Manual CarboQC

78
                                                                                                                                                                                                                                                                                        Instruction Manual CarboQC Lab and At-line Beverage Carbonation Meter Firmware Version : V2.00E

description

Manual para el manejo de medición de CO2

Transcript of Manual CarboQC

Page 1: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 1/78

           

     

                                      

         

                             

             

                        

                                  

                                    

Instruction Manual

CarboQC

Lab and At-line Beverage Carbonation Meter Firmware Version : V2.00E

Page 2: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 2/78

 

Page 3: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 3/78

 

Instruction Manual

CarboQC

Lab and At-line Beverage Carbonation Meter Firmware Version: V2.00E

Page 4: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 4/78

 

 Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.

Nor is any liability assumed for damages resulting from information contained in the document.

 Anton Paar GmbH reserves the right to perform changes to the content. This also extends to changes

to delivery volumes or any features of delivered parts.

 All rights reserved (including translation). This document, or any part of it, may not be reproduced,

changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm

or any other process) without prior written permission by Anton Paar GmbH.

Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked

as such. They are the property of their respective owner.

Published by Anton Paar Printed: Anton Paar, Austria

Copyright © 2012 Anton Paar GmbH, Graz, Austria

 Address: Anton Paar GmbH Anton-Paar-Str. 20 A-8054 Graz / Austria – EuropeTel: +43 316 257-0Fax: +43 316 257-257

E-Mail: [email protected]: www.anton-paar.com

Date: May 21, 2012

Document number: C26IB001EN-B

Page 5: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 5/78

C26IB001EN-B 5

Contents

1 About the Inst ruct ion Manual ......................................................................................................... 8

2 Safety Inst ruct ions........................................................................................................................... 9

2.1 General Safety Instructions ..................................................................................................... 9

2.2 Special Safety Instructions for CarboQC in Combination with PFD Filling Device .................11

2.3 Warning Signs and Mandatory Signs on the PFD Filling Device............................................11

3 About CarboQC and PFD Fil ling Device ...................................................................................... 13

4 Checking the Suppl ied Parts ........................................................................................................ 14

5 Functional Components ................................................................................................................ 16

5.1 Front View.............................................................................................................................. 16

5.2 CarboQC Components .......................................................................................................... 165.2.1 Measuring Chamber............................................................................................... 16

5.2.2 Display.................................................................................................................... 17

5.2.3 Keypad ................................................................................................................... 18

5.3 Rear View.............................................................................................................................. 19

5.4 Integrated Battery .................................................................................................................. 19

6 Putting CarboQC into Operat ion .................................................................................................. 20

6.1 Installing CarboQC ................................................................................................................ 20

6.2 Switching the Instrument On and Off..................................................................................... 23

7 Inst rument Settings ....................................................................................................................... 25

7.1 Menu Access ......................................................................................................................... 25

7.2 Setting Date and Time ........................................................................................................... 25

7.3 Setting the Units .................................................................................................................... 25

8 Measurement .................................................................................................................................. 26

8.1 Preconditions......................................................................................................................... 26

8.2 Preparing CarboQC............................................................................................................... 26

8.3 Transferring the Sample into the Measuring Chamber.......................................................... 27

8.4 Measurement Using the PFD Filling Device.......................................................................... 27

8.4.1 Setting the Correct Flow......................................................................................... 27

8.4.2 Measurement Procedure........................................................................................ 29

8.5 Checking for a Correct Measurement Procedure and Avoiding Errors.................................. 33

9 Checking the System..................................................................................................................... 34

9.1 Influencing Factors ................................................................................................................ 34

9.2 Zero Point Check................................................................................................................... 34

9.3 Check with Carbonated Liquid............................................................................................... 34

9.4 Beginning of the Shift/Day..................................................................................................... 36

9.5 Entering a Pressure Offset .................................................................................................... 37

10 Cleaning the System...................................................................................................................... 38

10.1 End of the Shift/Day............................................................................................................... 3810.2 Weekly Maintenance ............................................................................................................. 39

Page 6: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 6/78

6 C26IB001EN-B

10.3 Cleaning the Filter.................................................................................................................. 40

10.4 Cleaning the Measuring Chamber......................................................................................... 40

10.4.1 Required Devices ................................................................................................... 40

10.4.2 Rinsing ................................................................................................................... 40

10.4.3 Cleaning the Entire Measuring Chamber Volume .................................................. 41

10.5 Cleaning the Instrument Housing and Display....................................................................... 41

11 Measuring At-line........................................................................................................................... 42

11.1 Using CarboQC with Protective Casing................................................................................. 42

11.1.1 Overview ................................................................................................................ 42

11.1.2 Mounting the Protective Casing ............................................................................. 44

11.1.3 Connecting CarboQC in the Protective Casing to the Production Line or Tank ..... 45

11.1.4 Measurement ......................................................................................................... 46

11.1.5 Cleaning the Measuring Chamber.......................................................................... 47

11.2 Using CarboQC without Protective Casing............................................................................ 4811.2.1 Connecting CarboQC to the Production Line or Tank ............................................ 48

11.2.2 Measurement ......................................................................................................... 49

11.2.3 Cleaning the Measuring Chamber.......................................................................... 50

12 Menu Operat ion.............................................................................................................................. 51

12.1 Using the Keys on the Keypad .............................................................................................. 51

12.2 Menu Description and Function............................................................................................. 51

12.2.1 Displaying the Measurement Data ......................................................................... 51

12.2.2 Accessing the Menu............................................................................................... 52

12.2.3 "adjustment" Submenu........................................................................................... 5212.2.4 "instrument" Submenu............................................................................................ 53

12.2.5 "methods" Submenu............................................................................................... 55

12.2.6 "data memory" Submenu ....................................................................................... 59

12.2.7 "service" Submenu ................................................................................................. 60

13 Techn ical Background................................................................................................................... 61

14 CarboQC Methods ......................................................................................................................... 62

14.1 Available Methods with CarboQC.......................................................................................... 62

14.1.1 Standard Methods .................................................................................................. 62

14.1.2 Customer-Specific Methods ................................................................................... 62

14.2 Establishing Customer-Specific Formulae............................................................................. 62

14.2.1 Converting the CO2 Result into a Customer-Specific Result.................................. 62

14.2.2 Converting the CO2 Concentration into a Customer-Specific Unit ......................... 64

14.2.3 Using a Customer-Specific Solubility Coefficient.................................................... 64

14.3 Transformation of the Determined Air Content ...................................................................... 65

14.4 Calculation of the Package Pressure at a Reference Temperature....................................... 66

 Appendix A: Technical Data ........................................................................................................... 68

 A.1 CarboQC ............................................................................................................. 68

 A.2 PFD Filling Device............................................................................................... 69

 Appendix B: Wetted Par ts ............................................................................................................... 70

Page 7: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 7/78

C26IB001EN-B 7

 Appendix C: Interface Commands for Communicat ion between a PC and CarboQC .............. 71

 Appendix D: Error Messages ......................................................................................................... 72

 Appendix E: Troubleshooting ........................................................................................................ 73

 Appendix F: Firmware Versions .................................................................................................... 74

 Appendix G: Declarat ion o f Conformi ty ........................................................................................ 75

 Appendix H: Menu Tree................................................................................................................... 76

Index..................................................................................................................................................... 77

Page 8: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 8/78

1 About the Instruction Manual

8 C26IB001EN-B

1 About the Instruction Manual

This instruction manual informs you about the installation and the safe handling

and use of the product. Pay special attention to the safety instructions and

warnings in the manual and on the product.

The instruction manual is a part of the product. Keep this instruction manual for

the complete working life of the product and make sure it is easily accessible to

all people involved with the product. If you receive any additions or revisions to

this instruction manual from Anton Paar GmbH, these must be treated as part of

the instruction manual.

Conventions for safety messages

The following conventions for safety messages are used in this instruction

manual:

TIP Tip gives extra information about the situation at hand.

Typographical conventions

The following typographical conventions are used in this instruction manual:

DANGER

Danger indicates a hazardous situation which, if not avoided, will result in

death or serious injury.

WARNING

Warning indicates a hazardous situation which, if not avoided, could result in

death or serious injury.

CAUTION

Caution indicates a hazardous situation which, if not avoided, could result in

minor or moderate injury.

NOTICE

Notice indicates a situation which, if not avoided, could result in damage to

property.

Convention Description

<key> The names of keys and buttons are

written inside angle brackets.

Page 9: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 9/78

2 Safety Instructions

C26IB001EN-B 9

2 Safety Instructions

2.1 General Safety Instructions

• Read this instruction manual before using CarboQC.

• Follow all hints and instructions contained in this instruction manual toensure the correct use and safe functioning of CarboQC.

Liability

• This instruction manual does not claim to address all safety issues

associated with the use of the instrument and samples. It is your

responsibility to establish health and safety practices and determine the

applicability of regulatory limitations.

• Anton Paar GmbH only warrants the proper functioning of CarboQC if no

adjustments have been made to the mechanics, electronics, and firmware.

• Only use CarboQC for the purpose described in this instruction manual.

 Anton Paar GmbH is not liable for damages caused by incorrect use of

CarboQC.

Installation and Use

• CarboQC is not an explosion-proof instrument and therefore must not be

operated in areas with risk of explosion.

• The installation procedure shall only be carried out by authorized personnel

who are familiar with the installation instructions.

• Do not use any accessories or wearing parts other than those supplied or

approved by Anton Paar GmbH.

• Make sure all operators are trained to use the instrument safely and correctly

before starting any applicable operations.

• In case of damage or malfunction, do not continue operating CarboQC. Do

not operate the instrument under conditions which could result in damage to

goods and/or injuries and loss of life.

"Menu Level 1 > Menu Level 2" Menu paths are written in bold, inside

straight quotation marks. The menu

levels are connected using a closing

angle bracket.

Convention Description

Page 10: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 10/78

2 Safety Instructions

10 C26IB001EN-B

• Check CarboQC for chemical resistance to the samples and cleaning

agents.

• Neither CarboQC nor, if applicable, the PFD Filling Device shall be used for

or filled with acids and alkalis above 5 %w/w, hydrofluoric acid, solvent and

other poisonous liquids.

• Do not place CarboQC on a wet or humid bench because humidity entering

the housing could damage the instrument.

• Place CarboQC and, if applicable, the PFD Filling Device in a way so that

there is no danger of accidentally dropping the equipment or tearing off the

pressurized hoses and fittings. Dropping the equipment increases the risk of

the pressurized container bursting.

• Do not expose CarboQC and the PFD Filling Device to direct sunlight for

extended periods of time.

Maintenance and Service

• Due to the nature of the measurement, the measuring results not only

depend on the correct use and functioning of the CarboQC and, if applicable,

the PFD Filling Device, but may also be influenced by other factors. We

therefore recommend you have the results checked (e.g. plausibility tested)

by skilled personnel before consequential actions are taken.

• Service and repair procedures may only be carried out by authorized

personnel or by Anton Paar GmbH.

Disposal

• Dispose of CarboQC and, if applicable, the PFD Filling Device according to

the country-specific legal requirements.

• Dispose of samples and cleaning liquids according to the country-specific

legal requirements.

Page 11: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 11/78

2 Safety Instructions

C26IB001EN-B 11

2.2 Special Safety Instructions for CarboQC in

Combination with PFD Filling Device

• Do not supply compressed gas with more than 6.5 bar (95 psi) relative

pressure to the PFD Filling Device.

• Use only clean and oil-free compressed air ("instrument air").

• In the unlikely event that a bottle or can bursts, immediately push the

activating lever of the PFD Filling Device to the de-energized position. Turn

off the compressed gas supply and then disconnect the PFD Filling Device

from the compressed gas supply. Renew the safety shield and discard the

old one. Damage of the instrument or parts thereof may be the consequence

of a burst container. Therefore, do not continue to use the PFD Filling Device

unless it has been checked by authorized service personnel. Contact your

local Anton Paar representative.

• The safety shield of the PFD Filling Device is subject to aging, and has to be

replaced after 5 years at the latest.

2.3 Warning Signs and Mandatory Signs on the PFD

Filling Device

Warning sign on the rear side of the PFD Filling Device

WARNING

If a glass bottle bursts inside the PFD, the safety shield protects the user from

glass chips. However, some small glass particles may still be released by the

rear gap of the safety shield which could lead to injuries of eye or face.

Hearing damage is possible in case a bottle or can bursts. Also damage of the

safety shield may occur.

• Therefore, place the PFD Filling Device with a distance of 0.5 m (20inches) with its back to the wall.

• The PFD Filling Device has to be placed so nobody can stand betweenthe PFD Filling Device and the wall.

• When pressurizing a glass bottle inside the PFD, only the operator,equipped with appropriate protective clothing, may be located directly infront of the PFD Filling Device. All other personnel has to keep a distanceof at least 3 m to the PFD Filling Device.

• If a container bursts, the vicinity might get sprayed with sample. The PFDFilling Device has to be positioned so that there are no power outlets orventilation slots behind the PFD Filling Device.

• In case of damage, the safety shield has to be replaced by trained servicepersonnel.

Page 12: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 12/78

2 Safety Instructions

12 C26IB001EN-B

On the front side of the PFD Filling Device

Mandatory signs

WARNING

If a glass bottle bursts inside the PFD, the safety shield protects the user fromglass chips. However, some small glass particles may still be released by the

rear gap of the safety shield which could lead to injuries of eye or face.

Hearing damage is possible in case a bottle or can bursts. Also damage of the

safety shield may occur.

• Do not pressurize empty bottles or cans. The safety shield is onlydimensioned for gas volumes resulting from the depth of immersion of thesample tube.

• Do not pressurize glass bottles with a volume higher than 1 L.Otherwise,the protection through the safety shield is not ensured.

• Do not pressurize bottles/cans which show cracks or substantialscratches. Adhere to specifications of bottle manufacturers.

• In the unlikely case of a bottle/can bursting, the safety shield may getdamaged and therefore has to be replaced by a new one. Renew thesafety shield and discard the old one.

Wear safety goggles:

Wear safety goggles when handling the PFD Filling Device.

Wear ear protection:

Wear ear protection when handling the PFD Filling Device.

Page 13: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 13/78

3 About CarboQC and PFD Filling Device

C26IB001EN-B 13

3 About CarboQC and PFD Filling

Device

CarboQC uses the patented "Multiple Volume Expansion" (MVE) method toeliminate the influence of dissolved air on the determined carbon dioxide content.

The analysis is based on highly accurate absolute pressure and temperature

measurements.

CarboQC can determine the CO2 content of sugar based soft drinks, soft drinks

with low sugar concentration, diet drinks, beer, strong beer, sparkling wine and

mineral water.

CarboQC can be used in a laboratory environment as well as at-line. For at-line

measurements, a special protective casing is available.

The instrument is equipped with rechargeable batteries to provide the required

power for a limited period of time.

To perform a measurement, connect the instrument to the PFD Filling Device in

the laboratory or to a production line or tank, select the proper method, and press

the <Start> key. Within approx. 2 minutes the CO2 concentration is shown on the

display and transferred to the data memory. The measuring result can also be

printed out immediately or later from the data memory.

TIP For detailed information about the PFD Filling Device, see the PFD Filling Device

instruction manual.

Page 14: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 14/78

4 Checking the Supplied Parts

14 C26IB001EN-B

4 Checking the Supplied Parts

CarboQC was tested and packed carefully before shipment. However, damage

may occur during transport.

1. Keep the packaging material (box, foam piece, transport protection) for

possible returns and further questions from the transport and insurance

company.

2. Check the delivery for completion by comparing the supplied parts to those

noted in Table 4.1.

3. For checking the supplied parts of the PFD Filling Device, see the PFD

Filling Device instruction manual.

4. If a part is missing, contact your Anton Paar representative.

5. If a part is damaged, contact the transport company and your Anton Paar

representative.

Table 4.1: Supplied parts

Symbol Pcs. Article Description Mat. No.

1 CarboQC 14805

1 Instruction manual

English or 

German

16605

16606

1 Power adapter incl. cable 14869

4m Hose 2.5x4 polyurethane

transparent

15272/

1m

1 Filter incl. 1# sieve

1# Sieve D = 8 mm for filter 

20054

16118

1 Straight fitting R1/8" 4 mm 15274

Page 15: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 15/78

4 Checking the Supplied Parts

C26IB001EN-B 15

1 T-Hose connector 4 mm 17756

1 Wrench 10 3636

Table 4.1: Supplied parts

Symbol Pcs. Article Description Mat. No.

Optional accessories Mat. No.

Printer RS-232C 64434Printer interface cable for CarboQC 5-pole incl. Dsub adapter 25pin/

9pin

22434

Power adapter IP76 incl. cable

Splash-proof power supply with European standard power plug

23672

Customer functions upon

request

Fruit pulp filter incl. housing and 10# filter cartridges

10# Filter cartridges for fruit pulp filter 

21167

21168

 All-purpose filter incl. 1# filter cartridge

1# Filter cartridge for all-purpose filter 

25483

25391

Filter with male hose filling nozzle for at-line operation 18829

Carrying strap 16012

Protective boot CarboQC 90079

Carry case for CarboQC, 440 x 250 x 170 mm 23888

Protective casing CarboQC 102189

Page 16: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 16/78

5 Functional Components

16 C26IB001EN-B

5 Functional Components

5.1 Front View

Fig. 5 - 1 Front view

1 ... Flow valve2 ... Sample outlet

3 ... Sample inlet4 ... Lugs for carrying strap5 ... Keypad6 ... Display7 ... Measuring chamber 

5.2 CarboQC Components

5.2.1 Measuring Chamber 

Fig. 5 - 2 Measuring chamber 

1

2

3

4

5

6

7

Flow valve

Sample outlet

Sample inlet

Piston

 to adjust the sample flow

with integrated stirrer 

Page 17: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 17/78

5 Functional Components

C26IB001EN-B 17

5.2.2 Display

Fig. 5 - 3 Alphanumerical LCD display (4 lines, 16 characters)

1 ... Activated method2 ... Battery status symbol3 ... Sample number 4 ... CO2 concentration

5 ... Air content6 ... Package pressure (pt) or measuring temperature (tm)

The 4th line either displays the measuring temperature or the package pressure.

The package pressure is calculated from the determined concentration of CO2

and other dissolved gases.

To switch the display from package pressure to measuring temperature, press the

<Menu> key (see Fig. 5 - 4) for approx. 5 seconds.

Battery status symbol

  /

(blinking)

The battery is being charged.

The battery is fully charged and the power adapter is connected.

The battery is fully charged and the power adapter is not

connected.

The battery is charged approx. 66 %.

The battery is charged approx. 33 %.

(blinking)

The battery is empty. The battery has to be charged again by

connecting CarboQC to the mains via the power adapter.

1 2 3

4

5

6

Page 18: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 18/78

5 Functional Components

18 C26IB001EN-B

5.2.3 Keypad

Fig. 5 - 4 Keypad

In the measuring mode the functions <Sample>, <Menu>, <Print>, <Method>,

<Rinse>, <Start> and <Stop> are active.

In the menu mode the functions <ESC>, <>, <> and <Enter> are active.

Key functions in the measuring mode

Special key func tions activated with a long push (approx. 5 sec.)

Menu/ To access the menu.

Method/Enter  To select one of 7 predefined or one of 17 customer-

specific methods.

Print/ To start a printout of the actual measuring result.

Rinse To rinse the measuring chamber. Further short pushes

activate/deactivate the integrated stirrer.Sample/ESC For entering a sample number. This number will be

increased each time you press <Start>.

Start To start a measurement.

Stop To stop a measurement, rinsing or cleaning procedure.

Menu To switch between measuring temperature and package

pressure.

Print To print out all measuring results stored in the data memory ofCarboQC.

Rinse To rinse the measuring chamber according to the settings in the

menu "instrument > rinse".

Start To clean the measuring chamber at its maximum volume

expansion.

Stop To switch to method selection mode. While keeping the key

pressed, CarboQC displays the available methods in intervals of

2 seconds. Releasing the key when the desired method is

displayed selects this method and the instrument switches back

to measurement mode.

Page 19: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 19/78

5 Functional Components

C26IB001EN-B 19

Special key combination (simultaneously pressed)

5.3 Rear View

Fig. 5 - 5 Rear view

1 ... Handle2 ... COM1 interface for PC or printer 3 ... Type plate with serial number 4 ... On/off switch5 ... COM2 interface for printer or data transfer to a PC with AP-SoftPrint6 ... Cover  7 ... Supply inlet

5.4 Integrated Battery

CarboQC has an integrated rechargeable battery. Up to 200 measurements can

be performed with a completely charged battery.

TIP To extend the lifetime of the internal battery, perform a battery service procedure

once every 3 months (see Chapter 12.2.4).

 A battery service can be performed on-site.

• To perform this service, select the menu "instrument > battery"   and set

"service"  to "on" .

The battery service will take a few hours and is best performed overnight. Itswitches itself off upon completion.

Start and Stop To start a rinsing procedure that has to be stopped by

pressing the <Stop> key.

1

2

34

5

6

7

Page 20: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 20/78

6 Putting CarboQC into Operation

20 C26IB001EN-B

If the battery has to be exchanged by an authorized service technician, make sure

that the battery service is performed, even if the newly installed battery is fully

charged.

6 Putting CarboQC into Operation

6.1 Instal ling CarboQC

CarboQC can be used in the laboratory as well as at-line. In any case, sample

has to be transferred into the measuring chamber of CarboQC without the loss of

carbon dioxide during the transfer. Subsequently, the installation of CarboQC in

combination with a PFD Filling Device is described.

The PFD Filling Device pierces the bottle closure or the base of the can and

forces the beverage into the measuring chamber of CarboQC using compressed

gas (6±0.5 bar relative/87±7 psi). In the unlikely event that a bottle or can bursts,

the operator is protected by a safety shield which is moved vertically over a bottle

or can. A safety pin on the bottom of the safety shield ensures that the system

cannot be pressurized as long as the safety shield is not in the protective position.

WARNING

If a glass bottle bursts inside the PFD, the safety shield protects the user from

glass chips. However, some small glass particles may still be released by therear gap of the safety shield which could lead to injuries of eye or face.

Hearing damage is possible in case a bottle or can bursts. Also damage of the

safety shield may occur.

• If a container bursts, chippings may be hurled through the gap on the backof the PFD Filling Device. Therefore place the PFD Filling Device with adistance of 0.5 m (20 inches) with its back to the wall. The PFD FillingDevice has to be placed so nobody can stand between the PFD FillingDevice and the wall.

WARNING

Compressed gas is used. Loose hoses may lead to injuries, especially of eyes

and face.

• If CarboQC is operated in combination with a PFD Filling Device, do notactivate the compressed gas supply before all hoses are properlyconnected.

NOTICE

• Do not operate CarboQC without the filter being installed.

• Do not place CarboQC on a wet or humid bench because humidityentering the housing could damage the instrument.

Page 21: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 21/78

6 Putting CarboQC into Operation

C26IB001EN-B 21

TIP If required, fasten the PFD Filling Device to the bench with clamps using the

opening in the front of the base plate or with a bolt through the corresponding hole

in the base plate.

Compressed gas supply

CarboQC and PFD Filling Device are best operated with compressed air. In case

a dissolved oxygen meter is also connected, use compressed nitrogen as

compressed air might influence the oxygen result.

If there is no in-house supply of compressed gas available, the PFD Filling Device

can also be operated with compressed gas from a gas cylinder.

The hoses of the PFD Filling Device have an outer diameter of 6 mm. If the

compressed gas outlet does not match the required diameter for PFD Filling

Device operation, check for pressure adapters at your local supplier.

Mounting the hoses

For the hose connections between CarboQC and PFD Filling Device, use the

delivered polyurethane hose 2.5 x 4 mm.

To connect a hose

• To connect a hose to a fitting, push the hose into the fitting as far as it will go.

TIP When reconnecting a hose, cut off a few millimeters of the hose if the color of the

hose is changed to achieve tight seal.

To disconnect a hose

• To remove the connection, press the ring of the fitting towards the fitting andthen pull out the hose.

NOTICE

Keep the hoses between the instruments as short as possible. This reduces

sample consumption as well as the risk of CO2 diffusing through the hoses

thus influencing the results.

NOTICE

Only take apart connections if it is really necessary to do so as frequent

handling may destroy the fitting.

Page 22: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 22/78

6 Putting CarboQC into Operation

22 C26IB001EN-B

Fig. 6 - 1 CarboQC and PFD Filling Device

1 ... PFD Filling Device sample tube outlet2 ... Connection to compressed gas supply 

2a PFD Filling Device outlet for compressed gas 2b PFD Filling Device inlet for compressed gas

3 ... Hose coming from the PFD Filling Device sample tube outlet4 ... Filter  5 ... CarboQC sample inlet6 ... CarboQC sample outlet7 ... Hose with T-hose connector to be led to the waste vessel

**) according to your requirements

2

4

5

7

1

1

2a 2b3

6

HoseInner

diameter 

Outer 

diameter Connects

Recommended

length

Nylon hose4.0 mm

(0.16")

 6.0 mm

(0.24")

PFD Filling Device compressed

gas inlet with compressed gas

supply

**)

PFD Filling Device compressed

gas outlet with T-hose connector

to be led into the waste vessel

**)

Polyurethane hose,

transparent

2.5 mm

(0.1")

 4.0 mm

(0.16")

PFD Filling Device sample tube

outlet with filter inlet

1000 mm (40")

Filter outlet with

CarboQC sample inlet

100 mm (4")

CarboQC sample outlet with

T-hose connector to be led into

the waste vessel

**)

Page 23: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 23/78

6 Putting CarboQC into Operation

C26IB001EN-B 23

To connect the PFD Filling Device to the compressed gas supp ly

1. Connect the compressed gas supply to the inlet (2b) of the PFD Filling

Device with a nylon hose 4x6 mm of the required length.

2. Connect another nylon hose 4x6 mm to the exhaust (2a) of the PFD FillingDevice.

3. Connect the other end to a T-hose connector 6 mm (supplied with the PFD

Filling Device) and lead the hose into the waste vessel (omission to connect

the T-hose connector will cause the hose to jump when releasing the

pressure from the PFD Filling Device).

To connect the PFD Filling Device with CarboQC

1. Connect a polyurethane hose 2.5 x 4 mm of approx. 1000 mm (40") length to

the sample tube outlet (1) of the PFD Filling Device and to the inlet of thefilter (4).

Mount the filter in the correct flow direction (indicated on the filter with anarrow).

2. Connect a polyurethane hose 2.5 x 4 mm of approx. 100 mm (4") length to

the outlet of the filter and the sample inlet (5) of CarboQC.

3. Connect a polyurethane hose 2.5 x 4 mm to the sample outlet (6) of

CarboQC.

4. Connect the T-hose connector 4 mm (supplied with CarboQC) to the otherend (7) and put it into the waste vessel.

5. When all hoses are connected properly, activate the compressed gas supply.

6.2 Switching the Instrument On and Off 

WARNING

Serious injuries are possible due to high voltage at parts of the instrument/

measuring system unless the following instructions are followed:

• Make sure that the non-fused earth conductor of the power cord isconnected to earth.

• For proper insulation to the mains only use the original power adapterwhich is supplied with the instrument.

• The power adapter is not suitable for operation under wet at-line oroutdoor conditions! Do not connect the power adapter of CarboQC to themains under wet ambient conditions.

Page 24: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 24/78

6 Putting CarboQC into Operation

24 C26IB001EN-B

1. Connect the power adapter to the supply inlet of CarboQC and the mains.

TIP When operating under laboratory conditions, the power adapter can be

permanently connected to CarboQC.

Fig. 6 - 2 Supply inlet and on/off switch

2. To switch CarboQC on, use the on/off switch.

3. If the display does not immediately light up, press the on/off switch for 10

seconds.

4. To switch CarboQC off, use the on/off switch.

If CarboQC is operated from the battery, it switches off automatically after 10

minutes if not used.

NOTICE

Before connecting the power adapter to the mains and switching the

instrument on, make sure that the correct line voltage is available (AC 100 to

240 V, 50 to 60 Hz). If large voltage fluctuations are to be expected, werecommend using a constant voltage source (UPS).

NOTICE

When connecting the plugs to CarboQC, tighten them with your fingers solely!

Do not use any tools for tightening!

Power supplyinlet

To themains

Power adapter On/off switch

Page 25: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 25/78

7 Instrument Settings

C26IB001EN-B 25

7 Instrument Settings

7.1 Menu Access

 A password has to be entered to access the menu. The default password is

"0000".

1. Press <Menu>, <>, select "menu-access"  and press <Enter>.

2. To enter the password, enter the characters individually using the <> and

<> keys, and confirm each setting with <Enter>.

3. After finishing all the required entries or changes in the menu, exit the menu

mode by pressing <ESC> up to 3 times.

7.2 Sett ing Date and Time

1. Set date and time in the menu "instrument > date & time".

2. Select the formats "mmddyy"  or "ddmmyy"  for the date and "24h"  or "am/

pm"  for the time.

7.3 Set ting the Uni ts

• Different units for the temperature (°C or °F), pressure (bar or psia =

"absolute psi"), and CO2 (g/L or another unit) can be selected in the menu

"instrument > unit" .

Page 26: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 26/78

8 Measurement

26 C26IB001EN-B

8 Measurement

8.1 Preconditions

CarboQC determines the amount of dissolved carbon dioxide in a beverage. To

get a correct measurement, the instrument has to be in good working condition

(see Chapter 9.2).

In addition, the sample package has to be in equilibrium. As carbon dioxide in a

package can be found in the liquid phase as well as in the head space,

equilibration is essential for repeatable and well defined measurement results.

Equilibration can be achieved by gently shaking the bottle horizontally about 15

times back and forth. This will ensure that the net transfer of CO 2 between liquid

and head space is zero, in short, in equilibrium.

8.2 Preparing CarboQC

CarboQC is equipped with 7 standard methods to account for the different

solubility of carbon dioxide in different beverages (see Chapter 14.1.1). The

selected method is displayed in the top left corner of the display and has to be

selected manually prior to a measurement by pressing the <Method> key,

selecting the desired method and confirming with <Enter>.

If necessary, press <Method> on CarboQC to select the appropriate method for

the sample to be measured from 7 standard methods or 17 customer-specific

methods. For customer-specific methods, see also Chapter 14.

The 7 standard methods are:

The standard methods do not compensate for any temperature effect that

influences the CO2 distribution between headspace and liquid of the bottles or

cans. Therefore, they should only be applied within a small range (e.g. 3 °C/5 °F

around reference temperature) of package temperatures.

Method Sample Concentration range

00 MINERAL Mineral water  

02 DIET Diet drinks 0-3 °Brix

04 BEER Beer 10-14 °Plato

06 ST.BEER Strong beer >14 °Plato

08 COLA Sugar-based soft drinks 8-12 °Brix

10 SPARKL. Sparkling wine 9-13% vol. alc.

12 MID Sugar-based soft drinks 3-8 °Brix

Page 27: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 27/78

8 Measurement

C26IB001EN-B 27

Compensation of the headspace-related temperature effect and therefore

measurements in a wider range of package temperatures can be achieved with

the 17 customer-specific methods once they are properly set up. For more

information, see Chapter 12.2.5.

8.3 Transferring the Sample into the Measuring

Chamber 

Sample transfer always carries the risk of loss of carbon dioxide or introduction of

other dissolved gases. Thus, sample transfer is best performed in combination

with a filling system that avoids these risks.

The PFD Filling Device is equipped with a seal to close off a small area of the

package, bottle or can. This small area is pierced with a knife to allow introduction

of a sample tube.

 Applying pressurized gas to the head space of the container allows the transfer

of sample to the CarboQC measuring chamber without the loss of CO 2.

Subsequently, the measurement of CO2 with CarboQC in combination with a PFD

Filling Device is described in detail.

TIP Piercing the package with a PFD Filling Device might result in the pierced disk

falling into the container. The filter between PFD Filling Device and CarboQC will

keep this particle from entering the measuring chamber.

8.4 Measurement Using the PFD Filling Device

8.4.1 Setting the Correct Flow

The flow through the measuring chamber should be such that 100 to 150 mL of

sample are consumed when the filling time is set to 20 seconds. Considerably

lower or higher flow rates may cause erroneous measurements due to sample

carry-over or gas bubbles in the measuring chamber.

Fig. 8 - 1 Measuring chamber and flow valve

Flow valve 

Page 28: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 28/78

8 Measurement

28 C26IB001EN-B

To adjust the flow, insert a suitable tool (e.g. screwdriver, ... ) into the cavity of the

flow valve and turn it clockwise for decreased and counterclockwise for increased

sample flow.

With a graduated vessel

1. Use a beaker or graduated cylinder to check the correct flow.

2. Insert the T-hose connector of the sample outlet hose into the beaker.

3. Select the menu "instrument > rinse"  and set the rinsing time to 20

seconds.

4. Position the flow valve (see Fig. 8 - 1) at about 10% flow, i.e. close to the

minimum flow position.

5. Carry out a measurement as described in Chapter 8.4.2.

6. Check the volume of liquid consumed: It should be 100 to 150 mL.

7. If the volume is not in the required range, slightly open or close the flow valve

(see Fig. 8 - 1) and repeat steps 5 to 7 until the volume/weight is within the

required range.

With a marked bottl e

1. Make volume markings "0" and "150 mL" on a transparent 0.5 L (16 oz) or

similar size PET bottle.

Use water and a graduated cylinder or a balance to correctly place thesemarkings on the bottle.

2. Fill the bottle up to the "0" mark with deionized water and close it.

3. Insert it into the PFD Filling Device.

4. Perform steps 3 to 7 as described above.

Via weighing

1. Weigh a full beverage bottle or can and make a note of the weight.

2. Insert the package into the PFD Filling Device.

3. Perform steps 3 to 5 as described above.

4. Remove the package from the PFD Filling Device and weigh it again.

The weight should have decreased by 100 to 150 g.

Page 29: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 29/78

8 Measurement

C26IB001EN-B 29

8.4.2 Measurement Procedure

 At CarboQC

1. Check the sample flow and adjust, if necessary (see Chapter 8.4.1).

2. Select the required method on CarboQC.

3. Select the sample number, if necessary.

 At the PFD Fil ling Device

1. Move the sample tube of the PFD Filling Device to the highest position (see

the PFD Filling Device instruction manual).

2. Push the safety shield upwards completely.

3. Insert/remove the PET bottle adapter, if required.

4. Adjust the position of the piercing head, if required.

5. Make sure that the bottle/can is equilibrated (see Chapter 8.1).

6. Insert the PET bottle/can/glass bottle.

7. Push the safety shield down until the safety pin engages.

8. Hold the safety shield down and pull the activating lever to pierce the

package.

9. Lower the sample tube of the PFD Filling Device into the package.

10. If the sample tube touches the bottom of the package, lift it slightly.

11. Fasten the sample tube.

NOTICE

Make sure that the bottles/cans are in equilibrium prior to putting them into thePFD Filling Device. Equilibrium is e.g. achieved by shaking the bottle/can 15

times horizontally.

Wear safety goggles:

Wear safety goggles when handling the PFD Filling Device.

Wear ear protection:

Wear ear protection when handling the PFD Filling Device.

Page 30: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 30/78

8 Measurement

30 C26IB001EN-B

 At CarboQC

1. Press <Start> to start the measurement.

The measuring chamber is first rinsed and filled with 100 to 150 mL of the

new sample.

2. Check the measuring chamber visually for bubble-free sample flow.

It must be free of bubbles before the actual analysis starts. If the bubbles donot disappear after approx. 15 seconds, clean the filter (see Chapter 10.3).

3. Check the pressure of the compressed gas supply. It shall be 60.5 bar

relative (877 psi).

Once the rinsing is finished, the analysis will start. The piston will move downto a first halt and then to a second halt with the stirrer permanently inoperation.

4. Wait approx. 90 seconds for the measurement result.

The measurement result will be displayed, stored in the memory andoptionally printed out. The piston will then move up, but the measuringchamber will remain closed.

TIP If a measurement was carried out with an incorrect method, the current result can

be recalculated using the correct method as long as this result is still displayed.

Press <Method>, select the correct method and press <Enter>. The recalculated

result will be shown on the display. To store and optionally print out this

recalculated result, press <Print>.

 At the PFD Fil ling Device

1. Move the sample tube up to the highest position and fasten it.

2. Push the activating lever to de-energize the system.

3. Move the safety shield up to the highest position and remove the package.

4. If the package sticks to the piercer, rotate it counterclockwise and pull it down

simultaneously.

NOTICE

If the sample has a high content of particles (e.g. pulp), the sieve of the filter

tends to clog. This will lower the pressure in the measuring chamber. Check

the pressure indicated at CarboQC during sample filling. When using a PFD

Filling Device, the pressure should not go below 5 bar/70 psia. If the sieve is

completely clogged, bubbles and foam form in the hose between filter and

measuring chamber and in the measuring chamber itself. Clean the filter as

described in Chapter 10.3.

Page 31: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 31/78

8 Measurement

C26IB001EN-B 31

5. Optionally, check with a balance whether the correct amount of sample was

taken from the bottle or can.

Now CarboQC and the PFD Filling Device are ready for the nextmeasurement.

6. If no further measurements are performed, clean the system (see Chapter  

10).

NOTICE

If the instrument is not in use for more than 30 minutes after the last sample

measurement, rinse the instrument with deionized water.

Page 32: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 32/78

8 Measurement

32 C26IB001EN-B

Measurement Procedure

Equilibrate the package.

Check the sample flow.

Select the required method on CarboQC and the sample number, if

necessary.

Move the sample tube to the highest position.

Push the safety shield upwards completely.

Insert/remove the PET bottle adapter, if required.

 Adjust the position of the piercing head, if required.

Insert the PET bottle/can/glass bottle.

Push the safety shield down until the safety pin engages. Hold it down.

Pull the activating lever.

Lower the sample tube into the package.

Press <Start> on CarboQC.

Wait approx. 90 seconds for the measurement result and optional printout.

Move the sample tube up and push back the activating lever.

Move the safety shield up and remove the package.

Perform the next measurement or clean the system.

Page 33: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 33/78

8 Measurement

C26IB001EN-B 33

8.5 Checking for a Correct Measurement Procedure

and Avoiding Errors

Zero point checks

Regular zero point checks of CarboQC with deionized water are recommended

to survey faultless functionality of the instrument. The water check with CarboQC,

when carried out correctly with deionized water, is the precondition for a correct

measurement result. The result has to be 0± 0.02 vol (0± 0.03 g/L).

If the zero point check and adjustment are carried out with water that contains

CO2, the consequence will be an unjustified offset. Consequences are systematic

errors of results.

Sample equilibration

Only transfer equilibrated sample to the measuring chamber of CarboQC. If

samples are not equilibrated (shaken) prior to analysis, the consequence might

be erroneous results or poor repeatabilities.

Correct measurement method

Check whether the correct measuring method is chosen on CarboQC. A wrong

method will lead to a wrong result.

Correct sample flow

Check the correct flow. The volume consumption per measurement is

recommended to be 120 to 150 mL, and for identical samples at least 100 mL.

Too low sample volume consumption causes sample carry-over.

Correct filling pressure

The filling pressure displayed on the CarboQC display has to be 5.5 to 6 bar

relative (79 psi to 87 psi) and the measuring chamber has to be filled without

bubbles. Too low sample pressures during sample filling causes bubbles in the

sample chamber. As a consequence, the starting volume will consists partially of

gas bubbles thus yielding incorrect results.

Measuring chamber filled without bubbles

Bubbles in the measuring chamber can falsify the result. The most common

cause for bubbles is a clogged filter. To clean the filter, see Chapter 10.3.

Page 34: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 34/78

9 Checking the System

34 C26IB001EN-B

9 Checking the System

9.1 Influencing Factors

Mechanical functionality of CarboQC is essential for a correct result. CarboQC

has three parameters that act as a precondition for a correct result.

• pressure• temperature• volume expansion of the measuring chamber 

9.2 Zero Point Check

 A zero point check is necessary before starting the determinations. The zero pointcheck is a measurement with deionized (or distilled) water. The CO2 result has to

be 0±0.02 vol (0±0.03 g/L).

The deionized water must not contain any dissolved carbon dioxide and is ideally

free of ions such as calcium that could cause deposits in the measuring chamber

of CarboQC.

Sometimes, deionized water contains CO2 absorbed from the ambience and

CarboQC displays the amount of the CO2 absorbed. Either heat up and boil the

water and cover it up while cooling down or do the water check with tap water. In

case of using tap water, rinse the measuring chamber with deionized waterafterwards.

In case the measurement of deionized water does give a CO 2  reading out of

tolerance, always check the water first before inserting a pressure offset (see

Chapter 9.5) into CarboQC.

In general, the zero point check with CO2-free water is sufficient to ensure correct

functioning of the instrument in-between service intervals.

9.3 Check with Carbonated Liquid

Sometimes checkups are required with test samples at elevated carbon dioxide

contents. Usually these checkups are provided within the periodical service and

maintenance procedure. In case more frequent tests are required, continuous

surveillance can be performed on a day-to-day basis.

There are two ways to perform an instrument check with carbonated liquids.

The first is the use of a CO2 standard. This procedure shall only be carried out by

an authorized Anton Paar representative or service technician.

The second is the use of a control chart and internal standards.

Page 35: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 35/78

9 Checking the System

C26IB001EN-B 35

Cans of diet soft drink or other cans of a non-sucrose based, non-inverting soft

drink with > 3 vol (6 g/L) CO2  content are suitable to be used as internal

standards.

If filling cans, provide 36 cans for initial preparation of the standards plus a

sufficient number of cans for one check measurement per day over the time span

between instrument calibrations by Anton Paar service representatives. Only use

cans from the same production minute(s).

If purchasing cans, make sure that the required number of cans is from the same

batch.

To p repare in ternal standards

1. Immediately after calibration of CarboQC by an Anton Paar service

representative, draw 6 cans from the beginning, 6 cans from the middle, and

6 cans from the end section of a production run and measure them.

2. Calculate the mean values for every 6 measurements for the beginning,

middle and end sections.

The 3 mean values may only differ by ±0.02 vol (±0.04 g/L).

3. If the mean values are suitable, provide a sufficient number of cans from this

production run to be kept as internal standards.

4. If the mean values are not suitable, measure another 3 times 6 cans as

above.

5. Determine the 3 mean values of the 12 respective measurements (for each

mean value take the 6 measurements from the first selection plus the 6

measurements from the second selection).

These 3 mean values may only differ by ±0.015 vol (±0.03 g/L).

6. If the mean values are suitable, provide a sufficient number of cans from this

production run to be kept as internal standards.

7. If the mean values are not suitable, the cans are not suitable as internal

standards and cans from a different, more stable production run have to be

used.

NOTICE

Only use cans as internal standards. Do not use bottles as their CO2 content

might change.

Page 36: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 36/78

9 Checking the System

36 C26IB001EN-B

To draw and use the cont rol chart

1. Calculate the mean m and standard deviation sd of all 18 (or 36)

measurements.

The standard deviation shall be 0.025 vol (0.05 g/L) maximum based on theabove limitations.

2. Draw a control chart with sample number on the x-axis and CO2

concentration on the y-axis ranging from (m - 3 * sd) to (m + 3 * sd) as the

control limits.

3. Draw control limit lines, mean line and auxiliary lines as appropriate. This

can either be done on paper or using a suitable computer software.

4. Measure one can of the internal standards per production day and draw a

corresponding point in the control chart.

5. If a clear trend away from the mean becomes visible over time or the control

limit of 3 * sd is exceeded successively, call for service immediately.

9.4 Beginning of the Shift /Day

Check for the correct sample flow with

deionized water. A volume of 100 to 150 mL isrecommended.

NOK  Adjust the sample flow.

See Chapter 8.4.1.

OK

Check the zero point: Fill a clean 0.5 L PET

bottle with deionized water, close the bottle and

measure it (see Chapter 8.4.2). The CO2 result

has to be 0±0.02 vol (0±0.03 g/L).

  OK

NOK Rinse with glassware cleaner (pH <10) andthen with deionized water.

OK

Check the zero point with deionized water again

(see Chapter 8.4.2). The CO2 result has to be

0±0.02 vol (0±0.03 g/L).

CarboQC is ready for

routine measurements.

OK

NOK

Contact your Anton Paar representative.

Page 37: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 37/78

9 Checking the System

C26IB001EN-B 37

9.5 Entering a Pressure Offset

With the menu "auto p-offset" , an offset for the pressure sensor can be

determined semiautomatically. Enter a pressure offset if the CO2 result of a zero

point check with deionized water is out of specification (see Chapter 9.2).

TIP Sometimes, deionized water contains CO2 which has been absorbed from the air,

and this CO2  content is displayed. Before carrying out an automatic pressure

offset, make sure that the displayed CO2 is not caused by the water, but by the

pressure sensor. Boiling and cooling the water before measurement will eliminate

dissolved CO2.

Example:

 A zero point check with deionized water yields (even after cleaning the system) a

CO2 value of 0.05 g/L. The maximum allowed deviation for deionized water is 0.03

g/L. Therefore, the pressure sensor must be adjusted:

1. Select the menu "adjustment > auto p-offset" .

In the first line, the previously measured CO2 value (0.05 g/L) is displayed. In

the second line, the pressure value is displayed (e.g. +0.020 bar).

2. Press <Enter>.

The displayed CO2 value flashes.

3. Change the CO2 value using the <> and <> keys until the value is insidespecification (0 ±0.03 g/L).

The pressure value is changing corresponding to the CO2 value.

4. When the required CO2 value is displayed, press the <Enter> key.

The newly calculated pressure offset is automatically stored in the CarboQCmemory and is used for subsequent measurements.

5. To change to the measuring window, press <ESC> three times.

Page 38: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 38/78

10 Cleaning the System

38 C26IB001EN-B

10 Cleaning the System

10.1 End of the Shift/Day

Rinse the measuring chamber with deionized water.

Rinse with glassware cleaner (pH <10) by pressing <Start> for approx. 5

seconds to clean the entire measuring chamber. Wait 2 minutes for this to

finish and then press <Stop>.

Rinse the measuring chamber with deionized water.

Clean sample residues from the PFD Filling Device and CarboQC with a

wet tissue. Dry with a soft dry tissue. If the PFD Filling Device is very dirty,

spray it with (warm) water, move movable parts up and down in between

spraying and then dry with a soft tissue.

NOTICE

If the instrument is not used for some time (e.g. over the weekend) or if

shipping the instrument, always make sure that the measuring chamber is

about half filled with deionized water to avoid damage of the O-rings.

Page 39: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 39/78

10 Cleaning the System

C26IB001EN-B 39

10.2 Weekly Maintenance

Clean the filter (see Chapter 10.3).

Rinse the measuring chamber with warm

deionized water.

Rinse the measuring chamber with glassware

cleaner (pH <10).

Rinse the measuring chamber with deionized

water.

Rinse the measuring chamber with a 60 to 90%

solution of ethanol. Do not use denatured

ethanol!

Rinse the measuring chamber with deionized

water.

Fill deionized water by pressing <Start> for

approx. 5 seconds to clean the entire measuring

chamber. Wait 2 minutes for this to finish.

Check whether gas bubbles enter the measuring

chamber.

Yes Contact your Anton

Paar representative.

No

Press <Stop> and leave deionized water in the

measuring chamber.

Clean sample residues from the PFD Filling

Device and CarboQC with a wet tissue. Dry with

a soft dry tissue. If the PFD Filling Device is very

dirty, spray it with (warm) water, move movable

parts up and down in between spraying and then

dry with a soft tissue.

Page 40: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 40/78

10 Cleaning the System

40 C26IB001EN-B

10.3 Cleaning the Filter 

TIP Make sure that the activating lever of the PFD Filling Device is in the "off" position

before disconnecting the filter.

1. Press <Rinse> on CarboQC and wait 20 seconds for the measuring

chamber to open.

2. Disconnect the filter from the hoses and connect it in the reverse direction to

the hose coming from the PFD Filling Device.

3. Put a 0.5 L or similar size PET bottle with deionized water into the PFD

Filling Device (see Chapter 8.4.2).

4. Hold the filter outlet over a sink or waste container and fix it in this position.

5. Pull down the safety shield and pull the activating lever.

 Air with sample residues will spray out of the filter.

6. Lower the sample tube to flush the filter with approx. 200 mL of deionized

water.

7. Move the sample tube up to let air flow through the filter for about one

minute.

8. Push the activating lever to the "off" position.

9. Reconnect the filter in the correct flow direction.

10.4 Cleaning the Measuring Chamber 

10.4.1 Required Devices

• To clean the measuring chamber of CarboQC, fill a 0.5 L or similar size PET

bottle with deionized water or glassware cleaner (pH <10) which removes

protein.

• Insert this bottle into the PFD Filling Device the same way as a sample to be

measured.

10.4.2 Rinsing

1. Press the <Rinse> key for approx. 5 seconds.

The measuring chamber is rinsed for the set rinse time.

2. Always rinse with deionized water after rinsing with glassware cleaner.

Page 41: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 41/78

10 Cleaning the System

C26IB001EN-B 41

3. Leave only deionized water in the measuring chamber for extended periods

of time.

10.4.3 Cleaning the Entire Measuring Chamber Volume

1. Press the <Start> key for approx. 5 sec.

 After rinsing, the piston will descend to the maximum measuring chambervolume expansion with the stirrer turned on. This will clean the entirechamber volume.

If deionized water is used, the pressure in the measuring chamber willbecome very low, far below ambient pressure. Air bubbles will enter themeasuring chamber if it is not completely tight. Therefore, this function canalso be used to detect leaks caused by damaged O-rings or valves.

2. Press <Stop> to exit this mode.

3. Always rinse with deionized water after rinsing with glassware cleaner.

4. Leave only deionized water in the measuring chamber for extended periods

of time.

10.5 Cleaning the Instrument Housing and Display

If sample was spilled onto the housing of CarboQC, always wipe it off

immediately.

• To clean the instrument housing and/or the keypad, use a soft tissue and

(warm) water.

• Dry with a soft, dry tissue.

Page 42: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 42/78

11 Measuring At-line

42 C26IB001EN-B

11 Measuring At-line

11.1 Using CarboQC with Protective Casing

11.1.1 Overview

The protective casing for CarboQC is especially designed for use in harsh

process environments. The robust polyurethane casing protects CarboQCagainst mechanical impact and splash water. Special foam parts inside the casing

dampen mechanical shocks on the instrument if the instrument falls down or

bumps into an obstacle.

The protective casing is mounted and dismounted fast and easily. The most

important standard tasks during a measurement can be carried out using the keys

on the protective casing:

• switching the instrument on and off • selecting a method• starting and stopping a measurement

• reading the display

 A stainless steel handle allows easy carrying of the casing with the CarboQC

inside. Also opening and closing is done comfortably by draw latches.

Technical data

WARNING

Serious injuries through high voltage at instrument parts possible.

• The power supply is not suitable for operation under wet at-lineconditions.

• Charge the battery of CarboQC only under dry indoor conditions.• If CarboQC is not used for extended periods of time (> 2h), store the

instrument under dry indoor conditions only.

WARNING

Operation under pressure. Serious injuries and damage of goods possible if

the maximum allowable pressure is exceeded.

• Do not operate CarboQC with sample pressures higher than 10 bar (145psi) absolute.

Weight 1.95 kg (4.30 lbs)

Dimensions L x W x H 240 x 165 x 395 mm (9.45 x 6.5 x 15.55 in)

 Ambient temperature 0 °C to 40 °C (32 °F to 104 °F)

 Air humidity 10 to 95 % relative humidity

Protection class IP64

Page 43: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 43/78

11 Measuring At-line

C26IB001EN-B 43

Front view and rear view

Fig. 11 - 1 Front view and rear view

1 ... Handle2 ... Viewing window3 ... <Start> key4 ... <Stop> key5 ... Inlet of the protective casing6 ... Outlet of the protective casing

Side views

Fig. 11 - 2 Side views

1 ... Draw latches with secondary catch2 ... On/off key

Material housing Polyurethane

Wetted parts Polyurethane (hoses), brass (connections)

1

2

34

5 6

1

2

Page 44: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 44/78

11 Measuring At-line

44 C26IB001EN-B

11.1.2 Mounting the Protective Casing

Fig. 11 - 3 Opening the protective casing

1. To open the casing, press the secondary catch (1) on the draw latches on

both sides of the casing.

2. Loosen the upper part of the draw latches (2).

3. Now the draw latches are open (3) and you can remove the upper part of the

casing.

Fig. 11 - 4 Connecting the hoses to CarboQC

4. Put the ends of the hoses out of the casing and leave the little bag (4) with

silica gel in the casing.

5. Put the CarboQC inside the housing so that the on/off switch is positioned at

the correct place.

6. Connect the hose coming from the outlet opening (5) to the CarboQC

sample outlet.

7. Connect the hose coming from the inlet opening (6) to the CarboQC sample

inlet.

8. Close the casing with the draw latches.

Now CarboQC inside the protective casing is ready for use at the line.

1 2 3

4

5

6

Page 45: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 45/78

11 Measuring At-line

C26IB001EN-B 45

11.1.3 Connecting CarboQC in the Protective Casing to the

Product ion Line or Tank

CarboQC can be connected to the line via filter or filter with male hose filling

nozzle (at-line adapter, Mat. No. 18829). This at-line adapter allows for thickerdiameter hoses between adapter and line.

Fig. 11 - 5 Optional at-line adapter (filter with male hose filling nozzle, Mat. No.

18829)

TIP For at-line measurements, keep the hoses between the line and the filter as short

as possible to reduce pressure drops and minimize the risk of the sample

outgassing. The correct flow direction is marked with an arrow.

Fig. 11 - 6 Connecting the instrument to the line

1. Connect the outlet of the line/tank to the inlet of the filter (standard filter or at-

line adapter).

Line

Samplevalve

Filter 

from theoutlet

to thewaste

to theinlet

Page 46: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 46/78

11 Measuring At-line

46 C26IB001EN-B

2. Connect the outlet of the filter to the inlet of the protective casing using a

polyurethane hose 2.5 x 4 mm.

3. Connect a polyurethane hose transparent 2.5 x 4 mm to the outlet of the

protective casing.

4. Place the other end of this hose in the waste vessel.

Setting the correct flow

To set the correct flow, use the PFD Filling Device and proceed as described in

Chapter 8.4.1.

11.1.4 Measurement

TIP The minimum pressure for at-line operation of CarboQC has to be 0.5 bar (7.2 psi)

above saturation pressure of the dissolved gases. Proper functioning is best

determined by practical tests.

1. To select the method, press the <Stop> key on the protective casing and

hold it pressed until the desired method is displayed. Then release the key.

The method is selected and CarboQC switches back to measurement mode.

2. Open the sample valve of the line or tank.

3. Start the measurement by pressing the <Start> key on the protective casing.

4. Watch the measuring chamber.

 After approx. 15 seconds rinsing time it should be free of bubbles.

TIP If the system pressure of the line is too low, the measuring chamber will not be

free of bubbles in time. Fill the measuring chamber of CarboQC manually byrinsing it before pressing <Start>.

• Press the <Start> key and <Stop> key simultaneously and let a minimum of

100 mL flow through the measuring chamber, then press <Stop> to stop the

rinsing procedure.

• Make sure the measuring chamber is free of gas bubbles and press <Start>.

The measuring result is automatically displayed after approx. 90 secondsand stored in the memory of CarboQC (max. 99 measuring results).

NOTICE

Do not disconnect any hoses from the protective casing during the

measurement.

Page 47: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 47/78

11 Measuring At-line

C26IB001EN-B 47

11.1.5 Cleaning the Measuring Chamber 

By connecting CarboQC to a tap with process water 

• Attach the inlet valve of the protective casing to a tap with process water(NOT with tap water as the limestone might harm the inlet and outlet valves).

• Then press simultaneously <Start> and <Stop> on the protective casing to

open the valves and rinse the measuring chamber.

• Press <Stop> to stop the rinsing procedure.

With a peristaltic pump

• Connect the outlet tubing of a peristaltic pump to the filter or at-line adapter

that is connected to the inlet of the protective casing.

• Then press simultaneously <Start> and <Stop> on the protective casing to

open the valves and rinse the measuring chamber.

• Then operate the pump as required and rinse the measuring chamber with

detergent and/or deionized water. If detergent is used, the measuring

chamber has to be rinsed carefully with process water or deionized water

afterwards.

• Press <Stop> on the protective casing to stop the rinsing procedure.

By connecting CarboQC to the process line

• If the process line is cleaned with water or detergent, leave CarboQC inside

the protective casing connected.

• Then press simultaneously <Start> and <Stop> on the protective casing to

open the valves and rinse the measuring chamber.

• Press <Stop> on the protective casing to stop the rinsing procedure.

• If detergent is used, the measuring chamber has to be rinsed carefully with

process water or deionized water afterwards.

Page 48: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 48/78

11 Measuring At-line

48 C26IB001EN-B

11.2 Using CarboQC without Protective Casing

11.2.1 Connecting CarboQC to the Production Line or Tank

TIP If CarboQC is frequently used for at-line measurements at different locations, we

recommend equipping CarboQC with a protective boot (Mat. No. 90079) for

protection of the instrument during transport.

CarboQC can be connected to the line via filter or filter with male hose filling

nozzle (at-line adapter, Mat. No. 18829). This at-line adapter allows for thicker

diameter hoses between adapter and line (see Fig. 11 - 5).

TIP For at-line measurements, keep the hoses between the line and the filter as short

as possible to reduce pressure drops and minimize the risk of the sample

outgassing. The correct flow direction is marked with an arrow.

1. See also Fig. 11 - 6: Connect the outlet of the line/tank to the filter inlet or the

inlet of the at-line adapter or standard filter.

2. Connect the outlet of the filter to the inlet of CarboQC using a polyurethane

hose 2.5 x 4 mm.

3. Connect a polyurethane hose transparent 2.5 x 4 mm to the outlet of

CarboQC.

4. Place the other end of this hose in the waste vessel.

WARNING

Serious injuries through high voltage at instrument parts possible.

• The power supply is not suitable for operation under wet at-lineconditions.

• Charge the battery of CarboQC only under dry indoor conditions.• If CarboQC is not used for extended periods of time (> 2h), store the

instrument under dry indoor conditions only.

WARNING

Operation under pressure. Serious injuries and damage of goods possible if

the maximum allowable pressure is exceeded.• Do not operate CarboQC with sample pressures higher than 10 bar (145

psi) absolute.

NOTICE

Do not place CarboQC on a wet or humid bench because humidity entering

the housing could damage the instrument.

Page 49: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 49/78

11 Measuring At-line

C26IB001EN-B 49

TIP  A carrying strap (Mat. No. 16012) is optionally available for safe and convenient

operation under at-line conditions.

Setting the correct flow

To set the correct flow, use the PFD Filling Device and proceed as described in

Chapter 8.4.1.

11.2.2 Measurement

TIPThe minimum pressure for at-line operation of CarboQC has to be 0.5 bar (7.2 psi)above saturation pressure of the dissolved gases. Proper functioning is best

determined by practical tests.

1. Press the <Method> key on CarboQC and select the required method.

2. If necessary, change the sample number by pressing the <Sample> key.

3. Open the sample valve of the line or tank.

4. Start the measurement by pressing the <Start> key on CarboQC.

5. Watch the measuring chamber.

 After approx. 15 seconds rinsing time it should be free of bubbles.

TIP If the system pressure of the line is too low, the measuring chamber will not be

free of bubbles in time. Fill the measuring chamber of CarboQC manually by

pressing the <Rinse> key before pressing <Start>. After the measuring chamber

opens,

• press the <Rinse> key periodically to turn the stirrer on and off,

• make sure that a minimum of 100 mL flows through the measuring

chamber,

• make sure the measuring chamber becomes free of gas bubbles.The measuring result is automatically displayed after approx. 90 secondsand stored in the memory of CarboQC (max. 99 measuring results).

NOTICE

Do not disconnect any hoses from CarboQC during the measurement.

Page 50: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 50/78

11 Measuring At-line

50 C26IB001EN-B

11.2.3 Cleaning the Measuring Chamber 

By connecting CarboQC to a tap with process water 

• Attach the inlet valve of CarboQC to a tap with process water (NOT with tapwater as the limestone might harm the inlet and outlet valves).

• Then hit the <Rinse> key on the instrument to open the valves and rinse the

measuring chamber.

With a peristaltic pump

• Connect the outlet tubing of a peristaltic pump to the filter or at-line adapter

that is connected to the inlet valve of CarboQC.

• Hit the <Rinse> key on CarboQC.

• Then operate the pump as required and rinse the measuring chamber with

detergent and/or deionized water. If detergent is used, the measuring

chamber has to be rinsed carefully with process water or deionized water

afterwards.

By connecting CarboQC to the process line

• If the process line is cleaned with water or detergent, leave CarboQC

connected and hit the <Rinse> key.

• If detergent is used, the measuring chamber has to be rinsed carefully withprocess water or deionized water afterwards.

Page 51: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 51/78

12 Menu Operation

C26IB001EN-B 51

12 Menu Operation

12.1 Using the Keys on the Keypad

If the measuring window is displayed, the functions <Sample>, <Menu>, <Print>,

<Method>, <Rinse>, <Start> and <Stop> are active.

In the menu mode (after pressing the <Menu> key) the functions <ESC>, <>,

<> and <Enter> are active.

The key functions in the menu mode accomplish the following tasks:

12.2 Menu Description and Function

12.2.1 Displaying the Measurement Data

The stored measurement data can be displayed. To display the next or previous

data set, press the <> or <> key.

Fig. 12 - 1 Displaying the measurement data

ESC • Return to the measuring mode• Move to the next higher menu level

• Ignore the last entry and return to the next higher menulevel• Ignore the last entry and return to the previous digit or

character 

 and • Move to a menu item• Decrease or increase a digit or a character 

Enter  • Select a menu item• Confirm an input or a query• Accept the last entry and return to the next higher menu

level• Accept the last entry and move to the next digit or

character 

m e n u

d a t a m e m o r y

m e n u a c c e s s

Page 52: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 52/78

12 Menu Operation

52 C26IB001EN-B

12.2.2 Accessing the Menu

Press the <Menu> key on CarboQC to initiate the menu access.

Fig. 12 - 2 Main menu (scrolls when arrow is moved down)

For the complete menu tree, see the last pages of this instruction manual.

12.2.3 "adjustment" Submenu

Fig. 12 - 3 Menu/adjustment

 menu-access

password  A password has to be entered to

access the menu. The default

password is: 0000.

If you use a different password, press

the <Enter> key. Enter the characters

using the <> and <> keys, and

confirm each setting with <Enter>.

m e n u

a d j u s t m e n t

i n s t r u m e n t

m e t h o d s

m e n u - a d j u s t m e n t

c o m p e n s a t i o n

a u t o p - o f f s e t

compensation

  comp air 

on To compensate the influence of

dissolved air. CarboQC automatically

performs two volume expansions.

of f  To deactivate the air compensation.

One volume expansion has to be set.

vexp To set the volume expansion

between 2 and 40%. Default: 10%.

Only valid when comp air is set to off.

Page 53: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 53/78

12 Menu Operation

C26IB001EN-B 53

12.2.4 "instrument" Submenu

Fig. 12 - 4 Menu/instrument

  auto p-offset Offset for the absolute pressure

sensor (values between 0 and ±0.5

bar (0 and ±7 psi)). For details, see

Chapter 9.5.

C: x.xxx g/L (or respective selected unit)

p: x.xxx bar (psia)

m e n u - i n s t r u m e n t

r i n s e

u n i t

s e r i a l c o m

  rinse

  time: 03 to 200 seconds for rinsing the

measuring chamber. Default: 20

seconds.

unit To select the units of the displayed

CO2 content, pressure andtemperature as well as the number of

places after the decimal point.

t To select °C or °F of the temperature.

p To select bar or psia (=absolute psi)

of pressure.

co2 To select g/L or an arbitrary unit of

CO2 content.

co2dec To select 2 or 3 places after the

decimal point for the displayed CO2 

result, and 0 or 1 places after thedecimal point for the displayed air

content, respectively.

serial com The settings in this menu are valid for

both the printer and  PC interfaces.

baud

1200

2400

4800

  9600

19200  

Page 54: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 54/78

12 Menu Operation

54 C26IB001EN-B

format

 E71 Even parity, 7 data bits, 1 stop bit

 N81 No parity, 8 data bits, 1 stop bit

 O71 Odd parity, 7 data bits, 1 stop bit

password

new To define a new password for the

menu access. 4 characters/digits are

available. Default: 0000.

date & time Date and time and the corresponding

formats can be set.

date Two different formats for the date are

available (default: ddmmyy)

 ddmmyy Day.month.year, e.g. 13.10.03

 mmddyy Month-day-year, e.g. 10-13-03 year  To set the current year.

month To set the current month.

day To set the current day.

time Two formats for time are available.

24hrs E.g. 17:08

am/pm E.g. 05:10 pm

hour  To set the hour.

min To set the minute.

battery

service

on To perform a battery service

procedure. With the power adapter

connected to the CarboQC supply

inlet and the mains, the battery is

discharged and charged again. This

takes approx. 12 hours.

of f  Battery service procedure inactive.

device info  All the values given under device info

are read only.

print Output of the device info to printer or

PC via the corresponding interface.

of f 

printer 

PC

Page 55: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 55/78

12 Menu Operation

C26IB001EN-B 55

12.2.5 "methods" Submenu

In this submenu a list of customer adaptable methods is displayed. The names as

specified here are only valid if no changes have been made. Once the method-

specific names have been changed, the changed names will appear in the list.The corresponding coefficients for these methods will only be explained for the

first method.

The last 10 methods (CM01PROD - CM10PROD) allow to enter a solubility

coefficient ks , too. They can therefore be adapted to customer-specific

beverages which exceed the predefined methods range in °Brix, °Plato or % alc.

vol. of beverages.

Fig. 12 - 5 Menu/customer 

m e n u - m e t h o d s

0 1 M I N E R A L c

0 3 D I E T c c c c

0 5 B E E R c c c c

01 MINERALc Customer-specific method for mineral

water.

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit  An arbitrary unit can be entered (see

Chapter 14.2.2). Factory default: g/L.

fu Division factor which is used to convert

the determined result (in g/L) into a

customer-specific unit, e.g. enter 1.976 to

convert g/L to vol. For calculating

customer-specific polynomials.

k0 Coefficients k0 - k4 convert the CO2

results based on the built-in solubility

formula to customer-specific results

based on customer-specific tables/

formulae. These parameters are also

used to compensate for the headspace

effect in bottles and cans which causes a

lower amount of dissolved CO2 at higher

temperatures (see Chapter 14.2.1).

k1

Page 56: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 56/78

12 Menu Operation

56 C26IB001EN-B

k2

k3

k4

k5 Reference temperature [°C].

k6 Coefficients k6 - k7 are used to transform

the air or nitrogen content as determined

into related properties (see Chapter

14.3).

k7

03 DIETcccc Customer-specific method for diet

products (0 to 3°Brix).

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit For the unit, an arbitrary unit can be

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

05 BEERcccc Customer-specific method for beer (10 to

14°Plato).

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit For the unit, an arbitrary unit can be

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

07 ST. BEERc Customer-specific method for strong beer

(>14°Plato).

name To define a customer-specific name for

this method.

sel

Page 57: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 57/78

12 Menu Operation

C26IB001EN-B 57

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit For the unit, an arbitrary unit can beentered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

09 COLAcccc Customer-specific method for sugar

based beverages (8 to 12°Brix).

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit For the unit, an arbitrary unit can be

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

11 SPARKLcc Customer-specific method for sparkling

wine (9 to 13% vol. alc.).

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit For the unit, an arbitrary unit can be

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

13 MIDccccc Customer-specific method for sugar-

based soft drinks (3 to 8 °Brix).

Page 58: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 58/78

12 Menu Operation

58 C26IB001EN-B

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit  An arbitrary unit with 6 digits maximum

(e.g. mg/cm3) can be selected and

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

17 CM01PROD Customer-specific method including

selectable solubility of CO2.

name To define a customer-specific name for

this method.

sel

man sel The method is available in the method

list.

disable The method is not available in the

method list.

unit  An arbitrary unit with 6 digits maximum

(e.g. mg/cm3) can be selected and

entered (see Chapter 14.2.2). Factory

default: g/L.

fu See "MINERALc".

k0 See "MINERALc".

....

k7

ks Enter the CO2 gas absorption coefficient

in g/(L * bar) of CO2 at 5 bar and 20 °C

(e.g. 1.61 g/(L * bar) for a 5±2 °Brix soft

drink).

ka

....

kd

18 CM02PROD See "CM01PROD".

...

26 CM10PROD See "CM01PROD".

Page 59: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 59/78

12 Menu Operation

C26IB001EN-B 59

12.2.6 "data memory" Submenu

Fig. 12 - 6 Menu/data memory

m e n u - d a t a m e m o r y

b r o w s e m e m o r y

c l e a r m e m o r y

browse memory To display the stored measurement

data.

Browse through the memory using

the <> and <> keys, starting with

the most recent. To indicate that the

memory browser is activated, the

name of the activated method and

"memory" are alternately displayed in

the upper left corner. A stored

measurement contains the following

information:

• Name of the selected method• Sample No.• CO2 result

• Dissolved air  • Measuring temperature or

package pressure99 measurements can be stored. The

latest entry overwrites the oldest

when the memory is full.

clear memory

clear 

 yes To clear all measuring results in the

memory of CarboQC.

 no

Page 60: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 60/78

12 Menu Operation

60 C26IB001EN-B

12.2.7 "service" Submenu

In this menu, information about the device can be displayed. None of the values

can be changed.

device

SerNo Serial number of the instrument.

FW Firmware version of the instrument.

v-k Slope parameter: counts per 1 %

increase of the volume of the

measuring chamber.

v-I0 Zero volume expansion parameter:

counts at zero volume expansion, i.e.

when measuring chamber valves

close.

v-h Slope parameter for the

determination of the pressure drop at

zero volume expansion for zero

adjustment.

cycle Total number of measurement cycles

performed.

lb0 Solubility parameters for air/nitrogen.

lb1

lb2

ccor  Internal correction factor.

t dis Initial or final (internal reference

temperature).

  offset

t Offset for the temperature sensor

(values between 0 and ±5 °C (0 and

±9 °F)).

p Offset for the absolute pressure

sensor (values between 0 and ±0.5

bar (0 and ±7 psi)).

gain

t Gain factor for the temperature

sensor (values between 0 and 2).

p Gain factor for the pressure sensor

(values between 0 and 2).

Page 61: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 61/78

13 Technical Background

C26IB001EN-B 61

13 Technical Background

The working principle of CarboQC is based on the Multiple Volume Expansion

Method and allows the determination of the true CO2 content.

Fig. 13 - 1  shows the principle of the Multiple Volume Expansion method for

eliminating the influence of dissolved air or nitrogen on the measuring result: The

CO2 content is measured at two different volume expansions of the measuring

chamber, i.e. at 10 and 30%. If the two results are identical, there is no dissolved

air or nitrogen present and therefore no correction is required. Dissolved air or

nitrogen makes the second result lower than the first. The difference between the

two results is used to calculate a correction which completely eliminates the

influence of dissolved air or nitrogen on the measuring result.

Fig. 13 - 1 Compensation of dissolved air 

CO2 with air 

CO2 without air 

Page 62: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 62/78

14 CarboQC Methods

62 C26IB001EN-B

14 CarboQC Methods

14.1 Available Methods with CarboQC

14.1.1 Standard Methods

The 7 standard methods do not allow to change any parameter at all. It can be

seen from Chapter 8.2  which method is recommended for which beverage

sample. These methods do not provide any means for compensating temperature

effects due to the headspace volume of beverage packages. They are therefore

suitable for determining the true dissolved carbon dioxide content at a given

package temperature.

14.1.2 Customer-Specific Methods

CarboQC provides 17 customer-specific methods. 7 methods, ending with "c",

use solubility coefficients identical to the standard methods, and 10 customer-

specific methods CMxxPROD also allow the entry of the solubility coefficient.

The 7 user-specific methods are named identically to the standard methods with

additional letters "cc" to obtain eight digits altogether (DIETcccc; SPARKLcc and

so on). These 7 methods allow modification of the constants k0 to k7, but these six

methods do not allow access to the solubility coefficient for CO2 in the respective

sample. The solubility coefficients in these user-specific methods are identical tothe ones in the respective standard methods. The coefficients k0 to k7 can be used

to compensate headspace related package temperature effects and to adapt the

measurement results to those based on customer-specific solubility tables or

formulae.

There are also 10 CMxxPROD methods; xx = 01, ..., 10. These 10 customer-

specific methods allow the modification of the constants k0  to k7, but also the

alteration of the solubility coefficient ks of CO2 in the sample.

14.2 Establishing Customer-Specific Formulae

14.2.1 Converting the CO2 Result into a Customer-Specific

Result

In the customer-specific methods the calculated g/L results can be converted for

user-specific CO2  tables or formulae. This conversion can also correct the

temperature-dependent changes of the dissolved CO2 content resulting from the

headspace to liquid volume ratio of the bottles or cans.

Page 63: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 63/78

14 CarboQC Methods

C26IB001EN-B 63

This equation is used for custom-tailoring a CO2  result from the original CO2

content c as determined by CarboQC.

Ccust is the customer-specific CO2 result. It is the modified carbon dioxide content

that is calculated and displayed when customer-specific coefficients are used.

k0 will always be added to the result c. The factory setting of k0 = 0.

k1 is a multiplication factor. The factory setting of k1 = 1.

k5 is a reference temperature that needs to be keyed in if the result is to be related

to a specific temperature. The factory setting of k5 is zero, which means that k5

equals 0 °C, regardless of the temperature units selected in the menu.

 A simple way of applying a reasonable temperature correction for the headspaceinfluence is to use the customer-specific methods in CarboQC, rename them

properly and set k5 equal to the average measuring temperature, and for k2 a

value of 0.003 (typical range: 0.002 … 0.004) to get the temperature

compensation to take effect.

k3 and k4 can be used for better fine tuning of the final customer-specific result.

The factory setting is 0.

Example:

If the result given by CarboQC is 3 g/L and the desired or expected result is 3.3g/L, there are two ways CarboQC can be programmed in order to transfer the

CarboQC result into the customer-specific result:

(1) k0 is set to 0.3

The value ccust will be 0.3 g/L higher than the CarboQC result. This will also affect

the zero determination with deionized water. A k0 of 0.3 will cause an offset as all

results will be 0.3 higher than the determined value: instead of 0 the displayed

result will be 0.3; instead of 2 the displayed result will be 2.3 and so forth.

Therefore setting a positive k0 offset is a reasonable means to adapt the CarboQC

results to readings of previous methods which were offset by a fairly constant aircontent of the samples.

Ccust.......................... Customer-specific CO2 result [g/L]

k0, k1, k2, k3, k4  ........ Customer-specific coefficients

k5 ............................. Reference temperature [°C]

c .............................. Original CO2 result [g/L]

t ............................... Measured temperature [°C]

Ccust k0 k1*c k2*c* t k5 –  k3*c* t k5 –  2++ k4*c

2+ +=

Page 64: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 64/78

14 CarboQC Methods

64 C26IB001EN-B

(2) k1 is set to 1.1

The CarboQC result will be multiplied by 1.1. Instead of 3 the result will be 3.3 and

instead of 2 the result will be 2.2. If the zero point adjustment works correctly, the

zero reading will be close to zero and multiplication by 1.1 will not affect the zero

reading noticeably. As an example, a factor of approximately 1.1 is often

necessary if the customer-specific solubility table applied is actually a CO2

solubility table for water, but it is applied to 10 °Brix soft drinks.

14.2.2 Converting the CO2 Concentration into a Customer-

Specific Unit

For all customer-specific methods, the measured CO2  concentration can be

converted from g/L into a customer-specific unit.

Proceed as follows:

1. In the menu "methods" , select the desired customer-specific method.

2. Select the menu item "unit"  and enter a name for the desired unit (maximum

length of 6 characters).

3. In the menu "methods" , change the coefficient "k1"   so that the desired

conversion is applied. E.g. setting k1 = 2, results in multiplying the result by 2.

4. Change to the measuring window and select the changed method using the

<Method> key.

The newly set unit is displayed in the line for the CO2  content. When

measuring, the conversion is carried out according to the entered coefficientand the calculated result is displayed.

5. The factor "fu"   is a division factor which divides a customer-specific

polynomial. E.g. using fu = 1.976 leads to converting the measurement result

from g/L into vol.

14.2.3 Using a Customer-Specific Solubility Coefficient

The 10 CMxxPROD methods allow entry of k1 to k7 and, in addition, the solubility

coefficient ks in g/(L * bar) at a temperature of 20 °C. These 10 methods need to

be used when a sample does not match the ranges as given in Chapter 8.2; e.g.

1.61 g/(L * bar) for a 5 ±2 °Brix soft drink.

Page 65: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 65/78

14 CarboQC Methods

C26IB001EN-B 65

14.3 Transformation of the Determined Air Content

The following formula serves two purposes:

1. It allows to correct the ppm result with a different slope k6.2. It allows to compensate for the temperature dependent gas distribution in the

beverage container using k5, k7. The term in brackets is set to 1 if k7  0.

This formula is used if the "air ppm" result needs to be corrected or the amount of

gas in the head space needs to be taken into consideration. ctrans is the air content

as a function of head space.

The factory setting for the "slope" k6 is 1.

cair  is the air content in ppm as determined by CarboQC.

k5 is the reference temperature and Lb2 is the solubility of air at the determined

measuring temperature T2[K].

Lb(k5) is the solubility of air at the reference temperature k5. This value can betaken from a table.

k7 is the head space to liquid ratio of the beverage container. If k7  0 then the

term in brackets equals 1; if k7 > 0 then the term in brackets is calculated.

Ctrans ........... Air content as a function of head space

k5  ............... Reference temperature [°C]

k6  ............... Slope

k7  ............... Headspace to liquid ratio of can/bottle

cair  ............. . Air/nitrogen content in ppm as measured

T2 ............... Measuring temperature [K]

Lb2 ............. Solubility of air/nitrogen at measuring temperature [-]

Lb(k5) .......... Solubility of air/nitrogen at reference temperature k5 [-]

Ctrans k6*cair *

k7*273.1

T2

--------------- Lb2+

k7*273.1

273.1 k5+--------------------------- Lb k5 +

---------------------------------------------------------------=

Page 66: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 66/78

14 CarboQC Methods

66 C26IB001EN-B

14.4 Calculation of the Package Pressure at a

Reference Temperature

The 10 CMxxPROD methods CM01PROD to CM10PROD not only allow entry of

coefficients k0  to k7  and ks, but also of the coefficients ka, kb, kc  and kd. The

coefficients ka, kb, kc and kd can be modified if the package pressure, determined

at sample temperature, needs to be converted into the package pressure at a

reference temperature.

TIP The value for the pressure at a reference temperature pref  can be obtained as

relative pressure if the ambient pressure in [bar] is keyed in. If kd  is zero, the

obtained value of pref  will correspond to the absolute pressure.

Example: Calculation of the relative reference pressure of sparkling wine at 20°C

(68°F)

1. Select the menu "methods" .

2. Choose any CMxxPROD method.

3. Re-name the method and give it a significant denotation, e.g. SPARKL20.

4. Leave "f u = 1".

5. Leave "k0"  to "k4"  and "k6"  to "k7"  unchanged.

6. Select the reference temperature "k5"  and change the entry to "+20.0000" .

When setting "k5" , always enter the numerical value of °C, even if the unit Fis selected.

7. Set "ks = 1.51"  (ks for SPARKL is 1.51 g/(L*bar)).

pref .............. Package pressure at reference temperature

pt ................ Package pressure as determined by CarboQC

ka ................ Conversion factor for pressure unit

Factory setting = [bar]. If the unit [bar] is desired, set ka = 1.

kb, kc ........... Factor for temperature compensation

Factory setting: +0.00000

Recommendation:

kb = -0.028 (negative sign!) and kc = 0.00075

kd ................ Ambient pressure in [bar]. Factory setting: +0.00000

k5 ................ Reference temperature in °Ct .................. Sample temperature as determined by CarboQC

pref  ka pt 1 kb t k5 kc t k5 2 kd –  – + – + =

Page 67: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 67/78

14 CarboQC Methods

C26IB001EN-B 67

8. Set "ka"  to "kc"  as follows:

• ka = 1• kb  = -0.028• kc  = 0.00075

9. Change the preset value of "kd"  and enter the average ambient pressure in

[bar].

Then

pref, 20 = 1 * {ptas determined at t * [1 - 0.028 (t - 20) + 0.00075 * (t - 20)²] - 1}

is the relative package pressure had it been measured at 20 °C (68 °F).

Page 68: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 68/78

 Appendix A: Technical Data

68 C26IB001EN-B

 Appendix A: Technical Data

 A.1 CarboQC

Measuring range

CO2: 0 to 12 g/L (0 to 6 vol.) at 30 °C (86 °F)

0 to 20 g/L (0 to 10 vol.) < 15 °C (59 °F)

Temperature: -3 to 30 °C (27 to 86 °F)

Pressure: 0 to 10 bar relative (0 to 145 psi)

Repeatabili ty, s. d.: 0.01 g/L (0.005 vol.) CO2

Reproducibili ty, s. d.: 0.05 g/L (0.025 vol.) CO2

Resolution: 0.01/0.001 g/L (vol.) selectable

 Accuracy of the pressure sensor: ±0.01 bar accuracy of adjustment required Accuracy of the temperature sensor: ±0.2 °C

Sample volume: > 100 mL

Typical measuring time per sample: approx. 90 seconds

Environmental conditions

 Ambient temperature: +0 to +40 °C (+32 to +104 °F)

 Air humidity (CarboQC): 10 to 95 % relative humidity

Degree of protection (CarboQC

without power supply): IP 64

Overvoltage category:

(power supply)

II

Dimensions (L x W x H): 190 x 120 x 305 mm (7.5 x 4.7 x 12 inches)

Weight: approx. 3.1 kg (6.8 lbs)

Power 

Power adapter: Input: AC 100 to 240 V, 1.5 A, 50 to 60 Hz

Output: DC 15 V, 3.0 A

Indoor use only

Battery: Nickel metal hydride

7.2 V, 1900 mAh

Integrated protection

Charging time: approx. 2 hours

WARNING

Serious injuries are possible through high voltage if the following hints are not

adhered to.

• Only connect devices to the interfaces that comply with PELV (protectiveextra-low voltage) according to EN 61140 or with SELV (safety extra-lowvoltage) according to EN 60950.

Page 69: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 69/78

 Appendix A: Technical Data

C26IB001EN-B 69

 A.2 PFD Fil ling Device

PC/printer in terface

COM1 and COM2:

RS232; 1200 to19200 Baud; 1 stop bit; 7 or 8

data bits; no, odd or even parity

Factory default settings: data bits: 7; stop bits: 1; parity: even; baud rate:9600

TXD ... transmit data

RXD ... receive data

GND ... signal ground (connected to earth)

TXD

RXD

GND

Specified for the following containers:

Can:

Glass bottle:

PET bottle:

Max. volume:

0.5 L

1 L

3 L

Compressed gas:

Necessary quality of air if used as compressed

gas supply:

6±0.5 bar (87±7 psi)

Class 5 from ISO 8573-1

max. particle size: 40 µm

max. pressure dew point: +7°C (44.6 °F)

max. oil content: 25 mg/m3

Environmental conditions

 Ambient temperature

Operation:

Transportation/Storage:

+0 to +40 °C (+32 to +104 °F)

-20 to +70 °C (-4 to +158 °F)

 Air humidity

Operation/Transportation/

storage: 0 to 100 % relative humidity

Degree of protection: IP 64

Dimensions (L x W x H):

Maximum height with piercing head in highest

position:

190 x 270 x 670 mm (7.5 x 10.6 x 26.4 inches)

1020 mm (40.2 inches)

Weight: 10.1 kg (22.3 lbs)

Page 70: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 70/78

 Appendix B: Wetted Parts

70 C26IB001EN-B

 Appendix B: Wetted Parts

The following materials are in contact with the samples to be measured and with

the cleaning agents:

CarboQC:

PFD Filling Device:

Material Part

PU (Polyurethane) Hoses

PSU (Polysulfone) Measuring chamber  

Stainless steel Inlet-outlet valve, stirrer, piston

PEEK (Polyetheretherketon) Stirrer bearing

EPDM (Ethylene propylene diene

Terpolymer)

O-rings

Material Part

PU, Nylon Hoses

Stainless steel Sample tube, guide poles

EPDM O-ring

Page 71: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 71/78

 Appendix C: Interface Commands for Communication between a PC and CarboQC

C26IB001EN-B 71

 Appendix C: Interface Commands for

Communication between a

PC and CarboQC

Measurement data stored in the memory can be transferred to a PC via the PC

interface of CarboQC.

1. Connect the CarboQC PC interface to the PC interface using an appropriate

interface cable.

2. Make sure the interface settings of CarboQC and PC agree.

3. Test the communication using a simple interface program, e.g. Windows

Terminal, Procomm or Hyper Terminal.

• For data communication software, please contact your local distributor or

 Anton Paar GmbH.

PC command CarboQC response Comments

help GetId

GetDataHead

GetDataUnit

GetData

ResetData

If the command "help" is sent

from the PC, CarboQC

responds with a list of available

commands.

get id id:00921503;CarboQC;V2.00A Readout of serial number and

firmware version.

get data head head:date;time;sample;method;

CO2;CO2;air;tm;pt;p0;condition;sol

Response. P0 = system

pressure.

get data unit unit:;;;;g/l;g/l ;ppm;°C;bar;bar;;

g/lbar

The units of the measuring

results are transferred,

separated by a semicolon. The

units are displayed according to

the units selected in the

submenu.get data no new data available

data:10.07.08;17:27; 1;

CM03PROD; 0.332; 0.332;26.9;

28.00; 1.54; 0.747;valid;1.448

No new data available.

Transfer of the first available

measuring result. One result

per entry "get data", starting

with the last result.

reset data reset data successful Resets read data to not fetched

status e.g. for second data

transfer.

Page 72: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 72/78

 Appendix D: Error Messages

72 C26IB001EN-B

 Appendix D: Error Messages

Display Cause Correction

air: "error" Air value out of specification. • Select the proper method for yoursample.

• Rinse the measuring chamber.Repeat the measurement. If the errorpersists, contact your Anton Paarrepresentative.

• Make sure that the CO2 concentration

of the sample is within specifications(Appendix A).

"CO2 - - - - - g/L" CO2 value out of range.

• Bubbles in the measuring cell. 

• The measuring chamber is nottightly closed duringmeasurement.

• Make sure that the measuringchamber is free of bubbles.

• Contact your Anton Paarrepresentative.

"air: - - - - - ppm" Air value is not determined, e.g if

"compensation"  is set "off" .

 Activate the compensation in the menu

"adjustment > compensation" .

Page 73: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 73/78

 Appendix E: Troubleshooting

C26IB001EN-B 73

 Appendix E: Troubleshooting

Q The CO2 reading of the water check with deionized water shows >0.03 g/L CO2.

 A Sometimes, deionized water contains absorbed CO2 and displays the amountof CO2 absorbed. Either heat up and boil the water and cover it up while cooling

down or do the water check with tap water – but do not forget to rinse the

measuring chamber with deionized water afterwards!

Q The CO2 reading of the first sample after a water check is low.

 A If you go from samples containing very low to very high CO2 contents, the first

result for the high CO2 content is slightly lower than the following ones. If only

limited amounts of sample are available, it is a good idea to condition the

instrument by e.g. measuring mineral water after the water check before

measuring the first sample.

Page 74: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 74/78

 Appendix F: Firmware Versions

74 C26IB001EN-B

 Appendix F: Firmware Versions

Firmware

version

Date of

release

Document

number Comments

V 1.00A 09.03.2004 C26IB02A First released firmware.

V1.00B 16.04.2004 C26IB02B Zero adjustment improved (v-h).

V1.00C 09.09.2004 C26IB02C • Measuring times at 10% volume expansion and30% volume expansion optimized.

• Total pressure display inside containerintroduced.

• Printout includes total pressure inside container.• Piston slows down before returning to starting

position.

V1.00D 15.02.2005 C26IB02D • First released firmware version for PBA-S.

V1.00E 24.06.2005 C26IB02E

C26IB02F

• Improved polynomial function for calculating theair content.

V2.00A 28.07.2006 C26IB02G • New microcontroller for increased performance.• Firmware update simplified.• New method "MID" for soft drinks with 3 to 8

°Brix.• 10 instead of 4 "CMxxPROD" methods.• Arbitrary selection of CO2 units.

• Selectable if methods are displayed in themethod list ("man sel/disable" ).

• Semiautomatic adjustment of the pressureoffset in the menu "adjustment" .

• Data memory can be viewed without entering apassword.

V2.00B 28.07.2008 C26IB02H • Simplified firmware update for future firmwareversions.

• Calculation of package pressure at referencetemperature.

V2.00C 13.01.2010 C26IB02I • Improved display controller.

V2.00D 19.05.2010 C26IB001EN-A • Rinse time selectable up to 200 seconds.• Numbers added to the methods.

V2.00E 10.05.2012 C26IB001EN-B • New function after long push of the <Stop> key:Method selection.

• New key combination: When <Start> and<Stop> are simultaneously pressed, a rinsingprocedure is started.

Page 75: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 75/78

 Appendix G: Declaration of Conformity

C26IB001EN-B 75

 Appendix G: Declaration of Conformity

Page 76: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 76/78

 Appendix H: Menu Tree

76 C26IB001EN-B

 Appendix H: Menu Tree

adjustment compensation comp air: on | off  

vexp: 02, ..., 40 %

auto p-offset CO2: x.xxx g/l

p: x.xxx bar (psia)

instrument rinse time: 03, ..., 60 sec

unit t: °C | °F

p: bar | psia

CO2: g/l | vol

CO2dec: 2dec | 3dec

serial com baud: 9600 | 4800 | 2400 | 1200 | 19200

format: E71 | N81 | O71

password new: 0000

date & time date: ddmmyy | mmddyy

year: xx

month: xx

day: xx

time: 24hrs | am/pm

hour: xx

min: xx

battery service: off | on

device info print: off | printer | PC

methods 01 MINERALc03 DIETcccc

05 BEERcccc

07 ST.BEERc

09 COLAcccc

11 SPARKLcc

13 MIDccccc

name: MINERALc (xxxxxxxx)sel: man sel | disable

unit: g/l (xxxxxx)

fu, k1, ..., k7

17 CM01PROD, ...,

26 CM10PROD

name: MINERALc (xxxxxxxx)

sel: man sel | disable

unit: g/l (xxxxxx)

fu, k1, ..., k7, ks, ka, ..., kd

00 MINERAL

02 DIET04 BEER

06 ST.BEER

08 COLA

10 SPARKL.

12 MID

name: MINERAL

sel: man sel | disable

data memory browse memory

clear memory clear: no | yes

service device

offset

gain

Page 77: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 77/78

Index

C26IB001EN-B 77

Index

 A

 Absorption coefficient 58 Activating lever 30, 40 Air content 65 Air humidity 68, 69 Ambient temperature 68, 69

B

Battery 17, 19, 42, 48, 54, 68Battery status 17Bubbles 30, 39, 41, 46, 49

C

Can 28CO2 content 53, 61, 62CO2 result 34, 36, 53, 55, 59, 63Coefficient 55, 63Compressed gas 11, 20, 69Compressed gas supply 30Customer specific 18, 27, 55, 56, 57, 62

D

Date 25, 54Deionized water 36, 38, 39, 40Display 17, 41, 59, 72Dissolved air 52, 59, 61

E

Exhaust 23

F

Filter 14, 22, 23, 39, 40Flow valve 28

G

Glassware cleaner 36, 38, 39, 40, 41

H

Headspace 26, 55, 62Hose 14, 70

K

Keypad 18, 41, 51

M

Marking 28Measurement 27, 30, 46, 49, 60Measurement data 59, 71Measuring chamber 16, 18, 30, 38, 39, 40, 41, 46, 49, 53, 60, 61, 70Measuring mode 18, 51Measuring range 68Measuring temperature 59, 65Memory 18, 30, 46, 49, 59, 71Menu 18, 25, 51, 52Menu access 52, 54Menu mode 18, 25, 51Method 18, 30, 49, 55, 56, 57, 59, 62Multiple Volume Expansion method 61

N

Nitrogen 56, 60, 61, 65

O

Offset 60On/off switch 24

P

Password 25, 52, 54PC interface 53, 69, 71PET bottle 28, 40PFD 27, 28, 38, 39, 40, 69Piston 30, 41, 70Power adapter 14, 17, 23, 54, 68Pressure 23, 25, 30, 41, 46, 49, 53, 68Pressure sensor 53, 60Print 18, 30, 54

Printer interface 53Printout 18Purified water 38

R

Reference pressure 66Rinse 18, 36, 38, 39, 40, 53

S

Sample flow 36Sample inlet 23Sample number 18, 49Sample outlet 23

Page 78: Manual CarboQC

7/17/2019 Manual CarboQC

http://slidepdf.com/reader/full/manual-carboqc 78/78

Index

Sample pressure 42, 48Sample residues 38, 39, 40Sample tube 30, 40Sample volume 68Service 19, 54

Slope 60, 65Solubility 60, 65Solubility coefficient 55Start key 18Stirrer 18, 30, 41, 49, 70Stop key 18Supply inlet 24, 54

T

Temperature 25, 27, 53, 63, 65Temperature sensor 60

Time 25, 54

U

Unit 25, 53

V

Volume expansion 52, 60, 61

W

Waste vessel 22, 46, 48

Z

Zero point 36