Measurement of Flow - WordPress.comThis formula highlights Bernoulli's principle since if the speed...
Transcript of Measurement of Flow - WordPress.comThis formula highlights Bernoulli's principle since if the speed...
MEASUREMENT OF FLOW
TYPES OF FLOW
Two types of flow measurement
1. Gas Flow
2. Liquid Flow
FEW BASIC CONCEPTS
GAS FLOW IN BODY
Mechanical ventilation and
breathing monitoring are very
important practice to mechanically
assist patients or to perform
essential diagnoses. During both
artificial ventilation and function
respiratory monitoring the use of
flowmeters is required to perform
an accurate and continuous
monitoring of gas exchange.
GASES
• Gas volumes (and associated flows) can be measured directly using a
system with a calibrated chamber, but in clinical practice,
measurement is usually made indirectly, using a property of the gas
that changes in parallel to flow or volume and which can be more
easily determined.
GASES: SCIENTIFIC PRINCIPLES
Flow by definition is the quantity of the gas or fluid which passes a point in
unit time.
Relationship between flow and volume: for gas delivery systems (e.g.
oxygen flowmeters), flow is generally constant and the relationship
between flow and volume is expressed by
F = Q / t
F is equal to the mean flow, Q = the quantity (mass or volume) and T =
time.
TYPES
Flow can be laminar or turbulent. Several factors determine which
type of flow predominates, and these are combined into Reynolds’
number (Re), a dimensionless value that can be calculated for a
smooth parallel-sided tube by:
where v is linear velocity, p is density, l is diameter of tube and μ is
viscosity. At an Re less than 2000 flow will probably be laminar; an
Re over 2000 indicates flow is likely to be turbulent.
TYPES
Reynolds numbers for some real-life examples
• Blood flow in brain ~ 100
• Blood flow in aorta ~ 1000
• Typical pitch in major league baseball ~ 200000
• Blue whale swimming ~ 300000000
TYPES
Laminar flow is efficient, with layers passing
smoothly over each other producing a
parabolic (bullet-shaped) flow profile, with the
greatest velocity centrally. It is determined by
the Hagen-Poiseuille formula:
where: P is pressure drop, r is the radius of the tube and l is the length of the tube.
Important features are that flow is:
• directly proportional to the pressure drop
• proportional to the fourth power of the radius
• related to the viscosity but not the density of the gas.
TYPES
Turbulent flow is less efficient, with multiple eddy currents
occurring in the overall direction of flow. Because of the variable
nature of turbulence, there is no precise and comprehensive
equation to calculate flow, but turbulent flow is related to the:
• square root of the pressure drop
• density of the gas rather than its viscosity.
If flow is being calculated by measuring one of the above variables,
unforeseen changes in any of the others will compromise accuracy.
Changes could occur directly, such as by alteration in gas
composition, or indirectly, such as temperature variation causing a
change in viscosity.
BERNOULLI'S PRINCIPLE
pressure on the wider/slower side P1 has to be larger than the pressure on
the narrow/faster side.
"For incompressible, inviscid flow, Static Pressure + Dynamic Pressure + Hydrostatic
Pressure = CONSTANT along the STREAMLINE".
Assume that there is no change in the height of the fluid,
the pgh terms cancel if we subtract them from both sides.
This formula highlights Bernoulli's principle since if the speed v of a fluid is
larger in a given region of streamline flow, the pressure P must be smaller in
that region (which is Bernoulli's principle). An increase in speed v must be
accompanied by a simultaneous decrease in the pressure P in order for the
sum to always add up to the same constant number.
BERNOULLI'S PRINCIPLE
• This inverse relationship between the pressure and speed at a
point in a fluid is called Bernoulli's principle.
• Bernoulli's principle: At points along a horizontal streamline,
higher pressure regions have lower fluid speed and lower
pressure regions have higher fluid speed.
• It is the basics of most flow meters
PRINCIPALS FOR FLUID FLOW
METERING
PRINCIPALS FOR FLUID FLOW METERING
• Direct Measurement: gas volumes (and associated
flows) can be measured directly using bulk filling of an
enclosed space of known volume. Instruments using
direct measurement include the gas meter, and water-
displacement spirometer. Because of the logistical
problems of such devices, their use in clinical practice
is limited.
PRINCIPALS FOR FLUID FLOW METERING
Indirect measurement: for clinical use, measurement is usually made
indirectly, using a property of the gas that changes in parallel to flow or volume
and which can be more easily determined.
1. Pressure drop across a resistance – as flow occurs through a resistance,
a pressure change occurs. This effect can be used to calculate flow by:
keeping the resistance constant and measuring the pressure change as the
flow varies, as in a pneumotachograph having a constant pressure drop and
varying the resistance in a measurable way (e.g. bobbin rotameter).
2. Mechanical movement – flowing gas has kinetic energy related to its
velocity, which can be converted into a measurable value by: rotation of a
vane (e.g. a spirometer) bending a flexible obstruction, transducing this to
produce an electrical signal.
3. Heat transfer – gas flowing past a heated element acts to cool it, as in a
hot-wire anemometer.
4. Ultrasound interference – the velocity of an ultrasound signal is
increased by a gas flowing alongside it in the same direction, and
decreased if the gas is flowing against it.
PRINCIPALS FOR FLUID FLOW METERING
IDEAL FEATURES OF GAS FLOWMETER
Ideal features of a device used for gas flow or volume
measurement in clinical practice include:
• accurate across a wide range of flows
• unaffected by changes in gas temperature or composition
• low resistance so that it can be used in a spontaneously
breathing patient
• minimal impairment of performance with prolonged use.
MEASUREMENT SITE
For measuring tidal volume, it is important that the characteristics
of the breathing system are taken into consideration when deciding
where to measure.
e.g. Intensive care ventilators are non-rebreathing systems, and
measurements can be made accurately in the inspiratory or
expiratory parts of the system. In practice, both inspiratory and
expiratory tidal volumes are usually measured, because clinically
important discrepancies between them can occur, for example a
leak caused by lung pathology or the use of an uncuffed tube.
FLOWMETER TYPES:
1. Differential Pressure Flowmeters
In a differential pressure drop device the flow is calculated by measuring
the pressure drop over an obstructions inserted in the flow. The differential
pressure flowmeter is based on the Bernoullis Equation, where the
pressure drop and the further measured signal is a function of the
square flow speed.
Bernoulli Equation
FLOWMETER TYPES:
Differential Pressure Flowmeters: Types
Orifice Plate: The orifice plate is the simplest and cheapest. It is simply a plate with a hole of specified size and position cut in it, which can then clamped between flanges in a pipelineThe increase that occurs in the velocity of a fluid as it passes through the hole in the plate results in a pressure drop being developed across the plate.
Their accuracy are poor at low flow rates. A high accuracy depend on an orifice plate in good shape,
with a sharp edge to the upstream side. Wear reduces the accuracy.
THE ORIFICE PLATE
According to Bernoulli’s Principle, if you increase the speed of a flow you decrease its pressure
- a fact that we can exploit to measure flow using DP elements such as orifice plates. An orifice
plate is a precisely machined obstruction. As shown in the picture below, when your media
reaches the orifice, it is forced through a narrow hole in the center of the plate. By reducing
your flow path like this, you see changes in the speed and pressure of your media - on one side
of the plate the pressure is high, while on the other side the pressure is comparatively low. The
faster your media moves, the greater the difference in pressure.
ORIFICE PLATE
where phi is the volumetric flow rate,A1 is the inlet area of the flowmeter,
and A2 is the passage area of flow obstruction.
FLOWMETER TYPES:
Differential Pressure Flowmeters: Types
I. Orifice Plate
II. Venturi Tube
In the Venturi Tube the fluid flowrate is measured by reducing the cross
sectional flow area in the flow path, generating a pressure difference. After the
constricted area, the fluid is passes through a pressure recovery exit section,
where up to 80% of the differential pressure generated at the constricted area,
is recovered..
the Venturi tube flowmeter is often used
in applications where it's necessary with
lower pressure drops than the orifice
plate can provide.
Venturimeter: is a device used for measuring the rate of flow of a fluid flowing through a pipe. It
consists of three parts: • A short converging part • Throat • Diverging part
It consists of a cylindrical inlet section
equal to the pipe diameter; a converging
conical section in which the cross sectional
area decreases causing the velocity to
increase and the pressure head to
decrease;
a cylindrical throat section where the
velocity is constant so the decrease
pressure head can be measured; and a
diverging recovery cone where the velocity
decreased almost all of the original
pressure head is recovered.
1. The fluid enters the straight inlet section and a pressuretransmitter measures the static pressure at the high-pressure tap.
2. The fluid continues to the convergent section where the venturicross section reduces gradually. This causes the velocity of thefluid to increase and the static pressure to decrease.
3. A second tap allows pressure measurement at the center of thethroat, the narrowest section of the venturi. The flow rate of thefluid can be calculated from the static pressure drop between thehigh and the low pressure taps.
4. The divergent section of the venturi allows the fluid velocity todecrease and the pressure to recover most of its initial level.There is only a small permanent pressure loss over the venturi.
Venturi tubes, like all differential pressure flow producers, are based upon Bernoulli'stheory. They operate on the principal of a high pressure in the low velocity, largediameter inlet section.
Advantages and limitations of venturi tubes
Venturi tubes are highly accurate; they recover most of the pressure drop
they produce, and they are less susceptible to erosion than orifice plates
because of their smoother contour. However, venturi tubes are relatively
expensive and they require the use of a differential-pressure transmitter,
which contributes to the total cost of the flow measurement set up. They
tend to be voluminous and they may be difficult to install. Venturi tubes
also require a certain length of straight pipe both upstream and
downstream to ensure a flow that is undisturbed by fittings, valves, or
other equipment. However, the required pipe lengths are shorter than
those required for orifice plates.
FLOWMETER TYPES:
Differential Pressure Flowmeters: Types
I. Orifice Plate
II. Venturi Tube
III. Rotameters
Rotameters are widely used in gas delivery systems for
continuous flow measurement. Several designs have been
used, though now almost all use a vertical tapered tube
containing a bobbin or ball which is supported by the gas
flow as it passes upwards through the tube. The weight of
the ball (and thus the pressure drop required to support it) is
constant, but as the flow increases its position in the tube
rises, lowering the resistance as a larger pathway is created
alongside the ball.
• Rotameter continue….
Rotameters are known as gravity-type flowmeters because they
are based on the opposition between the downward force of
gravity and the upward force of the flowing fluid/Gas. When the
flow is constant, the float stays in one position that can be related
to the volumetric flow rate. That position is indicated on a
graduated scale.
Q = volumetric flow rate, e.g., liters per minute
k = a constant
A = annular area between the float and the tube wall
g = force of gravity
h = pressure drop (head) across the float
FLOWMETER TYPES:
Vane meters
The most common vane meter is the Wright’s
spirometer, in which the gas flow is directed
tangentially to strike a rotating vane in the gas
pathway. Originally this rotation was linked
mechanically to a needle and the volume read from
the adjacent dial, but modern versions use a light
source and photodetector positioned across the
vane to count its rotation. The Wright’s
spirometer tends to under-read at low flows
(because of friction) and to over-read at high
flows (because of momentum).Its clinical use is to assess conditions such as asthma, where the problem is largely confined to
airway resistance, which limits the expiratory flow rate. Although its use is limited in this respect,
it is a very simple and reliable bedside test.
• Vane meter: To convert rotations in voltage output
MECHANICAL FLOW TRANSDUCERS
3. Mechanical flow transducers
Another device using mechanical movement is the
flow transducer used in the Siemens intensive care
ventilators. The gas flow is split so that
measurement is made in a small side channel using
a thin metal disc supported on a flexible pin. The
disc is placed in the measuring channel at right
angles to the direction of gas flow, which results in it
being bent backwards by the flow. A strain gauge
situated immediately behind the pin is compressed
as it is bent, with a force dependent on the flow. The
resulting electrical signal is processed to calculate
the flow rate with a high degree of accuracy.
4. Hot wire anemometer
In this device an electrically
heated wire is placed in the
gas pathway, which is
cooled by the gas flow. The
degree of cooling depends
on the gas flow rate, which
can thus be calculated.
PRINCIPLES OF OPERATION
• Consider a thin wire mounted to supports and exposed to a velocity U. When a current is passed through wire, heat is generated ( I2Rw). In equilibrium, this must be balanced by heat loss (primarily convective) to the surroundings.
• If velocity changes,
convective heat transfer
coefficient will change, wire
temperature will change
and eventually reach a new
equilibrium.
Velocity U
Current I
Sensor (thin wire)
Sensor dimensions:length ~1 mmdiameter ~5 micrometer
Wire supports (St.St. needles)
PRINCIPLE OF OPERATION
HOT-WIRE STATIC TRANSFER FUNCTION
2
3
4
5
5 10 15 20 25 30 35 40
U m/sd
U/d
E/U
vo
lts^
-1
1,6
1,8
2
2,2
2,4
5 10 15 20 25 30 35 40
U m/s
E v
olts
Output voltage as fct. of velocity Voltage derivative as fct. of velocity
PROBE TYPES
1. Hot film , which is used in regions where a hot wire probe would quickly break such as in water flow measurements.
2. Hot wire , This is the type of hot wire that has been used for such measurements as turbulence levels in wind tunnels, flow patterns around models and blade wakes in radial compressors.
HOT WIRE SENSOR
10/48
HOT FILM SENSOR
11/48
MATERIAL USED IN MAKING SENSOR
• High Temperature Coefficient of resistance
• High Specific Resistance
• High Mechanical Strength
• Good Oxidation Resistance
• Low Thermal Conductivity
• Availability in small diameters
Tungsten : good strength, poor oxidation resistance
Platinum: good oxidation resistance, weak
Tungsten with thin platinum coating is generally used.
At high temperatures – Platinum‐iridium alloys, Platinumrhodium
alloys are used.
MODES OF ANEMOMETER OPERATION
1. Constant Current (CCA)
2. Constant Temperature (CTA)
INFUSION PUMPS
Infusion Pumps are an effective pathway for delivering fluid, blood, and
medicants to a patient's vital organs. Since the entire blood supply in a human
body completely circulates within 60 seconds, substances introduced into the
circulatory system are distributed rapidly.
An infusion device typically consists of three major components:
1. the fluid reservoir,
2. a catheter system for transferring the fluids into the body,
3. a device that combines electronics with a mechanism to generate
and regulate flow.
INFUSION PUMPS
Why are they used?
• To provide accurate and controllable flow over a prescribed period or on demand
What are they used for?
Wide range of drugs and therapies including
• Chemotherapy
• Pain management
• Total parental nutrition
• Anaesthesia/sedation
• Etc. etc.
Fluid container
Drip chamber
Roller Clamp
BASIC INFUSION SYSTEM
• Flow by gravity
• Flow controlled by roller clamp
• Difficult to set and control infusion
rate
AN ERROR WAITING TO HAPPEN!
• The infusion of drugs requires high flow rate accuracy and flow uniformity.
Sensors can be used to count the number of drops passing through the
drip chamber to provide flow feedback for automatic rate adjustment and
to detect any downstream occlusions directly below the pumping
mechanism. However, flow rate accuracy remains limited by the rate and
viscosity of the drip as well as improper angulation if in motion. Flow
uniformity can also suffer at low flow rates from the discrete nature of the
drop detector. Despite these limitations, a processor with an advanced
graphical user interface, smart and real-time physiological processing, and
wired and wireless connectivity options for patient monitoring and data
logging applications provides an additional level of safety by quickly
detecting complications and generating an alarm.
• One alternative to the drop sensor is a volumetric metering chamber. A
pump with a stepper or servo-controlled dc motor can be used to provide
the driving force for the fluid by mechanized displacement of the contents
in the volumetric chamber. The stepping resolution along with the
chamber elasticity can influence the flow uniformity. When the volume is
not uniform over the mechanism's cycle, software control can be used to
compensate for the variation.
INFUSION PUMPS
TWO BASIC TYPES
• Syringe Pumps
• Volumetric Pumps
SYRINGE PUMP
SYRINGE PUMPS
• Generally used for low volume, low flow rate infusions
• Good short term accuracy
• Long start up time at low flow rates
• Prime and purge line before connecting to patient
• Alarms: End/near end of infusion; drive disengaged, occlusion, battery low
• Specialised syringe pumps for ambulatory use, PCA, sedation, insulin etc
VOLUMETRIC PUMPS (PERISTALTIC PUMP)
Latch
Cam followers (fingers)
Pressure sensor
Air in line detector
VOLUMETRIC PUMPS
• Preferred for medium and high flow rates and large volumes
• Generally not suitable for rates < 5ml/h
• Variable short term accuracy
• Alarms: Latch/door open, set out, occlusion, battery low, air-in-line
• Specialised volumetric pumps for ambulatory use, epidural infusions etc.