Manual CarboQC
-
Upload
eunice-hdz-lomas -
Category
Documents
-
view
1.231 -
download
196
description
Transcript of Manual CarboQC
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 1/78
Instruction Manual
CarboQC
Lab and At-line Beverage Carbonation Meter Firmware Version : V2.00E
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 2/78
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 3/78
Instruction Manual
CarboQC
Lab and At-line Beverage Carbonation Meter Firmware Version: V2.00E
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 4/78
Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.
Nor is any liability assumed for damages resulting from information contained in the document.
Anton Paar GmbH reserves the right to perform changes to the content. This also extends to changes
to delivery volumes or any features of delivered parts.
All rights reserved (including translation). This document, or any part of it, may not be reproduced,
changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm
or any other process) without prior written permission by Anton Paar GmbH.
Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked
as such. They are the property of their respective owner.
Published by Anton Paar Printed: Anton Paar, Austria
Copyright © 2012 Anton Paar GmbH, Graz, Austria
Address: Anton Paar GmbH Anton-Paar-Str. 20 A-8054 Graz / Austria – EuropeTel: +43 316 257-0Fax: +43 316 257-257
E-Mail: [email protected]: www.anton-paar.com
Date: May 21, 2012
Document number: C26IB001EN-B
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 5/78
C26IB001EN-B 5
Contents
1 About the Inst ruct ion Manual ......................................................................................................... 8
2 Safety Inst ruct ions........................................................................................................................... 9
2.1 General Safety Instructions ..................................................................................................... 9
2.2 Special Safety Instructions for CarboQC in Combination with PFD Filling Device .................11
2.3 Warning Signs and Mandatory Signs on the PFD Filling Device............................................11
3 About CarboQC and PFD Fil ling Device ...................................................................................... 13
4 Checking the Suppl ied Parts ........................................................................................................ 14
5 Functional Components ................................................................................................................ 16
5.1 Front View.............................................................................................................................. 16
5.2 CarboQC Components .......................................................................................................... 165.2.1 Measuring Chamber............................................................................................... 16
5.2.2 Display.................................................................................................................... 17
5.2.3 Keypad ................................................................................................................... 18
5.3 Rear View.............................................................................................................................. 19
5.4 Integrated Battery .................................................................................................................. 19
6 Putting CarboQC into Operat ion .................................................................................................. 20
6.1 Installing CarboQC ................................................................................................................ 20
6.2 Switching the Instrument On and Off..................................................................................... 23
7 Inst rument Settings ....................................................................................................................... 25
7.1 Menu Access ......................................................................................................................... 25
7.2 Setting Date and Time ........................................................................................................... 25
7.3 Setting the Units .................................................................................................................... 25
8 Measurement .................................................................................................................................. 26
8.1 Preconditions......................................................................................................................... 26
8.2 Preparing CarboQC............................................................................................................... 26
8.3 Transferring the Sample into the Measuring Chamber.......................................................... 27
8.4 Measurement Using the PFD Filling Device.......................................................................... 27
8.4.1 Setting the Correct Flow......................................................................................... 27
8.4.2 Measurement Procedure........................................................................................ 29
8.5 Checking for a Correct Measurement Procedure and Avoiding Errors.................................. 33
9 Checking the System..................................................................................................................... 34
9.1 Influencing Factors ................................................................................................................ 34
9.2 Zero Point Check................................................................................................................... 34
9.3 Check with Carbonated Liquid............................................................................................... 34
9.4 Beginning of the Shift/Day..................................................................................................... 36
9.5 Entering a Pressure Offset .................................................................................................... 37
10 Cleaning the System...................................................................................................................... 38
10.1 End of the Shift/Day............................................................................................................... 3810.2 Weekly Maintenance ............................................................................................................. 39
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 6/78
6 C26IB001EN-B
10.3 Cleaning the Filter.................................................................................................................. 40
10.4 Cleaning the Measuring Chamber......................................................................................... 40
10.4.1 Required Devices ................................................................................................... 40
10.4.2 Rinsing ................................................................................................................... 40
10.4.3 Cleaning the Entire Measuring Chamber Volume .................................................. 41
10.5 Cleaning the Instrument Housing and Display....................................................................... 41
11 Measuring At-line........................................................................................................................... 42
11.1 Using CarboQC with Protective Casing................................................................................. 42
11.1.1 Overview ................................................................................................................ 42
11.1.2 Mounting the Protective Casing ............................................................................. 44
11.1.3 Connecting CarboQC in the Protective Casing to the Production Line or Tank ..... 45
11.1.4 Measurement ......................................................................................................... 46
11.1.5 Cleaning the Measuring Chamber.......................................................................... 47
11.2 Using CarboQC without Protective Casing............................................................................ 4811.2.1 Connecting CarboQC to the Production Line or Tank ............................................ 48
11.2.2 Measurement ......................................................................................................... 49
11.2.3 Cleaning the Measuring Chamber.......................................................................... 50
12 Menu Operat ion.............................................................................................................................. 51
12.1 Using the Keys on the Keypad .............................................................................................. 51
12.2 Menu Description and Function............................................................................................. 51
12.2.1 Displaying the Measurement Data ......................................................................... 51
12.2.2 Accessing the Menu............................................................................................... 52
12.2.3 "adjustment" Submenu........................................................................................... 5212.2.4 "instrument" Submenu............................................................................................ 53
12.2.5 "methods" Submenu............................................................................................... 55
12.2.6 "data memory" Submenu ....................................................................................... 59
12.2.7 "service" Submenu ................................................................................................. 60
13 Techn ical Background................................................................................................................... 61
14 CarboQC Methods ......................................................................................................................... 62
14.1 Available Methods with CarboQC.......................................................................................... 62
14.1.1 Standard Methods .................................................................................................. 62
14.1.2 Customer-Specific Methods ................................................................................... 62
14.2 Establishing Customer-Specific Formulae............................................................................. 62
14.2.1 Converting the CO2 Result into a Customer-Specific Result.................................. 62
14.2.2 Converting the CO2 Concentration into a Customer-Specific Unit ......................... 64
14.2.3 Using a Customer-Specific Solubility Coefficient.................................................... 64
14.3 Transformation of the Determined Air Content ...................................................................... 65
14.4 Calculation of the Package Pressure at a Reference Temperature....................................... 66
Appendix A: Technical Data ........................................................................................................... 68
A.1 CarboQC ............................................................................................................. 68
A.2 PFD Filling Device............................................................................................... 69
Appendix B: Wetted Par ts ............................................................................................................... 70
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 7/78
C26IB001EN-B 7
Appendix C: Interface Commands for Communicat ion between a PC and CarboQC .............. 71
Appendix D: Error Messages ......................................................................................................... 72
Appendix E: Troubleshooting ........................................................................................................ 73
Appendix F: Firmware Versions .................................................................................................... 74
Appendix G: Declarat ion o f Conformi ty ........................................................................................ 75
Appendix H: Menu Tree................................................................................................................... 76
Index..................................................................................................................................................... 77
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 8/78
1 About the Instruction Manual
8 C26IB001EN-B
1 About the Instruction Manual
This instruction manual informs you about the installation and the safe handling
and use of the product. Pay special attention to the safety instructions and
warnings in the manual and on the product.
The instruction manual is a part of the product. Keep this instruction manual for
the complete working life of the product and make sure it is easily accessible to
all people involved with the product. If you receive any additions or revisions to
this instruction manual from Anton Paar GmbH, these must be treated as part of
the instruction manual.
Conventions for safety messages
The following conventions for safety messages are used in this instruction
manual:
TIP Tip gives extra information about the situation at hand.
Typographical conventions
The following typographical conventions are used in this instruction manual:
DANGER
Danger indicates a hazardous situation which, if not avoided, will result in
death or serious injury.
WARNING
Warning indicates a hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Caution indicates a hazardous situation which, if not avoided, could result in
minor or moderate injury.
NOTICE
Notice indicates a situation which, if not avoided, could result in damage to
property.
Convention Description
<key> The names of keys and buttons are
written inside angle brackets.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 9/78
2 Safety Instructions
C26IB001EN-B 9
2 Safety Instructions
2.1 General Safety Instructions
• Read this instruction manual before using CarboQC.
• Follow all hints and instructions contained in this instruction manual toensure the correct use and safe functioning of CarboQC.
Liability
• This instruction manual does not claim to address all safety issues
associated with the use of the instrument and samples. It is your
responsibility to establish health and safety practices and determine the
applicability of regulatory limitations.
• Anton Paar GmbH only warrants the proper functioning of CarboQC if no
adjustments have been made to the mechanics, electronics, and firmware.
• Only use CarboQC for the purpose described in this instruction manual.
Anton Paar GmbH is not liable for damages caused by incorrect use of
CarboQC.
Installation and Use
• CarboQC is not an explosion-proof instrument and therefore must not be
operated in areas with risk of explosion.
• The installation procedure shall only be carried out by authorized personnel
who are familiar with the installation instructions.
• Do not use any accessories or wearing parts other than those supplied or
approved by Anton Paar GmbH.
• Make sure all operators are trained to use the instrument safely and correctly
before starting any applicable operations.
• In case of damage or malfunction, do not continue operating CarboQC. Do
not operate the instrument under conditions which could result in damage to
goods and/or injuries and loss of life.
"Menu Level 1 > Menu Level 2" Menu paths are written in bold, inside
straight quotation marks. The menu
levels are connected using a closing
angle bracket.
Convention Description
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 10/78
2 Safety Instructions
10 C26IB001EN-B
• Check CarboQC for chemical resistance to the samples and cleaning
agents.
• Neither CarboQC nor, if applicable, the PFD Filling Device shall be used for
or filled with acids and alkalis above 5 %w/w, hydrofluoric acid, solvent and
other poisonous liquids.
• Do not place CarboQC on a wet or humid bench because humidity entering
the housing could damage the instrument.
• Place CarboQC and, if applicable, the PFD Filling Device in a way so that
there is no danger of accidentally dropping the equipment or tearing off the
pressurized hoses and fittings. Dropping the equipment increases the risk of
the pressurized container bursting.
• Do not expose CarboQC and the PFD Filling Device to direct sunlight for
extended periods of time.
Maintenance and Service
• Due to the nature of the measurement, the measuring results not only
depend on the correct use and functioning of the CarboQC and, if applicable,
the PFD Filling Device, but may also be influenced by other factors. We
therefore recommend you have the results checked (e.g. plausibility tested)
by skilled personnel before consequential actions are taken.
• Service and repair procedures may only be carried out by authorized
personnel or by Anton Paar GmbH.
Disposal
• Dispose of CarboQC and, if applicable, the PFD Filling Device according to
the country-specific legal requirements.
• Dispose of samples and cleaning liquids according to the country-specific
legal requirements.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 11/78
2 Safety Instructions
C26IB001EN-B 11
2.2 Special Safety Instructions for CarboQC in
Combination with PFD Filling Device
• Do not supply compressed gas with more than 6.5 bar (95 psi) relative
pressure to the PFD Filling Device.
• Use only clean and oil-free compressed air ("instrument air").
• In the unlikely event that a bottle or can bursts, immediately push the
activating lever of the PFD Filling Device to the de-energized position. Turn
off the compressed gas supply and then disconnect the PFD Filling Device
from the compressed gas supply. Renew the safety shield and discard the
old one. Damage of the instrument or parts thereof may be the consequence
of a burst container. Therefore, do not continue to use the PFD Filling Device
unless it has been checked by authorized service personnel. Contact your
local Anton Paar representative.
• The safety shield of the PFD Filling Device is subject to aging, and has to be
replaced after 5 years at the latest.
2.3 Warning Signs and Mandatory Signs on the PFD
Filling Device
Warning sign on the rear side of the PFD Filling Device
WARNING
If a glass bottle bursts inside the PFD, the safety shield protects the user from
glass chips. However, some small glass particles may still be released by the
rear gap of the safety shield which could lead to injuries of eye or face.
Hearing damage is possible in case a bottle or can bursts. Also damage of the
safety shield may occur.
• Therefore, place the PFD Filling Device with a distance of 0.5 m (20inches) with its back to the wall.
• The PFD Filling Device has to be placed so nobody can stand betweenthe PFD Filling Device and the wall.
• When pressurizing a glass bottle inside the PFD, only the operator,equipped with appropriate protective clothing, may be located directly infront of the PFD Filling Device. All other personnel has to keep a distanceof at least 3 m to the PFD Filling Device.
• If a container bursts, the vicinity might get sprayed with sample. The PFDFilling Device has to be positioned so that there are no power outlets orventilation slots behind the PFD Filling Device.
• In case of damage, the safety shield has to be replaced by trained servicepersonnel.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 12/78
2 Safety Instructions
12 C26IB001EN-B
On the front side of the PFD Filling Device
Mandatory signs
WARNING
If a glass bottle bursts inside the PFD, the safety shield protects the user fromglass chips. However, some small glass particles may still be released by the
rear gap of the safety shield which could lead to injuries of eye or face.
Hearing damage is possible in case a bottle or can bursts. Also damage of the
safety shield may occur.
• Do not pressurize empty bottles or cans. The safety shield is onlydimensioned for gas volumes resulting from the depth of immersion of thesample tube.
• Do not pressurize glass bottles with a volume higher than 1 L.Otherwise,the protection through the safety shield is not ensured.
• Do not pressurize bottles/cans which show cracks or substantialscratches. Adhere to specifications of bottle manufacturers.
• In the unlikely case of a bottle/can bursting, the safety shield may getdamaged and therefore has to be replaced by a new one. Renew thesafety shield and discard the old one.
Wear safety goggles:
Wear safety goggles when handling the PFD Filling Device.
Wear ear protection:
Wear ear protection when handling the PFD Filling Device.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 13/78
3 About CarboQC and PFD Filling Device
C26IB001EN-B 13
3 About CarboQC and PFD Filling
Device
CarboQC uses the patented "Multiple Volume Expansion" (MVE) method toeliminate the influence of dissolved air on the determined carbon dioxide content.
The analysis is based on highly accurate absolute pressure and temperature
measurements.
CarboQC can determine the CO2 content of sugar based soft drinks, soft drinks
with low sugar concentration, diet drinks, beer, strong beer, sparkling wine and
mineral water.
CarboQC can be used in a laboratory environment as well as at-line. For at-line
measurements, a special protective casing is available.
The instrument is equipped with rechargeable batteries to provide the required
power for a limited period of time.
To perform a measurement, connect the instrument to the PFD Filling Device in
the laboratory or to a production line or tank, select the proper method, and press
the <Start> key. Within approx. 2 minutes the CO2 concentration is shown on the
display and transferred to the data memory. The measuring result can also be
printed out immediately or later from the data memory.
TIP For detailed information about the PFD Filling Device, see the PFD Filling Device
instruction manual.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 14/78
4 Checking the Supplied Parts
14 C26IB001EN-B
4 Checking the Supplied Parts
CarboQC was tested and packed carefully before shipment. However, damage
may occur during transport.
1. Keep the packaging material (box, foam piece, transport protection) for
possible returns and further questions from the transport and insurance
company.
2. Check the delivery for completion by comparing the supplied parts to those
noted in Table 4.1.
3. For checking the supplied parts of the PFD Filling Device, see the PFD
Filling Device instruction manual.
4. If a part is missing, contact your Anton Paar representative.
5. If a part is damaged, contact the transport company and your Anton Paar
representative.
Table 4.1: Supplied parts
Symbol Pcs. Article Description Mat. No.
1 CarboQC 14805
1 Instruction manual
English or
German
16605
16606
1 Power adapter incl. cable 14869
4m Hose 2.5x4 polyurethane
transparent
15272/
1m
1 Filter incl. 1# sieve
1# Sieve D = 8 mm for filter
20054
16118
1 Straight fitting R1/8" 4 mm 15274
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 15/78
4 Checking the Supplied Parts
C26IB001EN-B 15
1 T-Hose connector 4 mm 17756
1 Wrench 10 3636
Table 4.1: Supplied parts
Symbol Pcs. Article Description Mat. No.
Optional accessories Mat. No.
Printer RS-232C 64434Printer interface cable for CarboQC 5-pole incl. Dsub adapter 25pin/
9pin
22434
Power adapter IP76 incl. cable
Splash-proof power supply with European standard power plug
23672
Customer functions upon
request
Fruit pulp filter incl. housing and 10# filter cartridges
10# Filter cartridges for fruit pulp filter
21167
21168
All-purpose filter incl. 1# filter cartridge
1# Filter cartridge for all-purpose filter
25483
25391
Filter with male hose filling nozzle for at-line operation 18829
Carrying strap 16012
Protective boot CarboQC 90079
Carry case for CarboQC, 440 x 250 x 170 mm 23888
Protective casing CarboQC 102189
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 16/78
5 Functional Components
16 C26IB001EN-B
5 Functional Components
5.1 Front View
Fig. 5 - 1 Front view
1 ... Flow valve2 ... Sample outlet
3 ... Sample inlet4 ... Lugs for carrying strap5 ... Keypad6 ... Display7 ... Measuring chamber
5.2 CarboQC Components
5.2.1 Measuring Chamber
Fig. 5 - 2 Measuring chamber
1
2
3
4
5
6
7
Flow valve
Sample outlet
Sample inlet
Piston
to adjust the sample flow
with integrated stirrer
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 17/78
5 Functional Components
C26IB001EN-B 17
5.2.2 Display
Fig. 5 - 3 Alphanumerical LCD display (4 lines, 16 characters)
1 ... Activated method2 ... Battery status symbol3 ... Sample number 4 ... CO2 concentration
5 ... Air content6 ... Package pressure (pt) or measuring temperature (tm)
The 4th line either displays the measuring temperature or the package pressure.
The package pressure is calculated from the determined concentration of CO2
and other dissolved gases.
To switch the display from package pressure to measuring temperature, press the
<Menu> key (see Fig. 5 - 4) for approx. 5 seconds.
Battery status symbol
/
(blinking)
The battery is being charged.
The battery is fully charged and the power adapter is connected.
The battery is fully charged and the power adapter is not
connected.
The battery is charged approx. 66 %.
The battery is charged approx. 33 %.
(blinking)
The battery is empty. The battery has to be charged again by
connecting CarboQC to the mains via the power adapter.
1 2 3
4
5
6
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 18/78
5 Functional Components
18 C26IB001EN-B
5.2.3 Keypad
Fig. 5 - 4 Keypad
In the measuring mode the functions <Sample>, <Menu>, <Print>, <Method>,
<Rinse>, <Start> and <Stop> are active.
In the menu mode the functions <ESC>, <>, <> and <Enter> are active.
Key functions in the measuring mode
Special key func tions activated with a long push (approx. 5 sec.)
Menu/ To access the menu.
Method/Enter To select one of 7 predefined or one of 17 customer-
specific methods.
Print/ To start a printout of the actual measuring result.
Rinse To rinse the measuring chamber. Further short pushes
activate/deactivate the integrated stirrer.Sample/ESC For entering a sample number. This number will be
increased each time you press <Start>.
Start To start a measurement.
Stop To stop a measurement, rinsing or cleaning procedure.
Menu To switch between measuring temperature and package
pressure.
Print To print out all measuring results stored in the data memory ofCarboQC.
Rinse To rinse the measuring chamber according to the settings in the
menu "instrument > rinse".
Start To clean the measuring chamber at its maximum volume
expansion.
Stop To switch to method selection mode. While keeping the key
pressed, CarboQC displays the available methods in intervals of
2 seconds. Releasing the key when the desired method is
displayed selects this method and the instrument switches back
to measurement mode.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 19/78
5 Functional Components
C26IB001EN-B 19
Special key combination (simultaneously pressed)
5.3 Rear View
Fig. 5 - 5 Rear view
1 ... Handle2 ... COM1 interface for PC or printer 3 ... Type plate with serial number 4 ... On/off switch5 ... COM2 interface for printer or data transfer to a PC with AP-SoftPrint6 ... Cover 7 ... Supply inlet
5.4 Integrated Battery
CarboQC has an integrated rechargeable battery. Up to 200 measurements can
be performed with a completely charged battery.
TIP To extend the lifetime of the internal battery, perform a battery service procedure
once every 3 months (see Chapter 12.2.4).
A battery service can be performed on-site.
• To perform this service, select the menu "instrument > battery" and set
"service" to "on" .
The battery service will take a few hours and is best performed overnight. Itswitches itself off upon completion.
Start and Stop To start a rinsing procedure that has to be stopped by
pressing the <Stop> key.
1
2
34
5
6
7
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 20/78
6 Putting CarboQC into Operation
20 C26IB001EN-B
If the battery has to be exchanged by an authorized service technician, make sure
that the battery service is performed, even if the newly installed battery is fully
charged.
6 Putting CarboQC into Operation
6.1 Instal ling CarboQC
CarboQC can be used in the laboratory as well as at-line. In any case, sample
has to be transferred into the measuring chamber of CarboQC without the loss of
carbon dioxide during the transfer. Subsequently, the installation of CarboQC in
combination with a PFD Filling Device is described.
The PFD Filling Device pierces the bottle closure or the base of the can and
forces the beverage into the measuring chamber of CarboQC using compressed
gas (6±0.5 bar relative/87±7 psi). In the unlikely event that a bottle or can bursts,
the operator is protected by a safety shield which is moved vertically over a bottle
or can. A safety pin on the bottom of the safety shield ensures that the system
cannot be pressurized as long as the safety shield is not in the protective position.
WARNING
If a glass bottle bursts inside the PFD, the safety shield protects the user from
glass chips. However, some small glass particles may still be released by therear gap of the safety shield which could lead to injuries of eye or face.
Hearing damage is possible in case a bottle or can bursts. Also damage of the
safety shield may occur.
• If a container bursts, chippings may be hurled through the gap on the backof the PFD Filling Device. Therefore place the PFD Filling Device with adistance of 0.5 m (20 inches) with its back to the wall. The PFD FillingDevice has to be placed so nobody can stand between the PFD FillingDevice and the wall.
WARNING
Compressed gas is used. Loose hoses may lead to injuries, especially of eyes
and face.
• If CarboQC is operated in combination with a PFD Filling Device, do notactivate the compressed gas supply before all hoses are properlyconnected.
NOTICE
• Do not operate CarboQC without the filter being installed.
• Do not place CarboQC on a wet or humid bench because humidityentering the housing could damage the instrument.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 21/78
6 Putting CarboQC into Operation
C26IB001EN-B 21
TIP If required, fasten the PFD Filling Device to the bench with clamps using the
opening in the front of the base plate or with a bolt through the corresponding hole
in the base plate.
Compressed gas supply
CarboQC and PFD Filling Device are best operated with compressed air. In case
a dissolved oxygen meter is also connected, use compressed nitrogen as
compressed air might influence the oxygen result.
If there is no in-house supply of compressed gas available, the PFD Filling Device
can also be operated with compressed gas from a gas cylinder.
The hoses of the PFD Filling Device have an outer diameter of 6 mm. If the
compressed gas outlet does not match the required diameter for PFD Filling
Device operation, check for pressure adapters at your local supplier.
Mounting the hoses
For the hose connections between CarboQC and PFD Filling Device, use the
delivered polyurethane hose 2.5 x 4 mm.
To connect a hose
• To connect a hose to a fitting, push the hose into the fitting as far as it will go.
TIP When reconnecting a hose, cut off a few millimeters of the hose if the color of the
hose is changed to achieve tight seal.
To disconnect a hose
• To remove the connection, press the ring of the fitting towards the fitting andthen pull out the hose.
NOTICE
Keep the hoses between the instruments as short as possible. This reduces
sample consumption as well as the risk of CO2 diffusing through the hoses
thus influencing the results.
NOTICE
Only take apart connections if it is really necessary to do so as frequent
handling may destroy the fitting.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 22/78
6 Putting CarboQC into Operation
22 C26IB001EN-B
Fig. 6 - 1 CarboQC and PFD Filling Device
1 ... PFD Filling Device sample tube outlet2 ... Connection to compressed gas supply
2a PFD Filling Device outlet for compressed gas 2b PFD Filling Device inlet for compressed gas
3 ... Hose coming from the PFD Filling Device sample tube outlet4 ... Filter 5 ... CarboQC sample inlet6 ... CarboQC sample outlet7 ... Hose with T-hose connector to be led to the waste vessel
**) according to your requirements
2
4
5
7
1
1
2a 2b3
6
HoseInner
diameter
Outer
diameter Connects
Recommended
length
Nylon hose4.0 mm
(0.16")
6.0 mm
(0.24")
PFD Filling Device compressed
gas inlet with compressed gas
supply
**)
PFD Filling Device compressed
gas outlet with T-hose connector
to be led into the waste vessel
**)
Polyurethane hose,
transparent
2.5 mm
(0.1")
4.0 mm
(0.16")
PFD Filling Device sample tube
outlet with filter inlet
1000 mm (40")
Filter outlet with
CarboQC sample inlet
100 mm (4")
CarboQC sample outlet with
T-hose connector to be led into
the waste vessel
**)
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 23/78
6 Putting CarboQC into Operation
C26IB001EN-B 23
To connect the PFD Filling Device to the compressed gas supp ly
1. Connect the compressed gas supply to the inlet (2b) of the PFD Filling
Device with a nylon hose 4x6 mm of the required length.
2. Connect another nylon hose 4x6 mm to the exhaust (2a) of the PFD FillingDevice.
3. Connect the other end to a T-hose connector 6 mm (supplied with the PFD
Filling Device) and lead the hose into the waste vessel (omission to connect
the T-hose connector will cause the hose to jump when releasing the
pressure from the PFD Filling Device).
To connect the PFD Filling Device with CarboQC
1. Connect a polyurethane hose 2.5 x 4 mm of approx. 1000 mm (40") length to
the sample tube outlet (1) of the PFD Filling Device and to the inlet of thefilter (4).
Mount the filter in the correct flow direction (indicated on the filter with anarrow).
2. Connect a polyurethane hose 2.5 x 4 mm of approx. 100 mm (4") length to
the outlet of the filter and the sample inlet (5) of CarboQC.
3. Connect a polyurethane hose 2.5 x 4 mm to the sample outlet (6) of
CarboQC.
4. Connect the T-hose connector 4 mm (supplied with CarboQC) to the otherend (7) and put it into the waste vessel.
5. When all hoses are connected properly, activate the compressed gas supply.
6.2 Switching the Instrument On and Off
WARNING
Serious injuries are possible due to high voltage at parts of the instrument/
measuring system unless the following instructions are followed:
• Make sure that the non-fused earth conductor of the power cord isconnected to earth.
• For proper insulation to the mains only use the original power adapterwhich is supplied with the instrument.
• The power adapter is not suitable for operation under wet at-line oroutdoor conditions! Do not connect the power adapter of CarboQC to themains under wet ambient conditions.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 24/78
6 Putting CarboQC into Operation
24 C26IB001EN-B
1. Connect the power adapter to the supply inlet of CarboQC and the mains.
TIP When operating under laboratory conditions, the power adapter can be
permanently connected to CarboQC.
Fig. 6 - 2 Supply inlet and on/off switch
2. To switch CarboQC on, use the on/off switch.
3. If the display does not immediately light up, press the on/off switch for 10
seconds.
4. To switch CarboQC off, use the on/off switch.
If CarboQC is operated from the battery, it switches off automatically after 10
minutes if not used.
NOTICE
Before connecting the power adapter to the mains and switching the
instrument on, make sure that the correct line voltage is available (AC 100 to
240 V, 50 to 60 Hz). If large voltage fluctuations are to be expected, werecommend using a constant voltage source (UPS).
NOTICE
When connecting the plugs to CarboQC, tighten them with your fingers solely!
Do not use any tools for tightening!
Power supplyinlet
To themains
Power adapter On/off switch
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 25/78
7 Instrument Settings
C26IB001EN-B 25
7 Instrument Settings
7.1 Menu Access
A password has to be entered to access the menu. The default password is
"0000".
1. Press <Menu>, <>, select "menu-access" and press <Enter>.
2. To enter the password, enter the characters individually using the <> and
<> keys, and confirm each setting with <Enter>.
3. After finishing all the required entries or changes in the menu, exit the menu
mode by pressing <ESC> up to 3 times.
7.2 Sett ing Date and Time
1. Set date and time in the menu "instrument > date & time".
2. Select the formats "mmddyy" or "ddmmyy" for the date and "24h" or "am/
pm" for the time.
7.3 Set ting the Uni ts
• Different units for the temperature (°C or °F), pressure (bar or psia =
"absolute psi"), and CO2 (g/L or another unit) can be selected in the menu
"instrument > unit" .
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 26/78
8 Measurement
26 C26IB001EN-B
8 Measurement
8.1 Preconditions
CarboQC determines the amount of dissolved carbon dioxide in a beverage. To
get a correct measurement, the instrument has to be in good working condition
(see Chapter 9.2).
In addition, the sample package has to be in equilibrium. As carbon dioxide in a
package can be found in the liquid phase as well as in the head space,
equilibration is essential for repeatable and well defined measurement results.
Equilibration can be achieved by gently shaking the bottle horizontally about 15
times back and forth. This will ensure that the net transfer of CO 2 between liquid
and head space is zero, in short, in equilibrium.
8.2 Preparing CarboQC
CarboQC is equipped with 7 standard methods to account for the different
solubility of carbon dioxide in different beverages (see Chapter 14.1.1). The
selected method is displayed in the top left corner of the display and has to be
selected manually prior to a measurement by pressing the <Method> key,
selecting the desired method and confirming with <Enter>.
If necessary, press <Method> on CarboQC to select the appropriate method for
the sample to be measured from 7 standard methods or 17 customer-specific
methods. For customer-specific methods, see also Chapter 14.
The 7 standard methods are:
The standard methods do not compensate for any temperature effect that
influences the CO2 distribution between headspace and liquid of the bottles or
cans. Therefore, they should only be applied within a small range (e.g. 3 °C/5 °F
around reference temperature) of package temperatures.
Method Sample Concentration range
00 MINERAL Mineral water
02 DIET Diet drinks 0-3 °Brix
04 BEER Beer 10-14 °Plato
06 ST.BEER Strong beer >14 °Plato
08 COLA Sugar-based soft drinks 8-12 °Brix
10 SPARKL. Sparkling wine 9-13% vol. alc.
12 MID Sugar-based soft drinks 3-8 °Brix
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 27/78
8 Measurement
C26IB001EN-B 27
Compensation of the headspace-related temperature effect and therefore
measurements in a wider range of package temperatures can be achieved with
the 17 customer-specific methods once they are properly set up. For more
information, see Chapter 12.2.5.
8.3 Transferring the Sample into the Measuring
Chamber
Sample transfer always carries the risk of loss of carbon dioxide or introduction of
other dissolved gases. Thus, sample transfer is best performed in combination
with a filling system that avoids these risks.
The PFD Filling Device is equipped with a seal to close off a small area of the
package, bottle or can. This small area is pierced with a knife to allow introduction
of a sample tube.
Applying pressurized gas to the head space of the container allows the transfer
of sample to the CarboQC measuring chamber without the loss of CO 2.
Subsequently, the measurement of CO2 with CarboQC in combination with a PFD
Filling Device is described in detail.
TIP Piercing the package with a PFD Filling Device might result in the pierced disk
falling into the container. The filter between PFD Filling Device and CarboQC will
keep this particle from entering the measuring chamber.
8.4 Measurement Using the PFD Filling Device
8.4.1 Setting the Correct Flow
The flow through the measuring chamber should be such that 100 to 150 mL of
sample are consumed when the filling time is set to 20 seconds. Considerably
lower or higher flow rates may cause erroneous measurements due to sample
carry-over or gas bubbles in the measuring chamber.
Fig. 8 - 1 Measuring chamber and flow valve
Flow valve
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 28/78
8 Measurement
28 C26IB001EN-B
To adjust the flow, insert a suitable tool (e.g. screwdriver, ... ) into the cavity of the
flow valve and turn it clockwise for decreased and counterclockwise for increased
sample flow.
With a graduated vessel
1. Use a beaker or graduated cylinder to check the correct flow.
2. Insert the T-hose connector of the sample outlet hose into the beaker.
3. Select the menu "instrument > rinse" and set the rinsing time to 20
seconds.
4. Position the flow valve (see Fig. 8 - 1) at about 10% flow, i.e. close to the
minimum flow position.
5. Carry out a measurement as described in Chapter 8.4.2.
6. Check the volume of liquid consumed: It should be 100 to 150 mL.
7. If the volume is not in the required range, slightly open or close the flow valve
(see Fig. 8 - 1) and repeat steps 5 to 7 until the volume/weight is within the
required range.
With a marked bottl e
1. Make volume markings "0" and "150 mL" on a transparent 0.5 L (16 oz) or
similar size PET bottle.
Use water and a graduated cylinder or a balance to correctly place thesemarkings on the bottle.
2. Fill the bottle up to the "0" mark with deionized water and close it.
3. Insert it into the PFD Filling Device.
4. Perform steps 3 to 7 as described above.
Via weighing
1. Weigh a full beverage bottle or can and make a note of the weight.
2. Insert the package into the PFD Filling Device.
3. Perform steps 3 to 5 as described above.
4. Remove the package from the PFD Filling Device and weigh it again.
The weight should have decreased by 100 to 150 g.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 29/78
8 Measurement
C26IB001EN-B 29
8.4.2 Measurement Procedure
At CarboQC
1. Check the sample flow and adjust, if necessary (see Chapter 8.4.1).
2. Select the required method on CarboQC.
3. Select the sample number, if necessary.
At the PFD Fil ling Device
1. Move the sample tube of the PFD Filling Device to the highest position (see
the PFD Filling Device instruction manual).
2. Push the safety shield upwards completely.
3. Insert/remove the PET bottle adapter, if required.
4. Adjust the position of the piercing head, if required.
5. Make sure that the bottle/can is equilibrated (see Chapter 8.1).
6. Insert the PET bottle/can/glass bottle.
7. Push the safety shield down until the safety pin engages.
8. Hold the safety shield down and pull the activating lever to pierce the
package.
9. Lower the sample tube of the PFD Filling Device into the package.
10. If the sample tube touches the bottom of the package, lift it slightly.
11. Fasten the sample tube.
NOTICE
Make sure that the bottles/cans are in equilibrium prior to putting them into thePFD Filling Device. Equilibrium is e.g. achieved by shaking the bottle/can 15
times horizontally.
Wear safety goggles:
Wear safety goggles when handling the PFD Filling Device.
Wear ear protection:
Wear ear protection when handling the PFD Filling Device.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 30/78
8 Measurement
30 C26IB001EN-B
At CarboQC
1. Press <Start> to start the measurement.
The measuring chamber is first rinsed and filled with 100 to 150 mL of the
new sample.
2. Check the measuring chamber visually for bubble-free sample flow.
It must be free of bubbles before the actual analysis starts. If the bubbles donot disappear after approx. 15 seconds, clean the filter (see Chapter 10.3).
3. Check the pressure of the compressed gas supply. It shall be 60.5 bar
relative (877 psi).
Once the rinsing is finished, the analysis will start. The piston will move downto a first halt and then to a second halt with the stirrer permanently inoperation.
4. Wait approx. 90 seconds for the measurement result.
The measurement result will be displayed, stored in the memory andoptionally printed out. The piston will then move up, but the measuringchamber will remain closed.
TIP If a measurement was carried out with an incorrect method, the current result can
be recalculated using the correct method as long as this result is still displayed.
Press <Method>, select the correct method and press <Enter>. The recalculated
result will be shown on the display. To store and optionally print out this
recalculated result, press <Print>.
At the PFD Fil ling Device
1. Move the sample tube up to the highest position and fasten it.
2. Push the activating lever to de-energize the system.
3. Move the safety shield up to the highest position and remove the package.
4. If the package sticks to the piercer, rotate it counterclockwise and pull it down
simultaneously.
NOTICE
If the sample has a high content of particles (e.g. pulp), the sieve of the filter
tends to clog. This will lower the pressure in the measuring chamber. Check
the pressure indicated at CarboQC during sample filling. When using a PFD
Filling Device, the pressure should not go below 5 bar/70 psia. If the sieve is
completely clogged, bubbles and foam form in the hose between filter and
measuring chamber and in the measuring chamber itself. Clean the filter as
described in Chapter 10.3.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 31/78
8 Measurement
C26IB001EN-B 31
5. Optionally, check with a balance whether the correct amount of sample was
taken from the bottle or can.
Now CarboQC and the PFD Filling Device are ready for the nextmeasurement.
6. If no further measurements are performed, clean the system (see Chapter
10).
NOTICE
If the instrument is not in use for more than 30 minutes after the last sample
measurement, rinse the instrument with deionized water.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 32/78
8 Measurement
32 C26IB001EN-B
Measurement Procedure
Equilibrate the package.
Check the sample flow.
Select the required method on CarboQC and the sample number, if
necessary.
Move the sample tube to the highest position.
Push the safety shield upwards completely.
Insert/remove the PET bottle adapter, if required.
Adjust the position of the piercing head, if required.
Insert the PET bottle/can/glass bottle.
Push the safety shield down until the safety pin engages. Hold it down.
Pull the activating lever.
Lower the sample tube into the package.
Press <Start> on CarboQC.
Wait approx. 90 seconds for the measurement result and optional printout.
Move the sample tube up and push back the activating lever.
Move the safety shield up and remove the package.
Perform the next measurement or clean the system.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 33/78
8 Measurement
C26IB001EN-B 33
8.5 Checking for a Correct Measurement Procedure
and Avoiding Errors
Zero point checks
Regular zero point checks of CarboQC with deionized water are recommended
to survey faultless functionality of the instrument. The water check with CarboQC,
when carried out correctly with deionized water, is the precondition for a correct
measurement result. The result has to be 0± 0.02 vol (0± 0.03 g/L).
If the zero point check and adjustment are carried out with water that contains
CO2, the consequence will be an unjustified offset. Consequences are systematic
errors of results.
Sample equilibration
Only transfer equilibrated sample to the measuring chamber of CarboQC. If
samples are not equilibrated (shaken) prior to analysis, the consequence might
be erroneous results or poor repeatabilities.
Correct measurement method
Check whether the correct measuring method is chosen on CarboQC. A wrong
method will lead to a wrong result.
Correct sample flow
Check the correct flow. The volume consumption per measurement is
recommended to be 120 to 150 mL, and for identical samples at least 100 mL.
Too low sample volume consumption causes sample carry-over.
Correct filling pressure
The filling pressure displayed on the CarboQC display has to be 5.5 to 6 bar
relative (79 psi to 87 psi) and the measuring chamber has to be filled without
bubbles. Too low sample pressures during sample filling causes bubbles in the
sample chamber. As a consequence, the starting volume will consists partially of
gas bubbles thus yielding incorrect results.
Measuring chamber filled without bubbles
Bubbles in the measuring chamber can falsify the result. The most common
cause for bubbles is a clogged filter. To clean the filter, see Chapter 10.3.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 34/78
9 Checking the System
34 C26IB001EN-B
9 Checking the System
9.1 Influencing Factors
Mechanical functionality of CarboQC is essential for a correct result. CarboQC
has three parameters that act as a precondition for a correct result.
• pressure• temperature• volume expansion of the measuring chamber
9.2 Zero Point Check
A zero point check is necessary before starting the determinations. The zero pointcheck is a measurement with deionized (or distilled) water. The CO2 result has to
be 0±0.02 vol (0±0.03 g/L).
The deionized water must not contain any dissolved carbon dioxide and is ideally
free of ions such as calcium that could cause deposits in the measuring chamber
of CarboQC.
Sometimes, deionized water contains CO2 absorbed from the ambience and
CarboQC displays the amount of the CO2 absorbed. Either heat up and boil the
water and cover it up while cooling down or do the water check with tap water. In
case of using tap water, rinse the measuring chamber with deionized waterafterwards.
In case the measurement of deionized water does give a CO 2 reading out of
tolerance, always check the water first before inserting a pressure offset (see
Chapter 9.5) into CarboQC.
In general, the zero point check with CO2-free water is sufficient to ensure correct
functioning of the instrument in-between service intervals.
9.3 Check with Carbonated Liquid
Sometimes checkups are required with test samples at elevated carbon dioxide
contents. Usually these checkups are provided within the periodical service and
maintenance procedure. In case more frequent tests are required, continuous
surveillance can be performed on a day-to-day basis.
There are two ways to perform an instrument check with carbonated liquids.
The first is the use of a CO2 standard. This procedure shall only be carried out by
an authorized Anton Paar representative or service technician.
The second is the use of a control chart and internal standards.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 35/78
9 Checking the System
C26IB001EN-B 35
Cans of diet soft drink or other cans of a non-sucrose based, non-inverting soft
drink with > 3 vol (6 g/L) CO2 content are suitable to be used as internal
standards.
If filling cans, provide 36 cans for initial preparation of the standards plus a
sufficient number of cans for one check measurement per day over the time span
between instrument calibrations by Anton Paar service representatives. Only use
cans from the same production minute(s).
If purchasing cans, make sure that the required number of cans is from the same
batch.
To p repare in ternal standards
1. Immediately after calibration of CarboQC by an Anton Paar service
representative, draw 6 cans from the beginning, 6 cans from the middle, and
6 cans from the end section of a production run and measure them.
2. Calculate the mean values for every 6 measurements for the beginning,
middle and end sections.
The 3 mean values may only differ by ±0.02 vol (±0.04 g/L).
3. If the mean values are suitable, provide a sufficient number of cans from this
production run to be kept as internal standards.
4. If the mean values are not suitable, measure another 3 times 6 cans as
above.
5. Determine the 3 mean values of the 12 respective measurements (for each
mean value take the 6 measurements from the first selection plus the 6
measurements from the second selection).
These 3 mean values may only differ by ±0.015 vol (±0.03 g/L).
6. If the mean values are suitable, provide a sufficient number of cans from this
production run to be kept as internal standards.
7. If the mean values are not suitable, the cans are not suitable as internal
standards and cans from a different, more stable production run have to be
used.
NOTICE
Only use cans as internal standards. Do not use bottles as their CO2 content
might change.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 36/78
9 Checking the System
36 C26IB001EN-B
To draw and use the cont rol chart
1. Calculate the mean m and standard deviation sd of all 18 (or 36)
measurements.
The standard deviation shall be 0.025 vol (0.05 g/L) maximum based on theabove limitations.
2. Draw a control chart with sample number on the x-axis and CO2
concentration on the y-axis ranging from (m - 3 * sd) to (m + 3 * sd) as the
control limits.
3. Draw control limit lines, mean line and auxiliary lines as appropriate. This
can either be done on paper or using a suitable computer software.
4. Measure one can of the internal standards per production day and draw a
corresponding point in the control chart.
5. If a clear trend away from the mean becomes visible over time or the control
limit of 3 * sd is exceeded successively, call for service immediately.
9.4 Beginning of the Shift /Day
Check for the correct sample flow with
deionized water. A volume of 100 to 150 mL isrecommended.
NOK Adjust the sample flow.
See Chapter 8.4.1.
OK
Check the zero point: Fill a clean 0.5 L PET
bottle with deionized water, close the bottle and
measure it (see Chapter 8.4.2). The CO2 result
has to be 0±0.02 vol (0±0.03 g/L).
OK
NOK Rinse with glassware cleaner (pH <10) andthen with deionized water.
OK
Check the zero point with deionized water again
(see Chapter 8.4.2). The CO2 result has to be
0±0.02 vol (0±0.03 g/L).
CarboQC is ready for
routine measurements.
OK
NOK
Contact your Anton Paar representative.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 37/78
9 Checking the System
C26IB001EN-B 37
9.5 Entering a Pressure Offset
With the menu "auto p-offset" , an offset for the pressure sensor can be
determined semiautomatically. Enter a pressure offset if the CO2 result of a zero
point check with deionized water is out of specification (see Chapter 9.2).
TIP Sometimes, deionized water contains CO2 which has been absorbed from the air,
and this CO2 content is displayed. Before carrying out an automatic pressure
offset, make sure that the displayed CO2 is not caused by the water, but by the
pressure sensor. Boiling and cooling the water before measurement will eliminate
dissolved CO2.
Example:
A zero point check with deionized water yields (even after cleaning the system) a
CO2 value of 0.05 g/L. The maximum allowed deviation for deionized water is 0.03
g/L. Therefore, the pressure sensor must be adjusted:
1. Select the menu "adjustment > auto p-offset" .
In the first line, the previously measured CO2 value (0.05 g/L) is displayed. In
the second line, the pressure value is displayed (e.g. +0.020 bar).
2. Press <Enter>.
The displayed CO2 value flashes.
3. Change the CO2 value using the <> and <> keys until the value is insidespecification (0 ±0.03 g/L).
The pressure value is changing corresponding to the CO2 value.
4. When the required CO2 value is displayed, press the <Enter> key.
The newly calculated pressure offset is automatically stored in the CarboQCmemory and is used for subsequent measurements.
5. To change to the measuring window, press <ESC> three times.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 38/78
10 Cleaning the System
38 C26IB001EN-B
10 Cleaning the System
10.1 End of the Shift/Day
Rinse the measuring chamber with deionized water.
Rinse with glassware cleaner (pH <10) by pressing <Start> for approx. 5
seconds to clean the entire measuring chamber. Wait 2 minutes for this to
finish and then press <Stop>.
Rinse the measuring chamber with deionized water.
Clean sample residues from the PFD Filling Device and CarboQC with a
wet tissue. Dry with a soft dry tissue. If the PFD Filling Device is very dirty,
spray it with (warm) water, move movable parts up and down in between
spraying and then dry with a soft tissue.
NOTICE
If the instrument is not used for some time (e.g. over the weekend) or if
shipping the instrument, always make sure that the measuring chamber is
about half filled with deionized water to avoid damage of the O-rings.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 39/78
10 Cleaning the System
C26IB001EN-B 39
10.2 Weekly Maintenance
Clean the filter (see Chapter 10.3).
Rinse the measuring chamber with warm
deionized water.
Rinse the measuring chamber with glassware
cleaner (pH <10).
Rinse the measuring chamber with deionized
water.
Rinse the measuring chamber with a 60 to 90%
solution of ethanol. Do not use denatured
ethanol!
Rinse the measuring chamber with deionized
water.
Fill deionized water by pressing <Start> for
approx. 5 seconds to clean the entire measuring
chamber. Wait 2 minutes for this to finish.
Check whether gas bubbles enter the measuring
chamber.
Yes Contact your Anton
Paar representative.
No
Press <Stop> and leave deionized water in the
measuring chamber.
Clean sample residues from the PFD Filling
Device and CarboQC with a wet tissue. Dry with
a soft dry tissue. If the PFD Filling Device is very
dirty, spray it with (warm) water, move movable
parts up and down in between spraying and then
dry with a soft tissue.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 40/78
10 Cleaning the System
40 C26IB001EN-B
10.3 Cleaning the Filter
TIP Make sure that the activating lever of the PFD Filling Device is in the "off" position
before disconnecting the filter.
1. Press <Rinse> on CarboQC and wait 20 seconds for the measuring
chamber to open.
2. Disconnect the filter from the hoses and connect it in the reverse direction to
the hose coming from the PFD Filling Device.
3. Put a 0.5 L or similar size PET bottle with deionized water into the PFD
Filling Device (see Chapter 8.4.2).
4. Hold the filter outlet over a sink or waste container and fix it in this position.
5. Pull down the safety shield and pull the activating lever.
Air with sample residues will spray out of the filter.
6. Lower the sample tube to flush the filter with approx. 200 mL of deionized
water.
7. Move the sample tube up to let air flow through the filter for about one
minute.
8. Push the activating lever to the "off" position.
9. Reconnect the filter in the correct flow direction.
10.4 Cleaning the Measuring Chamber
10.4.1 Required Devices
• To clean the measuring chamber of CarboQC, fill a 0.5 L or similar size PET
bottle with deionized water or glassware cleaner (pH <10) which removes
protein.
• Insert this bottle into the PFD Filling Device the same way as a sample to be
measured.
10.4.2 Rinsing
1. Press the <Rinse> key for approx. 5 seconds.
The measuring chamber is rinsed for the set rinse time.
2. Always rinse with deionized water after rinsing with glassware cleaner.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 41/78
10 Cleaning the System
C26IB001EN-B 41
3. Leave only deionized water in the measuring chamber for extended periods
of time.
10.4.3 Cleaning the Entire Measuring Chamber Volume
1. Press the <Start> key for approx. 5 sec.
After rinsing, the piston will descend to the maximum measuring chambervolume expansion with the stirrer turned on. This will clean the entirechamber volume.
If deionized water is used, the pressure in the measuring chamber willbecome very low, far below ambient pressure. Air bubbles will enter themeasuring chamber if it is not completely tight. Therefore, this function canalso be used to detect leaks caused by damaged O-rings or valves.
2. Press <Stop> to exit this mode.
3. Always rinse with deionized water after rinsing with glassware cleaner.
4. Leave only deionized water in the measuring chamber for extended periods
of time.
10.5 Cleaning the Instrument Housing and Display
If sample was spilled onto the housing of CarboQC, always wipe it off
immediately.
• To clean the instrument housing and/or the keypad, use a soft tissue and
(warm) water.
• Dry with a soft, dry tissue.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 42/78
11 Measuring At-line
42 C26IB001EN-B
11 Measuring At-line
11.1 Using CarboQC with Protective Casing
11.1.1 Overview
The protective casing for CarboQC is especially designed for use in harsh
process environments. The robust polyurethane casing protects CarboQCagainst mechanical impact and splash water. Special foam parts inside the casing
dampen mechanical shocks on the instrument if the instrument falls down or
bumps into an obstacle.
The protective casing is mounted and dismounted fast and easily. The most
important standard tasks during a measurement can be carried out using the keys
on the protective casing:
• switching the instrument on and off • selecting a method• starting and stopping a measurement
• reading the display
A stainless steel handle allows easy carrying of the casing with the CarboQC
inside. Also opening and closing is done comfortably by draw latches.
Technical data
WARNING
Serious injuries through high voltage at instrument parts possible.
• The power supply is not suitable for operation under wet at-lineconditions.
• Charge the battery of CarboQC only under dry indoor conditions.• If CarboQC is not used for extended periods of time (> 2h), store the
instrument under dry indoor conditions only.
WARNING
Operation under pressure. Serious injuries and damage of goods possible if
the maximum allowable pressure is exceeded.
• Do not operate CarboQC with sample pressures higher than 10 bar (145psi) absolute.
Weight 1.95 kg (4.30 lbs)
Dimensions L x W x H 240 x 165 x 395 mm (9.45 x 6.5 x 15.55 in)
Ambient temperature 0 °C to 40 °C (32 °F to 104 °F)
Air humidity 10 to 95 % relative humidity
Protection class IP64
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 43/78
11 Measuring At-line
C26IB001EN-B 43
Front view and rear view
Fig. 11 - 1 Front view and rear view
1 ... Handle2 ... Viewing window3 ... <Start> key4 ... <Stop> key5 ... Inlet of the protective casing6 ... Outlet of the protective casing
Side views
Fig. 11 - 2 Side views
1 ... Draw latches with secondary catch2 ... On/off key
Material housing Polyurethane
Wetted parts Polyurethane (hoses), brass (connections)
1
2
34
5 6
1
2
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 44/78
11 Measuring At-line
44 C26IB001EN-B
11.1.2 Mounting the Protective Casing
Fig. 11 - 3 Opening the protective casing
1. To open the casing, press the secondary catch (1) on the draw latches on
both sides of the casing.
2. Loosen the upper part of the draw latches (2).
3. Now the draw latches are open (3) and you can remove the upper part of the
casing.
Fig. 11 - 4 Connecting the hoses to CarboQC
4. Put the ends of the hoses out of the casing and leave the little bag (4) with
silica gel in the casing.
5. Put the CarboQC inside the housing so that the on/off switch is positioned at
the correct place.
6. Connect the hose coming from the outlet opening (5) to the CarboQC
sample outlet.
7. Connect the hose coming from the inlet opening (6) to the CarboQC sample
inlet.
8. Close the casing with the draw latches.
Now CarboQC inside the protective casing is ready for use at the line.
1 2 3
4
5
6
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 45/78
11 Measuring At-line
C26IB001EN-B 45
11.1.3 Connecting CarboQC in the Protective Casing to the
Product ion Line or Tank
CarboQC can be connected to the line via filter or filter with male hose filling
nozzle (at-line adapter, Mat. No. 18829). This at-line adapter allows for thickerdiameter hoses between adapter and line.
Fig. 11 - 5 Optional at-line adapter (filter with male hose filling nozzle, Mat. No.
18829)
TIP For at-line measurements, keep the hoses between the line and the filter as short
as possible to reduce pressure drops and minimize the risk of the sample
outgassing. The correct flow direction is marked with an arrow.
Fig. 11 - 6 Connecting the instrument to the line
1. Connect the outlet of the line/tank to the inlet of the filter (standard filter or at-
line adapter).
Line
Samplevalve
Filter
from theoutlet
to thewaste
to theinlet
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 46/78
11 Measuring At-line
46 C26IB001EN-B
2. Connect the outlet of the filter to the inlet of the protective casing using a
polyurethane hose 2.5 x 4 mm.
3. Connect a polyurethane hose transparent 2.5 x 4 mm to the outlet of the
protective casing.
4. Place the other end of this hose in the waste vessel.
Setting the correct flow
To set the correct flow, use the PFD Filling Device and proceed as described in
Chapter 8.4.1.
11.1.4 Measurement
TIP The minimum pressure for at-line operation of CarboQC has to be 0.5 bar (7.2 psi)
above saturation pressure of the dissolved gases. Proper functioning is best
determined by practical tests.
1. To select the method, press the <Stop> key on the protective casing and
hold it pressed until the desired method is displayed. Then release the key.
The method is selected and CarboQC switches back to measurement mode.
2. Open the sample valve of the line or tank.
3. Start the measurement by pressing the <Start> key on the protective casing.
4. Watch the measuring chamber.
After approx. 15 seconds rinsing time it should be free of bubbles.
TIP If the system pressure of the line is too low, the measuring chamber will not be
free of bubbles in time. Fill the measuring chamber of CarboQC manually byrinsing it before pressing <Start>.
• Press the <Start> key and <Stop> key simultaneously and let a minimum of
100 mL flow through the measuring chamber, then press <Stop> to stop the
rinsing procedure.
• Make sure the measuring chamber is free of gas bubbles and press <Start>.
The measuring result is automatically displayed after approx. 90 secondsand stored in the memory of CarboQC (max. 99 measuring results).
NOTICE
Do not disconnect any hoses from the protective casing during the
measurement.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 47/78
11 Measuring At-line
C26IB001EN-B 47
11.1.5 Cleaning the Measuring Chamber
By connecting CarboQC to a tap with process water
• Attach the inlet valve of the protective casing to a tap with process water(NOT with tap water as the limestone might harm the inlet and outlet valves).
• Then press simultaneously <Start> and <Stop> on the protective casing to
open the valves and rinse the measuring chamber.
• Press <Stop> to stop the rinsing procedure.
With a peristaltic pump
• Connect the outlet tubing of a peristaltic pump to the filter or at-line adapter
that is connected to the inlet of the protective casing.
• Then press simultaneously <Start> and <Stop> on the protective casing to
open the valves and rinse the measuring chamber.
• Then operate the pump as required and rinse the measuring chamber with
detergent and/or deionized water. If detergent is used, the measuring
chamber has to be rinsed carefully with process water or deionized water
afterwards.
• Press <Stop> on the protective casing to stop the rinsing procedure.
By connecting CarboQC to the process line
• If the process line is cleaned with water or detergent, leave CarboQC inside
the protective casing connected.
• Then press simultaneously <Start> and <Stop> on the protective casing to
open the valves and rinse the measuring chamber.
• Press <Stop> on the protective casing to stop the rinsing procedure.
• If detergent is used, the measuring chamber has to be rinsed carefully with
process water or deionized water afterwards.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 48/78
11 Measuring At-line
48 C26IB001EN-B
11.2 Using CarboQC without Protective Casing
11.2.1 Connecting CarboQC to the Production Line or Tank
TIP If CarboQC is frequently used for at-line measurements at different locations, we
recommend equipping CarboQC with a protective boot (Mat. No. 90079) for
protection of the instrument during transport.
CarboQC can be connected to the line via filter or filter with male hose filling
nozzle (at-line adapter, Mat. No. 18829). This at-line adapter allows for thicker
diameter hoses between adapter and line (see Fig. 11 - 5).
TIP For at-line measurements, keep the hoses between the line and the filter as short
as possible to reduce pressure drops and minimize the risk of the sample
outgassing. The correct flow direction is marked with an arrow.
1. See also Fig. 11 - 6: Connect the outlet of the line/tank to the filter inlet or the
inlet of the at-line adapter or standard filter.
2. Connect the outlet of the filter to the inlet of CarboQC using a polyurethane
hose 2.5 x 4 mm.
3. Connect a polyurethane hose transparent 2.5 x 4 mm to the outlet of
CarboQC.
4. Place the other end of this hose in the waste vessel.
WARNING
Serious injuries through high voltage at instrument parts possible.
• The power supply is not suitable for operation under wet at-lineconditions.
• Charge the battery of CarboQC only under dry indoor conditions.• If CarboQC is not used for extended periods of time (> 2h), store the
instrument under dry indoor conditions only.
WARNING
Operation under pressure. Serious injuries and damage of goods possible if
the maximum allowable pressure is exceeded.• Do not operate CarboQC with sample pressures higher than 10 bar (145
psi) absolute.
NOTICE
Do not place CarboQC on a wet or humid bench because humidity entering
the housing could damage the instrument.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 49/78
11 Measuring At-line
C26IB001EN-B 49
TIP A carrying strap (Mat. No. 16012) is optionally available for safe and convenient
operation under at-line conditions.
Setting the correct flow
To set the correct flow, use the PFD Filling Device and proceed as described in
Chapter 8.4.1.
11.2.2 Measurement
TIPThe minimum pressure for at-line operation of CarboQC has to be 0.5 bar (7.2 psi)above saturation pressure of the dissolved gases. Proper functioning is best
determined by practical tests.
1. Press the <Method> key on CarboQC and select the required method.
2. If necessary, change the sample number by pressing the <Sample> key.
3. Open the sample valve of the line or tank.
4. Start the measurement by pressing the <Start> key on CarboQC.
5. Watch the measuring chamber.
After approx. 15 seconds rinsing time it should be free of bubbles.
TIP If the system pressure of the line is too low, the measuring chamber will not be
free of bubbles in time. Fill the measuring chamber of CarboQC manually by
pressing the <Rinse> key before pressing <Start>. After the measuring chamber
opens,
• press the <Rinse> key periodically to turn the stirrer on and off,
• make sure that a minimum of 100 mL flows through the measuring
chamber,
• make sure the measuring chamber becomes free of gas bubbles.The measuring result is automatically displayed after approx. 90 secondsand stored in the memory of CarboQC (max. 99 measuring results).
NOTICE
Do not disconnect any hoses from CarboQC during the measurement.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 50/78
11 Measuring At-line
50 C26IB001EN-B
11.2.3 Cleaning the Measuring Chamber
By connecting CarboQC to a tap with process water
• Attach the inlet valve of CarboQC to a tap with process water (NOT with tapwater as the limestone might harm the inlet and outlet valves).
• Then hit the <Rinse> key on the instrument to open the valves and rinse the
measuring chamber.
With a peristaltic pump
• Connect the outlet tubing of a peristaltic pump to the filter or at-line adapter
that is connected to the inlet valve of CarboQC.
• Hit the <Rinse> key on CarboQC.
• Then operate the pump as required and rinse the measuring chamber with
detergent and/or deionized water. If detergent is used, the measuring
chamber has to be rinsed carefully with process water or deionized water
afterwards.
By connecting CarboQC to the process line
• If the process line is cleaned with water or detergent, leave CarboQC
connected and hit the <Rinse> key.
• If detergent is used, the measuring chamber has to be rinsed carefully withprocess water or deionized water afterwards.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 51/78
12 Menu Operation
C26IB001EN-B 51
12 Menu Operation
12.1 Using the Keys on the Keypad
If the measuring window is displayed, the functions <Sample>, <Menu>, <Print>,
<Method>, <Rinse>, <Start> and <Stop> are active.
In the menu mode (after pressing the <Menu> key) the functions <ESC>, <>,
<> and <Enter> are active.
The key functions in the menu mode accomplish the following tasks:
12.2 Menu Description and Function
12.2.1 Displaying the Measurement Data
The stored measurement data can be displayed. To display the next or previous
data set, press the <> or <> key.
Fig. 12 - 1 Displaying the measurement data
ESC • Return to the measuring mode• Move to the next higher menu level
• Ignore the last entry and return to the next higher menulevel• Ignore the last entry and return to the previous digit or
character
and • Move to a menu item• Decrease or increase a digit or a character
Enter • Select a menu item• Confirm an input or a query• Accept the last entry and return to the next higher menu
level• Accept the last entry and move to the next digit or
character
m e n u
d a t a m e m o r y
m e n u a c c e s s
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 52/78
12 Menu Operation
52 C26IB001EN-B
12.2.2 Accessing the Menu
Press the <Menu> key on CarboQC to initiate the menu access.
Fig. 12 - 2 Main menu (scrolls when arrow is moved down)
For the complete menu tree, see the last pages of this instruction manual.
12.2.3 "adjustment" Submenu
Fig. 12 - 3 Menu/adjustment
menu-access
password A password has to be entered to
access the menu. The default
password is: 0000.
If you use a different password, press
the <Enter> key. Enter the characters
using the <> and <> keys, and
confirm each setting with <Enter>.
m e n u
a d j u s t m e n t
i n s t r u m e n t
m e t h o d s
m e n u - a d j u s t m e n t
c o m p e n s a t i o n
a u t o p - o f f s e t
compensation
comp air
on To compensate the influence of
dissolved air. CarboQC automatically
performs two volume expansions.
of f To deactivate the air compensation.
One volume expansion has to be set.
vexp To set the volume expansion
between 2 and 40%. Default: 10%.
Only valid when comp air is set to off.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 53/78
12 Menu Operation
C26IB001EN-B 53
12.2.4 "instrument" Submenu
Fig. 12 - 4 Menu/instrument
auto p-offset Offset for the absolute pressure
sensor (values between 0 and ±0.5
bar (0 and ±7 psi)). For details, see
Chapter 9.5.
C: x.xxx g/L (or respective selected unit)
p: x.xxx bar (psia)
m e n u - i n s t r u m e n t
r i n s e
u n i t
s e r i a l c o m
rinse
time: 03 to 200 seconds for rinsing the
measuring chamber. Default: 20
seconds.
unit To select the units of the displayed
CO2 content, pressure andtemperature as well as the number of
places after the decimal point.
t To select °C or °F of the temperature.
p To select bar or psia (=absolute psi)
of pressure.
co2 To select g/L or an arbitrary unit of
CO2 content.
co2dec To select 2 or 3 places after the
decimal point for the displayed CO2
result, and 0 or 1 places after thedecimal point for the displayed air
content, respectively.
serial com The settings in this menu are valid for
both the printer and PC interfaces.
baud
1200
2400
4800
9600
19200
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 54/78
12 Menu Operation
54 C26IB001EN-B
format
E71 Even parity, 7 data bits, 1 stop bit
N81 No parity, 8 data bits, 1 stop bit
O71 Odd parity, 7 data bits, 1 stop bit
password
new To define a new password for the
menu access. 4 characters/digits are
available. Default: 0000.
date & time Date and time and the corresponding
formats can be set.
date Two different formats for the date are
available (default: ddmmyy)
ddmmyy Day.month.year, e.g. 13.10.03
mmddyy Month-day-year, e.g. 10-13-03 year To set the current year.
month To set the current month.
day To set the current day.
time Two formats for time are available.
24hrs E.g. 17:08
am/pm E.g. 05:10 pm
hour To set the hour.
min To set the minute.
battery
service
on To perform a battery service
procedure. With the power adapter
connected to the CarboQC supply
inlet and the mains, the battery is
discharged and charged again. This
takes approx. 12 hours.
of f Battery service procedure inactive.
device info All the values given under device info
are read only.
print Output of the device info to printer or
PC via the corresponding interface.
of f
printer
PC
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 55/78
12 Menu Operation
C26IB001EN-B 55
12.2.5 "methods" Submenu
In this submenu a list of customer adaptable methods is displayed. The names as
specified here are only valid if no changes have been made. Once the method-
specific names have been changed, the changed names will appear in the list.The corresponding coefficients for these methods will only be explained for the
first method.
The last 10 methods (CM01PROD - CM10PROD) allow to enter a solubility
coefficient ks , too. They can therefore be adapted to customer-specific
beverages which exceed the predefined methods range in °Brix, °Plato or % alc.
vol. of beverages.
Fig. 12 - 5 Menu/customer
m e n u - m e t h o d s
0 1 M I N E R A L c
0 3 D I E T c c c c
0 5 B E E R c c c c
01 MINERALc Customer-specific method for mineral
water.
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit An arbitrary unit can be entered (see
Chapter 14.2.2). Factory default: g/L.
fu Division factor which is used to convert
the determined result (in g/L) into a
customer-specific unit, e.g. enter 1.976 to
convert g/L to vol. For calculating
customer-specific polynomials.
k0 Coefficients k0 - k4 convert the CO2
results based on the built-in solubility
formula to customer-specific results
based on customer-specific tables/
formulae. These parameters are also
used to compensate for the headspace
effect in bottles and cans which causes a
lower amount of dissolved CO2 at higher
temperatures (see Chapter 14.2.1).
k1
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 56/78
12 Menu Operation
56 C26IB001EN-B
k2
k3
k4
k5 Reference temperature [°C].
k6 Coefficients k6 - k7 are used to transform
the air or nitrogen content as determined
into related properties (see Chapter
14.3).
k7
03 DIETcccc Customer-specific method for diet
products (0 to 3°Brix).
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit For the unit, an arbitrary unit can be
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
05 BEERcccc Customer-specific method for beer (10 to
14°Plato).
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit For the unit, an arbitrary unit can be
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
07 ST. BEERc Customer-specific method for strong beer
(>14°Plato).
name To define a customer-specific name for
this method.
sel
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 57/78
12 Menu Operation
C26IB001EN-B 57
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit For the unit, an arbitrary unit can beentered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
09 COLAcccc Customer-specific method for sugar
based beverages (8 to 12°Brix).
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit For the unit, an arbitrary unit can be
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
11 SPARKLcc Customer-specific method for sparkling
wine (9 to 13% vol. alc.).
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit For the unit, an arbitrary unit can be
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
13 MIDccccc Customer-specific method for sugar-
based soft drinks (3 to 8 °Brix).
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 58/78
12 Menu Operation
58 C26IB001EN-B
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit An arbitrary unit with 6 digits maximum
(e.g. mg/cm3) can be selected and
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
17 CM01PROD Customer-specific method including
selectable solubility of CO2.
name To define a customer-specific name for
this method.
sel
man sel The method is available in the method
list.
disable The method is not available in the
method list.
unit An arbitrary unit with 6 digits maximum
(e.g. mg/cm3) can be selected and
entered (see Chapter 14.2.2). Factory
default: g/L.
fu See "MINERALc".
k0 See "MINERALc".
....
k7
ks Enter the CO2 gas absorption coefficient
in g/(L * bar) of CO2 at 5 bar and 20 °C
(e.g. 1.61 g/(L * bar) for a 5±2 °Brix soft
drink).
ka
....
kd
18 CM02PROD See "CM01PROD".
...
26 CM10PROD See "CM01PROD".
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 59/78
12 Menu Operation
C26IB001EN-B 59
12.2.6 "data memory" Submenu
Fig. 12 - 6 Menu/data memory
m e n u - d a t a m e m o r y
b r o w s e m e m o r y
c l e a r m e m o r y
browse memory To display the stored measurement
data.
Browse through the memory using
the <> and <> keys, starting with
the most recent. To indicate that the
memory browser is activated, the
name of the activated method and
"memory" are alternately displayed in
the upper left corner. A stored
measurement contains the following
information:
• Name of the selected method• Sample No.• CO2 result
• Dissolved air • Measuring temperature or
package pressure99 measurements can be stored. The
latest entry overwrites the oldest
when the memory is full.
clear memory
clear
yes To clear all measuring results in the
memory of CarboQC.
no
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 60/78
12 Menu Operation
60 C26IB001EN-B
12.2.7 "service" Submenu
In this menu, information about the device can be displayed. None of the values
can be changed.
device
SerNo Serial number of the instrument.
FW Firmware version of the instrument.
v-k Slope parameter: counts per 1 %
increase of the volume of the
measuring chamber.
v-I0 Zero volume expansion parameter:
counts at zero volume expansion, i.e.
when measuring chamber valves
close.
v-h Slope parameter for the
determination of the pressure drop at
zero volume expansion for zero
adjustment.
cycle Total number of measurement cycles
performed.
lb0 Solubility parameters for air/nitrogen.
lb1
lb2
ccor Internal correction factor.
t dis Initial or final (internal reference
temperature).
offset
t Offset for the temperature sensor
(values between 0 and ±5 °C (0 and
±9 °F)).
p Offset for the absolute pressure
sensor (values between 0 and ±0.5
bar (0 and ±7 psi)).
gain
t Gain factor for the temperature
sensor (values between 0 and 2).
p Gain factor for the pressure sensor
(values between 0 and 2).
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 61/78
13 Technical Background
C26IB001EN-B 61
13 Technical Background
The working principle of CarboQC is based on the Multiple Volume Expansion
Method and allows the determination of the true CO2 content.
Fig. 13 - 1 shows the principle of the Multiple Volume Expansion method for
eliminating the influence of dissolved air or nitrogen on the measuring result: The
CO2 content is measured at two different volume expansions of the measuring
chamber, i.e. at 10 and 30%. If the two results are identical, there is no dissolved
air or nitrogen present and therefore no correction is required. Dissolved air or
nitrogen makes the second result lower than the first. The difference between the
two results is used to calculate a correction which completely eliminates the
influence of dissolved air or nitrogen on the measuring result.
Fig. 13 - 1 Compensation of dissolved air
CO2 with air
CO2 without air
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 62/78
14 CarboQC Methods
62 C26IB001EN-B
14 CarboQC Methods
14.1 Available Methods with CarboQC
14.1.1 Standard Methods
The 7 standard methods do not allow to change any parameter at all. It can be
seen from Chapter 8.2 which method is recommended for which beverage
sample. These methods do not provide any means for compensating temperature
effects due to the headspace volume of beverage packages. They are therefore
suitable for determining the true dissolved carbon dioxide content at a given
package temperature.
14.1.2 Customer-Specific Methods
CarboQC provides 17 customer-specific methods. 7 methods, ending with "c",
use solubility coefficients identical to the standard methods, and 10 customer-
specific methods CMxxPROD also allow the entry of the solubility coefficient.
The 7 user-specific methods are named identically to the standard methods with
additional letters "cc" to obtain eight digits altogether (DIETcccc; SPARKLcc and
so on). These 7 methods allow modification of the constants k0 to k7, but these six
methods do not allow access to the solubility coefficient for CO2 in the respective
sample. The solubility coefficients in these user-specific methods are identical tothe ones in the respective standard methods. The coefficients k0 to k7 can be used
to compensate headspace related package temperature effects and to adapt the
measurement results to those based on customer-specific solubility tables or
formulae.
There are also 10 CMxxPROD methods; xx = 01, ..., 10. These 10 customer-
specific methods allow the modification of the constants k0 to k7, but also the
alteration of the solubility coefficient ks of CO2 in the sample.
14.2 Establishing Customer-Specific Formulae
14.2.1 Converting the CO2 Result into a Customer-Specific
Result
In the customer-specific methods the calculated g/L results can be converted for
user-specific CO2 tables or formulae. This conversion can also correct the
temperature-dependent changes of the dissolved CO2 content resulting from the
headspace to liquid volume ratio of the bottles or cans.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 63/78
14 CarboQC Methods
C26IB001EN-B 63
This equation is used for custom-tailoring a CO2 result from the original CO2
content c as determined by CarboQC.
Ccust is the customer-specific CO2 result. It is the modified carbon dioxide content
that is calculated and displayed when customer-specific coefficients are used.
k0 will always be added to the result c. The factory setting of k0 = 0.
k1 is a multiplication factor. The factory setting of k1 = 1.
k5 is a reference temperature that needs to be keyed in if the result is to be related
to a specific temperature. The factory setting of k5 is zero, which means that k5
equals 0 °C, regardless of the temperature units selected in the menu.
A simple way of applying a reasonable temperature correction for the headspaceinfluence is to use the customer-specific methods in CarboQC, rename them
properly and set k5 equal to the average measuring temperature, and for k2 a
value of 0.003 (typical range: 0.002 … 0.004) to get the temperature
compensation to take effect.
k3 and k4 can be used for better fine tuning of the final customer-specific result.
The factory setting is 0.
Example:
If the result given by CarboQC is 3 g/L and the desired or expected result is 3.3g/L, there are two ways CarboQC can be programmed in order to transfer the
CarboQC result into the customer-specific result:
(1) k0 is set to 0.3
The value ccust will be 0.3 g/L higher than the CarboQC result. This will also affect
the zero determination with deionized water. A k0 of 0.3 will cause an offset as all
results will be 0.3 higher than the determined value: instead of 0 the displayed
result will be 0.3; instead of 2 the displayed result will be 2.3 and so forth.
Therefore setting a positive k0 offset is a reasonable means to adapt the CarboQC
results to readings of previous methods which were offset by a fairly constant aircontent of the samples.
Ccust.......................... Customer-specific CO2 result [g/L]
k0, k1, k2, k3, k4 ........ Customer-specific coefficients
k5 ............................. Reference temperature [°C]
c .............................. Original CO2 result [g/L]
t ............................... Measured temperature [°C]
Ccust k0 k1*c k2*c* t k5 – k3*c* t k5 – 2++ k4*c
2+ +=
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 64/78
14 CarboQC Methods
64 C26IB001EN-B
(2) k1 is set to 1.1
The CarboQC result will be multiplied by 1.1. Instead of 3 the result will be 3.3 and
instead of 2 the result will be 2.2. If the zero point adjustment works correctly, the
zero reading will be close to zero and multiplication by 1.1 will not affect the zero
reading noticeably. As an example, a factor of approximately 1.1 is often
necessary if the customer-specific solubility table applied is actually a CO2
solubility table for water, but it is applied to 10 °Brix soft drinks.
14.2.2 Converting the CO2 Concentration into a Customer-
Specific Unit
For all customer-specific methods, the measured CO2 concentration can be
converted from g/L into a customer-specific unit.
Proceed as follows:
1. In the menu "methods" , select the desired customer-specific method.
2. Select the menu item "unit" and enter a name for the desired unit (maximum
length of 6 characters).
3. In the menu "methods" , change the coefficient "k1" so that the desired
conversion is applied. E.g. setting k1 = 2, results in multiplying the result by 2.
4. Change to the measuring window and select the changed method using the
<Method> key.
The newly set unit is displayed in the line for the CO2 content. When
measuring, the conversion is carried out according to the entered coefficientand the calculated result is displayed.
5. The factor "fu" is a division factor which divides a customer-specific
polynomial. E.g. using fu = 1.976 leads to converting the measurement result
from g/L into vol.
14.2.3 Using a Customer-Specific Solubility Coefficient
The 10 CMxxPROD methods allow entry of k1 to k7 and, in addition, the solubility
coefficient ks in g/(L * bar) at a temperature of 20 °C. These 10 methods need to
be used when a sample does not match the ranges as given in Chapter 8.2; e.g.
1.61 g/(L * bar) for a 5 ±2 °Brix soft drink.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 65/78
14 CarboQC Methods
C26IB001EN-B 65
14.3 Transformation of the Determined Air Content
The following formula serves two purposes:
1. It allows to correct the ppm result with a different slope k6.2. It allows to compensate for the temperature dependent gas distribution in the
beverage container using k5, k7. The term in brackets is set to 1 if k7 0.
This formula is used if the "air ppm" result needs to be corrected or the amount of
gas in the head space needs to be taken into consideration. ctrans is the air content
as a function of head space.
The factory setting for the "slope" k6 is 1.
cair is the air content in ppm as determined by CarboQC.
k5 is the reference temperature and Lb2 is the solubility of air at the determined
measuring temperature T2[K].
Lb(k5) is the solubility of air at the reference temperature k5. This value can betaken from a table.
k7 is the head space to liquid ratio of the beverage container. If k7 0 then the
term in brackets equals 1; if k7 > 0 then the term in brackets is calculated.
Ctrans ........... Air content as a function of head space
k5 ............... Reference temperature [°C]
k6 ............... Slope
k7 ............... Headspace to liquid ratio of can/bottle
cair ............. . Air/nitrogen content in ppm as measured
T2 ............... Measuring temperature [K]
Lb2 ............. Solubility of air/nitrogen at measuring temperature [-]
Lb(k5) .......... Solubility of air/nitrogen at reference temperature k5 [-]
Ctrans k6*cair *
k7*273.1
T2
--------------- Lb2+
k7*273.1
273.1 k5+--------------------------- Lb k5 +
---------------------------------------------------------------=
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 66/78
14 CarboQC Methods
66 C26IB001EN-B
14.4 Calculation of the Package Pressure at a
Reference Temperature
The 10 CMxxPROD methods CM01PROD to CM10PROD not only allow entry of
coefficients k0 to k7 and ks, but also of the coefficients ka, kb, kc and kd. The
coefficients ka, kb, kc and kd can be modified if the package pressure, determined
at sample temperature, needs to be converted into the package pressure at a
reference temperature.
TIP The value for the pressure at a reference temperature pref can be obtained as
relative pressure if the ambient pressure in [bar] is keyed in. If kd is zero, the
obtained value of pref will correspond to the absolute pressure.
Example: Calculation of the relative reference pressure of sparkling wine at 20°C
(68°F)
1. Select the menu "methods" .
2. Choose any CMxxPROD method.
3. Re-name the method and give it a significant denotation, e.g. SPARKL20.
4. Leave "f u = 1".
5. Leave "k0" to "k4" and "k6" to "k7" unchanged.
6. Select the reference temperature "k5" and change the entry to "+20.0000" .
When setting "k5" , always enter the numerical value of °C, even if the unit Fis selected.
7. Set "ks = 1.51" (ks for SPARKL is 1.51 g/(L*bar)).
pref .............. Package pressure at reference temperature
pt ................ Package pressure as determined by CarboQC
ka ................ Conversion factor for pressure unit
Factory setting = [bar]. If the unit [bar] is desired, set ka = 1.
kb, kc ........... Factor for temperature compensation
Factory setting: +0.00000
Recommendation:
kb = -0.028 (negative sign!) and kc = 0.00075
kd ................ Ambient pressure in [bar]. Factory setting: +0.00000
k5 ................ Reference temperature in °Ct .................. Sample temperature as determined by CarboQC
pref ka pt 1 kb t k5 kc t k5 2 kd – – + – + =
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 67/78
14 CarboQC Methods
C26IB001EN-B 67
8. Set "ka" to "kc" as follows:
• ka = 1• kb = -0.028• kc = 0.00075
9. Change the preset value of "kd" and enter the average ambient pressure in
[bar].
Then
pref, 20 = 1 * {ptas determined at t * [1 - 0.028 (t - 20) + 0.00075 * (t - 20)²] - 1}
is the relative package pressure had it been measured at 20 °C (68 °F).
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 68/78
Appendix A: Technical Data
68 C26IB001EN-B
Appendix A: Technical Data
A.1 CarboQC
Measuring range
CO2: 0 to 12 g/L (0 to 6 vol.) at 30 °C (86 °F)
0 to 20 g/L (0 to 10 vol.) < 15 °C (59 °F)
Temperature: -3 to 30 °C (27 to 86 °F)
Pressure: 0 to 10 bar relative (0 to 145 psi)
Repeatabili ty, s. d.: 0.01 g/L (0.005 vol.) CO2
Reproducibili ty, s. d.: 0.05 g/L (0.025 vol.) CO2
Resolution: 0.01/0.001 g/L (vol.) selectable
Accuracy of the pressure sensor: ±0.01 bar accuracy of adjustment required Accuracy of the temperature sensor: ±0.2 °C
Sample volume: > 100 mL
Typical measuring time per sample: approx. 90 seconds
Environmental conditions
Ambient temperature: +0 to +40 °C (+32 to +104 °F)
Air humidity (CarboQC): 10 to 95 % relative humidity
Degree of protection (CarboQC
without power supply): IP 64
Overvoltage category:
(power supply)
II
Dimensions (L x W x H): 190 x 120 x 305 mm (7.5 x 4.7 x 12 inches)
Weight: approx. 3.1 kg (6.8 lbs)
Power
Power adapter: Input: AC 100 to 240 V, 1.5 A, 50 to 60 Hz
Output: DC 15 V, 3.0 A
Indoor use only
Battery: Nickel metal hydride
7.2 V, 1900 mAh
Integrated protection
Charging time: approx. 2 hours
WARNING
Serious injuries are possible through high voltage if the following hints are not
adhered to.
• Only connect devices to the interfaces that comply with PELV (protectiveextra-low voltage) according to EN 61140 or with SELV (safety extra-lowvoltage) according to EN 60950.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 69/78
Appendix A: Technical Data
C26IB001EN-B 69
A.2 PFD Fil ling Device
PC/printer in terface
COM1 and COM2:
RS232; 1200 to19200 Baud; 1 stop bit; 7 or 8
data bits; no, odd or even parity
Factory default settings: data bits: 7; stop bits: 1; parity: even; baud rate:9600
TXD ... transmit data
RXD ... receive data
GND ... signal ground (connected to earth)
TXD
RXD
GND
Specified for the following containers:
Can:
Glass bottle:
PET bottle:
Max. volume:
0.5 L
1 L
3 L
Compressed gas:
Necessary quality of air if used as compressed
gas supply:
6±0.5 bar (87±7 psi)
Class 5 from ISO 8573-1
max. particle size: 40 µm
max. pressure dew point: +7°C (44.6 °F)
max. oil content: 25 mg/m3
Environmental conditions
Ambient temperature
Operation:
Transportation/Storage:
+0 to +40 °C (+32 to +104 °F)
-20 to +70 °C (-4 to +158 °F)
Air humidity
Operation/Transportation/
storage: 0 to 100 % relative humidity
Degree of protection: IP 64
Dimensions (L x W x H):
Maximum height with piercing head in highest
position:
190 x 270 x 670 mm (7.5 x 10.6 x 26.4 inches)
1020 mm (40.2 inches)
Weight: 10.1 kg (22.3 lbs)
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 70/78
Appendix B: Wetted Parts
70 C26IB001EN-B
Appendix B: Wetted Parts
The following materials are in contact with the samples to be measured and with
the cleaning agents:
CarboQC:
PFD Filling Device:
Material Part
PU (Polyurethane) Hoses
PSU (Polysulfone) Measuring chamber
Stainless steel Inlet-outlet valve, stirrer, piston
PEEK (Polyetheretherketon) Stirrer bearing
EPDM (Ethylene propylene diene
Terpolymer)
O-rings
Material Part
PU, Nylon Hoses
Stainless steel Sample tube, guide poles
EPDM O-ring
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 71/78
Appendix C: Interface Commands for Communication between a PC and CarboQC
C26IB001EN-B 71
Appendix C: Interface Commands for
Communication between a
PC and CarboQC
Measurement data stored in the memory can be transferred to a PC via the PC
interface of CarboQC.
1. Connect the CarboQC PC interface to the PC interface using an appropriate
interface cable.
2. Make sure the interface settings of CarboQC and PC agree.
3. Test the communication using a simple interface program, e.g. Windows
Terminal, Procomm or Hyper Terminal.
• For data communication software, please contact your local distributor or
Anton Paar GmbH.
PC command CarboQC response Comments
help GetId
GetDataHead
GetDataUnit
GetData
ResetData
If the command "help" is sent
from the PC, CarboQC
responds with a list of available
commands.
get id id:00921503;CarboQC;V2.00A Readout of serial number and
firmware version.
get data head head:date;time;sample;method;
CO2;CO2;air;tm;pt;p0;condition;sol
Response. P0 = system
pressure.
get data unit unit:;;;;g/l;g/l ;ppm;°C;bar;bar;;
g/lbar
The units of the measuring
results are transferred,
separated by a semicolon. The
units are displayed according to
the units selected in the
submenu.get data no new data available
data:10.07.08;17:27; 1;
CM03PROD; 0.332; 0.332;26.9;
28.00; 1.54; 0.747;valid;1.448
No new data available.
Transfer of the first available
measuring result. One result
per entry "get data", starting
with the last result.
reset data reset data successful Resets read data to not fetched
status e.g. for second data
transfer.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 72/78
Appendix D: Error Messages
72 C26IB001EN-B
Appendix D: Error Messages
Display Cause Correction
air: "error" Air value out of specification. • Select the proper method for yoursample.
• Rinse the measuring chamber.Repeat the measurement. If the errorpersists, contact your Anton Paarrepresentative.
• Make sure that the CO2 concentration
of the sample is within specifications(Appendix A).
"CO2 - - - - - g/L" CO2 value out of range.
• Bubbles in the measuring cell.
• The measuring chamber is nottightly closed duringmeasurement.
• Make sure that the measuringchamber is free of bubbles.
• Contact your Anton Paarrepresentative.
"air: - - - - - ppm" Air value is not determined, e.g if
"compensation" is set "off" .
Activate the compensation in the menu
"adjustment > compensation" .
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 73/78
Appendix E: Troubleshooting
C26IB001EN-B 73
Appendix E: Troubleshooting
Q The CO2 reading of the water check with deionized water shows >0.03 g/L CO2.
A Sometimes, deionized water contains absorbed CO2 and displays the amountof CO2 absorbed. Either heat up and boil the water and cover it up while cooling
down or do the water check with tap water – but do not forget to rinse the
measuring chamber with deionized water afterwards!
Q The CO2 reading of the first sample after a water check is low.
A If you go from samples containing very low to very high CO2 contents, the first
result for the high CO2 content is slightly lower than the following ones. If only
limited amounts of sample are available, it is a good idea to condition the
instrument by e.g. measuring mineral water after the water check before
measuring the first sample.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 74/78
Appendix F: Firmware Versions
74 C26IB001EN-B
Appendix F: Firmware Versions
Firmware
version
Date of
release
Document
number Comments
V 1.00A 09.03.2004 C26IB02A First released firmware.
V1.00B 16.04.2004 C26IB02B Zero adjustment improved (v-h).
V1.00C 09.09.2004 C26IB02C • Measuring times at 10% volume expansion and30% volume expansion optimized.
• Total pressure display inside containerintroduced.
• Printout includes total pressure inside container.• Piston slows down before returning to starting
position.
V1.00D 15.02.2005 C26IB02D • First released firmware version for PBA-S.
V1.00E 24.06.2005 C26IB02E
C26IB02F
• Improved polynomial function for calculating theair content.
V2.00A 28.07.2006 C26IB02G • New microcontroller for increased performance.• Firmware update simplified.• New method "MID" for soft drinks with 3 to 8
°Brix.• 10 instead of 4 "CMxxPROD" methods.• Arbitrary selection of CO2 units.
• Selectable if methods are displayed in themethod list ("man sel/disable" ).
• Semiautomatic adjustment of the pressureoffset in the menu "adjustment" .
• Data memory can be viewed without entering apassword.
V2.00B 28.07.2008 C26IB02H • Simplified firmware update for future firmwareversions.
• Calculation of package pressure at referencetemperature.
V2.00C 13.01.2010 C26IB02I • Improved display controller.
V2.00D 19.05.2010 C26IB001EN-A • Rinse time selectable up to 200 seconds.• Numbers added to the methods.
V2.00E 10.05.2012 C26IB001EN-B • New function after long push of the <Stop> key:Method selection.
• New key combination: When <Start> and<Stop> are simultaneously pressed, a rinsingprocedure is started.
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 75/78
Appendix G: Declaration of Conformity
C26IB001EN-B 75
Appendix G: Declaration of Conformity
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 76/78
Appendix H: Menu Tree
76 C26IB001EN-B
Appendix H: Menu Tree
adjustment compensation comp air: on | off
vexp: 02, ..., 40 %
auto p-offset CO2: x.xxx g/l
p: x.xxx bar (psia)
instrument rinse time: 03, ..., 60 sec
unit t: °C | °F
p: bar | psia
CO2: g/l | vol
CO2dec: 2dec | 3dec
serial com baud: 9600 | 4800 | 2400 | 1200 | 19200
format: E71 | N81 | O71
password new: 0000
date & time date: ddmmyy | mmddyy
year: xx
month: xx
day: xx
time: 24hrs | am/pm
hour: xx
min: xx
battery service: off | on
device info print: off | printer | PC
methods 01 MINERALc03 DIETcccc
05 BEERcccc
07 ST.BEERc
09 COLAcccc
11 SPARKLcc
13 MIDccccc
name: MINERALc (xxxxxxxx)sel: man sel | disable
unit: g/l (xxxxxx)
fu, k1, ..., k7
17 CM01PROD, ...,
26 CM10PROD
name: MINERALc (xxxxxxxx)
sel: man sel | disable
unit: g/l (xxxxxx)
fu, k1, ..., k7, ks, ka, ..., kd
00 MINERAL
02 DIET04 BEER
06 ST.BEER
08 COLA
10 SPARKL.
12 MID
name: MINERAL
sel: man sel | disable
data memory browse memory
clear memory clear: no | yes
service device
offset
gain
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 77/78
Index
C26IB001EN-B 77
Index
A
Absorption coefficient 58 Activating lever 30, 40 Air content 65 Air humidity 68, 69 Ambient temperature 68, 69
B
Battery 17, 19, 42, 48, 54, 68Battery status 17Bubbles 30, 39, 41, 46, 49
C
Can 28CO2 content 53, 61, 62CO2 result 34, 36, 53, 55, 59, 63Coefficient 55, 63Compressed gas 11, 20, 69Compressed gas supply 30Customer specific 18, 27, 55, 56, 57, 62
D
Date 25, 54Deionized water 36, 38, 39, 40Display 17, 41, 59, 72Dissolved air 52, 59, 61
E
Exhaust 23
F
Filter 14, 22, 23, 39, 40Flow valve 28
G
Glassware cleaner 36, 38, 39, 40, 41
H
Headspace 26, 55, 62Hose 14, 70
K
Keypad 18, 41, 51
M
Marking 28Measurement 27, 30, 46, 49, 60Measurement data 59, 71Measuring chamber 16, 18, 30, 38, 39, 40, 41, 46, 49, 53, 60, 61, 70Measuring mode 18, 51Measuring range 68Measuring temperature 59, 65Memory 18, 30, 46, 49, 59, 71Menu 18, 25, 51, 52Menu access 52, 54Menu mode 18, 25, 51Method 18, 30, 49, 55, 56, 57, 59, 62Multiple Volume Expansion method 61
N
Nitrogen 56, 60, 61, 65
O
Offset 60On/off switch 24
P
Password 25, 52, 54PC interface 53, 69, 71PET bottle 28, 40PFD 27, 28, 38, 39, 40, 69Piston 30, 41, 70Power adapter 14, 17, 23, 54, 68Pressure 23, 25, 30, 41, 46, 49, 53, 68Pressure sensor 53, 60Print 18, 30, 54
Printer interface 53Printout 18Purified water 38
R
Reference pressure 66Rinse 18, 36, 38, 39, 40, 53
S
Sample flow 36Sample inlet 23Sample number 18, 49Sample outlet 23
7/17/2019 Manual CarboQC
http://slidepdf.com/reader/full/manual-carboqc 78/78
Index
Sample pressure 42, 48Sample residues 38, 39, 40Sample tube 30, 40Sample volume 68Service 19, 54
Slope 60, 65Solubility 60, 65Solubility coefficient 55Start key 18Stirrer 18, 30, 41, 49, 70Stop key 18Supply inlet 24, 54
T
Temperature 25, 27, 53, 63, 65Temperature sensor 60
Time 25, 54
U
Unit 25, 53
V
Volume expansion 52, 60, 61
W
Waste vessel 22, 46, 48
Z
Zero point 36