Medical Engineering and Physics - DBIOM · and improve its performance [9,11,16–19,22–24]....
Transcript of Medical Engineering and Physics - DBIOM · and improve its performance [9,11,16–19,22–24]....
Medical Engineering and Physics 38 (2016) 671–678
Contents lists available at ScienceDirect
Medical Engineering and Physics
journal homepage: www.elsevier.com/locate/medengphy
Communication
A robust approach for ECG- based analysis of cardiopulmonary
coupling
Jiewen Zheng
a , Weidong Wang
b , Zhengbo Zhang
b , c , ∗, Dalei Wu
d , Hao Wu
b , Chung-kang Peng
e , f
a The Quartermaster Research Institute of the General Logistic Department, Beijing, P R China b Department of Biomedical Engineering, Chinese PLA (People’s Liberation Army) General Hospital, Beijing, P R China c Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA 02139, USA d Department of Computer Science and Engineering, The University of Tennessee at Chattanooga, Chattanooga, TN 37403, USA e Center for Dynamical Biomarkers and Translational Medicine, National Central University, Chungli, Taiwan f Division of Interdisciplinary Medicine & Biotechnology and Margret & H.A. Rey Institute for Nonlinear Dynamics in Medicine, Beth Israel Deaconess
Medical Center, Harvard Medical School, Boston, MA 02215, USA
a r t i c l e i n f o
Article history:
Received 8 January 2015
Revised 8 December 2015
Accepted 22 February 2016
Keywords:
ECG-derived respiration
Cardiopulmonary coupling
Adaptive filtering
Phase synchronization
Ensemble empirical mode decomposition
a b s t r a c t
Deriving respiratory signal from a surface electrocardiogram (ECG) measurement has advantage of simul-
taneously monitoring of cardiac and respiratory activities. ECG-based cardiopulmonary coupling (CPC)
analysis estimated by heart period variability and ECG-derived respiration (EDR) shows promising ap-
plications in medical field. The aim of this paper is to provide a quantitative analysis of the ECG-based
CPC, and further improve its performance. Two conventional strategies were tested to obtain EDR signal:
R-S wave amplitude and area of the QRS complex. An adaptive filter was utilized to extract the com-
mon component of inter-beat interval (RRI) and EDR, generating enhanced versions of EDR signal. CPC
is assessed through probing the nonlinear phase interactions between RRI series and respiratory signal.
Respiratory oscillations presented in both RRI series and respiratory signals were extracted by ensemble
empirical mode decomposition for coupling analysis via phase synchronization index. The results demon-
strated that CPC estimated from conventional EDR series exhibits constant and proportional biases, while
that estimated from enhanced EDR series is more reliable. Adaptive filtering can improve the accuracy
of the ECG-based CPC estimation significantly and achieve robust CPC analysis. The improved ECG-based
CPC estimation may provide additional prognostic information for both sleep medicine and autonomic
function analysis.
© 2016 IPEM. Published by Elsevier Ltd. All rights reserved.
1
e
[
t
s
i
p
k
e
c
w
G
a
u
t
s
n
R
i
p
d
c
n
a
h
1
. Introduction
Human cardiovascular and respiratory systems interact with
ach other and present effects of modulation and synchronization
1,2] . The cardiovascular and cardiorespiratory systems are charac-
erized by a complex interplay of several linear and nonlinear sub-
ystems [3] . Analysis of cardiopulmonary coupling (CPC) provides
nformation about the coupling dynamics between the cardiores-
iratory systems at different time scales [4] , leading to improved
nowledge of the interacting regulatory mechanisms under differ-
nt physiological and pathophysiological conditions [1] . Currently,
ardiac inter-beat (R–R) interval (RRI) dynamics analysis has been
idely investigated in the evaluation of autonomic nervous system
∗ Corresponding author. Department of Biomedical Engineering, Chinese PLA
eneral Hospital, 28 Fuxing Rd., Beijing, PR China. Tel.: + 86 10 66937921.
E-mail address: [email protected] (Z. Zhang).
p
s
p
d
ttp://dx.doi.org/10.1016/j.medengphy.2016.02.015
350-4533/© 2016 IPEM. Published by Elsevier Ltd. All rights reserved.
ctivities [5] . Compared with RRI dynamics analysis, CPC analysis
tilizes information presented in both cardiovascular and respira-
ory variables. It has been demonstrated that CPC analysis during
leep can characterize different sleep stages and identify sleep ap-
ea events, overcoming some potential limitations encountered in
RI dynamics analysis techniques [6,7] . A recent study shows the
mportance of monitoring CPC, proving that the strength of CPC
rogressively decreased as a function of sympathetic activation in-
uced by head-up tilt test [8] . Therefore, CPC analysis might over-
ome or at least complement traditional univariate analysis tech-
iques such as RRI dynamics analysis.
For the investigation of cardiorespiratory interaction, bivariate
pproaches are commonly applied [1] . For CPC analysis, both heart
eriod variability and respiratory signal are required. Respiratory
ignal is usually measured by motion or volume change during res-
iration, or by airflow [9] . Compared with other techniques, ECG-
erived respiration (EDR) is a more convenient and cost-effective
672 J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678
Fig. 1. Structure of adaptive estimation of respiratory signal by RLS algorithm, with
EDR series as the primary input and RRI series as the reference input.
m
i
h
w
a
0
2
n
o
7
[
a
r
r
p
y
c
a
2
2
p
f
i
t
e
i
i
2
t
Q
t
E
w
e
d
m
2
i
a
p
o
t
t
s
a
alternative for respiration monitoring, which derives respiratory
signal from ECG recording without using special-purpose hardware
[10] . Hence, with solely ECG recording, information of heart period
variability and respiratory signal can be derived and CPC analysis
can be achieved. Recent studies have demonstrated the effective-
ness of EDR and ECG-based CPC analysis as a screening or diag-
nostic tool for sleep disordered breathing [6,7,11,12] .
EDR is a signal-processing technique deriving respiratory wave-
forms from ordinary ECG [10,13] . Recently, many algorithms have
been developed such as amplitude modulation of ECG wave [14] ,
autoregressive model-based method on heart rate series [15] ,
band-pass filter on ECG waveform [16] , and other methods com-
bining both beat morphology and HR information [17,18] . For sin-
gle ECG recording, amplitude-based EDR algorithms have been re-
ported with satisfactory performance, especially in the context of
sleep apnea monitoring [19,20] . Two forms of amplitude-based al-
gorithms are commonly studied and implemented: R-S wave am-
plitude and area of the QRS complex [10,16,21] . As EDR is an in-
direct recovery of respiratory waveform from recorded ECG, it is
relatively noisy. Many studies have been performed to evaluate
and improve its performance [9,11,16–19,22–24] . However, most of
these studies focused on the accuracy of breathing rate detection
[9,16,18,22,24] . Little attention has been paid on the performance
of EDR for CPC analysis [6,12] .
The aim of our work is to provide a quantitatively evaluation
on the ECG-based CPC analysis, compared to CPC values assessed
from real respiratory signal such as respiratory inductive plethys-
mography (RIP). As EDR is a surrogate respiratory signal and rel-
atively noisy, a recursive least squares (RLS) adaptive filter is em-
ployed to improve its performance for CPC analysis. In this study,
CPC was assessed through probing the nonlinear phase interac-
tions between the corresponding components extracted from RRI
series and respiratory signals. The performance of the ECG-based
CPC analysis was tested on paced breathing (PB) dataset and Fan-
tasia Dataset [25] . In the PB dataset, a stepwise PB procedure was
performed to generate different levels of cardiorespiratory interac-
tion, testing the degree of agreement between the ECG-based CPC
and the gold standard through the range of measurements. As el-
derly subjects usually exhibit attenuated amplitude of respiratory
sinus arrhythmia (RSA) [26] , we are curious to explore the effec-
tiveness of the RLS adaptive filter in improving CPC estimation for
elderly subjects based on the Fantasia dataset. Two datasets both
have respiratory signals recorded by RIP. Compared with other res-
piration monitoring techniques, RIP has the advantages of greater
accuracy and better sensitivity [27] .
2. Materials and methods
2.1. Participants and protocol
2.1.1. PB dataset
We collected the PB dataset by a stepwise PB protocol, which
was reviewed and approved by the Ethics Committee of Chinese
PLA General Hospital. Twenty healthy volunteers (8 females and
12 males, with an average age of 30) participated in this experi-
ment. Volunteers were excluded when they had a history of respi-
ratory, cardiovascular or neurological diseases. During the exper-
iment, each volunteer underwent a stepwise PB procedure in a
sitting position at rest. The stepwise PB procedure consisted of
one 3-min spontaneous breathing session and six PB sessions. Each
PB session had a pre-defined breathing rate with a constant in-
spiration to expiration ratio (1:2), lasting 3 min . The six PB ses-
sions were arranged in a stepwise order with guiding breathing
rate changing from high to low in a protocol of [14, 12.5, 11,
9.5, 8 and 7] breath/min (BPM). During the stepwise PB procedure,
participants were guided by six melodies to perform the six PB
aneuvers. Each melody comprises two different tones, with a ris-
ng tone guiding inhalation while a lower tone accompanying ex-
alation. Physiological variables of ECG (lead II) and respiration
ere recorded during the experiment. All signals were sampled
t 10 0 0 Hz with amplifier bandwidths of 0.05–100 Hz for ECG and
.05–20 Hz for respiration [28] .
.1.2. Fantasia dataset
The Fantasia dataset includes continuous recording of sponta-
eous respiration and ECG signals from twenty young (21 –34 years
ld, mean age: 26) and twenty elderly (68 –85 years old, mean age:
5) rigorously screened healthy subjects in supine resting position
25] . Each subgroup of subjects includes equal numbers of men
nd women. In this dataset, each record includes ECG and respi-
ation sampled at 250 Hz, lasting 120 min . In our study, from each
ecord, we randomly extracted 5-min stable signals to compare the
erformance of the ECG-based CPC analysis to the reference anal-
sis from real respiratory signal. For the breath-by-breath phase
oupling algorithm used in our study (see Section 2.3 ), 5 min data
re enough to produce accurate CPC estimation.
.2. Signal pre-processing
.2.1. QRS complex detection
To achieve accurate QRS complex detection and EDR extraction,
ower line interference and baseline wander were first removed
rom raw ECG waveforms. ECG beat detection was performed us-
ng Hamilton & Tompin’s QRS detector [29] , and each beat annota-
ion was verified by visual inspection. With this algorithm, we got
ach R wave location and RRI time series. Normal-to-normal sinus
ntervals, extracted from the R –R interval time series, were linearly
nterpolated at an even interval of 0.2 s (i.e., 5 Hz).
.2.2. Deriving two basic types of EDR
The two commonly used EDR algorithms: R-S wave ampli-
ude (with EDR series noted as EDR_RS) [16,21] and area of the
RS complex [10] (with EDR series noted as EDR_area) were
ested to obtain EDR on a beat-to-beat basis. Both EDR_RS and
DR_area series were interpolated to 5 Hz. These two EDR series
ill further be processed by an adaptive filter to generate two
nhanced types of EDR. RIP signals in both the PB and Fantasia
atasets were also down-sampled to 5 Hz through a decimation
ethod.
.2.3. Enhanced types of EDR by adaptive filtering
The structure of the adaptive filter used in our study is shown
n Fig. 1 . In our study, we chose EDR series as the primary input
nd RRI series as the reference input to the adaptive filter. The out-
ut of the adaptive filter (EDR
′ in Fig. 1 ) is the common component
f respiration-induced oscillations presented in both EDR and RRI
ime series. RLS algorithm was employed by taking into account
he advantages of fast convergence speed and without eigenvalue
pread problem [18] . Based on the two basic types of EDR (EDR_RS
nd EDR_area), two enhanced types of EDR (noted as EDR_RS_filt
J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678 673
Fig. 2. Time series of RRI, RIP, EDR_RS, EDR_area, EDR_RS_filt and EDR_area_filt under various paced breathing conditions from one subject.
f
R
t
2
r
c
e
or EDR_RS and EDR_area_filt for EDR_area) were derived by the
LS adaptive filter, respectively. In our case, the forgetting factor of
he RLS algorithm was set to 0.996 and the length of the filter to
0 [18] .
With the aforementioned signal pre-processing steps, time se-
ies of RRI, RIP, EDR_RS, EDR_area, EDR_RS_filt and EDR_area_filt
an be acquired. Fig. 2 shows the time series of RRI and differ-
nt types of respiratory signals (RIP and EDR based) under various
674 J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678
Fig. 3. An example of the original time series and their corresponding dominant respiratory component (DRC) extracted from the original time series in the PB dataset. (a)
The original time series of RIP, RRI and different types of EDR, (b) DRCs extracted from the original time series by EEMD approach.
2
t
d
p
a
w
v
s
p
p
s
P
w
a
t
n
W
t
O
w
i
c
t
(
d
s
2
p
p
T
a
v
i
t
w
paced breathing conditions from one subject. For comparison and
clarity, all time series were normalized.
2.2.4. Extraction of respiratory component by EEMD
Cardiovascular and cardiorespiratory systems are characterized
by a complex interplay of several linear and nonlinear subsystems
[3] , and interactions between heart rate variability and respira-
tion are characterized by alternation of periods of quasiperiodicity
(weak coupling) with periods of relatively high coordinated activ-
ity [30] . Therefore, CPC is an intermittent phenomenon character-
ized by non linear dynamics and non-stationarities. To address this
challenge, ensemble empirical mode decomposition (EEMD) was
applied to all the cardiorespiratory series and respiration induced
oscillations presented in the respiratory band were extracted to as-
sess the CPC [31] .
EEMD is a noise-assisted EMD algorithm [31] , which was de-
signed to extract dynamical information from non-stationary and
nonlinear signals at different time scales [32,33] . The essence of
the EMD is to decompose a signal into different components called
intrinsic mode function (IMF) that satisfy the necessary conditions
for a meaningful Hilbert transform calculation. Compared with
EMD, EEMD can be used without subjective intervention and pro-
duces a set of IMFs that bear the full physical meaning by elim-
inating the issue of mode mixing [31] . In our case, since ventila-
tion is the driving factor for respiration-induced oscillations pre-
sented in all the series, we selected the IMF in the respiratory
band as the dominant respiratory component (DRC) [34] . For the
Fantasia dataset, it was observed that respiration-related oscilla-
tions in all cardiorespiratory series lie in IMFs 3-5, we identified
DRC in each cardiorespiratory series from IMFs 3-5 according to
the average frequency of each IMF. For the PB dataset, it was found
that respiration-related oscillations in all series lie in IMFs 3-6
during the PB procedure, however, the dominant respiratory IMF
might differ between different PB sessions. Therefore, for the PB
dataset, after decomposing each time series by EEMD, we further
segmented each series into seven subsections (one spontaneous
breathing subsection and six PB subsections) according the step-
wise PB protocol. Thus, for each series in each subsection, IMF in
the respiratory band was determined as the DRC. Fig. 3 shows the
original cardiorespiratory series and their corresponding DRCs ex-
tracted from the PB dataset.
.3. Phase coupling analysis
The concept of phase synchronization has been developed for
he identification of interactions between respiratory and car-
iovascular systems [1,35] , and recent studies have shown that
hase dynamics are more sensitive to coupling strength than other
mplitude-derived markers [34,36] . Therefore, in our study, CPC
as estimated from Hilbert transformation of the extracted DRC
ia phase synchronization index. With the extracted DRC from each
eries, Hilbert transformation was applied to provide instantaneous
hase of signal, and point-by-point phase difference between res-
iratory signal and RRI was calculated. We quantified the phase
ynchronization as
= ( cos ϕ i ) 2 + ( sin ϕ i )
2 (1)
here ϕ i is the point-by-point phase difference of paired series,
nd ( cos ϕ i ) and ( sin ϕ i ) represent the average value during a cer-
ain time interval [37] .
If phase difference ϕ i = ϕ Resp − ϕ RRI ( ϕ Resp for respiratory sig-
al and ϕ RRI for RRI) is constant, both series are synchronized.
hen respiratory signal and RRI are highly synchronized, the dis-
ribution of phase difference ϕ i should be within a narrow range.
therwise, for non-synchronized signals, the distribution should be
ide. The phase synchronization was quantified by this normalized
ndex ranging from 0 to 1, with larger values indicating higher syn-
hronization level between cardiorespiratory systems. Fig. 4 shows
he extracted DRCs from each time series at the PB rate of 14 BPM
from the same subject as in Fig. 2 ), along with the corresponding
istribution of the phase difference between RRI and respiratory
ignal (RIP and different types of EDR).
.4. Statistical analysis
To determine the performance of the ECG-based CPC analysis,
hase couplings assessed from different types of EDR were com-
ared with the gold standard assessed from RIP using both paired
-test and Wilcoxon rank sum test. The effectiveness of the RLS
daptive filter in improving the CPC estimation was tested. A p
alue of less than 0.05 (two-sided) was used as the level of signif-
cance. Results are reported as mean ± standard deviation. To test
he degree of agreement between the ECG-based CPC estimation
ith the gold standard, Bland –Altman plot was also performed
J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678 675
Fig. 4. An example of the extracted DRCs from each cardio-respiratory time series at the PB rate of 14 BPM, and the corresponding distribution of phase difference (a)
between RRI and respiratory signal ; (b) between RRI and RIP; (c) between RRI and EDR_area; (d) between RRI and EDR_RS; (e) between RRI and EDR_area_filt; (f) between
RRI and EDR_RS_filt .
[
R
3
3
d
E
e
c
d
d
s
C
s
t
b
T
f
t
e
S
s
s
t
n
3
d
e
t
c
s
e
b
t
i
m
38] . All statistical analyses were performed using MATLAB version
2010b (MathWorks, Natick, Massachusetts, USA).
. Results
.1. CPC estimation
The results of CPC estimation are shown in Table 1 for the PB
ataset. From Table 1 , it can be seen that the two basic types of
DR underestimate CPC in all breathing conditions, while the two
nhanced types of EDR produce more accurate CPC values very
lose to the gold standard estimated from RIP. As one subject un-
ertook seven breathing maneuvers during the stepwise PB proce-
ure, for each subject we took the mean value of seven CPC mea-
ures for statistical analysis. It was found that the differences in
PC estimation between the two basic types of EDR and the gold
tandard are relatively large and statistically significant. However,
here are no statistically significant differences in CPC estimation
etween the two enhanced types of EDR and the gold standard.
he results clearly indicate that adaptive filter can improve the per-
ormance of the ECG-based CPC estimation significantly for both
wo basic EDR algorithms. Moreover, there are no significant differ-
nces in CPC estimation between the two enhanced types of EDR.
imilar conclusions can be observed from the Fantasia dataset, as
hown in Table 2 . The results from the Fantasia dataset demon-
trated that even for aged group with potential attenuated ampli-
ude of RSA, ECG-based CPC estimation can also be improved sig-
ificantly by adaptive filtering.
.2. Bland –Altman plot of the CPC estimation
Fig. 5 shows the Bland –Altman plot of CPC estimation in the PB
ataset. It is very clear that both two enhanced types of EDR gen-
rate lower biases in CPC estimation (as shown in Fig. 5 (c) and (d))
han the basic types of EDR (as shown in Fig. 5 (a) and (b)). Similar
onclusion goes to the Fantasia dataset (Bland –Altman plot was not
hown). Moreover, from Fig. 5 (a) and (b), it was found that there
xists a proportional bias in CPC estimation from either of the two
asic types of EDR. The basic types of EDRs tend to produce bet-
er CPC estimations when the breathing rate gets lower. The ex-
stence of the proportional bias indicates that ECG-based CPC esti-
ation from either of the two basic types of EDR does not perform
676 J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678
Table 1
Comparison of CPC estimation between RRI and different types of respiratory signal in the PB dataset at different breathing conditions, with
CPC values assessed from RIP signal as gold reference. The results are reported as mean ± standard deviation.
Respiratory
signals
Different breathing conditions ( n = 20) Combined
results
( n = 20) SB 14 BPM 12.5 BPM 11 BPM 9.5 BPM 8 BPM 7 BPM
RIP 0.86 ± 0.10 0.93 ± 0.04 0.93 ± 0.05 0.93 ± 0.05 0.90 ± 0.08 0.88 ± 0.09 0.82 ± 0.15 0.89 ± 0.09
EDR_area 0.69 ± 0.13 0.80 ± 0.18 0.83 ± 0.11 0.83 ± 0.17 0.81 ± 0.16 0.79 ± 0.17 0.78 ± 0.15 0.79 ± 0.16 ∗
EDR_RS 0.75 ± 0.12 0.78 ± 0.14 0.82 ± 0.12 0.86 ± 0.09 0.84 ± 0.16 0.81 ± 0.20 0.80 ± 0.15 0.81 ± 0.14 §
EDR_area_filt 0.80 ± 0.17 0.86 ± 0.13 0.92 ± 0.08 0.93 ± 0.08 0.90 ± 0.12 0.89 ± 0.10 0.82 ± 0.13 0.87 ± 0.13 ♦
EDR_RS_filt 0.85 ± 0.08 0.90 ± 0.08 0.93 ± 0.06 0.93 ± 0.06 0.92 ± 0.07 0.89 ± 0.10 0.82 ± 0.16 0.89 ± 0.10 #
Note : SB: spontaneous breathing; BPM: breathing per minute. ∗ p < 0.001 for comparison of CPC estimation between RIP and EDR_area (both by paired T -test and Wilcoxon rank sum test) . § p < 0.001 for comparison of CPC estimation between RIP and EDR_RS (both by paired T -test and Wilcoxon rank sum test) . ♦ p = 0.302 by paired T -test and p = 0.352 by Wilcoxon rank sum test for comparison of CPC estimation between RIP and EDR_area_filt . # p = 0.757 by paired T -test and p = 0.979 by Wilcoxon rank sum test for comparison of CPC estimation between RIP and EDR_RS_filt .
Table 2
Comparison of CPC estimation between RRI and different types of respiratory signal in the Fanta-
sia dataset, with CPC values assessed from RIP signal as gold reference. The results are reported as
mean ± standard deviation.
Different groups RIP EDR_area EDR_RS EDR_area_filt EDR_RS_filt
Young group ( n = 20) 0.80 ± 0.10 0.58 ± 0.20 0.56 ± 0.14 0.80 ± 0.11 0.77 ± 0.12
Elderly group ( n = 20) 0.69 ± 0.19 0.48 ± 0.20 0.50 ± 0.18 0.71 ± 0.13 0.70 ± 0.12
Total ( n = 40) 0.75 ± 0.15 0.53 ± 0.20 0.53 ± 0.16 0.76 ± 0.13 0.74 ± 0.12
o
a
e
d
s
t
t
p
e
t
a
t
d
p
b
g
a
e
t
s
s
p
a
c
l
t
p
s
s
a
t
d
r
g
e
r
i
v
consistently through the range of measurements. In contrast, for
the enhanced types of EDR, there are no obvious proportional bi-
ases in CPC estimation. This result indicates that the enhanced
types of EDRs can produce robust ECG-based CPC estimations in
different breathing conditions.
4. Discussion
Prior work has documented the effectiveness of EDR for breath-
ing rate detection [16,22–24] . Several techniques for improving the
signal quality of EDR have been developed [24] . However, little at-
tention has been paid on the performance of EDR for CPC analy-
sis. In this study, we not only provided a quantitatively analysis on
the ECG-based CPC, but also used the RLS adaptive algorithm to
improve its performance. Moreover, to achieve more reliable phase
coupling analysis, EEMD was applied to cardiorespiratory signals to
extract respiration-related oscillations adaptively, instead of using a
low-pass filter with a constant cut-off frequency [37] . Our results
demonstrate that the basic EDR algorithms underestimate CPC in
all breathing conditions, while the enhanced types of EDR by adap-
tive filtering can produce more accurate and robust CPC estimation.
The Bland –Altman plots in Fig. 5 show that there are propor-
tional biases in CPC estimation for both two basic types of EDR.
The biases at high breathing rates are larger than those at low
breathing rates. This phenomenon is easy to explain. As breath-
ing rate gets lower, the synchronization between heart rate and
respiration becomes stronger. Therefore, basic form of EDR tends
to perform better in CPC analysis at low breathing frequency. The
enhanced types of EDR not only reduced the biases in CPC analy-
sis, but also eliminated the proportional biases. Thus, the enhanced
types of EDR can achieve robust ECG-based CPC analysis in differ-
ent breathing conditions.
As elderly subjects exhibit reduced amplitude of RSA [26] , the
effectiveness of adaptive filter in improving CPC estimation was
tested on elderly subjects by using the Fantasia dataset. The re-
sults clearly indicated that even for elderly subject group, CPC es-
timation could also be improved significantly by adaptive filter. In
this study, the forgetting factor of RLS filter was set to 0.996 and
the length of the filter to 20. Better results might be achieved by
ptimizing the parameters of the forgetting factor and filter length
ccording to the clinical applications under consideration.
One interesting phenomena was found that the strength of CPC,
ither estimated from the direct respiratory measurement or from
ifferent types of EDR-based measurements, was improved over
pontaneous breathing during the first three PB procedures, and
hen progressively deteriorated when the breathing rate was lower
han 9.5 BPM. Amplitude and phase dynamics are two major as-
ects in CPC investigation, and amplitude dynamics was consid-
red in most of studies [4] . In our previous study, we investigated
he effect of PB on the amplitude of RRI and demonstrated that the
mplitude-modulation effect of PB becomes more evident when
he breathing rate decreases [39] . The variation of phase coupling
uring the PB procedure indicates that in contrast to the RRI am-
litude, the PB rate might have an optimal range to achieve the
est coherent state between respiration and heart rate.
Our results provide compelling evidence that the basic EDR al-
orithms underestimate CPC, and the enhanced types of EDR by
daptive filtering are effective in improving CPC estimation. How-
ver, some limitations are worth noting. It should be noted that
he CPC investigated in our study is different from the ECG-based
leep spectrogram proposed by Thomas et al. [6] The ECG-based
leep spectrogram combines both amplitude and frequency cou-
ling, taking into account of coherence and cross spectrum of RRI
nd EDR [6] . The CPC investigated in our study reflects the syn-
hronization between respiration and RSA. Hence, it should be re-
ated to the coherence in the frequency domain. Therefore, to test
he value of our algorithm in sleep medicine, information of am-
litude coupling may also need to be included for coupling analy-
is. The future work of our study is to calculate the ECG-derived
leep spectrogram described in [6] by integrating adaptive filter
nd EEMD algorithms, testing whether these approaches can lead
o improvement in sleep quality measurement and sleep apnea
etection. Second, since the RLS filter uses the RRI series as one
eference input to enhance EDR quality, the performance of our al-
orithm might be compromised when the RRI series contains inad-
quate amount of oscillations caused by RSA. Therefore, an index
eflecting the amount of RSA oscillations in RRI series or an index
ndicating the EDR signal quality should be developed to super-
ise the performance of the algorithm during applications. Finally,
J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678 677
Fig. 5. Comparison of CPC estimation from different type of EDRs and from RIP signal, with CPC values assessed from RIP signal as gold reference. (a) Difference in CPC
estimation between RIP and EDR_area; (b) Difference in CPC estimation between RIP and EDR_RS; (c) Difference in CPC estimation between RIP and EDR_area _filt; (d)
Difference in CPC estimation between RIP and EDR_RS_filt.
f
r
v
5
h
e
p
a
p
a
a
i
C
E
p
t
J
A
F
F
e
o
2
R
uture work need to conduct to investigate the applications of the
obust ECG-based CPC analysis in the evaluation of autonomic ner-
ous system activity.
. Conclusion
In conclusion, CPC estimated from conventional EDR series ex-
ibits constant and proportional biases, while that estimated from
nhanced EDR series is more reliable. Adaptive filtering can im-
rove the accuracy of the ECG-based CPC estimation significantly,
chieving reliable and robust result. The methods developed in this
aper may lead to improved performance of the ECG-based CPC
nalysis for cardiorespiratory interaction and autonomic function
nalysis, providing additional prognostic information in sleep qual-
ty measurement and sleep apnea detection.
onflict of interest
The authors declare that they have no conflict of interest.
thical approval
We collected a paced breathing dataset according to a stepwise
aced breathing protocol, which was reviewed and approved by
he local ethics committee of Chinese PLA General Hospital. The
udgement’s reference number is IRB-BME-028.
cknowledgments
This project was supported by the National Natural Science
oundation of China (grant no. 61471398 ), Beijing Natural Science
oundation (grant nos. 3102028 , 31220343 ), General Logistics Sci-
nce Foundation (grant no. CWS11C108), and National Key Technol-
gy Research and Development Program (grant nos. 2013BAI03B04,
013BAI03B05).
eferences
[1] Schulz S , Adochiei FC , Edu IR , Schroeder R , Costin H , Bar KJ , et al. Cardiovas-cular and cardiorespiratory coupling analyses: a review. Philos Trans A Math
Phys Eng Sci 2013;371:20120191 . [2] Garcia AJ , Koschnitzky JE 3rd , Dashevskiy T , Ramirez JM . Cardiorespiratory cou-
pling in health and disease. Auton Neurosci 2013;175:26–37 .
678 J. Zheng et al. / Medical Engineering and Physics 38 (2016) 671–678
[
[
[
[
[
[
[
[
[3] Voss A , Schulz S , Schroeder R , Baumert M , Caminal P . Methods derived fromnonlinear dynamics for analysing heart rate variability. Philos Trans A Math
Phys Eng Sci 2009;367:277–96 . [4] Bahraminasab A , Kenwright D , Stefanovska A , Ghasemi F , McClintock PV . Phase
coupling in the cardiorespiratory interaction. IET Syst Biol 2008;2:48–54 . [5] Heart rate variability: standards of measurement, physiological interpreta-
tion and clinical use. Task Force of the European Society of Cardiologyand the North American Society of Pacing and Electrophysiology. Circulation
1996;93(5):1043–65 .
[6] Thomas RJ , Mietus JE , Peng CK , Goldberger AL . An electrocardiogram-basedtechnique to assess cardiopulmonary coupling during sleep. Sleep
2005;28:1151–61 . [7] Harrington J , Schramm PJ , Davies CR , Lee-Chiong TL Jr . An electrocardio-
gram-based analysis evaluating sleep quality in patients with obstructive sleepapnea. Sleep Breath 2013;17:1071–8 .
[8] Porta A , Bassani T , Bari V , Tobaldini E , Takahashi AC , Catai AM , et al. Mod-
el-based assessment of baroreflex and cardiopulmonary couplings duringgraded head-up tilt. Comput Biol Med 2012;42:298–305 .
[9] O’Brien C , Heneghan C . A comparison of algorithms for estimation of arespiratory signal from the surface electrocardiogram. Comput Biol Med
2007;37:305–14 . [10] Moody GB , Mark RG , Zocola A , Mantero S . Derivation of respiratory signals
from multi-lead ECGs. In: Proceedings of IEEE conference on computers in car-
diology, vol. 12; 1985. p. 113–16 . [11] Khandoker AH , Palaniswami M . Modeling respiratory movement signals dur-
ing central and obstructive sleep apnea events using electrocardiogram. AnnBiomed Eng 2011;39:801–11 .
[12] Schramm PJ , Thomas R , Feige B , Spiegelhalder K , Riemann D . Quantitative mea-surement of sleep quality using cardiopulmonary coupling analysis: a retro-
spective comparison of individuals with and without primary insomnia. Sleep
Breath 2013;17:713–21 . [13] Wang RC , Calvert TW . A model to estimate respiration from vectorcardiogram
measurements. Ann Biomed Eng 1974;2:47–57 . [14] Dobrev D , Daskalov I . Two-electrode telemetric instrument for infant heart rate
and apnea monitoring. Med Eng Phys 1998;20:729–34 . [15] Thayer JF , Peasley C , Muth ER . Estimation of respiratory frequency from autore-
gressive spectral analysis of heart period. Biomed Sci Instrum 1996;32:93–9 .
[16] Boyle J , Bidargaddi N , Sarela A , Karunanithi M . Automatic detection of respi-ration rate from ambulatory single-lead ECG. IEEE Trans Inf Technol Biomed
2009;13:890–6 . [17] Leanderson S , Laguna P , Sornmo L . Estimation of the respiratory frequency us-
ing spatial information in the VCG. Med Eng Phys 2003;25:501–7 . [18] Varanini M , Emdin M , Allegri F , Raciti M , Conforti F , Macerata A , et al. Adap-
tive filtering of ECG signal for deriving respiratory activity. Comput Cardiol
1990:23–6 . [19] Pogach MS , Punjabi NM , Thomas N , Thomas RJ . Electrocardiogram-based sleep
spectrogram measures of sleep stability and glucose disposal in sleep disor-dered breathing. Sleep 2012;35:139–48 .
[20] de Chazal P , Heneghan C , Sheridan E , Reilly R , Nolan P , O’Malley M . Automatedprocessing of the single-lead electrocardiogram for the detection of obstructive
sleep apnoea. IEEE Trans Biomed Eng 2003;50:686–96 .
[21] Nemati S , Malhotra A , Clifford GD . Data fusion for improved respiration rateestimation. EURASIP J Adv Signal Process 2010;2010:926305 .
22] Bailon R , Sornmo L , Laguna P . A robust method for ECG-based estimationof the respiratory frequency during stress testing. IEEE Trans Biomed Eng
2006;53:1273–85 . [23] Langley P , Bowers EJ , Murray A . Principal component analysis as a tool for an-
alyzing beat-to-beat changes in ECG features: application to ECG-derived res-piration. IEEE Trans Biomed Eng 2010;57:821–9 .
[24] Widjaja D , Varon C , Dorado AC , Suykens JA , Van Huffel S . Application of ker-
nel principal component analysis for single-lead-ECG-derived respiration. IEEETrans Biomed. Eng 2012;59:1169–76 .
25] Saeed M , Villarroel M , Reisner AT , Clifford G , Lehman LW , Moody G , et al. Mul-tiparameter intelligent monitoring in intensive care II: a public-access inten-
sive care unit database. Crit Care Med 2011;39:952–60 . 26] De Meersman RE . Aging as a modulator of respiratory sinus arrhythmia. J
Gerontol 1993;48:B74–8 .
[27] Carry PY , Baconnier P , Eberhard A , Cotte P , Benchetrit G . Evaluation of respi-ratory inductive plethysmography: accuracy for analysis of respiratory wave-
forms. Chest 1997;111:910–15 . 28] Zhang Z , Zheng J , Wu H , Wang W , Wang B , Liu H . Development of a respiratory
inductive plethysmography module supporting multiple sensors for wearablesystems. Sensors (Basel) 2012;12:13167–84 .
29] Hamilton PS , Tompkins WJ . Quantitative Investigation of Qrs detection
rules using the Mit/Bih arrhythmia database. IEEE Trans Biomed Eng1986;33:1157–65 .
[30] Schafer C , Rosenblum MG , Kurths J , Abel HH . Heartbeat synchronized withventilation. Nature 1998;392:239–40 .
[31] Wu Z , Huang NE . Ensemble empirical mode decomposition: a noise-assisteddata analysis method. Adv Adapt Data Anal 2009;1:1–41 .
32] Huang NE , Shen Z , Long SR , Shen SS , Qu WD , Gloersen P , et al. The empirical
mode decomposition and the Hilbert spectrum for nonlinear and non-station-ary time series analysis. Proc R Soc London Ser A 1998;454:903–93 .
[33] Janusauskas A , Marozas V , Lukosevicius A . Ensemble empirical mode de-composition based feature enhancement of cardio signals. Med Eng Phys
2013;35:1059–69 . [34] Manor BD , Hu K , Peng CK , Lipsitz LA , Novak V . Posturo-respiratory synchro-
nization: effects of aging and stroke. Gait Posture 2012;36:254–9 .
[35] Pereda E , De la Cruz DM , De Vera L , Gonzalez JJ . Comparing generalized andphase synchronization in cardiovascular and cardiorespiratory signals. IEEE
Trans Biomed Eng 2005;52:578–83 . 36] Hu K , Peng CK , Czosnyka M , Zhao P , Novak V . Nonlinear assessment of cere-
bral autoregulation from spontaneous blood pressure and cerebral blood flowfluctuations. Cardiovasc Eng 2008;8:60–71 .
[37] Cysarz D , Bussing A . Cardiorespiratory synchronization during Zen meditation.
Eur J Appl Physiol 2005;95:88–95 . [38] Bland JM , Altman DG . Statistical methods for assessing agreement between
two methods of clinical measurement. Lancet 1986;1:307–10 . 39] Wang B , Zhang Z , Wang W . Research on the respiratory sinus arrhythmia in
the process of guided breathing. J Biomed Eng 2012;29:45–50 69 .