12/11/2006 1
Met One Instruments BAM-1020 “2X”David Gobeli, Ph.D.
Met One Instruments
David Gobeli, Ph.D.
Met One Instruments
![Page 2: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/2.jpg)
12/11/2006 2
Agenda
� BAM-1020 “2X” Product Upgrade
� BAM-1020 “2X” Designation Test
• PM2.5
• PM10-2.5
• Operational advantages
• Sesmi-continuous v. continuous
• BAM v. other
![Page 3: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/3.jpg)
12/11/2006 3
BAM-1020 “2X” Upgrade
� Technical Modifications
� Performance Specifications
� Impact of changes on Designation
� PM10
� PM2.5
� PM10-2.5
![Page 4: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/4.jpg)
12/11/2006 4
BAM-1020 “2X”
� Improve sensitivity without sacrificing accuracy or simplicity/reliability of previous design
� Maintains simplicity/reliability of BAM-1020
� Maintains accuracy of BAM-1020
� Improves sensitivity of BAM-1020 by approximately 2x
![Page 5: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/5.jpg)
12/11/2006 5
BAM-1020 “2X” Reliability
� Technology is “stable”
� “2X” upgrade is more evolutionary than revolutionary
� BAM-1020 “footprint” does not change
![Page 6: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/6.jpg)
12/11/2006 6
BAM-1020 “2X” Simplicity
� Lower limit of detection (2σ) < 3.0 µg/m3
(1 hour)
� BAM-1020 “2X” not require auxiliary
equipment such as nephalometer, Nafion
dryers, complex plumbing, etc., etc., etc.
![Page 7: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/7.jpg)
12/11/2006 7
BAM-1020 “2X” Simplicity
BAM-1020 Flow Diagram Competing Technology
![Page 8: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/8.jpg)
12/11/2006 8
Technical Improvements
� Sensitivity enhanced by ~ 2 x
� Software� User-selectable beta count (4, 6 or 8
minutes)
� Temperature compensation added
� Hardware� Tape drive optimized for added sensitivity
� Source moved closer to detector
� Temperature compensation
![Page 9: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/9.jpg)
12/11/2006 9
BAM-1020 2X Specifications
14C < 100 µCiSource Composition and
Activity
< 0.3 µg/m3Root mean square
(RMS) error (24-hours)
< 1.5 µg/m3Root mean square
(RMS) error (1-hour)
< 0.6 µg/m3Lower Limit of Detection
(2σ) 24-hours
< 3.0 µg/m3Lower Limit of Detection
(2σ) 1 hour
4, 6 or 8 minutesCounting period per
measurement cycle
40-50 minutes user
selectable depending on
counting period per cycle
Sampling period per
measurement cycle
1 hour standard, others
available
Measurement Cycle
0-1,000 µg/m3 standardMeasurement Ranges
Performance SpecificationParameter
![Page 10: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/10.jpg)
12/11/2006 10
Additional Information
� All BAM-1020 monitors sold after mid-1999 can be upgraded
� PM10 Designation has been amended to reflect “2X” upgrade
� Upgrade price ~$2,800 plus PMT replacement if necessary.
![Page 11: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/11.jpg)
12/11/2006 11
BAM-1020 PM10 Designation
� Existing PM10 designation amended to reflect improvements (10/2006)
� 4, 6 or 8 minute count time are designated for PM10 equivalency
� Smart heater settings 35-45% RH
� ∆T control off
![Page 12: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/12.jpg)
12/11/2006 12
BAM-1020 PM2.5 Designation
� Met One will seek PM2.5 Class 3 FEM designation on BAM-1020 “2X”
� Site selection is in progress
� Winter sites needed in Midwest, California, and Rocky Mountain West cities
� Summer sites needed in California and Northeast cities
![Page 13: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/13.jpg)
12/11/2006 13
BAM-1020 PM2.5 Designation
� Current plans are to run designation testing at ≤ 35% moisture, ∆T off
� VSCC will be necessary
� No current plans to seek REM Designation
� States seeking REM should contact Met One on this matter
![Page 14: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/14.jpg)
12/11/2006 14
BAM-1020 PM2.5 Designation
� All BAM-1020 monitors sold after 1999 will be upgradeable to PM2.5 FEM designation once it is achieved
� Factory modification and recalibration will be required
![Page 15: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/15.jpg)
12/11/2006 15
BAM-1020 “2X” PM10-2.5
Designation� 2 BAM-1020 “2X” can be used with
optional software to provide � results
� Met One has also developed a single device capable of measuring PM10-2.5
� Met One is currently evaluating long term market potential for both approaches
![Page 16: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/16.jpg)
12/11/2006 16
![Page 17: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/17.jpg)
12/11/2006 17
Theory of Operation
� Beta rays are measured across clean filter tape. The measured value – I0
� Air containing particulate matter is sampled and deposited on the filter tape.
� Beta rays are measured across dirty filter tape. The measured value – I
![Page 18: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/18.jpg)
12/11/2006 18
BAM-1020 Calibration
� Calibration occurs once at the factory against a transfer standard in a “smoke room”
� There are NO environmental correction factors EVER that need to be employed
� Zero and Span are checked once per hour during operation. No changes are ever made however
![Page 19: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/19.jpg)
12/11/2006 19
Mass Calculation
0
0
1ln
xI I e
Ix
I
AxC
V
µ
µ
−=
=−
=
![Page 20: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/20.jpg)
12/11/2006 20
Mass Calculation
� DEFINITIONS:
� I – beta measurement across clean filter
� I0 – beta ray measurement across dirty filter
� µ – calibration coefficient (m2/µg)
� x – mass density on filter tape (µg/m2)
� A – dust spot deposition area on filter tape (m2 )
� V – sampled volume (m3)
� C – particulate matter concentration (µg/m3)
![Page 21: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/21.jpg)
12/11/2006 21
Semi-Continuous v. Continuous
� Aerosol chemistry is complex and depends on:
� Temperature
� Humidity
� Chemical composition
� Two step measurement process superior under these conditions
� Higher reliability
� Higher sensitivity
![Page 22: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/22.jpg)
12/11/2006 22
Semi-Continuous v. Continuous
� Tape advancement on regular basis reduces the “memory” issue
� Measurement is taking place in a controlled environment
� Source may be moved closer to the detector, thereby enhancing sensitivity
� Expensive, complex add-ons needed on monitors with concurrent sampling to overcome sensitivity problem
![Page 23: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/23.jpg)
12/11/2006 23
Semi-Continuous v. Continuous
� For customers requiring continuous monitoring, Met One can offer them the E-BAM or the ACE-BAM
![Page 24: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/24.jpg)
12/11/2006 24
E-BAM
![Page 25: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/25.jpg)
12/11/2006 25
ACE-BAM
![Page 26: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/26.jpg)
12/11/2006 26
FEM Designation
� US-EPA established uniform Reference Methods for PM10 and PM2.5 gravimetric measurement.
� 24-Hour average for automated methods must closely match Reference Methods in order to be considered accurate
![Page 27: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/27.jpg)
12/11/2006 27
Beta Gauge Advantages
� Beta attenuation technique is simpler, easier to use than other popular methods.
� Less expensive, lower cost of operation
� Filter replacement is predictable, less labor intensive
� Time proven technology: beta gauges have been in use for over 50 years.
� Beta attenuation has demonstrated superior accuracy compared to other popular methods.
![Page 28: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/28.jpg)
12/11/2006 28
Beta Gauge Advantages
� No make up gas required
� Technology is “stable”.
� Filter replacement is needed only once every two months under routine operating conditions
� Fewer site trips required
� Excellent accuracy compared to manual gravimetric technique
![Page 29: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/29.jpg)
12/11/2006 29
Beta Gauge Advantages
� Seasonal “Correction Factors” are not needed.
� Geographical “Correction Factors” are not needed.
� It is not necessary to ever calibrate the unit in the field
� BAM provides excellent agreement with all reference standards
![Page 30: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/30.jpg)
12/11/2006 30
Requires seasonal correction factors
Accurate under all climate conditions
Accuracy
Larger – especially with FDMS Option
Smaller“Footprint”
ComplexVery simpleComplexity
TEOM 1400ABAM-1020
BAM-1020 vs. TEOM Comparison
![Page 31: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/31.jpg)
12/11/2006 31
California Air Resources Board Data Summary
![Page 32: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/32.jpg)
12/11/2006 32
![Page 33: Met One Instruments BAM-1020 “2X” - US EPA · 2015-08-28 · BAM-1020 “2X” PM 10-2.5 Designation 2 BAM-1020 “2X” can be used with optional software to provide results](https://reader034.fdocuments.in/reader034/viewer/2022043002/5f7e395566d623427a3654db/html5/thumbnails/33.jpg)
12/11/2006 33
Questions?