Microchannel Plate Photomultiplier - University of...
Transcript of Microchannel Plate Photomultiplier - University of...
UNIVERSITY OF LJUBLJANA
FACULTY OF MATHEMATICS AND PHYSICS
DEPARTMENT OF PHYSICS
Seminar
Microchannel Plate Photomultiplier
Author: Matija Dmitrovič Petrič
Supervisor: prof. dr. Peter Križan
Abstract
Microchannel plate photomultiplier (MCP - PMT) is considered to be the most promising
candidate for photon detectors which are required by the Cherenkov detector. This seminar
describes the basic structure of MCP - PMT and its characteristics.
Ljubljana, march 2008
1
Table of Contents
1. Introduction ......................................................................................................................................... 2
2. Cherenkov radiation ............................................................................................................................ 3
3. Microchannel Plate Photomultiplier .................................................................................................... 4
3.1 Structure of MCPs ......................................................................................................................... 5
3.2 Structure of MCP-PMTs ............................................................................................................... 7
4. Characteristics of MCP-PMTs ............................................................................................................ 7
4.1 Gain characteristics ....................................................................................................................... 7
4.2 Time characteristics ....................................................................................................................... 9
4.2.1 Rise/fall times ......................................................................................................................... 9
4.2.2 Transit time ............................................................................................................................. 9
4.2.3 TTS (transit time spread) ...................................................................................................... 10
4.3. Dark current ................................................................................................................................ 10
4.4 Magnetic characteristics .............................................................................................................. 10
5. Multianode MCP-PMTs .................................................................................................................... 12
6. References: ........................................................................................................................................ 14
2
1. Introduction
One of the unsolved theoretical questions in physics is why the universe is made chiefly of
matter, rather than consisting of equal parts of matter and antimatter. The Big Bang should
have produced equal amounts of matter and antimatter, so in time there should have been total
cancellation of both. In other words, protons should have cancelled with anti-protons,
electrons with positrons, neutrons with anti-neutrons, and so on for all elementary particles.
This would have resulted in a sea of photons in the universe with no matter. Since this is quite
evidently not the case, after the Big Bang, physical laws must have acted differently for
matter and antimatter.
The most plausible explanation for this is the CP violation in particle physics. CP is the
product of two symmetries: C for charge conjugation, which transforms a particle into its
antiparticle, and P for parity, which creates the mirror image of a physical system. The strong
interaction and electromagnetic interaction seem to be invariant under the combined CP
transformation operation, but this symmetry is slightly violated during certain types of weak
decay.
One experiment that investigates this violation is the Belle experiment at the High Energy
Accelerator Research Organisation (KEK) in Tsukuba, Japan. For the Belle particle
identification system upgrade a new Cherenkov detector is being considered. It requires
photon detectors that work in high magnetic fields of about 1.5 T, have pad size of few mm
and timing resolution around 50 ps. The most promising candidate is a microchannel plate
photomultiplier (MCP - PMT). In this seminar I will describe how Cherenkov detector works,
and why MCP PMT has all of the above advantages.
Figure 1: Typical MCP – PMT is a few cm big
3
2. Cherenkov radiation
In a medium such as water, the speed of light is considerably less than the speed of light in
vacuum (c). In a medium with refractive index n the velocity of light is vlight = c/n. If the
refractive index is greater than one it is possible for a particle to travel through water (nwater
= 1.3) or other media at a speed faster than the speed of light in that media. When a charged
particle does so (v > vlight), a faint radiation is produced from the medium. The charged
particle excites the water molecules which then return to their normal state emitting photons
of blue light. Because the particle is moving faster than the speed of light in water, it can
trigger a cascade of photons which are in phase with each other and can constructively
interfere to form the visible blue glow. The light propagates away in a cone forward of the
region where the interaction took place. The half angle θ of the cone can easily be derived in
terms of the velocity v of the particle by looking at where wave fronts emitted from the track
of the particle constructively interfere.
cosθ = vlight/v = c/(vn)
The emitted radiation is called Cherenkov radiation after P. Cherenkov who received the 1958
Nobel Prize for Physics with Igor Y. Tamm and Ilya M. Frank for their investigation of the
phenomenon. The Cerenkov angle is zero at the threshold velocity for the emission of
Cerenkov radiation. The angle takes on a maximum as the particle speed approaches the speed
of light. It is possible to detect the Cherenkov radiation as it forms circles on a surface and can
be used to measure the speed and direction the particle was travelling in. It is therefore a very
useful means of studying the products of particle collisions and cosmic rays.
The equivalent phenomenon in acoustics is the formation of the acoustical shock wave
generated by a body moving with supersonic velocity.
Figure 2: Identification of particles by measuring Cherenkov angle and momentum for two
different refractive indexes
4
Cherenkov radiation is commonly used in experimental particle physics for particle
identification. One could measure (or put limits on) the velocity of an electrically charged
elementary particle by the properties of the Cherenkov light it emits in a certain medium. If
the momentum of the particle is measured independently, one could compute the mass of the
particle by its momentum and velocity, and hence identify the particle.
The most advanced type of a detector is the RICH, or ring imaging Cherenkov detector,
developed in 1980s. In a RICH detector, a cone of Cherenkov light is produced when a high
speed charged particle traverses a suitable gaseous or liquid medium, often called radiator.
This light cone is detected on a position sensitive planar photon detector, which allows
reconstructing a ring or disc, the radius of which is a measure for the Cherenkov emission
angle.
Figure 3: Cherenkov radiation from a charged particle nearly at the speed of light moving in a
medium with index of refraction n > 1
3. Microchannel Plate Photomultiplier
MCP-PMT is a planar component used for detection of photons, especially of blue light. It is
closely related to an electron multiplier, as both intensify single particles or photons by the
multiplication of electrons via secondary emission. MCP-PMTs, photomultiplier tubes that
incorporate an MCP in place of the conventional discrete dynodes, offer wide-bandwidth
measurements down to the picosecond level as well as low-light-level detection at the photon
counting level.
5
3.1 Structure of MCPs
The MCP consists of a two-dimensional array of a great number of glass capillaries
(channels) bundled in parallel and formed into the shape of a thin disk (Figure 4). MCPs are
typically 2 mm thick. Each channel has an internal diameter ranging from 6 to 100 microns
with the inner wall processed to have the proper electrical resistance and secondary emissive
properties. Accordingly, each channel acts as an independent electron multiplier. Being
accelerated by the electric field created by the voltage VB (usually between -1000 and –3600
V) applied across both ends of the MCP, these secondary electrons bombard the channel wall
again to produce additional secondary electrons (Figure 4). This process is repeated many
times along the channel and as a result, a large number of electrons are released from the
output end.
Figure 4: Schematic structure of an MCP
PHOTOELECTRONS
6
Figure 5: Schematic of a single channel of an MCP showing the mechanism of secondary
electron production
MCPs are quite different in structure and operation from conventional discrete dynodes and
therefore offer the following outstanding features:
- High gain despite compact size (104 – 10
7)
- Fast time response (< 100 ps)
- Two-dimensional detection with high spatial resolution
- Stable operation even in high magnetic fields (up to 2 T)
There are various types of detectors that utilize the advantages offered by MCPs, for example
image intensifiers for low-light-level imaging, fast time response photomultiplier tubes that
incorporate an MCP (MCP-PMTs), position-sensitive multianode photomultiplier tubes,
streak tubes for ultra-fast photometry, and photon counting imaging tubes for ultra-low light
level imaging.
Photoelectron
7
3.2 Structure of MCP-PMTs
Figure 6 shows the cross section of a typical MCP-PMT. This MCP-PMT consists of an input
window, photocathode, dual MCP, and anode. The photoelectrons emitted from the
photocathode enter the channels of the MCP and impinge on the inner wall where they are
multiplied by means of secondary emission. This process is repeated along the channels, and
finally a large number of electrons are collected by the anode as an output signal. The
photocathode to MCP distance is approximately 2 millimetres, forming a close-proximity
structure. Two MCPs are stacked to obtain sufficient gain. A thin film called "ion barrier" is
usually formed on the photoelectron input side of the MCP in order to prevent ions generated
inside the MCP from returning to the photocathode.
Figure 6: Cross section of a typical MCP-PMT
4. Characteristics of MCP-PMTs
4.1 Gain characteristics
The gain of an MCP-PMT depends on the number of MCPs incorporated in the tube. Figure 7 shows
the typical gain versus supply voltage characteristics of an MCP-PMT.
8
Figure 7: Typical gain of an MCP-PMT (using a two-stage MCP of 6 µm channel diameter)
The gain (µ) of an MCP is determined by the length-to-diameter ratio 𝛼 (= 𝐿 / 𝑑) of a channel, and
approximated as follows:
µ = 𝑒𝐺𝛼
where G is the secondary emission characteristics called the gain factor. This gain factor is an inherent
characteristic of the channel wall material and is a function of the electric field intensity inside the
channel.
In general, a higher gain can be obtained as α is made greater, though the gain rising point moves to
the higher supply voltage side. However, if the gain becomes higher than 104, noise begins to increase
significantly due to ion feedback effects, which causes a serious problem. To avoid this, α is usually
selected to be around 40 so that a single MCP provides a gain of about 104 at 1 kV supply voltage.
As shown in Figure 7 above, a higher gain can be obtained from a two-stage MCP-PMT. This gain
level enables photon counting measurements.
9
4.2 Time characteristics
The signal pulse can broaden during the multiplication process from the photocathode to the
anode. This is due to the emission-angle distribution and initial-velocity distribution of
photoelectrons and secondary electrons, as well as the effects of the focusing lens. In a MCP-
PMT, a strong electric field is applied parallel from the photocathode to MCPin and the
MCPout to anode, so that the emission-angle distribution and initial-velocity distribution of
photoelectrons can be almost ignored. Furthermore, since MCP is used in place of
conventional dynodes, the electron transit time in the secondary electron multiplication
process is very short, allowing a dramatic improvement in the transit time spread. Due to
these features, the MCP-PMT offers time response characteristics that are the best among
currently available photomultiplier tubes
4.2.1 Rise/fall times
The rise and fall times of an MCP-PMT are evaluated from the output waveform when the
MCP-PMT detects a light pulse whose width is sufficiently short compared to the time
response of the MCP-PMT. These parameters are especially important when observing the
waveform of ultra-short pulsed light. Figure 8 shows an actual waveform obtained with an
MCP-PMT.
Figure 8: Pulse response waveform of MCP-PMT
4.2.2 Transit time
The transit time is the time delay between the input of a light pulse at the photomultiplier tube
and the appearance of the output pulse from the photomultiplier tube.
10
4.2.3 TTS (transit time spread)
When one photon enters the MCP-PMT, the photocathode converts it into an electron which
travels to the anode while being multiplied. The transit time of an electron bunch differs
depending on each input photon. The distribution of this transit time is referred to as the
transit time spread or TTS. This TTS is an important parameter, especially in the time-
correlated photon counting technique where the measurement of timing is of prime
consideration.
Figure 9: Transit time spread fitted with double Gaussian function. The timing distribution
has a main peak with σ of about 40 ps and characteristically long tail due to the backscattering
of photoelectrons. Contributions of electronics and laser to the width of the main peak are σ ≈
15 ps and σ ≈ 12 ps respectively.
4.3. Dark current
As with normal photomultiplier tubes, the dark current of MCP-PMT greatly depends on the
photocathode type and operating temperature. In particular, the dark current of a multialkali
photocathode with enhanced red sensitivity are relatively high at room temperatures, so MCP-
PMTs using such a photocathode may need to be cooled during operation.
4.4 Magnetic characteristics
The following points are essential for good magnetic characteristics:
- The distance between the photocathode, dynodes and anode should be shortened to
minimize the electron transit distance.
- The electrodes should be designed to apply a parallel electric field from the photocathode to
the anode so that the secondary electrons do not converge but travel in parallel to the tube
axis.
- A high electric field intensity should be applied.
11
Because of the small dimensions and correspondingly high electric filed strengths, the MCP-
PMT meets all the above requirements and provides superior magnetic characteristics. The
extent of the effect of a magnetic field on the output depends on the direction of the magnetic
field with respect to the MCP axis. In magnetic fields parallel to the tube axis, the MCP-PMT
can operate at up to 2.0 T, but in magnetic fields perpendicular to the tube axis, the output
drops drastically if fields exceed 70 mT (Figure 10 and 11).
Figure 10: Typical magnetic characteristics of an MCP-PMT when in magnetic fields parallel
to tube axis.
Figure 11: Typical magnetic characteristics of an MCP-PMT when in magnetic fields
perpendicular to tube axis.
12
5. Multianode MCP-PMTs
The previous sections mainly discussed MCP-PMTs having a single anode. A variety of
MCP-PMTs with independent multianodes have been developed and put to practical use.
These multianode MCP-PMTs offer simultaneous, two-dimensional (or one-dimensional)
detection as well as fast response speed and low-light-level detection.
Figure 12: 4-channel multianode MCP-PMT
Surface response of PMTs is fairly uniform. Multiple counting is observed at pad boundaries
due to charge sharing effect. Slice of 2D distribution shows uniform response within the pads,
short range cross-talk due to charge sharing and long range photoelectron backscattering
cross-talk (Figure 13)
13
Figure 13: Individual pad count rates as a function of position across 4 channels in linear
(upper) and logarithmic (lower) scale.
14
6. References:
J. L. Wiza, Microchannel plate detectors, Nuclear Instruments and Methods 162 (1979), 587 - 601
S. Korpar et al., Timing and cross-talk properties of BURLE multi-channel MCP PMTs
Hamamatsu Photonics, Photomultiplier tubes
http://math.ucr.edu/home/baez/physics/Relativity/SpeedOfLight/cherenkov.html
http://en.wikipedia.org/wiki/Cherenkov_radiation
http://commons.wikimedia.org/wiki/Image:Cherenkov2.svg
http://en.wikipedia.org/wiki/Micro-channel_plate
http://hea-www.harvard.edu/HRC/mcp/mcp.html
http://belle.kek.jp