Medical Image Analysis Medical Imaging Modalities Figures come from the textbook: Medical Image...
-
Upload
aubrey-jennings -
Category
Documents
-
view
215 -
download
0
Transcript of Medical Image Analysis Medical Imaging Modalities Figures come from the textbook: Medical Image...
Medical Image AnalysisMedical Image AnalysisMedical Imaging Modalities
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Anatomical or structural◦X-ray radiology, X-ray mammography,
X-ray CT, ultrasound, Magnetic Resonance Imaging
Functional or metabolic◦Functional MRI, (Single Photon
Emission Computed Tomography) SPECT, (Positron Emission Tomography) PET, fluorescence imaging
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
X-ray ImagingX-ray Imaging
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
39 P50N K
LO
N
Ejected Electron
Incident Electron
X-ray Photon
Figure 4.1. Atomic structure of a tungsten atom. An incident electron with energy greater than K-shell binding energy is shown interacting with a K-shell electron for the emission of an X-ray photon.
X-ray ImagingX-ray ImagingTungsten
◦K-shell binding energy level: 69.5 keV◦L-shell binding energy level: 10.2 keV◦An emision of X-ray photon of 59.3 keV
X-ray generation◦Electrons are released by the source
cathode and are accelerated toward the target anode in a vacuum under the potential difference ranging from 20,000 to 150,000 volts
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
X-ray 2-D Projection X-ray 2-D Projection ImagingImagingDiagnostic radiology
◦2-D projection of the three-dimensional anatomical structure of the human body
◦Localized sum of attenuation coefficients of material: air, blood, tissue, bone
◦Film or 2-D array of detectorsDigital radiographic system
◦Use scintillation crystals optically coupled with photomultiplier
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
X-ray Source
X-ray ScreenFilmX-ray Screen
3-D Object orPatient
2-D ProjectionImage
Anti-scatter Grid
Figure 4.2. (a). A schematic diagram of a 2-D X-ray film-screen radiography system. A 2-D projection image of the 3-D object is shown at the bottom. (b). X-ray radiographic image of a normal male chest.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
X-ray 2-D Projection X-ray 2-D Projection ImagingImagingScattering
◦Create artifacts and artificial structures
Reduce scattering◦Anti-scattered grids and collimators
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
X-ray MammographyX-ray MammographyTarget material
◦Molybdenum: K-, L-, M-shell binding energies levels are 20, 2.8, 0.5 keV. The characteristic X-ray radiation is around 17 keV.
◦Phodium: K-, L-, M-shell binding energies levels are 23, 3.4, 0.6 keV. The characteristic X-ray radiation is around 20 keV.
A small focal spot of the order of 0.1mm
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
X-ray Source
X-ray ScreenFilmX-ray Screen
CompressedBreast
MovingAnti-scatter Grid
CompressionDevice
Figure 4.3. A film-screen X-ray mammography imaging system.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.4. X-ray film-screen mammography image of a normal breast.
X-ray Computed X-ray Computed TomographyTomography3-D
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
dxzyx
inout ezxyIzxyI),,(
),;(),;(
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
y
x
zX-Y Slices
Figure 4.5. 3-D object representation as a stack of 2-D x-y slices.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
x
z
y
Iin(x; y,z)Iout(x; y,z)
(x,y; z)
11
22 92
15
12 42 52 62 72 82
Figure 4.6. Source-Detector pair based translation method to scan a selected 2-D slice of a 3-D object to give a projection along the y-direction.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.7: The translate-rotate parallel-beam geometry of first generation CT scanners.
X-ray Computed X-ray Computed TomographyTomographyGenerations
◦First: an X-ray source-detector pair that was translated in parallel-beam geometry
◦Second: a fan-beam geometry with a divergent X-ray source and a linear array of detectors. Use translation to cover the object and rotation to obtain additional views
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Generations◦Third: a fan-beam geometry with a
divergent X-ray source and an arc of detectors. Without translation. Additional views are obtained by simultaneous rotation of the X-ray source and detector assembly. “Rotate only”
◦ Fourth: use a detector ring around the object. The X-ray source provides a divergent fan-beam of radiation to cover the object
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.8. The first generation X-ray CT scanner
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Ring of Detectors
Source
SourceRotation Path
X-rays
Object
Figure 4.9. The fourth generation X-ray CT scanner geometry.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.10. X-ray CT image of a selected slice of cardiac cavity of a cadaver.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.11. The pathological image of the selected slice shown with the X-ray CT image in Figure 4.10
Magnetic Resonance Magnetic Resonance ImagingImagingNuclear magnetic resonance
◦The selected nuclei of the matter of the object
◦Blood flow and oxygenation◦Different parameters: weighted,
weighted, Spin-density◦Advance: MR Spectroscopy and
Functional MRI◦Fast signal acquisition of the order of
a fraction of a secondFigures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
1T 2T
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.12. MR images of a selected cross-section that are obtained simultaneously using a specific imaging technique. The images show (from left to right), respectively, the T1-weighted, T-2 weighted and the Spin-Density property of the hydrogen protons present in the brain.
Magnetic Resonance Magnetic Resonance ImagingImaging1H: high sensitivity and vast
occurrence in organic compounds13C: the key component of all
organic15N: a key component of proteins
and DNA19F: high relative sensitivity31P: frequent occurrence in organic
compounds and moderate relative sensitivity
Adapted from the Wikipedia, www.wikipedia.org.
MR SpectroscopyMR Spectroscopy
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
MR SpectroscopyMR Spectroscopy
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Functional MRIFunctional MRI
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
MRI PrinciplesMRI Principles : spin-lattice relaxation time : spin-spin relaxation time : the spin density
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
1T
2T
MRI PrinciplesMRI Principles
1. Great web sites1. Simulations from BIGS - Lernhilfe
für Physik und Technik2. http://www.cis.rit.edu/class/
schp730/bmri/bmri.htm
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MRI PrinciplesMRI PrinciplesSpin
◦A fundamental property of nuclei with odd atomic numbers is the possession of angular moment
Magnetic moment◦The charged protons create a
magnetic field around them and thus act like tiny magnets
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MRI PrinciplesMRI Principles : the spin angular moment : the magnetic moment : a gyromagnetic ratio, MHz/T
A hydrogen atom◦ :42.58 MHz/T
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
J
J
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
NN
SS
JJ
JJ
Figure 4.13. Left: A tiny magnet representation of a charged proton with angular moment, J. Right: A symbolic representation of a charged proton with angular moment, J and a magnetic moment, μ.
MRI PrinciplesMRI PrinciplesPrecession of a spinning proton
◦The interaction between the magnetic moment of nuclei with the external magnetic field
◦Spin quantum number of a spinning proton: ½
◦The energy level of nuclei aligning themselves along the external magnetic field is lower than the energy level of nuclei aligned against the external magnetic field
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.14 (a) A symbolic representation of a proton with precession that is experienced by the spinning proton when it is subjected to an external magnetic field. (b) The random orientation of protons in matter with the net zero vector in both longitudinal and transverse directions.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MRI PrinciplesMRI PrinciplesEquation of motion for isolated
spin
Solution:
J
kHHdt
Jd
00
kHdt
d
0
00 H
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Longitudinal Vector OX at the transverse position X
Net LongitudinalVector: Zero
Net TransverseVector: Zero
Net LongitudinalVector: Zero
Net TransverseVector: Zero
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Lower Energy
Level
Higher Energy
Level
S
N
H
Lower Energy
Level
Higher Energy
Level
S
N
H0
Figure 4.15 (a). Nuclei aligned under thermal equilibrium in the presence of an external magnetic field. (b). A non-zero net longitudinal vector and a zero transverse vector provided by the nuclei precessing in the presence of an external magnetic field.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Non-zero Net Longitudinal Vector
x
y
z
x
y
zH0
Net Zero Transverse Vector
MRI PrinciplesMRI PrinciplesThe precession frequency
◦Depends on the type of nuclei with a specific gyromagnetic ratio and the intensity of the external magnetic field
◦This is the frequency on which the nuclei can receive the Radio Frequency (RF) energy to change their states for exhibiting nuclear magnetic resonance
◦The excited nuclei return to the thermal equilibrium through a process of relaxation emitting energy at the same precession frequency
MRI PrinciplesMRI Principles90-degree pulse
◦Upon receiving the energy at the Larmor frequency, the transverse vector also changes as nuclei start to precess in phase
◦Form a net non-zero transverse vector that rotates in the x-y plane perpendicular to the direction of the external magnetic field
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
S
N
S
N x
y
z
Figure 4.16. The 90-degree pulse causing nuclei to precess in phase with the longitudinal vector shifted clockwise by 90-degrees as a result of the absorption of RF energy at the Larmor frequency.
MRI PrinciplesMRI Principles180-degree pulse
◦If enough energy is supplied, the longitudinal vector can be completely flipped over with a 180-degree clockwise shidf in the direction against the external magnetic field
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
S
N
S
N
x
y
z
Figure 4.17. The 180-degree pulse causing nuclei to precess in phase with the longitudinal vector shifted clockwise by 180-degrees as a result of the absorption of RF energy at the Larmor frequency.
MRI PrinciplesMRI PrinciplesRelaxation
◦The energy emitted during the relaxation process induces an electrical signal in a RF coil tuned at the Larmor frequency
◦The free induction decay of the electromagnetic signal in the PF coil is the basic signal that is used to create MR images
◦The nuclear excitation forces the net longitudinal and transverse magnetization vectors to move
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MRI PrinciplesMRI PrinciplesA stationary magnetization
vector
The total response of the spin system
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
N
nnM
1
1
0
2
)(
T
kMM
T
jMiMHM
dt
Md zzyx
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Dephasing
RF Pulse
Random Phase (Zero Transverse Vector)
In Phase Spin
Relaxation
Figure 4.18. The transverse relaxation process of spinning nuclei.
MRI PrinciplesMRI PrinciplesThe longitudinal and transverse
magnetization vectors with respect to the relaxation times
where
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
tiTtyxyx eeMtM 02/
,, )0()(
11 //0 )0()1()( Ttz
Ttzz eMeMtM
piyxyx eMM 0)0()0( ',',
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
t
t
Mx,y (t)
Mz (t)
Figure 4.19. (a) Transverse and (b) longitudinal magnetization relaxation after the RF pulse.
MRI PrinciplesMRI PrinciplesThe RF pulse causes nuclear
excitation changing the longitudinal and transverse magnetization vectors
After the RF pulse is turned off, the excited nuclei go through the relaxation phase emitting the absorbed energy at the same Larmor frequency that can be detected as an electrical signal, called the Free Induction Decay (FID)
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MRI PrinciplesMRI PrinciplesThe NMR spin-echo signal (FID
signal)
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
dxdydzezyxMS zyxizyx
zyx )(0 ),,(),,(
zyxzyxi
zyx dddeSMzyx zyx )(0 ),,(),,(
MR InstrumentationMR InstrumentationThe stationary external magnetic
field◦Provided by a large superconducting
magnet with a typical strength of 0.5 T to 1.5 T
◦Housing of gradient coils◦Good field homogeneity, typically on
the order of 10-50 parts per million◦A set of shim coils to compensate for
the field inhomogeneity Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Gradient Coils
Magnet
Gradient Coils
RFCoils
PatientPlatform
MonitorMonitor Data-Acquisition
System
Figure 4.20. A general schematic diagram of a MR imaging system.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
MR InstrumentationMR InstrumentationAn RF coil
◦To transmit time-varying RF pulses◦To receive the radio frequency
emissions during the nuclear relaxation phase
◦Free Induction Decay (FID) in the RF coil
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MR Pulse SequencesMR Pulse SequencesNMR signal
◦The frequency and the phaseSpatial encoding in MR imaging
◦Frequency encoding and phase encoding
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
x
z
y
y
xx
y
z
z
Sagital
Coronal
Axial
Figure 4.21 (a). Three-dimensional object coordinate system with axial, sagittal and coronal image views. (b): From top left to bottom right: Axial, coronal and sagittal MR images of a human brain.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
MR Pulse SequencesMR Pulse Sequences
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Z Gradient
90 RF Pulse(Slice Selection)
X Gradient
Phase-Encoding(x-scan selection)
Z Gradient
180 RF Pulse(Slice Echo Formation)
Y Gradient
Frequency Encoding(Read-Out Pulse)
Figure 4.22. (a): Three-dimensional spatial encoding for spin-echo MR pulse sequence. (b): A linear gradient field for frequency encoding. (c). A step function based gradient field for phase encoding.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Varying Spatially DependentLarmor Frequency
S
N
S
N
Linear Gradient
Precessing Nuclei
External Magnet
Positive PhaseChange
Negative PhaseChange
Phase -EncodingGradientStep
MR Pulse SequencesMR Pulse SequencesThe phase-encoding gradient
◦Applied in steps with repeated cycles◦If 256 steps are to be applied in the
phase-encoding gradient, the readout cycle is repeated 256 times, each time with a specific amount of phase-encoding gradient
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Spin Echo ImagingSpin Echo Imaging :
◦Between the application of the 90 degree pulse and the formation of echo (rephasing of nuclei
:◦Between the 90 degree pulse and
180 degree pulse
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
ET
2/ET
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
RF Energy: 90 Deg Pulse
Zero Net Vector:Random Phase
RelaxationDephasing
RF Energy: 180 Deg Pulse
Echo -Formation
RF Energy: 90 Deg Pulse
Zero Net Vector:Random Phase
In Phase
Rephasing
Echo -Formation
In Phase
Figure 4.23. The transverse relaxation and echo formation of the spin echo MR pulse sequence.
Spin Echo ImagingSpin Echo ImagingK-space
◦The placement of raw frequency data collected through the pulse sequences in a multi-dimensional space
◦By taking the inverse Fourier transform of the k-space data, an image about the object can be reconstructed in the spatial domain
◦The NMR signals collected as frequency-encoded echoes can be placed as horizontal lines in the corresponding 2-D k-space
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Spin Echo ImagingSpin Echo Imaging : the cycle repetition time weighted
◦A long and a long weighted
◦A short and a shortSpin-density
◦A long and a short
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
RT
2T
RT ET
1T
RT ET
RT ET
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90 deg RF pulse
180 deg RF pulse
Gz: Slice Selection Frequency EncodingGradient
Gx: Phase EncodingGradient
Gy: Readout Frequency EncodingGradient
TE /2
TE /2
TE
RF pulseTransmitter
NMRRF FIDSignal
Figure 4.24. A spin echo pulse sequence for MR imaging.
Spin Echo ImagingSpin Echo Imaging
The effective transverse relaxation time from the field inhomogeneities
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
12 1),,(),,( 0T
T
T
T RE
eezyxzyx
2
11
2*
2
H
TT
Spin Echo ImagingSpin Echo ImagingThe effective transverse
relaxation time from a spatial encoding gradient
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
2
11*
2**
2
Gd
TT
Echo Planar ImagingEcho Planar ImagingA single-shot fast-scanning
methodSpiral Echo Planar Imaging (SEPI)
◦where
)(1
)( tdt
dtG xx
)(1
)( tdt
dtG yy
tttx cos)(
ttty sin)(
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90 deg RF pulse
90 deg RF pulse
Gz: Slice Selection Frequency EncodingGradient
Gx: OscillatingGradient
Gy: Readout Gradient
RF pulseTransmitter
NMRRF FIDSignal
Figure 4.25. A single shot EPI pulse sequence.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
gy
gx
x
y
Figure 4.26. The k-space representation of the EPI scan trajectory.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
x
y
SEPI Trajectory
Data SamplingPoints
Figure 4.27. The spiral scan trajectory of SEPI pulse sequence in the k-space.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90 deg RF pulse
180 deg RF pulse
Gz: Slice Selection Frequency EncodingGradient
Gx Gradient
Gy Gradient
TE /2
TE /2
TE
RF pulseTransmitter
NMRRF FIDSignal
TD
Figure 4.28. The SEPI pulse sequence
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.29. MR images of a human brain acquired through SEPI pulse sequence.
Gradient Echo ImagingGradient Echo ImagingFast low angle shot (FLASH) imaging
◦Utilize low-flip angle RF pulses to create multiple echoes in repeated cycles to collect the data required for image reconstruction
◦A low-flip angle (as low as 20 degrees)◦The readout gradient is inverted to re-
phase nuclei leading to the gradient echo during the data acquisition
◦The entire pulse sequence time is much shorter than the spin echo pulse sequence
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Low Flip Angle RF pulse
Gz: Slice Selection Frequency EncodingGradient
RF pulseTransmitter
Gx: Phase EncodingGradient
Gy: Readout Frequency EncodingGradient TE
NMRRF FIDSignal
Figure 4.30. The FLASH pulse sequence for fast MR imaging.
Flow ImagingFlow ImagingTracking flow
◦Diffusion (incoherent flow) and perfusion (partially coherent flow)
◦The FID signal generated in the RF receiver coil by the moving nuclei and velocity-dependent factors
MR angiography
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90 deg RF pulse
180 deg (selective)RF pulse
Gz: Slice Selection Frequency EncodingGradient
Gx: Phase EncodingGradient
Gy: Readout Frequency EncodingGradient
TE /2
TE
RF pulseTransmitter
NMRRF FIDSignal
Figure 4.31. A flow imaging pulse sequence with spin echo.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.32: Left: A proton density image of a human brain. Right: The corresponding perfusion image.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90 degree RF pulse
Gz: Slice Selection Frequency EncodingGradient
RF pulseTransmitter
Gx: Phase EncodingGradient
Gy: Readout Frequency EncodingGradient
TE NMRRF FIDSignal
Next 90 degree RF pulse
TR
Figure 4.33. Gradient echo based MR pulse sequence for 3-D MR volume angiography.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.34. An MR angiography image.
Nuclear Medicine Imaging Nuclear Medicine Imaging ModalitiesModalitiesRadioactivity decay
Half-life of a radionuclide decay
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
teNtN )0()(
693.0
halfT
Nuclear Medicine Imaging Nuclear Medicine Imaging ModalitiesModalitiesThe radioactivity of a
radionuclide◦The average decay rate
◦Curie (CI) disintegrations per second
(dps)
◦Becquerel (Bq) One dpsFigures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Ndt
dN
10107.3
Single Photon Emission Single Photon Emission Computed TomographyComputed TomographyRadioisotope
◦The radioisotopes are injected in the body through administration of radiopharmaceutical drugs that metabolize with the tissue
Gamma rays◦The gamma rays from the tissue pass
through the body and are captured by the detectors surrounding the body to acquire raw data for defining projections
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Single Photon Emission Single Photon Emission Computed TomographyComputed TomographyRadionuclides
◦Thallium ◦Technetium◦Iodine◦Gallium
Gamma ray◦Decay by emitting gamma rays with
photon energy ranging from 135 keV to 511 keV
Attenuation xd eII
0
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Object EmittingGamma Photons
Scintillation Detector Arrays Coupled with
Photomultiplier Tubes
Figure 4.35. A schematic diagram of detector arrays of SPECT scanner surrounding the patient area.
Single Photon Emission Single Photon Emission Computed TomographyComputed TomographyScintillation detector
◦Barium fluoride◦Cesium iodide◦Bismuth germinate BGO
Photomultiplier tube
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.36. A 99Tc SPECT image of a human brain
Single Photon Emission Single Photon Emission Computed TomographyComputed TomographyAttenuation and scattering
◦Photoelectric absorption and Compton scattering
Poor in structural information◦Attenuation and scattering
Assessment of metastases or characterization of a tumor
Lower cost than PET
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Positron Emission Positron Emission TomographyTomographyConcept
◦Simultaneous detection of two 511keV energy photons traveling in the opposite direction
Radionuclides◦Decay by emitting positive charged
particles called positrons◦Fluorine 18-F◦Oxygen 15-O◦Nitrogen 13-N◦Carbon 11-C
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Object EmittingPositrons
Scintillation Detector Arrays
Point of Positron Emission
Point of Positron Annihilation
Position Dependent Photomultiplier Tubes
CoincidenceDetection
SystemComputer
Display
Detector Ring
Figure 4.37. A schemtaic diaggram of PET scanner.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Positron Emission Positron Emission TomographyTomographyAfter emission
◦Travel typically for 1-3 mm, losing some of its kinetic energy
◦The annihilation of the positron with the electron
◦Cause the formation of two gamma photons with 511keV traveling in opposite directions
◦Coincidence detection◦The point of emission of a positron is
different from the point of annihilation with an electron
Positron Emission Positron Emission TomographyTomographyRadiopharmaceutical
◦Fluorodeoxyglucose (FDG)◦Resolution and sensitivity of PET
imaging is significantly better than SPECT
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.38: Serial images of a human brain with FDG PET imaging.
Ultrasound ImagingUltrasound ImagingDiagnostic imaging
◦Anatomical structures, blood flow measurements and tissue characterization
◦Safety, portability, low-cost
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Ultrasound ImagingUltrasound ImagingVelocity
Relative intensity in dB
Shorter waves◦Better imaging resolution
Frequencies: 2 MHz to 5 MHz are common
c
2
110log10I
I
Reflection and Reflection and TransmissionTransmissionAcoustic impedance
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
cZ 0
21
122,1 ZZ
ZZR
21
22,1
2
ZZ
ZT
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
x1 x2 x3
I0
R0
T1,2
T3,4
T2,3
T4,3
T3,2
T5,4
Z1
Z2Z3
Z4 Z5
T2,1
Figure 4.39. A path of a reflected sound wave in a multilayered structure.
RefractionRefractionSnell’s law
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
1
2sinsinc
cit
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Piezoelectric crystal
Acoustic absorbers
Blockers
ImagingObject
Transmitter/Receiver
Circuit
ControlCircuit
PulseGeneration and
Timing
Data-Acquisition Analog to
Digital Converter
Computer Imaging Storage and Processing
Display
Figure 4.40. A schematic diagram of a conventional ultrasound imaging system.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure comes from the Wikipedia, www.wikipedia.org.
Ultrasound ImagingUltrasound ImagingA-mode
◦Records the amplitude of returning echoes from the tissue boundaries with respect to time
◦Perpendicular incident angle◦Basic method
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Ultrasound ImagingUltrasound ImagingM-mode
◦Variations in signal amplitude due to object motion
◦X-axis represents the time, while the y-axis indicates the distance of the echo from the transducer
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.41. M-Mode display of mitral valve leaflet of a beating heart.
Ultrasound ImagingUltrasound ImagingB-mode
◦Two-dimensional images representing the changes in acoustic impedance of the tissue
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.42. The “B-Mode” image of a beating heart with mitral stenosis.
Ultrasound ImagingUltrasound ImagingDoppler ultrasound imaging
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
c
cos2 ffdoppler
Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Figure 4.43. A Doppler image of the mitral valve area of a beating heart. Figures 4.4.3-5 are taken from the website http://www2.umdnj.edu/~shindler/ms.html.