Mr. Rodney Bryant VCP-CA Section, MC-221 Remediation ... · CES APAR Second Zone DUS.Doc . 2...
Transcript of Mr. Rodney Bryant VCP-CA Section, MC-221 Remediation ... · CES APAR Second Zone DUS.Doc . 2...
11231 Richmond Avenue
Suite D104 Houston, TX 77082 Tel (832) 916-3690
October 16, 2017
Project No. 4020
Mr. Rodney Bryant
VCP-CA Section, MC-221
Remediation Division
Texas Commission on Environmental Quality
P. O. Box 13087
Austin, Texas 78711-3087
SUBJECT: VOLUNTARY CLEANUP PROGRAM (VCP)
CES GRIGGS ROAD VCP SITE
4900 GRIGGS ROAD, 4904 GRIGGS ROAD, AND 5910 WAYLAND STREET
(VCP No. 2751; CN604824904; RN108370644)
RESPONSE TO JUNE 16, 2017 COMMENTS ON REVISED APAR – COMMENT 4
ADDITIONAL GROUNDWATER ASSESSMENT
Dear Mr. Bryant:
Pastor, Behling & Wheeler, LLC (PBW), on behalf of the CES Griggs Road PRP Group (the “Group”), has
prepared this response to Comment 4 on the revised Affected Property Assessment Report (APAR). The
APAR summarized the environmental assessment of the CES Griggs Road VCP Site (the “Site”) pursuant to
the requirements of VCP Agreement No. 2751. The VCP Agreement was executed between the Group and
the Texas Commission on Environmental Quality (TCEQ) on June 10, 2015.
The revised APAR was submitted to TCEQ on April 17, 2017. Comments were provided to the Group by the
TCEQ on June 16, 2017. As quoted below, TCEQ requested an additional assessment of the second sand
(approximately 40-60 feet below ground surface (bgs)) in Comment 4.
“In addition, the TCEQ understands that cone penetrometer tests (CPT) were conducted to identify
the presence of a laterally continuous clay layer that could possibly separate the upper groundwater
bearing unit (GWBU) from a second sand unit. The TCEQ notes that the data located in the boring
logs for the CPT test do no indicate a continuous clay layer. The 20’ [foot] thick clay layer
identified contains numerous intermittent sands and clayey sands which may allow for vertical
migration of contamination and connection of the first and second sands. Additional assessment of
this second sand (approximately 40-60’ bgs) should be conducted to determine whether the
contamination identified in the shallow sand is also present in the second sand.”
In response to TCEQ’s comment, the Group conducted CPT sampling with a piezocone on October 10, 2017
to assess if chlorinated volatile organic compounds (cVOCs) detected in the on-Site upper GWBU had
migrated vertically into the second GWBU. The CPT location, CPT-7-2, was advanced in the presumed
source area between MW-1 and MW-5. A groundwater sample was collected after the second zone was
identified by CPT. Figure 1 presents the sample location.
The first GWBU was encountered at approximately 13 feet bgs in CPT-7-2. The top of the confining clay
underlying the first GWBU was encountered approximately 27 feet bgs with a thickness of approximately 16
Mr. Rodney Bryant
October 16, 2017
Page 2 of 2
feet. These depths are consistent with the shallow GWBU depths observed from previous investigations. The
second GWBU was encountered at approximately 43 feet bgs. This depth is consistent with the previous CPT
investigation. The CPT was advanced to approximately 55 feet bgs to ensure the groundwater sample was
collected within the second sand zone (see CPT log in the Attachments). A groundwater sample and a
duplicate were collected (see the Photolog in the Attachments). The following table provides the results:
CPT-7-2 Chemical of
Concern
cis-1,2-
Dichloroethene Tetrachloroethene
trans-1,2-
dichloroethene Trichloroethene
Vinyl
chloride Concentration
(mg/L) < 0.00020 < 0.00030 < 0.00020 < 0.00020 < 0.00020
Duplicate
concentration
(mg/L) < 0.00020 < 0.00030 < 0.00020 < 0.00020 < 0.00020
As shown in this table, all targeted cVOCs were below the limits of detection. The laboratory report is
provided in the Attachments.
Based on these results, the Group requests agency concurrence that the data presented in this letter report
demonstrate that additional assessment of the second sand zone at the CES Site has been completed.
With this letter report, coupled with the previous Group submittals dated September 21 and October 11, 2017,
the Group has submitted complete responses to all agency comments as provided by the TCEQ comment
letter dated June 16, 2017. Consequently, the Group requests agency approval of the revised APAR in order
to proceed with preparation of the Response Action Plan.
If you have any questions or comments, please call (832) 916-3691 or email [email protected].
Sincerely,
PASTOR, BEHLING & WHEELER, LLC
Brenda P. Basile, Ph.D.
Senior Consulting Scientist
cc: CES Griggs Road PRP Group
Attachments: Figure 1 On-Site Property Map
CPT Log
Photolog
Laboratory Report
Mr. Rodney Bryant October 16, 2017
FIGURE
01
11
12
14
13
10
17
18
16
09
19
20
21
23
08
07
01
22
15
02
03
01
03
04
02
06
05
06
03
02
24
25
27
28
29
30
29A
14
18
13
12
10
17
11
03A
16
15
07
04A
09
08
MW-02
MW-04
MW-03
04
05
26
MW-01
MW-05
31
32
MW-08
MW-06
MW-07
CPT-05
CPT-06
CPT-04
CPT-01
CPT-03
CPT-02
HA-02
HA-01
HA-04
HA-05
HA-03
1
2
3
4
5
6
7
8
9
10
11
12
14
13
4900 Griggs Rd.
4904 Griggs Rd.
5910 Wayland St.
120600
Scale in Feet
DATE:
PROJECT:
PASTOR, BEHLING & WHEELER, LLC
CONSULTING ENGINEERS AND SCIENTISTS
REVISIONS
CHECKED:
BY:
Figure 1
AJD
BBJULY, 2017
4020
Source:
Imagery taken from Google Earth, photography dated July 31, 2015.
Approx. Property Boundary
PBW Sample Location
PBW Sample Location (Jan. 2017)
PBW Monitoring Well Location
PBW Monitoring Well Location (Jan. 2017)
CPT Location
Hand Auger Location
CES GRIGGS ROAD VCP SITE
HOUSTON, TEXAS
Office Building
Training Center/Locker Room (2008)/
Administrative (2010)
Industrial Waste Processing & Storage (2008)/
Main Processing (2010)
Old Cleaning Rack (1996)/
Tank Trucking Cleaning (2008)/Safe Wash (2010)
Equipment Cleaning (2008)/
Drum Cleaning (2010)
New Shop (1996)/Truck Maintenance (2008)
Cleaning Rack (1996)/Tank Truck Cleaning (2008)/
Equipment & Fleet Maintenance (2015)
Oil/Water Separating Tanks (2010)/
North Tank Farm (2015)
Oil/Water Separating Tanks (2010)/
South Tank Farm (2015)
Welding Shop
Fuel Storage Shed
WMU 126-128 (Frac Tanks)
Former Mechanic Shop (2010)/Field Services (2015)
1
2
3
4
5
6
7
8
9
11
12
13
14
Hot Oil Heat System (2008)/
Drum-Tote Storage (2015)
10
Approximate Location Second Zone Groundwater Sample
Mr. Rodney Bryant October 16, 2017
CPT LOG
0
10
20
30
40
50
60
Job Number: Operator: Location:
CPT Number:Date:Elevation:
Coordinates:Cone Number:
04.19170065Bobby BrandtHouston, TX
CPT-7-210-Oct-20170.00
29.69850 -95.34361CP15-CF75PB7SN2-P1E1 1629
Page 1 of 1
Sleeve Friction Fs (TSF) Cone Resistance qc (TSF) Pore Pressure U2 (TSF) Friction Ratio Rf (%)Soil Behavior
Type
(1) sensitive fine grained (OL-CH)
(2) organic material (OL-OH)
(3) clay (CH)
(4) silty clay to clay (CL-CH)
(5) clayey silt to silty clay (MH-CL)
(6) sandy silt to clayey silt (ML-MH)
(7) silty sand to sandy silt (SM-ML)
(8) sand to silty sand (SM-SP)
(9) sand (SW-SP)
(10) gravel to gravelly sand (SW-GW)
(11) very stiff fine grained* (CH-CL)
(12) sand to clayey sand* (SC-SM)
Robertson et al. 1986 *Overconsolidated or Cemented
0 4 0 300 -3 0 12 0 10 0 12
Mr. Rodney Bryant October 16, 2017
PHOTOLOG
PHOTOGRAPHIC LOG
Client Name: CES Griggs Road PRP Group
Site Location: CES Environmental Services, 4904 Griggs Road, Houston, Texas
Project No. 4020
Photo No.
1 Date:
10/09/17 Looking northwest at the location of CPT-7-2. The red arrow points to the location of the second zone groundwater sampling point. The orange cone is sitting on top of MW-5.
Photo No.
2 Date:
10/10/17
View of the screen used during groundwater sampling at CPT-7-2.
Mr. Rodney Bryant October 16, 2017
LABORATORY REPORT
Data Usability Summary
To: Tony Saturni, CES PRP Group Date: October 13, 2017
From: Brenda Basile, Ph.D. File: CES APAR SECOND ZONE DUS.doc
RE: Second Zone Groundwater for 4904 Griggs CC:
Pastor, Behling & Wheeler, LLC (PBW) reviewed one laboratory report for the analyses of groundwater samples collected by PBW (Houston, Texas) at the CES Griggs Road VCP Site on October 10, 2017. The report was reviewed for conformance to the requirements of SW-846 Test Methods for Evaluating Solid Waste Physical/Chemical Methods (SW-846) and the Texas Commission on Environmental Quality (TCEQ) guidance document Review and Reporting of COC Concentration Data (RG-366/TRRP-13). The purpose of this sample event was to provide second zone groundwater data for the Affected Property Assessment Report (APAR).
Samples were analyzed using SW-846 8260B - Volatile Organic Compounds by Gas Chromatography-Mass Spectrometry (GC/MS). Quality control (QC) data were reviewed as described in RG-366/TRRP-13 and the results of the review are discussed in this data usability summary (DUS).
Introduction
The August 2017 samples were analyzed for chlorinated VOCs. Table 1 lists the sample identifications cross-referenced to laboratory identifications and the analyses performed for each sample. No data were qualified due to exceedances of QC criteria.
QC Results
ACCREDITATION
ALS Environmental is accredited under Texas certificate T104704231-17-19 for the matrices, methods and analytes discussed in this DUS.
PRESERVATION AND HOLDING TIMES
Samples were received at the laboratory at temperatures less than 6C. Samples were prepared and
analyzed within method holding times.
BLANKS
No analytes were detected in field or laboratory quality control.
SURROGATE RECOVERIES
Surrogate recoveries were within the RG-366/TRRP-13 acceptance criteria of 60-140 percent recovery (%R) for VOCs.
INTERNAL STANDARD RECOVERIES
According to the LRC, internal standard recoveries were within the method acceptance criteria.
LABORATORY CONTROL SAMPLES
Laboratory control sample (LCS) recoveries (%R) were within the acceptance criteria of 60-140% criteria for VOCs.
CES APAR Second Zone DUS.Doc 2
October 13, 2017
MATRIX SPIKE/MATRIX SPIKE DUPLICATE
Batch or non-project sample data were not evaluated. No project samples were spiked.
FIELD PRECISION
Analytes were not detected in the field duplicate samples so precision was not calculated. Field duplicate data are acceptable.
SUMMARY
Analytical data are usable for determining concentrations in second zone groundwater samples collected at the CES Griggs Road VCP Site.
Table 1 Cross-Reference Field Sample Identifications and Laboratory Identifications
Field Identification Laboratory Identification VOCs Comment
Trip Blank 171010 HS17100583-01 X Trip Blank
CPT-7-2 171010 HS17100583-02 X
CPT-7-2 171010 Dup HS17100583-03 X Field duplicate of CPT-7-2 171010
October 13, 2017
Brenda BasilePastor, Behling & Wheeler, LLC11231 Richmond AvenueSuite D104Houston, TX 77082
The analytical data provided relates directly to the samples received by ALS Environmental and for only the analyses requested. Results are expressed as "as received" unless otherwise noted.
QC sample results for this data met EPA or laboratory specifications except as noted in the Case Narrative or as noted with qualifiers in the QC batch information. Should this laboratory report need to be reproduced, it should be reproduced in full unless written approval has been obtained by ALS Environmental. Samples will be disposed in 30 days unless storage arrangements are made.
If you have any questions regarding this report, please feel free to call me.
Sincerely,
ALS Environmental received 3 sample(s) on Oct 10, 2017 for the analysis presented in the following report.
Laboratory Results for: CES Griggs Rd Quarterly GW
Dear Brenda,
Work Order: HS17100583
Generated By: Dayna.Fisher
Dane J. Wacasey
10450 Stancliff Rd. Suite 210 Houston, TX 77099 T: +1 281 530 5656 F: +1 281 530 5887
www.alsglobal.comRight Solutions • Right Partner
Page 1 of 18
Client:TRRP Laboratory Data Package Cover PageProject:
WorkOrder:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWHS17100583
This data package consists of all or some of the following as applicable:
This signature page, the laboratory review checklist, and the following reportable data:
R1 Field chain-of-custody documentation;
R2 Sample identification cross-reference;
R3 Test reports (analytical data sheets) for each environmental sample that includes:a) Items consistent with NELAC Chapter 5, b) dilution factors, c) preparation methods, d) cleanup methods, and e) if required for the project, tentatively identified compounds (TICs).
R4 Surrogate recovery data including:a) Calculated recovery (%R), and b) The laboratory’s surrogate QC limits.
R5 Test reports/summary forms for blank samples;
R6 Test reports/summary forms for laboratory control samples (LCSs) including:a) LCS spiking amounts, b) Calculated %R for each analyte, and c)The laboratory’s LCS QC limits.
R7 Test reports for project matrix spike/matrix spike duplicates (MS/MSDs) including:a) Samples associated with the MS/MSD clearly identified, b) MS/MSD spiking amounts, c) Concentration of each MS/MSD analyte measured in the parent and spiked samples, d) Calculated %Rs and relative percent differences (RPDs), and e) The laboratory’s MS/MSD QC limits.
R8 Laboratory analytical duplicate (if applicable) recovery and precision:a) the amount of analyte measured in the duplicate, b) the calculated RPD, and c) the laboratory’s QC limits for analytical duplicates.
R9 List of method quantitation limits (MQLs) and detectability check sample results for eachanalyte for each method and matrix.
R10 Other problems or anomalies. The Exception Report for each “No” or “Not Reviewed (NR)” item in Laboratory Review Checklist andfor each analyte, matrix, and method for which the laboratory does not hold NELAC accreditation underthe Texas Laboratory Accreditation Program.
ALS Group USA, Corp Date: 13-Oct-17
Page 2 of 18
Client:TRRP Laboratory Data Package Cover PageProject:
WorkOrder:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWHS17100583
Dane J. Wacasey
Release Statement: I am responsible for the release of this laboratory data package. This laboratory isNELAC accredited under the Texas Laboratory Accreditation Program for all the methods, analytes andmatrices reported in this data package except as noted in the Exception Reports. The data have beenreviewed and are technically compliant with the requirements of the methods used, except where noted bythe laboratory in the attached exception reports. By my signature below, I affirm to the best of myknowledge, all problems/anomalies, observed by the laboratory have been identified by the laboratory inthe Laboratory Review Checklist, and no information affecting the quality of the data has been knowinglywithheld.
Check, if applicable: [NA] This laboratory meets an exception under 30 TAC §25.6 and was last inspectedby [ ] TCEQ or [ ] ______________ on (enter date of last inspection). Any findings affecting the data inthis laboratory data package are noted in the Exception Reports herein. The official signing the cover pageof the report in which these data are used is responsible for releasing this data package and is by signatureaffirming the above release statement is true.
ALS Group USA, Corp Date: 13-Oct-17
Page 3 of 18
Laboratory Review Checklist: Reportable Data
Laboratory Name: ALS Laboratory Group LRC Date: 10/13/2017
Project Name: CES Griggs Rd Quarterly GW Laboratory Job Number: HS17100583
Reviewer Name: Dane Wacasey Prep Batch Number(s): R303310
#1 A2 Description Yes No NA3 NR4 ER#5
R1 OI Chain-of-custody (C-O-C)
Did samples meet the laboratory’s standard conditions of sample acceptability
upon receipt? X
Were all departures from standard conditions described in an exception report? X
R2 OI Sample and quality control (QC) identification
Are all field sample ID numbers cross-referenced to the laboratory ID numbers? X
Are all laboratory ID numbers cross-referenced to the corresponding QC data? X
R3 OI Test reports
Were all samples prepared and analyzed within holding times? X
Other than those results < MQL, were all other raw values bracketed by
calibration standards? X
Were calculations checked by a peer or supervisor? X
Were all analyte identifications checked by a peer or supervisor? X
Were sample detection limits reported for all analytes not detected? X
Were all results for soil and sediment samples reported on a dry weight basis? X
Were % moisture (or solids) reported for all soil and sediment samples? X
Were bulk soils/solids samples for volatile analysis extracted with methanol per
SW-846 Method 5035? X
If required for the project, TICs reported? X
R4 O Surrogate recovery data
Were surrogates added prior to extraction? X
Were surrogate percent recoveries in all samples within the laboratory QC
limits? X
R5 OI Test reports/summary forms for blank samples
Were appropriate type(s) of blanks analyzed? X
Were blanks analyzed at the appropriate frequency? X
Were method blanks taken through the entire analytical process, including
preparation and, if applicable, cleanup procedures? X
Were blank concentrations < MQL? X
R6 OI Laboratory control samples (LCS):
Were all COCs included in the LCS? X
Was each LCS taken through the entire analytical procedure, including prep and
cleanup steps? X
Were LCSs analyzed at the required frequency? X
Were LCS (and LCSD, if applicable) %Rs within the laboratory QC limits? X
Does the detectability data document the laboratory’s capability to detect the
COCs at the MDL used to calculate the SDLs? X
Was the LCSD RPD within QC limits? X
R7 OI Matrix spike (MS) and matrix spike duplicate (MSD) data
Were the project/method specified analytes included in the MS and MSD? X
Were MS/MSD analyzed at the appropriate frequency? X
Were MS (and MSD, if applicable) %Rs within the laboratory QC limits? X
Were MS/MSD RPDs within laboratory QC limits? X
R8 OI Analytical duplicate data
Were appropriate analytical duplicates analyzed for each matrix? X
Were analytical duplicates analyzed at the appropriate frequency? X
Were RPDs or relative standard deviations within the laboratory QC limits? X
R9 OI Method quantitation limits (MQLs):
Are the MQLs for each method analyte included in the laboratory data package? X
Do the MQLs correspond to the concentration of the lowest non-zero calibration
standard? X
Are unadjusted MQLs and DCSs included in the laboratory data package? X
R10 OI Other problems/anomalies
Are all known problems/anomalies/special conditions noted in this LRC and
ER? X
Were all necessary corrective actions performed for the reported data? X
Was applicable and available technology used to lower the SDL and minimize
the matrix interference affects on the sample results? X
Is the laboratory NELAC-accredited under the Texas Laboratory Program for
the analytes, matrices and methods associated with this laboratory data package? X
Page 4 of 18
Laboratory Review Checklist: Supporting Data
Laboratory Name: ALS Laboratory Group LRC Date: 10/13/2017
Project Name: CES Griggs Rd Quarterly GW Laboratory Job Number: HS17100583
Reviewer Name: Dane Wacasey Prep Batch Number(s): R303310 #1 A2 Description Yes No NA3 NR4 ER#5
S1 OI Initial calibration (ICAL)
Were response factors and/or relative response factors for each analyte within QC
limits? X
Were percent RSDs or correlation coefficient criteria met? X
Was the number of standards recommended in the method used for all analytes? X
Were all points generated between the lowest and highest standard used to
calculate the curve? X
Are ICAL data available for all instruments used? X
Has the initial calibration curve been verified using an appropriate second source
standard? X
S2 OI Initial and continuing calibration verification (ICCV and CCV) and
continuing calibration blank (CCB)
Was the CCV analyzed at the method-required frequency? X
Were percent differences for each analyte within the method-required QC limits? X
Was the ICAL curve verified for each analyte? X
Was the absolute value of the analyte concentration in the inorganic CCB < MDL? X
S3 O Mass spectral tuning:
Was the appropriate compound for the method used for tuning? X
Were ion abundance data within the method-required QC limits? X
S4 O Internal standards (IS):
Were IS area counts and retention times within the method-required QC limits? X
S5 OI
Raw data (NELAC section 1 appendix A glossary, and section 5.12 or ISO/IEC
17025 section
Were the raw data (for example, chromatograms, spectral data) reviewed by an
analyst? X
Were data associated with manual integrations flagged on the raw data? X
S6 O Dual column confirmation
Did dual column confirmation results meet the method-required QC? X
S7 O Tentatively identified compounds (TICs):
If TICs were requested, were the mass spectra and TIC data subject to appropriate
checks? X
S8 I Interference Check Sample (ICS) results:
Were percent recoveries within method QC limits? X
S9 I Serial dilutions, post digestion spikes, and method of standard additions
Were percent differences, recoveries, and the linearity within the QC limits
specified in the method? X
S10 OI Method detection limit (MDL) studies
Was a MDL study performed for each reported analyte? X
Is the MDL either adjusted or supported by the analysis of DCSs? X
S11 OI Proficiency test reports:
Was the laboratory's performance acceptable on the applicable proficiency tests or
evaluation studies? X
S12 OI Standards documentation
Are all standards used in the analyses NIST-traceable or obtained from other
appropriate sources? X
S13 OI Compound/analyte identification procedures
Are the procedures for compound/analyte identification documented? X
S14 OI Demonstration of analyst competency (DOC)
Was DOC conducted consistent with NELAC Chapter 5C or ISO/IEC 4? X
Is documentation of the analyst’s competency up-to-date and on file? X
S15 OI
Verification/validation documentation for methods (NELAC Chap 5 or
ISO/IEC 17025 Section 5)
Are all the methods used to generate the data documented, verified, and validated,
where applicable? X
S16 OI Laboratory standard operating procedures (SOPs):
Are laboratory SOPs current and on file for each method performed? X Items identified by the letter “R” must be included in the laboratory data package submitted in the TRRP-required report(s). Items identified by the letter “S” should be retained and made available upon request for the appropriate retention period. O = Organic Analyses; I = Inorganic Analyses (and general chemistry, when applicable); NA = Not Applicable; NR = Not Reviewed; R# = Exception Report identification number (an Exception Report should be completed for an item if “NR” or “No” is checked).
Page 5 of 18
Laboratory Review Checklist: Exception Reports
Laboratory Name: ALS Laboratory Group LRC Date: 10/13/2017
Project Name: CES Griggs Rd Quarterly GW Laboratory Job Number: HS17100583
Reviewer Name: Dane Wacasey Prep Batch Number(s): R303310
ER#5 Description
No Exceptions
Items identified by the letter “R” must be included in the laboratory data package submitted in the TRRP-required report(s). Items identified by the letter “S” should be retained and made available upon request for the appropriate retention period. O = Organic Analyses; I = Inorganic Analyses (and general chemistry, when applicable); NA = Not Applicable; NR = Not Reviewed; R# = Exception Report identification number (an Exception Report should be completed for an item if “NR” or “No” is checked).
Page 6 of 18
Client: Pastor, Behling & Wheeler, LLC
Work Order: HS17100583Project: CES Griggs Rd Quarterly GW SAMPLE SUMMARY
Lab Samp ID Client Sample ID Collection DateMatrix TagNo Date Received Hold
HS17100583-01 10-Oct-2017 10:45VBLKW-100617-01
10-Oct-2017 14:10Trip Blank 171010 Water
HS17100583-02 10-Oct-2017 10:00 10-Oct-2017 14:10CPT-7-2 171010 Groundwater
HS17100583-03 10-Oct-2017 10:05 10-Oct-2017 14:10CPT-7-2 171010 DUP Groundwater
ALS Group USA, Corp 13-Oct-17Date:
Page 7 of 18
Client:Project:Sample ID:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWTrip Blank 171010
WorkOrder:Lab ID:
Collection Date:
HS17100583HS17100583-01
10-Oct-2017 10:45 Matrix:Water
ANALYTICAL REPORT
ANALYSES RESULT MQLDILUTION FACTORUNITS
DATE ANALYZEDQUAL SDL
LOW LEVEL VOLATILES BY SW8260C Method:SW8260 Analyst: AKP1mg/L 12-Oct-2017 19:460.00020cis-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 19:460.00030Tetrachloroethene 0.0010U
1mg/L 12-Oct-2017 19:460.00020trans-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 19:460.00020Trichloroethene 0.0010U
1mg/L 12-Oct-2017 19:460.00020Vinyl chloride 0.0010U
Surr: 1,2-Dichloroethane-d4 1%REC 12-Oct-2017 19:46114 70-126
Surr: 4-Bromofluorobenzene 1%REC 12-Oct-2017 19:46100 81-113
Surr: Dibromofluoromethane 1%REC 12-Oct-2017 19:46103 77-123
Surr: Toluene-d8 1%REC 12-Oct-2017 19:46106 82-127
13-Oct-17Date: ALS Group USA, Corp
Note: See Qualifiers Page for a list of qualifiers and their explanation.
Page 8 of 18
Client:Project:Sample ID:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWCPT-7-2 171010
WorkOrder:Lab ID:
Collection Date:
HS17100583HS17100583-02
10-Oct-2017 10:00 Matrix:Groundwater
ANALYTICAL REPORT
ANALYSES RESULT MQLDILUTION FACTORUNITS
DATE ANALYZEDQUAL SDL
LOW LEVEL VOLATILES BY SW8260C Method:SW8260 Analyst: AKP1mg/L 12-Oct-2017 20:110.00020cis-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 20:110.00030Tetrachloroethene 0.0010U
1mg/L 12-Oct-2017 20:110.00020trans-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 20:110.00020Trichloroethene 0.0010U
1mg/L 12-Oct-2017 20:110.00020Vinyl chloride 0.0010U
Surr: 1,2-Dichloroethane-d4 1%REC 12-Oct-2017 20:11115 70-126
Surr: 4-Bromofluorobenzene 1%REC 12-Oct-2017 20:1198.8 81-113
Surr: Dibromofluoromethane 1%REC 12-Oct-2017 20:11106 77-123
Surr: Toluene-d8 1%REC 12-Oct-2017 20:11107 82-127
13-Oct-17Date: ALS Group USA, Corp
Note: See Qualifiers Page for a list of qualifiers and their explanation.
Page 9 of 18
Client:Project:Sample ID:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWCPT-7-2 171010 DUP
WorkOrder:Lab ID:
Collection Date:
HS17100583HS17100583-03
10-Oct-2017 10:05 Matrix:Groundwater
ANALYTICAL REPORT
ANALYSES RESULT MQLDILUTION FACTORUNITS
DATE ANALYZEDQUAL SDL
LOW LEVEL VOLATILES BY SW8260C Method:SW8260 Analyst: AKP1mg/L 12-Oct-2017 20:370.00020cis-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 20:370.00030Tetrachloroethene 0.0010U
1mg/L 12-Oct-2017 20:370.00020trans-1,2-Dichloroethene 0.0010U
1mg/L 12-Oct-2017 20:370.00020Trichloroethene 0.0010U
1mg/L 12-Oct-2017 20:370.00020Vinyl chloride 0.0010U
Surr: 1,2-Dichloroethane-d4 1%REC 12-Oct-2017 20:37118 70-126
Surr: 4-Bromofluorobenzene 1%REC 12-Oct-2017 20:3797.2 81-113
Surr: Dibromofluoromethane 1%REC 12-Oct-2017 20:37104 77-123
Surr: Toluene-d8 1%REC 12-Oct-2017 20:37104 82-127
13-Oct-17Date: ALS Group USA, Corp
Note: See Qualifiers Page for a list of qualifiers and their explanation.
Page 10 of 18
Client:CES Griggs Rd Quarterly GWPastor, Behling & Wheeler, LLC
WorkOrder:Project:
HS17100583DATES REPORT
Collection Date Prep Date Analysis DateClient Samp IDSample ID TCLP Date DF
Batch ID R303310 Test Name : LOW LEVEL VOLATILES BY SW8260C Matrix: Groundwater
12 Oct 2017 20:11HS17100583-02 10 Oct 2017 10:00 1CPT-7-2 171010
12 Oct 2017 20:37HS17100583-03 10 Oct 2017 10:05 1CPT-7-2 171010 DUP
Batch ID R303310 Test Name : LOW LEVEL VOLATILES BY SW8260C Matrix: Water
12 Oct 2017 19:46HS17100583-01 10 Oct 2017 10:45 1Trip Blank 171010
13-Oct-17Date: ALS Group USA, Corp
Page 11 of 18
ALS Group USA, Corp Date: 13-Oct-17
WorkOrder: HS17100583
Test Code: 8260_LL_WInstrumentID: VOA4
METHOD DETECTION / REPORTING LIMITS
AqueousMatrix:Test Number: SW8260Test Name: Low Level Volatiles by SW8260C
Units: mg/L
Type Analyte DCSCAS MDL PQLDCS SpikeA 0.00078156-59-2 0.00020cis-1,2-Dichloroethene 0.00100.00050
A 0.00064127-18-4 0.00030Tetrachloroethene 0.00100.00050
A 0.00078156-60-5 0.00020trans-1,2-Dichloroethene 0.00100.00050
A 0.0008379-01-6 0.00020Trichloroethene 0.00100.00050
A 0.0007475-01-4 0.00020Vinyl chloride 0.00100.00050
S 017060-07-0 01,2-Dichloroethane-d4 0.00100
S 0460-00-4 04-Bromofluorobenzene 0.00100
S 01868-53-7 0Dibromofluoromethane 0.00100
S 02037-26-5 0Toluene-d8 0.00100
Page 12 of 18
Client:Project:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GW
WorkOrder: HS17100583
QC BATCH REPORT
Batch ID: R303310 Instrument: VOA4 Method: SW8260
Sample ID: VBLKW-171012 Units: ug/L Analysis Date: 12-Oct-2017 11:24
Run ID: VOA4_303310 SeqNo: 4261531 PrepDate: DF: 1
Analyte SPK ValMQLResultSPK Ref
Value %RECControl
LimitRPD Ref
Value %RPDRPD Limit Qual
MBLK
Client ID:
cis-1,2-Dichloroethene U 1.0
Tetrachloroethene U 1.0
trans-1,2-Dichloroethene U 1.0
Trichloroethene U 1.0
Vinyl chloride U 1.0
54.35 50 0 109 70 - 1231.0Surr: 1,2-Dichloroethane-d4
49.1 50 0 98.2 82 - 1151.0Surr: 4-Bromofluorobenzene
51.8 50 0 104 73 - 1261.0Surr: Dibromofluoromethane
53.26 50 0 107 81 - 1201.0Surr: Toluene-d8
Sample ID: VLCSW-171012 Units: ug/L Analysis Date: 12-Oct-2017 10:34
Run ID: VOA4_303310 SeqNo: 4261529 PrepDate: DF: 1
Analyte SPK ValMQLResultSPK Ref
Value %RECControl
LimitRPD Ref
Value %RPDRPD Limit Qual
LCS
Client ID:
cis-1,2-Dichloroethene 53.8 50 0 108 75 - 1221.0
Tetrachloroethene 51.83 50 0 104 76 - 1191.0
trans-1,2-Dichloroethene 48.62 50 0 97.2 72 - 1271.0
Trichloroethene 50.42 50 0 101 77 - 1211.0
Vinyl chloride 53.96 50 0 108 70 - 1301.0
47.59 50 0 95.2 70 - 1301.0Surr: 1,2-Dichloroethane-d4
51.48 50 0 103 82 - 1151.0Surr: 4-Bromofluorobenzene
50.22 50 0 100 73 - 1261.0Surr: Dibromofluoromethane
53.76 50 0 108 81 - 1201.0Surr: Toluene-d8
ALS Group USA, Corp Date: 13-Oct-17
Note: See Qualifiers Page for a list of qualifiers and their explanation.
Page 13 of 18
Client:Project:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GW
WorkOrder: HS17100583
QC BATCH REPORT
Batch ID: R303310 Instrument: VOA4 Method: SW8260
Sample ID: HS17100397-07MS Units: ug/L Analysis Date: 12-Oct-2017 17:37
Run ID: VOA4_303310 SeqNo: 4261541 PrepDate: DF: 1
Analyte SPK ValMQLResultSPK Ref
Value %RECControl
LimitRPD Ref
Value %RPDRPD Limit Qual
MS
Client ID:
cis-1,2-Dichloroethene 47.21 50 0 94.4 70 - 1281.0
Tetrachloroethene 46.15 50 0 92.3 70 - 1301.0
trans-1,2-Dichloroethene 48.79 50 0 97.6 70 - 1301.0
Trichloroethene 45.3 50 0 90.6 70 - 1291.0
Vinyl chloride 44.71 50 0 89.4 70 - 1301.0
54.06 50 0 108 70 - 1261.0Surr: 1,2-Dichloroethane-d4
51.52 50 0 103 81 - 1131.0Surr: 4-Bromofluorobenzene
50.13 50 0 100 77 - 1231.0Surr: Dibromofluoromethane
52.85 50 0 106 82 - 1271.0Surr: Toluene-d8
Sample ID: HS17100397-07MSD Units: ug/L Analysis Date: 12-Oct-2017 18:03
Run ID: VOA4_303310 SeqNo: 4261542 PrepDate: DF: 1
Analyte SPK ValMQLResultSPK Ref
Value %RECControl
LimitRPD Ref
Value %RPDRPD Limit Qual
MSD
Client ID:
cis-1,2-Dichloroethene 47.57 50 0 95.1 70 - 128 47.21 0.749 201.0
Tetrachloroethene 51.78 50 0 104 70 - 130 46.15 11.5 201.0
trans-1,2-Dichloroethene 45.52 50 0 91.0 70 - 130 48.79 6.95 201.0
Trichloroethene 43.65 50 0 87.3 70 - 129 45.3 3.72 201.0
Vinyl chloride 43.63 50 0 87.3 70 - 130 44.71 2.44 201.0
50.62 50 0 101 70 - 126 54.06 6.56 201.0Surr: 1,2-Dichloroethane-d4
53.49 50 0 107 81 - 113 51.52 3.76 201.0Surr: 4-Bromofluorobenzene
49.19 50 0 98.4 77 - 123 50.13 1.89 201.0Surr: Dibromofluoromethane
53.93 50 0 108 82 - 127 52.85 2.02 201.0Surr: Toluene-d8
The following samples were analyzed in this batch: HS17100583-01 HS17100583-02 HS17100583-03
ALS Group USA, Corp Date: 13-Oct-17
Note: See Qualifiers Page for a list of qualifiers and their explanation.
Page 14 of 18
QUALIFIERS, ACRONYMS, UNITS
Client:Project:WorkOrder:
Pastor, Behling & Wheeler, LLCCES Griggs Rd Quarterly GWHS17100583
Qualifier Description* Value exceeds Regulatory Limit
a Not accredited
B Analyte detected in the associated Method Blank above the Reporting Limit
E Value above quantitation range
H Analyzed outside of Holding Time
J Analyte detected below quantitation limit
M Manually integrated, see raw data for justification
n Not offered for accreditation
ND Not Detected at the Reporting Limit
O Sample amount is > 4 times amount spiked
P Dual Column results percent difference > 40%
R RPD above laboratory control limit
S Spike Recovery outside laboratory control limits
U Analyzed but not detected above the MDL/SDL
Acronym DescriptionDCS Detectability Check Study
DUP Method Duplicate
LCS Laboratory Control Sample
LCSD Laboratory Control Sample Duplicate
MBLK Method Blank
MDL Method Detection Limit
MQL Method Quantitation Limit
MS Matrix Spike
MSD Matrix Spike Duplicate
PDS Post Digestion Spike
PQL Practical Quantitaion Limit
SD Serial Dilution
SDL Sample Detection Limit
TRRP Texas Risk Reduction Program
Unit Reported Descriptionmg/L Milligrams per Liter
ALS Group USA, Corp Date: 13-Oct-17
Page 15 of 18
CERTIFICATIONS,ACCREDITATIONS & LICENSES
Agency Number Expire Date
Arkansas 17-027-0 27-Mar-2018
California 2919 2016-2018 31-Jul-2018
Illinois 004112 09-May-2018
Kentucky 123043 30-Apr-2018
Louisiana 03087 2017-2017 30-Jun-2018
North Carolina 624-2017 31-Dec-2017
North Dakota R193 2017-2017 30-Apr-2018
Oklahoma 2017-088 31-Aug-2018
Texas T104704231-17-19 30-Apr-2018
13-Oct-17Date: ALS Group USA, Corp
Page 16 of 18
CTB
10-Oct-2017 14:10Date/Time Received:
HS17100583
PBW - Houston
Work Order:
Client Name:
Sample Receipt Checklist
Received by:
Shipping container/cooler in good condition?
Custody seals intact on shipping container/cooler?
Custody seals intact on sample bottles?
Chain of custody present?
Chain of custody signed when relinquished and received?
Chain of custody agrees with sample labels?
Samples in proper container/bottle?
Sample containers intact?
Sufficient sample volume for indicated test?
All samples received within holding time?
Container/Temp Blank temperature in compliance?
Temperature(s)/Thermometer(s):
Cooler(s)/Kit(s):
Date/Time sample(s) sent to storage:
Water - VOA vials have zero headspace?
Water - pH acceptable upon receipt?
pH adjusted?
pH adjusted by:
Login Notes:
No Not Present
Yes No Not Present
Yes No Not Present
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
1.4c/0.6c uc/c IR154261010/10/2017 1900
Yes No No VOA vials submitted
Yes No N/A
Yes No N/A
CPT-7-2 171010 COC states 3 bottles, received 4. Extra vial is ~80% sediment
Checklist completed by: Cesar A. LiraDateeSignatureDateeSignature
11-Oct-201711-Oct-2017
ALS.HSWater Carrier name:Matrices:
Reviewed by: Dane J. Wacasey
Client Contacted: Date Contacted: Person Contacted:
Contacted By: Regarding:
Comments:
Corrective Action:
Yes
NoYesTX1005 solids received in hermetically sealed vials? N/A
ALS Group USA, Corp 13-Oct-17Date:
Page 17 of 18
Page 18 of 18