Close-out location for license# 21-11553-01, Mclaren Lapeer Region, Lake Orion Cardiology site
1. A close-out survey was performed on 07/25/13 of our Lake Orion Cardiology, 720 N. Lapeer Rd, Lake Orion, Ml 48362 address and is enclosed for your review. We have also enclosed a map of all areas wipe tested.
2. We did not dispose of radioactive materials by release to sewers or incineration. In addition, there were no spills or leaking sources of any long-lived radioactive material or on-site burials of radioactive materials. A leak test is enclosed for your viewing.
3. Any further correspondence may be sent to Mclaren Lapeer Region, 1375 N. Main St., Lapeer, Ml 48446.
4. We have enclosed confirmation of the sealed sources received at Mclaren Lapeer Region, 1375 N. Main St., Lapeer, Ml 48446 from the Lake Orion closed-out site.
Cordially,
Barton Buxton, CEO Mclaren Lapeer Region
Lake Orion Cardiology 2
Close-out July 2013
RECEIVED AUG 0 2 2013
~JMcLaren LAPEER REGION
1375 North Main St. Lapeer, Michigan 48446-1350
tel (810) 667-5500
mclaren.org
Nuclide Type CalibrationActivity Calibration Date Model Number
Co-57 Flood 10 mCi 4/1/2012
Ba-133 Vial 265 uCi 3/2/2012
Cs-137 Vial 198.5 uCi 5/1/2006
These sources were received by Mclaren Lapeer Region on 07/25/13
from Lake Orion Cardiology, 720 N. Lapeer Rd., Lake Orion, Ml 48362
IPL
BM06E
IPL
SeriaiNumber
1584-181
BM06133E 1205
1145-61-19
Close-out Survey
Date Performed: 07/25/13
Performed By: Laura Luna, Physics Consultant
~?MPC / ~ " MEDICAL. PHYSICS "-'"..-' '"..~ c:.ONSUI. l"AN"T S. lNG
Wipe tests analyzed with a CRC-15W well counter. Efficiency for Co-57 is 83%. MDA is 21 dpm.
\
Survey performed with Atomic Product survey meter modei14C with end window probe, S/N 85232, calibrated on 12/19/12 with a check source reading of 0.9 mR/hr. Range used was 0.1 mR/hr.
All sealed radioactive sources were transported to Mclaren Lapeer Region, 1375 N. Main St., Lapeer, Ml 48446. There is no history of any leaking sources. Documentation of received sources and a current leak test is enclosed.
History of Radionuclides Used
Tc-99m was last used 07/22/13. No other isotopes, except sealed sources: Cs-137, Ba-133 and Co-57, were used at this facility.
Visual Check
The area was checked to ensure that all radioactive waste had been removed. No radioactive material was located.
Radiation Level Survey
No area within the department demonstrated radiation levels in excess of the background reading of 0.02 mR/hr.
Lake Orion Cardiology Close-out July 2013 3
Removable Contamination
1 minute counts of a wipe sample covering at least a1 00 square centimeter area
Area Results(dpm) 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 10 0 11 0 12 0 13 0 14 0 15 0 16 0 17 0 18 0 19 0
Area 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38
Results(dpm) 0 8 8 8 8 8 0 0 0 0 0 12 12 12 12 12 0 0 0
Maximum removable contamination: 12 net dpm/100 cm2
Conclusion
As of 07/25/12 all radioactive materials were removed from the site and no removable contamination was present.
Lake Orion Cardiology Close-out July 20 13 4
~1MPC t~~ MEDICAL Pi<YSICS CONSUL "T AN·r S, lNG
Lake Orion cardiology 720 N. Lapeer Rd. Lake Orion, Ml 48362
Outside
12'9" 6
18 14 10
Staff 23 11 7 19 15
Lounge Imaging Room
12'10" 20 12 8 16
22 21 13 9 17 Outside
2 5 Hallway Hot Lab 6'4"
1 3 4
8'0"
38 37 36 35 34
32 31 Echo Room 33 Stress Room 30 29
10'0'
28 27 26 25 24
12'9" Office
Lake Orion Cardiology Close-out July 2013 5
. ( \
April19, 2013
Carissa Wessel, R.T. Manager of Diagnostic Imaging Lapeer Regional Medical Center 1375 North Main Lapeer, Michigan 48446
Dear Carissa:
MEDICAL PHYSICS CONSULTANTS, INC .
Enclosed, please find the report on the evaluation at LAKE ORION CARDIOLOGY for the 2"d quarter visit of 2013. I have listed my comments and recommendations concerning licensed activities at your facility below:
1. The following documentation is enclosed for Dr. Tarr to review and sign: the dose calibrator, accuracy test, linearity test, sealed source inventory, leak test and this letter.
2. Nuclear Medicine personnel exposures for the first quarter of 2013 (through 03/19/13) were below the ALARA Levell limits of 125 mrem/quarterforthe whole body, 1250 mrem/quarter for the extremity and skin, and 375 mrem for the lens of eye.
3. During my visit, I was able to perform both an auto calibration and an energy test on the well counter. The results were within the range of expected values.
4. There were no reported radioactive spills or accidental administrations since the previous visit.
5. As a reminder, your survey meter (SN #226161) is being located after being sent to the previous owners address. MPC has supplied a loaner meter for the interim.
6. The documentation for the daily surveys, weekly wipe tests, dose calibrator constancy, waste disposal and package logs were well maintained and on file since my previous visit.
2 14 E. Huron Street Ann Arbor, Ml48104 (734) 662-9224 Fax (734) 662-3197
'\'·1-, <Y:. 70 E. 9lst Street, Suite 106
Indianapolis, IN 46240 (317) 581-1931 Fax (317) 581-!911
t-.1 '1 r;>~ / Li\ 4806 Mile Iligh Drive .:;y~
Salt Lake City, trr 84124 (80 I) 272-2952 Office & Fax
www.mpcphysics.com
N7375 Crystal Ridge Drive Beavc,· Dam, Wl53916 (920) 885-9872 Fax (920) 885-9870
(
McLaren Lake Orion Cardiology 2nd qtr. 2013 2
7. The Philips SPECT camera NEMA numbers, COR values and bar (resolution) tests are being done according to manufacturer's specifications and are acceptable.
8. Worker instruction is due for stress and clerical personnel.
It has been a pleasure to be of service to you. If you have any questions concerning this report, please feel free to contact me at 734-662-3197 or email me at [email protected] ..
Sincerely,
'f~~.L Laura M. Luna, B.S. Medical Physics Consultant
Kenneth Tarr, Radiation Safety Officer: fr...:..;/};..;" .:...__" _, -~....,t:." -----Carissa Wessel, Management: Co A." J \.A.)~ b };.~Q
"'"'"' Jenny Jenuwine, Technologist:._,"-+··tfC1/W.!....
1
~--·'"'-4--+___,.;~1~. =..~..:...:::.... ____ _ {I ::; v---
Date: 04/19/13
Location: Mclaren Lapeer Region-Lake Orion Cardiology ~7MJ:~s mCONSULTANTS.INC
NUCLEAR MEDICINE AUDIT
IDENTIFICATION
Licensee: Mclaren Lapeer Region-Lake Orion Cardiology
License Number: 21-11553-01
42
Expiration Date: 12/31/2013
11/16/2011 Latest Amend Number Amendment Date:
Authorized User(s): AS LISTED ON AMENDMENT #43
Actions Filed: REMOVING DR. KOTLAROV
Radiation Safety Officer Kenneth P. Tarr, D.O.
RSO - Brachytherapy:
Most Recent State Inspection: Most Recent Inspection- NRC:
State Registration Number: Registration Date
Accreditation Expiration Date:
Reporting Physicist: LML
DOCUMENTS AND NOTICES
NRC
10 CFR 19, 20 and 35- Current, Posted or Referenced and on file?
License+ Application and Amendments- Posted or Referenced and on file?
"Notice of Violation", Proposed Civil Penalty or "order"- Posted?
'Notice to Employees' form NRC-3 - Current and Posted?
Management agreement(s) for amend, renewals, AU's and RPP changes?
ManagemenURSO "Letter of Understanding:- Available?
Statement of RSO Authority, Duties and Responsibilities- Available?
STATE Operating Procedures- Posted or Referenced and on file?
Radioactive Materials Registration Certificate- Current and on file?
Ionizing Radiation Rules - Posted or Referenced and on file?
'Notice of Violation', Proposed Civil Penalty or 'Order' - Posted?
State 'Notice to Employees' form- Posted?
NM 2003 (C) MPC 02/11 Page 1
10CFR19.11
10CFR19.11
10CFR19.11
10CFR19.11
10 GFR 35.24(a)
1 0 CFR 35.24(b )(g)
10 CFR 35.24(e)
YES
YES
YES
YES
N/A
YES
YES
N/A
N/A
N/A
N/A
N/A
Date: 04/19/13
Location: Mclaren Lapeer Region-Lake Orion Cardiology
POSTING AND LABELING
Doors and/or Rooms - Properly Posted?
Areas - Properly Posted?
Sealed Source Area/Rm. Exemption <5 mR/hr@ 30 em?
Containers - Properly Labeled?
Syringes or Syringe Shields -Properly Labeled?
10 CFR 20.1902 YES
10CFR20.1902 YES
10CFR20.1903(c) N/A
10 CFR 20.1904(a) and 35.69 YES
10 CFR 35.69 YES
AUTHORIZED RADIONUCLIDE USE AND POSSESSION
NRC Authorized Radionuclides
10CFR35.100
1 OCFR35.200
Xenon Excluded
Generators Excluded 10CFR35.300
Thyroid CA Tx Excluded
1 OCFR35.400
10CFR35.500
1 OCFR35.600
1 OCFR35.1 000
10CFR31.11
Phosphorus-32
Strontium -90
Cesium-137
Gadolinium-153
lridium-192
Uranium-238 depleted
Possession Limits
AS NEEDED
AS NEEDED
NO
NO
UNDER 1 Cu
YES
1000 mCi
State Registered Radionuclides
Fluorine-18
Cobalt-57
Sealed
Unsealed
Gallium-67
lndium-111
lodine-123
Thallium-201
Palladium-1 03
Possession Limits
AS NEEDED
AS NEEDED
Radioactivity in possession within allowed limits?
Unit doses properly measured and documented? 10 CFR 35.63(a)(b) and 35.2063 YES
Non-unit doses properly measured and documented? 10 CFR 35.63(c) and 35.2063 N/A
Prescribed dosage list or dosage range available/posted and followed? 10 cFR 35.63(d) YES
Accidental administrations properly documented? N/A
Mo-99 and or Sr-82/Sr-85 assays properly recorded? 10 CFR 35.204 N/A
IND protocols properly managed and documented?
SECURITY OF RADIOACTIVE MATERIAL
Rooms/Areas containing Stored RAM actively secured?
Stored RAM, i.e., Sharps Containers, actively secured from removal?
Control and constant surveillance maintained over RAM NOT in storage?
NM 2003 Ccl MPC 02/11 Page2
10 CFR 20.1801 YES
10CFR20.1801 YES
10 CFR 20.1802 YES
Date: 04/19/13
Location: Mclaren Lapeer Region-Lake Orion Cardiology ~7 MJ:~~s ~CONSULTANTS, INC
RADIATION PROTECTION PROGRAM
Radiation Protection Program (RPP) Developed and Implemented? 10 cFR 20.11o1(a) YES
RPP Procedures Documented and Available for Review? 10 CFR2o.21o2 YES
RPP Reflects Current Operations? YES
Annual Review of Written RPP Documented? 10 CFR 20.1101(c) and 20.2102(aX1) YES
Annual Review Date: 12/31/13
Annual Review of Operational RPP Documented? 10 CFR 20.1101 and 1 OCFR 35.24 YES
Annual Review Date: 12/31/13
Audits of Radiation Protection Program Documented? 10 CFR 20.2102(a)(2) YES
Radiation Safety Committee established as per requirements of 10 CFR 35.24(f)?
Radiation Safety Committee membership in accordance with 10 CFR 35.24(f)?
Last Radiation Safety Committee meeting:
Procedures and Engineering Controls for ALARA?
Periodic Performance Reviews of ALARA?
Environmental Airborne Emissions TEDE <10 mrem/yr?
OPERATING PROCEDURES
Radiation Safety Regulations - Posted?
Emergency Liquid Spill Prodecures - Posted?
Emergency Xenon-133 Spill Procedures and Clearance Times/Posted?
Aerosol Precautions - Current and Followed?
Documentation of Radioactive Spills?
Radioactive material order and delivery procedures available?
Opening Packages of Radioactive Material - Current and Followed?
Radioactive Waste Disposal- Current and Followed?
10 CFR 20. 1101(b) YES
10 CFR20.1101(b) YES
10 CFR 20.1101(d)
YES
YES
10 CFR 20.1101(d) N/A
10 CFR 20.1101(d) N/A
N/A
10 CFR 20.1906(a) YES
10 CFR 20.1906(e) YES
10 CFR 20.2001 YES
PERFORMANCE AND DOCUMENTATION OF REQUIRED PROCEDURES
Survey and monitoring reasonable to evaluate levels, cone. and hazards?
Hands and clothing monitoring (mR/hr)?
Daily surveys writtn directive prep and admin areas with surveyor name?
Sealed source safety and handling instructions available and followed?
Semiannual sealed source leak test with RSO signature? 10 CFR 35.67(b)
Semiannual inventory of sealed sources with RSO signature? 10 CFR 35.67(g)
NM 2003 (CJ MPC 02/11 Page 3
01/11/13
07/18/12
01/11/13
07/18/12
10 CFR 20.1501(a)(2) YES
NUREG 1556.V9 YES
10 CFR 35.70(a) YES
10 CFR 35.67(a) YES
04/19/13 YES
10/18/12
04/19/13 YES
10/18/12
Date: 04/19/13
Location: Mclaren Lapeer Region-Lake Orion Cardiology
'"·-~ERSONNEL MONITORING
Whole body badges?
Extremity TLD's Available and Worn Appropriately?
External Occupational Exposure Levels Reviewed?
~ZMJ::Cs t)"'-.JCONSULTANTS, INC
1 0 CFR 20. 1201 YES
ALARA I ALARA II Reportable:
Fetal Exposure to Declared Pregnancies?
Annual reports provided to supervised staff?
Previous Occupational Dose - Records Available?
Internal Occupational Dose Monitoring Required or Performed?
Sum of External and Internal Dose Required?
Records of compliance with TEDE to Public (100 mrem)?
TRAINING PROGRAM
10CFR20.1208 N/A
10CFR 19 N/A
10 CFR 20.1201(f) and 2104 YES
10CFR20.1502(b) N/A
10 CFR 20.1202 N/A
10 CFR 20.2107 N/A
Supervised staff familiar with operating procedures and license conditions?
Supervised staff familiar with emergency spill procedures?
10 CFR 35.27 YES
10 CFR 35.27 YES
10 CFR 35.27 YES
10 CFR 35.27 N/ A
10 CFR 19.12 YES
10CFR19.12 YES
Supervised staff familiar with radiation protection procedures?
Supervised staff familiar with written directive procedures?
New supervised staff instructed?
Workers likely to receive >1 00 mrem/year instructed?
Technical Staff: Annual: 01/13 Hazmat: 03/12 Written Directive: N/A
Stress: DUE
Clerical: DUE
GENERAL HEALTH PHYSICS
Supervised staff practice safe handling techniques, i.e. tongs?
Supervised staff wear buttoned lab coats or other protective clothing?
Supervised staff wear gloves as appropriate?
Supervised staff use syringe shields as appropriate?
Work areas appropriate for radiation safety purposes?
Storage areas appropriate for radiation safety purposes?
Refrigerator Available?
Fume Hood Available and Operational?
Consolidated "Spill Kit" available?
Area Shielding Available and Appropriate:
Lead Bricks NO
Syringe Shields YES
Lead Vial Containers
Lead Syringe Carriers
NUREG-1556 Vol. 9
NUREG-1556 Vol. 9
NUREG-1556 Vol. 9
10CFR20.1406
10 CFR 20.1406
Lead-Glass Face Shield Leaded Waste Containers
NO
YES
YES
NM 2003 Ccl MPC02/11 Page4
YES
YES
YES
YES
YES
YES
NO
N/A
Date: 04/19/13
Location: McLaren Lapeer Region-Lake Orion Cardiology ~ 7MJ:~s tNcoNSULTANTS,oNc
RADIOACTIVE MATERIAL PACKAGE RECEIPT
Visual Inspection?
Radiation Level surveys (mR/hr) with Trigger Levels:
One (1) meter or Three (3) feet?
Radioactive Contamination Monitoring (dpm/300 cmA2) with Trigger Levels
Wipe test 300 cmA2 external package surface< 22 dpm/cmA2?
WASTE DISPOSAL
Return to supplier used?
Survey external package surface appropriate for DOT labeling?
Wipe test 300 cmA2 external package surface <22 dpm/cmA2?
Confirmation of sealed source return from supplier?
Transfer of Radioactive Material to Authorized Recipient used?
Manifest Tracking System Available?
49 CFR 173.441 YES
49 CFR 173.443 YES
YES
10 CFR 71 YES
49 CFR 173.443 YES
YES
10 CFR 20.2001(aX1) YES
10 CFR 20.2006(aX2) NO
Disposal by Decay in Storage used? 10 CFR 20.2001(aX2) and 35.92
Waste Tracking System Available?
Waste Segregated by half-life?
Radiation levels at background at final disposal? 10 CFR 35.92(aX1)
Release in Effluents used? 10 CFR 20.2001(aX3)
Effluent Cone. (uCi/ml) below Table2, Appendix B 10 CFR 20?
Documented DAC calculations for Xe-133, Tc-99m DTPA aero. and 1-131 available?
Last Ventillation Evaluation:
Xenon trap monitoring (uCi)?
Release into Sanitary Sewerage used?
Average monthly volume of water released is known?
Sewer Cone. (uCi/ml/month) below Table 3, Appendix B 10 CFR 20?
<5 Ci of H-3 and <1 Ci of C-14 and < 1 Ci all other released?
Disposal of H-3 and C-14 as "Deregulated Waste"?
Appropriate documentation to confirm deregulated status available?
10CFR20.1101(d) N/A
10 CFR 20.2003 NO 10 CFR 20.2003 N/A
N/A
10 CFR 20.2003(aX4) N/A
1 0 CFR 20.2005 N/ A
N/A
Labels Defaced or Removed Prior to Disposal? 10 CFR 20.1904(b) and 35.92(b) YES
BREAST -FEEDING
Breast feeding release criteria known to staff and/or posted?
Breast-feeding instructions provided and release records on file?
FACILITY DECOMMISSIONING
Recordkeeping requirements of Decommissioning on file?
NM 2003 (Cl MPC 02/11 Page 5
YES
10 CFR 35.75(d) N/A
10 CFR 30.35(gXi-iv) N/A
/ '
~7MJ~Cs tNcoNSULTANTS.INC
Licensee: Mclaren Lapeer Region-Lake Orion Cardiology Date: 04/19/13
Model #: CRC-15W
Serial #: 171588
Instrument ID:
Installation Date: 02/27/12
Calibration in accordance with manufacturers recommendation
Constancy checks on all commonly used settings
Linearity tests acceptable (required quarterly)?
First quarter: 01 /11/13
Third quarter: 07/18/12
Accuracy tests acceptable?
Geometrical variation test acceptable?
Second quarter:
Fourth quarter:
04/19/13
02/27/12
Dose Calibrator Instrument Evaluation 1
10 CFR 35.60(b) YES
YES
YES
04/19/13
10/18/12
YES
YES