Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE...

28
Managing Radiation Dose: Managing Radiation Dose: IHE REM Profile IHE REM Profile Kevin O’Donnell Kevin O’Donnell Toshiba Medical Systems Toshiba Medical Systems Co-chair, IHE Radiology Planning Co-chair, IHE Radiology Planning Cmte Cmte Member, DICOM Standards Cmte & WG- Member, DICOM Standards Cmte & WG- 6 6 Member, MITA X-Ray Section Member, MITA X-Ray Section

Transcript of Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE...

Page 1: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

Managing Radiation Dose:Managing Radiation Dose:IHE REM ProfileIHE REM Profile

Kevin O’DonnellKevin O’DonnellToshiba Medical SystemsToshiba Medical Systems

Co-chair, IHE Radiology Planning CmteCo-chair, IHE Radiology Planning CmteMember, DICOM Standards Cmte & WG-6Member, DICOM Standards Cmte & WG-6Member, MITA X-Ray SectionMember, MITA X-Ray Section

Page 2: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

2RSNA 2008RSNA 2008

MotivationMotivation

X-Ray based imaging can provide tremendously useful information

Patient Dose is an important consideration

Potential benefit > potential risk

… but the risk should still be managed

Page 3: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

3RSNA 2008RSNA 2008

Why Measure / MonitorWhy Measure / Monitor

“I think patientdose is improving.”

“I think it’sgetting worse.”

Managing in the presence of data is far better and easier than managing in its absence. – Robert Glass

Page 4: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

4RSNA 2008RSNA 2008

Why Measure / MonitorWhy Measure / Monitor

Technology is constantly changing

Technique is constantly evolving

Understanding continues to improve

Page 5: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

5RSNA 2008RSNA 2008

Why Measure / MonitorWhy Measure / Monitor

ALARA Guidelines for Physicians (Patient Dose) As Low As Reasonably Achievable

Patient Dose < = Reasonably Achievable Dose

A < = B

Should be easy / automatic / routine

Can we make the data readily available…

Page 6: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

6RSNA 2008RSNA 2008

… … A New IHE ProfileA New IHE Profile

What’s an IHE Profile?

Page 7: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

7RSNA 2008RSNA 2008

IHE in One SlideIHE in One Slide

IHE helps vendors implement & test functions that span multiple systems

Profiles are implementation guides how to use existing standards to address a specific problem scenario

Connectathons are test events managed testing of Profile implementations

IHE helps users purchase & integrate multi-system solutions list required IHE Profile support in RFPs

Page 8: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

8RSNA 2008RSNA 2008

IHE ConnectathonIHE Connectathon

One Week;

70+ vendors;

350+ engineers;

Thousandsof tests;

Managed process;

Find & resolve issues.

Results @www.ihe.net

Page 9: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

9RSNA 2008RSNA 2008

IHE Radiation Exposure Monitoring Profile

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

National Registry

Archive

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

Numer 12.2 14.5 9.5 10.9

Dose Analysis& Reporting

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9

Outlier: #Performing Phys.Over Target: 12.2%

Outlier: #Performing Phys.Over Target: 12.2%

Outlier: #Performing Phys.Over Target: 12.2%

Page 10: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

10RSNA 2008RSNA 2008

A New IHE ProfileA New IHE Profile

IHE Radiation Exposure Monitoring Profile Integration of systems reporting dose and systems

which receive, store, or process those reports Modalities, PACS, RIS, Workstations, Registries

Facilitate compliance with Euratom 97/43,ACR Guidelines, etc.

Directly based on DICOM Dose Reports Creation, Collection, Distribution, Processing

Page 11: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

11RSNA 2008RSNA 2008

DICOM Dose ReportsDICOM Dose Reports

“SR Objects” – DICOM Structured Reports Easily ingested (and regurgitated) by PACS

Granularity : “Irradiation Event” + Accumulated Dose over Study, Series

Templates: CT, Projection X-Ray (Mammo, Fluoro)

Not addressed: NM, RT

Page 12: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

12RSNA 2008RSNA 2008

What about using MPPS?What about using MPPS?

Modality Performed Procedure Steps Status messages back to RIS / PACS

Transient Information designed for workflow; not persistent archiving

Lacks complete dose details

Page 13: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

13RSNA 2008RSNA 2008

What about Image Headers?What about Image Headers?

Mostly Exposure details (for interpretation)

No image; no dose deleted due to quality / patient motion

More images; more dose extra reconstructions; post-processing

Lacks complete dose details

Page 14: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

14RSNA 2008RSNA 2008

What To MeasureWhat To Measure

So many choices… Exposure, Dose, Dose Index, Estimated Dose,

Effective Dose, Estimated Effective Dose, … CTDIw, CTDIvol, CTDI100, CTDIfreeair, …

Solution: Collaborated with IEC Subcommittee 62B Established a baseline & a pipeline Upgrade when necessary

Page 15: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

15RSNA 2008RSNA 2008

Key MeasurementsKey Measurements

CT Dose DLP, CTDIvol, kVP, mA, sec Effective Dose [ Optional ]

(Reference estimation method)

Projection X-Ray Dose DAP, Dose@RP, kVP, mA, sec Fluoro Dose, Fluoro Time

Mammography Dose AGD, Entrance Exposure @ RP, kVP, mA, sec Compression, Half Value Layer

ftp://medical.nema.org/medical/dicom/2008/08_16pu.pdf

ftp://medical.nema.org/medical/dicom/final/cp874_ft.pdf

Page 16: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

16RSNA 2008RSNA 2008

Other Details in Dose SROther Details in Dose SR

Full Patient / Order / Study Details

Unique ID for each Irradiation Event

Equipment ID

Patient Position, Anatomy imaged

Imaging Geometry (projection)

Collimation

X-Ray Filters, Anode Target Material

Calibration, Phantom, Dosimeter

Page 17: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

17RSNA 2008RSNA 2008

DICOM CP 874DICOM CP 874

Updated details for REM Implementers Will apply to 2009 Connectathon

Minor adjustments to DICOM templates Require identification of irradiating system Require Target Region Optional references to persons who ordered / performed Code to differentiate interventional procedures from diagnostic etc.

Page 18: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

18RSNA 2008RSNA 2008

Creating SR Dose ReportsCreating SR Dose Reports

Modalities CT, XR, XA, MX

“Readers” CR, DX, Film Digitizers

3rd Party Workstations

RIS, PACS

Page 19: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

19RSNA 2008RSNA 2008

Storing SR Dose ReportsStoring SR Dose Reports

Usually stored in the Study folder Archive, Backup, Reconciliation Query / Retrieve

Can be sent to other destinations

Page 20: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

20RSNA 2008RSNA 2008

Using SR Dose ReportsUsing SR Dose Reports

Possible applications:

Radiation QA Periodically Query / Retrieve Reports from Archive Set policies/standards and flag deviations Set goals for improvement and track; Implement protocol changes and compare difference in dose

Patient Impact Evaluation e.g. if Patient identified as pregnant post-facto

Dose Mapping Store data in realtime from Modality to Mapping Workstation

Page 21: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

21RSNA 2008RSNA 2008

Using SR Dose ReportsUsing SR Dose Reports

National Registries Anonymize and submit Dose Reports to Register Compile Population Risk Estimations Derive Dose Reference Levels Provide Site-Site Comparisons

Individual Dose Record Collect Dose Reports over time

Clinical Trials Collect Dose together with Images Demonstrate both improved detection &

reduced dose

Page 22: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

22RSNA 2008RSNA 2008

REM Profile: Actors & TransactionsREM Profile: Actors & Transactions

Page 23: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

23RSNA 2008RSNA 2008

2009 Connectathon Participation2009 Connectathon Participation

Vendors Planning to Test IHE REM Profile

GE

Siemens

Toshiba

Results will be posted: http://ihe.univ-rennes1.fr/con_result/

CMI

Infimed

Krucom

Page 24: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

24RSNA 2008RSNA 2008

What to Ask ForWhat to Ask For

IHE Radiation Exposure Monitoring Profile (REM) Check Product: IHE Integration Statement

DICOM X-ray Radiation Dose SR Storage

• CT Radiation Dose Template, or• Projection X-Ray Radiation Dose Template

Check Product: DICOM Conformance Statement

Page 25: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

25RSNA 2008RSNA 2008

TakeawayTakeaway

Start planning: How will this support your Radiation Safety Plan

IHE REM Profile www.ihe.net www.ihe.net/Technical_Framework/index.cfm#radiology

Page 26: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

26RSNA 2008RSNA 2008

Questions ?Questions ?

Page 27: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

27RSNA 2008RSNA 2008

TimelineTimeline

Nov. 2007 DICOM CT Dose Reports published (DICOM Proj. X-Ray Dose Reports already published)

June 2008 IHE REM Profile published http://www.ihe.net/Technical_Framework

Aug. 2008 Vendors sign up to test REM Implementations

Feb. 2009 Connectathon Testing of REM Profile (Chicago) Results posted: http://ihe.univ-rennes1.fr/con_result/

Page 28: Managing Radiation Dose: IHE REM Profile Kevin O’Donnell Toshiba Medical Systems Co-chair, IHE Radiology Planning Cmte Member, DICOM Standards Cmte & WG-6.

28RSNA 2008RSNA 2008

Users

IHE

Vendors

RFP

(Profiles &Actors)

Standards

(DICOM,HL7, etc)

Standards

(DICOM,HL7, etc)

Standards

(DICOM,HL7, etc)

IHEIntegrationStatement

(Profiles &Actors)

Results

SelectIHE

Profiles

IHEIntegration

Profiles

IHEIntegration

Profiles

IHEIntegration

Profiles

IHE TechnicalFramework

Productwith IHE

IHEConnectathons

IHEDemonstrations

Real-World

Integration

Problems

The IHE ProcessThe IHE Process