Module 3 Use, calibration and maintenance
of laboratory equipment
1
Learning objectives
At the end of this module you will be able to: describe briefly the operating characteristics of
biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators;
correctly use biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators;
describe the main maintenance operations for biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators.
2
Content outline
Principles, use and maintenance of:
• biological safety cabinet (BSC)• centrifuge• autoclave• inspissator (coagulator)• pH meter• incubator
3
Biological safety cabinet (BSC)
A ventilated contained area providingprotection for the operator and theenvironment against infectious aerosolsduring the handling of hazardous microorganisms
The most important equipment in all diagnostic mycobacteriology laboratories is the BSC.
4
HEPA filter
HEPA = high-efficiency particulate air
The HEPA filter traps and remove 99.97% of airborne particles of diameter 0.3 µm or more.
5
• BSCs should be ducted or vented to the outside.
• No recirculation into the room is allowed.
• The BSC should be connected to a UPS (Universal Power Supply).
BSC class I and II – main differences
BSC class I and II
BSC class Minimal facevelocity(m/s)
Air flow (%)
Re- Exhaustedcirculated
Exhaust system
I 0.38 0 100 Hard duct
IIA1 0.38 70 30 Thimble connection
IIA2 0.51 70 30 Thimble connection
IIB1 0.51 30 70 Hard duct
IIB2 0.51 0 100 Hard duct6
Class I BSC Protection of the user and the environment
Air is ducted to the outside of the building.
HEPA filter
7
Class II BSC
Protection of the user, the environment and the product
Class IIA BSC Class IIB BSC
8
Laminar flow hood for product (not user) protection
• Air blows towards workers.• Used only to protect product for sterile preparations
(media preparation).
Laminar vertical flowHorizontal airflow
9
BSC location
A B
CD
A , B – not recommendedD, C – recommended
10
BSC – operations for correct use• Switch on the BSC airflow and the UV light (if
present) 5–15 minutes before use.• Check the airflow velocity on the display.• Complete the logbook, recording details of operation
of the BSC.• Switch off the UV lamp.
Never work with the UV lamp on..• Disinfect the work surface, interior walls and interior
window surface with 70% ethanol or other suitable compounds.Do not use corrosive substances..
11
BSC – rules for correct use• Organize all items needed for work in the BSC to avoid
frequent displacements.• Organize the work to minimize arm movements. • Do not overcrowd the working area – this disturbs the
airflow. • Conduct all manipulations within the BSC as far towards
the back of the unit as possible (at least 15cm from the front grille).
• In class II BSCs, never allow the front grille to be covered with anything.
• Do not use large open flames in the BSC.• Do not accumulate waste in the BSC; remove it when
activities are finished.12
BSC – after use
• Autoclave the waste.
• Wipe down the inside of the BSC and work surface with 70% alcohol.
• Switch on the UV light for 1 hour (if the lamp is in good working order).
• Switch off the BSC fan.
13
Certification and maintenance
Regular certification and maintenance, with replacement of filters, are needed and represent a major challenge in low-income countries. If maintenance is not carried out regularly, the filters may become clogged and tubercle bacilli may be blown into the face of the operator. A BSC that is not well maintained is more hazardous than protective.
14
BSC – evaluation of performancePerformance of the BSC should be evaluated by the manufacturer or a qualified professional, with specific, well-calibrated equipment.
15
BSC – maintenance• Modern cabinets are equipped
with airflow indicators and warning devices.
• Maintenance, with replacement of filters, should be carried out by professionals when the airflow falls below the minimum recommended level.
The BSC should not be used until this maintenance has been carried out.
Out of order
16
BSC – maintenanceAirflow smoke pattern tests are performed to determine:
– whether the airflow along the entire perimeter of the work access opening is inward;
– whether airflow within the work area is downward with no dead spots or refluxing;
– whether ambient air passes onto or over the work surface;
– whether there is refluxing to the outside at the window wiper gasket and side seals.
17
BSC – maintenance
A cabinet leak test is carried out:– to determine whether exterior surfaces of all
plenums, welds, gaskets, and plenum penetrations or seals are free of leaks;
– before initial installation or annually on fully installed cabinets
18
BSC – fumigation• Decontamination of the BSC with formaldehyde gas.
• Hydrogen peroxide.
• Fumigation should be performed only by professionals.
System DECONTAKIT 19
BSC – fumigation
Fumigation of the BSC should be carried out:
– before release of the BSC for use after a major biohazardous spill;
– before replacement of HEPA filters;– before repair work requiring access to the sealed plenum;– before service or replacement of the circulation fan or
components;– before maintenance work in contaminated areas;– before performance tests requiring entry into contaminated
areas;– before movement of the BSC to another laboratory or area of
use;– before changing work programmes, e.g. to non-TB work;– before release of the BSC for resale or salvage.
20
BSC – logbook
21
BSC – maintenance logbook
22
Centrifuge – selection of proper equipment for TB culture
• The centrifuge should:
• Have aerosol-free (O-ring sealed) swing buckets.
• Have locking mechanism for protection to prevent opening before rotation has ceased.
• Be able to generate 3000g.
• Be refrigerated.
23
Centrifuge – rotors
24
Inadequate centrifuges and tubes
25
Prevent accidents by regular maintenance
26
Centrifuge – use• Inspect the centrifuge chamber for liquid or signs of
corrosion. • Check that there are shock-absorbing pads in the bottom
of the centrifuge buckets. • Balance the sample tubes and insert them in the buckets
in the appropriate position.–Never add water to a specimen to balance paired tubes
–Use an empty tube filled with water as a “balance” .
• Select the required time and speed.• Stop centrifuge immediately if any abnormal noise is
noticed.• Never operate a centrifuge with the lid open.• Open the sealed buckets in the BSC.
27
Centrifuge tubes – adaptors
• Centrifuge tubes must tolerate g-forces of at least 3000g.
• Centrifuge tubes must be used with appropriate rubber or plastic cushions matched to the tube and bucket-holder.
• Use only adaptors recommended by the manufacturer of the centrifuge in use.
28
RCF, g, RPM
RCF (or g) = relative centrifugal force
rpm = revolutions per minute
RCF = 1.12 * 10−6 * r * v2
or
RCF (g) = 1118 x r x (v/1000)2
r = radius in mm v = speed in rpm
rpm = 1000 *√RCF/1.12 * r
Always work in RCF, not rpm: newcentrifuges can convert automatically.
r (mm)
29
Centrifuge – installation
• Ensure that the centrifuge is installed on a rigid, flat, level surface.
• Because centrifuges produce vibration during operation, they must not be installed next to balances or other sensitive equipment.
• Allow sufficient free space around the centrifuge for ventilation to prevent overheating.
• Initial calibration should be performed only by a qualified service technician .
30
Centrifuge – daily maintenance
• Wipe the inside bowl with disinfectant solution and rinse thoroughly.
• If a refrigerated centrifuge is turned off at night, open the top to allow the bowl to dry. During the day when the unit is under refrigeration, keep the top closed to avoid condensation and ice build-up.
31
Centrifuge – monthly maintenance
• Clean the centrifuge housing, rotor chamber, rotors and rotor accessories with a neutral cleaning agent.
• Clean plastic and non-metal parts with a fresh solution of 5% sodium hypochlorite (bleach).
32
Centrifuge – annual maintenance
Annual service should be carried out by a qualified service technician: centrifuge brushes, timer, speed and electrical leaks should all be checked.
33
Centrifuge – logbook
34
Centrifuge – maintenance logbook
ITEM IDENTIFICATION
Equipment CENTRI-FUGE
Brand name
Purchase date
Model/type
Location within lab.
Serial no.
Warranty expiration date
Manufacturer
Address
Tel. Contact person
Technical service representative
Tel.
PERIODICITY :
Date Maintenance operation Operator
FAILURE EVENTS
Date Event Corrective action taken
Operator
35
Autoclave
The autoclave using saturated steam under pressure is the most efficient means of:•sterilization in a diagnostic TB laboratory;•decontamination of biological material consisting of infectious waste.
36
Autoclave
For optimum function:
• All of the air in the chamber should be replaced by steam.
• The temperature must be 121 ºC.
• Materials to be sterilized must be packed loosely.
37
Tips for loading the autoclave
• Only appropriate containers should be used.• Discarded cultures should be in solid-bottomed
containers no more than 40 cm deep.• Leave a large air space around each container.• Never cover containers.• If autoclaving closed containers, add water to
generate steam.
Separate autoclaves should be used for sterilization of solutions or glassware (clean materials) and for decontamination of infectious materials.
38
Autoclaves suitable for laboratory use
Pressure-cooker laboratory autoclave
Autoclave with air discharge by gravity displacement
39
Examples of autoclaves
40
Autoclave – use• Prepare material for autoclaving with thermosensitive paper.
• Do not load material to be sterilized with material to be decontaminated.
• Fill the bottom of the autoclave with distilled water.
• Open the air outlet valve.
• Turn on the heating.
• Close the outlet valve.
• Do not touch the drainage tap or the outlet or the safety valve while heating under pressure.
• When the required time is up, turn off the heating completely.
• When the temperature falls below 100 ºC, open the outlet valve slowly.
41
Autoclave – use• Never unscrew the lid clamps and open the lid
until the hissing sound has stopped.• Leave the sterilized/decontaminated material to
cool before removing it from the autoclave.• Check whether the autoclave tape has turned
black and the covering paper has turned brown. If they have not, the material is not decontaminated. Check the autoclave for malfunction.
• Biological indicators should be used after 40 hours of use.
42
Autoclave – suggested protocols
• Solid material for sterilization: 121 ºC (appropriate pressure 115 kPa), 20 min.
• Solid material for decontamination: 121 ºC (appropriate pressure 115 kPa), 30 min.
• Liquid material for sterilization: 15 min at 121 ºC if “liquid cycle” is available.
43
Autoclave – maintenance
• Check door gaskets for cracks and pitting.
• Check for proper operation of door clamps and door locks.
• Inspect valve discs and seats for signs of wear or cutting.
• Check pressure-release safety valves and thermometers.
44
Autoclave – logbook
45
Autoclave – maintenance and incidents logbook
46
Inspissator• Essential when egg media are “home-made” (made in the
laboratory).
47
Inspissator – use
• Carefully control the amount of heating.• Ensure the required temperature has been
reached before loading the tubes• Heat the tubes at constant temperature of
80–85 ºC for 45 minutes.• Tubes or bottles loaded at an angle of 5–
10º ( 10° ).• Always check the quality of media after
cooling.48
Inspissator – maintenance
• Check temperature at each use.• Clean after preparation of each batch of culture
media.
49
Inspissator – logbook
50
Inspissator – maintenance and incidents logbook
51
pH meter
52
pH meter – use
• Calibrate the pH meter before use at three pH values (pH 4.0, 7.0 and 10.0), using the calibration solutions provided by the manufacturer.
• Immerse the electrode in the test solution.
• Rinse the electrode after use.
• Cover the pH meter after use.
53
pH meter – calibration
• The pH meter must be calibrated according to the manufacturer’s instructions.
• Calibration must be performed once daily or, in case of infrequent use, at least on the day of pH testing. It must be performed before the first measurement of the day.
• Standardize with pH 4.0 and 7.0 buffers before each test or series of tests.
• Discard contaminated or cloudy standard buffers.
• Calibration results are acceptable if the pH values of the reference buffer solutions are within 0.1 pH units of the expected pH.
54
pH meter – Maintenance
• Keep the instrument clean. Cover it after each use.
• Rinse the electrode after use.
• Make sure that the electrode is always filled with electrolyte according to the manufacturer’s instructions.
• Do not touch the electrode membrane, which can be easily damaged.
• A glass electrode should be kept immersed in a standard salt solution for long-term storage.
• Calomel electrodes must be kept in a potassium chloride (KCl) buffer solution when not in use.
• Date buffer solutions and discard when unsatisfactory.55
Calibration solution Solution test
Date Temp.(ºC)
Electrode serial no.
pH 4 pH 7 pH8* Signature Solution/buffer
pH read Signature
pH-meter – calibration and record sheet
*pH 10 can be used if commercially purchased56
pH-meter – logbookITEM IDENTIFICATION
Equipment pHmeter Brand name
Purchase date
Model/type
Location within the lab
Serial no.
Warranty expiration dateManufacturer
Address
Tel. Contact person
Technical service representative
Tel.
FAILURE EVENTS
Date Event Corrective action taken
Operator
57
IncubatorsIncubators ensure optimum growth conditions for culture of M. tuberculosis.
Incubators are available in various sizes, from small (benchtop) models to large closets up to several hundred liters. Walk-in incubator rooms with circulating fans to obtain an homogeneous temperature are also frequently used.
58
Incubator – use
• Keep door(s) closed to prevent heat loss to the environment.
• Ensure that rack positions are clearly marked.
• Switch the incubator off when not in use.
• Do not overload.
59
Incubator – maintenance
• Record temperatures daily on appropriate forms.
• Check that the actual temperature corresponds with the thermostat setting.
• Clean every 14 days with 70% ethanol.
• Clean immediately after any infectious spills.
60
Incubator – calibration
Calibrate:
–before use;–after temperature changes have been detected and rectified;–following a power failure;–after cleaning of spillages.
61
Incubator – calibration• Ensure that the door is closed and that the incubator is
switched on.• Set the required temperature using the temperature
control and leave the incubator to run for 1 hour.• Place a thermometer in the centre of the incubator with
the probe away from the heating element.• Take the temperature reading after 30 minutes. If the
temperature is not 36 ± 1 ºC, adjust the control. • Repeat this process every 30 minutes until the required
temperature is reached.• Continue to take readings until two consecutive readings
(30 minutes apart) are 36 ± 1 ºC.
62
Incubator – maintenance form
63
Incubator logbook
Trimester : Year :
Month Temp. Visa Month Temp. Visa Month Temp. Visa
°C (Initials) °C (Initials) °C (Initials)
1 1 1
2 2 2
3 3 3
4 4 4
5 5 5
6 6 6
7 7 7
8 8 8
9 9 9
10 10 10
11 11 11
12 12 12
13 13 13
14 14 14
15 15 15
16 16 16
17 17 17
18 18 18
19 19 19
20 20 20
21 21 21
22 22 22
23 23 23
24 24 24
25 25 25
26 26 26
27 27 27
28 28 28
29 29 29
30 30 30
31 31 31
64
True and false exercise
1. The biological safety cabinet is the most important equipment for TB culture laboratories.
2. Only class II BSCs can be used for TB culture.
3. Centrifuge speed should always be expressed in rpm.
4. Centrifugation may produce hazardous aerosols.
5. Checking the autoclave regularly is unnecessary if it is used only for decontamination.
6. Equipment maintenance is crucial for optimal performance.
7. Maintenance can always be performed in-house.65
Module review: take-home messages
1. Equipment is expensive: all equipment must be treated with care and properly maintained.
2. Quality-control checks on equipment should be run regularly and results should be recorded on appropriate forms.
3. The biological safety cabinet is the most important item of equipment for TB culture under safe conditions.
4. Read carefully and always follow manufacturers’ instructions for routine work and maintenance.
5. Familiarize yourself with all working parts of all the instruments.
6. Call for help when troubleshooting any equipment malfunction.
66
Self-assessment
• Describe the differences between the two types of BSC.• What precautions must you take when working in a
BSC? • Describe the operations to be undertaken before starting
a centrifuge. • Explain the principles of sterilization using an autoclave.
67