S by SR Instruments, Inc. Model SR465 In-Floor Platform Scale System OWNERS MANUAL SR Instruments, Inc. 600 Young Street, Tonawanda, NY, 14150 Tel: 888 WEIGH ME Fax: (716)693-5854 Part No. MAN465 rev: 000706

Transcript of Model SR465 In-Floor Platform Scale SystemModel SR465 In-Floor Platform Scale System OWNERS MANUAL...

SSby SR Instruments, Inc.

Model SR465In-Floor Platform

Scale System

OWNERS MANUAL

SR Instruments, Inc. 600 Young Street, Tonawanda, NY, 14150Tel: 888 WEIGH ME Fax: (716)693-5854

Part No. MAN465 rev: 000706

Table of Contents

SECTION IOPERATION and MAINTENANCE

System Description and Specifications Page 1

System Operation and button Functions Page 2

SECTION IITECHNICAL SERVICE INFORMATION

Calibration and Maintenance Page 4

Care and Cleaning Page 5

Schematic Diagram Page 6

Part List and Component Layout Page 7

Platform Installation Instructions Page 8

Part No. MAN465 rev: 000706

MODEL SR465FLOOR SCALE SYSTEM

The SR465 is a Precision Digital Scale System specifically designed to be installedin the floor so that the weighing surface is flush. The multi-purpose design offers theversatility of being used as a Stand-On scale for weighing ambulatory patients aswell as a Wheelchair / Chair / Walker Scale for semi or non-ambulatory applications.

This Scale System employs the latest in microprocessor and load cell technologyto provide accurate and repeatable weight data. Four (4) identically matched loadcell transducers are strategically placed to ensure an accurate representation ofthe patient's weight regardless of weight distribution.

The low power microprocessor circuitry allows the SR465 to derive it's power from six (6) common"D" cell alkaline batteries which will provide over 20,000 weight reading before needing replacement.This eliminates the need for an external battery charger or an A.C. outlet.

The Patient's weight is displayed on a user friendly 16 character dot matrix liquidcrystal display. This display uses bright green L.E.D. back lighting for superior visibilityin all lighting conditions. With the push of a button, patient weight may be viewed inboth POUNDS and KILOGRAMS with a displayed resolution of 0.1 #/Kg (100 grams).

SPECIFICATIONS

Maximum Capacity 1000 LB's / 454Kg'sPlatform Size 32" x 36"

Display Resolution 0.1 LB's / 0.1 Kg's (100 grams)Accuracy 0.1% +/- 1 digit of reading

for calibrated range.Display Type 16 character dot matrix L.C.D.

ZERO One Button Auto-TareDisplay Time Auto power-down 60 sec.

Damping Automatic digital filterPower Supply 6 -"D" cell batteries (LR20)

Warranty Two Years Parts and Labor

page 1

Part No. MAN465 rev: 000706

MODEL SR465 FLOOR SCALE SYSTEM OPERATION

INITIALIZING THE SYSTEM ZERO:Before the patient is positioned on to the scale system you must first set the system zero. This

is done as follows;

1. Press the "O" button. The displayed message will indicate " HANDS OFF ". Makesure that nothing is in contact with the scale system while zeroing. In a few secondsthe display will indicate "WEIGHT = 0.0 LB ".

Note: When the unit is first turned on the mode will default to the last mode used.

The patient may now be positioned on to the weighing surface to view the weight data.

BUTTON FUNCTIONS :

The "0" button is used to initialize the system zero before placing a patient on to thescale system. When the "0" button is pressed, the display message will indicate"HANDS OFF". At this point you should make sure that nothing is in contact with theweighing surface so that the system zero will be accurate. In a few moments thedisplay will read - "WEIGHT = 0.0 Lb".

The display mode will automatically default to the last mode used

If after positioning the patient on to the scale system, the display shuts off beforeyou are able to view the weight data, simply press the "WEIGH" button to WAKE thedisplay and view the patient's weight.

NOTE: It is recommended that you ZERO the system before each weighing toassure accuracy.

The "HOLD" button will freeze the displayed patient weight and store it away inmemory. To recall the LAST WEIGHT after the system has shut off, press the"HOLD" button BEFORE pressing any other. The system only recalls the LASTWEIGHT that was stored by HOLD.

Weight data may be viewed in either Pounds or Kilograms. Pressing the "MODE"button allows you to toggle the POUNDS and KILOGRAMS display modes. Bothmodes will display a resolution of 0.1 pounds or kilograms.

page 2

Part No. MAN465 rev: 000706

ZERO0

WEIGH

MODE

LBKG

HOLD

ZERO0

SECTION II

TECHNICAL SERVICE INFORMATION

For Sales, Product Informationand Technical Assistance Please Call:

1(800)654-6360

Part No. MAN465 rev: 000706

SYSTEM CALIBRATION and MAINTENANCE SIMPORTANT : The following calibration procedure should only be performed by qualifiedservice personnel. The SR465 load cell, itself, has no user serviceable components and shouldnot be tampered with for any reason. Recalibration is generally not required, but may be checkedperiodically to ensure accuracy.

CALIBRATION PROCEDURE :

NOTE: Make sure that nothing is in contact with the scale system during this procedure. Be sureto remove your hands from the system when noting the displayed calibration results.

1. Press and Hold both buttons simultaneously.2. The display will read "HOLD TO CAL" as the right hand digit counts down toenter the CAL MODE.3. When in the CAL MODE, press the "ZERO" button to zero the display.4. Place a known calibrated weight on to the weighing surface and compare it tothe displayed reading.5. Use the LEFT CAL button to make "+" corrections and the RIGHT button tomake "-" corrections to the displayed weight. The displayed value should be within0.1% of the calibrated weight plus or minus 1 digit of reading.

When adjustments are completed:press the "WEIGH" button to ABORT your changes and exit the CAL mode orpress the "HOLD" button to SAVE all changes and exit the CAL mode.

CALIBRATION TOLERANCE TABLE

LOW APPLIED HIGHLIMIT LOAD LIMIT99.9 100.0 100.1199.8 200.0 200.2299.7 300.0 300.3399.6 400.0 400.4499.5 500.0 500.5599.4 600.0 600.6699.3 700.0 700.0799.2 800.0 800.8

899.1 900.0 900.9999.0 1000.0 1001.0

page 4

Part No. MAN465 rev: 000706

ZERO0

WEIGH

by SR Instruments, Inc.

HOLD MODE

LBKG

SS

HIDDEN "CAL" BUTTONS+ -

SYSTEM CARE and CLEANINGCLEANING and DISINFECTING:

The readout housing for the SR465 Scale Systems is primarily made of aluminum. You needonly exercise caution when cleaning the display window of the electronics as this is made of clearpolyester . We recommend mild soap and water for general cleaning.

The floor transducer is covered by a Stainless Steel plate and may be vacuumed or swept.

BATTERY REPLACEMENT (model SR465 only):

When battery replacement is necessary the display will indicate "REPLACE BATTERY".To replace the batteries

1. Remove the instruction cover plate to expose the battery compartment..2. Replace ALL SIX (6) Batteries as shown.3. Replace the cover plate.

page 5

Part No. MAN465 rev: 000706

+ + +

+++COVER PLATE

SCHEMATIC and COMPONENT LAYOUT

Part No. MAN465 rev: 000706

PARTS LISTPart No. MAN465 rev: 000706

Parent Item No_ No_ Description Quantity per Position Position 2 Position 3SE1857D255 EPCB1857D P.C.BOARD 1857 REVISION "D" 1SE1857D255 EC347RM .0047MFD #167472J400A(MALLORY) 1 C4SE1857D255 EC510RF 0.1MFD #SR205C104KAA (AVX) 5 C10,C11, C12,C13, C14SE1857D255 EC747RE016 47MFD #UVX1V470MPA(NICHICON) 1 C1SE1857D255 EC710RE016 10MFD #UVX1C100MAA (NICHICON) 1 C2SE1857D255 EC810RE016 100MFD 16V CAP #UVX1C101MPA 1 C3SE1857D255 EC510RM050 .1MFD #R68104J63B (NISSEI) 4 C5,C8,C9, C22SE1857D255 EC610RM050 1MFD 168105J50G(B32529-105-63J 1 C6SE1857D255 EC210RP 100PF POLY CAPACITOR 1 C7SE1857D255 EC847RE016 470MFD RADIAL (UVX1C471M) 2 C20,C21SE1857D255 ED1N4001 DIODE 1N4001 (RECTRON) 1 D1SE1857D255 ESRBE085 SWITCH (RAYCHEM) RBE085A1 1 F1SE1857D255 EIICL7660C IC #ICL7660CPA (SEE TC7660CPA) 1 IC1SE1857D255 EIMAX135CP IC #MAX135CPI (MAXIM) 1 IC2SE1857D255 EICD4078BE IC #CD4078BE 1 IC4SE1857D255 EINM93C46N IC #93LC46B-I/P (MICROCHIP) 1 IC5SE1857D255 EIICL7652A IC #ICL7652CPA (MAXIM) 1 IC6SE1857D255 EITLE2426C IC #TLE2426CLP (TEXAS INSTR.) 1 IC7SE1857D255 EIMAX667CP IC #MAX667CPA (MAXIM) 1 IC9SE1857D255 FCBSW11404 TERMINAL STRIP #BSW-114-04-G-S 1 J5ASE1857D255 FCTSW11409 TERMINAL STRIP #TSW-114-09-G-S 1 J5BSE1857D255 ETZVNL110A TRANSISTOR,VOLTAGE REF.ZVNL110 1 Q1SE1857D255 ET2N3906 TRANSISTOR 2N3906 (MATOROLA) 1 Q2SE1857D255 ER17505W3 75 OHM 3 WATT 5% RESISTOR 1 R2SE1857D255 EV5KKVSF82 5 K POTENTIOMETER 1 R1SE1857D255 ER41005QF 10K 1/4 WATT 5% RESISTOR 2 R3,R4SE1857D255 ER62001EF 2MEG 1% RES 1 R5SE1857D255 ER57503EJ 750K 1/8 WATT 1% J RESISTOR 1 R12SE1857D255 ER44531EJ 45.3K 1/8 WATT 1% J RESISTOR 2 R6,R8SE1857D255 ER43241EJ 32.4K 1/8 WATT 1% J RESISTOR 1 R7SE1857D255 ER72205QF 22M 1/4 WATT 5% RESISTOR 1 R9SE1857D255 ER51301EJ 130K 1/8 WATT 1% J RESISTOR 2 R10,R11SE1857D255 EV10K3299W 10K POT 3299W-1-103 (BOURNS) 1 R13SE1857D255 ER64705QF 4.7M 1/4 WATT 5% RESISTOR 1 R14SE1857D255 ER61205QF 1.2M 1/4 WATT 5% RESISTOR 1 R15SE1857D255 ER54991EJ 499K 1/8 WATT 1% J RESISTOR 1 R16SE1857D255 ER1005QF 100K 5% RES 1 R17SE1857D255 ER61001EJ 1M 1/8 WATT 1% J RESISTOR 2 R18,R19SE1857D255 ER53321EJ 332K 1/8 WATT 1% J RESISTOR 1 R20SE1857D255 ER42205QF 22K 1/4 WATT 5% RESISTOR 1 R21SE1857D255 ER770101 RESISTOR NETWORK 100K #770-101 4 RSIP1,2 RSIP3,4SE1857D255 ESPD60401 SWITCH (ITT CANON) #D6-04-01 4 SW3,SW4, SW7,SW8SE1857D255 ESPD60501 SWITCH, PUSH (D6-05-01) 4 SW1,SW2, SW5,SW6SE1857D255 EXSE3201 CRYSTAL 32.768 KHZ (EPSON) 1 X1SE1857D255 EXPX200 CERAMIC RESONATOR #PX200 1 X2SE1857D255 EIDIP8 SOCKET 8 PIN (BURNDY) 4SE1857D255 EIDIP14 SOCKET,14 PIN #DILB-14P-223T 1SE1857D255 EIDIP28 SOCKET 28 PIN #DILB-28P-223T 1SE1857D255 EIDIP40 SOCKET 40 PIN #DILB-40P-223T 1SE1857D255 0

Model SR465 Platform Installation ProcedureNOTE: PIT FRAME OUTSIDE DIMENSIONS- 3" HIGH 38" WIDE 42" LONG

1. Carefully unpackage the scale platform and inspect to ensure there has been no shipping damage.2. Prepare the pit area to accept the scale platform and readout as described in fig. 1, below. 3. Mount the readout on the wall and run the connector cable down the conduit to the pit as described in fig. 1 below.4. Remove the platform deck by using the 1/4" hex wrench (supplied) to remove the 4 mounting screws on the cover.5. Place the platform into the pit so that the connector for the readout cable is nearest to the conduit.6. Plug the readout cable into the connector on the platform.7. Position the platform in the pit so that the frame is equally spaced from all sides of the pit.8. Using a 9/16" socket wrench, adjust the 4 corner leveler bolts (A) so that the platform base is one inch (1") below the

surface of the floor on all sides. This will ensure that the deck is even with the floor surface when installed. NOTE: If youare planning on tiling or carpeting the deck be sure to compensate for the thickness of material. You may place the deckon the platform to check the level as you make your adjustments.

9. When the 4 corners are done, adjust the remaining level bolts so that they make contact with the floor of the pit thenback them off a 1/4 turn.

10. When your level adjustments are complete tighten the locking nuts on all 6 level bolts to lock them in place.11. Adjust the 6 side clearance adjustment bolts (B) so that the platform is spaced equally on all sides of the pit then tighten

the locking nuts on all six (6)bolts.12. Set the deck on to the platform and secure in place with the 4 top mounting screws.

Ifyouhaveanyquestionspleasecallthefactoryat

1(800)654-6360

Part No. MAN465 rev: 000706

8 SIDE CLEARANCEADJUSTMENT BOLTS

4 CORNERLEVEL BOLTS

Part No. MAN465 rev: 000706