EMC seminar Page 1 Misure di pre-compliance EMI Roberto Sacchi Application Engineer [email protected]

Transcript of Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility...

Page 1: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Page 1

Misure di pre-compliance EMI

Roberto Sacchi

Application [email protected]

Page 2: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Agenda

Introduzione alle misure EMI Terminologia;

Sistema di misura (antenna, LISN, ricevitore, etc.);

Detectors;

Normative europee ed internazionali

Misure di pre-compatibilita’ elettromagnetica Misure di emissioni radiate;

Misure di emissioni condotte

Soluzioni Agilent Introduzione agli analizzatori serie-X;

Software applicativo per le misure di pre-compatibilita’ EMI

Page 2

Page 3: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Comparison of precompliance and full compliance

measurements

Page 3

Precompliance MeasurementsEvaluate the conducted and radiated

emissions of a device using correct

detectors and bandwidths before going

to a test house for compliance testing

Full Compliance measurements

Full compliance testing requires a receiver

that meets all the requirements of CISPR

16-1-1 (response to a CISPR pulse gen),

a qualified open area test site or semi

anechoic chamber and an antenna tower

and turntable to maximize EUT signals.

Page 4: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

What is EMC?

Electromagnetic Compatibility (EMC): The ability of equipment to function

satisfactorily in its electromagnetic environment without introducing intolerable

disturbances into that environment or into other equipment.

Combination of Interference and Immunity.

Electromagnetic Interference (EMI):Electromagnetic energy emanating from one device which causes another device to

have degraded performance.

Electromagnetic Immunity (Susceptibility, EMS): Tolerance in the presence

of electromagnetic energy (Performance degradation due to electromagnetic energy).

Compliance measurements require a receiver that meets the requirements of

CISPR part 16 (for commercial) or MIL-STDd 461 (for military).

All EMI receivers require a pre-selector at lower frequencies to limit the input energy

and maintain sufficient dynamic range to meet the CISPR 16 requirements.

Page 5: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Definitions

EMC –

ElectroMagnetic Compatibility

EMI –

ElectroMagnetic

Interference

EMS –

ElectroMagnetic

Susceptibility

(aka Immunity)

Page 6: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Page 6

Page 7: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

OverviewWhat is Signal, Vector and Spectrum Analysis?

Page 7

•Display and measure amplitude versus frequency for RF & MW signals

•Separate or demodulate complex signals into their base components (sine waves)

Spectrum Analysis

Page 8: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

OverviewTypes of Tests Made

Page 8

Modulation

Noise

Distortion

EMC

Page 9: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

9

Architecture of Modern Spectrum/Signal Analyzers

What does “Modern” mean?

Digitize the IF output, not detector output

FFT and swept capability (neither one is optimum for everything)

Digitized data output available

Connectivity

Automated measurement features

What we hope it doesn’t mean

Incompatibility

What can it mean

Ability to use new features to duplicate or expand necessary old ones

Complete spectrum analyzer & vector signal analyzer

Page 10: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Theory of OperationSwept Spectrum Analyzer Block Diagram

Page 10

Pre-Selector

Or Low Pass

Input Filter

Crystal

Reference

Oscillator

Log

Amp

RF input

attenuator

mixer

IF filter

(RBW)envelope

detector

video

filterlocal

oscillator

sweep

generator

IF gain

Input

signal

ADC, Display

& Video

Processing

Page 11: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

11

Traditional Spectrum Analyzer

Scalar analysis

Digitizing the video signal

Classic superheterodyne swept spectrum analyzer

Product detector

loss of phase

information

Page 12: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

12

Digital IF Spectrum/Signal Analyzer

Vector data CAN be preserved (mag/phase or I/Q)

Digitizing the IF Signal

Some troublesome operations

and conversions are now

fast, accurate DSP

Page 13: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

OverviewDifferent Types of Analyzers

Page 13

Parallel filters measured

simultaneouslyA

ff1 f2

FFT Analyzer

A

ff1 f2

Filter 'sweeps' over

range of interest

Swept Analyzer

Page 14: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

SpecificationsResolution: RBW Type Determines Sweep Time

280 sec

134 sec

13.5 sec

8563E Analog RBW

PSA Digital RBW

PSA FFT RBW

Page 15: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

15

Speed Improvements

Useful comparisons highly specific, many factors

PXA mode switching typically faster than PSA

Where speed is critical, consider modifying measurement routines to

include features such as list sweep

Benchmark PXA PSASpeed

improvement

Preset (*RST) 28 ms 168 ms 6x

Marker peak search 6.5 ms 78 ms 12x

Local Update 13 ms 17 ms 1.3x

CF Tune and Transfer (4 - 5GHz) 109 ms 186 ms 1.7x

Remote sweep and trace transfer 18 ms 30 ms 1.67x

Nominal speed comparison, PSA example:

Page 16: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Modern spectrum analyzer

Resolution BW Selectivity or Shape Factor

Page 16

3 dB

60 dB

60 dBBW

60 dB BW

3 dB BW

3 dB BW

Selectivity =

Determines resolvability of unequal amplitude signals

Page 17: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Specifications

Resolution: RBW Type and Selectivity

Page 17

DIGITAL FILTER

ANALOG FILTER

SPAN 3 kHzRES BW 100 Hz

Typical

Selectivity

Analog 15:1

Digital ≤5:1

Page 18: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

18

Digital Filter Shape

Better shape factor, biggest selectivity benefit for different signal levels

Equivalent selectivity at a wider, faster-sweeping RBW

digital filters swept an additional 3-4x faster

30 kHz Digital Filter

Page 19: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

CISPR Bandwidth Requirements

Measurement Range CISPR Band CISPR Bandwidth

9 KHz – 150KHz A 200 Hz

150 KHz – 30 MHz B 9 KHz

30 MHz – 1 GHz C/D 120 KHz

> 1GHz E 1 MHz

Page 19

Bandwidth -6dB

-20dB

Page 20: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

MIL-STD-461 Bandwidth Requirements

Measurement Range -6dB Bandwidth

30Hz - 1 KHz 10 Hz

1 KHz -10 KHz 100 Hz

10 KHz - 150 KHz 1 KHz

150 KHz - 30MHz 10 KHz

30 MHz - GHz 100 KHz

> 1GHz 1 MHz

Page 20

Page 21: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

21

Some modern analyzers approach accuracy of power meter + sensor

• Even better for low-level signals, with narrower noise bandwidth and

the benefit of frequency selectivity

Some factors determining uncertainty:

• Input connector (mismatch)

• RF input attenuator

• Mixer and input filter (flatness)

• IF gain/attenuation (reference level)

• RBW filters

• Display scale fidelity

• Calibrator

Modern Spectrum Analyzer Accuracy

Page 22: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

22

Modern Spectrum Analyzer Accuracy Examples

Page 23: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Line Impedance Stabilization Networks (LISN)

Page 23

Purpose of a LISN:

1. Isolates the power mains from the

equipment under test. The power

supplied to the EUT must be as clean as

possible. Any noise on the line will be

coupled to the X-Series signal analyzer

and interpreted as noise generated by

the EUT.

2. Isolates any noise generated by the EUT

from being coupled to the power mains.

Excess noise on the power mains can

cause interference with the proper

operation of other devices on the line.

3. The signals generated by the EUT are

coupled to the X-Series analyzer using a

high-pass filter, which is part of the LISN.

Signals that are in the pass band of the

high-pass filter see a 50-Ω load.

Page 24: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

LISN

Page 24

Page 25: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

LISN

Page 25

@ Electrical Network Frequency

@ 150 kHz to 30 MHz

Page 26: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Transient Limiter

The purpose of the limiter is to protect the input of the EMC analyzer from

large transients when connected to a LISN. Switching EUT power on or off

can cause large spikes generated in the LISN.

The Agilent 11947A transient limiter incorporates a limiter, high-pass filter,

and an attenuator. It can withstand 10 kW for 10 μsec and has a frequency

range of 9 kHz to 200 MHz. The high-pass filter reduces the line frequencies

coupled to the EMC analyzer.

Page 26

DUT

LimiterLISN

Page 27: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Field Strength and Antenna factors

Radiated EMI emissions tests measure the electric field. The

field strength is calibrated in dBμV/m.

Antenna factors is the ratio of the electric field (V/m) present

at the plane of the antenna versus the voltage out of the

antenna connector.

Log units:

AF(dB/m) = E(dBμV/m) - V(dBμV)

E(dBμV/m) = V(dBμV) + AF(dB/m)

Notes:

Antenna factors are not the same as antenna gain.

dBμV = dBm + 107

Page 27

Page 28: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Field Strength Unit

Page 28

Radiated EMI emissions measurements measure the

electric field. The field strength is calibrated in dBμV/m.

Pt = total power radiated from an isotropic radiator

Pd = the power density at a distance from the isotropic radiator

(far field >λ/2π)

24 r

PP t

d

120R

R

EPd

2

2

2

4 r

P

R

E t

r

PE

t 30 [V/m]

[ohm]

Page 29: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Antennas used in EMI emission measurements

Page 29

Page 30: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

30

Detectors: Convert IF Samples to Display Bins or

“Buckets”

Multiple simultaneous detectors

Screen Shot “Detector 3types”

Time

Volts

Peak

Neg Peak

Sample

Display points or

buckets

Normal, Average, Neg Peak

Peak, Neg Peak, Sample

Page 31: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Detectors

Page 31

Most radiated and conducted limits are based on quasi-peak

detection mode.

Page 32: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Peak vs. Quasi-peak vs. Average

Page 32

time

VPeak Detection

Quasi-Peak Detection

Average Detection

time

VPeak Detection

Quasi-Peak DetectionAverage Detection

Page 33: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Peak ≧ QP ≧AveragePeak Detector

• Initially used

• Faster than QP and Average modes

• If all signals fall below the limit, then the product passes and no future

testing is needed.

QP

• For CW signal, Peak = QP

• Much slower by 2 or 3 order magnitude compared to using Peak detector

• Charge rate much faster than discharge rate

– the higher repetition rate of the signal, the higher QP reading

Average

• Radiated emissions measurements above 1 GHz are performed using

average detection

Page 33

Page 34: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Close field probe

Page 34

Measures the magnetic field H strength at the center

of its sense loop. The plane of the probe tip loops

must be perpendicular to the radiating magnetic field

Page 35: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Test example

Page 35

Page 36: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Emissions Regulations (Summary)

Page 36

Page 37: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

European Norms example

EN55014 (CISPR 14)

This standard applies to electric motor-operated and thermal

appliances for household and similar purposes, electric tools

and electric apparatus.

Limit line use depends upon the power rating of the item.

EN55014 distinguishes between household appliances, motors

less than 700W, less than 1000W and greater than 1000W.

Limits for conducted emissions are 150 kHz to 30 MHz, and

limits for radiated emissions are 30 MHz to 300 MHz.

Page 37

Page 38: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

The pre-compliance measurement process

Before making measurements on your product, some

preliminary questions must be answered.

1. Where will the product be sold (for example, Europe, United

States, Japan)?

2. What is the classification of the product?a. Information technology equipment (ITE)

b. Industrial, scientific or medical equipment (ISM)

c. Automotive or communication

d. Generic (equipment not found in other standards)

3. Where will the product be used (for example home,

commercial, light industry or heavy industry)?

With the answers to these questions, you can determine

which standard your product must be tested against.

Page 38

Page 39: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

General Process for Making EMI Measurements

Page 39

Determine the country or countries in which the product

will be sold which in turn identifies the regulator agency.

Select the limit lines to be tested to (conducted/radiated).

Select the band to be used.

Correct for transducer loses and amplifiers gains.

Identify signals above the limit that must be evaluated.

Zoom in on failed signal and perform quasi-peak or

average measurements.

Page 40: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Conducted Emissions Measurements

Page 40

1. Connect DUT to the test system

2. Set the proper frequency range

3. Load limit lines and correction factors for LISN and limiter

4. View the ambient emissions with DUT OFF

5. Switch on the DUT and find signals above limits by using peak detector

6. Measure all signals above limits with quasi-peak and average detectors

Page 41: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Page 41

Page 42: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

Radiated Emissions are difficult to

measure because of multiple

dimensions (five) and the use of

quasi-peak detection below 1GHz

41.2563MHz

218.120MHz

1500.260MHz

1 - Azimuth

2 - Antenna Height

3 - Field Strength

4 - Frequency

5 -Time

The challenge of measuring radiated emissions

Page 43: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Radiated Emissions Measurements

Page 43

1. Connect the antenna to the EMI receiver and separate the antenna from

the DUT as specified by the regulation requirements

2. Set the proper frequency range and bandwidth

3. Load limit lines and correction factors for antenna and cable.

4. With DUT OFF, measure the ambient emissions and store them

5. Switch on the DUT and find signals above limits by using peak detector

(only those not present during the ambient scan)

6. Measure all signals above limits with quasi-peak and average detectors

Page 44: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

1. Select the measurement range

Page 44

Page 45: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

2. Load Corrections factors

Page 45

Amplitude at

point circled

Amplitude

referenced to

blue line

Page 46: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

3. Load Limit line

Page 46

Circle indicates

the position of

the amplitude

frequency pair

Page 47: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

4. Scan for signals above the limits with peak

detector

Page 47

Page 48: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

5. Quasi-peak and average measurements

Page 48

Page 49: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Troubleshooting

Page 49

Use the close-field probe to locate the sources of the radiated signals

exceeding the limit lines

Page 50: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Agilent Solutions

Page 50

Page 51: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Agilent X-Series Signal Analyzers

Multiple instruments in one box: Swept spectrum analyzer;

FFT analyzer;

RF and Baseband Vector Signal analyzer;

Noise Figure analyzer.

Fastest signal analysis measurements

Broadest set of applications and demodulation capabilities

Upgradeable HW

Most advanced user interface & world-class connectivity

Page 52: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

Instrument ArchitectureModern Spectrum Analyzers Architecture (PSA, X-Series)

RF Section IF Section BB Section

•Attenuation

•Filtering

•Downconversion• RBW Filtering

• Envelope Detection

• Log Conversion

• VBW Filtering

• Peak/sample/rms

detection

• Averaging

ADCIF/BB Section

on ASIC

“All Digital” IF Architecture

Page 53: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Modern Spectrum Analyzer Block Diagram

YIG ADC

Analog IF

FilterDigital IF Filter

Digital Log Amp

Digital Detectors

FFT

Swept vs. FFTAttenuation

Pre-amp

Replaced

by

Page 54: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

“All Digital IF” Advantages

RF Section ADCIF/BB Section

on ASIC

Flexibility:

RBW filtering in 10% steps

Filters with better selectivity

Multiple operation modes (Swept, FFT, VSA, NFA)

Accuracy:

Log conversion practically ideal

No drift errors; increased repeatability

Speed:

When Swept mode is slow, go FFT

FFT

Page 55: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

55

Techniques for Reducing DANL, Improving Dynamic

Range

Reduce attenuation

Add preamp

Reduce RBW

Add external filtering

Better/shorter cables, connectors

Move analyzer closer

Time averaging (where possible, not measurement avg.)

Measurement processing (take advantage of Moore’s Law)

• Noise power subtraction/noise correction/NNC

• Noise floor extension (NFE) leverages deep knowledge of

analyzer/circuit behavior

Page 56: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

56

CW Signal Measured Near Analyzer Noise Floor

Actual S/N

Displayed

S/N

CW Signal

Apparent

Signal

This is

fundamental, and

often missedAmpl & Freq

Axes Expanded

Example: No noise subtraction or near noise correction

Page 57: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

57

Noise Subtraction, “Noise Floor Extension”

New PXA technique “NFE” improves D.A.N.L.

analyzer noise power calculated/subtracted real time

3 dB error

without NFE

“No” error

Improved noise floor

or displayed average

noise level

Page 58: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

58

Noise-Like Signal and Noise

1 MSymbol/sec QPSK, 1.9 GHz

Signal accurately measured, but noise biased higher by analyzer

noise power (no NFE)

Average detector, slower sweep to measure signal and noise,

reduce variance

Page 59: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

59

Result from Noise Subtraction

Implemented in the Agilent PXA Signal Analyzer

Blue trace shows more accurate measurement due to removal of analyzer noise power

Note increased variance of result

Page 60: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

60

Analyzer Noise Floor without NFE

Source switched off, pink trace shows analyzer noise level, no NFE

Other measurement conditions unchanged

PXA DANL (pink) adds to source power (blue) for first meas. result (yellow)

Note that noise level variance (pink trace) is smaller without NFE

Page 61: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

61

Analyzer Noise Floor with NFE

Source still off, green trace shows analyzer noise level with NFE

Other measurement conditions unchanged

Note high variance result from subtraction of small, noisy numbers

Analyzer DANL now far enough below source for minimal(0.2 - 0.4 dB) error

Page 62: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

62

A Closer Look

Pink trace adds to blue trace; result is yellow trace (NFE not used)

Green trace is included in blue trace but resulting error very small

Source noise Level, no NFE

Source Noise Level, with NFE

Analyzer Noise, no NFE

Analyzer Noise with NFE

Page 63: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

63

Signal Type vs. EffectivenessSignal/Noise Variation with RBW

Amplitude envelope vs. time

Best RBW is one matched to signal

Best ability to separate analyzer noise from signal

RBW (log)

SNR

(dB)

Noise-like

Real signals

can be one

type or

combination

Page 64: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

64

Noise Floor EnhancementCW Example

95% confidence interval, 2 dB tolerance

3.5 dB improvement for CW signal

Page 65: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

65

Noise Floor EnhancementNoise-Like Signal Example

95% confidence interval, 1 dB tolerance

9.1 dB improvement for noise-like signal

Page 66: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

66

Noise Floor EnhancementPulsed RF Example

95% confidence interval, 3 dB tolerance

10.8 dB improvement for pulsed-RF signal

Page 67: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

67

Low Noise Path

To μW

Converters

To Low Band

Page 68: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

68

Alternate “Low Noise Path”

3 dB

@ 3.6 GHz

10 dB

@ 26 GHz

Example:

Spur Search

20-50x faster

at 18 GHz

Page 69: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

69

Combining Noise Floor Extension and Low Noise Path

3.6 - 26.5 GHz, preamplifier off

Low noise path incompatible with preamp, >3.6 GHz only

Page 70: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

70

Noise Figure Measurements with a Spectrum/Signal

Analyzer and NFE

Noise Figure Measurement Application

Perf. Comparable to Dedicated Noise Figure Analyzer

NFE Offers an Additional Calibration Type

Faster/easier but less precise

NF

un

ce

rta

inty

(d

B)

NF Uncertainty vs. Cal Type

Page 71: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

71

Noise Floor Subtraction

Analyzer noise adds incoherently to any signal to be measured

Power calculations are performed on a linear power scale

(watts, not dBm) and results typically are shown in dBm

PobsS+N = PobsN + PS

PS = PobsS+N − PobsN

Page 72: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMI Roadmap

10/25/2010Page 72

EMC Features standard in X-Series:(today)

• Limit Lines (2000 pts)

• Amplitude correction (2000 pts)

• 40001 sweep points

Page 73: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

Page 73

Option EMC in X-Series:(available today)

CISPR 16-1-1 detectors

(to latest spec)

Quasi Peak

EMI Average (“CISPR-AVG”)

RMS Average (“CISPR-RMS”)

EMI Bandwidths (CISPR & MIL STD)

EMI Presets

Tune & Listen

Measure at Marker

EMI Peak, EMI Average, and

Quasi Peak measurements

displayed together

Page 74: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Page 74

Full Featured Pre-compliance Application

Ship Nov 2010

Available in all X-Series models

W/N6141A EMC measurement application

Page 75: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Corrections factors edit display

Page 75

Amplitude at

point circled

Amplitude

referenced to

blue line

Page 76: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Limit line edit display

Page 76

Circle indicates

the position of

the amplitude

frequency pair

Page 77: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

Page 77

Log Display

Peak List

Auto-detect peaks

Limit Delta

Realtime

Meters

with any 3

Simultaneous

Detectors

Page 78: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Page 78

N6141A measurement: Frequency Scan with Log Display

- same functionality as E7400 Signal List

Meters tune

to selected

signal

Page 79: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMI Roadmap

10/25/2010Page 79

N6141A measurement: Strip Chart

• Time record

of zero span

data scrolls

to left

• Up to three

different

detectors

• Can be used

to make

“click”

measure-

ments

Click measurements are made on home appliances

Patent

Applied

For

Page 80: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Option EDP (Enhanced Display Package)

for the SA- available November 2010 • Spectrogram

• Trace Zoom

• Zone Span

Group/Presentation Title

Agilent Restricted

Page 80

Page 81: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Summary

Page 81

Pre-compliance

solutions

PXA, MXA, EXA,CXA

N6141A EMC advanced

measurement application

Page 82: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Per documentazione su prodotti ed applicazioni EMI/EMC visitare il sito

http://www.agilent.com/find/EMC

Contatti:

Agilent Technologies ItaliaGiuseppe SavoiaSignal Analysis and Generation SpecialistE-mail: [email protected]

Agilent Contact CenterE-mail: [email protected]: 02 9260 8484

Page 83: Misure di pre-compliance EMI · PDF fileEvaluate the conducted and radiated ... Susceptibility (aka Immunity) EMC seminar ... Antennas used in EMI emission measurements Page 29

EMC seminar

Setup instructions for the N6141A EMC

measurement application

Press [Mode], more, {EMI Receiver}

Press [Meas], {Frequency Scan}

Press [Meas Setup], {Scan Table}, {Range 3 on}, {Range 5 off}

[Return] {Signal list}, {Delete Signals}, {Delete all} OK

Press [Mode Setup], {meter control},{meters}, {Select meter 1},

{Meter on}, {Select meter 2}, {Meter on}

Press [Meas Setup], {Scan Sequence}, {Scan only}, more

{Search criteria}, {Peak criteria and limits}, {Limits}, {Limits

on}, {Margin -6 dB}