Mass Spectrometry. How is MS Done? Basic MS Instrumentation.
-
date post
19-Dec-2015 -
Category
Documents
-
view
221 -
download
1
Transcript of Mass Spectrometry. How is MS Done? Basic MS Instrumentation.
3D Quadrupole ion trap Linear Quadrupole ion trap
3D radio-frequency ion traps: ions are trapped in center point (single point)2D Linear trap: rf field is zero along the whole central line of the electrode configuration
Quadrupolar ion traps (Paul trap)
3 sets of parallel plates in a uniform magnetic field
In a 7-Tesla magnetic field an ion with m/z =100 will spin 1,000,000 cycles (travel ~ 30 km) in a 1 sec. observation period
Fourier transform ion cyclotron resonance (FTICR) MS
Mass Analyzers
Instrument
Mass analyzer 2D trap QTOF FT-ICR/Orbitrap
Mass accuracy 500 ppm 10 ppm 1 ppm
Resolution 1,000 10,000 100,000
Sensitivity 0.1 fmol 1 fmol 0.1 fmol
Dynamic range 100 50 10,000
Price $250K $450K $600K
Resolution/Sensitivity/Mass Accuracy
MR =
ΔM
Mass
Mass Difference
5000 is typical
unit resolution up to mass 5000
Proteins: 100 AAM = 15,000R = 15,000 for unit resolution
Exact Mass/Mass Defect/Isotopic Abundance
C3H8O 60.05754C2H8N2 60.06884
C2H4O2 60.02112
CH4N2O 60.03242
R > 5300 required