Module I-3 Pressure Instruments

download Module I-3 Pressure Instruments

of 41

Transcript of Module I-3 Pressure Instruments

  • 8/13/2019 Module I-3 Pressure Instruments

    1/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 1

    -

    Practical Demonstration of

    PRESSURE INSTRUMENTS

    MODULE NO. : I-3

    MODULE SUBJ.: Pressure Instruments

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    2/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 2

    Tasks:

    I-3.1 Demonstrate servicing and calibration of different types of pressure

    gauges.

    I-3.2 Service and calibrate a high/ low pressure switch.

    I-3.3 Service and calibrate a differential pressure switch

    I-3.4 Describe the meaning of the terms, zero, span, linearity when

    calibrating a pressure transmitter.

    I-3.5 Demonstrate servicing and calibrating a pneumatic pressure transmitter.

    I-3.6 Demonstrate calibration of a smart transmitter using a hand held

    communicator.

    I-3.7 Demonstrate servicing and calibrating an electronic transmitter

    I-3.8 Demonstrate isolation procedure from process and electrical supply on a

    pneumatic/electronic pressure transmitter.

    I-3.9 Check function / adjustment / maintenance of pressure regulators for

    process lines.

    I-3.10 Service and calibrate a pneumatic indicating controller.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.1

    Demonstrate servicing and calibration of different

    types of pressure gauges.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    3/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 3

    Reference: OJT Instructor to arrange reference catalogue forpressure gauges relevant to each working area.

    Materials: 1. Different types of pressure gauges as models.2. Cleaning rags

    Equipment & Tools: 1. Tool Box2. Pressure gauges tool kit

    Conditions: None.

    Requirements By Trainee:

    To study the task and familiarize himself. Be able to identify pressure gauges types. Be able to calibrate different types of pressure gauges. Be able to disassemble and reassemble any pressure gauge. Perform adjustments and service the pressure gauges. Describe an understanding to his trainer. Write observations and procedures in his work book.

    TASK No.: I-3.1 Continue

    Details:

    Pressure Gauges

    A pressure gauge is a device, which senses pressure and provides

    a visual representation of that pressure. Most pressure gauges have bourdon tube

    sensors. Vacuum gauges and low-range gauges often use bellows sensors.

    Differential-pressure gauges can use piston or bellows sensors. The sensor type is

    usually dictated by the preferred manufacturer and the required range.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    4/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 4

    Pressure gauges should be selected so that the expected operating pressure is in

    the center third of the gauge range. It is also important that the highest pressure

    that will ever be applied to the gauge be below the maximum reading. Usually, the

    gauge can be selected so that the gauge maximum is above the set pressure of the

    system relief valve and the normal pressure is in the readable range.

    Pressure gauges are sometimes liquid filled. This is to protect the gauge dial and

    movement from the atmosphere. The liquid fill also provides some pulsation or

    vibration dampening. Glycerin is the most common fill liquid.

    Pressure gauges lose accuracy when exposed to hot fluids. When the process

    temperature is above approximately 180F (82C) a siphon should be installed. If

    the process fluid will not condense, at ambient temperature, the siphon can be

    filled with a suitable fluid such as ethylene glycol or glycerin.

    Differential-pressure gauges are useful when a pressure difference that is small

    compared to the static pressure needs to be measured. Differential-pressure gauges

    differ from static-pressure gauges in that they have two pressure connections.

    Differential gauges must be installed with an equalizing valve so that they will not

    be overranged while disconnecting.

    TASK No.: I-3.1 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    5/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 5

    Calibration Check

    To check the calibration of a gauge it must be removed from the line or equipment

    and installed on a dead-weight tester, or test gauge manifold. Check calibration as

    illustrated at 1/3, 1/2, and 2/3 of the dial range. If the readings are outside of the

    tolerance limits, the instrument requires recalibration. Remove the front ring, glassfront (lens) and gasket. Then recalibrate as follows:

    TASK No.: I-3.1 Continue

    Recalibration

    There are three specific adjustments, these are:A.Zero adjustment,B.Range adjustment, andC.Linearity adjustment.(A) Zero Adjustment

    If the gauge shows an equal amount of incorrect reading, either too high or too

    low over the entire range, adjustment is made by resetting the indicating pointer.

    These are furnished in three types, i.e. micrometer, slotted and plain.

    The micrometer type is adjusted by holding the pointer firmly with one hand and

    turning the micrometer screw with the other. Follow the same procedure for the

    slotted type, except rotate the slotted hub with a screwdriver. For gauges equipped

    with plain pointer make minor adjustments by rotation the slotted scale. Where

    this adjustment is not sufficient, remove the pointer (using a hand jack ), and reset

    in desired position.

    (B) Range Adjustment

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    6/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 6

    If the gauge readings show a progressively increasing (plus or minus) error, from

    the lower to the upper end of the range, loosen the two slide screws (1) and

    proceed as follows: For plus error, the slide (2) must be moved outward; for minus

    error, the slide must be moved inward. After attaining proper adjustment, retighten

    the slide screws (1).

    (C) Linearity Adjustment

    Assuming the gauge reads accurately at high and low points (1/3 and 2/3 of gauge

    range) but inaccurate at the center of the scale linearity adjustment is required.

    A linearity adjustment, being the most difficult adjustment of the three due to the

    geometric function of all components, may in some instances be accomplished in

    the field by loosening the two movement mounting screws (3) and rotating the

    TASK No.: I-3.1 Continue

    movement. Without a technical gauge background this would be a trial and error

    adjustment.

    After each adjustment (which should require no more than a few degrees), the

    gauge should be re-tested through the complete range, after setting pointer at

    approximately 5% of scale range. Make a zero and/or range adjustment if

    required.

    Pointer Adjustment, Removal and Reassemble.

    In case where pointer adjustment is necessary, the following procedure should be

    followed:

    For liquid filled gauges drain the liquid from the case.

    Loosen the ring using a strap wrench or equivalent.

    A slotted screwdriver blade is required so that the screwdrivers blade will

    properly engage the slot on the pointer hub and clear the tip of the pinion shaft

    protruding from the hub. While holding the tail of the pointer, a clockwise rotation

    of the pointers hub will result in a counterclockwise pointer motion and viceversa.

    Care must be taken so as not to apply an excessive axial force to the pointer.

    If liquid filling is required, hold the gauge in a horizontal position and slowly fill

    the gauge with filling fluid.

    Should it be necessary to remove the pointer, the standard hand jack set should be

    used. When reassembling the pointer on lower connected gauges care must be

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    7/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 7

    taken to support the pinion shaft. This is accomplished by removing the fill plug in

    the back of the case and placing a support against the pinion shaft from the back

    side. On back connected gauges, position the driver against the pointer hub and

    push until the pinion stops against the socket. Contact will be felt. Hold the driver

    in this position and strike it lightly.

    In both cases, care must be taken to ensure the pinion shaft is not bent by heavy

    blows on the pointer driver. Gauge calibration must be checked after reinstalling

    the pointer.

    Tighten the ring to the case using a strap wrench.

    MODULE No.: I-3 Pressure Instruments

    TASK No.:I-3.2

    Service and calibrate a high/ low pressure switch

    Reference: OJT Instructor to arrange reference catalogue / servicemanual for pressure switches relevant to each working

    area.

    Materials: Cleaning rags.

    Equipment & Tools: 1. Tool Box2. Digital Maltimeter3. Pneumatic / Hydraulic pressure calibrator.

    Conditions: Work permit

    Requirements By Trainee:

    To study the task and familiarise himself. To select the proper tools / equipment to perform this task. Draw / Sketch the calibration set-up in his work book.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    8/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 8

    Recognises the difference between high & low switch in calibration procedure. Be able to perform routine service of a pressure switch. Discuss an understanding to his trainer.

    Write observations and procedure in his work book.

    TASK No.: I-3.2 Continue

    Details:

    Pressure Switches

    A) Electric Pressure Switches

    An electric pressure switch senses pressure and opens

    or closes a switch element at a set pressure to signal another electrical device.Electric pressure switches are available in a wide variety of styles.

    Most pressure switches trip at a pressure above atmospheric, and are called gauge

    pressure or simply pressure switches. Switches can also be manufactured to trip at

    a pressure referenced to a complete vacuum and are called absolute pressure

    switches. Those set to trip below atmospheric pressure are called vacuum

    switches and those which can be set either above or below atmospheric pressure

    are called compound switches. Some switches are manufactured so that the trip

    point is factory set, while others are field adjustable.

    Pressure switches are set to trip at a certain point with rising or falling pressure.

    When the pressure is returned to within the acceptable range, the switch does not

    reset at exactly the same point that it tripped. The difference in the trip

    point and the set point is called dead band or reset. The electricalswitch is usually either single-pole, double-throw or double-pole, double-throw.

    Figure 23 shows these types, as well as others less frequently used. The number of

    poles determines the number of separate circuits that can be controlled by the

    switch, single pole for one circuit and

    double-pole for two circuits. The double-throw term means that a common

    terminal is connected to either of two other terminals, normally open or normally

    closed.

    B) Pneumatic Pressure Switches/Pressure Pilots

    A pneumatic pressure switch senses pressure and

    opens or closes a small valve at a set pressure to supply or vent a pneumatic signal

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    9/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 9

    to another pneumatic device. Pneumatic pressure switches are commonly known

    as pressure pilots. They are frequently used when pneumatic shutdown and control

    TASK No.: I-3.2 Continue

    systems are selected. Often, pressure pilots are used in Division 1 areas, such as

    on wellheads even when the primary process control is electronic.

    Devices, which are similar to electric pressure switches, are called pneumatic

    pressure switches. Pneumatic pressure switches are equipped with a two-way or

    three-way valve instead of an electrical switch. The two-way valve is either open

    or closed. A three-way valve connects a common port with one of two other ports,

    depending on whether the switch is tripped or not.

    Devices, which have been designed to be pneumatic, are usually called pressurepilots. The most common types of pilots are the piston-actuated, known as stick

    pilots, and the bourdon tube actuated pilots. Stick pilots are more often used on

    wellheads and bourdon tube pilots are more often used on process equipment.

    Dead band or reset is equally important for pneumatic pressure switches /pressure

    pilots as for electric pressure switches. Pneumatic devices tend to have an even

    larger dead band than electric devices because more movement is required for

    actuation.

    Most pressure pilots are equipped with three-way pneumatic valves so that theycan be used either as a high-pressure pilot or a low-pressure pilot depending on

    how they are connected.

    Pressure Switch Calibration;

    Switch operating set point shall be verified at intervals to make sure the

    healthiness of the switch and the associated circuit.

    For any switch with a tag number prefix ending LL or HH, a maintenance by passis provided on the shut down panel. Care must be taken to operate the correct

    switch by an authorised person.

    Switch must be isolated, depressurised and drained in order to hook up the test

    equipment.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    10/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 10

    TASK No.: I-3.2 Continue

    TASK No.: I-3.2 Continue

    To calibrate a pressure switch, follows:

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    11/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 11

    Normally Closed:

    1. Clean the switch.2. Connect the switch to Wallace calibrator.3. Adjust the Multimeter to measure Ohm.4. Connect the red and brown leads of the switch to the meter.5. Increase the pressure slowly till you reach the fixed set point, (the meter reads

    open circuit).

    6. Decrease the pressure slowly and observe the Multimeter indication, record thepressure value when meter read closed circuit.

    7. Calculate the dead band value.Normally Open:

    1. Clean the switch.2. Connect the switch to Wallace calibrator.3. Select resistance measurement OHM for the multimeter.4. Connect the multimeter to the switch to red and brown leads.5. The meter reads open circuit.6. Increase the pressure more than the set point, till the meter reads zero OHM.7. Decrease the pressure slowly till the meter reads open circuit.8. Check the set point with the name plate set point.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.3

    Service and calibrate a differential pressure switch

    Reference: OJT Instructor to arrange reference catalogue / servicemanual for diff. Pressure switches relevant to each

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    12/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 12

    working area.

    Materials: Cleaning rags.

    Equipment & Tools: 1. Tool Box,2. Digital Multimeter, and3. Pneumatic / Hydraulic pressure calibrator.

    Conditions: Work permit

    Requirements By Trainee:

    To study the task and familiarise himself. To select the proper tools / equipment to perform this task. Draw / Sketch the calibration set-up in his work book. Recognises the difference between high & low switch in calibration procedure.

    Be able to perform routine service of a differential pressure switch. Discuss an understanding to his trainer. Write observations and procedure in his work book.

    TASK No.: I-3.3 Continue

    Details:

    Deferential Pressure Switches

    Description;

    These switches act on the difference between pressures acting on either diaphragm

    or bellows. There are two kinds of differential pressure switches:

    Single switches; these switches control only one differential pressure to one limit.

    Double Switches; these switches control the differential pressure to two limits.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    13/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 13

    TASK No.: I-3.3 Continue

    Differential Pressure Switch

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    14/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 14

    Calibration Procedure

    1. Open the low pressure port to the atmosphere.2. Apply the setting pressure to the high pressure port.3. Adjust the multimeter to (OHM) position.4. Connect the multimeter leads to the terminal strip. (Normally open).5. Use a suitable open-end wrench to turn HEX adjusting nut.6. Adjust the setting point until the multimeter measure the continuity of the

    micro switch.

    7. Repeat step 6 till you get a precise set point.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.4

    Describe the meaning of the terms, zero, span,

    linearity when calibrating a pressure transmitter.

    Calibration Set-up of Diff. Press. Switch

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    15/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 15

    Reference:

    Materials: None.

    Equipment & Tools: None.

    Conditions: None.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to define Zero, Span and Linearity of a pressure transmitter, Discuss an understanding with his trainer, and Write the definition of each in his work book.

    TASK No.: I-3.4 Continue

    Details:

    Zero Pressure

    Is the bottom scale value of measuring range, for example if measuring range of a

    press transmitter is from (-2 to 40 psi), zero press. is equal to (-2 psi).

    Span

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    16/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 16

    The difference between the Upper Range Value (URV) and the Lower Range

    Value (LRV) of an instrument, on instrument starting at zero, the span equal to the

    range, for above example the span = 42 psi.

    Linearity

    Linearity is the degree to which the calibration curves of matches a straight line,

    the linearity error is generally the greatest departure from the best straight line that

    can be drawn through the measured calibration points.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.5

    Demonstrate servicing and calibrating a pneumatic

    pressure transmitter

    Reference: OJT Instructor to arrange reference catalogue / Service

    manual for pneumatic pressure transmitter modelrelevant to each working area.

    Materials: Cleaning rags with solvent.

    Equipment & Tools: 1. Tool Box,2. Hydraulic or Pneumatic pressure calibrator, and

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    17/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 17

    3. Standard output gauge.

    Conditions: Work permit.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the operation principle of the pneumatic press. transmitter, Be able to identify the main parts of a pneumatic transmitter, Perform periodic adjustments / calibration of a pneumatic transmitter, Perform periodic and corrective maintenance or replace parts of a pneumatic

    transmitters,

    Draw / Sketch calibration set-up of a pneumatic press. transmitter, Discuss an understanding with his trainer, and Write observation and procedures in his work book.

    TASK No.: I-3.5 Continue

    Details:

    Pressure Transmitters

    Pressure transmitters are used when the controller, recorder,

    or indicator needs to be located in a control room or panel where it is undesirable

    to pipe the process fluid. They are also used when several devices are to be

    operated from a single measurement or when elevated zero is required. The output

    is usually 4-20, milliamps for electronic transmitters or 3-15 psig (20-100 kPa) for

    pneumatic transmitters. Other signals can be used if required by the receiver, but

    these are the most common and should be used if possible.

    Elevated zero occurs when the base value of the measured variable is not at

    atmospheric pressure. Most transmitters have this as an option, while most

    controllers do not. Elevated zero is used when the pressure range of

    interest is narrow and at a fairly high level.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    18/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 18

    Suppressed zero, where the base value of the measured variable is below

    atmospheric pressure, is sometimes available. Usually, the zero is as near to

    perfect vacuum as possible and the unit is called a absolute pressure transmitter.

    The use of pneumatic transmitters is decreasing; however, a number of

    manufacturers still make them for the replacement market and some newinstallations are still being made. Pneumatic transmission may be advantageous

    when existing equipment is pneumatic with which operating personnel are already

    familiar.

    The next figure shows the main parts of a pneumatic pressure transmitter, these

    are:

    1. Bellows Capsule as a sensing element,2. Force Bar,3. Leaf Spring,4. Range Adjusting Bar,5. Booster Relay, and6. Feedback Bellows.TASK No.: I-3.5 Continue

    TASK No.: I-3.5 Continue

    Clibration

    Calibration is required if the transmitter has been taken apart for cleaning or for

    parts replacement, if a change of range is desired, or if the amount of zero

    elevation or suppression is changed substantially.

    If the capsule was removed or the flexure cap screw loosened, before calibrating,

    make the flexure cap screw adjustment above.

    The transmitter to be calibrated to 3 to 15 psi signal pressure range, at which the

    receiver signal is used.

    Calibration procedures; Detailed bench calibration and in-line calibration

    procedures of the pneumatic pressure transmitter are listed in the reference

    catalogue or service manual of the applied model. Consult your trainer.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    19/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 19

    Maintenance and Servicing

    Maintenance and servicing of pneumatic pressure transmitter are limited to clean

    or replace its parts, such as:

    1. Supply Air Filter blow out at least once a day,2. Replace Screen Filter of the process inlet,3. Clean Nozzle Assembly,4. Clean booster relay Restrictor,5. Replace Booster Relay,6. Change Range Bar,7. Remove / Replace Bellows Capsule, and8. Adjusting Flexure Cap Screw.Disassembly; normal servicing of the transmitter does not require the removal of

    any parts other than those already mentioned. Further disassembly is notrecommended because of possible loss of accuracy or damage to the transmitter,

    detailed servicing procedures are mentioned in maintenance section of the selected

    transmitter model, Consult your trainer.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.6

    Demonstrate calibration of a smart transmitter using a

    hand held communicator.

    Reference: OJT Instructor to arrange reference catalogue / servicemanual for smart transmitter model relevant to each

    working area..

    Materials: None.

    Equipment & Tools: 1. Function generator,2. Digital Multimeter, and3. Pneumatic / Hydraulic pressure calibrator.4. HHC for Smart Transmitter model.

    Conditions: None.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    20/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 20

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe using of HHC keypad,

    Be able to define the difference between the electronic press. transmitter andsmart type,

    Perform adjustments and calibration of Smart transmitters, Be able to maintain and troubleshoot Smart pressure transmitters, Draw / Sketch calibration set-up, Discuss an understanding to his trainer, and Write observations and procedures in his work book.

    TASK No.: I-3.6 Continue

    Details:Smart Transmitters

    Microprocessor-based pressure transmitters, sometimes called

    smart transmitters are offered by several companies. These transmitters can be

    remotely programmed for any desired range, either by the main process computer

    or by a special programmer. These transmitters are usually used in sophisticated

    computer control schemes.Smart transmitters uses an integrated circuit sensor to measure the process

    variable . This integrated circuit contains microprocessor-based electronics and

    digital communication circuitry to add functions over the same two wires used to

    transmit power and the measurement signal. The next figure shows electronics

    assembly of Foxboro 863DP smart transmitter.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    21/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 21

    TASK No.: I-3.6 Continue

    Smart Press. Transmitter Calibration Set-up;

    Calibration equipment should be a minimum of three times more accurate than the

    desired accuracy of the transmitter.

    Smart Press. Transmitter, Electronic Parts.

    4 to 20 mA Calibration Setu

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    22/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 22

    Calibration is performed by simulating the process pressure. This done by

    applying a pressure equal to the transmitter range pressure.

    Field Calibrationis performed without disconnecting the process piping.

    To perform field calibration, refer to detail procedure in the reference doc.,

    Consult your trainer.

    Bench Calibrationset-up requires disconnecting the process piping

    To calibrate the transmitter , use the detailed procedure of the reference doc.,

    consult your trainer.

    TASK No.: I-3.6 Continue

    Fault Analysis;

    Fault analysis of Foxboro Smart Press. Transmitter (863 DP) is accomplished

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    23/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 23

    through the following flow shart.

    TASK No.: I-3.6 Continue

    Fault Analysis Flowchart Continue

    TASK No.: I-3.6 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    24/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 24

    Maintenance;

    These transmitters when used in hazardous area, must be deenergised before

    removing the covers or performing any maintenance. Maintenance is limited to the

    replace electronics modules assembly, sensor and housing assembly, cover O-rings, and optional indicator.

    Detailed procedure to replace these parts, refer to the reference catalogue or

    maintenance manual of the smart transmitter model, consult your trainer.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.7

    Demonstrate servicing and calibrating an electronic

    transmitter

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    25/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 25

    Reference: OJT Instructor to arrange reference catalogue / Servicemanual for electronic pressure transmitter model

    relevant to each working area.

    Materials: Cleaning rags with solvent.

    Equipment & Tools: 1. Tool Box,2. Hydraulic or Pneumatic pressure calibrator, and3. Function Generator and digital Multimeter.

    Conditions: Work permit.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the operation principle of the electronic presser transmitter, Be able to identify the main parts of an electronic transmitter, Perform periodic adjustments / calibration of an electronic transmitter, Perform periodic and corrective maintenance or replace parts of an electronic

    transmitters,

    Draw / Sketch calibration set-up of the electronic press. transmitter, Discuss an understanding with his trainer, and Write observation and procedures in his work book.

    TASK No.: I-3.7 Continue

    Details:

    Electronic Pressure Transmitters

    Electronic transmitters with 4-20 milliamp outputs are the most common.

    Pressure transmitters are available in a variety of ranges. The ranges available

    vary from one manufacturer to another, so it is necessary to consult manufacturer's

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    26/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 26

    literature before selection. The range and the span are two different parameters.

    The span is the actual pressure range to be measured after the transmitter has been

    adjusted. The range is the pressure range within which the span can be adjusted.

    Most transmitters have two adjustments, zero and span. The zero is adjusted so

    that the output is minimum (4 ma, 3 psig, or 20 kPa) at zero pressure, or thebottom of the span if elevated zero is used. The span is adjusted so that the output

    is maximum (20 ma, 15 psig, or 100 kPa) when the pressure is at the top of the

    span.

    But the zero must always be rechecked after the span has been adjusted and the

    span checked after the zero adjustment.

    Next figure shows the main parts of an electronic pressure transmitter;

    Sensing module, and Electronic housing contains; header assembly board, DC ampilfier board, and

    calibration board

    TASK No.: I-3.7 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    27/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 27

    Electronic Press. Transmitter, Main Parts.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    28/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 28

    TASK No.: I-3.7 Continue

    Principle of operation;

    This electrical block diagram illustrates the operation of the electronic pressure

    transmitter.

    The process pressure pressure is transmitted through an isolating diaphragmand oil fill fluid to a sensing diaphragm located in the centre of the cell. The

    position of the sensing diaphragm is detected by the capacitance plates , the

    sensor is driven by an oscillator.

    Demodulator consists of a diode bridge, which rectifies the ac signal to dc todrive the oscillator.

    Voltage regulator provide a constant voltage of 6.4 V dc for the reference and 7V dc to supply the oscillator.

    Current control amplifier drives the current control circuitry to a level thatcauses the current detector to feedback a signal.

    The current limit prevent output current from exceeding 30 mA in case ofoverpressure .

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    29/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 29

    TASK No.: I-3.7 Continue

    This figure shows the signal terminals which are located in the electrical housing

    in a separate compartment. All power to the transmitter is supplied over the signal

    wiring and there is no additional wiring required. Signal wiring need to beshielded, but twisted pair should be used for best result. Output current is limited

    to 30 mA dc on the 4-20 mA dc unit.

    Calibration

    Span Adjustment Range; is continuously adjustable to allow calibration any

    where between maximum span and 1/6 of maximum span.

    Zero Adjustment Range; Transmitters zero is the lowest value of the transmitter

    range, at which the output of the electronic pressure transmitter is 4 mA.

    The zero and span adjustment potentiometers are accessible externally and are

    located behind the name plate of the transmitter.

    TASK No.: I-3.7 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    30/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 30

    The recommended procedure for zero and span adjustment detailed in the

    reference catalogue / service manual of the electronic pressure transmitter model,

    consult your trainer.

    Linearity Adjustment; this is a factory calibration adjusted for optimum

    performance over the calibrated range of the instrument and is not normallyadjusted in the field, however only maximise linearity over particular range be

    applicable.

    Damping Adjustment; the output amplifier boards are designed to permit

    damping of rapid pulsations in the pressure source by adjusting the control marked

    Damping located on the solder side of the amplifier board.

    Calibration procedures for adjusting or changing ranges are outlined in the

    calibration section of the transmitter catalogue, consult your trainer to proceed.

    Maintenance

    Maintenance is a technique for checking out the transmitter components, the

    method for disassembly and reassembly, and a trouble shooting guide.

    Test Terminal; to assure that their is no leakage current through the diode when a

    test reading is resistance of the test connection or meter should not exceed 10

    ohms.

    Sensing Module Checkout; the sensing module is not a field repairable and mustbe replaced if found to be defective. If no obvious defect, such as a punctured

    isolating diaphragm or loss of fill fluid is observed, the sensing module may be

    checked. The detail procedure to checkout the sensing module are listed in the

    maintenance section of the transmitters catalogue, consult your trainer.

    Circuit Boards Checkout; the printed circuit boards Amplifier board and

    Calibration board can most easily be checked for a malfunction by substituting

    into the circuit

    TASK No.: I-3.7 Continue

    TROUBLESHOOTING

    1) SYMPTOM: HIGH OUTPUT

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    31/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 31

    Potential Source and Corrective Action.

    Impulse Piping

    Check for leaks or blockage.

    Check for liquid in dry lines.

    Check for sediment in transmitter process flanges.

    Transmitter Electronics Connections

    Make sure bayonet connectors are clean and check out the sensor connections.

    Check that bayonet pin #8 is properly grounded to the case.

    Transmitter Electronics Failure

    Determine faulty circuit board by trying spare boards.

    Replace faulty circuit board.

    Power Supply

    Check output of power supply.

    2) SYMPTOM: LOW OUTPUT OR NO OUTPUT

    Potential Source and Corrective Action

    Loop Wiring

    Check for shorts and multiple grounds.

    Check polarity of connections.

    Check loop impedance.

    Impulse PipingCheck that pressure connection is correct.

    Check for leaks or blockage.

    Check for entrapped gas in liquid lines.

    Check for sediment in transmitter process flanges.

    TASK No.: I-3.7 Continue

    Transmitter Electronics Connections

    Check for shorts in sensor lead.

    Make sure bayonet connectors are clean and check the sensor connections.

    Check that bayonet pin #8 is properly grounded to the case.

    Test Diode Failure

    Replace test diode or jumper test terminals.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    32/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 32

    Transmitter Electronics Failure

    Determine faulty circuit board by trying spare boards.

    Replace faulty circuit board.

    3) SYMPTOM: ERRATIC OUTPUT

    Potential Source and Corrective Action

    Loop Wiring

    Check for intermittent shorts, open circuits and multiple grounds.

    Process Fluid Pulsation

    Adjust electronic damping pot (4-20 mA dc only).

    Impulse Piping

    Check for entrapped gas in liquid lines and for liquid in dry lines.

    Transmitter Electronics Connections

    Check for intermittent shorts or open circuits.

    Make sure that bayonet connectors are clean and check the sensor connections.

    Check that bayonet pin#8 is properly grounded to the case.

    Transmitter Electronics Failure

    Determine faulty circuit board by trying spare boards.

    Replace faulty circuit board.

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.8

    Demonstrate isolation procedure from process and

    electrical supply on a pneumatic/electronic pressure

    transmitter.

    Reference:

    Materials: 1. Soap Solution, and2. Cleaning Rags

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    33/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 33

    Equipment & Tools: 1. Tool Box, and2. Pneumatic or Hydraulic Pressure Source.

    Conditions: Work Permit

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to demonstrate safe process isolation procedure for the pressure

    devices,

    Be able to perform safe electrical isolation procedure for the pressure devices, Discuss an understanding to his trainer, and Write observations and isolation procedures in his work book.

    TASK No.: I-3.8 Continue

    Details:

    Preparations

    1. Fill in work permit.2. Use soap solution and brush to check any gas or air leakage.3. If there is any leakage from fittings, tighten it by a suitable open-end wrench.Flushing the transmitter impulse line; the figure below show typical installationof the pressure transmitter for either liquid or gas services, follow the next

    procedure for safe process isolation.

    1. Close isolating valve A.2. Slowly open drain valve C.3. Slowly open isolating valve A to flush the impulse line.4. After flushing the impulse line, close valve C, slowly open A, and put the

    transmitter on operation.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    34/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 34

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.9

    Check function / adjustment / maintenance of pressure

    regulators for process lines.

    Reference: OJT Instructor to arrange reference catalogue / Servicemanual for pressure regulators models in each working

    area.

    Materials: Cleaning Rags

    Equipment & Tools: 1. Tool Box,2. Standard test gauges, and3. Continuous air supply pressure.

    Conditions: Work permit.

    Typical Pressure Transmitter Piping

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    35/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 35

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to identify pressure regulators types, Be able to describe the principle of operation of continuous bleed regulator, State the advantages and disadvantages of continuous and continuous bleed

    regulators,

    Be able to dismantle, inspect, clean and replace a regulator, Discuss an understanding to his trainer, and Write observations in his work book.

    TASK No.: I-3.9 Continue

    Details:

    Pressure Regulators

    If a steady constant pressure is required either for instruments supply or process

    operations. This is ensured by fitting a pressure regulator in the line.

    A standard pressure regulator consists of a body in which is fitted a flexible

    diaphragm. The diaphragm is connected to a valve and is top loaded by a spring

    which can be adjusted.

    System or primary air pressure is led to the underside of the diaphragm. The

    diaphragm flexes, opening the valve and allowing air to pass to the outlet side

    where it is known as secondary air. A small amount of secondary air is bled to the

    top side of the diaphragm which, together with the spring, forces the diaphragm

    down and operates the valve. Any change in primary air pressure will cause

    movement of the diaphragm opening or closing the valve and stabilising thesecondary air pressure. Permanent changes in secondary air pressure can be made

    by adjusting the pressure on the spring with a screw.

    There are two types of pressure regulators;

    1. Continuous bleed , and2.Non continuous bleed

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    36/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 36

    TASK No.: I-3.9 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    37/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 37

    MODULE No.: I-3 Pressure Instruments

    TASK No.: I-3.10

    Service and calibrate a pneumatic indicating controller

    Reference: OJT Instructor to arrange reference catalogue / Servicemanual for pneumatic controllers models relevant to

    each working area.

    Materials: 1. Plastic / St.St. Tubing,2. Regulated Air Supply.

    Equipment & Tools: 1. Pneumatic calibrator bench or Druck Pump,2. Standard test gauges,3. Service/ Repair Kit, and4. Tool Box.

    Conditions: None.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to identify the major components of pneumatic controller, To describe the pricible of operation of pneumatic indicating controller, Be able to perform periodic adjustments / calibration of the controller, To perform previntive maintenance, service, parts replacement of the

    controller,

    Draw/ Sketch calibration set-up in his work book, Discuss an understanding to his trainer, and

    Continuous Bleed Press. Regulator

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    38/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 38

    Write observations and procedure in his work book.

    TASK No.: I-3.10 Continue

    Details:

    Pressure Controllers

    A pressure controller is a device, which senses the pressure

    in the process and vary an output which controls a device to regulate that pressure.

    The control device, or end element, is usually a pneumatic-control valve. The

    controller output is usually either a 3-15 or 6-30 psig (20-100 or 40-200 kPa)

    pneumatic signal.

    Pressure controllers can be categorized either as indicating or blind. The indicating

    controller has a mechanism so that the operator can read the process pressuredirectly from the controller. The blind controller has no direct-reading mechanism

    and the operator must rely on an adjacent pressure gauge or other device to know

    the process pressure. The indicating controller set point is usually marked on the

    indicator, thus it is easy to adjust to the desired point.

    Adjustment of the blind controller is more of a trial and error process. Indicating

    controllers are somewhat more expensive than blind controllers, but the cost

    difference is moderate if a pressure gauge can be eliminated.

    Pressure controllers must provide an output to control the end element. This can

    be an electric or pneumatic signal, but is most often pneumatic for field-mounted

    controllers. The pneumatic signal is usually 3-15 psig (20-100 kPa), but it can be

    6-30 psig (40-200 kPa) if required to reduce the control valve actuator size. The

    control action needed for pressure control is proportional plus integral, or P and I

    (integral is also referred to as reset by some manufacturers).

    The proportional action varies the output in proportion to the difference between

    the measured pressure and the set pressure. The integral or reset action gradually

    increases the amount of the correction until the measured pressure is returned tothe set point. A more extensive discussion of control modes and controller tuning

    can be found in the controllers catalogue or service manual, consult your trainer.

    TASK No.: I-3.10 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    39/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 39

    A common option for pressure controllers is an auto/manual switch. This is a

    valve, which allows the output of a manual regulator to be directed to the end

    element (valve actuator) instead of the controller's automatic output. The transfer

    can be either bumbles where the outputs are automatically matched to each other

    when the auto/manual switch is transferred, or manual balance where the operator

    must match the manual regulator output to the automatic output transfer to manualor the set point to the process variable before transfer to automatic.

    Pressure controllers are either surface, panel, pipe-stand, or yoke mounted.

    Surface-mounted controllers are fastened to a wall or other vertical surface. Panel-

    mounted, also called flush-mounted, controllers are mounted in a cutout in a

    control panel. Pipe-stand mounting occurs where a vertical or horizontal pipe

    support is constructed and the controller is provided with a bracket and U-bolts to

    attach it to a two-inch pipe-stand. It is not a good idea to support controllers on

    process piping. Yoke mounted controllers are fastened to the valve yoke withspecial brackets. Yoke mounting is convenient when the valve is accessible.

    TASK No.: I-3.10 Continue

    Principle of operation

    This figure explains in steps how the pneumatic indicator controller works;

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    40/41

    TASK DETAILING MANUAL

    Module I-3 www.arfanali.webs.com Page 40

    TASK No.: I-3.10 Continue

    Clibration

    Calibration is required if the controller has been taken apart for cleaning or for

    parts replacement, if a change of range is desired, or if the amount of zero is

    changed substantially.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 8/13/2019 Module I-3 Pressure Instruments

    41/41

    TASK DETAILING MANUALIf the capsule was removed or the flexure cap screw loosened, before calibrating,

    make the flexure cap screw adjustment above.

    The controller to be calibrated to 3 to 15 psi signal pressure range, at which the

    receiver signal is used.

    Calibration procedures; Detailed bench calibration and in-line calibration

    procedures of the pneumatic indicating controllerer are listed in the reference

    catalogue or service manual of the applied model. Consult your trainer.

    Maintenance and Servicing

    Maintenance and servicing of pneumatic indicating controller are limited to clean

    or replace its parts, such as:

    1. Supply Air Filter blow out at least once a day,2. Replace Screen Filter,3. Clean Nozzle Assembly,4. Clean booster relay Restrictor,5. Replace Booster Relay,6. Change measuring Range ,7. Remove Measurement Pointer or Set Point Index, and8. Adjusting controllers Alignment.Disassembly; normal servicing of the controller does not require the removal of

    any parts other than those already mentioned. Further disassembly is not

    recommended because of possible loss of accuracy or damage to the controller,

    detailed servicing procedures are mentioned in maintenance section of the selected

    pneumatic indicating controller model. Consult your trainer.