653 Bioanalysis (2015) 7(6), 653–659 ISSN 1757-6180 Methodology part of 10.4155/BIO.14.310 © Phenomenex LTD Aims: A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. Materials & Methods: An average blood volume of 10.6 μl was absorbed by the samplers across the different HCTs investigated (20–65%). Results: No notable change of volume absorbed was noted across the HCT range. Furthermore, the variation in blood sample volumes across six different laboratories was within acceptable limits. Conclusion: The novel volumetric absorptive microsampling device has the potential to deliver the advantages of dried blood spot sampling while overcoming some of the issues associated with the technology. The use of dried blood spot (DBS) sampling for the collection of samples for the determina- tion of drug concentrations has received a lot of attention recently within the pharmaceutical industry [1,2] . This approach has been widely used for the determination and discovery ani- mal pharmacokinetic (PK), preclinical toxi- cokinetic, clinical PK and human therapeutic drug monitoring data [3–11] . The interest in the technique has been driven by its advantages over conventional plasma sampling for these study designs. These advantages include; Reduced blood volumes (<20 μl com- pared with >200 μl), leading to ethical benefits in rodent PK and toxicokinetic studies (reductions in the numbers of animals used and refinements in the warming procedures required to encour- age blood flow) and its applicability to pediatric study designs; Simplified sample collection workflows, obviating the requirements for centrifu- gation, plasma transfer and frozen sample storage and transfer; Increased or suitable stability for some analytes without requiring frozen storage [12–16] ; Ability to obtain high-quality samples in locations not previously readily amenable for collection (e.g., patient homes and remote locations); Cost savings associated with the shipping and storage of study samples at ambi- ent room temperature rather than frozen, and a requirement for reduced amounts of experimental drug substances, particularly for discovery studies. However, recent publications have high- lighted a number of issues that have the potential to adversely affect the quality of the quantitative data obtained from DBS samples when a subpunch is taken from a sample [17–21] . Most practitioners prefer to use the subpunch method for DBS sam- pling as it simplifies the initial collection and spotting of the sample by removing the need to use accurate volume spotting techniques. Instead, the animal technician or clinician can spot an approximate vol- ume from which a fixed-diameter subpunch can be taken in order to realize an accurate volume at the point of analysis. However, this approach relies on the blood spreading evenly when initially spotted. Recent stud- ies have clearly demonstrated that the blood A device for dried blood microsampling in quantitative bioanalysis: overcoming the issues associated with blood hematocrit Neil Spooner* ,1 , Philip Denniff 1 , Luc Michielsen 2 , Ronald De Vries 2 , Qin C Ji 3 , Mark E Arnold 3 , Karen Woods 4 , Eric J Woolf 5 , Yang Xu 5 , Valérie Boutet 6 , Patricia Zane 7 , Stuart Kushon 8 & James B Rudge 9 1 Bioanalytical Science & Toxicokinetics, Drug Metabolism & Pharmacokinetics, GlaxoSmithKline Research & Development, Ware, UK 2 Bioanalysis Department, Janssen Research & Development, Turnhoutseweg 30, 2340 Beerse, Belgium 3 Bioanalytical Science & Selective Integration, Bristol-Myers Squibb, Co., Princeton, NJ 08543, USA 4 Pre Clinical Bioanalysis & Toxicokinetics AstraZeneca, Alderley Park, UK 5 Merck Research Laboratories, PPDM- Global Bioanalytics, WP 75B-300, West Point, PA 19486, USA 6 Sanofi Recherche & Developpement, 3 digue d’Alfortville, Bâtiment Claude Bernard, 94140, Alfortville, France 7 Sanofi, 55 Corporate Drive, Bridgewater, NJ 08807, USA 8 Neoteryx, LLC, 421 Amapola Avenue, Torrance, CA 90501, USA 9 Phenomenex, 411 Madrid Avenue, Torrance, CA 90501, USA *Author for correspondence: Tel.: +44 1920 882550 Fax: +44 1920 884140 [email protected] For reprint orders, please contact [email protected]

Transcript of Methodology A device for dried blood microsampling in ... · 2015 Aims: A cross-laboratory...

653Bioanalysis (2015) 7(6), 653–659 ISSN 1757-6180

Methodology

part of

10.4155/BIO.14.310 © Phenomenex LTD

Bioanalysis

Methodology7

6

2015

Aims: A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. Materials & Methods: An average blood volume of 10.6 μl was absorbed by the samplers across the different HCTs investigated (20–65%). Results: No notable change of volume absorbed was noted across the HCT range. Furthermore, the variation in blood sample volumes across six different laboratories was within acceptable limits. Conclusion: The novel volumetric absorptive microsampling device has the potential to deliver the advantages of dried blood spot sampling while overcoming some of the issues associated with the technology.

The use of dried blood spot (DBS) sampling for the collection of samples for the determina-tion of drug concentrations has received a lot of attention recently within the pharmaceutical industry [1,2]. This approach has been widely used for the determination and discovery ani-mal pharmacokinetic (PK), preclinical toxi-cokinetic, clinical PK and human therapeutic drug monitoring data [3–11]. The interest in the technique has been driven by its advantages over conventional plasma sampling for these study designs. These advantages include;

• Reduced blood volumes (<20 μl com-pared with >200 μl), leading to ethical benefits in rodent PK and toxicokinetic studies (reductions in the numbers of animals used and refinements in the warming procedures required to encour-age blood flow) and its applicability to pediatric study designs;

• Simplified sample collection workflows, obviating the requirements for centrifu-gation, plasma transfer and frozen sample storage and transfer;

• Increased or suitable stability for some analytes without requiring frozen storage [12–16];

• Ability to obtain high-quality samples in locations not previously readily amenable for collection (e.g., patient homes and remote locations);

• Cost savings associated with the shipping and storage of study samples at ambi-ent room temperature rather than frozen, and a requirement for reduced amounts of experimental drug substances, particularly for discovery studies.

However, recent publications have high-lighted a number of issues that have the potential to adversely affect the quality of the quantitative data obtained from DBS samples when a subpunch is taken from a sample [17–21]. Most practitioners prefer to use the subpunch method for DBS sam-pling as it simplifies the initial collection and spotting of the sample by removing the need to use accurate volume spotting techniques. Instead, the animal technician or clinician can spot an approximate vol-ume from which a fixed-diameter subpunch can be taken in order to realize an accurate volume at the point of analysis. However, this approach relies on the blood spreading evenly when initially spotted. Recent stud-ies have clearly demonstrated that the blood

A device for dried blood microsampling in quantitative bioanalysis: overcoming the issues associated with blood hematocrit

Neil Spooner*,1, Philip Denniff1, Luc Michielsen2, Ronald De Vries2, Qin C Ji3, Mark E Arnold3, Karen Woods4, Eric J Woolf5, Yang Xu5, Valérie Boutet6, Patricia Zane7, Stuart Kushon8 & James B Rudge9

1Bioanalytical Science & Toxicokinetics,

Drug Metabolism & Pharmacokinetics,

GlaxoSmithKline Research &

Development, Ware, UK 2Bioanalysis Department, Janssen

Research & Development,

Turnhoutseweg 30, 2340 Beerse,

Belgium 3Bioanalytical Science & Selective

Integration, Bristol-Myers Squibb, Co.,

Princeton, NJ 08543, USA 4Pre Clinical Bioanalysis & Toxicokinetics

AstraZeneca, Alderley Park, UK 5Merck Research Laboratories, PPDM-

Global Bioanalytics, WP 75B-300, West

Point, PA 19486, USA 6Sanofi Recherche & Developpement,

3 digue d’Alfortville, Bâtiment Claude

Bernard, 94140, Alfortville, France 7Sanofi, 55 Corporate Drive, Bridgewater,

NJ 08807, USA 8Neoteryx, LLC, 421 Amapola Avenue,

Torrance, CA 90501, USA 9Phenomenex, 411 Madrid Avenue,

Torrance, CA 90501, USA

*Author for correspondence:

Tel.: +44 1920 882550

Fax: +44 1920 884140

[email protected]

For reprint orders, please contact [email protected]

654 Bioanalysis (2015) 7(6) future science group

Methodology Spooner, Denniff, Michielsen et al.

does not spread homogenously on the DBS material (substrate). Furthermore, the hematocrit (HCT) of the blood affects its viscosity and so gives rise to dif-ferent-sized DBS spots, with high-HCT blood giving smaller blood spots and low-HCT blood giving larger spots. Thus, the volume of blood in a fixed-diameter subpunch taken from these samples is difficult to determine, resulting in lower-quality drug concentra-tion data for samples, particularly in which the blood HCT varies markedly from that of the control blood used to prepare the calibrant and quality control (QC) samples.

A number of laboratories have published manuscripts describing novel approaches to overcoming these issues regarding HCT and homogeneity with DBS [22–25]. Unfortunately, these approaches rely on spotting an accurate volume of blood, measuring another blood component or are not readily commercially avail-able. Thus, their implementation for day-to-day study support may be impractical.

A novel dried blood sampler, termed the volumetric absorbtive microsampler (VAMS), has been designed in order to deliver the benefits of DBS sampling while overcoming the issues associated with HCT and homogeneity and also enabling further simplification of the sample collection and processing/extraction work-flows [26]. The VAMS consists of an absorbent polymeric tip designed to take up a fixed volume of blood (nomi-nally 10 μl) by capillary action. The tip is attached to a handle by a plastic pin (Figure 1). The handle has been designed to be ergonomic to hold and has fins to help locate it within 96-well blocks during extraction and to minimize the possibility of the sampler tip coming into contact with surfaces during storage and shipping. The sampler is filled by holding the handle part of the sam-pler and dipping only the leading surface of the tip into a pool of blood and allowing it to fill. The tip of the sam-pler must not be completely submerged into the blood sampler, as this may cause overfilling. When the tip has turned completely red, it is full, which takes 2–5 s, mak-ing the device self-indicating. The last part of the tip to turn red is the shoulder. In addition, the sampler is engineered to fit a standard laboratory air-displacement pipette or automated liquid handler in order to allow for automation during extraction of the blood sample.

This article describes results from a cross-company and -laboratory gravimetric experiment to determine the volume of blood absorbed by the sampler at vari-ous blood HCT values (∼20, 45 and 65%). Further-more, in order to convert the weights of the blood absorbed by the sampler into volumes, the density of the blood at each HCT was determined by each laboratory.

Materials & methodsEquipment & reagentsVAMSs were supplied by Phenomenex, Inc. (CA, USA; exclusive distributor for the Mitra™ micro-sampler, manufactured by Neotryx, LLC, CA, USA). Control EDTA blood was obtained from the fol-lowing suppliers in accordance with each company’s current policies on informed consent and ethical approval: human blood was obtained from: Glaxo-SmithKline (GSK Stevenage, UK) from in-house facilities and used within 7 h of draw; Merck from Biological Specialties (PA, USA) and used within 10 h of draw; Janssen from the Clinical Pharmacol-ogy Unit (Belgium) and in-house and used within 4 h of draw; and Sanofi from Biopredic Interna-tional (Saint-Gregoire, France) and used within 48 h of draw. Rat blood was obtained from Bioreclamation (Brussels, Belgium) and used within 2 weeks of draw (Bristol-Meyers Squibb [BMS]) or from in-house facilities used within 7 h of draw (AstraZeneca [AZ]). Blood samples with different nominal HCT values of 20, 45 and 65% (and additionally at 60 and 69% for two laboratories) were prepared at each participating laboratory by determining the HCT of the control blood by centrifugation and then either removing the appropriate volumes of plasma from centrifuged blood or adding plasma to blood that had not been centrifuged, followed by gentle mixing. The HCT of the modified blood samples was then confirmed using centrifugation [19].

All experimental sites used microbalances that mea-sured to at least five decimal places of a gram. GSK used a Sartorius R200D, Janssen used a Sartorius CPA225D, BMS used a Sartorius Cubis Precision Lab Balance, AZ used a Sartorius MC210P, Merck used a Mettler Toledo SAG 285 and Sanofi used a Mettler Toledo AT261. The pipettes used were Gilson Micro-man M10 (GSK), Eppendorf Reference (Janssen), Biohit mline (10–100 μl; BMS), Gilson Microman M25 (AZ), Eppendorf positive displacement repeaters equipped with a 0.1-ml ‘Combitip’ (Merck) and Bio-hit Proline (0.5–10 μl; Sanofi). Scintillation vials were obtained from Thermo Fisher (20 ml: GSK and BMS; 7 ml: Merck), Fiolax (20 ml: Janssen) and Perkin Elmer (20 ml: AZ and Sanofi).

Key terms

Hematocrit: Ratio of the volume of red blood cells to the volume of whole blood.

Volumetric absorbtive microsampler: A technique for the collection of dried biological samples for quantitative bioanalysis by absorbing a fixed accurate volume of blood onto an absorbent tip. The commercial name for these samplers is Mitra™.

www.future-science.com 655

Figure 1. Volumetric absorbtive microsampler before (left) and after (right) filling with blood.

Tip shoulder

Absorbant tip

Handle

future science group

A device for dried blood microsampling in quantitative bioanalysis Methodology

Gravimetric determination of blood volume absorbedPrior to performing the blood absorption experi-ments, the balance and weighing areas were cleaned and dried to ensure there was no loose liquid on the balance plate, as evaporation would lead to poor bal-ance stability. Moreover, to improve precision during weighing, vials were placed in the same place on the balance. Finally, to increase humidity and reduce the rate of evaporation, a wet tissue was placed in a beaker and placed within the balance enclosure, but not on the balance plate.

Before weighing, each balance was set to zero and a 2-ml aliquot of the 20, 45 or 65% (including 60 and 69% for two laboratories) HCT blood was placed in a scintillation vial, which was capped and placed in the center of the balance plate.

To determine the weight of an accurately pipet-ted volume (control protocol) at each HCT level, the initial weight of a capped vial containing 2 ml of blood was recorded. The vial was then uncapped and 10 μl of blood was removed using a pipette. The vial was then recapped, reweighed and the weight was recorded. This procedure was conducted six times per HCT at each laboratory. Once completed, the average density of each HCT at each laboratory was calculated as follows:

To determine the weight of the blood absorbed by the VAMS tip at each HCT level (VAMS protocol), the initial weight of a capped vial containing 2 ml of blood was recorded. The vial was then uncapped and a fresh VAMS was carefully ‘dipped’ into the blood such that only the tip was exposed to the blood. The tip was held in the blood until it appeared to be full (i.e., no white portions were observed) and then for an extra approximately 2 s before being removed. Care was taken when filling the tips not to immerse the tip past the shoulder, as this could result in excess blood being retained and hence introduce a positive bias to any measurements. Furthermore, upon removing the tip after filling, care was taken to ensure the sampler or tip had not contacted the walls of the vial, as this could have caused errors in the result. The VAMS was then discarded. As in the control experiment, each vial was recapped, reweighed and the weight was recorded. This procedure was conducted six times per HCT at each laboratory. The blood volume absorbed on the VAMS tip for each HCT at each laboratory was calculated as follows:

Laboratories 1, 4, 5 and 6 performed the operation to determine the mean weight of blood in a 10-μl aliquot for the above experiments as described earlier, while laboratories 2 and 3 performed the operation by weigh-ing a vial containing 2 ml of blood and then dispensed 10-μl aliquots of blood into it from a separate vial con-taining blood at the same HCT as that being weighed, and reweighing the target vial after each addition.

Results & discussionVolume of blood absorbed at different HCTsThe volume of human and rat blood with different HCTs absorbed by the VAMS was investigated in six different laboratories by determining changes in weight after dipping the device into control blood (Figure 2). The average volume of human or rat blood collected across the approximate HCT range of 20–65% was 10.6 ± 0.4 μl (error determined as [2 × standard devia-tion]/square root of the count). This is in good agree-ment with the 10.5 ± 0.1 μl determined using radio-active 14C caffeine spiked into human blood followed by measurement of CO

2 after oxidation of the dried

tip [26]. Furthermore, the slope of the blood volume against HCT plot is close to zero, demonstrating that the blood volume collected by the tip showed little variance with different blood HCT values.

The variability in the blood volume collected across laboratories and HCTs was minimal (CV: 8.7%) and well within the 15% that is routinely accepted for the validation of quantitative bioanalyical methods. However, there was notable variance in the minimum and maximum volumes absorbed (between 9.1 and 13.1 μl). While the reasons for this variance are not

Blood density (ng/ml)

10Mean weight of blood in 10 l aliquot (g)

1000µ

#

=

Volume of blood in VAMS Tip (µl)

Mean weight of blood in 10 l aliquot (g)Mean weight of blood absorbed by VAMS Tip (g)

µ

=

656 Bioanalysis (2015) 7(6)

Figure 2. Gravimetric determination of the volume of blood absorbed by volumetric absorbtive microsampler tips at various hematocrit values. Determinations were performed at six different laboratories. Each data point is the average of six determinations made at each laboratory.

Blo

od

vo

lum

e (µ

l)

Hematocrit (%)

14

12

10

8

6

4

2

0

0 20 40 60 80

Laboratory 1

Laboratory 2

Laboratory 4

Laboratory 5

Laboratory 6

Average

Laboratory 3(1)

Laboratory 3(2)y = -0.0892x + 10.668R2 = 0.0009

future science group

Methodology Spooner, Denniff, Michielsen et al.

known, one possible explanation may originate from the approaches taken by the different participating laboratories to pipette 10 μl of blood, the weight of which is required for the calculation of both the vol-ume of blood in the VAMS tip and the blood density. Laboratories 1, 4, 5 and 6 derived the value for the weight of the blood by removal with a pipette, while laboratories 2 and 3 derived it by the addition of blood with a pipette. It is possible that a small amount of blood may remain in the pipette tip when dispensing, resulting in the weights for the determination of the volume values being marginally higher in the labora-tories using blood removal compared with those using addition. This is in turn reflected in the lower density (Figure 3) and higher blood volume (Figure 2) values for laboratories 2 and 3, where addition was used, than those for the other laboratories, where removal was used. Alternative explanations for these differ-ences may be either a variation in manufacturing of the tip or differences in which the VAMS device was used for sampling (i.e., dipping the tip too deeply into the blood pool). Based upon this, it is important that laboratories perform additional experiments as part of the development and validation of a quantitative bioanalytical assay when using these blood samplers. These may include the inclusion of tips from various production batches and that more than one operator should prepare QC samples. It is also important that users of the technology be thoroughly trained before using them to collect study samples.

ConclusionA device for consistently sampling approximately 10 μl of whole blood is demonstrated. The sampler displays the same benefits as those associated with DBS sampling. Furthermore, the device offers a number of additional benefits over that sampling system:

• Simplified sampling: the VAMS is used for accurate volume sample collection and storage/transport. This compares well with DBS, which requires a separate sample collection device (often a capillary) from which the blood is applied to the DBS substrate surface;

• Simplified analysis: the whole VAMS tip is extracted without further manipulation, whereas for DBS, the spot must be punched (either in its entirety or partially) prior to extraction. Further-more, the design of the VAMS is readily amena-ble to the tip-based automation systems that are present in most bioanalytical laboratories.

The effect of blood HCT on the volume of blood absorbed by the sampler was investigated in the laboratories of six different companies. This dem-onstrated minimal variation in the volume of blood collected with different HCTs and acceptable differ-ences in blood volumes collected between the partic-ipating laboratories using both human and rat blood. These results indicate that the issues associated with

www.future-science.com 657

Figure 3. Variation in the density of blood with hematocrit determined by each of the six participating laboratories. Laboratories 1, 2, 3 and 6 used human blood while laboratories 4 and 5 used rat blood. Comparisons to published values for human blood density are also illustrated [27]. Each data point is the average of six determinations made at each laboratory.

Blo

od

den

sity

(g

/ml)

Hematocrit (%)

1.300

1.200

1.100

1.000

0.900

0.800

0.700

0.600

0 20 40 60 80

Laboratory 1

Laboratory 2

Laboratory 4

Laboratory 5

Laboratory 6

Average

Published

Laboratory 3(1)

Laboratory 3(2)

future science group

A device for dried blood microsampling in quantitative bioanalysis Methodology

variations in the volume of blood analyzed with dif-ferent HCTs when using DBS sampling are mini-mized or eliminated with the novel VAMS.

While the VAMS device demonstrates much promise for the accurate collection of small blood volumes for quantitative bioanalysis and overcomes a number of issues associated with DBS sampling, further investigation is required in order to demon-strate its quantitative bioanalytical performance and practical use in busy clinical and laboratory settings. As such, a number of other characterizations of the VAMS for drug concentration measurement have been undertaken and will be reported in separate publications.

Future perspectiveFurther thorough investigations in order to fully understand the reality of the benefits and any possible issues associated with the bioanalytical and practical field performance of the VAMS device are required before its widespread practical application and regu-latory acceptance is realized. If successful, the VAMS has the potential to deliver simplified blood collec-tion, shipping, storage and analytical approaches for the quantitative bioanalysis of pharmaceuticals and biomarkers compared with currently accepted wet and dry processes. In particular, it is likely that the device will see rapid adoption for the collection of high-quality samples that are problematic for cur-rent processes (i.e., pediatric studies, subjects in

remote geographic locations, home sampling or serial sampling from rodent studies).

AcknowledgementsJanssen would like to thank D Van Roosbroek. BMS would like

to thank S Basdeo, C D’Arienzo, L Discenza and T Olah. Merck

would like to thank I Xie, L Xue and M Wang. The Neoteryx

and Phenomenex authors would like to formally acknowl-

edge GlaxoSmithkline’s N Spooner and P Dennif for their sig-

nificant contribution and support in the development of the

volumetric absorbtive microsampler technology. The authors

would like to formally acknowledge the scientific contribu-

tions of our colleagues in this pilot study who were involved

in the development of the volumetric absorbtive microsampler

technology and the Mitra™ microsampling device.

Financial & competing interests disclosureFinancial support for this work was provided by Neoteryx and

Phenomenex. The authors have no other relevant affiliations

or financial involvement with any organization or entity with

a financial interest in or financial conflict with the subject

matter or materials discussed in the manuscript apart from

those disclosed.

No writing assistance was utilized in the production of this

manuscript.

Open accessThis work is licensed under the Creative Commons Attribution-

NonCommercial 3.0 Unported License. To view a copy of this

license, visit http://creativecommons.org/licenses/by-nc-nd/3.0/

658 Bioanalysis (2015) 7(6) future science group

Methodology Spooner, Denniff, Michielsen et al.

ReferencesA paper of special note has been highlighted as: ••ofconsiderableinterest

1 Spooner N. A dried blood spot update: still an important bioanalytical technique. Bioanalysis 5(8), 879–883 (2013).

2 Timmerman P, White S, Cobb Z et al. Update of the EBF recommendation for the use of DBS in regulated bioanalysis integrating the conclusions from the EBF DBS microsampling consortium. Bioanalysis 5(17), 2129–2136 (2013).

3 Beaudette P, Bateman KP. Discovery stage pharmacokinetics using dried blood spots. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 809, 153–158 (2004).

4 Clark GT, Haynes JJ, Baylis MA, Burrows L. Utilization of DBS within drug discovery: development of a serial microsampling pharmacokinetic study in mice. Bioanalysis 2(8), 1477–1488 (2010).

5 Discenza L, Obermeier MT, Westhouse R, Olah TV, D’Arienzo CJ. A bioanalytical strategy utilizing dried blood spots and LC-MS/MS in discovery toxicology studies. Bioanalysis 4(9), 1057–1064 (2012).

6 Rahavendran SV, Vekich S, Skor H et al. Discovery pharmacokinetic studies in mice using serial microsampling, dried blood spots and microbore LC-MS/MS. Bioanalysis 4(9), 1077–1095 (2012).

7 Barfield M, Spooner N, Lad R, Parry S, Fowles S. Application of dried blood spots combined with HPLC-MS/MS for the quantification of acetaminophen in toxicokinetic studies. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 870, 32–37 (2008).

8 Spooner N, Lad R, Barfield M. Dried blood spots as a sample collection technique for the determination of pharmacokinetics in clinical studies: considerations for the validation of a quantitative bioanalytical method. Anal. Chem. 81, 1557–1563 (2009).

9 Spooner N, Ramakrishnan Y, Barfield M, Dewit O, Miller S. Use of DBS sample collection to determine circulating drug concentrations in clinical trials: practicalities and considerations. Bioanalysis 2(8), 1515–1522 (2010).

10 Xu Y, Woolf E, Agrawal NGB, Kothare P, Pucci V, Bateman KP. Merck’s perspective on the implementation of dried blood spot technology in clinical drug development – what, when and how. Bioanalysis 5(3), 341–350 (2013).

11 Edelbroek PM, van der Heijden J, Stolk LM. Dried blood spot methods in therapeutic drug monitoring: methods assays and pitfalls. Ther. Drug Monit. 31, 327–336 (2009).

12 D’Arienzo CJ, Li QC, Discenza L et al. DBS sampling can be used to stabilize prodrugs in drug discovery rodent studies without the addition of esterase inhibitors. Bioanalysis 2(8), 1415–1422 (2010).

13 Heinig K, Bucheli F, Hartenbach R, Gajate-Perez A. Determination of mycophenolic acid and its phenyl glucuronide in human plasma, ultrafiltrate, blood, DBS and dried plasma spots. Bioanalysis 2(8), 1423–1436 (2010).

14 Bowen C, Hemberger M, Kehler J, Evans C. Utility of dried blood spot sampling and storage for increased stability of photosensitive compounds. Bioanalysis 2(11), 1823–1828 (2010).

15 Liu G, Ji QC, Jemal M, Tymiak AA, Arnold ME. Approach to evaluating dried blood spot sample stability during drying process and discovery of a treated card to maintain analyte stability by rapid on-card pH modification. Anal. Chem. 83, 9033–9038 (2011).

16 Bowen CL, Volpatti J, Cades J, Licea-Perez H, Evans CA. Evaluation of glucuronide metabolite stability in dried blood spots. Bioanalysis 4(23), 2823–2832 (2012).

17 Ren X, Paehler T, Zimmer M, Guo Z, Zane P, Emmons G. Impact of various factors on radioactivity distribution in different DBS papers. Bioanalysis 2(8), 1469–1475 (2010).

18 O’Mara M, Hudson-Curtis B, Olson K, Yueh Y, Dunn J, Spooner N. The effect of hematocrit and punch location on assay bias during quantitative bioanalysis of dried blood spot samples. Bioanalysis 3(20), 2335–2347 (2011).

19 Denniff P, Spooner N. The effect of hematocrit on assay bias when using DBS samples for the quantitative analysis of drugs. Bioanalysis 2(8), 1385–1395 (2010).

20 Cobb Z, de Vries R, Spooner N et al. In-depth study of homogeneity in DBS using two different techniques: results from the EBF DBS-microsampling consortium. Bioanalysis 5(17), 2161–2169 (2013).

21 De Vries R, Barfield M, van de Merbel N et al. The effect of hematocrit on bioanalysis of DBS: results from the EBF DBS-microsampling consortium. Bioanalysis 5(17), 2147–2160 (2013).

22 Li F, Zulkoski J, Fast D, Michael S. Perforated dried blood spots: a novel format for accurate microsampling. Bioanalysis 3(20), 2321–2333 (2011).

23 Fan L, Lee JA. Managing the effect of hematocrit on DBS analysis in a regulated environment. Bioanalysis 4(4), 345–347 (2012).

Executive summary

• A blood sampler has been developed that simplifies the sample collection of a specific fixed volume and bioanalytical workflows compared to dried blood spot (DBS) and conventional plasma samples.

• Issues associated with sample homogeneity for DBS sampling are eliminated as the entire sample is extracted.• The device minimizes the effect of hematocrit on the volume of the sample analyzed that is observed for DBS

samples.• While the between-laboratory variation in human and rat blood volumes collected was within acceptable

limits, there is notable interoperator variability, which will need to be further investigated, and appropriate operator training will need to be conducted.

www.future-science.com 659future science group

A device for dried blood microsampling in quantitative bioanalysis Methodology

24 Meesters RJW, Zhang J, van Huizen NA, Hoof GP, Gruters RA, Luider TM. Dried matrix on paper discs: the next generation DBS microsampling technique for managing hematocrit effect in DBS analysis. Bioanalysis 4(16), 2027–2035 (2012).

25 Capiau S, Stove VV, Lambert WE, Stove CP. Prediction of the hematocrit of dried blood spots via potassium measurement on a routine clinical chemistry analyzer. Anal. Chem. 85, 404–410 (2013).

26 Denniff P, Spooner N. Volumetric absorptive microsampling: a dried sample collection technique for quantitative bioanalysis. Anal. Chem. 86, 8489–8495 (2014).

•• Firstreportofvolumetricabsorptivemicrosamplingforthecollectionoffixedbloodvolumesregardlessofhematocrit.

27 Density of Blood: The Physics Factbook. http://hypertextbook.com