Mechanical Haemolysisand Systemic Pressure
Thomas Ryzlewicz
![Page 2: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/2.jpg)
Measurement of the Systemic Pressure
= Pressure at the Entry of the Dialyzer
![Page 3: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/3.jpg)
why is the Systemic Pressure Measurement simply necessary?
• to detect a critical Elevation of the mechanical Resistance early in the entire Dialysis System (Dialyzer and Bloodline)
• Condensation of Labor and Experiece had developed unfavourably in nursing
![Page 4: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/4.jpg)
the Additional Mechanical Resistance in the Dialysis System
Resistance in and in front of the Dialyzer
Resistance of the Tubule
Resistance of the Vessel
Additional Resistance, comparable to ET Impedance
![Page 5: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/5.jpg)
the Additional Mechanical Resistance in the Dialysis System
Resistance in and in front of the Dialyzer
Resistance of the Tubule
Resistance of the Vessel
here no Alarm Monitoring(!)
Alarm Monitoring by venous Pressure
![Page 6: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/6.jpg)
. . . and now the Resistance in the Dialyzer increases . . .
Resistance in and in front of the Dialyzer
Resistance of the Tubule
Resistance of the Vessel
here no Alarm Monitoring(!)
Alarm Monitoring by venous Pressure
![Page 7: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/7.jpg)
. . . and now the Resistance in the Dialyzer increases . . .
Resistance in and in front of the Dialyzer
Resistance of the Tubule
Resistance of the Vessel
here no Alarm Monitoring(!)
Alarm Monitoring by venous Pressure
HAEMOLYSIS
![Page 8: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/8.jpg)
. . . und wenn jetzt der Widerstand im Dialysator steigt . . .
Widerstand im und vor dem Dialysator
Kanülen-Widerstand Gefäß-Widerstand
hier Alarmüberwachunghier keine Alarm-überwachung (!)
HAEMOLYSIS
![Page 9: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/9.jpg)
. . . und wenn jetzt der Widerstand im Dialysator steigt . . .
Widerstand im und vor dem Dialysator
Kanülen-Widerstand Gefäß-Widerstand
hier Alarmüberwachunghier keine Alarm-überwachung (!)
HÄMOLYSE !
![Page 10: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/10.jpg)
Online-HDF
Thomas Ryzlewicz
![Page 11: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/11.jpg)
![Page 12: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/12.jpg)
Low-flux-HD High-flux-HD
online-HDF
Pressures in the Dialyzer
![Page 13: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/13.jpg)
Low-flux-HD High-flux-HD
online-HDF
Pressures in the Dialyzer
![Page 14: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/14.jpg)
Hydraulics at Highflux-HD
![Page 15: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/15.jpg)
Hydraulics at online-HDF
Solvant Drag
![Page 16: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/16.jpg)
online-Therapie ist Herstellung von Infusionslösung!
European Pharmacopoeia 2005 kennt online-Herstellung von Infusionslösung nicht(!)
klar ist die Keimfreiheit für Infusionslösungen
ISO-Norm 11663 2009 Qualität von Dialysierflüssigkeiten und verwandten Therapien (steril und pyrogen-frei (=>EU<0,03/ml)
SAL > 6 (sterility assurance
level) entspr.10-8
Keimred./ml 3 Filter (ultra) 10-
11/ml
![Page 17: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/17.jpg)
online-Therapy is Production of Infusion-Fluid!
Tap-Water
Permat Standard
HD-Fluid
Ultra-pure Dialysis Fluid.
sterile u. pyrogen-free Infusion-Fluid
Softening + Reverse-Osmosis
Konzentrats
Ultrafiltration
Ultrafiltration
bakteriolog. Quality
CFU/mlEU/ml
Employment in the Dialysis
< 10-2
< 0,25
< 102
< 0,50
< 10-
1
< 0,03
SAL > 6
< 0,03
Basics for every Dialysis Fluid
Lowflux synthetic
Highflux-HD u. Low.-Vol.-HDF
online-HDF/HF Infusion-Fluid
According to I. LedeboISO-Norm 11663-
2009
![Page 18: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/18.jpg)
online-Therapy is Production of Infusion-Fluid!
Tap-Water
Permat Standard
HD-Fluid
Ultra-pure Dialysis Fluid.
sterile u. pyrogen-free Infusion-Fluid
Softening + Reverse-Osmosis
Konzentrats
Ultrafiltration
Ultrafiltration
bakteriolog. Quality
CFU/mlEU/ml
Employment in the Dialysis
< 10-2
< 0,25
< 102
< 0,50
< 10-
1
< 0,03
SAL > 6
< 0,03
Basics for every Dialysis Fluid
Lowflux synthetic
Highflux-HD u. Low.-Vol.-HDF
online-HDF/HF Infusion-Fluid
According to I. LedeboISO-Norm 11663-
2009
![Page 19: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/19.jpg)
online-Therapy is Production of Infusion-Fluid!
the Ultra-System:
one step Ultrafiltration
CFU Reduktion 10-5 SAL 3
two step Ultrafiltration
CFU Reduktion 10-8 SAL 6
three step Ultrafiltration(with U-2000-Ultrafilter)
CFU Reduktion 10-11 SAL 9
![Page 20: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/20.jpg)
how does a HDF-Regime look like?
Qb = 400 cc/min.
Qd = 500 cc/min. in total,
400 cc/min. of it diffusiv
and 100 cc/min. konvektiv
Infusate 6,0 Litres/h.
![Page 21: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/21.jpg)
HDF: Prädilution or Postdilution?
Disadvatage: Reduction of the Conzentration Gradient
Disadvantage: Lengthening of the distances of Diffusion by the packed RBC‘s
Disadvantage: high Sekundary Membrane by Protein
![Page 22: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/22.jpg)
HDF: Prädilution or Postdilution?
Prä-Dilution
Post-Dilution
Lengthening of the way of Diffusion
Sekundary Membrane
Reduction of the
Conzentration Gradient
![Page 23: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/23.jpg)
![Page 24: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/24.jpg)
![Page 25: Mechanical Haemolysis and Systemic Pressure Thomas Ryzlewicz.](https://reader034.fdocuments.in/reader034/viewer/2022042610/570491ca1a28ab14218dda64/html5/thumbnails/25.jpg)