Measurement Bench · Amplitude Modulation (AM) Waveform generator(8) 33220A (Agilent) The AM is a...
Transcript of Measurement Bench · Amplitude Modulation (AM) Waveform generator(8) 33220A (Agilent) The AM is a...
Dr. L.Scucchia
Measurement Bench
• Accessories
• Power supply
• Wave form generator
• Multimetre
• Oscilloscope
Dr. L.Scucchia
ACCESSORIES
Dr. L.Scucchia
Accessories (1)
Cables with BNC-croccodile connectors:
This kind of cable will be used to provide the
AC signal to the circuit to be measured.
Cables with Banana-croccodile connectors:
This kind of cables will be used to provide the
supply voltage to the circuit.
Dr. L.Scucchia
Compensate Probes
They are used to connect the oscilloscope to
the circuit to be measured.
They permit to cancel the effect of the equivalent
capacitance at the oscilloscope input.
Typically the probe is provided with a slide switch with two possible connections:
1X no compensation
10X the signal is attenuated by 10 times and there is compensation
Accessories (2)
10
1 9
20
9 1 20
Oscill . Signal
XOscill . Signal
X
X
MV f V
M M
C pFV f V
C pF pF
M C pF M pF
Dr. L.Scucchia
The breadboard or card for prototypes is a device easy to use. It allows you to
create circuit prototypes without soldering, with the advantage, then, you
can easily modify the prototypes.
Accessories (3)
Dr. L.Scucchia
Inside the device consists of several series of metal clips.
The arrangement of the clip on the breadboard produces rows of five hole
connected together (socket strip) and some collector lines (bus strip): the first
are normally used for the connection of the electronic components whereas the
latter are employed for the connections of supply or ground
Accessories (4)
The collector lines typically are
disconnected to the center.
Dr. L.Scucchia
Accessories (5)
The component connection is achieved by inserting the
connectors of the components in holes electrically
connected. The figure shows the case of two
resistors connected in series.
Particular care must be taken to connect
components with three terminals. For
example in the case of the transistor; C, B
and E must be connected to 3 different
rows of holes.
The components with dual in-line
package must be placed across
the center line of the
breadboard, in this way each leg
may be connected to a different
row of holes. E B
C
E B C
Dr. L.Scucchia
POWER SUPPLY
The power supply is a dc voltage generator, and it is the tool through which it
is possible to provide the electric power necessary for the proper operation of
the circuit to be measured.
Dr. L.Scucchia
Power supply (5)
E3630A (Agilent)
1. Power Switch
2. To view the 0 +6 V output
3. To view the 0 +20 V output
4. To view the 0 -20 V output
5. Voltage output from 0 to +6 V
6. Common terminal
7. Voltage output from 0 to +20 V
8. Voltage output from 0 to -20 V
9. Terminal for the ground connection
10.Alignment control knob
11.± 20V Control knob output
12.6VControl knob output
13.Overload LED
14.Current display
15.Voltage display
1
2 3 4
9 8 7 6 5
12 11 10 14 13
15
Dr. L.Scucchia
• This power supply provides:
2 outputs in tracking mode with a maximum voltage ±20V (0.5A),
1 output with a maximum voltage +6 V (2.5A).
• The outputs in tracking mode can be used in series to provide a voltage of 40V
(0.5A).
• if a load change causes the current limit to be exceeded, the OVERLOAD
LED is lighted.
• The three outputs share the COM terminal, isolated from the grounding of the
chassis.
• The reference mass is achieved by connecting the terminal COM to the
terminal ┴
Alimentatori (6)
E3630A (Agilent)
Dr. L.Scucchia
V
Vmax V’
I
Voltage control
Overload
Imax
Dr. L.Scucchia
WAVEFORM GENERATOR
Dr. L.Scucchia
Waveform generator (1)
It generates a periodic waveform that can be injected at the input of the circuit
to be measured.
Connecting the oscilloscope probes at the input and output of the circuit one can
evaluate the behavior of the circuit with respect to the injected waveform.
The output terminal of the function generators commonly consists of a BNC
connector with an impedance of 50
Typical signals which can be obtained are:
Sinusoidal waveforms.
Triangular waveforms.
Rectangular waveforms.
Dr. L.Scucchia
Waveform generator(2)
33220A (Agilent)
9 Manual Trigger Key
10 Output Enable/Disable Key
11 Knob
12 Cursor keys
13 Sync Connector
14 Output Connector
15 Numeric pad
16 Display
1 On/Off Switch
2 Graph Mode Key
3 Modulation/Sweep/Burst Keys
4 State Storage Menu Key
5 Utility Menu Key
6 Help Menu Key
7 Menu Operation Softkeys
8 Waveform Selection Keys
2
10 1 4
16 11
12
5 6
8
7
9 13 14 3
15
Dr. L.Scucchia
1. Press and hold any function key or softkey to display a help topic for that key or
feature. Or press Help to display the Help Menu.
2. Lighted keys indicate active keys or functions such as the currently active
waveform .
3. No signal is output unless the output key is lighted.
4. To select dc volts from the front panel, press Utility and select the DC On softkey.
5. The key Graph toggles between Graph Mode (lighted) and Menu Mode.
6. In Menu Mode, the softkeys allow you to select parameters and functions as
shown in the softkey menu at the bottom of the display. Some softkeys toggle
between related parameters. For example, the left softkey toggles between Freq
and Period below:
Referent points
Waveform generator(3)
33220A (Agilent)
Dr. L.Scucchia
Quick Start
Waveform generator(4)
33220A (Agilent)
1. Press the On/Off Switch. After a few seconds,
the instrument assumes the default values.
2. Select a waveform.
3. Change Waveform Parameters.
Use the six softkeys to select parameters.
Some softkeys toggle between related
parameters (for example: Freq and Period).
4. Parameter value can be changed using
the knob. First, use the cursor keys ( )
to select the first digit. Then use the knob to
change the first digit value
Dr. L.Scucchia
Waveform generator(5)
33220A (Agilent)
6 Now, press Offset to select that
parameter. Use the cursor keys to select
digits. Use the knob to set the required
offset value.
7. Now let‟s use the numeric keypad to set
the Period. Press the left softkey once to
select Freq.
8. Then press it again to select Period.
9. Key a value on the keypad. The display
changes. For example, enter 1.5.
10.Press the softkey for the desired unit (for
example, msec) to enter the value.
Dr. L.Scucchia
Waveform generator(6)
33220A (Agilent)
If RLoad = 50 , VLoad= 1.000Vpp
VOutput=2.000 Vpp
ROutput=50
RLoad +
VLoad
-
The 33220A has a fixed output impedance (ROutput) of 50 at the output connector.
If the load impedance (RLoad) is different from 50, the displayed amplitude and offset
levels will be incorrect.
For example, selecting a voltage 1.00 0Vpp, Voutput is equal to 2.000 Vpp
If Rload is different from 50
RLoad = 100 VLoad= 1.333Vpp
RLoad = VLoad= 2.000Vpp
Dr. L.Scucchia
If you do not know the load value,
you must estimate the voltage levels by using
the oscilloscope.
Dr. L.Scucchia
Waveform generator(7)
33220A (Agilent)
To Change the Output Termination
1 Press.
2 Navigate the menu to set the output termination. Press the Output Setup
softkey and then select the Load softkey.
3 Select the desired output termination.
Dr. L.Scucchia
Amplitude Modulation (AM)
Waveform generator(8)
33220A (Agilent)
The AM is a technique used in communication in order to transmit the
information via a carrier waveform (535 kilohertz to 1.7 megahertz).
The information is contained in the modulating waveform.
In AM, the amplitude of the carrier is varied by the modulating waveform.
carrier waveform
modulating waveform
Dr. L.Scucchia
1 Select the function, frequency, and amplitude of the carrier.
40.0
50.00
5 Select the modulating waveform shape: Press the Shape softkey to select the
shape of the modulating waveform. For this example, select a sine wave.
2 Select AM: Press and then select “AM” using the
Type softkey.
Waveform generator(9)
33220A (Agilent)
3 Set the modulation depth: Press the AM Depth
softkey and then set the value to 40% using the
numeric keypad or the knob and cursor keys.
4 Set the modulating frequency. Press the AM
Freq softkey and then set the value to 50 Hz using
the numeric keypad or the knob and cursor keys
Modulated Waveform at Output
Dr. L.Scucchia
MULTIMETRE
Dr. L.Scucchia
Multimetre (1)
The Multimeter is a measuring instrument that basically allows to measure currents,
voltages and resistances. There are two types of multimetres:
analog (MA) and
digital (MD).
The first employs a mechanical tracking system that shows the desired size on a
suitable scale, the latter uses a digital system to convert electrical quantities at
the input in a reading on the display.
Technically speaking the MA are less accurate than the digital ones (they are subject
to an error typically higher than 3%), are harder to read. Despite these limitations,
MA are better for measurements on circuits that have high electrical noise.
In the following, digital
multimetre will be considered, in
particular the 34401A (Agilent).
Dr. L.Scucchia
Multimetre (2)
34401A (Agilent)
1 2
3 4 5 6 7 8
1 Measurement Function keys
2 Math Operation keys
3 Single Trigger/Autotrigger/
Reading Hold key
4 Shift / Local key
5 Front / Rear Input Terminal Switch
6 Range / Number of Digits
Displayed keys
7 Menu Operation keys
8 Power switch
Turn on the multimeter.
Switching the instrument on, by pressing the 8 key, the display lights and the
meter performs a self test. To control the appearance of the display, hold down
the Shift key (4). To complete the self-test and hold for 5 seconds the Shift key.
Pressing the Shift key you can select the alternate function (in blue above the
key).
Dr. L.Scucchia
Multimetre(2)
34401A (Agilent) To Measure Voltage
Ranges: 100 mV, 1 V, 10 V, 100 V,
1000 V (750 Vac)
Maximum resolution: 100 nV (on 100
mV range)
AC technique: true RMS, ac-coupled
The 34401A provides “true RMS” (Root Mean Square) measurements. They are achieved
by the dissipated power in a resistor.
This power is proportional to the square of the true RMS voltage (VRMS), and it is
independent of waveshape.
A common way to describe signal waveshapes is crest factor. Crest factor is the ratio of the
peak value to RMS value of a waveform.
2
0
1 T
RMSV V t dtT
2
RMSdis
VP
R
ˆˆ. . ;
RMS
VC F V peak value
V
Dr. L.Scucchia
To Measure Resistance
Ranges: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1
MΩ, 10 MΩ, 100 MΩ
Maximum resolution: 100 μΩ (on 100
ohm range)
The Agilent 34401A offers two methods for measuring resistance: 2-wire and 4-wire.
For both methods, the test current flows from the input HI terminal ad then through the resistor
being measured.
For 2-wire, the voltage drop across the resistor to be
measured is sensed internal to the multimeter.
Therefore, cable resistances are also measured.
For 4-wire, separate cable connections are required.
Since no current flows in the measurement cables, these
does not give a measurement error. This type of
measurement is more accurate since it allows to eliminate
the resistance introduced by the connecting cables
Multimetre(3)
34401A (Agilent)
Dr. L.Scucchia
To Measure Current
Ranges: 10 mA (dc only), 100 mA (dc
only), 1 A , 3 A
Maximum resolution: 10 nA (on 10 mA
range)
AC technique: true RMS, ac-coupled
Multimetre(4)
34401A (Agilent)
To Measure Frequency (or Period)
Measurement band: 3 Hz to 300 kHz
(0.33 sec to 3.3 μsec)
Input signal range: 100 mVac to 750
Vac
Dr. L.Scucchia
To Test Continuity
Test current source: 1 mA
Maximum resolution: 0.1 Ω (range is fixed
at 1 kohm)
Beeper threshold: 1 Ω to 1000 Ω (beeps
below adjustable threshold)
To Check Diodes
Test current source: 1 mA
Maximum resolution: 100 μV (range is
fixed at 1 Vdc)
Beeper threshold: 0.3 volts ≤ Vmeasured ≤
0.8 volts (not adjustable)
Multimetre(5)
34401A (Agilent)
Dr. L.Scucchia
Math Operations
There are five math operations available, only one of which can be enabled at a
time. Each math operation performs a mathematical operation on each reading or
stores data on a series of readings.
The table shows the math/measurement function combinations allowed.
Null: value subtracted from the measurement.
Min/Max: displays minimum and maximum values for the triggered set
of readings.
dB: measurement relative to value in „dB Relative‟ register
dBm: measurement relative to a resistance referenced to 1 mW
Multimetre (6)
34401A (Agilent)
The math operation Limit can be enabled using the LIMIT TEST command in the
MATH MENU.
Dr. L.Scucchia
To Select a Range
The measurement range depends on the selected function:
Voltage 100 mV, 1 V, 10 V, 100 V, 1000 V
Current 10 mA (dc only), 100 mA (dc only), 1 A , 3 A
Frequency 3 Hz to 300 kHz
Resistance 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ
You can let the multimeter automatically select the range using autoranging or
you can select a fixed range using manual ranging.
Multimetre (6)
34401A (Agilent)
Dr. L.Scucchia
To Set the Resolution
Resolution is the number of digits the multimeter can measure or display.
You can set the display resolution to 4 ½, 5 ½, or 6 ½ digits either to optimize
measurement speed or noise rejection.
The most significant digit (leftmost on the display) is the “1⁄2” digit, since it can only
be a “0” or “1.”
Multimetre(6)
34401A (Agilent)
Press the shift key Selects 4 ½ digits
Selects 5 ½ digits
Selects 6 ½ digits
To Set the Trigger
Measurements are triggered from the front panel using the keys:
Single Trig: disables auto-triggering (default) and issues a single trigger to the
34401A each time the key is pressed.
Auto: enables auto-triggering if the 34401A is in single trigger mode.
Hold: displays the measurement after three consecutive readings.
Dr. L.Scucchia
Front panel menu
It provides access to a subset of the 34401A functionality. The menu is organized in
a top-down tree structure with three levels (menus, commands, and parameters).
Multimetre (6)
34401A (Agilent)
You move down ∨ or up ∧ the menu tree to get
from one level to the next. Each of the three
levels has several horizontal choices which you
can view by moving left < or right > .
Dr. L.Scucchia
Multimetre(6)
34401A (Agilent)
Dr. L.Scucchia
OSCILLOSCOPE
Dr. L.Scucchia
Oscilloscope (1)
The oscilloscope allows to display a voltage (vertical axis - Y axis) versus
time (horizontal axis - X axis).
The oscilloscope can also display two voltages on both the axes (XY mode).
The quantities shown on the screen are voltages.
To facilitate reading of the magnitudes, on the screen there is a grid
consisting of some horizontal and vertical divisions.
The oscilloscopes in the laboratory are DSO3202A by Agilent. Digital
Storage Oscilloscope (DSO)
A digital oscilloscope uses an analog-to-digital converter (ADC) to convert
the measured voltage into digital information.
The digitized data are stored in a “channel memory”
Dr. L.Scucchia
Oscilloscope (2)
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
Yellow = Channel specific blocks
Blue = System blocks
Dr. L.Scucchia
Oscilloscope (3) A probe feeds the input signal into the oscilloscope where the
attenuator and the amplifier are used to compensate for input
levels for various voltage per division settings. The trace position
is adjusted changing the DC Offset.
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
Yellow = Channel specific blocks
Blue = System blocks
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
Dr. L.Scucchia
Oscilloscope (4)
The ADC provides a discrete form of the signal:
• in the time domain (sampling) and
• in the amplitude domain (quantization).
The ADC converter operates continuously at the maximum sampling frequency.
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
Yellow = Channel specific blocks
Blue = System blocks
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
Dr. L.Scucchia
Oscilloscope (5)
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
Yellow = Channel specific blocks
Blue = System blocks
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
The memory is organized as a circular buffer.
The write pointer advances at the rate of the ADC. write
pointer
Dr. L.Scucchia
Oscilloscope (6)
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
Yellow = Channel specific blocks
Blue = System blocks
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
The trigger signal, through the timebase system,
causes the start of the reading cycle of the channel
memory
write
pointer read
pointer
Dr. L.Scucchia
Trigger level
0 k n-1
read
pointer
Oscilloscope (7)
The trigger event defines a reading interval.
The trigger determines when captured data are stored and displayed.
On the display the result is a
signal portion preceding and
following the trigger event
Dr. L.Scucchia
Oscilloscope (8)
Trigger level
0 k n-1
0 k n-1
0 k n-1
read
pointers
Trigger problems (different trigger events)
Two or more trace on the display.
A trace in movement on the display
Dr. L.Scucchia
Oscilloscope (10)
Trigger
Coupling
Attenuator Amplifier ADC Channel
Memory
Display
DSP
Trigger
Logic
Timebase
System
CPU
System
Input
BNC
Yellow = Channel specific blocks
Blue = System blocks
DSO Block Diagram
Trigger
Comp
DC
Trigger
Level
+
-
+
DC
Offset
-
Trigger
signal
Newer DSOs use custom DSPs to quickly process data
and then send waveform data into display memory.
Dr. L.Scucchia
Oscilloscope (11)
DSO3202A (Agilent)
Power switch
Measure controls
Vertical controls
Horizontal Controls
MENU’ ON/OFF
Run controls
Menù controls
Trigger controls
Menu defined buttons
Inputs Compensation
terminals
Display
Entry knob
Waveform controls
Dr. L.Scucchia
Low Frequency Compensation
1. Set the Probe menu attenuation to 10X.
2. Attach the probe tip to the probe compensation
connector.
3. Press the Autoscale front panel button.
4. If the waveform is not rectangular you must
compensate.
5. Through a nonmetallic tool adjust the screw
located near the probe up to get a signal perfectly
compensated.
High Frequency Compensation
1. Using the BNC adapter, connect the probe to a
square wave generator.
2. Set the square wave generator to a frequency of 1
MHz and an amplitudeof 1 Vp-p.
3. Press the Autoscale front panel button.
4. If the waveform does not appear like the Correctly
Compensated waveform shown in figure, then
adjust the 2 high frequency compensation
adjustments on the probe for the flattest square
wave possible.
Oscilloscope (12)
DSO3202A (Agilent)
Dr. L.Scucchia
Oscilloscope (13)
DSO3202A (Agilent)
Display
Dr. L.Scucchia
Oscilloscope (14)
DSO3202A (Agilent)
Vertical controls
Each channel has a vertical control menu activated
by pressing the 1 or 2 keys on the front panel. The
characteristics of the track are modified using the
knobs Scale and Location.
The menu is activated by pressing MENU„ ON/OFF.
• Channel Coupling Control
• Bandwidth Limit Control
• Probe Attenuation Control
• Invert Control
• Digital Filter Controls
Mathematical operations are activated by pressing
Math.
The waveforms may be stored using the menu Ref
Dr. L.Scucchia
Connection control - pushing the button to the right of Coupling you can choose
between three possible connections:
AC for a connection via capacitor
(display the AC component)
DC for a direct connection
(display full signal)
GND for the check of the reference level
(ground voltage)
Bandwidth Limit Control - allows to
remove, from the waveform examined,
the high frequency components not
relevant to the measurement made. Press
the key corresponding to BW Limit
function is activated (ON) and all
components at frequencies greater than
20 MHz are rejected.
Oscilloscope (15)
DSO3202A (Agilent)
Dr. L.Scucchia
Control probe attenuation - Allows you to specify the
attenuation on the probe selected. The probe attenuation
control changes the attenuation factor for the probe.
The attenuation factor changes the vertical scaling of the
oscilloscope so that the measurement results reflect the
actual voltage levels at the probe tip.
Invert control - The invert control inverts the displayed waveform with respect to the
ground level.
Digital filter controls - Pressing the Digital Filter menu key displays the Filter Controls.
The filter controls set the digital filter
used to filter the sampled waveform
data. The types of filters that are
available are shown in the table.
Oscilloscope (16)
DSO3202A (Agilent)
Dr. L.Scucchia
Oscilloscope (17)
DSO3202A (Agilent)
Horizontal controls
The oscilloscope shows the time per division
in the status bar, only one value for all
waveforms.
The controls Scale and Position allow to
change scale and position of the waveform.
The menu associated with the time base is
activated by pressing the Main/Delayed.
Delayed ON/OFF
Time Base Y-T/X-Y
Trig–Offset Reposition the track
Holdoff Change the instant trigger
Holdoff Reset
Dr. L.Scucchia
With Delayed command you can enable or
disable delayed sweep mode. In this mode,
the screen is divided into two parts, the upper
part shows the original waveform while the
lower part shows an enlarged view. The
enlarged portion is called the delayed sweep
window. and the monopole of scale and
horizontal position control the relevant
parameters of the waveform.
The key for Time Base allows you to choose the
format TY and XY. In XY format, the vertical
displacement of the trace on the screen depends
on the signal on channel 1 while the horizontal
deviation depends on the signal on channel 2. In
XY format many as the functions are not
supported.
Oscilloscope (18)
DSO3202A (Agilent)
Dr. L.Scucchia
The holdoff time is the period of waiting
before starting a new trigger to change the
duration of this interval is sufficient to press the
key next to Holdoff and adjust the time using
the knob added. Pressing the Reset holdoff the
holdoff time is set to its minimum value of
100ns.
The Trig-Offset button allows you to relocate the center of the track displayed
on the screen.
Oscilloscope (19)
DSO3202A (Agilent)
Dr. L.Scucchia
Oscilloscope (20)
DSO3202A (Agilent)
Trigger controls
Level: it changes the voltage level of the trigger signal.
Force: it forces the acquisition.
50%: voltage level of the trigger signal is set equal to
the average value.
Ext Trig: external trigger.
Mode/Coupling: Trigger menu is activated.
• Mode Edge, Pulse e Video (metods)
• Source CH1, CH2, ext, …
• Slope
• Sweep Auto, Normal
• Coupling AC, DC, LFrejet e HFrejet
Dr. L.Scucchia
Oscilloscopio (21)
3202A (Agilent)
Waveform
Acquire
Controlls
Display
Controls
Type: Vectors, Dots
Grid:
Persist: Infinite, OFF
Clear:
Menù Display: 1s,2s,5s…
Screen: Normal,Invert
Mode: Normal, Average, Peak_detect
Sampling methods:Real time, Equ time
Averages: 2-256
Sequence:
The Average mode allows to remove random noise from the waveforms thus
improving the accuracy of the measurement. To avoid aliasing is useful acquiring
type Peak Detect. Real time is recommended to observe non-periodic signals,
whereas Equ-time allows to visualize high frequency signals better. The
command Sequence allows to record the waveform input from the channel 1 or 2,
with a depth of acquisition of 1000 frames.
Pressing the Display button the corresponding menu is shown where you can
specify the display characteristics of the track and the grid.
Dr. L.Scucchia
Real-time sampling
In the real-time sampling mode, single waveforms are
sampled at same intervals.
The digitizer works at maximum speed to acquire as
many points as possible in one sweep, and the 3000
Series oscilloscopes provide sampling rates up to 1
GSa/s.
In real-time sampling the trigger event happens on a
particular feature of the waveform (amplitude). In this
type of data acquisition, the sample rate of the ADC
determines the sample spacing and the number of points
that will be displayed.
Oscilloscopio (22)
DSO3202A (Agilent)
Equivalent time Sampling
The input signal is only sampled once per
trigger.
At the next triggered, a small delay is added
and another sample is taken.
The number of samples determines the
necessary number of cycles to reproduce the
waveform.
Dr. L.Scucchia
Sourcee: CH1, CH2
Voltage: Vpp,Vmax, Vmin, Vavg,...
Time Freq, Period,
Clear
Display All: OFF, ON Measure
Controls Manual: Menù
Track: Menù
Auto Measure: Menù
Cursor
Meas
Oscilloscope (23)
DSO3202A (Agilent)
Meas button on the front panel activates the automatic measurement system, in
particular this enables oscilloscope to perform 20 different measures including:
Vpp, Vmax, Vmin, Vamp, Vavg, Vrms, Freq, Period, risetime and Fall Time.
The Cursor button on the front panel activates the menu corresponding to the
measures concerning the marker. There are three modes: Manual, Track and Auto
Measure.
• Manual, the screen shows two parallel cursors that can be moved on the track in
order to obtain the measures of voltage or of time desired. The
values corresponding to the cursors are shown in the upper part of the screen.
• Track, two sliders are automatically activated that can be adjusted using the
added knob.
• Auto Measure, available when the measurement system is automatically
activated, the oscilloscope displays the cursor in relation to the latest measures
used.
Dr. L.Scucchia
Voltage measurements:
• Vpp (Peak-to-Peak Voltage)
• Vmax (Maximum Voltage)
• Vmin (Minimum Voltage)
• Vavg (Average Voltage)
• Vamp (Amplitude Voltage = Vtop - Vbase)
• Vtop (Top Voltage)
• Vbase (Base Voltage)
• Vrms (True Root-Mean-Square Voltage)
• Overshoot (Measure the overshoot voltage in percent (Vmax-Vtop)/Vamp. Overshoot is a waveform
distortion which follows a major edge transition)
• Preshoot (Measure the preshoot voltage in percent (Vmin-Vbase)/Vamp,. Preshoot is a waveform
distortion which precedes a major edge transition)
Oscilloscope (24)
DSO3202A (Agilent)
Dr. L.Scucchia
Time Measurements:
• Frequency
• Period
• Rise Time
• Fall Time
• +Width
• -Width
• +Duty
• -Duty
• Delay 1→2
• Delay 1→2
Oscilloscopio (25)
3202A (Agilent)
Dr. L.Scucchia
t1 t2
Channel 2
50%
50%
Channel 1
Oscilloscope (26)
DSO3202A (Agilent)
Delay 1-2, rising edges
Channel 2
50%
50%
Channel 1
Delay 1-2, falling edges
Delay From channel 1 to channel 2
rising edges ( = t2-t1)
Delay From channel 1 to channel 2
Falling edges ( = t2-t1)
t1 t2
Dr. L.Scucchia
Oscilloscope(27)
DSO3202A (Agilent)
Utility
Controlls
Mask Test: Menù
I/OSetup Menù
Language: English,…..
Sound: ,
System Info
Self-Cal
Self-Test
Save and
Recall
Controlls
Save/Recall: Waveforms,
Setups
Default Setup
Waveform: 1-10
Setup: 1-10
Load
Save
Controls Autoscale and Run/Stop
The Autoscale key is used to retrieve and set automatically the oscilloscope controls
necessary for a good display of the waveform at the input.
The Run/Stop key on the front panel of the oscilloscope starts or stops the acquisition
system of the oscilloscope. When the acquisition is stopped, the button is red and the
vertical and horizontal scales can be changed within a fixed interval
Dr. L.Scucchia
V p
-p
Period
Horizontal = 1 µs/div
V m
ax
Ground level (0.0 V)
indicator
Vertical = 1 V/div
Period (T) = 4 divisions x 1 µs/div = 4 µs, Freq = 1/T = 250 kHz.
Vp-p = 6 divisions x 1 V/div = 6 V p-p
Vmax = +4 divisions x 1 V/div = +4 V
Vmin = Vmax-Vp-p = -2 V
Making Measurements by visual estimation
Oscilloscope(28)
DSO3202A (Agilent)
Dr. L.Scucchia
Oscilloscope (29)
DSO3202A (Agilent)
The math functions control allows the selection of the math functions add,
subtract, multiply, and FFT (Fast Fourier Transform) for CH1 and CH2. The
mathematical result can be measured visually and also using the cursor controls.
The FFT of a waveform that has a DC component or offset can cause incorrect FFT
waveform magnitude values. To minimize the DC component, choose AC Coupling on the
source waveform.
To reduce random noise and aliasing components in repetitive or single-shot waveforms, set
the oscilloscope acquisition mode to averaging.
To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms
scale displays component magnitudes using a log scale.
The FFT math function mathematically converts a
time-domain waveform into its frequency
components.
Dr. L.Scucchia
Selecting an FFT Window
FFT based measurements are subject to errors from an effect known as leakage.
This effect occurs when the FFT is computed from of a block of data which is not
periodic. To correct this problem appropriate windowing functions must be applied.
The window is shaped so that it is exactly zero at the beginning and end of the data
block and has some special shape in between.
For the 3202A there are 4 FFT windows. Each window has trade-offs between
frequency resolution and amplitude accuracy. Your source waveform characteristics
along with your measurement priorities help determine which window to use.
Oscilloscope (30)
DSO3202A (Agilent)