Mass spectrometry

111
By Kommineni.vidyachowdhary M.pharm Vaagdevi pharmacy college warangal 1 MASS SPECTROMETRY MASS SPECTROMETRY

Transcript of Mass spectrometry

Page 1: Mass spectrometry

ByKommineni.vidyachowdhary M.pharm

Vaagdevi pharmacy collegewarangal

1

MASS MASS SPECTROMETRYSPECTROMETRY

Page 2: Mass spectrometry

Mass spectrometry is the most accurate method for determining the molecular mass of the compound and its elemental composition.

It is also called as positive ion spectra or line spectra.

04/15/232

Page 3: Mass spectrometry

Introduction• Mass spectrometry (Mass Spec or MS) uses high energy electrons to

break a molecule into fragments.

• It does not involve the absorption or emission of light.• A beam of high-energy electrons breaks the molecule apart.

• The masses of the fragments and their relative abundance reveal information about the structure of the molecule.

• Separation and analysis of the fragments provides information about:

– Molecular weight

– Structure

Page 4: Mass spectrometry

Mass Spec Principles

IonizerIonizer

Sample

+_

Mass AnalyzerMass Analyzer DetectorDetector

Page 5: Mass spectrometry

04/15/23 5

Schematic of Mass Spectrometry

Ionizer

↓Mass-to-charge ratio Analyzer

Detector

Page 6: Mass spectrometry

How does a mass spectrometer work?

• Ionization method– MALDI– Electrospray(Proteins must be charged

and dry)

• Mass analyzer– MALDI-TOF

• MW – Triple Quadrapole

• AA seq– MALDI-QqTOF

• AA seq and MW– QqTOF

• AA seq and protein modif.

Create ions Separate ions Detect ions

• Mass spectrum

• Database analysis

Page 7: Mass spectrometry

04/15/23 7

What is a Mass Spectrometer?

A Mass Spectrometer is a machine that weighs molecules ! (by measuring the mass to charge ratio of ions)

SourceEICIESIAPCIAPPIMALDI

DispersionTOFFT-ICRSectorQuadTrap

DetectorFaraday CupChanneltronMCP

Page 8: Mass spectrometry

04/15/23 8

Mass Spectrometry Categorization• Based on ionization:

Matrix-assisted laser desorption/ionization (MALDI) mass spectrometryElectrospray ionization (ESI) mass spectrometryInduction-coupled plasma mass spectrometry (ICP-MS)Atmospheric Ionization Mass SpectrometrySecondary Ionization Mass Spectrometry (SIMS)

• Based on M/Z Separation:Quadrupole Mass spectrometryIon trap mass spectrometrymagnetic sector mass spectrometrytime-of-flight mass spectrometryFourier Transform Ion Cyclotron Resonance Mass SpectrometryIon Mobility Mass Spectrometry

Page 9: Mass spectrometry

04/15/23 9

Mass Spectrometry Categorization(continued)

• Based on Applications:Environmental Mass SpectrometryBiological Mass SpectrometryCell Mass SpectrometryPortable Mass Spectrometry

• Based on Configuration:Tandem Mass Spectrometry

• Based on Sample Introduction:GCMS; LCMS, Electrophoresis Mass Spectrometry

Page 10: Mass spectrometry

04/15/23 10

MS Operation• Nearly all mass spectrometers need to operate under

high vacuum condition with the pressure less than 10 -5 Torr with the only exceptions of an ion trap mass spectrometer (milli-Torr) and an ion mobility mass spectrometer (Torr).

• Never turn on a mass spectrometer without knowing the chamber pressure.

• A tour to major mass spectrometry facilities in Genomic Research Center, Sinica will be arranged.

Page 11: Mass spectrometry

Mass spectrometersMass spectrometersLinear Tim e Of Flight tube

Reflector Tim e Of Flight tube

detector

reflector

ion source

ion source

detector

tim e of flight

tim e of flight

• Time of flight (TOF) (MALDI)Time of flight (TOF) (MALDI)– Measures the time required for ions to fly down the length

of a chamber. – Often combined with MALDI (MALDI-TOF) Detections from

multiple laser bursts are averaged. Multiple laser

• Tandem MS- MS/MSTandem MS- MS/MS-separation and identification of compounds in complex mixtures- induce fragmentation and mass analyze the fragment ions. - Uses two or more mass analyzers/filters separated by a

collision cell filled with Argon or Xenon

• Different MS-MS configurations– Quadrupole-quadrupole (low energy)– Magnetic sector-quadrupole (high)– Quadrupole-time-of-flight (low energy)– Time-of-flight-time-of-flight (low energy)

Page 12: Mass spectrometry

Typical Mass Spectrometer

Page 13: Mass spectrometry

LC/LC-MS/MS-Tandem LC, Tandem MS

Page 14: Mass spectrometry

Typical Mass Spectrum• Characterized by sharp, narrow peaks

• X-axis position indicates the m/z ratio of a given ion (for singly charged ions this corresponds to the mass of the ion)

• Height of peak indicates the relative abundance of a given ion (not reliable for quantitation)

• Peak intensity indicates the ion’s ability to desorb or “fly” (some fly better than others)

Page 15: Mass spectrometry

m/z ratio:

Molecular weight divided by the Charge on this protein

All proteins are sorted based on a

mass to charge ratio (m/z)

Page 16: Mass spectrometry

Typical Mass Spectrum

aspirin

Relative Relative AbundanceAbundance

120 m/z-for singlysingly charged ion this is the mass

Page 17: Mass spectrometry

Resolution & Resolving PowerResolution & Resolving Power• Width of peak indicates the resolution of the MS

instrument

• The better the resolution or resolving power, the better the instrument and the better the mass accuracy

• Resolving power is defined as:M is the mass number of the observed mass (M) is the

difference between two masses that can be separated

Page 18: Mass spectrometry

Resolution in MSResolution in MS

Page 19: Mass spectrometry

Resolution in MSResolution in MS

QTOF

783.455

784.465

785.475

783.6

Page 20: Mass spectrometry

04/15/2320

Page 21: Mass spectrometry

Introduction to Mass Spectrometry

Sample introduction

IonizationMinimize collisions, interferences

Separatemasses

Count ionsCollect results

Nier-type mass spec

Page 22: Mass spectrometry

The sample cone isolates thetorch from the interior.

The torch box of an Agilent 7500 ICPMS

spray chambertorchAr feed

RF coil

Page 23: Mass spectrometry

04/15/2323

Page 24: Mass spectrometry

04/15/2324

Page 25: Mass spectrometry

N

SSource magnets

Filament

Collector

Sample Inlet+

+ ++

+

+++

+

++

+

+

+

+

+

Extraction lenses

Electron ionization

04/15/2325

Page 26: Mass spectrometry

In this type anode and cathode are arranged with a very

fine gap (0.5 to 2mm) which may serve as a slit.

The gaseous sample introduced at the anode points where

the electric field is concentrated.

The ionisation of the sample takes place by extraction of

electrons from the sample by microtips of the anode.

04/15/2326

Page 27: Mass spectrometry

N

SSource magnets

Filament

Collector

Sample Inlet+

+ ++

+

+++

+

++

+

+

+

+

+

Extraction lenses

+

+++

++

++

+ +

+

+

+

+

Chemical ionization

04/15/2327

Page 28: Mass spectrometry

These are the devices used to separate the ions produced in the ion source into their individual m/z ratios and focus them on the detector.

A number of different mass analysers exist in mass spectrometry.

04/15/2328

Page 29: Mass spectrometry

Magnetic sector – single focussing - double focussing

Quadrupole analysers

Ion Trap (Quistor) devices

Time-of-Flight (TOF) analysers

04/15/2329

Page 30: Mass spectrometry

Magnetic sectors – single focussing•Generally bulky and expensive.•Earliest type of analyser and still popular.•High accelerating potential especially in comparison with other methods (usually 4-10kV).•Magnets wedge shaped and must provide homogeneous fields.•Ion beam enters and exits at exactly 90˚.•Focuses ions according to their momentum.

04/15/2330

Page 31: Mass spectrometry

04/15/2331

Page 32: Mass spectrometry

Magnetic sectors – double focussing

•Adds a second electric sector to provide energy focussing of ions independent of mass.•Double focussing means that both the energy and momentum focus is designed to coincide at the collector slit.•Very high mass resolution can be achieved by this arrangement.•Very bulky and expensive but high performance.

04/15/2332

Page 33: Mass spectrometry

04/15/2333

Page 34: Mass spectrometry

Quadrupole Mass Analyzer

04/15/2334

+

+

++

Page 35: Mass spectrometry

The quadrupole consists of four parallel rods. The opposing rodshave the same polarity whilst adjacent rods have opposite polarity.

Each rod is applied with a DC and an RF voltage.Ions are scanned by varying the DC/Rf quadrupole voltages.

Only ions with the selected mass tocharge ratio will have the correct oscillatory pathway in the Rf field.

04/15/2335

Page 36: Mass spectrometry

• Consists of ring electrode and two end caps

• Principle very similar to quadrupole

• Ions stored by RF & DC fields• Scanning field can eject ions

of specific m/z• Advantages - MS/MS/MS…..

- High sensitivity full scan MS/MS

04/15/2336

Page 37: Mass spectrometry

04/15/2337

Page 38: Mass spectrometry

• In this type of analyser the sorting of the ions is done in absence of magnetic field.

• It operates on the principle that, if the ions produced are supplied with equal energy and allowed to travel predetermined distance then they will acquire different velocities depending of their masses.

04/15/2338

Page 39: Mass spectrometry

• The detector records the charge induced when an ion passes by or hits a surface

• Electron Multipliers (EM)*– Most common detector– -Can Detect positive and negative ions

04/15/2339

Page 40: Mass spectrometry

• Faraday Cup– Least expensive detector– Captured ions transfer charge to cup– used to calibrate other MS detectors

04/15/2340

Page 41: Mass spectrometry

• Photographic detection: This detector system is most

sensitive than any other detector because the photoplate integrates the ion signal over a period of time.

The photoplates are processed by the usual photographic techniques and read with the aid of densitometer.

04/15/2341

Page 42: Mass spectrometry

Different Ionization Methods• Electron Impact (EI - Hard method)

– small molecules, 1-1000 Daltons, structure

• Fast Atom Bombardment (FAB – Semi-hard)– peptides, sugars, up to 6000 Daltons

• Electrospray Ionization (ESI - Soft)– peptides, proteins, up to 200,000 Daltons

• Matrix Assisted Laser Desorption (MALDI-Soft)– peptides, proteins, DNA, up to 500 kD

Page 43: Mass spectrometry

Electron Impact IonizationElectron Impact Ionization

• Sample introduced into instrument by heating it until it evaporates

• Gas phase sample is bombarded with electrons coming from rhenium or tungsten filament (energy = 70 eV)

• Molecule is “shattered” into fragments (70 eV >> 5 eV bonds)

• Fragments sent to mass analyzer

Page 44: Mass spectrometry
Page 45: Mass spectrometry

EI Fragmentation of CHEI Fragmentation of CH33OHOH

CH3OH CH3OH+

CH3OH CH2O=H+ + H

CH3OH + CH3 + OH

CHO=H+ + HCH2O=H+

Why wouldn’t Electron Impact be suitable for analyzing proteins?

Page 46: Mass spectrometry

Why You Can’t Use EI For Analyzing Why You Can’t Use EI For Analyzing ProteinsProteins

• EI shatters chemical bonds

• Any given protein contains 20 different amino acids

• EI would shatter the protein into not only into amino acids but also amino acid sub-fragments and even peptides of 2,3,4… amino acids

• Result is 10,000’s of different signals from a single protein -- too complex to analyze

Page 47: Mass spectrometry

Soft Ionization MethodsSoft Ionization Methods

337 nm UV laser

MALDI

cyano-hydroxycinnamic acid

Gold tip needle

Fluid (no salt)

ESI

+_

Page 48: Mass spectrometry

Soft IonizationSoft Ionization • Soft ionization techniques keep the molecule of interest

fully intact

• Electro-spray ionization first conceived in 1960’s by Malcolm Dole but put into practice in 1980’s by John Fenn (Yale)

• MALDI first introduced in 1985 by Franz Hillenkamp and Michael Karas (Frankfurt)

• Made it possible to analyze large molecules via inexpensive mass analyzers such as quadrupole, ion trap and TOF

Page 49: Mass spectrometry
Page 50: Mass spectrometry

Ionization methodsIonization methods• Electrospray mass spectrometry (ESI-MS)

– Liquid containing analyte is forced through a steel capillary at high voltage to electrostatically disperse analyte. Charge imparted from rapidly evaporating liquid.

• Matrix-assisted laser desorption ionization (MALDI)– Analyte (protein) is mixed with large excess of matrix (small organic molecule)– Irradiated with short pulse of laser light. Wavelength of laser is the same as absorbance max

of matrix.

Page 51: Mass spectrometry

Electrospray IonizationElectrospray Ionization• Sample dissolved in polar, volatile buffer (no salts)

and pumped through a stainless steel capillary (70 - 150 m) at a rate of 10-100 L/min

• Strong voltage (3-4 kV) applied at tip along with flow of nebulizing gas causes the sample to “nebulize” or aerosolize

• Aerosol is directed through regions of higher vacuum until droplets evaporate to near atomic size (still carrying charges)

Page 52: Mass spectrometry

Electrospray (Detail)Electrospray (Detail)

Page 53: Mass spectrometry

Electrospray IonizationElectrospray Ionization• Can be modified to “nanospray” system with flow < 1

L/min

• Very sensitive technique, requires less than a picomole of material

• Strongly affected by salts & detergents

• Positive ion mode measures (M + H)+ (add formic acid to solvent)

• Negative ion mode measures (M - H)- (add ammonia to solvent)

Page 54: Mass spectrometry

Positive or Negative Ion Mode?Positive or Negative Ion Mode?

• If the sample has functional groups that readily accept H+ (such as amide and amino groups found in peptides and proteins) then positive ion detection is used-PROTEINS

• If a sample has functional groups that readily lose a proton (such as carboxylic acids and hydroxyls as found in nucleic acids and sugars) then negative ion detection is used-DNA

Page 55: Mass spectrometry

Matrix-Assisted Laser Desorption Matrix-Assisted Laser Desorption IonizationIonization

337 nm UV laser

MALDI

cyano-hydroxycinnamic acid

Page 56: Mass spectrometry

MALDIMALDI

• Sample is ionized by bombarding sample with laser light

• Sample is mixed with a UV absorbant matrix (sinapinic acid for proteins, 4-hydroxycinnaminic acid for peptides)

• Light wavelength matches that of absorbance maximum of matrix so that the matrix transfers some of its energy to the analyte (leads to ion sputtering)

Page 57: Mass spectrometry

HT Spotting on a MALDI Plate

Page 58: Mass spectrometry

MALDI IonizationMALDI Ionization

++

+

+

-

--

++

+

+

-

---++

Analyte

Matrix

Laser

+

++

• Absorption of UV radiation by chromophoric matrix and ionization of matrix

• Dissociation of matrix, phase change to super-compressed gas, charge transfer to analyte molecule

• Expansion of matrix at supersonic velocity, analyte trapped in expanding matrix plume (explosion/”popping”)

+

+

+

Page 59: Mass spectrometry

MALDIMALDI• Unlike ESI, MALDI generates spectra that have just a singly

charged ion

• Positive mode generates ions of M + H

• Negative mode generates ions of M - H

• Generally more robust that ESI (tolerates salts and nonvolatile components)

• Easier to use and maintain, capable of higher throughput

• Requires 10 L of 1 pmol/L sample

Page 60: Mass spectrometry

Principal for MALDI-TOF MASS Principal for MALDI-TOF MASS

++

+++

++

+ ++

p u lse dU V o r IR la se r(3 -4 n s )

d e te c to r

vacuum

strong electric field

Time Of Flight tube

p e p tid e m ixtu ree m b e d de d in ligh t a b so rb in g ch em ica ls (m a trix )

c lo u d o fp ro ton a te dp e p tid e m o lecu le sa ccV

Page 61: Mass spectrometry

Principal for MALDI-TOF MASSPrincipal for MALDI-TOF MASS

Linear Time Of Flight tube

Reflector Time Of Flight tube

detector

reflector

ion source

ion source

detector

tim e of flight

tim e of flight

Page 62: Mass spectrometry

MALDI MALDI == SELDISELDI

337 nm UV laser

MALDI

cyano-hydroxycinnaminic acid

Page 63: Mass spectrometry

MALDI/SELDI SpectraMALDI/SELDI Spectra

Normal

Tumor

Page 64: Mass spectrometry

Background of fragmentation

• The impact of a stream of high energy electrons causes the molecule to lose an electron forming a radical cation.– A species with a positive charge and one unpaired

electron

+ e-C H

H

HH H

H

H

HC + 2 e-

Molecular ion (M+)

m/z = 16

Page 65: Mass spectrometry

Background

• The impact of the stream of high energy electrons can also break the molecule or the radical cation into fragments.

(not detected by MS)

m/ z = 29

molecular ion (M+) m/ z = 30

+ C

H

H

H

+ H

HH C

H

H

C

H

H

H C

H

H

C

H

H

H C

H

H

+ e-H C

H

H

C

H

H

H

m/z = 15

Page 66: Mass spectrometry

Background

• Molecular ion (parent ion):– The radical cation corresponding to the mass of the

original molecule

• The molecular ion is usually the highest mass in the spectrum– Some exceptions w/specific isotopes– Some molecular ion peaks are absent.

HH

H

HC H C

H

H

C

H

H

H

Page 67: Mass spectrometry

Background

• Mass spectrum of ethanol (MW = 46)M+

Page 68: Mass spectrometry

Background

• The cations that are formed are separated by magnetic deflection.

Page 69: Mass spectrometry

Background

• Only cations are detected.– Radicals are “invisible” in MS.

• The amount of deflection observed depends on the mass to charge ratio (m/z).– Most cations formed have a charge of +1 so the

amount of deflection observed is usually dependent on the mass of the ion.

Page 70: Mass spectrometry

Background

• The resulting mass spectrum is a graph of the mass of each cation vs. its relative abundance.

• The peaks are assigned an abundance as a percentage of the base peak. – the most intense peak in the spectrum

• The base peak is not necessarily the same as the parent ion peak.

Page 71: Mass spectrometry

Background

M+base peak

The mass spectrum of ethanol

Page 72: Mass spectrometry

Background

• Most elements occur naturally as a mixture of isotopes.– The presence of significant amounts of heavier

isotopes leads to small peaks that have masses that are higher than the parent ion peak.

• M+1 = a peak that is one mass unit higher than M+

• M+2 = a peak that is two mass units higher than M+

Page 73: Mass spectrometry

Easily Recognized Elements in MS

• Nitrogen:– Odd number of N = odd MW

CH3CNM+ = 41

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/2/09)

Page 74: Mass spectrometry

Easily Recognized Elements in MS Bromine:

M+ ~ M+2 (50.5% 79Br/49.5% 81Br)

2-bromopropane

M+ ~ M+2

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/1/09)

Page 75: Mass spectrometry

Easily Recognized Elements in MS• Chlorine:

– M+2 is ~ 1/3 as large as M+

Cl

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/2/09)

M+2

M+

Page 76: Mass spectrometry

• Sulfur:– M+2 larger than usual (4% of M+)

Easily Recognized Elements in MS

M+

Unusually large M+2

S

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/1/09)

Page 77: Mass spectrometry

Easily Recognized Elements in MS• Iodine

– I+ at 127– Large gap

Large gap

I+

M+

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/2/09)

I CH2CN

Page 78: Mass spectrometry

Fragmentation Patterns

• The impact of the stream of high energy electrons often breaks the molecule into fragments, commonly a cation and a radical.– Bonds break to give the most stable cation.– Stability of the radical is less important.

Page 79: Mass spectrometry

Fragmentation Patterns

• Alkanes– Fragmentation often splits off simple alkyl groups:

• Loss of methyl M+ - 15• Loss of ethyl M+ - 29• Loss of propyl M+ - 43• Loss of butyl M+ - 57

– Branched alkanes tend to fragment forming the most stable carbocations.

Page 80: Mass spectrometry

Fragmentation Patterns• Mass spectrum of 2-methylpentane

Page 81: Mass spectrometry

Fragmentation Patterns• Alkenes:

– Fragmentation typically forms resonance stabilized allylic carbocations

Page 82: Mass spectrometry

Fragmentation Patterns• Aromatics:

– Fragment at the benzylic carbon, forming a resonance stabilized benzylic carbocation (which rearranges to the tropylium ion)

M+

CH

H

CH Br

HC

H

H

or

Page 83: Mass spectrometry

Fragmentation Patterns• Alcohols

– Fragment easily resulting in very small or missing parent ion peak

– May lose hydroxyl radical or water• M+ - 17 or M+ - 18

– Commonly lose an alkyl group attached to the carbinol carbon forming an oxonium ion.• 1o alcohol usually has prominent peak at m/z =

31 corresponding to H2C=OH+

Page 84: Mass spectrometry

Fragmentation Patterns

• MS for 1-propanol

M+M+-18

CH3CH2CH2OH

H2C OH

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/28/09)

Page 85: Mass spectrometry

Fragmentation Patterns

• Amines– Odd M+ (assuming an odd number of nitrogens

are present)– -cleavage dominates forming an iminium ion

CH3CH2 CH2 N

H

CH2 CH2CH2CH3 CH3CH2CH2N CH2

H

m/ z =72

iminium ion

Page 86: Mass spectrometry

Fragmentation Patterns86

CH3CH2 CH2 N

H

CH2 CH2CH2CH3

72

Page 87: Mass spectrometry

Fragmentation Patterns

• Ethers– -cleavage forming oxonium ion

– Loss of alkyl group forming oxonium ion

– Loss of alkyl group forming a carbocation

Page 88: Mass spectrometry

Fragmentation Patterns

H O CHCH3

MS of diethylether (CH3CH2OCH2CH3)

CH3CH2O CH2H O CH2

Page 89: Mass spectrometry

Fragmentation Patterns

• The impact of the stream of high energy electrons often breaks the molecule into fragments, commonly a cation and a radical.– Bonds break to give the most stable cation.– Stability of the radical is less important.

Page 90: Mass spectrometry

Fragmentation Patterns

• Alkanes– Fragmentation often splits off simple alkyl groups:

• Loss of methyl M+ - 15• Loss of ethyl M+ - 29• Loss of propyl M+ - 43• Loss of butyl M+ - 57

– Branched alkanes tend to fragment forming the most stable carbocations.

Page 91: Mass spectrometry

Fragmentation Patterns• Mass spectrum of 2-methylpentane

Page 92: Mass spectrometry

Fragmentation Patterns• Alkenes:

– Fragmentation typically forms resonance stabilized allylic carbocations

Page 93: Mass spectrometry

Fragmentation Patterns• Aromatics:

– Fragment at the benzylic carbon, forming a resonance stabilized benzylic carbocation (which rearranges to the tropylium ion)

M+

CH

H

CH Br

HC

H

H

or

Page 94: Mass spectrometry

Fragmentation PatternsAromatics may also have a peak at m/z = 77 for the benzene ring.

NO2

77M+ = 123

77

Page 95: Mass spectrometry

Fragmentation Patterns• Alcohols

– Fragment easily resulting in very small or missing parent ion peak

– May lose hydroxyl radical or water• M+ - 17 or M+ - 18

– Commonly lose an alkyl group attached to the carbinol carbon forming an oxonium ion.• 1o alcohol usually has prominent peak at m/z =

31 corresponding to H2C=OH+

Page 96: Mass spectrometry

Fragmentation Patterns

• MS for 1-propanol

M+M+-18

CH3CH2CH2OH

H2C OH

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/28/09)

Page 97: Mass spectrometry

Fragmentation Patterns

• Amines– Odd M+ (assuming an odd number of nitrogens

are present)– -cleavage dominates forming an iminium ion

CH3CH2 CH2 N

H

CH2 CH2CH2CH3 CH3CH2CH2N CH2

H

m/ z =72

iminium ion

Page 98: Mass spectrometry

Fragmentation Patterns86

CH3CH2 CH2 N

H

CH2 CH2CH2CH3

72

Page 99: Mass spectrometry

Fragmentation Patterns

• Ethers– -cleavage forming oxonium ion

– Loss of alkyl group forming oxonium ion

– Loss of alkyl group forming a carbocation

Page 100: Mass spectrometry

Fragmentation Patterns

• Aldehydes (RCHO)– Fragmentation may form acylium ion

– Common fragments:

• M+ - 1 for • M+ - 29 for

RC O

R (i.e. RCHO - CHO)

RC O

Page 101: Mass spectrometry

Fragmentation Patterns• MS for hydrocinnamaldehyde

M+ = 134C C C H

H

H

H

H

O

133

105

91105

91

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/28/09)

Page 102: Mass spectrometry

Fragmentation Patterns

• Ketones– Fragmentation leads to formation of acylium ion:

• Loss of R forming

• Loss of R’ forming

RC O

R'C O

RCR'

O

Page 103: Mass spectrometry

Fragmentation Patterns• MS for 2-pentanone

CH3CCH2CH2CH3

O

M+

CH3CH2CH2C O

CH3C O

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/28/09)

Page 104: Mass spectrometry

Fragmentation Patterns

• Esters (RCO2R’)– Common fragmentation patterns include:

• Loss of OR’ – peak at M+ - OR’

• Loss of R’ – peak at M+ - R’

Page 105: Mass spectrometry

Frgamentation Patterns

M+ = 136

C

O

O CH3

105

77 105

77

SDBSWeb : http://riodb01.ibase.aist.go.jp/sdbs/ (National Institute of Advanced Industrial Science and Technology, 11/28/09)

Page 106: Mass spectrometry

Rule of Thirteen• The “Rule of Thirteen” can be used to identify

possible molecular formulas for an unknown hydrocarbon, CnHm.

– Step 1: n = M+/13 (integer only, use remainder in step 2)

– Step 2: m = n + remainder from step 1

Page 107: Mass spectrometry

Rule of Thirteen• Example: The formula for a hydrocarbon with

M+ =106 can be found:

– Step 1: n = 106/13 = 8 (R = 2)

– Step 2: m = 8 + 2 = 10

– Formula: C8H10

Page 108: Mass spectrometry

Rule of Thirteen

• If a heteroatom is present, – Subtract the mass of each heteroatom from the

MW– Calculate the formula for the corresponding

hydrocarbon– Add the heteroatoms to the formula

Page 109: Mass spectrometry

Rule of Thirteen

Example: A compound with a molecular ion peak at m/z = 102 has a strong peak at 1739 cm-

1 in its IR spectrum. Determine its molecular formula.

Page 110: Mass spectrometry

References

• Chatwal GR, Anand SK. Instrumental method of chemical analysis, Himalaya publishing house.

• Sharma YR. Elementary organic spectroscopy.• Willard,merritt,dean. Instumental methods of

analysis. • www.google.com/images