing inappropriate pacingacceleration or inhibition. Abandon or fractured leads more prone to tip heating.
MR Conditional
Labeling for MRI
*ASTM standard F2503: Standard Practice for Marking Medical Devices and Other Items for Safety in the
Magnetic Resonance Environment
*ASTM standard F2503: Standard Practice for Marking Medical Devices and Other Items for Safety in the
Magnetic Resonance Environment
...an item that has been demonstrated to pose
no known hazards in a specified MRI
environment with specified conditions of use.
Minimize ferromagnetic content
Isolated circuit board
Circuit component change
Designed for MRI Environment
Unintended cardiac stimulation
Device interactions
Force and Torque
MR Conditional Pacing Leads
Approved by Health Canada:
Newly developed specific MRI Conditional
Leads
Retroactive MR Conditional Labeling
Health Canada Scanning
ConditionsBiotronik ProMRI
Pre-Scan
Conditions
The following requirements must always be fulfilled in order to perform an MR scan
using BIOTRONIK’s Entovis ProMRI® System:
• The device system consists of a pacemaker with one or two pacing leads in
combination to constitute an MR conditional device system.
• There must be no other implanted medical devices that may interact with MRI, such
as:
-- Abandoned pacemaker/ICD leads
-- Lead extensions
-- Other active medical devices
-- Non-MRI compatible devices
• The absence of phrenic nerve stimulation at 4.8 V at 1.0 ms.
• The leads have been implanted for at least 6 weeks.
• The device system is implanted pectorally.
• The measured pacing threshold is not above 2.0 V at 0.4 ms pulse width.
• The pacing system should be functioning normally prior to an MRI
• The pacemaker is reprogrammed to a special MRI mode immediately prior to the MR
scan.
MRI Scanner
Limitations
The MRI scanner has to meet the following conditions:
• Use of a clinical MRI system with a cylindrical bore or elliptical bore and a static
magnetic field strength of 1.5 Tesla.
• The slew rate of the MRI scanner’s gradient fields should not exceed 200 T/m/s per
axis.
• No additional local transmitting coils are used.
Health Canada Scanning
ConditionsBiotronik ProMRI
Restrictions During
the MRI Scan
The following conditions must be met during the MR scan:
• The mean specific absorption rate (SAR) for the whole body displayed by the MR
scanner must not exceed 2.0 W/kg.
• The head absorption rate displayed by the MRI scanner must not exceed 3.2 W/kg.
• Emergency equipment for resuscitation must be kept at hand and properly certified
staff must be readily available.
• The patient should be continuously monitored in an appropriate manner during the
entire MR scan. Among others, the following parameters can be observed for this
purpose:
-- Blood oxygen saturation
-- Blood pressure
-- ECG
• Adherence to the permissible positioning ranges. The isocenter of the high-frequency
coil should not be below eye level or above the hip bone. In practice, this means that
the marker line of the laser light localizer, which is used for subsequent positioning of
the patient within the MRI scanner, should not be set below eye level (lower edge of eye
socket) and not above the hip bone (two fingers above the symphysis). These areas
have to be adhered to during the MR scan.
Permissible
Positioning Zone
and Scan
Exclusion Zone
The first permissible positioning zone starts at the top of the skull and ends at eye level.
The second permissible positioning zone starts at the hip bone level and ends at the
patient’s feet. The eyes and hip bone serve as the maximum allowed positioning marks
for the isocenter of the MR scanner. These visible marks can be marked with a laser
during MR scanner positioning. Figure 1 below illustrates the permissible positioning
zone and scan exclusion zone of the Entovis MR Conditional system.
Must be programmed into the appropriate
mode by a cardiologist / electrophysiologist or
designate prior to scanning.
Printed Documentation for MRI Record
Confirms pacers required
settings and functionality
Key Conditions: MRI
Patient monitoring
Verbal and Visual
Pulse Ox and/or
EKG
RV
LV
Can artifact
Lead
Signal alteration
Short axis Long axis
T2 DE
MR Conditional Systems
Manuals
Facility requirements for performing MR
imaging in patients with MR-conditional CIEDs
The imaging facility must have a protocol for MR scanning
of patients with CIEDs, developed via a collaborative effort
of MR and CIED specialists. Ideally, it should consist of an
onsite CIED clinic to interrogate and program the CIED
systems. In some cases, it might be possible for a
radiology suite to establish close collaboration with an
offsite CIED clinic where patients are assessed before and
after MR. However, on the day of MR scan, a member of
the CIED team (technician, nurse, or physician) should be
readily accessible for device troubleshooting or
reprogramming, if required.
MRI in patients with conventional
pacemakers
The MagnaSafe Registry 1500 exams, no deaths or poor patient outcomes.
Insufficient data to determine safety, and possible long-term effects are unknown.
Performing MRI in patients with conventional pacemakers cannot be considered a routine procedure. Appropriate multidiscipline risk Vs benefit discussion is paramount.
Special precautions must be taken including: careful patient selection and device programming, rigorous monitoring, reanimation readiness during the procedure, and exclusion of thoracic MRI.
Summary
Multiple MR Conditional Pacemaker systems available for 1.5T MRI systems in Canada.
Very specific Pre-scanning, Scanning and Post-scanning conditions and procedures are necessary.
Present systems require the system to be programmed into MRI mode by Cardiologist/electrophysiologist prior to scanning and reprogrammed after scanning. Collaborative “team approach” protocol should be in place.
Canadian Heart Rhythm Society (CHRS) and CanadianAssociation of Radiologists (CAR) Joint Consensus Statement on Magnetic Resonance Imaging with Cardiac Implantable Electronic Devices.
http://www.carjonline.org/article/S0846-5371%2814%2900074-6/pdf.
Thank You
Questions?