Page 1: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Edward F. Jackson, PhD

Department of Imaging Physics

Challenges in Quantification of DCE-MRI

ACRIN DCE-MRI Workshop – October 2, 2009

Page 2: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Outline

• General MR quantification challenges

– Lack of standards • Varying measurement results between vendors

– Highly variable quality control procedures• Varying measurement results between centers

• DCE-MRI quantification challenges

– Data acquisition

– Data processing

– Quality control

– Vendor specific issues

– Multicenter trial issues

Page 3: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

General Challenges in MR Quantification

Arbitrary (and spatially- / temporally-dependent) signal intensity units

– Magnitude and homogeneity of Bo

– Magnetic field gradient nonlinearity and/or miscalibration

– RF coil dependency: RF coil type, B1 sensitivity profiles, subject positioning within the coil

– Slice profile variations (with RF pulse shape, flip angle, etc.)

– System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

Page 4: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

Bo Homogeneity

•In general: Larger B0 => higher signal-to-noise

•B0 inhomogeneity yields spatially variant signal intensities in general and spatially variant fat suppression when chemically selective saturation methods are utilized.

Page 5: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Arbitrary (and spatially- / temporally-dependent) signal intensity units

– Magnitude and homogeneity of Bo

– Magnetic field gradient nonlinearity and/or miscalibration

– RF coil dependency: RF coil type, B1 sensitivity profiles, subject positioning within the coil

– Slice profile variations (with RF pulse shape, flip angle, etc.)

– System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

General Challenges in MR Quantification

Page 6: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Gradient Field Nonlinearity Effects

Reference: Sumanaweera TS et al., Neurosurgery 35(4):696-703, 1994.

In-Plane Through-Plane Through-Plane

Slice at isocenter

Page 7: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Isocenter

Errormax with correction: < 1 mm @ 10 cm

Errormax w/o correction: ~ 4.5mm @ 10 cm

20 cm off isocenter

Errormax with correction: < 2 mm @ 10 cm

Errormax w/o correction: ~ 5.5cm @ 10 cm

20 cm FOV, white: w/correction, black: w/o correction

Gradient Field Nonlinearity Effects

Page 8: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Primary Limits on Spatial Accuracy

• System Limitations– Poor Bo homogeneity

– Linear scale factor errors in the gradient fields

– Field distortion due to induced eddy currents

– Nonlinearities of the gradient fields

• Object-Induced– Chemical shift effects (fat / water displacement, in-plane and slice)

– Intravoxel magnetic susceptibility differences (particularly air-tissue)

– Effects are minimized with non-vendor specific appropriate acquisition parameters (increased BW, smaller FOV), but at the expense of SNR. [Importance of Technique!]

Page 9: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Arbitrary (and spatially- / temporally dependent) signal intensity units

– Magnitude and homogeneity of Bo

– Magnetic field gradient nonlinearity and/or miscalibration

– RF coil dependency: RF coil type, B1 coil sensitivity profiles, subject positioning within the coil

– Slice profile variations (with RF pulse shape, flip angle, etc.)

– System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

General Challenges in MR Quantification

Page 10: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

B1 Coil Response Non-Uniformity

B1 response non-uniformity

1.5T

3.0T

Page 11: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Arbitrary (and spatially- / temporally-dependent) signal intensity units

– Magnitude and homogeneity of Bo

– Magnetic field gradient nonlinearity and/or miscalibration

– RF coil dependency: RF coil type, B1 sensitivity profiles, subject positioning within the coil

– Slice profile variations (with RF pulse shape, flip angle, etc.)

– System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

General Challenges in MR Quantification

Page 12: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Slice profile variations (with RF pulse shape, flip angle, etc.)

5 mm SE 5 mm fast GRE

5.78 mm5.07 mm

Typically, faster imaging sequences use increasingly truncated RF pulses resulting in thicker slice profiles for a given prescribed slice thickness. This gives rise to increased partial volume averaging effects and a concomitant loss of spatial resolution.

Flip angle calibrations can also be negatively affected.

General Challenges in MR Quantification

Page 13: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Arbitrary (and spatially- / temporally-dependent) signal intensity units

– Magnitude and homogeneity of Bo

– Magnetic field gradient nonlinearity and/or miscalibration

– RF coil dependency: RF coil type, B1 sensitivity profiles, subject positioning within the coil

– Slice profile variations (with RF pulse shape, flip angle, etc.)

– System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

General Challenges in MR Quantification

Page 14: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

System stability issues (RF & gradient subsystems, Bo, RF coils, etc.)

For quantitative imaging, particularly in longitudinal studies, a rigorous quality control program is critical.

Key components of frequent QC tests:• Geometric accuracy

• Slice thickness

• Signal-to-noise ratio (or low contrast object detectability)

• Uniformity

• High contrast spatial resolution

• Center frequency

• Transmit gain

• Contrast response

General Challenges in MR Quantification

Page 15: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

ADNI

• Multicenter, multivendor study

• Optimized pulse sequence / acquisition parameters for each platform

• MagPhan/ADNI phantom scan at each measurement point

• Access to vendor gradient correction parameters

• With full correction for gradient nonlinearities and optimized acquisition strategies, spatial accuracies of ~0.3 mm can be obtained over a ~180 mm spherical volume

Page 16: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Quantification Challenges

Page 17: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Outline

• General MR quantification challenges

– Lack of standards • Varying measurement results between vendors

– Highly variable quality control procedures• Varying measurement results between centers

• DCE-MRI quantification challenges

– Data acquisition

– Data processing

– Quality control

– Vendor specific issues

– Multicenter trial issues

Page 18: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Dynamic Contrast Enhanced-MRI

3.0 T 3D FSPGR

20 3.5-mm sections every 4.5 s

0.86-mm in-plane resolution

Ktrans Parametric Map

Page 19: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

Two Compartment Pharmacokinetic Model

CP = [Gd] in plasma (mM) = Cb / (1-Hct)

CEES = [Gd] in extravascular, extracellular space (mM)Ktrans = endothelial transfer constant (min-1)

kep = reflux rate (min-1)

vP = fractional plasma volume, ve = fractional EES volume (= Ktrans / kep)

Standardized parameters as proposed by Tofts et al., J Magn Reson Imaging, 10:223-232, 1999.

Ktrans

Plasma Flow

PlasmaCP, vP

EESCEES, ve

Endothelium

kep

( ')

0( ) ( ') 'ep

t k t ttransEES PC t K C t e dt CL(t) = vP CP(t) + CEES(t)

0 2 4 6 8 100

1

2

3

Time (min)

[GdD

TP

A]

(mM

)

CP(t)

CL(t)

MDACC MR Research

Measured Measured

Page 20: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Signal intensity data from tumor

and vascular ROIs

'

0'

1

ept k t ttrans

P P P

L

K C t e v C tC t

Hct

0,1

0

01

0,10,1

0,10,1

1cos1

1coscos1

ln1

Re

SS

e

eSS

e

TRR

RTRRTR

RTRRTR

1

1

1

0,11][r

R

r

RRGd

Determine vP, Ktrans, and kep, from non-linear Marquardt-Levenberg fitting of CL(t) and CP(t) data

DCE-MRI

Page 21: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

[GdDTPA] (mM)

(S -

S0)

/ S

0 (%

)

Spin-Echo: TE/TR = 10/400 ms

FSPGR: TR/TR = 2/10 ms, =250

r1/r2 = 4.5 / 5.5 mM-1 s-1

T1,0 =1250 ms ; T2,0 = 125 ms

FSPGR

SE

DCE-MRI – Pulse Sequences

Several different pulse sequences and variations can be used– Most common: fast spoiled gradient echo, with or without magnetization

preparation pulses

Page 22: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Data Acquisition Challenges

• Pulse sequence

– Contrast response must be well characterized

• Temporal resolution

– Must match choice of pharmacokinetic model and parameters of interest• Must be rapid (~4-6 s) for generalized kinetic model with estimation of vp

• Recommended to be ≤15 s for any pharmacokinetic model

• Spatial resolution

– Must be adequate for target lesion size and application

• Anatomic coverage

– Should fully cover target lesion(s) and include an appropriate vascular structure

Page 23: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Data Acquisition Challenges

• Effects of motion

– Liver / lung / breast• Respiratory motion

• Cardiac motion

– Head & neck• Swallowing

• Tongue / jaw movement

– Brain and other “fixed” anatomy• Involuntary motion, e.g., when contrast agent is injected

• Effects should be prospectively mitigated, if possible, or retrospectively addressed (rigid body or deformable registration)

Page 24: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Data Acquisition Challenges

• Vascular input function (VIF)

– Needed to account for variations in cardiac output and contrast agent delivery parameters (i.v. access location, gauge of needle, position of arm, injection rate for contrast agent and flush, etc.)

– Requires good temporal resolution

– Acquisition strategy should:

• Minimize motion artifacts due to pulsatile flow

• Minimize the impact of flow related enhancement (in axial acquisition transverse to major vessel) on VIF, e.g., use in-plane reference vessel, spatial saturation pulse (at a cost of temporal resolution), etc.

Page 25: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Data Acquisition Challenges

• T1 measurement

– Required if contrast agent concentration, as opposed to signal intensity change, is used in the pharmacokinetic modeling

– Must be accomplished in a reasonable acquisition time

– Options widely available on scanners:• Inversion recovery --- accurate, but too slow

• Saturation recovery --- also slow

• Variable flip angle fast gradient echo techniques --- can be acquired in acceptable scan times, but can yield widely varying T1 estimates depending on specific pulse sequence and acquisition parameters

– T1 measurement “noise” is propagated into estimates of Ktrans and other pharmacokinetic model parameters.

Page 26: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

VFA T1 MeasurementsMDACC VFA MDACC VFA

Region 20090907* Flip + 5** Average Diff T1

1 738.22 1124.59 931.41 386.37

2 422.02 563.15 492.59 141.13

3 556.35 770.53 663.44 214.18

4 1356.10 1406.91 1381.51 50.81

5 468.94 639.36 554.15 170.42

6 704.67 1052.80 878.74 348.13

7 955.35 1234.30 1094.83 278.95

8 405.00 513.88 459.44 108.88

9 490.49 659.58 575.04 169.09781.24 110.936 <=stdev

27.8 217.43 <=1.96*stdev207.55 <=mean difference

*Flip angles: 2, 5, 10, 15, 20, 25, 30 - per QIBA protocol

**Flip angles: 7, 10, 15, 20, 25, 30, 35

3D FSPGR TE=1ms, TR=5.1ms

Page 27: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Ktrans Simulations – T1 Dependence

Ktrans vs Calculated Lesion T1

0.00

0.04

0.08

0.12

0.16

0.20

1000 1100 1200 1300 1400

T1 (ms)

Ktr

an

s (

min

-1)

Page 28: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

DCE-MRI Data Acquisition Challenges• Quality Control

– To reduce system-level variance in longitudinal measurements, a robust quality control program is needed

– The ACR MR Accreditation Program is the most widely implemented MR QC program, but does not address all issues important in DCE-MRI, e.g., contrast response and signal stability.

• Use of multiple “matched” scanners

• “Upgrade Dilemma”

– If a system is upgraded during an ongoing longitudinal study, a thorough evaluation of the impact(s) of such an upgrade on DCE-MRI sequences and acquisition parameters (and contrast response, signal stability, etc.) must be performed.

Page 29: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

From Single- to Multi-Vendor Studies

• Major challenges:

– Protocol matching

• Pulse sequence and acquisition parameter selection for matched:– contrast response– temporal resolution– spatial resolution– anatomic coverage

• Can be achieved, but requires effort at start up and, subsequently, constant monitoring for changes in hardware/software (need for ongoing quality control)

– Vendors focused on “competitive advantage” in diagnostic radiology, not on quantitative imaging applications

Page 30: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

From Single- to Multi-Center Studies

• Major challenges:

– Matched protocols

– Distribution and activation of protocols• Distribute/load electronically (ADNI)

• Provide expert training and initial protocol load/test

• Rely on local expertise

– Compliance with protocol• Local radiologists, technologists

– Widely varying quality control• Ranging from specific for DCE-MRI, to ACR MRAP, to none

– “Scanner upgrade dilemma”

– Data management / submission

Page 31: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

• Data analysis implementation strategies are as variable as acquisition strategies

• Currently, DCE-MRI analysis tools based upon pharmacokinetic modeling are developed by the users as no such tools are currently commercially available from the MR vendors

• Choice of model must match data acquisition strategy, e.g., temporal resolution of the acquired data

• To facilitate testing/validation of various analysis packages, readily available, standardized test / retest data and analysis results (Ktrans, kep, ve, vp) are needed.

DCE-MRI Data Analysis Challenges

Page 32: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

• Many choices to be made:

– Vascular input selection

• Manual ROI, automated identification vascular structure pixels, etc.

• Reproducibility

– Lesion region of interest(s)

• Definition criteria

• Reproducibility

– Fits of single averaged pixel uptake curve or pixel-by-pixel fits

– Modeling of: [Gd](t) (requiring T1 mapping) or SI(t) data

DCE-MRI Data Analysis Challenges

Page 33: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Quantitative Imaging Biomarkers – Who’s Interested?

NIST USMS Workshop 2006

Representative Agencies / Organizations

Page 34: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

General As Yet Unmet Needs

To move MR imaging biomarkers from exploratory / secondary endpoints to primary endpoints / surrogate markers:

– There exists a need for standardized acquisition pulse sequences and analysis techniques for MR “imaging biomarker” studies.

– Validated phantoms and test data need to be available to users in order to test new releases of pulse sequences and analysis software.

– Specific phantoms should be available to validate each vendor’s acquisition techniques for the particular MR biomarker technique (lesion morphology, perfusion, diffusion, MR spectroscopy, etc.).

Page 35: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

General As Yet Unmet Needs

To move MR imaging biomarkers from exploratory / secondary endpoints to primary endpoints / surrogate markers:

– Imaging biomarker to tissue-based and outcome measure comparisons are needed for validation.

– Reproducibility (test/retest) studies are lacking in several key areas.

Page 36: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Multi-center trial design = Variations in imaging

Hardware & Software algorithms

• Standardize imaging techniques across multi-centers • Cross-consistency - within & across imaging systems

• Don’t assume – assess!– Discover & adjust for differences

George Mills, MD, MBA (former Director FDA/CDER Division of Medical Imaging & Radiopharmaceutical Drug Products)

Need for Standardized Imaging Techniques

Page 37: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

• “Precision is the goal”

• Implement the same, detailed imaging acquisition protocols at all clinical sites

• Optimize image processing & reconstruction software– Avoid manual techniques as possible– Select & develop semi-automated => automated– Reduce human interactions

• “Established” vs. “cutting edge”

Need for Standardized Imaging Techniques

George Mills, MD, MBA (former Director FDA/CDER Division of Medical Imaging & Radiopharmaceutical Drug Products)

Page 38: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Quantitative MR Efforts

• NCI: RIDER / SAIC Contracts

• IRAT: MR Committee / Phantom Development

• RSNA:Quantitative Imaging Biomarker Alliance DCE-MRI Cmte

• ISMRM: Ad Hoc Committee for Standards for Quantitative MR

• AAPM: Working Group for Standards for Quantitative MR Measures

Page 39: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

NCI RIDER

• NCI Cancer Imaging Program RIDER – Reference Image Database to Evaluate Response*

Collaborative project for development and implementation of a caBIG public resource

Series of manuscripts in Translational Oncology in Dec 2009 and

data made publically available through NCIA (phantom and anonymized clinical trial imaging and meta data)

Page 40: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

NCI RIDER Reports

• Clarke L, Croft B, Nordstrom R, Zhang H, Kelloff G. Editorial: Quantitative imaging for evaluation of response to cancer therapy

• Meyer CR, Armato III SG, Fenimore CP, McLennan G, Bidaut LM, Barboriak DP, Gavrielides MA, Jackson EF, McNitt-Gray MF, Kinahan PE, Petrick N, Zhao B. Quantitative imaging to assess tumor response to therapy: common themes of measurement, truth data and error sources

• McNitt-Gray MF, Bidaut LM, Armato III SG, Meyer CR, Gavrielides MA, McLennan G, Petrick N, Zhao B, Reeves AP, Beichel R, Kim H-JG, Kinnard L. CT assessment of response to therapy: tumor volume change measurement, truth data and error

• Kinahan PE, Doot RK, Wanner-Roybal M, Bidaut LM, Armato III SG, Meyer CR, McLennan G. PET/CT assessment of response to therapy: tumor change measurement, truth data and error

• Jackson EF, Barboriak DP, Bidaut LM, Meyer CR.  MR assessment of response to therapy: tumor change measurement, truth data and error sources

Page 41: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

NCI RIDER DCE-MRI Phantom Data

Gel-filled compartments with varying T1 relaxation times

Eurospin TO5 – DiagnosticSonar, Ltd.

Page 42: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

RIDER T1 Measurements – Vendor 1

AMR7 Run 1 MultiFlip vs IR

y = 1.0907x - 15.548

R2 = 0.9981

0

500

1000

1500

2000

0 500 1000 1500 2000

IR T1 (ms)

Ave

Mu

ltif

lip

T1

(ms)

Week 0 Run 1 vs Run 2 (AMR7)

y = 0.9935x + 2.1855

R2 = 0.9999

0

500

1000

1500

2000

0 500 1000 1500 2000

Run 1 T1 (ms)

Ru

n 2

T1

(ms)

Week 0 vs Week 1 (AMR7)

y = 0.9838x + 2.5057

R2 = 0.9999

0

500

1000

1500

2000

0 500 1000 1500 2000

Week 0 T1 (ms)

Wee

k 1

T1

(ms)

Run 1 = baseline Run 2 = 2 hrs post baseline Week 1 = 1 week post baseline

Bosca & Jackson, AAPM 2009; Jackson et al., Trans Oncol, Dec 2009

Page 43: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

RIDER T1 Measurements – Vendor 2

Run 1 = baseline Run 2 = 2 hrs post baseline Week 1 = 1 week post baseline

Bosca & Jackson, AAPM 2009; Jackson et al., Trans Oncol, Dec 2009

Page 44: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Imaging Response Assessment Teams

• IRAT: MR Committee / Phantom Development

– Current status:

• MR phantom based on the commercially available MagPhan/ADNI phantom (The Phantom Laboratory).

• Phantom distributed to IRAT MR sites to obtain multicenter, single vendor data.

• Similar to, but pre-dates, RSNA QIBA DCE-MRI project acquisition and data analysis strategy

Page 45: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

RSNA QI Biomarker Alliance

• RSNA QIBA: DCE-MRI Technical Committee

– Multiple subcommittees:

• Phantom development/selection

• Scan protocol / data analysis

• Synthetic DCE-MRI test data

– MR phantom based on the IRAT DCE-MRI phantom

– Phantoms to be distributed to multiple MR sites to obtain multicenter (N=6), multivendor (N=3) data

– Data analysis provided by core lab

http://qibawiki.rsna.org/index.php?title=DCE-MRI

Page 46: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Phantom Scans• Phantom and “cuboid” scanned using torso phased array coil to mimic

typical Phase I clinical trial scans (liver mets / lung mets).

RSNA QI Biomarker Alliance

Page 47: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

• RSNA QIBA: DCE-MRI Technical Committee

– Phantom measurements:

• Phased array acquisition

• Body coil acquisition

• SNR acquisition

• VFA T1 measurement acquisition

• DCE acquisition

– Each of the above acquisitions repeated with phantom rotated by 90, 180, 270, and 360o

– All acquisitions repeated one week later

Ratio map correction for B1

sensitivity characteristics

RSNA QI Biomarker Alliance

Page 48: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

RSNA QIBA DCE-MRI

Difference in T1 from each contrast sphere, week 1 minus week 0.

Difference in R1 from each contrast sphere, week 1 minus week 0.

Page 49: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

RSNA QIBA DCE-MRI

Comparison of R1 vs Signal Intensity at Rotation “A”

Page 50: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

ISMRM Ad Hoc Committee

• ISMRM: Ad Hoc Committee for Standards for Quantitative MR

– Membership includes MR physicists, technologists, radiologists, NIST representative, NCI/CIP representatives, vendors, pharma. Expertise in research trials using quantitative MR.

– Current status:

• White paper on quantitative MR

• Design specifications & construction of a MR system phantom (collaboration with and funding by NIST)

• Initial multicenter/multivendor phantom pilot studies to begin in late 2009 / early 2010.

http://wiki.ismrm.org/twiki/bin/view/QuantitativeMR/

Page 51: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

ISMRM SQMR System PhantomSpatial accuracy

Contrast response

Section thickness

High contrast resolution

0.7 0.8, 0.9, 1.0, 1.1 mm

All materials fully characterized by NIST.

Page 52: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

ISMRM SQMR System PhantomT1 Compartments T2 Compartments

PD Compartments

Page 53: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

AAPM Working Group

• AAPM: Working Group for Standards for Quantitative MR Measures– Charge

• Development of quantitative MR phantoms and acquisition techniques / parameters specific to a given quantitative measure. Obtaining repeat measures on each phantom at multiple centers and using current high-field scanners from all major vendors.

• Develop signal response characteristic standards for relevant pulse sequences. These response characteristic standards will allow ongoing QC at individual centers on a given platform (across upgrades) and comparisons between vendor platforms (and hardware/software versions.

• Develop sets of annotated test data that can be publicly distributed (via the NCIA, for example) and used to validate image analysis software tools.

Page 54: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

RSNA QI Initiatives

RSNA

TQI

(Toward Quantitative Imaging)

Education / Clinical Practice

QIBA

(Quantitative Imaging Biomarker

Alliance)

Biomarker precision / instrumentation

Imaging Biomarker Roundtable

Communication amongst various

societies & groups

CTSA Imaging Working Group

Translational research

infrastructure

Based on information from Dan Sullivan, MD

Page 55: MDACC MR Research Edward F. Jackson, PhD Department of Imaging Physics Challenges in Quantification of DCE-MRI ACRIN DCE-MRI Workshop – October 2, 2009.

MDACC MR Research

Quantitative Imaging Initiatives

Uniform Protocols for Imaging in Clinical Trials

(UPICT - CTSA)

Imaging Biomarker Quality Control / Phantom

Development Groups

(NIST, FDA, Inter-Society and Inter-Agency WGs)

NCI caBIG Imaging Workspace

(NCIA, LIDC, RIDER)

Imaging Response Assessment Teams / Quantitative Imaging

Network (NCI)

NCI / FDA / NIST

Scientific Societies

Imaging Scientists / Animal Imaging Cores

Imaging Equipment Vendors, Pharma, & CROs