Modern Radiotherapy
DR :- Omer Hashim
![Page 2: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/2.jpg)
Conformal &IMRT
In two-dimensional (2D) radiotherapy singlebeams from one or two directions were used. Beam setups were usually quite simple and treatment plans frequently consisted of opposed lateral fields [1, 2].
![Page 3: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/3.jpg)
three-dimensional (3D) conformal radiation therapy is a technique where the beams of radiation used in treatment are shaped to match the tumor. Previously, radiation treatment matched the height and width of the tumor, meaning that healthy tissue was exposed to the beams.
![Page 4: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/4.jpg)
In 3D & IMRT With the advent of CT imaging, which took into account axial Anatomical and complex tissue contours, three-dimensional conformal radiotherapy (3D-CRT) had arrived
![Page 5: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/5.jpg)
Conformal 3D:-The technique aiming to geometrical shaping of the fields alone
IMRT :-The technique to modulate the intensity of the fluency acrossThe geometrically shaped fields
![Page 6: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/6.jpg)
![Page 7: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/7.jpg)
Planning steps :-
CT scanning Refer
image
Volume delineati
on
3D image Dose
analysis
Treatment QA
Positioning
Treatment delivery
![Page 8: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/8.jpg)
1) Positioning &immobilization
This is the two most important steps in planningOf the 3D. Should be include the following :-1)Comfortable2)Reproducible3)Minimal beam attenuating 4)Affordable
![Page 9: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/9.jpg)
Positioning
supine prone
![Page 10: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/10.jpg)
Immobilization Frame base
Invasive
Nom invasive
Frame less
![Page 11: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/11.jpg)
Data acquisition
CT scanning is the corner stone in the conformal RT CT scanning provides detailed cross-sectional anatomy of the
normal organs, as wellas 3D tumor information. These images provide density data for
radiation dose calculation It different from diagnosis CT in the:_1)wider borne70—85 cms 2) CT couch must be flat3) using supporting aids and immobilization devices andaligned using tattoos and midline and lateral laser
![Page 12: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/12.jpg)
![Page 13: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/13.jpg)
In the ct room we do after position fixation of pts specify the organ to be treated thenContrast – laser -- tech SSD or SAD then we takeCT cut 3—5 cms according to part to scanning
![Page 14: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/14.jpg)
Volume delineation
Then referring data to the planning room and Target Volume delineation take place (contouring) common international language for describing target volumes is found in International Commission on Radiation Unit (ICRU) . Report 50 (1993), 62 (1999) and 71 (2004 )
![Page 15: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/15.jpg)
GTV(Gross tumor volume) :-The primary tumor or other tumor mass shown by clinical examination or images
CTV (Clinical target volume) :- consist of the GTV + Margin for subclinical microscopic disease. The GTV - CTV margin depend on the site of the tumors and alsoderived from biological characteristics of the tumor in
head &neck5—10 mms. LNs CTV = 3—5 mms.PTV (Planning target volume) :- this margin for
physiological organ motion (internal margin). variations in patient positioning. alignment of treatment beams (set-up
• margin)
![Page 16: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/16.jpg)
ITV:- additional margin CTV/GTV for physiological organ movement .OAR Organs at risk :- organs around the treated Volume that will be affected by radiation and have limited dose e.g- spinal cord .
![Page 17: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/17.jpg)
![Page 18: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/18.jpg)
![Page 19: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/19.jpg)
Drawing
Usually done in planning room by doctors with Special program. Selection for the color then drown GTV – add margin to CTV – add margin to PTV . THE drown OAR
![Page 20: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/20.jpg)
MRI
Also most drown program done for CT .MRI have the following:-Superior soft tissue resolution.Ability to assist neural and bone morrow infiltration .Ability to obtain image in any position coronal seggitallMetabolic effect of tumor – vascular effect -No radiation during scanning
![Page 21: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/21.jpg)
Image Registration & fusion
The techniques in which two identical point in imageIs determined (CT&MRI). Then fusion of MRI &CTBenefit form MRI advantage
![Page 22: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/22.jpg)
MRI-CT image fusion
MRI :- show soft tissue & tumor will CT :- show bone structure will
![Page 23: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/23.jpg)
CT/MR registration
![Page 24: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/24.jpg)
Fusing of the CT & MRI volumes,
![Page 25: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/25.jpg)
![Page 26: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/26.jpg)
![Page 27: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/27.jpg)
![Page 28: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/28.jpg)
Dose objective Total dose
Total time
Total fraction /N
Organ at risk dose
Number of beam
Beam &couch angle
planning techqunlogy
Forwardplanning inverse
planning
![Page 29: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/29.jpg)
Virtual simulation
type of simulation in which we use the data obtain by CT . Determine of the field then beam entering Is done in simulation room but not 3D
![Page 30: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/30.jpg)
IMRT
IMRT not depend on the shaping of the tumor Only as done by 3D but also use MLC to define the beam intensity independently in different regions of each incident beam by use of micro leaf lead to Creating of micro beam with different density
![Page 31: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/31.jpg)
Process of IMRT
Target delineation
Treatment planning and optimization
Quality assurance
Treatment deliver
![Page 32: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/32.jpg)
to produce the desired uniform distributionof dose, or a deliberate non-uniform dose distribution, in the target volume. Theposition of the leaves of the MLC can be varied in time with a fixed or movinggantry. IMRT can be delivered using dose compensation, multiple static fields, step-and-shoot, dynamic MLC or tomotherapy
![Page 33: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/33.jpg)
![Page 34: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/34.jpg)
Dynamic IMRT
![Page 35: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/35.jpg)
Type Intensity modulation method
Segmental MLC (step and shoot) Multiple MLC segments delivered from each treatment direction
Dynamic MLC (slidingwindow)
Leaves slide across the field at different rates
Intensity-modulated arc therapy (IMAT) Leaves move while the gantry is rotating. Can require multiple rotation arcs.
Serial tomotherapy Gantry rotates around the patient with the couch fixed. Binary leaves modulate a fan beam. Upon completion of each rotation, the couch ismoved in a step-wise fashion
Helical tomotherapy Gantry and couch move synchronously. Binary leaves modulate a fan beam
![Page 36: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/36.jpg)
Forward planning
The technology where the planner will try to variety of companion of the beam Engle .couch angle, beam weight and beam modifying to reach to the optimum dose destitution .all trial done manually Until reach to the optimum solution is reach
![Page 37: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/37.jpg)
Inverse planning
Planning when selection of the beam, couch angle, beam weight and beam modifying to reach to the optimum dose destitution done by the computer. Inverse planning requires specification of doseprescription to GTV, PTV and PRV in terms of dose–volume constraints, fluence
![Page 38: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/38.jpg)
![Page 39: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/39.jpg)
The Magic Bullet“Conformality”
Prostate
Rectum
Prostate
Rectum
Prostate
Rectum
Pre-IMRT IMRT IGRT
Radiation Field •Improved Outcome•Reduced Side effects
![Page 40: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/40.jpg)
avnatage of IMRT• Conformality• Concave shaped dose distribution
• Decreased penumbra• OAR sparing with sharp dose fall-off• Dose escalation• Lower rate of complication• Reduced cost of patient care
• Multiple simultaneous treatments• Least sensitive to physical planning
parameters• Only secondary field shaping devices-MLCs
![Page 41: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/41.jpg)
Delineation of OARs
OAR
: parallel
series
parotids, ears, TMJs, duodenum, liver, GI
: spinal cod, brainstem, optic chiasm, nerves, lens, brain, eyes, mandible, temporal lobes
![Page 42: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/42.jpg)
For serial organs, Planning organ at risk volume(PRV) should be delineated
![Page 43: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/43.jpg)
9 beams
80
120
150180210
240
280
7 beams
•Depends on target shape and location
Beam Placement
![Page 44: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/44.jpg)
• Optimum beam angles can be derived by using beam angle optimization available with TPS
Beam Angle Optimization
![Page 45: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/45.jpg)
Planning Objectives (Constraints)
• Required by the inverse planning process – dose or dose-volume constraints for all structures• A trial and error process to come up with the proper dose
or dose-volume constraints.• Don’t ask the impossible – set realistic goals – improperly
specified constraints will result in inferior plans.• Create site-specific protocols which can be used for
similar cases.
![Page 46: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/46.jpg)
Optimization
Definition:Systematic computerized process to generate a large number of plans rapidly and to evaluate and rank them according to some specified criteria
![Page 47: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/47.jpg)
2D fluence profiles of various beam ports is the free treatment parameterOther parameters like beam entry angles, beam energy, the type of radiation, number of beams are decided by the planner and are fixedEach treatment field is divided into independent spatial subunits called as ‘bixels’No of bixels depend on
Size of lesion to be irradiatedChoice of resolution of the fluence mapsSelected to correlate with the width of MLC usedFor MLC width of 10mm at isocentre, bixel 10 mmX10mm
![Page 48: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/48.jpg)
Traditional Optimization
Contour regions of interest
Define constraints and select beam parameters
Compute dose
Plan accepted
IMRT Optimization
No
Contour regions of interest
Define constraints and establish an objective function
Compute beamlet weights or beam segment shapes and weights
Optimization requirement meet
Plan acceptedNo
No
![Page 49: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/49.jpg)
IGRT The process of frequency image (2D, 3D) before and
During course of radiotherapy .high accurateAnd Precise treatment.IGRT is use because of organ movement:-Interfraction motionIntrafraction motion – Reduce severity and risk of therapy-induced
complications.– Increase both quality and probability of success.
![Page 50: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/50.jpg)
RT Targeting Uncertainty in RT plan • Setup Variation• Internal Organ Displacement• Volume Change and
![Page 51: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/51.jpg)
et up error :-Patient position out of plan –wt loss Un voluntary movement as reparation –patient
discomfort –boil gas motionOrgan motion :-bladder rectum filling . Intra abd-
pressure
![Page 52: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/52.jpg)
![Page 53: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/53.jpg)
![Page 54: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/54.jpg)
![Page 55: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/55.jpg)
![Page 56: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/56.jpg)
![Page 57: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/57.jpg)
![Page 58: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/58.jpg)
![Page 59: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/59.jpg)
![Page 60: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/60.jpg)
![Page 61: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/61.jpg)
![Page 62: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/62.jpg)
![Page 63: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/63.jpg)
![Page 64: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/64.jpg)
![Page 65: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/65.jpg)
![Page 66: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/66.jpg)
![Page 67: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/67.jpg)
![Page 68: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/68.jpg)
![Page 69: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/69.jpg)
![Page 70: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/70.jpg)
![Page 71: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/71.jpg)
IGRT
USG
BETSonoarray
I-beamResitu
Video based
Align RTPhotogrammetryReal 2 video G
IMRTVideo subtraction
Planner xray
KV xray OBI
Gantry mounted Room mounted
CT
Fan beam Cone beam
MRI
Mv CT KV CTSiemens Mobile CT arm
Varian OBI
Eelekta
Siemens inline
![Page 72: Modern Radiotherapy](https://reader034.fdocuments.in/reader034/viewer/2022051503/587627011a28ab8b7b8b5cbb/html5/thumbnails/72.jpg)