Portable gasanalyzer POLARIS 1001 series"Methane- CH4" Usermanual KDYUSH.413327.017 UM

Transcript of Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User...

Page 1: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

Portable gasanalyzer

POLARIS1001 series"Methane-CH4"

UsermanualKDYUSH.413327.017 UM

IGMInstruments Saint-Petersburg

2015

Page 2: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

Table of contentsIntroduction 31 Description 41.1 Application notes 41.2 Specification 41.2.1 Common specification 41.2.2 Performance 51.2.3 Supply 51.2.4 Data output 51.3 Delivery set 61.4 Description and operation 61.4.1 Gas analyzer principle 61.4.2 Functional diagram 81.4.3 Construction design 91.5 Marking and sealing 101.6 Package 102 Using of gas analyzer 112.1 Preparation for use 112.1.1 Precautions 112.1.2 Preparations 112.2 Proper use 112.2.1 Turning on and measuring 112.3 Maintenance 132.3.1 General notes 132.3.2 Visual inspection 132.3.3 Zeroing and span gas calibration 133 Troubleshooting 133.1 Troubleshooting list 133.2 Malfunction self-test 144 Storage 145 Transportation 146 Warranty 157 Repairing 15

Appendix AChecking procedureAppendix B RS-232 Communication protocolAppendix C Gas feeding diagram required for zeroing and span gascalibration

2

Page 3: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

3

Introduction

This User manual is designed to study the arrangement, design and operation of a portable gasanalyzer "POLARIS" 1001 series "Methane CH4" (hereinafter, gas analyzer orGA).

TheUsermanualcontainsthebasictechnicaldataandcharacteristicsofGA,recommendationson maintenance, as well as other information necessary for the properoperations.

Example entries in order and documentation of other products: "Portable gas analyzer "POLARIS"1001 series" Methane CH4 "KDYUSH. 413327.017 TU."

The changes can be made in electronic circuit and construction design of GA during themanufacturing process without affecting technical and metrological characteristics of device. The manufacturer reservesthe right not to include these changes in Usermanual.

Page 4: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

4

1 Description1.1 Application notes

Main features- GA has built-in sensor and internal rechargeablebattery.

- GA is designedfor: measuring the volume fraction of methane at the approximately safe level ofimpact (ASLI) in

the ambient air of populatedareas; detecting of methane leaks and output alarms in excess of the measured the value ofthe

threshold.- Main application of GA is mobile environmental laboratories. The gas analyzer is designed tobe

used in non-hazardousareas.- Information about the measured values of methane volume fraction displayson:

digital LCD with up to 0.1 ppmresolution. green and red LED bargraphs; external PC via RS-232C serial interface(additionally).

Примечание: на лицевой панели ГА и ЖКИ единица измерения объемной долиметана "млн-1" обозначена"ppm".

Способ забора пробы – принудительный. Метод измерения концентрации метана- оптический(инфракрасный).

По устойчивости к воздействию климатических факторов ГА соответствуетисполнению УХЛкатегории3.1.поГОСТ15150.ГАявляетсяпереносным,восстанавливаемымвусловиях предприятия-изготовителяизделием.

Конструктивно ГА имеет моноблочное исполнение со встроеннымпервичнымпреобразователем и аккумуляторной батареей, защищенное от попадания внутрь ГАтвердых тел и воды. По степени защиты ГА соответствует IP 32 по ГОСТ14254.

ГА может эксплуатироваться в периодическом режиме с необходимыми поусловиямэксплуатации включениями ивыключениями.

Operatingconditions:

Temperature range,0С - 10 ...40Relative humidity at 25 0С, Atmosphericpressure:Ingress Protection

45 ...9584,0 107 kPa (630800mmhg)

- Ingress ProtectionIP 32

Composition of testenvironments:

Air environmentcomponent Concentration values of components underoperation conditions

1 Oxygen 21 %Vol.2Nitrogen 78 %Vol.3 Сarbondioxide 0 % Vol. ... 1,0 %Vol.4 Сarbonicoxide 0 ppm ... 100ppm5Methane 0 ppm ... 10000ppm6 CnHm heavy hydrocarbons (propane cross sensitivity) 0 ppm ... 1000ppm7Dust 0 mg/m³ ... 2,0mg/m³8 Sulfurvapor 0 mg/m³ .... 0,31mg/m³

1.2 Specification

1.2.1. Commonspecification

User, 04.06.15,
Перевод этого текста отсутствует в англ.версии руководства для модели 1011, он был удален намеренно?
Sergeev Stas, 05.06.15,
да, убираем
Page 5: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

5

1) Outline dimensions H x W x L, mm, not morethan: 140 x 120 x 2802) Weight, kg, not morethan: 43) Index of protection acc. IEC529-89: IP 324) Supply is carried by embedded rechargeablebattery.5) Average life time 5years.6) MTBF, without regard to batteries replacements, hours, not lessthan: 10000

Operating principle: Optical(IR)Categories,operating,storageandtransportationconditionsastoenvironmentclimaticaspects influence:

NF 3.1 category according to GOST 15150 (32 ГОСТ14254).GA is portable device, recoverable under the manufacturer'sterritory.

1) Average service life of GA must be at least 5 years.2) MTBF of GA without regard to replaceable elements of batteries must be h, at least 10,000

1.2.2. Performance1) Gas samplingmethod: Forced-feed (gaspump)

2) Gas feeding. Flow rate, l/min, not lessthan: 23) Measuring range of methane volume fraction,ppm: 0 20004) Indication range of methane volume fraction,ppm: 0 99995) LCD resolution, methane volume fraction,ppm: 16) Accuracy о, methane volume fraction,ppm: (5 +0,05СIN) Where СIN – GA input methane volumefraction.7) GA has sound and light alarm indication with thresholdsetup:- by manufacturer, methane volume fraction,ppm: 100- by user, in range, methane volume fraction,ppm: 0 20008) Alarm operation accuracy, methane volume fraction,ppm: 59) Additional temperature error, on each 10

оС: 0,5о

10) Additional error by impact of non-measured components in airenvironment 0,5о

11) Additional pressure error, on each 3.3kPa: 0,5о

12) Time of response (Т90 ), sec, not morethan: 413) Operating period without requalification, month , not morethan: 614) Recovery time to gas overload impact (up to 100% of methane concentration during 10 minwithcontinuous gas feeding), sec, not morethan: 2015) Warm-up time, min, not morethan: 5

1.2.2. SupplyInternal rechargeable battery output voltage,V 10.60.6Minimumworkingtimeonfullychargedbattery(batterysuppliedfromthechargerconnectedtothe 220 V 50 Hz during 16 hours) , hours, not lessthan: 6

1.2.3. Data outputMeasured valued of methane concentration is displayed:

- On LCD with 0.1 ppm resolution for 1 ppm range and with 1 ppm resolution for 100…9999 ppm range.

- On green LED bar graph in range 0…2000 ppm with regard to scale coefficient.- On red LED bar graph in range 0…2000 ppm with regard to alarm threshold settings.

GA provides data on the serial digital interface RS-232C. Communication protocol and transmission parameters are given in Appendix B.During operation, GA may provide additional light and sound signals: LEDs glow red bar-graph LED and buzzer - methane concentration above the threshold level (100 ppm - the default factory setting).

NOTES: On red LED bar graph displays the readings of methane concentration in the same manner as in the green LED bar graph - see article2.2.1.5.

1.3. DeliverysetTable 1 Deliveryset

Sergeev Stas, 06/05/15,
да
User, 06/04/15,
В англ.переводе модели 1011 Предел допускаемой дополнительной погрешности звучит как Additionalerror - ??
User, 06/04/15,
В англ.переводе модели 1011 номер ГОСТа 14254 изменен на 15150 - ??
Page 6: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

6

Наименование Обозначение Количествошт.

Примечания

Газоанализатор«Полярис» KDYUSH. 413327.017 1Charging unit АС-220-S-15-1500 TU6589-004-39491876-99 1

RS-232 serialcable 1User manual KDYUSH. 413327.017 UM 1Calibration procedure Appendix А

1.4. Description andoperation

1.4.1. Gas analyzerprincipleTheprincipleofGAoperationisbasedonselectiveabsorptionofinfraredradiationmethane molecules in the wavelength 3.2-3.4 umregion.

Infrared radiation of source (black body) passes through the measuring gas cell, which ispumped by sample gas. Than the IR light gets into electrically controlled spectral filter, which transmission spectrumismatchedtotheabsorptionspectrumofthemeasuredgas,andthenentersaphotodetector (receiver),stabilizedinsensitivitytotemperaturebythermoelectricbattery(Peltierelement).

Thetransitionofaspectralfilteraccordingtoanappliedelectricalcontrolsignaleither superimposed or not coincide with the absorption spectra of methane. Thus, at the receiver, a signal with a frequency modulation control appears, and its modulationamplitude is described in accordance withequation:

Iр - Iо / 0.5*( Iр + Iо) = 1-ехр { - [ К(р) - К (о)]СL}, (1)

where:К () - absorption coefficient at a givenwavelength;L - optical length of thecell;С - the measured concentration ofgas;Ip-theamplitudeofthesignalonreceiveratthemomentofmatchingelectricallycontrolledspectral filter with the spectral absorption region ofmethane.Io - the amplitude of the signal on receiver at the moment of mismatching electrically controlledspectral filter with the spectral absorption region ofmethane.

Unknown concentration of the gas is givenby:

С = -Ln[ 1-((Iр – Iо) / 0.5*( Iр + Iо))] /(L [ K (р) - К (о)]), (2)

Asusedspectralcorrelationregistrationtechniqueeliminatestheinfluenceofwatervapor,dirt andotherelementsofopticalnonselectiveinfluences(includingotherhydrocarboncompounds)aswell as the instability associated with the optoelectronic elements of GAitself.

Functional structure is shown onFig.1

Sergeev Stas, 06/05/15,
да
User, 06/04/15,
Транслитерация с русского
Page 7: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

Rechargeable battery

9

11Gassample

23

6

MCU

74

+

51

Gaspump

10

8

7

1.4.2. Functional diagram

RS232

Fig.1 GAprinciples

Operation is controlled by a microprocessor unit (7) MCU, which generates a clock pulsesequence to control the master oscillator (6). MCU digitalizes the amplitudes of signals coming from the synchronousdetectorunit(4).Theclockgenerator(6)enablestheformationofacontrolvoltagetothe electrically controlled spectral filter (5), and a synchronous timing detector (4).

Infraredsource(1)isenergizedbystabilizedcurrentfromthepowersupplyvoltageunit(9).The infraredsourceemitsintheband,thatoverlapsdesiredandreferencesignalwavelengths.Reciever(2) passes radiation in the band in the bandwidth 3.2-3.6 um, provided by interference filter, which isdirectly mountedtothehousingofphotodetector.Temperaturestabilization(3)byPeltierelementisintegrated and controlled by current going from thermal stabilization unit (8). The thermal stabilization controlsignal goes from thermistor mounted inside ofphotodetector.

MCU produces mathematical processing and calculates concentration of the target gasin accordancewiththeequationgivenabove.Informationabouttheconcentrationoutputsontheserial port (10) (MAX 242 chip) in the standard RS-232 protocol (see Appendix1).

Information about the measured concentration is displayed on indication unit (11), consisting ofa matrix LCD and two LED bar indicators - green andred.

The LCD can display information about the operability of GA, based on measurementsof additional signals and the operatingvoltage.

Power supply unit (9) provides GA with an input voltage +(10.60.6) VDC from thebattery.

Page 8: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

8

1.4.3. ConstructiondesignGA has design as it shown on Fig.2.Following parts are located in front side of gasanalyzer:

LCD display with backlighting Bar graph LEDs (red andgreen) LEDs for indication of embedded gas pump and audiblealarm LEDs for indication of current “2.0”, “20.0”, “200.0ppm”. The keyboard with followingbuttons:

- on-off power- on-off pump

on-off buzzer - save set up values -"ВВОД" set up zero - > 0< setting up alarm threshold -"ПОРОГ" values selection - connector for the charger and serial RS-232plug.

Page 9: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

9

Fig. 2 GAdesign

Input and output fittings are located on the side face ofGA.Gas sampling device with probe connects to quick-detachable inputfitting.To replace the battery the bottom cover can be dismounted by removing the two barsand unscrewing four fixingscrews.

1.5. Marking andsealingFace sidemarking:- Device name;- Manufacturertrademark. Side facemarking:- Serial number and productiondate. Device is sealed by themanufacturer.

1.6. PackageGA is packed in a bag, then in carton box. The box with the GA keepstechnical documentation in a plastic bag and charger. The box fits in an additional polyethylene bagfortransportation.

Page 10: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

10

2 Using of gasanalyzer

2.1.Preparation foruse

2.1.1. PrecautionsUsingofGAisallowedonlyforpersonswhoareappropriatelytrainedonsafety,whohave studiedthisUMandhavequalificationgrouponelectricalsafety.

Whenworkingwiththegasmixtureinvesselsunderpressureitmustbeproceedwithfullcompliance to gas process, pipeline and transportationsrequirements.

2.1.2. Preparations2.1.2.1. Unpack theGA.2.1.2.2. Charge the battery by charger, which comes in delivery set, with the followinginstructions:

- Remove the cover on the front of thecharger- Connect the charger connector (connector has a key and admits a unique ability toconnect).- Connect the charger to power line 220 VAC 50 Hz, then press the power button on thecharger.TheGAshouldbelityellowLED(REMcharge)intheareaofthechargeconnector,andthenduringthe chargingprocess,theyellowLEDgoesoffandthegreen.BothLEDsareextinguishedwhenthebatteryis fullycharged.TheyellowLEDshouldbehighlighted(rapidchargestage).Duringthechargingprocess,theyellowLED goes off and the green LED lights. Both LEDs are off when the battery is fullycharged.Lighting of red LED indicates malfunction of the battery or open circuit ofcharge. Battery charging time is according to the instructions to thecharger.

2.1.2.3. Turn off the charger and disconnect it from power line 220 VAC 50 Hz, then disconnect thecharge fromGA.

2.2. Properuse

2.2.1. Turning on andmeasuring2.2.1.1.Turn on GA by pressing the button In LCD screen it displays “CH4 Working” during 20 sec. Then “WARM UP” appears in upper raw andduring this time squares sequence from 1 to 8 is displaying on lowerraw.After that measured concentration readings displays in upper raw (1-5 digits with “ppm”symbol) In

lower raw GA indicates operation information about the checkupresults.

Example:

Where:ХХХХХ ppm - measured methane concentration value in tworaws:

- in upper raw – with 2.5 sec accumulation and updatetime- in lower raw - with 0.2 sec accumulation and updatetime

OК - electrical circuit functionality test is successfully completed and battery voltage isvalid.NOTE:ThesymbolsequencelistedaboveindicatesthatGAisworkingproperly.Seefor troubleshooting list in

article 2.3.4 if other symbolsappears in right downcorner.

Обозначение символов наЖКИ:ХХХХХ ppm - измеряемая концентрацияметана100 - порогсигнализацииOК - контроль работоспособности электрической схемы и напряженияаккумуляторнойбатареиПоявление (высвечивание) на ЖКИ упомянутых выше символов свидетельствуетобисправности

электрической схемыГА.

XXX.X

ppmХХ.Х

Page 11: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

11

Появление (высвечивание) на ЖКИсимволов свидетельствует о неисправности электрической схемыГА.

Появление (высвечивание) на ЖКИсимволов свидетельствует о необходимости заряда аккумуляторнойбатареи.

Появление (высвечивание) на ЖКИсимволов свидетельствуетонеобходимостидополнительногопрогреваприборадлявыходана требуемый метрологический уровень. Использование прибора возможно послеокончанияпрогрева и высвечиваниясимволов

2.2.1.2. Complete warm-up time is 5min.NOTE:Beforestartingupthefirstmeasuringthatisrecommendedtofollowzeroingprocedure,see section

2.3.3 fordetails.2.2.1.3. Move the end of the rubber tube running from the inlet fitting of the gas cell, toa controlled point of the

airvolume.2.2.1.4. Feed the gas cell of GA with 2 l/min flow of analyzed air by turning on the embeddedgas pump. To do this

requires topress button. When gas pump starts to work, red LED nearthe buttonis lighting. Press the button again to turn off the gas pump when it's necessary, that will off the red LED aswell.

2.2.1.5. Check the measured value of methane concentration of on LCD screen ofGA.On the green LED bar graph GA displays concentration with the number of lightingsegments proportional to volume fraction of methane with regard to scale coefficient. Full scale is indicatedby highlighting one of three green LEDs marked as "2.0 ppm", "20.0 ppm", "200.0ppm".On the red LED bar graph GA displays concentration the same ways as on greed LED bar graph,but the scale is specified by alarm threshold settings. The threshold of alarm is always as half scale of redLED bar and the median segment is constantlyhighlighted.By default, a setting of alarm threshold is 100 ppm, thus the full scale for the red LED bar graphis 200 ppm.The delay time of displaying information on red LED bar is 0.2sec.

2.2.1.6. Working time on fully charged battery without recharging is 6hours. When the level of methane alarm concentration (alarm threshold settings) is reached, the buzzer starts to sound and the number of highlighted red segments exceeds half of scale (middle segment of the scale at the same time goes off). The audio signal can be switched off if it is necessary by pressing the button then the LED nearest to the sound button (which lights up when you turn on GA) goes off.

XXXX ppm100 Fail

XXXX ppm

100 BATT

XXXX ppm

100 ETH

XXXX ppm

100 OК

Sergeev Stas, 06/05/15,
Нет, нужен перевод. Это специфика работы с 1001
User, 06/04/15,
Перевод этого текста отсутствует в РП модели 1011, его убрали намеренно?
Page 12: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

12

2.2.2. Reading data onPC1. ConnectGAtocompatiblePCwithRS-232cable.2. To work with GA you can use any software with ASCII charset and keyboardinput.3. PC comportsettings:

9600 baudrate 8-bitmessage, 1 stop bit, noparity, Communication protocol - see AppendixB.

2.3.Maintenance2.3.1. Generalnotes

MaintenanceshouldbeperformedinordertoensurethenormaloperationofGAduringits lifetime. Recommended types and dates ofservice:

- Visual inspection of the GA (article 2.3.2) - every 6months;- Zeroing and span gas calibration (article 2.3.3) - every 6months;- Cleaning the enclosure from dust and dirt - every 6months.

2.3.2. Visual inspectionCheck the enclosure of GA for cracks. Check theconnectors.

2.3.3. Zeroing and span gascalibration2.3.3.1. Prepare GA for operations according article2.2.1.1.2.3.3.2. Connect GA to gas feeding according to diagram shown in AppendixC.2.3.3.3. FeedtheGAwithzerogas(nitrogen,CGM№1seeAppendixA).Flowrateis0.5l/min. duration - oneminute.2.3.3.4. GAzeroing:- Hold “>0<” button during 1 sec. theLCDlights the symbols"ZERO?".- Press"ВВОД"buttontoconfirmzeroing.Thenewvalueofthezerostateiswrittentoflash memory.

NOTE Set to zero can also be submitting from PC by sending "ZERO"command.2.3.3.5. ConnectGAtoPCandfeedtheGAwithspangas(CGM№2or3seeAppendixA).Flow rate is 0.5 l / min.

duration - oneminute.2.3.3.6. Calibrate the GA by sending command "CALB XXXX" from PC, where XXXX is thenumerical value which is strongly corresponds to the concentration of CGMapplied.

2.3.3.7. Setting up alarmthresholds.Tosetupalarmthresholdpushthebutto

n"ПОРОГ"andpressortochoosedesirablevalue,whichisdisplayingonthebottomlineintheleftcornerofLCD."Press“ВВОД”tostorechosenalarm threshold in the memory. This will automatically change the scale of the red bar indicator. The newalarm threshold value corresponds to half of indicationscale.

User, 06/04/15,
Имеют ли кнопки уже утвержденный перевод?
Sergeev Stas, 06/05/15,
Клавитурки у Поляриса с кириллице, поэтому пока надо оставить так. В user manual для образцов -думаю сойдет, а в будущем, если закажет кто-нибудь закажет партию, клавиатуры купим новые и тогда подправим.
User, 06/04/15,
Имеют ли кнопки уже утвержденный перевод?
Page 13: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

13

3Troubleshooting3.1.Troubleshootinglist

Table 3 Troubleshootinglist

Faultevent Possiblereason RecommendationsBATT symbols lights on LCD Lowbattery. Charge thebattery.screen.

Fail symbols lights on LCDscreen. Electrical circuit malfunction, Enter the command FULV? from PCdetected by the systemself-test to check to possible malfunction

reasons (see article 3.2 or AppendixB)Save response log and sendthisinformation to manufacturer forproblemsolving.

No symbols on LCDscreen. Completely dischargedbattery Try to charge or replace thebattery.

OPT symbols lights in right down Optics in gas cell isdirty Enter the command FULV? from PCcorner of LCDscreen. to check to possible malfunction

reasons (see article 3.2 or AppendixB)Save response log and sendthisinformation to manufacturer forproblemsolving.

Red LED lighting during the Batteryfail Replace thebattery.batterycharging.

EHT symbols lights in rightdown Lowtemperature Additional heating of GA requiredcorner of LCDscreen. achieving proper performance.After

the heating the using of the GAispossible when warm-up processdone and device displays OKsymbols in right downcorner.

NOTE: Any malfunctions requires breaking GA sealing must be fixed bymanufacturerauthorized personalonly.

3.1. Malfunctionself-testGAhastheabilitytooutputdatatoPCinself-testmode.SendFULV?commandtoidentify possible problems. See Appendix B for details about command format. Contact to manufacturer formore information.

Page 14: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

14

4 Storage GAshouldbestoredinaheatedplaceatatemperatureoffrom5to40°Candrelative humidity of

80% at 25 °C. Device should be stored in the factorypacking. Storage time is 3months. Amountofdust,vaporsofacidsandalkalis,corrosivegasesandothersubstancesthat may

cause corrosion in storage places should not exceed the values of its content required of type 1 according to GOST 15150-69 (IEC 721-2-1, IEC 721-3-1 IEC 721-3-7,IEC68-1).

5 TransportationTransportationofGAshouldbedoneinconsumerpackaging,transportbox(woodenor cardboard).GA for transportation must be packed in the transportbox. Transportation is possibleif:1) ambient temperature from minus 30 to plus 50 °C;2) maximum relative humidity of 98% at 35 °C.GAcanbetransportedbyalltypesoftransport,includingairtransport,inheatedandsealed places with

protection from direct exposure to rain anddust.

6 WarrantyThe manufacturer guarantees compliance with all the requirements of the GA technicalconditions

iftheCustomerobservesoperation,transportationandstoragewithinthewarrantyperiod-12months from the date of take intooperation.

The starting point for calculating the warranty period is the date of sale to theConsumerWarrantyrepairorreplacementshallbesubjecttotheGAofimproperuse,transportationand storage.The warranty period is extended for repeating time till the Consumer put in use repaireddevice.

7 RepairingIn case of GA failure, as well as the detection of incompleteness of its delivery set, theConsumer must

return device to the Manufacturer with the followingnotice:- Model of gas analyzer, serial number, production date and date ofsale;- Information about availability of factoryseals;- Defect (or incompleteness)description.

Page 15: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

16

Appendix AChecking procedure

Sergeev Stas, 06/05/15,
Это не нужно. В методике поверки ВНИИМа - все на русском и для русских.
Page 16: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

17

Appendix B RS-232 communicationprotocolCommunicationparameters:

9600 baudrate 8-bit message, 1 stop bit, noparity, output rate: 1 time per 4 sec. ASCII

Commands:Commands with multiple (continues) outputting of dataresponses:

CDAT?– requests concentration slow output. Response format is AAAAA 0DH AAAAA is methane concentration in ppm*10 with regard to scale coefficientset.FULV? — requestsself-test results. Response is informat:

AAAAA09HBBBBB09HCCCCC09HDDDDD09HEEEEE09HFFFFF09HGGGGG09HHHHHH09HIIIII 09H JJJJJ 09H KKKKK0DH

Where:

AAAAA T amb – ambient temperature ofGABBBBB Signal value of measuring channel (Umeas) , accumulation period - 4secCCCCC Signal value of reference channel (Uref) , structurally stablescaleDDDDD U const – constant component - the lightintensityEEEEE T eth – referencetemperatureFFFFF T PD – photodetectortemperatureGGGGG Ua – battery voltagecontrolHHHHH U gen –clock generator voltagecontrolIIIII Concentration calculated from table inppm*10JJJJJ Concentration with regard to scale coefficient (see CALB XXXXX) inppm*10KKKKKККК GA serialnumber

Commands with singleresponse:DATA?— requests concentration output AAAAA0DHAAAAA is methane concentration in ppm*10 with regard to scale coefficientset.CDAT?– requests concentration slow output. Response format is AAAAA 0DH

AAAAA is methane concentration in ppm*10 with regard to scale coefficientset.

Commands for operation with constants which are written to the EEPROM of the deviceSRAL?– serialnumberrequest.INIT – reset device to factorysettings. ZERO – zeroing theGA. CALB XXXXX – save the scale (calibration) coefficient. XXXXX is applied concentration inppm*10.

AlarmsettingsALARM?— requests alarmthreshold.ALARM XXXXX— set up alarm threshold. XXXXX - concentrationinppm*10.

Page 17: Microsoft Word - POLYARIS RE _perenosnoy _ ver5.docigm-pribor.com/files/polaris-1001/User manual...  · Web viewMTBF, without regard to batteries replacements, hours, not lessthan:10000

18

Appendix C Gas feeding diagram required for zeroing and spangas calibration

12

3

Gas outputto exhaustsystem

4

1 - Control Gas Mixture (CGM ) vessel underpressure;2 - fine control gasvent;3 - flowmeter;4 –GA