Model UFM Compact Ultrasonic Flowmeter Installation/Setup

8
UFM Compact Ultrasonic Flowmeter Installation The materials included in this compilation are for the use of Dwyer Instruments, Inc. potential customers and current employees as a resource only. They may not be reproduced, published, or transmitted electronically for commercial purposes. Furthermore, the Company’s name, likeness, product names, and logos, included within

description

The materials included in this compilation are for the use of Dwyer Instruments, Inc. potential customers and current employees as a resource only. They may not be reproduced, published, or transmitted electronically for commercial purposes. Furthermore, the Company’s name, likeness, product names, and logos, included within these compilations may not be used without specific, written prior permission from Dwyer Instruments, Inc. ©Copyright 2014 Dwyer Instruments, Inc.

Transcript of Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Page 1: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

UFM Compact Ultrasonic Flowmeter

Installation

The materials included in this compilation are for the use of Dwyer Instruments, Inc. potential customers and current employees as a resource only. They may not be reproduced, published, or transmitted electronically for commercial purposes. Furthermore, the Company’s name, likeness, product names, and logos, included within these compilations may not be used without specific,

written prior permission from Dwyer Instruments, Inc. ©Copyright 2014 Dwyer Instruments, Inc.

Page 2: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Suitable Applications

Use with Water OnlyFully Developed Flow

Minimal Air in PipeMinimal Sludge in Pipe

Suitable Pipe Run Available

Page 3: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Installation Location

Install UFM at 45° angle Avoids Interference with Air and Sludge

Page 4: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Installation Location

Ensure Suitable Location on Pipe20 Pipe Diameters Upstream & 10 Downstream

Page 5: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Set-Up

Enter Password: 71360

Enter Pipe ID

Enter Pipe Material

Obtain Separation Code

Page 6: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Sensor Separation

Device Displays Sensor Location Code (e.g. A2)

Position Sensors According to Code

PASSWORD:71360

Page 7: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

Attach to PipeApply Silicon Grease to Sensors

Attached to Pipe

Snap on Top Assembly

Connect Sensor Wiring

PASSWORD:71360

Page 8: Model UFM Compact Ultrasonic Flowmeter Installation/Setup

The materials included in this compilation are for the use of Dwyer Instruments, Inc. potential customers and current employees as a resource only.  They may not be reproduced, published, or transmitted electronically for commercial purposes.  Furthermore, the Company’s name, likeness, product names, and logos, included within these compilations may not be used without specific, written prior permission from Dwyer Instruments, Inc.   ©Copyright 2014 Dwyer Instruments, Inc.