Mini Analyzer Manual

download Mini Analyzer Manual

of 115

Transcript of Mini Analyzer Manual

  • 8/6/2019 Mini Analyzer Manual

    1/115

    Mini AnalyserEdition 1.0

    User

    Manual

  • 8/6/2019 Mini Analyzer Manual

    2/115

    +61 7 4660 1888

    Greenspan Customer Service

    Technical Support When You Need It

    The correct choice of sensor should be supported by professional advice to ensure long term success

    in the field. Greenspan Technical Services is dedicated to customer support and provides

    assistance in the selection, installation, deployment and commissioning of sensors with a full range

    of consulting services.

    A full technical support and field advice service can be accessed by ringing Customer Service on

    +61 7 4660 1888 between 8am - 6pm, 5 days a week.

    All requests for information will be serviced within 24 hours.

    All Greenspan products are designed, developed and manufactured in Australia and can be

    supplied at short notice.

    Warranty Details

    Greenspan warrants all new Greenspan products against defects in materials and workmanship for12 months from the date of invoice. During the warranty period, we will repair or, at our option,

    replace at no charge a product that proves to be defective provided that it is returned, shipping

    prepaid, to Greenspan Technology Pty Ltd.

    Greenspans liability and obligations in connection with any defects in materials and workmanship

    are expressly limited to repair or replacement, and the sole and exclusive remedy in the event of

    such defects shall be repair or replacement. Greenspans obligations under this warranty are

    conditional upon it receiving prompt written notice of claimed defects within the warranty period

    and its obligations are expressly limited to repair or replacement.

    This warranty does not apply to products or parts thereof which have been altered or repaired

    outside of the Greenspan factory or other authorised service centre, or products damaged by

    improper installation or application, or subjected to misuse, abuse neglect or accident. This

    warranty also excludes items such as reference electrodes and Dissolved Oxygen membranes thatmay degrade during normal use.

    Greenspan Technology Pty Ltd will not be liable for any incidental or consequential damage or

    expense incurred by the user due to partial or complete inoperability of its products for any reason

    whatsoever or due to inaccurate information generated by its products.

    All Warranty service will be completed as soon possible. If delays are unavoidable customers will

    be contacted immediately.

    The sensors should not be dismantled unless under instruction from Greenspan. Incorrect

    handling will void the warranty.

  • 8/6/2019 Mini Analyzer Manual

    3/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD i

    TABLE OF CONTENTS

    TABLE OF CONTENTS I

    1 INTRODUCTION 1

    1.1 Applications 2

    2 SAFETY 3

    2.1 Guidelines For Working With Chemicals 3

    2.2 Other Safety Guidelines 3

    3 OVERVIEW OF MINI-ANALYSER 4

    3.1 The Greenspan Mini-Analyser 4

    3.2 Communications Interface 5

    3.3 Serial Communication Port 5

    3.4 Battery Back-Up 6

    3.5 Configuration File 6

    3.6 Sample Delivery 7

    3.7 Optional 4-20mA Output 7

    4 WATER QUALITY PARAMETERS 8

    4.1 Dissolved Oxygen 8

    4.2 Electrical Conductivity 8

    4.3 ORP 8

    4.4 pH 9

    4.5 Turbidity 9

    4.6 Ammonia and Ammonium 94.6.1 Ammonia 9

    4.6.2 Ammonium 10

    4.7 Nitrate 10

    4.8 Phosphate 10

    5 GETTING STARTED CHECKLIST 11

    6 INSTALLATION 12

  • 8/6/2019 Mini Analyzer Manual

    4/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD ii

    6.1 Choosing the site 126.1.1 Positioning the Analyser 126.1.2 Positioning the pump 12

    6.2 Setting up the Mini-Analyser 136.2.1 Mounting the cabinets 13

    6.2.2 Unpacking the battery and sensors 13

    6.3 Configuring the communications 146.3.1 Connecting the modem 14

    6.4 Installing the sample delivery module 146.4.1 Selecting a position for the module 146.4.2 Mounting the sample delivery module 15

    6.4.3 Connecting The sample delivery and Analyser 156.4.4 Connecting the pump to the sample delivery 16

    6.5 Installing the sample delivery pump control module 166.5.1 Checking the operation of the sample delivery 17

    6.6 Connecting or adding external devices 17

    6.7 Getting the Analyser commissioned 176.7.1 Running some test schedules 176.7.2 Testing a known sample 18

    7 COMMUNICATIONS 19

    7.1 Getting Started with the User Interface 19

    7.2 Installing a Modem 19

    7.2.1 Setting the modem initialisation string 20

    8 SCHEDULES 21

    8.1 Working With Schedules Using Analyser32 218.1.1 Setting up a schedule 218.1.2 Creating or editing schedules 22

    8.1.3 Entering a name 228.1.4 Viewing and editing tests in the schedule 228.1.5 The schedule duration 228.1.6 Timing a schedule 228.1.7 Enabling the schedule 22

    8.1.8 The scheduler priority 228.1.9 Using periodic triggering 238.1.10 Using timetable triggering 248.1.11 Using event triggering 248.1.12 Using the external trigger 258.1.13 Using initial triggering 258.1.14 Editing the tests for a schedule 26

    8.1.15 Adding fill and clean operations 26

    8.2 Starting a Schedule Using Analyser32 278.2.1 The current date/time display 278.2.2 Inserting comments 278.2.3 Operator identification 27

    8.2.4 Clear memory 28

  • 8/6/2019 Mini Analyzer Manual

    5/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD iii

    8.3 Stopping a Schedule 288.3.1 The last started display 288.3.2 Current date/time 28

    8.3.3 Inserting a comment 288.3.4 Operator identification 29

    8.3.5 Confirming the end of scheduling 29

    9 ALARMS 30

    9.1 Setting Alarms With Analyser32 309.1.1 Setting the sample alarms 319.1.2 Memory alarm 319.1.3 The battery alarm 329.1.4 The solutions alarms 32

    9.1.5 Water or liquid ingress 339.1.6 Enabling an external output 339.1.7 Alarms and the data log 339.1.8 Enabling the alarm system 33

    9.1.9 Dialout on alarm 349.1.10 Receiving dialout alarms 35

    9.2 Clearing Alarms With Analyser32 359.2.1 Clearing the alarm 369.2.2 Cancelling a dialout 36

    10 DOWNLOADING DATA 36

    10.1 Downloading Data using Analyser32 3610.1.1 Downloading the entire data log 3710.1.2 Downloading new data 37

    10.1.3 Downloading data between dates 37

    11 CHEMICAL STORE 38

    11.1 Changing Solution Bottles 38

    11.2 Updating Reagents Using Analyser32 40

    11.3 Waste Disposal 4111.3.1 Unmodified waste 4111.3.2 Modified waste 41

    12 CHECKING MEASUREMENTS AND SENSORS 42

    12.1 Dissolved Oxygen 4212.1.1 Analysing the DO data 42

    12.2 Electrical Conductivity 4412.2.1 Analysing the EC data 44

    12.3 Oxidation-Reduction Potential 46

    12.4 pH 4612.4.1 Analysing the pH data 46

    12.5 Turbidity 4812.5.1 Analysing the turbidity data 48

  • 8/6/2019 Mini Analyzer Manual

    6/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD iv

    12.6 Ammonia 5012.6.1 Analysing the ammonia data 50

    12.7 Ammonium 5212.7.1 Analysing the ammonium data 52

    12.8 Nitrate 5412.8.1 Analysing the nitrate data 54

    12.9 Phosphate 5612.9.1 Analysing the phosphate data 56

    13 MAINTENANCE 58

    13.1 General Mini-Analyser Maintenance 5813.1.1 Stopping or aborting the schedule 5813.1.2 Cabinet 58

    13.1.3 Pump tubing 59

    13.1.4 Setting the pump pressure 5913.1.5 Sample/chemical tubing 6013.1.6 Valve tubing 6113.1.7 Sample delivery pot 6113.1.8 The water ingress detector 62

    13.1.9 Running diagnostics 6213.1.10 The clean regime 64

    13.2 Sensor Maintenance 6513.2.1 Replacing sensor bodies and electrodes 6513.2.2 DO 6613.2.3 EC 67

    13.2.4 ORP 69

    13.2.5 pH 7013.2.6 Turbidity 7113.2.7 Ammonia 7213.2.8 Ammonium 7413.2.9 Nitrate 7513.2.10 Reference electrodes. 76

    13.2.11 Phosphate 77

    13.3 Sensor Calibration 7813.3.1 Dissolved oxygen 8013.3.2 Electrical conductivity 8213.3.3 Oxidation-reduction potential 8413.3.4 pH 86

    13.3.5 Turbidity 87

    13.4 Maintenance Schedule 90

    14 UPGRADING FIRMWARE 91

    14.1 Direct connection 91

    14.2 Modem connection. 92

    14.3 Uploading the firmware 93

    15 TROUBLESHOOTING 97

  • 8/6/2019 Mini Analyzer Manual

    7/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD v

    15.1 Common Mini-Analyser problems. 97

    15.2 Analyser Problems within the Mini-Analyser 98

    15.3 Communications Problems 99

    16 SPECIFICATIONS 101

    16.1 Mechanical 101

    16.2 Power Supply 101

    16.3 Communications 101

    16.4 Telephone Modem 101

    16.5 GSM Modem 101

    16.6 Operating Environment 102

    16.7 Housing Requirements 102

    16.8 Measurement Range 102

    16.9 Measurement Accuracy 102

    16.10 Measurement Times 103

    16.11 Sensor Life 103

    16.12 External inputs 103

    16.13 External Outputs 103

    16.14 Storage Conditions 103

    17 SPARE PARTS LIST. 104

    17.1 Sensors 104

    17.2 Electronics 105

    17.3 Cables 106

    17.4 Plastic parts 107

    17.5 Miscellaneous 108

  • 8/6/2019 Mini Analyzer Manual

    8/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 1

    1 INTRODUCTION

    This manual covers all sensor configurations available for the Mini-Analyser. The analyser

    you have purchased may not contain all these sensors depending on the configuration that you

    have purchased. Please refer to the sections that are relevant to your analyser.

    The setting of new environmental standards worldwide has created a demand for instruments

    that can provide water quality data for the monitoring of natural water systems. The

    Greenspan Mini-Analyser offers the maximum features and flexibility of this range of

    instruments.

    Aimed at the task of environmental and industrial monitoring the Mini-Analyser is capable of

    remote communications access and on-line measurement. The most important advantages of

    on-line measurement are speed and the integrity of results. Many properties can change if the

    sampled water is exposed to air or sunlight or is stored for a period of time before testing. The

    Mini-Analyser allows measurements to be taken quickly without the need to add preservativesor to freeze the sample.

    Between measurements the sensors are automatically cleaned and are either immersed in a

    suitable storage solution or stored under conditions designed to prolong sensor life.

    Permanently submerged sensors will generally have shorter lives and therefore need to be

    replaced more often.

    The Mini-Analyser is available in several configurations covering most of the important water

    quality parameters.

    PHYSICAL MEASUREMENTS

    Dissolved Oxygen

    Electrical Conductivity

    Oxidation-Reduction Potential

    pH

    Temperature

    Turbidity

    ION SPECIFIC MEASUREMENTS

    Ammonia or Ammonium

    Nitrate

    PHOTO-COLORIMETRICMEASUREMENTS

    Phosphate

    OTHER FEATURES

    8 external sensor inputs plus rain gauge

    input

    can be configured for multiple sample

    inlets

    external alarm output

    remote access, using telephone, mobile

    network or satellite phone

    direct access using serial

    communications port

    large data log capacity

    automatic calibration of

    ammonia/ammonium, nitrate and

    phosphate sensors

    temperature compensation for

    dissolved oxygen, electrical

    conductivity and pH

  • 8/6/2019 Mini Analyzer Manual

    9/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 2

    The Mini-Analyser can be thought of as an automated chemical laboratory. It consists of

    pumps, valves, tubing, chemical storage and sensors interconnected so that a wide range of

    tasks can be performed. Controlled by a microprocessor it provides for excellent accuracy,

    precision and flexibility. Figure 1 shows a simplified block diagram.

    1.1 Applications

    The Mini-Analyser has been designed as a stand-alone analyser for installation in a diverse

    range of applications, some of these include:

    River Monitoring: Greenspan Analysers are used extensively to monitor river water

    quality prior to extraction for treatment and distribution by the water supply network

    Wastewater: The on-line monitoring of nutrient levels in wastewater treatment plants

    Wetlands: Unattended operation in remote areas monitoring the impact of human and

    agricultural nutrient levels on wetlands

    Harbour Monitoring: Used in conjunction with external sensors linked to the system, to

    monitor physical and nutrient levels and alert to the possible pollution of the waters

    Aquaculture: The Mini-Analyser can assist farm operators to maintain the best possible

    levels of nutrients to maximise returns on their investments

    Figure 1: Mini-Analyser block diagram

  • 8/6/2019 Mini Analyzer Manual

    10/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 3

    2 SAFETY

    2.1 Guidelines For Working With Chemicals

    1. Always follow approved procedures when handling, storing, packaging or using chemicals

    and chemical waste.

    2. Personnel who handle chemicals and waste must be trained in these procedures and

    informed of the potential hazards associated with them.

    3. Always read the material safety data sheets before using a chemical. In particular read the

    first aid information. The material safety data sheet should also provide information on

    the correct storage and transport of that material.

    4.

    Wear protective clothing when handling chemicals. This includes eye protection, rubbergloves and protective coat or apron.

    5. Do not consolidate untreated Mini-Analyser waste with any other waste.

    6. Keep a record of chemical inventory. This does not apply to chemicals installed in the

    Mini-Analyser cabinet. Approximately every six months check that all chemicals are

    stored correctly, in particular check for damaged labels, outdated chemicals and damaged

    containers.

    7. Chemicals when stored should be segregated according to hazard class and compatibility.

    The ammonia ionic strength adjuster (ISA) reagent is classed as corrosive.

    8. The waste container should not be allowed to fill more than 90% full. The headspace is

    needed for ease of dispensing.

    2.2 Other Safety Guidelines

    Do not attempt to perform maintenance or repairs not covered in this manual. All repairs are

    to be carried out by Greenspan or Greenspan trained personnel.

  • 8/6/2019 Mini Analyzer Manual

    11/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 4

    3 OVERVIEW OF MINI-ANALYSER

    3.1 The Greenspan Mini-Analyser

    The Greenspan Mini-Analyser can provide a tailored solution to your specific monitoring

    needs. Designed specifically to suit certain monitoring situations such as environmental,industrial discharge or single point of source. The Mini-Analyser can be configured to suit a

    wide range of applications in a variety of locations.

    Figure 3.1: Mini-Analyser

  • 8/6/2019 Mini Analyzer Manual

    12/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 5

    3.2 Communications Interface

    The connections to and from the Mini-Analyser

    terminate at the communications interface as

    shown in Figure 3.2

    The communications interface has the following

    features:

    Modem interface. Modem power and data

    connections are made here.

    External sensor inputs. Up to 8 external 4 -

    20 mA sensors may be connected.

    Rain gauge input.

    External alarm output. An external output

    pulse can be generated from an alarm

    occurring within the Mini-Analyser. This

    can be used to trigger a warning bell or

    light.

    Pump control output. This controls the

    pump that draws sample water from the

    source (river, lake etc) into the sample pot.

    Sampler trigger output. This can control an

    external water sampler.

    4 20 mA parameter outputs. Normally 4

    but expandable to 8.

    External output control signals. The

    analyser can be configured with up to 20

    additional 12Vdc outputs for controlling

    external devices.

    3.3 Serial Communication Port

    To make a direct connection from a PC to the

    Mini-Analyser, use the serial port located on the

    lower right-hand side panel. The connector is

    labelled User. Figure 3.2 shows the location

    of this connector.

    The connector is a standard 9-pin female D-

    connector.Figure 3.2: Communications interface

  • 8/6/2019 Mini Analyzer Manual

    13/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 6

    3.4 Battery Back-Up

    The Mini-Analyser electronics are powered from a 12V DC lead-acid gel battery. The battery

    is kept charged from the AC mains power. In case of loss of AC mains power supply the

    battery can keep the Mini-Analyser running for many hours. The output of the battery charger

    is set to 14.7VDC on charge and 13.8VDC on float. Do not adjust the power supply to anyother voltage.

    3.5 Configuration File

    The configuration file is stored in the Mini-Analysers memory and is a very important file.

    The information it contains includes:

    The serial number

    Calibration data

    Machine control code - for the control of pumps and valves

    User configurable settings - site information, phone numbers, alarm settings and schedule

    details.

    Chemical store information - reagent levels, constituents etc.

    A default configuration file is loaded into the Mini-Analyser at the factory. The Analyser32

    software updates this file when user configurable settings are changed.

  • 8/6/2019 Mini Analyzer Manual

    14/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 7

    3.6 Sample Delivery

    When the Mini-Analyser requires sample for testing, the sample delivery system is initiated.

    Upon initiation of the sample delivery system the Mini-Analyser starts the sample delivery

    pump and pumps water through the sample pot for several minutes ensuring that the pot has

    been flushed and filled with fresh sample. During the sample pumping a small valve on top of

    the sample pot is closed. At the end of the sample-pumping period the pump stops and the

    valve is opened venting the top of the pot to atmosphere. The Mini-Analyser then primes the

    sample from the pot to the analysis area for several minutes. The system is now ready to

    begin testing.

    3.7 Optional 4-20mA Output

    The Mini-Analyser can be configured to transmit sensor results to other control and datacapture / display systems such as those used in water treatment plants. This data can be

    transmitted using an optional 4 - 20 mA current loop printed circuit assembly. Each assembly

    is capable of four data outputs.

    Voltage and current specifications: Supply voltage range 8 to 30VDC with 100R load.

    Reverse polarity protected to 30VDC

    Maximum load of 340R with 12V supply.

    Fig 3.6: Sample deliver system

  • 8/6/2019 Mini Analyzer Manual

    15/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 8

    4 WATER QUALITY PARAMETERS

    This section details the various parameters measured by the Mini-Analyser providing

    background information on:

    What the parameters are.

    How the Mini-Analyser measures these parameters.

    4.1 Dissolved Oxygen

    Dissolved oxygen refers to the amount of oxygen dissolved in the water. DO can be

    significantly influenced by factors such as temperature, presence of dissolved salts and the

    biological and chemical demand of the water.

    To measure DO, the Mini-Analyser uses a galvanic cell type oxygen sensor. The sensor is a

    lead-oxygen battery, which has a lead anode and gold cathode. Oxygen molecules enter thecell through a semi-permeable membrane. They then diffuse in the acid electrolyte and are

    reduced at the gold electrode. A temperature compensated voltage output is proportional to

    the partial pressure of oxygen concentration. The Mini-Analyser measures DO in %

    saturation.

    4.2 Electrical Conductivity

    Electrical conductivity is a measure of how well a solution conducts electrical current. It can

    be used to detect changes in the water system, ie new pollutants or rising salinity.

    To measure EC the Mini-Analyser uses a flow cell with two electrodes. The Mini-Analysermakes a resistance/conductivity measurement using alternating current. Alternating current is

    used in EC measurements to prevent polarisation occurring. The spacing and size of the

    electrodes determines the cell constant. The Mini-Analyser EC sensor has a cell constant of

    1 cm-1.

    4.3 ORP

    ORP, oxidation-reduction potential or redox measures the ability of a fluid to be oxidised or

    reduced (give or take electrons) and is dependent on the types and concentrations of ions

    present. Oxidation and reduction happen simultaneously; while one substance is being

    oxidised the other is being reduced. Oxidation and reduction reactions occur until equilibriumis reached. Unreacted agents give the water the ability or potential to oxidise or reduce.

    This potential is what is measured.

    The Mini-Analyser measures ORP using a noble metal electrode. When a metal electrode is

    exposed to varying concentrations of chemicals, an electric potential is developed. The Mini-

    Analyser uses a platinum electrode for this purpose. The potential of the platinum electrode is

    measured against that of a reference electrode made from Ag/AgCl. This reference electrode

    does not vary its potential when chemicals in solution change. ORP is an aggregate measure

    incorporating all ions that are oxidising or reducing agents. The result is given in mV.

  • 8/6/2019 Mini Analyzer Manual

    16/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 9

    4.4 pH

    The pH of a solution is a measure of hydrogen ion activity in a solution. It defines how acidic

    or alkaline the solution is. Measured on a scale from 0 to 14, a value of 7 is said to be pH

    neutral, less than 7, acidic and above 7, basic or alkaline.

    The Mini-Analyser uses a flat, glass electrode pH sensor with a combined double junction

    reference electrode. The glass has a special composition so that a potential is formed across it

    in the presence of hydrogen ions. Inside the glass is an electrolyte of constant pH and

    concentration and a reference electrode which gives a stable voltage as a comparison.

    Between measurements the pH electrode is stored in moist air to prolong electrode life.

    4.5 Turbidity

    Turbidity is a measure of the amount of suspended solids in a solution. In natural water,

    turbidity can be caused by clay, silt, small organic matter and microscopic organisms.

    Turbidity is measured by passing light through a transparent cell containing the sample and

    measuring the amount of scattered light with a detector. The Mini-Analyser uses a

    nephelometer in which the detector is orientated at 90 to the emitter. This result is expressed

    in NTU calibrated against Formazin Standards.

    4.6 Ammonia and Ammonium

    When ammonia is dissolved in water it reacts to form ammonium. Both species are normally

    present in a solution but the relative amounts of each are influenced by the pH and, to a lesser

    extent, the ionic strength of the solution.

    When the Mini-Analyser performs a measurement of ammonia or ammonium all of the

    species are converted to one type by pH adjustment.

    4.6.1 Ammonia

    Ammonia has the symbol NH3, and is a pungent colourless gas that is highly soluble in water.

    A saturated water solution of ammonia contains about 30% ammonia at room temperature. Its

    main uses are as a fertiliser, a refrigerant (or freezing agent) and for manufacturing nitric acid.

    The Mini-Analyser uses an ammonia electrode that has a membrane separating the sample

    from an internal solution. The membrane is permeable to gas and impermeable to water.

    Dissolved ammonia in the sample solution diffuses through the membrane until the partialpressure of ammonia is the same on both sides of the membrane. The diffused ammonia gas

    dissolves and its concentration in the form of ammonia is related to the internal concentration

    of hydroxide ions. The measurement is then similar to a pH probe. The Mini-Analyser uses

    pH 4 buffer to reset the response of the electrode.

    Between measurements the Mini-Analyser keeps the electrode stored in a solution appropriate

    to the sample range being measured. The check results are given in machine counts and the

    test results are given in ppm.

  • 8/6/2019 Mini Analyzer Manual

    17/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 10

    4.6.2 Ammonium

    Ammonium has the symbol NH4. Formed during the breakdown of protein molecules from

    organic matter by bacteria, NH4 occurs naturally in water systems.

    Ammonium is measured with an ion specific electrode (ISE) and a separate half-cell reference.The electrode has an ammonium ion selective membrane. When ammonium is present in the

    sample a voltage develops across the membrane and is measured with respect to the reference

    electrode. The sample result is calculated by applying a 2-point calibration each time the

    Mini-Analyser is used. The check results are given in machine counts and the test results are

    given in ppm.

    4.7 Nitrate

    The chemical symbol for the nitrate ion is NO3. Nitrate is a very soluble ion and an oxidising

    agent. It is often found with salts such as potassium (KNO3) or sodium (NaNO3). It is a

    common ingredient in agricultural fertilisers.

    The Mini-Analyser measures nitrate using an ion specific electrode (ISE) and a separate

    reference half-cell. The electrode has a nitrate ion selective membrane. When nitrate is

    present in the sample a voltage develops across the membrane and is measured with respect to

    the reference electrode. The sample result is calculated by applying a 2-point calibration each

    time the Mini-Analyser is used. The check results are given in machine counts and the test

    results are given in ppm.

    4.8 Phosphate

    Almost all phosphorous found in natural waters is in the form of phosphates. Phosphates arecompounds, which include the molecule PO4. They are normally produced from phosphoric

    acid H3PO4 by removing a number of hydrogen atoms and replacing them with a metal.

    Commercially phosphates are used in water softeners and fertilisers.

    The Mini-Analyser uses a colorimetric method to measure phosphate. Reagents added to the

    sample react with the dissolved phosphate present to produce a blue colour, the strength of

    which is relative to the amount of phosphate in the sample. A spectrophotometer is used to

    measure this colour development. The turbidity of the sample is compensated for by

    measuring an unreacted sample. As there is no filtering of the sample the phosphorus

    measured by this process is termed total reactive phosphorus. The check results are given in

    machine counts and the test results are given in ppb or ppm.

  • 8/6/2019 Mini Analyzer Manual

    18/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 11

    5 GETTING STARTED CHECKLIST

    The following is a checklist to follow when installing and using the Mini-Analyser for the first

    time. Sections 6-11 will guide you through these steps.

    1. Verify that the selected installation site and infrastructure is ready for the installation.

    2. Ensure that the chemicals have been mixed according to the supplied Greenspan recipes,

    placed in the bottles supplied with the analyser and are ready for the installation.

    3. Check that the computer to be used at the site has HyperTerminal installed. Refer to

    section 13.1.9 for detail on setting up HyperTerminal.

    4. Install the user interface (Analyser32) on a PC or notebook computer for use at the site.

    5. Carry out the installation of the Mini-Analyser and system components. See section 6

    6. Set up communications. Set up modems, cables, initialisation strings and telephone

    numbers. See section 7.

    7. Set password access. See Analyser32 Manual, Section 4.

    8. Start and stop the analyser. See section 8.

    9. Enter schedule details. See section 8.

    10.Set alarms and dialout phone numbers. See section 9.

    11.Download from the data store. See section 10.

    12.Check the chemical store. See section 11.

    13.Check measurements and sensors. See section 12.

  • 8/6/2019 Mini Analyzer Manual

    19/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 12

    6 INSTALLATION

    The instructions detailed here are for a typical mains powered installation with a landline

    telephone connection and one sample inlet. For the installation of optional components

    consult Greenspan Technology or your local Greenspan representative.

    6.1 Choosing the site

    In most cases, by the time the installation stage has been reached, the sites will have been

    chosen for the Analyser and the pump. To cover the possibility that your site has not been

    chosen and prepared prior to the installation we will recap on the preferred criteria.

    6.1.1 Positioning the Analyser

    The Analyser cabinet should be positioned in a location that offers some protection from the

    extremes of the weather. An air-conditioned room is ideal but a location on a wall or bench

    with shelter from the sun and rain is equally acceptable. Extremes of temperature should beavoided ie below 5 Deg C and above 40 Deg C

    The location should also be away from large electrical or magnetic fields and not be subjected

    to vibration. High humidity with condensing vapour as you would find at a turbulent outflow

    is another area to be avoided.

    6.1.2 Positioning the pump

    The installation of the pump, control module, cable and piping is usually the biggest part of

    the total installation and will require many hours. The time required will depend on the

    specific site and may vary considerably with each site but at least one full day may be required

    for even a relatively simple site.

    The purpose of the sample delivery pump is to supply a representative sample of the water

    body being monitored to the analyser location at a flow rate between 20 to 60 litres per

    minute. The position of the pump in the water source requires careful consideration. Take

    care to avoid areas where weed, turbulence or silt may foul the pump or have an undesirable

    influence on the parameters that are to be monitored.

    The availability of mains power to the pump location is another factor that may influence the

    siting of the pump. The practical limit for standard cable to supply power to the pump may be

    as little as 25 metres. This distance depends on the start and run current demands of the pump

    being used. For longer distances the cable size may need to be upgraded appropriate to the

    distance involved and this can become quite expensive for long distances.

    In a typical installation the pump should ideally be installed in a stilling well that is fixed

    firmly to a pile or solid structure in the water source. The pump should be installed at a depth

    that is not too close to the bottom to suck up silt or too near the top such that it may run dry

    should the water level drop appreciably.

    Security of the pump once installed also has to be taken into account.

  • 8/6/2019 Mini Analyzer Manual

    20/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 13

    6.2 Setting up the Mini-Analyser

    The Mini-Analyser has been designed to be able to be fixed to a wall using the optional Wall

    Mounting Kit available from Greenspan. The unit is also fitted with rubber feet for sitting

    directly onto a bench or tabletop. The primary requirement is that it be a level, stable surface.

    6.2.1 Mounting the cabinets

    To fix the cabinet to a wall the wall structure and fixing points must be able to support a

    weight of 30 Kg. Use the optional wall mounting kit referring to the enclosed engineering

    note for details on fitting the kit.

    If the analyser configuration includes the larger chemical storage module then the complete

    unit may be freestanding. The top and bottom modules are fixed together with four small

    bolts. The requirement when freestanding is that it be a firm, level, stable surface and located

    where the combined cabinets cannot be knocked over. The chosen location should also have

    mains power and telephone services available unless alternative options have been purchased.

    6.2.2 Unpacking the battery and sensors

    Once the cabinet is in place, unpack and install the lead-acid gel battery. If you have

    purchased the accessories cabinet fit the battery in that cabinet. If you do not have the

    accessories cabinet do not install the battery in the electronics compartment as corrosion may

    be a concern should the battery vent gas. Place the battery on the bench top or close to the

    analyser cabinet.

    Connect the supplied power plug-pack to the mains power outlet (100-240VAC 50-60Hz) and

    the analyser 12V Batt input on the right-hand side of the analyser cabinet. Connect the cable

    marked Battery to the battery terminals with black to the negative andred to the positive +.

    If a digital multimeter is available, set it to read 12VDC and connect to the 12V battery

    terminals. Note the voltage reading, it should be approximately 12.6V. Turn on the main

    power switch for the plug pack and note the voltage again. It should now read 13.8V

    indicating that the battery charger is on and operating correctly.

    Some of the sensors will have been removed from the flow cells and packed for storage and

    shipping. These will require unpacking and replacing in the flow cells in the analyser. The

    pH and ORP electrodes will have protective cover caps that have to be removed carefully.

    Wipe any excess liquid from the electrode. Check that it is undamaged and fit to the relevant

    marked flow cell.

    The ammonia electrode NH3 requires careful unpacking and preparation before use. Refer to

    section 13.2.7 Ammonia for details.

    Ammonium NH4 and nitrate NO3 are small replaceable tip electrodes that are fitted to an

    electrode holder. These tips are packed dry in small glass vials during shipping. Remove the

    tip from the vial and fit to the appropriate electrode holder. Both the glass vial and the

    electrode holder will be marked with the parameter name. Once fitted to the holder the

    complete electrode can be fitted to the parameter flow cell.

    For full details on dealing with the sensors refer to section 13.2 Sensor Maintenance.

  • 8/6/2019 Mini Analyzer Manual

    21/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 14

    The next step is to position the chemicals in the cabinet and connect the supply tubes. The

    chemicals should have been mixed by a local certified laboratory and placed in the bottles

    supplied with the analyser prior to the installation day. Place the solution bottles in the cabinet

    and connect the colour coded pick-up tubes to the bottles. Refer to section 11 for full details

    on the chemical store.

    6.3 Configuring the communications

    Communications with the Mini-Analyser can be done using the following options:

    A serial link using a standard 9 way serial cable.

    A landline modem or cellphone modem and the public telephone network.

    The first option requires that the PC be directly connected to the analyser with a 9-way serial

    cable. A notebook computer is a convenient choice of computer for this purpose.

    The second option can use standard telephone networks, cellular networks and radio or

    satellite phone systems as the communications link.

    When on-site for the installation the direct connection method is essential for the setup and

    testing that is required. The Greenspan supplied software, Analyser32 will need to be

    installed on the computer for the installation. The computer should also have HyperTerminal

    installed. HyperTerminal is a communications program included with Microsoft Windows but

    is not always installed by default. Check that it is installed on the computer prior to

    installation day.

    6.3.1 Connecting the modem

    The analyser will be supplied with a modem for connection to a telephone landline as

    standard. The modem is positioned external to the analyser cabinet in a location of your

    choice and connected to the analyser Modem connector on the right-hand side of the

    cabinet. Two cables are supplied with the modem for connection to the power and data

    connectors on the right-hand side of the analyser. These connectors are marked

    Phone/Modem Power and Modem. The telephone landline will also need to be connected

    to the modem using a modular connector.

    Direct connection to the analyser is done from the computer to the analyser using the supplied

    9-way serial cable. The connection to the analyser is on the lower right-hand side of the

    cabinet and is marked User.

    6.4 Installing the sample delivery module

    6.4.1 Selecting a position for the module

    The sample delivery module (sample pot) should be situated close to the analyser cabinet

    (within 2 metres) and at a height slightly lower than the cabinet. If it is positioned higher than

    the cabinet the possibility exists that sample may be able to syphon into the cabinet should

    there be a tubing failure within the cabinet. The sample pot should also be positioned higher

    than the water source so that the waste side can drain away once the pump is stopped. If the

    installation location does not allow these criteria to be met, contact Greenspan Technology oryour local Greenspan representative.

  • 8/6/2019 Mini Analyzer Manual

    22/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 15

    If the analyser is mounted to a wall, then the wall is also a good mounting point for the sample

    pot. When the analyser is bench mounted the sample pot may be attached to the edge of the

    bench top.

    When choosing the mounting point be aware that two 25mm pipes will connect to the sample

    pot from outside the building. One is the sample from the pump and the other is the samplewaste outflow back to the water source. The position should also allow for the removal of the

    white cylindrical portion of the pot assembly for cleaning. Refer to section 13.1.7 Sample

    Delivery Pot for details on maintenance.

    6.4.2 Mounting the sample delivery module

    The sample pot has a stainless steel mounting

    bracket attached to it and this bracket has four

    holes for fixing to a vertical surface. Use the

    screws provided for fixing to a wooden surface or

    the expanding bolts for a masonry surface. If your

    fixing needs differ from these two choices

    appropriate fixings will have to be obtained from

    your local hardware supplier.

    6.4.3 Connecting The sample delivery andAnalyser

    Once mounted in position several connections

    need to be made to the sample delivery module.

    There are two connectors on the top of the pot

    assembly. These two connectors are of differenttypes and cannot be interchanged. One connector

    is for the sample temperature sensor and is a small

    multi-pin connector on a short cable. The second connector is part of a small air valve on the

    top of the assembly.

    Connect the cable exiting from the analyser on the right-hand side marked Sample Temp to

    the temperature sensor on the top of the pot assembly. The second cable marked Air Valve

    should be connected on the top of the pot but left free at the other end. This will connect to

    the pump control module to be installed later.

    There are two small polyethylene tubes that also attach to the top of the pot assembly. One isthe sample pick-up that connects to the analyser and this tube will have the number 71 (violet

    and brown) and is connected to the mating luer fitting directly in the middle of the top of the

    pot.

    The second tube is an air vent that can also vent a small amount of sample at times. This

    2.5mm internal diameter tube will be supplied as a separate item (not connected to anything)

    and will not have any fittings attached. Connect one end of the tubing to the 2.5mm barb

    fitting on the top of the sample pot and run the free end of the tubing to a location where it can

    vent liquid without harm. Do not place the free end in liquid or an enclosed place as it does

    need to be open to clean air.

    Once these cables and tubes have been connected to the pot assembly and analyser support

    them by fixing to the wall or bench using appropriate clamps.

    Fig 6.4: Sample delivery module

  • 8/6/2019 Mini Analyzer Manual

    23/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 16

    6.4.4 Connecting the pump to the sample delivery

    Once the pump has been installed the pipe has to be run to the analyser position for connection

    to the sample delivery pot and another pipe of the same diameter or slightly larger run from

    the sample pot back to the water source or some suitable drain. It may be convenient to run

    both pipes and the power cable along the same route. If this is done ensure that the dischargepoint for the sample pot outflow does not influence the intake to the pump.

    Greenspan personnel have found that it is easier to connect from the sample delivery pot

    assembly to the sample supply and discharge pipes close to the analyser location with 25mm

    flexible hose. The use of flexible hose close to the analyser reduces stress on the connections

    and makes the task of routing the pipes to the sample delivery module easier. Usually this is

    done from the entry point to the room where the analyser has been installed.

    Connect the sample supply from the pump to the side of the sample pot marked inlet and the

    discharge hose to the side marked outlet. Ensure all fittings that are threaded have been

    assembled using teflon thread tape and have been tightened firmly.

    Refer to section 3.6 and look at the diagram for the sample delivery system to verify that the

    system installed conforms with this diagram.

    6.5 Installing the sample delivery pump control module

    In most cases the pump used will be powered from the mains power at the site. The analyser

    outputs a low voltage signal to control this pump so some form of interface and control

    module is required to do that task The Greenspan external pump control module provides this

    with a switched 12Vdc 3A output.

    The control module can be fixed to the wall or bench next to the sample delivery module or

    fitted in the optional accessories cabinet. The control module comes fitted with a Din rail clip

    so all that is required for mounting is that a 50mm long piece of Din rail be fixed at the

    preferred mounting point and the module clipped to that rail.

    Connect the cable from the sample delivery module to the connector marked Air Valve.

    The connector marked Analyser on the control module and the connector marked I/P 4&5 /

    Pump on the analyser communications interface are connected using the cable marked in the

    same way.

    Connect the cable from the power supply/charger marked Sample control 12V In to the

    mating connector on the control module.

    There is one connector left and this is the switched 12Vdc out of the control module. This

    output is normally at 0V and goes to 12Vdc when the analyser enables the sample pump

    control signal. A cable is supplied to connect to this output and interface with a suitable mains

    power switching circuit (user installed) to enable the pump. Greenspan suggests a relay

    appropriate to the voltage and current demands of the pump with a 12Vdc solenoid coil. This

    method has proven to be simple and reliable. A qualified electrician may be required to install

    the mains power relay and pump wiring.

  • 8/6/2019 Mini Analyzer Manual

    24/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 17

    6.5.1 Checking the operation of the sample delivery

    On completion of the installation of the pump, sample delivery and control module the full

    sample delivery system can be tested using the computer and diagnostic routines in the

    analyser. Refer to section 13.1.9 Running Diagnostics for details on setting up the computer

    and connection.

    Warning: If the pump is turned on manually (not using analyser control) without the

    tubing fitted to the air vent port sample will be vented strongly from the port as the air

    vent is normally open.

    When connected to the analyser and at the diagnostic menu, select G Set/Clear External

    output. This will give a selection of four external outputs to switch. Number 3 - Ext1 power

    is the control output for the sample delivery. Ensure that the sample delivery is ready to

    operate and press 3. This will toggle the output on and enable the pump at the water source.

    Pressing 3 again will toggle the output off.

    Enable the pump and verify that sample flow is happening and that the flow rate is as expected

    with no leaks. When satisfied with the sample delivery disable the pump. Should problems

    be encountered requiring assistance contact Greenspan Technology or your local Greenspan

    representative.

    6.6 Connecting or adding external devices

    The analyser has considerable capability for external sensors to be fitted and can capture the

    data from those sensors for inclusion in the on-board data logging and transmission. Because

    of the large range of devices and connector types available it is not possible here to detail

    fitting any of these devices to the analyser. Should you wish to connect any external sensorsto the analyser inputs contact Greenspan Technology or your local Greenspan representative.

    6.7 Getting the Analyser commissioned

    Now that the analyser, communications and sample delivery are in place, it is time to

    commission and check that the analyser is functioning correctly. Refer to the Analyser32

    manual section 18 and 18.1 for detail on commissioning the system.

    While the analyser is running the commission regime look around inside the wet chemistry

    side of the analyser to check that nothing has become disconnected during shipping to cause

    any leaks. If a problem is noted and the correction of the problem is not obvious contact

    Greenspan Technology or your local Greenspan representative.

    6.7.1 Running some test schedules

    After the analyser has been commissioned and checked for leaks it is time to run a test of the

    full system. Refer to section 8 Schedules for details on setting up and running a schedule.

    Complete details on working with the User Interface are in the Analyser32 software user

    manual.

  • 8/6/2019 Mini Analyzer Manual

    25/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 18

    Setup a schedule to run all tests and the sample delivery pump using periodic timing with a

    period start of 0 hours and a trigger period of 1 minute. This will have the effect of

    retriggering the schedule at the next 1 minute interval after the schedule stops effectively

    causing the analyser to run continuously. This is a good method to use when testing the

    analyser. Be aware that this timing is not recommended to be used under normal

    circumstances as it increases the chemical usage considerably.

    Allow this schedule to run several times before stopping the analyser as it will assist in the

    stabilising of the sensors after shipping. Refer to section 12 Checking Measurements and

    Sensors for details on how to check the sensors and analyse the results.

    6.7.2 Testing a known sample

    Now that everything is working and stabilised a schedule could be run using a sample with

    known parameter values. This will verify the correct function of the analyser and sensors and

    give confidence that the installation has been completed effectively.

    To setup this schedule refer to sections 8.1.14 and 8.1.15 to disable the sample delivery pump

    control for the period of this testing. Disconnect the sample pick-up tube number 71 from the

    top of the sample delivery pot and place it in the known sample.

    Edit the schedule to contain the tests for the parameters you wish to verify ensuring that the

    fill/clean operations are unchecked and run the schedule. Allow the schedule to run for

    several passes if possible and then download the data. Check the data using Aquagraph and

    compare the results to those expected. Refer to section 12 Checking Measurements and

    Sensors for assistance with evaluating the results.

    When you have completed the verification testing reconnect the sample inlet tube 71 to the

    port on the top of the sample delivery pot. If this is not done water will be vented strongly

    from the open port on the sample delivery the next time the sample pump is operated.

    Should there be problems that you cannot resolve, contact Greenspan Technology or your

    local Greenspan representative.

  • 8/6/2019 Mini Analyzer Manual

    26/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 19

    7 COMMUNICATIONS

    Communications with the Mini-Analyser can be done using the following options:

    Modem and communications network.

    Serial link using standard serial cable.

    The first option can use standard telephone networks, cellular networks and radio or satellite

    phone systems as the communications link. The second option requires a site visit. A

    notebook computer is a convenient choice of computer for this purpose.

    7.1 Getting Started with the User Interface

    Analyser32 is a fully featured program that allows the user to do tasks like setting up

    schedules and calibrating sensors. Sections 8 through 10 of this manual give a brief overview

    of the basic use of the interface. For more in depth information please refer to the Analyser32user manual

    You may also find it useful to have a terminal emulator program such as HyperTerminal or

    Telix. These programs are useful for specific tasks such as upgrading firmware (see Section

    14), incorporating new sensors or functions and when setting the modem initialisation string

    (see below). The terminal emulator program can also be used for diagnostic testing of the

    Mini-Analyser. See section 13.1.9.

    7.2 Installing a Modem

    Some recommended modems for use with the Mini-Analyser:

    Zoom Fax/Modem 56Kx.

    US Robotics Courier V Everything 56k.

    Banksia mymodem 56v

    Diamond SupraExpress 336e and 56e.

  • 8/6/2019 Mini Analyzer Manual

    27/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 20

    7.2.1 Setting the modem initialisation string

    The modem initialisation string needs to be set should you change the type of modem used

    with your Mini-Analyser. This can only be done at the site by a direct connection. Contact

    Greenspan or your service agent before undertaking this task to obtain the correct initialisation

    string for your modem.

    Procedure:

    1. Connect a PC to the Mini-Analyser serial communications port using a standard serial

    cable.

    2. Run a terminal emulator program such as HyperTerminal or Telix. For more information

    on setting up these two applications refer to section 13.1.9. Running diagnostics.

    3. Set the communication parameters to 19200 baud, 8 data bits, 1 stop bit and no parity.

    4. When the terminal program is setup and a connection made with the Mini-Analyser areturn string will be displayed on the screen similar to 002.A034Y49NY10491520;. This

    sequence may vary with different versions of firmware but will be similar.

    5. Refer to the return string example in step 4 and note the characters that are underlined.

    Look at the return string that has appeared on your PC screen after connection to the Mini-

    Analyser and note these three characters. These refer to the version of firmware installed

    in the Mini-Analyser.

    6. Press the keyboard sequence *& by pressing SHIFT-8 and SHIFT-7 on the PC

    keyboard. The factory menu will appear.

    7. Turn caps lock on. Press M for Set Modem init string. The current init string will be

    displayed and you will be prompted to change it or not. Press Y to change.

    8. You can choose from the two default strings provided, one for normal, and one for cellular

    phones or you can press E to edit. Choose a string or edit the current string. To type in

    a completely new string use the edit option.

    9. Press Y when asked to save the init string. Press any key and then exit the interface.

  • 8/6/2019 Mini Analyzer Manual

    28/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 21

    8 SCHEDULES

    Schedules control the operation of the Mini-Analyser chemical laboratory and external

    sensors. They describe what tests are performed and when they are performed. By using the

    Mini-Analyser user interface software, an operator can choose the tests to be performed by a

    schedule. To set up a schedule, the operator of the Mini-Analyser performs the following

    steps:

    1. Create a new schedule.

    2. Selects the tests to be performed and their order.

    3. Chooses the activation mechanism for the schedule.

    4. Sets the timer if required.

    5. Enables the Schedule.

    8.1 Working With Schedules Using Analyser32

    8.1.1 Setting up a schedule

    From the Main Menu press the Setup button. From the Setup Menu select Scheduling. TheSchedule Setup window shown in Figure 8.1.1 will appear.

    To create a new schedule, press theNew button. If a schedule name already exists and youwant to modify it, select that schedule, then press theEditbutton. It is possible to create more

    than one schedule.

    Figure 8.1.1 The Schedule Setup window

  • 8/6/2019 Mini Analyzer Manual

    29/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 22

    8.1.2 Creating or editing schedules

    To access the schedule editor, pressNew orEditin the Schedule Setup window.

    8.1.3 Entering a name

    Enter a name for your schedule in the box located at top left of the window. If no name isentered then the Mini-Analyser will create one. The name can be up to 24 characters in

    length.

    8.1.4 Viewing and editing tests in the schedule

    The Tests in Schedule list describes all of the tests to be performed in the schedule as well as

    the fill and clean operations. The maximum number of tests permitted in one schedule is 20.

    Fill operations bring the sample from the sample source to the Mini-Analyser chemical

    laboratory ready for testing. Clean operations send leftover sample back to the sample source

    and clean the line. If fill and clean operations are included in the schedule they are always run

    at the beginning and end of the schedule respectively. To edit what tests are performed presstheEdit Tests button. For more information on this see section 8.1.14.

    8.1.5 The schedule duration

    The time it takes for the schedule to execute is displayed in the top right corner of the window.

    This time is calculated by the software, based on which tests are selected. This information is

    useful to help plan the timing of multiple schedules.

    8.1.6 Timing a schedule

    To set the times when the schedule will execute there are five options:

    Periodic- A periodic schedule will execute at fixed time intervals.

    Timetable - A timetable schedule will execute at specified times on a daily basis.

    Event The schedule will start in response to a variation detected in another parameter.

    External Trigger A signal from an external line will start the schedule.

    InitialSchedule The Initial schedule will run immediately the Analyser is started and

    run once only. Note. The Initial schedule will run once every time the Analyser is started.

    8.1.7 Enabling the schedule

    To enable a schedule, tick the Schedule Enabledbox. A schedule will not run without this boxbeing ticked.

    8.1.8 The scheduler priority

    Note: ANY SCHEDULE, WHICH IS DUE TO COMMENCE WHILE ANOTHER ISIN PROGRESS, WILL NOT RUN.

    If two or more schedules are due to execute at the same time, the schedule

    earliest in the list will run. The other schedule(s) will not run.

  • 8/6/2019 Mini Analyzer Manual

    30/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 23

    8.1.9 Using periodic triggering

    Periodic timing is set by selectingPeriodic from the Scheduled Times box in the schedule edit

    window. See Figure 8.1.9.

    The schedule will run first at thestart time. It will run again depending on what time is usedfor a Trigger Period. If the Trigger Period is larger than the schedule duration then the

    schedule will run at the next trigger time. If the Trigger Period is less than the schedule

    duration it will run again at the next trigger time when the scheduler is not already running, ie

    this schedule will run continuously. If more than one schedule is required to run, avoid

    conflict by making sure this is not the case.

    Figure 8.1.9: Setting triggering to periodic

  • 8/6/2019 Mini Analyzer Manual

    31/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 24

    8.1.10 Using timetable triggering

    Timetable Triggering allows the user to set schedules to run at specific times rather than after

    specified periods.

    In the Schedule Editor, choose the timetable option under the heading, Scheduled Times.Next, press theEdit Timetable button. This brings up the timetable editor. See Figure 8.1.10

    below.

    Add times to the timetable by entering them in theEnter Time box and then pressing the

    Add>> button. The new time will be added to the list in chronological order. To remove atime, highlight it in the list and press theDelete button. The Clear Allbutton removes all of

    the times listed.

    8.1.11 Using event triggering

    Event triggering is used when you wish to have a schedule run in response to a change of acertain magnitude in another parameter, for example temperature.

    You can select from a list of parameters to use as a trigger by pressing the down arrow in the

    Triggerbox. The magnitude of change that is required to cause a trigger is entered into the

    box labelledData Variation Value. When the sample data for the triggering parameter

    changes by more than theData Variation Value, either up or down, it will trigger an Eventand start the second Event triggered schedule. The data value at this time becomes the new

    reference point for theData Variation Value.

    For this option to work a schedule that tests the trigger parameter must be enabled and started.

    This provides the sample testing that can cause the event that will trigger the secondschedule.

    Figure 8.1.10: The Timetable window

  • 8/6/2019 Mini Analyzer Manual

    32/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 25

    8.1.12 Using the external trigger

    The external trigger option allows a signal from an external source to start the schedule. This

    trigger can be a negative going pulse or relay contact closure. The pulse should be a minimum

    100S in duration, and pull the Mini-Analyser trigger input from 5VDC to ground. The relay

    contact should provide a short duration ground (less than the period of the schedule) on theMini-Analyser trigger input.

    This trigger could be from a SCADA system in response to a pump being activated or some

    other event where you may wish to synchronise the sampling and testing by the Mini-

    Analyser.

    8.1.13 Using initial triggering

    The Initial triggering is used to ensure that any regime that has to be run every time the

    Analyser is restarted can be set to run automatically at start up.

    The Initial schedule will run immediately after the Analyser is started and will run once onlyafter that start sequence until the Analyser is stopped. When the Analyser is stopped and

    started again the Initial schedule will again run once for that start / stop sequence.

    Any schedule setup to use the Initial trigger function should be created as the first schedule in

    the list. This will ensure that the Initial schedule always runs first, irrespective of the timing

    or triggering methods used for other schedules as the Analyser processes the schedules starting

    from the first schedule in the list. Refer to Figure 8.1.13.

    Figure 8.1.13: Setting the schedule to Initial

  • 8/6/2019 Mini Analyzer Manual

    33/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 26

    8.1.14 Editing the tests for a schedule

    This allows the schedules to be setup with a variety of tests. TheEdit Tests window is shown

    in Figure 8.1.14.

    The tests available are shown in a list on the left hand side of the window. To add a test to the

    current list of tests shown on the right, highlight the test to add and press theAdd Test >>button. The same test can be added more than once to the schedule.

    When the test is added to the list it is also given a channel number. The channel number is the

    code used for the test in the data log. A schedule can be rearranged to execute in any order by

    using the following buttons:

    Push Test Up Move the position of a highlighted test up the list.

    Push Test Down Move the position of a highlighted test down the list.

    Delete Test Remove a highlighted test.

    Clear All Tests Remove all tests from the list.

    8.1.15 Adding fill and clean operations

    These procedures bring the sample to the internal inlets of the Mini-Analyser and then clean

    out the lines afterwards. The fill operations occur at the beginning of a schedule and clean

    operations at the end of a schedule. If the automatic box is ticked fills and cleans will beautomatically added according to which internal inlets are used.

    Figure 8.1.14: The Edit Tests window

  • 8/6/2019 Mini Analyzer Manual

    34/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 27

    The fills and cleans can be manually deselected by unticking the Automatic box. There are a

    few cases where this is desirable. For example testing procedures where a known solution is

    used in place of a sample.

    8.2 Starting a Schedule Using Analyser32

    The Start Mini-Analyserwindow, shown in Figure 8.2, appears after pressing the Startbutton

    in the main menu to turn the scheduler on.

    8.2.1 The current date/time display

    This display shows the settings on both the Mini-Analyser and computer. If the computer

    settings are incorrect they should be reset. Use the Windows control panel to reset the time

    and date information. The Analyser32 software will update the Mini-Analyser time and date

    information with that of the PCs when the scheduler is started. It is important when using

    External Triggering that the Mini-Analyser time be synchronised with the SCADAsystem time.

    8.2.2 Inserting comments

    Type text into the Commentfield to insert a comment in the data log. The length can be up to

    128 characters. Comments are very helpful when trying to locate events in the data log or

    troubleshooting problems with the Mini-Analyser. Aquagraph can display this comment

    information.

    8.2.3 Operator identification

    You can type text in the Operator 1 and Operator 2 fields to record who started the scheduler.These fields will accept 3 characters.

    Figure 8.2: The Start Mini-Analyser

  • 8/6/2019 Mini Analyzer Manual

    35/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 28

    8.2.4 Clear memory

    Clears all the logger memory. A warning will be displayed before the memory can be cleared.

    Note: Once the memory is cleared all data that was in the logger is erased.

    Ensure that all important data is retrieved before clearing the logger

    memory.

    8.3 Stopping a Schedule

    To stop the Mini-Analyser press the Stop button in the main menu. The Stop Mini-Analyserwindow will appear as shown in Figure 8.3.

    8.3.1 The last started display

    A display at the top of this window gives the date and time the Mini-Analyser was last started.

    8.3.2 Current date/time

    This shows the settings on both the Mini-Analyser and computer. If the computer settings are

    incorrect they should be reset via the Windows control panel.

    8.3.3 Inserting a comment

    Type text in the Commentfield to put a comment in the data log. The comment might be why

    the Mini-Analyser was stopped. The length can be up to 128 characters. Again, commentsare very helpful when trying to locate events in the data log.

    Figure 8.3: The Stop Mini-Analyser

    window

  • 8/6/2019 Mini Analyzer Manual

    36/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 29

    8.3.4 Operator identification

    You can put text in the Operator 1 and Operator 2 fields to identify who stopped the

    scheduler. This information is recorded in the data log. The field will accept up to 3

    characters.

    8.3.5 Confirming the end of scheduling

    Once yourStop Mini-Analyserwindow details are correct press the Stop button. If thescheduler is currently performing operations the stop command will take affect as soon as the

    Mini-Analyser is finished the current schedule.

  • 8/6/2019 Mini Analyzer Manual

    37/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 30

    9 ALARMS

    Alarms bring important events to the attention of the user and are normally used to notify the

    user that action is required. An Mini-Analyser can be set to phone a remote location in

    response to an alarm. Alarms can be used to trigger an external output.

    9.1 Setting Alarms With Analyser32

    To go to the Alarms window:

    1. From the main menu choose Setup.

    2. From the Setup menu chooseAlarms. TheAlarms window is shown in Figure 9.1.

    There are 2 categories of alarm:

    Sample alarms a result from a sensor test is outside preset limits.

    System alarms part of the Mini-Analyser system requires attention. The 4 main system

    alarms are triggered by:

    Liquid volumes from the chemical store.

    Memory level of the data log.

    Battery level.

    Water ingress - leakage of liquid inside the Mini-Analyser.

    Figure 9.1: The Alarms window

  • 8/6/2019 Mini Analyzer Manual

    38/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 31

    9.1.1 Setting the sample alarms

    Sample alarms are triggered when an alarm threshold is reached. The channel alarm set

    window allows for a high and low threshold to be set. Once triggered an alarm cannot be

    activated again until the measured value crosses a re-enable threshold. This is to prevent the

    unwanted activation of alarms due to small fluctuations.

    To set a sample alarm, highlight the parameter of interest in the Channels box and press Set.

    A Channel Alarm window will appear as shown in Figure 9.1.1.

    To set an alarm, type in values for the Trigger Value and theRe-enable Value then tick the

    correspondingEnabledbox. Press the OKbutton to confirm the change.

    As a guide:

    The low alarm re-enable value should be higher than the low alarm trigger value.

    The high alarm re-enable value should be lower than the higher alarm trigger value.

    9.1.2 Memory alarm

    The memory alarm area is located in the top right corner of the Alarms window.

    A memory alarm will be activated if:

    The Memory alarm is enabled, and,

    The amount of stored data in the memory goes above the trigger level.

    Figure 9.1.1: The Channel Alarm window

  • 8/6/2019 Mini Analyzer Manual

    39/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 32

    When the memory reaches full capacity the Mini-Analyser will stop. The data log should

    be downloaded and then the memory cleared prior to restarting the Mini-Analyser.

    The memory alarm can also be used to trigger the external output by ticking the Memory

    External Outputbox.

    9.1.3 The battery alarm

    The battery alarm is located in the Channels box. The battery alarm does not have a re-enable

    value and can only be triggered from a low alarm.

    9.1.4 The solutions alarms

    The solution alarms are located in the middle of the alarms window and can be set by

    highlighting the relevant solution and pressing the Setbutton, a window like the one shown inFigure 9.1.4 will appear.

    The default value for the trigger is set to 10% at the factory.

    The Mini-Analyser will cease to operate when the volume remaining reaches 5%of the bottle value.

    The Solution alarms have a common external output enable box located in theAlarmswindow. If this box is ticked any solution alarm will trigger an external output. See section

    9.1.6.

    Figure 9.1.4: The Solution Alarm window

  • 8/6/2019 Mini Analyzer Manual

    40/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 33

    9.1.5 Water or liquid ingress

    This alarm detects water ingress into the internal sensor compartment of the Mini-Analyser.

    This alarm is triggered by a sensor that has two moisture-sensing electrodes and is located in

    the drip tray at the bottom of the chemistry module.

    Note: A water ingress alarm will cause the Mini-Analyser to cease operation.

    Fingers or metal objects that come into contact with the sensor can trigger this alarm.

    9.1.6 Enabling an external output

    The Mini-Analyser has an external output that is enabled in the software. This external output

    is in the form of a 500ms active low pulse. Other software or hardware can use this as a

    trigger for a program or event. This could be in the form of making a bell ring or a light shine.

    The steps to enable an external output are:

    1. From the Main Menu choose Setup.

    2. From the Setup Menu chooseAlarms.

    3. Choose a source for the alarm ie pH, phosphate, Battery.

    4. Highlight the source by clicking it and press the Setbutton.

    5. Tick theEnable External Output on Alarm box.

    The output (a 500 ms pulse active low) will occur if:

    An alarm occurs, and,

    The source of the alarm has itsEnable External Output on Alarm box ticked.

    9.1.7 Alarms and the data log

    When any alarm is triggered details of the alarm are written to the data log. The details logged

    are:

    The source of the alarm. For example, Nitrate, pH etc.

    The value which caused the alarm to trigger.

    The date and time of the alarm.

    9.1.8 Enabling the alarm system

    For any of the individual alarms to activate the Global Alarms Enabledbox needs to be ticked

    in the bottom right corner of theAlarms window.

  • 8/6/2019 Mini Analyzer Manual

    41/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 34

    9.1.9 Dialout on alarm

    The alarm dialout function allows the user to receive notification of alarms from remote sites.

    This feature is accessed by pressing theDialout Phone Numbers button in the alarms window.The window is shown in Figure 9.1.9.

    The user can specify up to 4 phone numbers for the Mini-Analyser to dial in response to an

    alarm. If an alarm is present the Mini-Analyser will look to see if the current time falls within

    any of the time windows. If one is found the Mini-Analyser will dial the corresponding phone

    number. For a dialout to happen a time window must be set.

    To specify the amount of time the Mini-Analyser modem is powered up after an alarm dialout:

    1. From the main menu choose Setup.

    2. From the Setup menu choose Communication Times.

    3. Enter the duration of power-up in minutes and seconds into the box labelled

    Duration of Power up following an Alarm Dialout.

    4. Press OK.

    This time is set to give the user a period in which to dial up the Mini-Analyser and check the

    cause of the alarm dialout.

    Figure 9.1.9: The DialOut Phone Numbers window

  • 8/6/2019 Mini Analyzer Manual

    42/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 35

    An alarm dialout will occur if:

    There is no direct connection with the Mini-Analyser.

    A valid telephone number and time window are set.

    The current time is inside a dialout window.

    An unacknowledged alarm exists.

    The Mini-Analyser will attempt to call the corresponding phone number a maximum of 3

    times.

    9.1.10 Receiving dialout alarms

    To receive a dialout alarm at a remote computer several conditions have to be met. The

    telephone number that the Mini-Analyser dials must be for a modem connected to the remote

    computer. This computer must also be powered on, running Windows and have theappropriate program (SmartStandby) running in the background. For further details on using

    SmartStandby contact Greenspan Technology or your local Greenspan representative.

    9.2 Clearing Alarms With Analyser32

    TheAcknowledge Alarm window, shown in Figure 9.2, is accessed by pressing the

    Acknowledge button in the alarm display area of the Main Menu.

    Figure 9.2: The Acknowledge Alarm window

  • 8/6/2019 Mini Analyzer Manual

    43/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 36

    9.2.1 Clearing the alarm

    Pressing the Clear Alarm button will bring up a message asking the user to confirm that the

    alarm is to be deleted. If the answer isyes the alarm is cleared and deactivated. A no answerwill return the user to the main menu.

    After checking the Mini-Analyser to determine and rectify the cause of the alarm you will

    need to re-enable the alarm. Refer to section 9.1.

    9.2.2 Cancelling a dialout

    Press theDialout Cancelbutton to stop the Mini-Analyser from continuing a modem dialoutsequence.

    10 DOWNLOADING DATA

    The data log is a record kept by the Mini-Analyser detailing the results of scheduled tests,

    alarms and reagent levels. This data can be downloaded for graphical analysis usingAquagraph or archived for future reference.

    10.1 Downloading Data using Analyser32

    To download data from the Mini-Analyser press theData button from the Main Menu. The

    Download Data window is shown in Figure 10.1.

    The data files that will be stored on the computer are in a Greenspan data format. These can

    be converted into a comma separated format for use with many common spreadsheets usingthe Aquagraph utility supplied with the Mini-Analyser.

    Figure 10.1: The Download Data window

  • 8/6/2019 Mini Analyzer Manual

    44/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 37

    Note: The Scheduler does not need to be stopped to download the data file.

    10.1.1 Downloading the entire data log

    The first option labelledAll Stored Data will download the whole data log. This option isuseful for archive and backup prior to clearing the data memory.

    10.1.2 Downloading new data

    SelectingData Since Last Downloadwill gather only the data recorded since the lastdownload. File markers are inserted where the data was last downloaded. These markers are

    set at the end of the file once data has been downloaded.

    10.1.3 Downloading data between dates

    When theData between Following Dates option is selected the edit boxes forData From and

    Data To are made available. The default From and To dates are the log start and end dates.

    Data can be downloaded from between any dates. The file markers for the data since last

    download are not adjusted when using the data between dates option.

  • 8/6/2019 Mini Analyzer Manual

    45/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 38

    11 CHEMICAL STORE

    The Mini-Analyser chemical store contains a variety of chemicals that can change from one

    installation to another depending on the water parameters being measured. Some of these

    chemicals are very corrosive, others can be hazardous to your health, which means that all of

    them need to be handled with care and attention all of the time so that you normally have safework practices. Rubber gloves and eye protection should be worn at all times when handling

    the chemicals to reduce the risk of contact and most importantly, work carefully.

    The Mini-Analyser is supplied with two complete bottle sets. This is done so that chemicals

    do not have to be poured into bottles at the site. With two sets, one can be in use at the Mini-

    Analyser and the other at the laboratory being refilled. The empty bottles should be washed at

    the laboratory before being refilled. When a chemical needs replenishing, the empty bottle is

    taken out and a new bottle installed using the procedure below. The empty or near empty

    bottles are then taken to the laboratory for refilling. This puts the hazardous handling of the

    chemicals into the laboratory and reduces the risk of accident at the Mini-Analyser site.

    The chemicals are connected to the Mini-Analyser by polyethylene tubing and Luer fittings.

    This type of fitting is used widely and is very reliable and chemical resistant. A numbering

    system is used to identify chemicals and tubes. When replacing reagents always check that

    the labelled tube has the same colour code and numbering as the bottle it connects to.

    11.1 Changing Solution Bottles

    All of the chemicals have a number and colour allocated to them and this numbering follows

    the colour code shown in table 11.1a.

    Table 11.1a: Chemical colour code

    Number 0 1 2 3 4 5 6 7 8 9

    Colour Black Brown Red Orange Yellow Green Blue Violet Grey White

    Table 11.1b: Chemicals and their allocated numbers

    Chemical Number Chemical Number

    pH/EC Standard 12 Phosphate 1 21

    Standard 1 15 Phosphate 2 20

    Standard 2 16 Phosphate 3 53

    Cleansing solution 1 50 Phosphate 4 54

    Air 51 Sample 1 71

    Nitrate 55

    Ammonia 1 23

    Ammonia 2 49 Combined Waste 87

    Ammonium 24 Non-Modified Waste 90

  • 8/6/2019 Mini Analyzer Manual

    46/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 39

    Procedure:

    14.Stop the scheduler if it is running, and wait for any remaining steps to finish.

    15.Put on safety equipment - a protective coat, gloves and eye protection.

    16.Connect a PC to the Mini-Analyser. Run Analyser32 and check which reagent requires

    replacement.

    17.Locate the bottle that requires replacement. Disconnect the tubing luer fitting.

    18.Remove the bottle with dropper cap from the Mini-Analyser.

    19.Unscrew the cap from the new bottle. Unscrew the dropper cap from the bottle being

    replaced.

    20.Check that the colour code on the bottle matches the colour code on the inlet tube.

    21.Swap the caps on each bottle. DO NOT TOP BOTTLES UP, ALWAYS REPLACE WITHFRESH BOTTLES.

    22.Place the new bottle back into position. Connect the tubing to the bottle and tighten the

    luer fitting to be finger tight. Do not over tighten this fitting. Firm pressure is all that is

    required.

    23.Reset the volume level for the replaced solution in the chemical store, using the PC. Refer

    to Section 11.2 in this manual to do this using Analyser32.

    24.Do a visual check that the bottle volumes match the recorded bottle volumes displayed onthe PC and if necessary, edit the volume level in the chemical store.

    25.Visually check the level of the waste bottle to see if it requires replacement.

    26.You may now restart the scheduler or start logging if desired.

    27.Exit the user interface, disconnect the serial cable from the Mini-Analyser.

  • 8/6/2019 Mini Analyzer Manual

    47/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 40

    11.2 Updating Reagents Using Analyser32

    The Mini-Analyser keeps record of the levels of solutions available in the chemical store.

    This is done to notify you when they are close to empty and also to prevent the Mini-Analyser

    from running when they are completely empty.

    Note: The chemical store can only be edited or updated with the scheduler OFF

    To access the chemical store database:

    1. From the Main Menu choose Setup.

    2. From the Setup Menu choose Chemical Store.

    The Setup Chemical Store window shown in Figure 11.2a will appear.

    If a complete refill of all bottles has been done you can quickly reset all volumes to full by

    pressing the Set All Solutions as Fullbutton. A message will appear asking you to confirm

    this.

    To change the level of an individual solution highlight that row by clicking it with the mouse

    and press theEditbutton, theEdit Reagentwindow shown in Figure 11.2b will appear.

    Figure 11.2a: The Setup Chemical Store window

  • 8/6/2019 Mini Analyzer Manual

    48/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 41

    If the bottle has been completely refilled tick theFull Solution Bottle Installedbox. To set the

    level to something else click on the Current Levelbox and edit the data. In this window youcan also edit the alarm volume level in the Trigger Volume box and theActive Constituent

    Concentration if applicable.

    The concentration value is used in calibration procedures and analysis calculations. These

    values would not normally be changed unless the concentration of the standard is changed.

    Press the OKbutton to confirm your changes.

    NOTE: Changing the active constituent values will change the data results. DO NOT

    change these values unless you have been instructed to do so.

    11.3 Waste Disposal

    11.3.1 Unmodified waste

    Unmodified waste has the same composition as the source. Therefore for disposal it can be

    pumped back to the water source, to a drain or combined with modified waste depending on

    the installation setup.

    11.3.2 Modified waste

    28.Modified waste is collected in a 20L plastic container. This waste must be treated by pH

    neutralisation before disposal as the pH of the combined waste may be from 1.5 to 12

    depending on the sensor installation in the Mini-Analyser.

    Figure 11.2b: The Edit Reagent window

  • 8/6/2019 Mini Analyzer Manual

    49/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 42

    12 CHECKING MEASUREMENTS AND SENSORS

    It is important to be able to verify the accuracy of measured results. This is normally done

    with the help of standards. Verification also tells you whether the sensor is operating properly

    or if it may need maintenance or replacing. View the sensor check values using Analyser32 or

    by viewing the data file using Aquagraph. Using Aquagraph to view a downloaded data file isbest as it allows you to see many data point and trends that cannot be seen with Analyser32.

    When analysing data and encountering problems, refer to section 15 Troubleshooting for

    assistance in locating the problem. For further assistance contact Greenspan Technology or

    your local Greenspan representative.

    12.1 Dissolved Oxygen

    Each time the Mini-Analyser does a test on the DO a correction to the sample result is made

    by applying a single point calibration. This is done to compensate for changes in the response

    of the electrode. The DO check result does not have this correction applied to it and can beused as an indicator of the sensor condition.

    Under normal operation the Mini-Analyser check will give a result for the dissolved oxygen of

    the air bubbled deionised water. This value is 105%.

    The Mini-Analyser check value can drift over time. This drift is usually a result of a change in

    sensor characteristics. Once this result has drifted out of an acceptable range (20 %), then the

    sensor may have to be cleaned, re-calibrated or replaced. Re-calibration may be achieved by

    following the sensor calibration procedure (see section 12.3). If results for the sample vary

    significantly from expected and the check results are acceptable then other causes may need to

    be investigated. Contact your Greenspan Technology Service agent.

    12.1.1 Analysing the DO data

    The easiest way to check if the analyser is functioning properly is to use Aquagraph to graph

    each process individually and check for unusual values.

    What to graph

    Graph both DO data series, DO 1 and DO check, on the same graph.

    Checking results

    When analysing the results, you should look for a steady check value around 105% and thedata returns expected results.

    What should the graph look like if the

    Analyser is working correctly?

    The check values in the graph you obtain

    should look something like this.

  • 8/6/2019 Mini Analyzer Manual

    50/115

    UM-010-0058 MAN GREENSPAN TECHNOLOGY PTY LTD 43

    What if the graph does not look like this?

    The check value should remain steady. If this is not the case, ie drift, sudden variation or

    unusual values, this may indicate that there is a problem. Data results will vary between sites

    and over time but extreme variation, ie sudden peaks, indicates a possible problem. In either

    case further action should be taken to discover the cause of unusual data.What to do if the graph is abnormal

    If there is an obvious problem that does not seem to have a cause that you can determine,

    contact your Greenspan Service Agent.

    Note that data from a period of time with an obvious problem may not be valid and including

    it in a data summary may lead to incorrect inferences.

    Some typical graphs

    Both the check value and the datavalue are very similar in value in graph

    1. As the check should normally be

    105% the analyser has applied

    compensation to the data result giving

    a value to the data that may be

    incorrectly high. The DO should be

    re-calibrated when the check value is

    this incorrect. Note that this has been

    done near the end of the graph and the

    data is now valid.

    In graph 2 the DO check value has

    gone very low and the automatic

    compensation has caused the result to

    go very high. The most likely reason

    is that the analyser has run out of

    de-ionised water and is still running

    (incorrect volume levels in the

    chemical store). When the DI water is

    refilled the check value returns nearnormal and the data is once again

    valid.