Model 8600 Series Forma -86C Ultra Low Temperature Freezer Operating and Maintenance Manual 7018612 Rev. 3

Transcript of Model 8600 Series - Thermo Fisher

Page 1: Model 8600 Series - Thermo Fisher

Model 8600 SeriesForma -86C Ultra Low Temperature FreezerOperating and Maintenance Manual7018612 Rev. 3

Page 2: Model 8600 Series - Thermo Fisher

Thermo Scientific

MANUAL NUMBER 7018612

3 25693/FR-2080 7/29/09 Chg’d drier from 209016 to 209017 (refrig schematics), -201 drawing ccs

3 FR-2055 7/29/09 Added “common to” to refrigeration schematics ccs

2 25411/FR-2040 4/30/09 Changed drier from 209020 to 209016 (refrigeration schematics) ccs

1 24567/FR-2000,24605/ 5/23/08 Clarified chamber probe and thermocouple callouts, Important info above, ccs

FR-2000, 24417/SI-9962 updated exploded parts drawing (-201-1) to current production, removed 8628,

24191 software updated - added ‘Test BUS Operation’ and Micro Board Failure Error

0 23977/FR-1962 7/13/07 Release 4 - Foamed-in-place (heat exchanger) ccs

REV ECR/ECN DATE DESCRIPTION BY

Preface

Model 8600 Series i

Models Capacity (cu. ft. ) Voltage

8612 12.7 230

8615 12.7 120

8618 17 120

8619 17 230

8620 20 230

8621 20 120

Covered in this manual:

Part Number Description Qty

34040 Key Ring 1

213F Key 2

380520 Neoprene Cap 2

510016 1/4-20 x 5-1/2” Bolt 2

402058 Vacu-Key 1

195763 Retaining Clip 1

370563 Remote Alarm Connector 1

Packing List

Important installer and user information:A redundant temperature sensing device has been included in this ULTfreezer. This device is a type “T” thermocouple. For convenient access,the thermocouple (Figure 1-3) terminates in an interconnect jack (Figure1-5) behind the base front cover. (May be located differently in chests.See Section 1.) It is strongly recommended that this thermocouple beattached to a redundant 24 hour 7 day monitoring system with alarmcapabilities. Connecting the sensor to a monitoring and alarm systemseparate from the freezer provides the utmost in product safety, shouldthe integral system fail. s

Page 3: Model 8600 Series - Thermo Fisher

Thermo Scientificii Model 8600 Series

Preface

Important Read this instruction manual. Failure to read, understand andfollow the instructions in this manual may result in damage to the unit,injury to operating personnel, and poor equipment performance. s

Caution All internal adjustments and maintenance must be performed byqualified service personnel. s

Material in this manual is for information purposes only. The contents andthe product it describes are subject to change without notice. ThermoScientific makes no representations or warranties with respect to thismanual. In no event shall Thermo be held liable for any damages, direct orincidental, arising out of or related to the use of this manual.

©2007 Thermo Scientific. All rights reserved.

Page 4: Model 8600 Series - Thermo Fisher

Thermo Scientific Model 8600 Series iii

Preface

Important operating and/or maintenance instructions. Read the accompanying text carefully.

Potential electrical hazards. Only qualified persons should perform procedures associated with thissymbol.

Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.

Hot surface(s) present which may cause burns to unprotected skin, or to materials which may bedamaged by elevated temperatures.

Marking of electrical and electronic equipment, which applies to electrical and electronic equipmentfalling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the marketafter 13 August 2005.

This product is required to comply with the European Union’s Waste Electrical & ElectronicEquipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Scientifichas contracted with one or more recycling/disposal companies in each EU Member State EuropeanCountry, and this product should be disposed of or recycled through them. Further information onThermo’s compliance with this directive, the recyclers in your country and information on Thermoproducts will be available at www.thermo.com.

4 Always use the proper protective equipment (clothing, gloves, goggles, etc.)

4 Always dissipate extreme cold or heat and wear protective clothing.

4 Always follow good hygiene practices.

4 Each individual is responsible for his or her own safety.

Page 5: Model 8600 Series - Thermo Fisher

Thermo Scientificiv Model 8600 Series

Preface

Do You Need Information or Assistance on

Thermo Scientific Products?

If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:

1-740-373-4763 Direct

1-866-984-3766 Toll Free, U.S. and Canada

1-740-373-4189 FAX

http://www.thermo.com Internet Worldwide Web Home Page

Service E-Mail Address

Thermo Scientific

Controlled Environment Equipment

401 Millcreek Road, Box 649

Marietta, OH 45750

Our staff can provide information on pricing and give you quotations. We can

take your order and provide delivery information on major equipment items or make

arrangements to have your local sales representative contact you. Our products are listed on the

Internet and we can be contacted through our Internet home page.

Our staff can supply technical information about proper setup, operation or

troubleshooting of your equipment. We can fill your needs for spare or replacement parts or

provide you with on-site service. We can also provide you with a quotation on our Extended

Warranty for your Thermo Scientific products.

Whatever Thermo Scientific products you need or use, we will be happy to discuss your

applications. If you are experiencing technical problems, working together, we will help you

locate the problem and, chances are, correct it yourself...over the telephone without a service

call.

When more extensive service is necessary, we will assist you with direct factory trained

technicians or a qualified service organization for on-the-spot repair. If your service need is

covered by the warranty, we will arrange for the unit to be repaired at our expense and to your

satisfaction.

Regardless of your needs, our professional telephone technicians are available to assist you

Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by

telephone or fax. If you wish to write, our mailing address is:

International customers, please contact your local Thermo Scientific distributor.

Sales Support

Service Support

[email protected]

Page 6: Model 8600 Series - Thermo Fisher

Model 8600 Series vThermo Scientific

Table of Contents

Installation and Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1Freezer Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1Control Panel Keys and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3Keypad Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4Freezer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5

Choose Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5Install Wall Bumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5RS-232 Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6Remote Alarm Contacts and Analog Output . . . . . . . . . . . . . . . . .1-7Attach Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7Connect Unit to Electrical Power . . . . . . . . . . . . . . . . . . . . . . . . . .1-8

Freezer Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8Set Operating Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9Set High Temperature Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9Set Low Temperature Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10

Run Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1Calibrate the Control Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1High Alarm Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1Low Alarm Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1System Battery Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2Bus Battery Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2Display Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3Clear High Stage Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3Set an Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3RS485 Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4Back-Up System Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4

Section 1

Section 2

Section 3

Page 7: Model 8600 Series - Thermo Fisher

vi Model 8600 Series Thermo Scientific

Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1Wrong Power Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1High Stage System Failure Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Voltage Compensation Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Multiple Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Micro Board Failure Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Probe Failure Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Lost Communication Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1Clean Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1Clean Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1

Clean Water-cooled Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2Defrost Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3Clean Lid Gasket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3Replace Battery(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4Prepare Unit for Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5

Factory Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1BUS (Back Up System) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1

Injection Assembly Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1Temperature Probe Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2Connect Probe/ Solenoid Harness . . . . . . . . . . . . . . . . . . . . . . . . . .6-3BUS Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4

Set Optional BUS SetPoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4Test the BUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5Disconnect BUS From Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . .6-5Chart Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5Datalogger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7Water-cooled Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1Refrigeration Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1Electrical Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1Warranty Handling Liquid NitrogenHandling Liquid CO2

Table of Contents

Section 4

Section 5

Section 6

Section 9

Section 11Appendix

Section 7Section 8

Section 10

Page 8: Model 8600 Series - Thermo Fisher

Model 8600 Series 1-1Thermo Scientific

Section 1 Installation and Start-Up

Figure 1-1• Control Panel - keypad, displays and indicators• BUS (Optional Back Up System) panel• Optional temperature recorder - 7 day, one pen or datalogger• Keylock - keyed lid lock

Figure 1-2• Remote alarm contacts and selectable analog output connection -

0-1V, 4-20mA (default), 0-5V• Power inlet for power cord connection• Optional BUS connections for probe and solenoid• RS-232 or RS-485 interface• Power Switch (mains disconnect)

Figure 1-1. Front View

Figure 1-2. Rear View

Freezer Components

Page 9: Model 8600 Series - Thermo Fisher

1-2 Model 8600 Series Thermo Scientific

Section 1Installation and Start-Up

aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa

C F

%RH

Alarm

Sensor

To remove

filter

Thermocouple

interconnect jack

O

Battery

Switch

Condenser

Figure 1-3. Probe Cover

Figure 1-3• Houses the control, optional recorder, datalogger, BUS, or 1535 alarm

probes.

Figures 1-4 and 1-5• Freezer Filter Location• Battery Switch• Thermocouple

Receptacle• Battery Mounting

Bracket• Freezer Bqattery• Optional BUS Battery

Figure 1-4. Receptacle, Filter Battery Switch and Condenser Locations

Figure 1-5. Left Side, Sidecar Removed

Page 10: Model 8600 Series - Thermo Fisher

Model 8600 Series 1-3Thermo Scientific

Section 1Installation and Start-Up

Mode Select Switch - Used to select Run, Settings, Calibrate and SystemConfiguration Modes.

Mode Select Indicators -Run: Run MenuSettings: Set Points MenuCalibrate: Calibrate MenuConfiguration: Configuration Menu

Temperature Display - Displays temperature in degrees Celsius.Alarm Indicator - Light pulses on/off during an alarm condition. Silence - Silences the audible alarm. Low Battery - indicates a low battery condition of the freezer battery.Hot Condenser - indicates a hot condenser condition.Message Center - displays system status and alarms. Scroll for Parameters Arrows - moves through the choices of the selectedmode.Up and Down Arrows - Increases or decreases values, toggles betweenchoices. Enter - Stores the value into computer memory.

Control Panel Keysand Indicators

Figure 1-6. Control Panel

Page 11: Model 8600 Series - Thermo Fisher

The Model 8600 Series freezer has four basic modes which allow freezersetup: Run, Settings, Calibrate and Configuration.

Run is the default mode for the freezer during normal operation. Settings is used to enter system set points for freezer operation. Calibrate is used to calibrate various system parameters.Configuration allows for custom setup of various options.

The chart below shows the selections under each of the modes.

Scroll for Parameters Arrows: Steps the operator through the parametersof Settings, Calibrate and Configuration Modes. The right arrow goes tothe next parameter, the left arrow returns to the previous parameter.

Up Arrow: Increases or toggles the parameter value that has been selectedin the Settings, Calibrate, and Configuration Modes.

Enter: Must press Enter key to save to memory all changed values.Down Arrow: Decreases or toggles the parameter values that have been

selected in the Settings, Calibrate, and Configuration Modes.Silence Key: Press to silence the audible alarm. See Section 4 for alarm

ringback times. Message Center: Displays the system status (Mode) at all times. Displays

SYSTEM OK during normal operation, or alarm messages if the systemdetects an alarm condition. See Section 4 - Alarms.

Caution If tipped more than 45°, allow the unit to sit upright for 24 hoursbefore start up. s

1-4 Model 8600 Series Thermo Scientific

Section 1Installation and Start-Up

Keypad Operation

Run Settings Calibrate Configuration

Default ModeSYSTEM OK

Control Set Point Control Probe High Alarm Test

LINE VOLTAGE High Alarm Set Point Optional Sample Probe Low Alarm Test

COMPENSATED VOLTAGE Low Alarm Set Point System Battery Test

HSHX TEMPERATURE Optional Back Up System Set Point BUS Battery Test

Display Temperature

Clear High Stage Alarm

Set Access Code

RS485 Address

BUS type CO2 or LN2

Page 12: Model 8600 Series - Thermo Fisher

To remove the freezer from the pallet, use the 7/16" wrench to remove allthe bolts securing the shipping bracket to the pallet.

Remove the shipping bracket. Remove the ramp boards from the palletand place the slotted end over the ramp brackets on the pallet. Thesupport blocks on the ramps will be facing down. Before moving thefreezer, make sure the casters are unlocked and moving freely. Align thecaster with the ramp boards. Use adequate personnel to roll the freezer offthe pallet.

The freezer can be easily pushed to the desired approved location,described in Section 1. If necessary, the doors and lower front panel maybe opened to move the unit through tight openings. When the freezer is inposition, set the front caster brakes.

Caution The freezer must not be moved with the product load inside. s

Locate the freezer on a firm, level surface in an area with an ambienttemperature between 18°C and 32°C. Provide ample room to reach themains disconnect switch (power switch) located on the rear of the freezer.

Caution For proper ventilation and airflow, a minimum clearance of 5” atthe rear and front and a clearance of 8” on the side of the freezer isrequired. Allow adequate space for lid opening. If ambient increases above36°C, clearance at the rear of the cabinet must be increased to 8”. s

The parts bag, located inside the cabinet, contains the following parts.

Install the bolts into the pre-tapped holes on the back of the compressorsection. Install a neoprene cap on each bolt. Refer to Figure 1-2 for thelocations of the pre-tapped holes.

Model 8600 Series 1-5Thermo Scientific

Section 1Installation and Start-Up

Freezer Installation

Choose Location

Install Wall BumpersQuantity Stock # Description Purpose

2 510016 1/4-20 x 5-1/2” Bolt Wall bumper

2 380520 Neoprene cap Cap protector

Page 13: Model 8600 Series - Thermo Fisher

The Model 8600 Series freezer has a data communications interface. Thefactory default setting is RS-232.

The wiring identification for the interfaceis shown in Figure 1-7. One nine pin, sub"D" style connector is located on the backof the freezer. See Figure 1-2 for thelocation of the connector on the freezer.

The freezer transmits temperature information every 60 minutes. Astandard DB9 serial extension cable can be used to connect the freezer to aserial device. Some serial devices may require a null modem adapter.

Data format:Baud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1200Data bits . . . . . .8 (7 bit ASCII with leading zero)Start bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1Stop bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2Parity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .none

The data transfer sequence is transmitted in the following format. X refersto numerical temperature data.

(NUL) (-) XXX (SP) C (SP) (OVERTEMP) (SP) (LF) (CR) (EOT) (SP)(UNDER TEMP)

The words "OVER TEMP" or "UNDER TEMP" are transmitted whenan alarm condition exists along with the temperature. If no alarmcondition exists, spaces will be sent. A total of 20 characters will be sent.

SP - SpaceLF - Line feedCR - Carriage returnEOT - End of text (4)NUL - Null character (00)

Note The RS232 is not compatible with Model 1535 Monitor/AlarmSystem. s

1-6 Model 8600 Series Thermo Scientific

Section 1Installation and Start-Up

RS-232 Communications

Figure 1-7. Pins

Page 14: Model 8600 Series - Thermo Fisher

The Model 8600 Series freezer has remote alarm contacts and analogoutput. See Figure 1-2 for the location of the remote alarm contacts. Theremote alarm connector is located in the parts bag provided with themanual. It must be installed if connecting the freezer to an alarm system.After installing the wiring from the alarm system to the connector, installthe connector to the freezermicroboard and secure with thetwo screws provided. The remotealarm provides a NO (normallyopen) output, a NC (normallyclosed) output and COM(common). The contacts will tripon a power outage, hightemperature alarm or lowtemperature alarm. Figure 1-8shows the remote contacts in alarmstate.

The analog output function allows the freezer to output signalsrepresenting the temperature of the freezer cabinet. The factory defaultsetting is 4-20 mA. Refer to the chart below for output specifications.

Insert the power cord into thepower inlet module. Place theretaining bracket (P/N 195763)over the connector. Tightenretaining screws to secure.

Model 8600 Series 1-7Thermo Scientific

Section 1Installation and Start-Up

Remote Alarm Contacts andAnalog Output

Figure 1-8. Alarm State

4-20mA 0-1V 0-5V

Temperature -100 to +50°C -100 to +50°C -100 to +50°C

Attach Power Cord

IMPORTANT USER INFORMATIONIMPORTANT USER INFORMATION

�������� ���� ����� ������ �� ��������� � �������� �� ������� ��������������� ���� ���� �� �������� �� ��������������� ��� �������� �� �� ����������� ���������� ��������� ������ ��!���������� �����

Figure 1-10. Power Cord Connection

Page 15: Model 8600 Series - Thermo Fisher

Connect Unit to ElectricalPower

Caution See the serial tag on the side of the unit for electrical specificationsor refer to the electrical schematics in this manual. s

The freezer should be operated on a dedicated grounded service. Check thevoltage rating on the serial tag of the unit and compare it with the outletvoltage. Then, with the power switch turned off, plug the line cord intothe wall outlet.

First, turn on the freezer power switch. Then open the lower front door bygrasping the bottom left corner. Locate the battery switch (Figure 1-4) andturn it to Standby mode ( ). During initial freezer start-up, the systembattery may require charging and the Low Battery message may appear inthe message center.

Caution Verify that the battery switch is turned to Standby mode ( ). Therechargeable batteries require 36 hours to charge at initial start-up. A “LowBattery” alarm may occur until the batteries are fully charged. Should apower failure occur during the initial start-up period, the electronics willhave limited operation. s

With the freezer properly installed and connected to power, system setpoints can be entered. The following set points can be entered in Settingsmode: Control temperature, high temperature alarm set point, lowtemperature alarm set point, and (optional) BUS set point. Default settingsare shown in the table below. See chart below for more detail.

1-8 Model 8600 Series Thermo Scientific

Section 1Installation and Start-Up

Freezer Start-Up

Control Set Point -80°C

High Temperature Alarm -70°C

Low Temperature Alarm -90°C

Optional BUS Set Point -60°C

Page 16: Model 8600 Series - Thermo Fisher

All Model 8600 Series freezers have an operating temperature range of-50°C to -86°C, depending on ambient temperature. The freezer is shippedfrom the factory with a temperature set point of -80°C. To change theoperating temperature set point:

1. Press the Mode key until the Settings indicator lights.

2. Press right arrow until “SET PT = -XX” displays in message center.

3. Press the up/down arrow key until the desired temperature set point isdisplayed.

4. Press Enter to save the set point.

5. Press the Mode key until the Run indicator lights for Run mode orpress the right/left arrow keys to go to next/previous parameter.

If no control keys are pressed, the freezer will automatically return to Runmode after 5 minutes.

Note If the set point is changed and the low temperature and hightemperature alarms are set 10° from the set point, the alarm set points willbe adjusted automatically to maintain a distance of at least 10° from setpoint. s

The high temperature alarm will activate an audible/visual warning whenthe freezer chamber temperature has reached or exceeded the hightemperature alarm set point. To set the high temperature alarm set point:

1. Press the Mode key until the Set indicator lights.

2. Press right arrow until “HI ALM = -XX” displays in message center.

3. Press the up or down arrow key until the desired high temperaturealarm set point is displayed.

4. Press Enter to save the setting.

5. Press the Mode key until the Run indicator lights or press the right orleft arrow to go to the next or previous parameter.

If no control keys are pressed, the freezer automatically returns to RUNmode after 5 minutes.

Note The high alarm set point must be set at least 5°C from the controlset point. s

Note At initial start-up, the high temperature alarm is disabled until thecabinet reaches set point or 12 hours elapse. s

Model 8600 Series 1-9Thermo Scientific

Section 1Installation and Start-Up

Set OperatingTemperature

Set High TemperatureAlarm

Page 17: Model 8600 Series - Thermo Fisher

The low temperature alarm will activate an audible/visual warning whenthe freezer chamber temperature has reached or decreased below the lowtemperature alarm set point. To set the low temperature alarm set point:

1. Press the Mode key until the Settings indicator lights.

2. Press the right arrow until “LO ALM = -XX” is displayed in themessage center.

3. Press the up or down arrow key until the desired low temperaturealarm set point is displayed.

4. Press Enter to save the setting.

5. Press the Mode key until the Run indicator lights or press the right orleft arrow to go to the next or previous parameter.

If no control keys are pressed, the freezer will automatically return to RUNmode after 5 minutes.

Note The low alarm set point must be set at least 5°C from the control setpoint. s

An access code can be set to prevent unauthorized change of settings inCalibrate, Configuration and Settings mode. (An access code of 000 isrequired to make changes.) If the access code is not at the default 000, youcan not leave RUN mode without entering a code. See Section 3 forinstructions on modifying the access code.

The Run mode is the default mode for the freezer. The run mode willdisplay the cabinet temperature on the temperature display and ‘SYSTEMOK’ on the message center under normal operating conditions. Inaddition, the Run mode allows display of the following information:

LINE VOLTAGECOMPENSATED VOLTAGEHSHX TEMPERATURE (heat exchanger temperature)

This information is scrolled individually by pressing the right arrow key. Ineach case, the message center returns to SYSTEM OK in 10 seconds if nokeys are pressed.

1-10 Model 8600 Series Thermo Scientific

Section 1Installation and Start-Up

Set Low TemperatureAlarm

Access Code

Run Mode

Page 18: Model 8600 Series - Thermo Fisher

Model 8600 Series 1-11Thermo Scientific

Section 1Installation and Start-Up

Press MODEto light SETTINGS

indicator

Mode

Mode

SET PT = -XX

Numbers increasePress MODE to moveto CALIBRATEmode

Numbers decrease

Press Enter

to save settingScroll for Parameters

Scroll for Parameters

Mode HI ALM = -XXScroll for Parameters

Press to returnto previous parameter

Press to returnto previous parameter

Press to returnto previous parameter

Low AlarmSet Point

High AlarmSet Point

ControlSet Point

To Set:

Settings Mode

Numbers increase

Numbers decrease

Press Enter

to save setting

Numbers increase

Numbers decrease

Press Enter

to save setting

Mode

Press to returnto previous parameter

OptionalBack Up System

Set Point

Numbers increase

Numbers decrease

Press Enter

to save setting

������ ��

LO ALM = -XX

BACKUP = -XX

Page 19: Model 8600 Series - Thermo Fisher
Page 20: Model 8600 Series - Thermo Fisher

Model 8600 Series 2-1Thermo Scientific

Section 2 Calibration

Once the freezer has stabilized, the control probe may need to becalibrated. Calibration frequency is dependent on use, ambient conditionsand accuracy required. A good laboratory practice would require at least anannual calibration check. On new installations, all parameters should bechecked after the stabilization period.

Caution Before making any calibration or adjustments to the unit, it isimperative that all reference instruments be properly calibrated. s

Plug a type T thermocouple reader into the receptacle located inside thelower door (Figure 1-5). Compare the control temperature set point to thetemperature of the measuring device. See Chart 2-1 at the end of thissection for more detail.

1. Press the Mode key until the Calibrate indicator lights.

2. Press the right arrow until “CONT T = -XX.X” appears in the messagecenter.

3. Press up/down arrow to match the display to calibrated instrument.

4. Press Enter to store calibration.

5. Press the Mode key to return to Run or the right/left arrow to go tonext/previous parameter.

See Chart 2-1 for calibration process functions.

Temperature Stabilization Periods

Startup - Allow 12 hours for the temperature in the cabinet to stabilizebefore proceeding.

Already Operating - Allow at least 2 hours after the display reaches setpoint for temperature to stabilize before proceeding.

Note During calibration, the temperature display will not be available. s

If no keys are pressed for approximately 5 minutes while in Calibrationmode, the system will reset to Run mode.

Calibrate the ControlProbe

Page 21: Model 8600 Series - Thermo Fisher

2-2 Model 8600 Series Thermo Scientific

Section 2Calibration

Press MODEto light CALIBRATE

indicator

Mode CONT T = -XX.X

Numbers increasePress MODE to moveto CONFIGURATIONmode

Numbers decrease

Press Enter

to save settingScroll for Parameters

Mode SAMPT = -XX.XScroll for Parameters

Press to returnto previous parameter

Press to returnto previous parameter

OptionalSample Probe

ControlProbe

To Calibrate:

Calibrate Mode

Numbers increase

Numbers decrease

Press Enter

to save setting

Chart 2-1

SYSTEM OK

Page 22: Model 8600 Series - Thermo Fisher

Model 8600 Series 3-1Thermo Scientific

Section 3 Configuration

The Configuration Mode is used for testing and custom setup of thefreezer. The configuration functions listed and described below may notbe necessary in all applications, but are available if needed. See Chart 3-1for more detail.

The high alarm test is used to verify the high alarm will activate, shouldthe freezer temperature equal or exceed the high alarm set point.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until HI ALRM TEST is displayed in the messagecenter.

3. Press Enter to initiate the test.

The temperature on the display will begin to increase until the high alarmset point has been reached. The audible alarm sounds and the alarmindicator flashes. Press the Silence key to silence the alarm.

The low alarm test is used to verify the low alarm will activate, should thefreezer temperature equal or become less than the low alarm set point.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until LO ALRM TEST is displayed in themessage center.

3. Press Enter to initiate the test.

The temperature on the display will begin to decrease until the low alarmset point has been reached. The audible alarm sounds and the alarmindicator flashes. Press the Silence key to silence the alarm.

High Alarm Test

Low Alarm Test

Page 23: Model 8600 Series - Thermo Fisher

3-2 Model 8600 Series Thermo Scientific

Section 3Configuration

System Battery Test To test the charge of the freezer battery:

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until SYS BAT TEST is displayed in the messagecenter.

3. Press Enter to initiate the test.

TESTING BATT will display during the testing period. Upon completionof the test, the message center will display BATT GOOD or BATT FAILOn test failure, the audible alarm sounds, the alarm indicator and the LowBattery indicator light. Press the Silence key and the alarm indicator willgo out. The Low Battery light will stay on until a future battery test isperformed and passed.

To test the charge of the BUS battery:

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until BUS BAT TEST is displayed in the messagecenter.

3. Press Enter to initiate the test.

TESTING BATT will display during the testing period. Upon completionof the test, the message center will display BBAT GOOD or BBAT FAILOn test failure, the audible alarm sounds, the alarm indicator and the LowBattery indicator light. Press the Silence key. The audible alarm and alarmindicator will go out. The Low Battery light will stay on. If this test fails, itis recommended to replace the BUS battery.

Bus Battery Test

Page 24: Model 8600 Series - Thermo Fisher

Model 8600 Series 3-3Thermo Scientific

Section 3Configuration

This function, available on freezers with the optional sample probe, allowsthe user to select which probe temperature is displayed in the temperaturedisplay window. The options are CONTROL or SAMPLE.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until DISP CONTROL or DISP SAMPLE isdisplayed in the message center.

3. Press up/down arrow to toggle between the two display selections.

4. Press Enter to save.

If Control probe is selected, the temperature display is on continuously. IfSample probe is selected, the temperature display is preceded with an ‘S’.

Should a high stage alarm occurr, it may become necessary to clear thealarm condition after the condition has been corrected.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until CLR HS ALARM is displayed in themessage center.

3. Press Enter to clear the alarm.

To set the Access Code:

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until “SET ACC CODE” is displayed in themessage center.

3. Press Enter.

4. The message center will display ACC CODE = 000. Press theup/down arrow until the desired access code is displayed (000 - 999).Press the left/right arrow to select digit 1, 2, 3.

Note Use the left and right arrow keys to move from the first through thethird digits within the access code. s

5. Press Enter to save the setting

6. Press the Mode key until Run indicator lights. A three-digit AccessCode can be entered to avoid unauthorized personnel changing setpoints, calibration, or configuration. A setting of 000 bypasses theaccess code. The factory setting is 000.

Display Temperature

Clear High StageAlarm

Set an Access Code

Page 25: Model 8600 Series - Thermo Fisher

If the freezer is configured for RS-485 communications, it will need tohave a unique identification address. This address is set through theConfiguration mode.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until RS485ADDR is displayed in the messagecenter.

3. Press Enter. The message center will display 485 ADDR XX.

4. Press up/down arrow to select the appropriate address for the freezer(1 - 24).

4. Press Enter to save.

This function, available on freezers with the optional BUS (back-upsystem), allows the user to select the type of gas injected into thefreezer chamber. The options are CO2 and LN2.

1. Press the Mode key until Configuration indicator lights.

2. Press the right arrow until BUS TYPE CO2 or BUS TYPE LN2 isdisplayed in the message center.

3. Press up/down arrow to toggle between the two display selections.

4. Press Enter to save.

3-4 Model 8600 Series Thermo Scientific

Section 3Configuration

RS485 Address

Back-Up System Type

Page 26: Model 8600 Series - Thermo Fisher

Model 8600 Series 3-5Thermo Scientific

Section 3Configuration

Press MODE

to light

CONFIGURATION

indicator

Mode

Mode SYS BAT TEST

LO ALRM TEST

Press MODE to move

to RUN modePress Enter

to initiate

test

Scroll for Parameters

Scroll for Parameters

Mode

HI ALRM TEST

Scroll for Parameters

Press to return

to previous parameter

Press to return

to previous parameter

Press to return

to previous parameter

Low Alarm

Test

High Alarm

Test

System Battery

Test

To Configure:

Mode

Scroll for Parameters

Chart 3-1

Press Enter

to initiate

test

Press Enter

to initiate

test

Press Enter

to initiate

test

Press to return

to previous parameter

Optional

BUS Battery

Test

BUS BAT TESTPress Enter

to initiate

test

Configuration Mode

SYSTEM OK

Page 27: Model 8600 Series - Thermo Fisher

3-6 Model 8600 Series Thermo Scientific

Section 3Configuration

Press Mode to exit.

Press right or left arrow

to move to next digit.

Numbers increase

Numbers decrease

Mode CLR HS ALARM

Press Enter to

clear the alarm

conditionScroll for Parameters

Press to return

to previous parameter

refer to previous page

Configuration Mode, Chart 3-1, Page 2 of 2

Clear High

Stage Alarm

To Configure:

DISP CONTROL

CONTROL probe

Press Enter to

save the settingMode

Scroll for Parameters

Press to return

to previous parameter

Display

Temperature SAMPLE probe

Mode

Scroll for Parameters

ModePress Enter to

save the settingScroll for Parameters

Mode BUS TYPE CO2 Press Enter to

save the settingScroll for Parameters

Press to return

to previous parameter

Optional

Back Up

System Type

Mode RS485 ADDR

Scroll through the

various options.

Press Enter to

save the setting

Scroll for Parameters

Press to return

to previous parameter

RS-485

Address

Type CO2

Type LN2

Access

Code

SET ACC CODE XXX

Page 28: Model 8600 Series - Thermo Fisher

Model 8600 Series 4-1Thermo Scientific

Section 4 Alarms

The Model 8600 Series freezer alarm system is shown in the table below.When an alarm is active, the message appears in the LED message center.Press the Silence key to mute the audible alarm for the ringback period.The visual alarm continue until the freezer returns to a normal condition.The alarms are momentary alarms only. When an alarm condition occurs,then returns to normal, the freezer automatically clears the alarm conditionand the message center.

If a 230V freezer is connected to a 120V power source or a 120V freezer isconnected to a 230V power source, the electronics detects an incorrectpower source connected to the freezer. Under this condition, the fans andcompressors will not turn on and an audible and visual alarm occur with"WRONG POWER" in the message center.

The “WRONG POWER” alarm may also occur if the battery switch isturned to Standby mode ( ) prior to applying power to the freezer. Theaudible and visual alarms remain until the freezer is connected to thecorrect power source.

Description Message Delay Ringback Relay

No alarm condition exists SYSTEM OK ---- ---- ----

Power Failure POWER FAIL 1 min. 15 min. Yes

High Temperature Alarm TEMP IS HIGH 1 min. 15 min. Yes

Low Temperature Alarm TEMP IS LOW 1 min. 15 min. Yes

Door Ajar DOOR IS OPEN 1 min. 15 min. No

Low Battery* LOW BATTERY 1 min. 12 hours No

Control Probe Failure PROBE 1 FAIL 1 min. 15 min. No

Heat Exchanger Probe Failure PROBE 2 FAIL 1 min. 15 min. No

Condenser Probe PROBE 3 FAIL 1 min. 15 min. No

Sample Probe Failure (optional) PROBE 4 FAIL 1 min. 15 min. No

High Stage System Failure HS SYST FAIL 1 min. 15 min. No

Condenser Hot Condition HOT CONDENSR 1 min. none No

Wrong Power WRONG POWER 0 min. none No

Voltage Compensation VCOMPEN FAIL 0 min. 15 min. No

Micro Board Failure MICRO FAIL 0 min. 15 min. No

All alarm delays and ringback times are ±30 seconds.*The automatic battery test runs 12 hours after initial start-up, then every 12 hours thereafter. A user-initiated batterytest can be performed from the Configuration menu. See Section 3.

Wrong Power Alarm

Page 29: Model 8600 Series - Thermo Fisher

4-2 Model 8600 Series Thermo Scientific

Section 4Alarms

High Stage SystemFailure Alarm

This condition is created when the high stage compressor and fans run for30 minutes and are not capable of cooling the interstage heat exchanger tothe proper temperature. Under this condition, the high stage compressorand fans will turn off after 30 minutes and an audible and visual alarm willoccur along with the "HS SYST FAIL" message in the LED messagecenter.

If the freezer is compensating for high or low line voltage, the system willmeasure the compensated AC voltage. If the voltage is incorrect, the unitwill stop attempting to compensate, and the compressors will run on directline voltage. Under this condition, the message center will display“VCOMPEN FAIL”.

When multiple alarm conditions occur, active messages are displayed in themessage center one at a time, updating at 5 second intervals. PressingSilence during multiple alarms causes all active alarms to be silenced and toring back in 15 minutes.

An internal communications failure has occurred with the micro board.During this alarm, the compressor(s) attempt to run continuously.However, with this type of failure, freezer operation becomesundependable.

The microprocessor in Model 8600 Series freezers continually scans allprobes including the control probe, heat exchanger probe, condenser probeand optional sample probe to ensure that they are operating properly.Should an error be detected, the "PROBE # FAIL" alarm occurs. If anerror is detected with the control probe (PROBE 1 FAIL), the high andlow stage compressors runs continuously. As a result, the cabinettemperature decreases until it reaches the lowest temperature that therefrigeration system can maintain. If an error is detected with the heatexchanger probe (PROBE 2 FAIL), the freezer cycles properly at itstemperature set point using a 5 minute step start between the high and lowstage compressors. If an error is detected with the condenser probe(PROBE 3 FAIL) or optional sample probe (PROBE 4 FAIL), there is noimpact on the performance of the freezer. However, the hot condenseralarm may also occur when the condenser probe fails. Contact theTechnical Services Department (1-888-213-1790) or your local distributor.

Communication between the micro board and display has been lost. Underthis condition, the visual alarm flashes with dashes (----) in the temperaturedisplay. “Lost Comm’ flashes in the message center. Call Technical Services.

VoltageCompensation Alarm

Multiple Alarms

Micro Board FailureAlarm

Probe Failure Alarms

Lost CommunicationAlarm

Page 30: Model 8600 Series - Thermo Fisher

Model 8600 Series 5-1Thermo Scientific

Section 5 Maintenance

Warning Avoid the excessive use of water around the control area due tothe risk of electrical shock. Damage to the controls may also result. s

Wipe down the freezer exterior using soap and water and a general uselaboratory disinfectant. Rinse thoroughly with clean water and dry with asoft cloth.

1. Open the door on the front of the unit, using the handle.

2. Locate the grille on the door. See Figure 1-4. Grasp the middle of thegrille material and gently pull out to remove.

3. Wash the filter material using water and a mild detergent.

4. Dry by pressing between two towels.

5. Install the filter back into the grille and attach the grille.

* Perform this procedure a minimum of four times per year. The clean filter alarmoccurs every three months as a reminder to clean the air filter. Depending uponenvironmental conditions, the filter may need to be cleaned or replaced morefrequently. If the filter becomes torn or excessively dirty, a replacement can bepurchased from Thermo. Order part number 760212.

1. Open the door on the front of the unit, using the handle.

2. Locate the condenser (Figure 1-4). Using a vacuum cleaner, exercisingcare to not damage the condenser fins, clean the condenser (minimumyearly*).

* Depending upon environmental conditions, the condenser may need to becleaned more frequently.

Clean Air Filter

Clean Condenser

Page 31: Model 8600 Series - Thermo Fisher

5-2 Model 8600 Series Thermo Scientific

Section 5Maintenance

Clean Water-cooledCondenser

The water-cooled condenser can be cleaned-in-place by using the CIPprocedure. Cleaning solutions can be used, depending on type of depositsor build-up to be removed.

Caution Do not use liquids that are corrosive to stainless steel or thebrazing material (copper or nickel). s

1. Disconnect the unit from the water supply.

2. Drain the unit.

3 . Rinse with fresh water and drain the unit again.

4. Fill with fresh water.

5. Add cleaning agent (solution and concentration dependent on depositsor build-up).

6. Circulate cleaning solution (if feasible).

7. Drain the cleaning solution.

8. Add and circulate a passivating liquid for corrosion inhibition of platesurfaces.

9. Drain this liquid.

10. Rinse with fresh water and drain.

11. Reconnect the water supply and fill the unit.

12 . Return to service.

CIP (Clean-In-Place) Procedure

Page 32: Model 8600 Series - Thermo Fisher

1. Remove all product and place it in another freezer.

2. Turn the unit off and disconnect it from the power source.

3. Turn off the battery switch(O). See Figure 5-1.

4. Open the lid and removesub-lids. Place towels on thechamber floor.

5. Allow the frost to melt andbecome loose.

6. Remove the frost with a softcloth.

7. After defrosting is complete, clean the interior with a non-chloridedetergent. Rinse thoroughly with clean water and dry with a soft cloth.

8. Plug unit in and turn power switch on.

9. Turn the battery power switch to Standby mode ( ).

10. Allow the freezer to operate empty overnight before reloading theproduct.

Using a soft cloth, remove any frost build-up from the gasket, sub-lids andlids (minimum monthly*). The clean gasket alarm occurs every threemonths as a reminder to remove frost build-up from the gasket and doors.Press the Silence key to mute the audible alarm.

*The lid gasket may need to be cleaned more frequently if dirt or excessive frostbuild-up prevents door from closing properly.

Model 8600 Series 5-3Thermo Scientific

Section 5Maintenance

Defrost Chamber

Figure 5-1. Locations

Clean Lid Gasket

Page 33: Model 8600 Series - Thermo Fisher

Caution For a consistent and dependable charge, replace the battery every2 years. Replacement batteries must be rechargeable and are available fromThermo. Refer to the parts list for stock number and description of thereplacement batteries (part number 400159). Dispose of the used batteriesin a safe manner and in accordance with good environmental practices. s

1. Open the front door by grasping the handle and pulling.

2. Locate the battery power switch (Figure 5-1). Turn the battery powerswitch to the Off position (O).

3. Remove the four screws holding the recorder bezel to gain access to thebattery.

4. Remove the three nuts securing the battery bracket. See Figure 5-1.

5. Remove the bracket and old battery. Discard the battery. Install thebracket and the new battery. Secure.

6. Reconnect the battery (red to positive and black to negative).

7. Replace the recorder bezel.

8. Turn the battery power switch to Standby mode ( ).

9. Close lower panel door.

Defrost the unit as described previously. This prepares the unit for storage.Turn off the battery power switch (O). Turn off the freezer power switch.

Warning If the unit has been in service, turn it off and disconnect thepower cord connector before proceeding with any maintenance. s

5-4 Model 8600 Series Thermo Scientific

Section 5Maintenance

Replace Battery(s)

Prepare Unit forStorage

Page 34: Model 8600 Series - Thermo Fisher

Model 8600 Series 5-5Thermo Scientific

Section 5Maintenance

Preventive MaintenanceModel 8600 Series Freezers

Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance isimportant to keep your unit functioning properly. The operator should perform routine cleaning andmaintenance on a regular basis. For maximum performance and efficiency, it is recommended that the unit bechecked and calibrated periodically by a qualified service technician.

The following is a condensed list of preventive maintenance requirements. See the specified section of theoperating manual for further details.

We have qualified service technicians, using NIST traceable instruments, available in many areas. For moreinformation on Preventive Maintenance or Extended Warranties, please contact us at the number below.

Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracyrequired.

Tips:• Fill an upright by starting at the bottom near the probe and add racks to one shelf at a time. Allow freezer to

recover to set point between shelves.• Fill a chest by starting at the left side near the probe. Filling with room temperature racks will result in a long

pull-down time.• Fill unit with frozen product to help overall performance; frozen water jugs, for example.• Always make certain the vacuum relief port is free of frost and ice, to allow for timely re-entry into the freezer

after a door opening.

Refer to ManualSection Action Monthly Yearly Every 2 Years

-- Verify ambient temperature, <90°F 4

-- * Adjust door handle for firm latching, as needed 4

Figure 1-4 for probe location5

Check and clean probe cover, gaskets hinges and lid(s) of ice andsnow

4

More frequent cleaningmay be required, depend-ing on use and environ-mental conditions

5. Check air filter. Clean or replace as needed 4

3 Check alarm back-up battery. 4 ** Replace

-- Check condenser fan motor for unusual motor noise or vibration 4

2 * Verify and document calibration, at the minimum, annually 4

5 * Clean condenser compartment and wipe off condenser 4

* Qualified service technicians only

** Dispose of properly, according to all state and federal regulations

Page 35: Model 8600 Series - Thermo Fisher
Page 36: Model 8600 Series - Thermo Fisher

Section 6 Factory Installed Options

Information about each of the factory installed options follow.

Before installation of BUS components, make sure the power to the freezeris disconnected, the battery switch is turned off (O) and the freezer haswarmed to ambient temperature.

The built-in BUS (back up system) will keep the freezer chambertemperature below the critical level in the event of a power or equipmentfailure. If power to the freezer fails, or temperature increases to the back upalarm set point, the BUS injects liquefied gas into the chamber to keep thechamber temperature within the specified range.

The BUS operates on an internal 12-volt, rechargeable battery which iskept charged during normal operation by the integral battery charger.

1. Locate the mounting hole forinstalling the injection teeassembly. See Figure 6-1.

Note Cover the open end ofinjection assembly with tape tokeep insulation from entering thenipple. s

2. Slide 3/8” flatwasher overopen end of nipple.

3. Insert the covered end of the injection assembly through exterior hole.

4. Remove the tape covering the end of the nipple and install the 1/8”NPT brass tee on the open end of the nipple. Place Permagum sealantbetween the brass tee and the chamber ceiling.

5. On the interior, seal around the injection assembly with Permagum.

Model 8600 Series 6-1Thermo Scientific

BUS (Back Up System)P/N 195943, 195945

Injection AssemblyInstallation

Figure 6-1. Mounting Hole Location

Page 37: Model 8600 Series - Thermo Fisher

6. Install the transfer hose connecting one end to the injection assembly,the other end to the solenoid valve. Install the solenoid valve to thesupply source. The solenoid mounting bracket is not required and maybe discarded.

Caution When selecting a CO2 supply cylinder, it must be equipped with asiphon tube. s

1. Plug the solenoid/probe connector into the BUS connection. Loop theprobe wire back into the base/side car. Secure the connector with ascrew on the ends of connector. The connector is keyed.

2. Route the temperature probe through the probe port. The probe portis located in the upper right corner (viewed from the side) of 12.7, 17and 20 cu. ft. models.

3. Carefully remove the existing Permagum sealant from around theprobe port opening.

6-2 Model 8600 Series Thermo Scientific

Section 6Factory Options

Figure 6-2. Injection Assembly

Temperature ProbeInstallation

probewires

Figure 6-3. Probe and Solenoid Wiring

Injection AssemblyInstallation (continued)

Page 38: Model 8600 Series - Thermo Fisher

Model 8600 Series 6-3Thermo Scientific

Section 6Factory Options

4. Open the freezer lid and locate the probe cover on the upper front leftwall. Remove the two Phillips head screws securing the probe cover(see Figure 6-4).

5. Route the BUS probe through the probe port (Figure 6-5),approximately 12”. Secure the back-up probe to the temperature probeusing a small tie wrap (Figure 6-4).

1. Seal around the interior and exterior opening of the probe port withPermagum sealant.

2. Reinstall the probe cover (Figure 6-5).

3. Carefully coil the extra probe lead in the compressor compartment,and secure it to the compartment wall with a tie wrap and tie wrapanchor provided. Additional tie wraps and anchors may be used tosecure the probe lead to the exterior back wall of the freezer.

4. Loosen the terminal screws on the solenoid. Slide the spade lugconnectors under the screws and tighten to secure.

5. Connect power to the freezer. Turn the freezer On, with battery switchOff (O).

a. The Solenoid Engaged light on the BUS control panel willilluminate (no injection occurs). This light stays on until the unitis below BUS setpoint.

b. The Low Battery indicator may also illuminate.

6. Turn the battery switch to Standby mode ( ) to charge both batteries.

Temperature ProbeInstallation (continued)

Connect Probe/ SolenoidHarness

Figure 6-5. Probe Access CoverFigure 6-4. Probe and Probe Cover

Page 39: Model 8600 Series - Thermo Fisher

Warning When activated, this unit injects liquid nitrogen or carbondioxide. Liquid nitrogen can cause serious freezing (frostbite) if it comesin contact with unprotected skin or eyes. Nitrogen suppresses oxygenlevels and may cause suffocation if area is not well ventilated. Refer toAppendix A for the proper handling of liquid LN2. s

Caution Make sure the pressure relief valve on any LN2 tank is adjusted to30 PSI maximum blow-off. s

Warning Carbon dioxide gas suppresses oxygen levels and may causesuffocation if area is not well ventilated. Refer to “Handling Liquid CO2 inAppendix B of this manual. s

Power - indicates the unit has AC power. Low Battery - battery charge is low. The battery needs replaced or

recharged.Solenoid Engaged - BUS has opened the solenoid so it can inject gas (CO2

or LN2).Press-To-Test - Activates the solenoid and injects LN2 or CO2 into the

freezer chamber as long as the button is depressed. The solenoid engagedindicator should light. If the Low Battery indicator lights during the test,replace the BUS battery.

Note Solenoid will not engage if lid is open. s

The optional back up system is designed to inject CO2 or LN2 into thefreezer compartment if the temperature rises above back up system setpoint. To set the BUS set point:

1. Press the Mode key until the Settings indicator lights.

2. Press the right arrow until “BACKUP = -XX” is displayed in themessage center.

3. Press up or down arrow key until desired BUS set point is displayed.

4. Press Enter to save the setting.

5. Press the Mode key until the Run indicator lights or press the right orleft arrow to go to the next or previous parameter.

If no control keys are pressed, the freezer automatically returns to RUNmode after 5 minutes.

6-4 Model 8600 Series Thermo Scientific

Section 6Factory Options

BUS Control Panel

Set Optional BUSSetPoint

Figure 6-6. BUS Control Panel

Page 40: Model 8600 Series - Thermo Fisher

After the freezer has stabilized and both batteries are fully charged, theBUS can be tested to verify proper operation.

1. Disconnect the AC power to the freezer by turning power switch off.

2. As the freezer warms up, verify the BUS injects at the desiredtemperature. Displayed temperature may vary by a few degrees frominject temperature due to the differences in probe locations.

Caution Changing the operating temperature set point can affect the BUSset point. The BUS set point will self adjust to maintain a temperature ofat least 10°C above the operating temperature set point. s

Caution The BUS set point can not be set any colder than the hightemperature alarm set point (see Section 1). If the back-up system isinstalled with CO2, then -65°C is the coldest BUS set point that can beused (if the cabinet set point is -75°C or colder). s

To disconnect the freezer back-up from the gas supply:

1. Close the supply valve.

2. Press the test button on the Back-Up System control box to remove thegas from the line.

3. Slowly disconnect the fitting assembly from the supply (in the eventthat any gas remains in the line).

Information for using the chart recorder follows.

1. Open the plastic door of the recorder and press button #3 until the penbegins to move outward.

2. Unscrew the knob at the center of the chart and remove the paper.

3. Install the new chart paper, position the paper to the correct time lineand replace the knob.

4. Remove the cap from the felt pen and press button #3.

Model 8600 Series 6-5Thermo Scientific

Section 6Factory Options

Disconnect BUS FromGas Supply

Chart Recorder

Install Chart Paper

Test the BUS

Page 41: Model 8600 Series - Thermo Fisher

The chart recorder contains eight temperature ranges and is factory-programmed for the freezer.

1. Press and hold button #3 for one second, then let the pen move off thechart paper.

2. Press and hold for five seconds either button #1 or button #2.

3. Release the button and the green LED will begin to flash. Count thenumber of flashes to determine the present program setting.

4. To change the program setting, press theleft or right arrows to increase or decreasethe count.

5. When the desired program number isflashing, press button #3 to bring the penarm back onto the chart. Recording willbegin in the new program.

Note Changing ranges may require an offset calibration. s

Note The recorder must be in service for 24 hours before performing thefollowing calibration procedure. s

1. Place an accurate thermometer in the chamber next to the recorderprobe.

2. Temperature probes for the recorder are located in the left front cornerof the freezer chamber (Figure 1-3).

3. After about three minutes, compare the thermometer reading with thechart recorder reading.

4. If an adjustment is necessary, press the #1 button to move the pen tothe left or the #2 to move the pen to the right. The button must beheld about five seconds before the pen begins to move. Release thebutton when the pen position matches the thermometer.

Note The felt-tip pen on the recorder requires periodic replacement.Usually the ink will appear to fade before replacement becomes necessary.Additional pen tips may be purchased from Thermo. s

6-6 Model 8600 Series Thermo Scientific

Section 6Factory Options

Change Recorder Range

Program From To

1 -40 30°C2 0 60°C3 -100 38°C4 -5 50°C5 0 100°C6 -100 200°C7 -115 50°C8 -10 70°C

Calibrate Recorder

Page 42: Model 8600 Series - Thermo Fisher

Dataloggers and ELPRO evaluation software provide monitoring anddocumentation of temperature and alarm conditions. The dataloggers havea memory capacity of 64,000 measured values or data points. Temperatureis measured, stored and displayed. Alarm conditions are recorded.Evaluation software permits data to be downloaded to a PC. A variety ofstatistical information is provided through calculations, analysis, graphsand printed reports. Refer to the ELPRO documentation for operatinginstructions for the datalogger.

The water-cooled condenser (P/N 195967) is a factory installed optionand requires a qualified technician at freezer installation. The installationshould include proper adjustment of the regulating valve, which controlsthe discharge pressure. Specifications are displayed in Table 6-1.

Model 8600 Series 6-7Thermo Scientific

Section 6Factory Options

Datalogger

Water-cooledCondenser

Water Source Tower City

Water Pressure Not to exceed 150 psig

Water Temperature Range Not to exceed 29.4C (85F)

Inlet Connection 0.5” compression

Outlet Connection 0.5” compression

Flow Rate Required 3.0 gallons (11.4 liters) per minute 1.0 gallon (3.8 liters) per minute

Drain Required No (return line is required) Yes

Table 6-1. Specifications for Water-cooled Condenser Option

Page 43: Model 8600 Series - Thermo Fisher
Page 44: Model 8600 Series - Thermo Fisher

Section 7 Specifications

Model 8600 Series 7-1Thermo Scientific

Model 8615 8612

Temperature Range -50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient

Exterior Dimensions72.0”W x 40.5” H x 28.8” F-B(182.9cm x 102.9cm x 73.2cm)Add 7.9” (20.1cm) to F-B for wall spacer/handle.

Interior Dimensions42.5”W x 28.0”H x 18.5” F-B(108.0cm x 71.1cm x 47.6cm)

Capacity 12.7 cu. ft. (360 liters)

Refrigeration Cascade system, (2) hermetically-sealed compressors

Insulation Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;1.0" (12.5cm) sub-lids

Electrical120VAC, 1 PH, 60 Hz, 16.0 FLAOperating Range: 108VAC-130VAC

230VAC, 1 PH, 50/60 Hz, 12.0 FLAOperating Range: 208VAC-240VAC

Breaker Requirements 20 Amp, 120VAC, Dedicated Circuit,20 Amp Time Delay Breaker

15 Amp, 230VAC, Dedicated Circuit,15 Amp Time Delay Breaker

Shipping Weight 716 lbs. ( 324.8 kg)

Model 8618 8619

Temperature Range -50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient

Exterior Dimensions87.6.0”W x 40.5” H x 28.8” F-B(222.5cm x 102.9cm x 73.2cm)Add 7.9” (20.1cm) to F-B for wall spacer/handle.

Interior Dimensions58.8”W x 28.0”H x 18.5” F-B(149.4cm x 71.1cm x 47.0cm)

Capacity 17.0 cu. ft. (481.4 liters)

Refrigeration Cascade system, (2) hermetically-sealed compressors

Insulation Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;1.0" (12.5cm) sub-lids

Electrical120VAC, 1 PH, 60 Hz, 16.0 FLAOperating Range: 108VAC-130VAC

230VAC, 1 PH, 50/60 Hz, 12.0 FLAOperating Range: 208VAC-240VAC

Breaker Requirements 20 Amp, 120VAC, Dedicated Circuit,20 Amp Time Delay Breaker

15 Amp, 230VAC, Dedicated Circuit,15 Amp Time Delay Breaker

Shipping Weight 821 lbs. ( 372.4 kg)

Page 45: Model 8600 Series - Thermo Fisher

CertificationsDeclaration of Conformity is available from the factory

Safety SpecificationsIndoor Use OnlyAltitude - up to 2,000 metersTemperature - 5°C to 43°CHumidity - Maximum RH 80% for temperatures up to 31°C, decreasing lin-early to 50% RH at 40°CMains Supply Fluctuations - Mains supply voltage fluctuations not to exceed±10% of the nominal voltageInstallation Category II 1

Pollution Degree 2 2

Class of Equipment I

1 Installation category (overvoltage category) defines the level of transient overvoltagewhich the instrument is designed to withstand safely. It depends on the nature of theelectricity supply and its overvoltage protection means. For example, in CAT II which isthe category used for instruments in installations supplied from a supply comparable topublic mains such as hospital and research laboratories and most industrial laboratories,the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120Vsupply.

2 Pollution degree describes the amount of conductive pollution present in theoperating environment. Pollution degree 2 assumes that normally only non-conductivepollution such as dust occurs with the exception of occasional conductivity caused bycondensation.

7-2 Model 8600 Series Thermo Scientific

Section 7Specifications

Model 8621 8620

Temperature Range -50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient

Exterior Dimensions96.0”W x 40.5” H x 28.8” F-B(243.8cm x 102.9cm x 73.2cm)Add 7.9” (20.1cm) to F-B for wall spacer/handle.

Interior Dimensions66.5”W x 28.0”H x 18.5” F-B(168.9cm x 71.1cm x 47.6cm)

Capacity 20.0 cu. ft. (566.3 liters)

Refrigeration Cascade system, (2) hermetically-sealed compressors

Insulation Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;1.0" (12.5cm) sub-lids

Electrical120VAC, 1 PH, 60 Hz, 16.0 FLAOperating Range: 108VAC-130VAC

230VAC, 1 PH, 50/60 Hz, 12.0 FLAOperating Range: 208VAC-240VAC

Breaker Requirements 20 Amp, 120VAC, Dedicated Circuit,20 Amp Time Delay Breaker

15 Amp, 230VAC, Dedicated Circuit,15 Amp Time Delay Breaker

Shipping Weight 833 lbs. ( 377.8 kg)

Page 46: Model 8600 Series - Thermo Fisher

Model 8600 Series 8-1Thermo Scientific

Section 8Spare Parts

Page 47: Model 8600 Series - Thermo Fisher

8-2 Model 8600 Series Thermo Scientific

Section 8Spare Parts

8615-200-1-B Rev. 1Page 1 of 2

CHEST FREEZER LID ASSEMBLY

Page 48: Model 8600 Series - Thermo Fisher

Model 8600 Series 8-3Thermo Scientific

Section 8Spare Parts

8615-200-1-B Rev. 1Page 2 of 2

CHEST FREEZER LID ASSEMBLY

Page 49: Model 8600 Series - Thermo Fisher

8-4 Model 8600 Series Thermo Scientific

Section 8Spare Parts

Page 50: Model 8600 Series - Thermo Fisher

Model 8600 Series 8-5Thermo Scientific

Section 8Spare Parts

Page 51: Model 8600 Series - Thermo Fisher

Exploded PartsDrawing

12.7, 17 & 20 cu ftChest Sidecar

8615-200-3-B Rev. 5Page 1 of 2

8-6 Model 8600 Series Thermo Scientific

Section 8Spare Parts

Page 52: Model 8600 Series - Thermo Fisher

Model 8600 Series 8-7Thermo Scientific

Section 8Spare Parts

Exploded PartsDrawing

12.7, 17 & 20 cu ftChest Sidecar

8615-200-3-B Rev. 5Page 2 of 2

Page 53: Model 8600 Series - Thermo Fisher

Exploded ViewDrawing

230V Relay EnclosureAssembly

8602-204-1-D Rev. 5Page 1 of 2

Part Number

Quantity

8-8 Model 8600 Series Thermo Scientific

Section 8Spare Parts

Page 54: Model 8600 Series - Thermo Fisher

Model 8600 Series 8-9Thermo Scientific

Section 8Spare Parts

Exploded ViewDrawing

230V Relay EnclosureAssembly

8602-204-1-D Rev. 5Page 2 of 2

Page 55: Model 8600 Series - Thermo Fisher

Exploded ViewDrawing

120V Relay EnclosureAssembly

8602-204-2-D Rev. 5Page 1 of 2

8-10 Model 8600 Series Thermo Scientific

Section 8Spare Parts

Page 56: Model 8600 Series - Thermo Fisher

Exploded ViewDrawing

120V Relay EnclosureAssembly

8602-204-2-D Rev. 5Page 2 of 2

Model 8600 Series 8-11Thermo Scientific

Section 8Spare Parts

Page 57: Model 8600 Series - Thermo Fisher
Page 58: Model 8600 Series - Thermo Fisher

Model 8600 Series 9-1Thermo Scientific

Section 9Refrigeration Schematics

Page 59: Model 8600 Series - Thermo Fisher

9-2 Model 8600 Series Thermo Scientific

Section 9Refrigeration Schematics

Page 60: Model 8600 Series - Thermo Fisher

Model 8600 Series 9-3Thermo Scientific

Section 9Refrigeration Schematics

Page 61: Model 8600 Series - Thermo Fisher
Page 62: Model 8600 Series - Thermo Fisher

Model 8600 Series 10-1Thermo Scientific

Section 10Electrical Schematics

Page 63: Model 8600 Series - Thermo Fisher

10-2 Model 8600 Series Thermo Scientific

Section 10Electrical Schematics

Page 64: Model 8600 Series - Thermo Fisher

Model 8600 Series 10-3Thermo Scientific

Section 10Electrical Schematics

Page 65: Model 8600 Series - Thermo Fisher

10-4 Model 8600 Series Thermo Scientific

Section 10Electrical Schematics

Page 66: Model 8600 Series - Thermo Fisher

Model 8600 Series 10-5Thermo Scientific

Section 10Electrical Schematics

Page 67: Model 8600 Series - Thermo Fisher

10-6 Model 8600 Series Thermo Scientific

Section 10Electrical Schematics

Page 68: Model 8600 Series - Thermo Fisher

Model 8600 Series 11-1Thermo Scientific

Section 11Warranty Information

THER

MO F

ISHE

R SC

IENT

IFIC

900 &

8600

SER

IES

ULT

FREE

ZER

WAR

RANT

YT

he

Warr

an

ty P

erio

d s

tart

s t

wo

we

eks f

rom

th

e d

ate

yo

ur

eq

uip

me

nt

is s

hip

pe

d f

rom

ou

r fa

cili

ty.

Th

is a

llow

s f

or

sh

ipp

ing

tim

e s

o t

he

wa

rra

nty

will

go

in

to e

ffe

ct

at

ap

pro

xim

ate

ly t

he

sa

me

tim

e y

ou

r e

qu

ipm

en

t is

de

live

red

. T

he

wa

rra

nty

pro

tectio

n

exte

nds t

o a

ny s

ubsequent

ow

ner

during t

he w

arr

anty

period.

Du

rin

g t

he

first

two

ye

ars

of

the

wa

rra

nty

pe

rio

d,

co

mp

on

en

t part

s p

rove

n t

o b

e n

on

-co

nfo

rmin

g in

ma

teria

ls o

r w

ork

ma

nsh

ip

will

be r

epaired o

r re

pla

ced a

t T

herm

o's

expense,

labor

inclu

ded.

The 9

00 S

eries U

LT

Fre

ezers

inclu

de a

n a

dditio

nal tw

o

ye

ar

wa

rra

nty

on

th

e c

om

pre

sso

rs,

part

s o

nly

, F.O

.B.

facto

ry.

Th

e 8

60

0 S

erie

s U

LT

Fre

eze

rs in

clu

de

an

ad

ditio

na

l th

ree

ye

ar

wa

rra

nty

on

th

e c

om

pre

sso

rs,

part

s o

nly

, F.O

.B.

facto

ry.

Insta

llatio

n a

nd

ca

libra

tio

n is n

ot

co

ve

red

by t

his

wa

rra

nty

ag

ree

-

me

nt.

Th

e T

ech

nic

al S

erv

ice

s D

epa

rtm

en

t m

ust

be

co

nta

cte

d f

or

wa

rra

nty

de

term

ina

tio

n a

nd

dire

ctio

n p

rio

r to

an

y w

ork

be

ing

pe

rfo

rme

d.

Exp

en

da

ble

ite

ms,

i.e

., g

lass,

filte

rs,

pilo

t lig

hts

, lig

ht

bu

lbs a

nd

do

or

ga

ske

ts a

re e

xclu

de

d f

rom

th

is w

ar-

ran

ty.

In a

dditio

n t

o t

he s

tandard

warr

anty

, th

e f

oam

ed-in-p

lace c

abin

et

desig

n c

arr

ies a

unit p

roduction lifetim

e w

arr

anty

(fo

am

ed-

in-p

lace

ca

bin

et,

eva

po

rato

r a

nd

fo

am

ed

-in

-pla

ce

do

or;

part

s o

nly

). P

lea

se

co

nta

ct

yo

ur

sa

les r

ep

rese

nta

tive

or

Th

erm

o f

or

additio

nal in

form

ation.

Re

pla

ce

me

nt

or

repa

ir o

f co

mp

on

en

t part

s o

r e

qu

ipm

en

t u

nd

er

this

wa

rra

nty

sh

all

no

t e

xte

nd

th

e w

arr

an

ty t

o e

ith

er

the

equip

ment

or

to t

he c

om

ponent

part

beyond t

he o

rigin

al tw

o y

ear

warr

anty

period.

The T

echnic

al S

erv

ices D

epart

ment

must

giv

e p

rior

appro

val fo

r th

e r

etu

rn o

f any c

om

ponents

or

equip

ment.

TH

IS W

AR

RA

NT

Y I

S E

XC

LU

SIV

E A

ND

IN

LIE

U O

F A

LL O

TH

ER

WA

RR

AN

TIE

S,

WH

ET

HE

R W

RIT

TE

N,

OR

AL

, O

R

IMP

LIE

D.

NO

WA

RR

AN

TIE

S O

F M

ER

CH

AN

TA

BIL

ITY

OR

FIT

NE

SS

FO

R A

PA

RT

ICU

LA

R P

UR

PO

SE

SH

AL

L A

PP

LY.

Therm

o s

hall

not

be lia

ble

for

any indirect

or

consequential dam

ages inclu

din

g,

without

limitation,

dam

ages r

ela

ting t

o lost

pro

fits

or

loss o

f pro

ducts

.

Your

local T

herm

o S

ale

s O

ffic

e is r

eady t

o h

elp

with c

om

pre

hensiv

e s

ite p

repara

tion info

rmation b

efo

re y

our

equip

ment

arr

ives.

Printe

d instr

uction m

anuals

care

fully

deta

il equip

ment

insta

llation,

opera

tion,

and p

reventive m

ain

tenance.

If e

quip

ment

serv

ice is r

equired,

ple

ase c

all

your

Technic

al S

erv

ices D

epart

ment

at

1-8

00-4

38-4

851 (

US

A a

nd C

anada)

or

1-7

40-3

73-4

763.

We're r

eady t

o a

nsw

er

your

questions o

n e

quip

ment

warr

anty

, opera

tion,

main

tenance,

serv

ice,

and

spe-

cia

l a

pp

lica

tio

ns.

Ou

tsid

e t

he

US

A c

on

tact

yo

ur

loca

l d

istr

ibu

tor

for

wa

rra

nty

in

form

atio

n.

ISO

9001

REGI

STER

EDR

ev.

4

4/0

9

Page 69: Model 8600 Series - Thermo Fisher

11-2 Model 8600 Series Thermo Scientific

Section 11Warranty Information

THER

MO FI

SHER

SCIEN

TIFIC

900 &

8600

SERIE

S ULT

FREE

ZER

INTER

NATIO

NAL D

EALE

R WA

RRAN

TYT

he

Warr

an

ty P

erio

d s

tart

s t

wo

mo

nth

s f

rom

th

e d

ate

yo

ur

eq

uip

me

nt

is s

hip

pe

d f

rom

ou

r fa

cili

ty.

Th

is a

llow

s f

or

sh

ipp

ing

tim

e

so

th

e w

arr

an

ty w

ill g

o in

to e

ffe

ct

at

ap

pro

xim

ate

ly t

he

sa

me

tim

e y

ou

r e

qu

ipm

en

t is

de

live

red

. T

he

wa

rra

nty

pro

tectio

n

exte

nds t

o a

ny s

ubsequent

ow

ner

during t

he w

arr

anty

period.

Deale

rs w

ho s

tock o

ur

equip

ment

are

allo

wed a

n a

dditio

nal fo

ur

month

s f

or

deliv

ery

and insta

llation,

pro

vid

ing t

he w

arr

anty

card

is c

om

ple

ted a

nd r

etu

rned t

o t

he T

echnic

al S

erv

ices

De

pa

rtm

en

t.

Du

rin

g t

he

first

two

ye

ars

of

the

wa

rra

nty

pe

rio

d,

co

mp

on

en

t part

s p

rove

n t

o b

e n

on

-co

nfo

rmin

g in

ma

teria

ls o

r w

ork

ma

nsh

ip

will

be r

epaired o

r re

pla

ced a

t T

herm

o's

expense,

labor

exclu

ded.

The 9

00 S

eries U

LT

Fre

ezers

inclu

de a

n a

dditio

nal tw

o

ye

ar

wa

rra

nty

on

th

e c

om

pre

sso

rs,

part

s o

nly

, F.O

.B.

facto

ry.

Th

e 8

60

0 S

erie

s U

LT

Fre

eze

rs in

clu

de

an

ad

ditio

na

l th

ree

ye

ar

wa

rra

nty

on

th

e c

om

pre

sso

rs,

part

s o

nly

, F.O

.B.

facto

ry.

Insta

llatio

n a

nd

ca

libra

tio

n is n

ot

co

ve

red

by t

his

wa

rra

nty

ag

ree

me

nt.

The T

echnic

al S

erv

ices D

epart

ment

must

be c

onta

cte

d f

or

warr

anty

dete

rmin

ation a

nd d

irection p

rior

to a

ny w

ork

bein

g p

er-

form

ed

. E

xp

en

da

ble

ite

ms,

i.e

., g

lass,

filte

rs,

pilo

t lig

hts

, lig

ht

bu

lbs a

nd

do

or

ga

ske

ts a

re e

xclu

de

d f

rom

th

is w

arr

an

ty.

In a

dditio

n t

o t

he s

tandard

warr

anty

, th

e f

oam

ed-in-p

lace c

abin

et

desig

n c

arr

ies a

unit p

roduction lifetim

e w

arr

anty

(fo

am

ed-in

-

pla

ce

ca

bin

et,

eva

po

rato

r a

nd

fo

am

ed

-in

-pla

ce

do

or;

part

s o

nly

). P

lea

se

co

nta

ct

yo

ur

sa

les r

ep

rese

nta

tive

or

Th

erm

o f

or

ad

di-

tional in

form

ation.

Re

pla

ce

me

nt

or

repair o

f co

mp

on

en

t part

s o

r e

qu

ipm

en

t u

nd

er

this

wa

rra

nty

sh

all

no

t e

xte

nd

th

e w

arr

an

ty t

o e

ith

er

the

eq

uip

-

me

nt

or

to t

he

co

mp

on

en

t part

be

yo

nd

th

e o

rig

ina

l tw

o y

ea

r w

arr

an

ty p

erio

d.

Th

e T

ech

nic

al S

erv

ice

s D

epa

rtm

en

t m

ust

giv

e

prior

appro

val fo

r th

e r

etu

rn o

f any c

om

ponents

or

equip

ment.

TH

IS W

AR

RA

NT

Y I

S E

XC

LU

SIV

E A

ND

IN

LIE

U O

F A

LL O

TH

ER

WA

RR

AN

TIE

S,

WH

ET

HE

R W

RIT

TE

N,

OR

AL

, O

R

IMP

LIE

D.

NO

WA

RR

AN

TIE

S O

F M

ER

CH

AN

TA

BIL

ITY

OR

FIT

NE

SS

FO

R A

PA

RT

ICU

LA

R P

UR

PO

SE

SH

AL

L A

PP

LY.

Therm

o s

hall

not

be lia

ble

for

any indirect

or

consequential dam

ages inclu

din

g,

without

limitation,

dam

ages r

ela

ting t

o lost

pro

fits

or

loss o

f pro

ducts

.

Your

local T

herm

o S

ale

s O

ffic

e is r

eady t

o h

elp

with c

om

pre

hensiv

e s

ite p

repara

tion info

rmation b

efo

re y

our

equip

ment

arr

ives.

Printe

d instr

uction m

anuals

care

fully

deta

il equip

ment

insta

llation,

opera

tion,

and p

reventive m

ain

tenance.

If e

qu

ipm

en

t se

rvic

e is r

eq

uire

d,

ple

ase

co

nta

ct

yo

ur

loca

l d

istr

ibu

tor

or

Th

erm

o (

1-8

00-4

38-4

851 in U

SA

and C

anada,

or

1-7

40-3

73-4

763).

We're r

eady t

o a

nsw

er

your

questions o

n e

quip

ment

warr

anty

, opera

tion,

main

tenance,

serv

ice,

and s

pecia

l

ap

plic

atio

ns.

Ou

tsid

e t

he

US

A,

co

nta

ct

yo

ur

loca

l d

istr

ibu

tor

for

wa

rra

nty

in

form

atio

n.

ISO

9001

REGI

STER

EDR

ev.

4

4/0

9

Page 70: Model 8600 Series - Thermo Fisher

Model 8600 Series 12-1Thermo Scientific

Appendix A Handling Liquid Nitrogen

Warning Contact of liquid nitrogen or cold gas with the skin or eyes maycause serious freezing (frostbite) injury. s

Handle liquid nitrogen carefully. The extremely low temperature can freeze human flesh very rapidly. Whenspilled on a surface the liquid tends to cover it completely and intimately,cooling a large area. The gas issuing from the liquid is also extremely cold.Delicate tissue, such as that of the eyes, can be damaged by an exposure tothe cold gas which would be too brief to affect the skin of the hands orface.

Never allow any unprotected part of your body to touch objects cooledby liquid nitrogen. Such objects may stick fast to the skin and tear the flesh when you attemptto free yourself. Use tongs to withdraw objects immersed in the liquid,and handle the object carefully.

Wear protective clothing. Protect your eyes with a face shield or safety goggles (safety glasses withoutside shields do not give adequate protection). Always wear gloves whenhandling anything that is, or may have been, in immediate contact withliquid nitrogen. Insulated gloves are recommended, but heavy leathergloves may also be used. The gloves should fit loosely, so that they can bethrown off quickly if liquid should splash into them. When handlingliquid in open containers, it is advisable to wear high-top shoes. Trousers(which should be cuffless if possible) should be worn outside the shoes.

Page 71: Model 8600 Series - Thermo Fisher

12-2 Model 8600 Series Thermo Scientific

Section 12Appendix A

The safe handling and use of liquid nitrogen in cryogenic refrigerators anddewar flasks is largely a matter of knowing the potential hazards and usingcommon-sense procedures based on that knowledge. There are twoimportant properties of liquid nitrogen that present potential hazards:

1. It is extremely cold. At atmospheric pressure, liquid nitrogen boils at -320°F (-196°C).

2. Very small amounts of liquid vaporize into large amounts of gas. Oneliter of liquid nitrogen becomes 24.6cu. ft. (700l) of gas.

The safety precautions in this appendix must be followed to avoid potentialinjury or damage which could result from these two characteristics. Donot attempt to handle liquid nitrogen until you read and fully understandthe potential hazards, their consequences, and the related safetyprecautions. Keep this booklet handy for ready reference and review.

Note Because argon is an inert gas whose physical properties are verysimilar to those of nitrogen, the precautions and safe practices for thehandling and use of liquid argon are the same as for liquid nitrogen. s

Use only containers designed for low temperature liquids. Cryogenic containers are specifically designed and made of materials thatcan withstand the rapid changes and extreme temperature differencesencountered in working with liquid nitrogen. Even these special containersshould be filled SLOWLY to minimize the internal stresses that occurwhen any material is cooled. Excessive internal stresses can damage thecontainer.

Do not cover or plug the entrance opening of any liquid nitrogen refrig-erator or dewar. Do not use any stopper or other device that wouldinterfere with venting of gas. These cryogenic liquid containers are generally designed to operate withlittle or no internal pressure. Inadequate venting can result in excessive gaspressure which could damage or burst the container. Use only the loose-fitting necktube core supplied or one of the approved accessories forclosing the necktube. Check the unit periodically to be sure that venting isnot restricted by accumulated ice or frost.

Introduction

Page 72: Model 8600 Series - Thermo Fisher

Model 8600 Series 12-3Thermo Scientific

Section 12Appendix A

Use proper transfer equipment. Use a phase separator or special filling funnel to prevent splashing andspilling when transferring liquid nitrogen into or from a dewar orrefrigerator. The top of the funnel should be partly covered to reducesplashing. Use only small, easily-handled dewars for pouring liquid. Forthe larger, heavier containers, use a cryogenic liquid withdrawal device totransfer liquid from one container to another. Be sure to followinstructions supplied with the withdrawal device. When liquid cylindersor other large storage containers are used for filling, follow the instructionssupplied with those units and their accessories.

Do not overfill containers. Filling above the bottom of the necktube (or specified maximum level) canresult in overflow and spillage of liquid when the necktube core or cover isplaced in the opening.

Never use hollow rods or tubes as dipsticks. When a warm tube is inserted into liquid nitrogen, liquid will spout fromthe top of the tube due to gasification and rapid expansion of liquid insidethe tube.

Warning Nitrogen gas can cause suffocation without warning! s

Store and use liquid nitrogen only in a well-ventilated place. As the liquid evaporates, the resulting gas tends to displace the normal airfrom the area. In closed areas, excessive amounts of nitrogen gas reducethe concentration of oxygen and can result in asphyxiation. Becausenitrogen gas is colorless, odorless and tasteless, it cannot be detected by thehuman senses and will be breathed as if it were air. Breathing anatmosphere that contains less than 18% oxygen can cause dizziness andquickly result in unconsciousness and death.

Note The cloudy vapor that appears when liquid nitrogen is exposed to theair is condensed moisture, not the gas itself. The issuing gas is invisible. s

Never dispose of liquid nitrogen in confined areas or places whereothers may enter. Disposal of liquid nitrogen should be done outdoors in a safe place. Pourthe liquid slowly on gravel or bare earth where it can evaporate withoutcausing damage. Do not pour the liquid on pavement.

Introduction(continued)

Page 73: Model 8600 Series - Thermo Fisher

12-4 Model 8600 Series Thermo Scientific

Section 12Appendix A

Appendix B Handling Liquid CO2

Warning High concentrations of CO2 gas can cause asphyxiation! OSHAStandards specify that employee exposure to carbon dioxide in any eight-hour shift of a 40-hour work week shall not exceed the eight-hour timeweighted average of 5000 PPM (0.5% CO2). The short term exposurelimit for 15 minutes or less is 30,000 PPM (3% CO2). Carbon dioxidemonitors are recommended for confined areas where concentrations ofcarbon dioxide gas can accumulate. s

Store and use liquid CO2 only in a well-ventilated place. As the liquid evaporates, the resulting gas tends to displace the normal airfrom the area. In closed areas, excessive amounts of CO2 gas reduce theconcentration of oxygen and can result in asphyxiation. Because CO2 gasis colorless, odorless and tasteless, it cannot be detected by the humansenses and will be breathed as if it were air. Breathing an atmosphere thatcontains less than 18% oxygen can cause dizziness and quickly result inunconsciousness and death.

Note The cloudy vapor that appears when liquid CO2 is exposed to the airis condensed moisture; not the gas itself. The issuing gas is invisible. s

Never dispose of liquid CO2 in confined areas or places where othersmay enter. Disposal of liquid CO2 should be done outdoors in a safe place. Pour theliquid slowly on gravel or bare earth where it can evaporate withoutcausing damage. Do not pour the liquid on pavement.

Page 74: Model 8600 Series - Thermo Fisher

Model 8600 Series 12-5Thermo Scientific

Section 12Appendix A

If a person seems to become dizzy or loses consciousness while workingwith liquid nitrogen or carbon dioxide, move them to a well-ventilatedarea immediately. If breathing has stopped, apply artificial respiration. Ifbreathing is difficult, give oxygen. Call a physician. Keep warm and atrest.

If exposed to liquid or cold gas, restore tissue to normal body temperature(98.6°F) as rapidly as possible, followed by protection of the injured tissuefrom further damage and infection. Remove or loosen clothing that mayconstrict blood circulation to the frozen area. Call a physician. Rapidwarming of the affected part is best achieved by using water at 108°F.Under no circumstance should the water be over 112°F, nor should thefrozen part be rubbed either before or after rewarming. The patient shouldneither smoke nor drink alcohol.

First Aid

Page 75: Model 8600 Series - Thermo Fisher

Model 8600 Series 13-1Thermo Scientific

Page 76: Model 8600 Series - Thermo Fisher

13-2 Model 8600 Series Thermo Scientific

Page 77: Model 8600 Series - Thermo Fisher
Page 78: Model 8600 Series - Thermo Fisher

Thermo ScientificControlled Environment Equipment401 Millcreek RoadMarietta, Ohio 45750United States

www.thermo.com