Maldi

59
Biological Mass Spectrometry A mini course

Transcript of Maldi

Page 1: Maldi

Biological Mass Spectrometry

A mini course

Page 2: Maldi

Biological Mass SpectrometryGeneral Mass Spectrometry

The General mass spectrometerAnalyzers typesCalibrations and mass AccuracyDetection

Matrix Assisted Laser Desorption (MALDI) TOF

Electrospray Ionization (ESI)Quadrupoles, Ion Traps, FTMS, and O-TOF

FragmentationPost source dissociation (PSD) (with MALDI)In source dissociation (with ESI)Collision induced dissociation CIDIRMPD and ESI

Page 3: Maldi

The General Mass Spectrometry

Ion S ou rceM A L D I

S p ray S ou rces

A n a lyzerTO F

Q u ad ru p o lesTrap p in g in s tru m en ts

D etec to r

M ass S p ec trom eter

Page 4: Maldi

Analyzers

• Time of Flights mass spectrometer

• Quadrupole mass spectrometer

• Quadrupole Ion traps

• FT-MS, FT-ICR

Page 5: Maldi

Time-of-flight Mass Spectrometers

• Basically a tube that ions travel through with lower mass ions traveling quicker through the tube.

• Reflectron increases resolution by compensating for energy and increasing path length.

• Delayed extraction increases resolution.

Page 6: Maldi

O-TOF

Ion BeamPulser

Grid

To TOF

Page 7: Maldi

Quadrupole Mass Spectrometer

+

+

-- ++

-

-

Rods

Ions

Page 8: Maldi

Quadrupole Ion Traps

+

+

-- ++

-

-

Ions

End Caps

Ring

Page 9: Maldi

FT-MS, FT-ICR

B+

+

Page 10: Maldi

Calibration

• All Mass Spectrometers MUST BE CALIBRATED. This adjust parameters in the mass assignment of the ions.

• In general, the closer in space and time the calibration is the more accurate the calibration.

• Default calibrations are usually set up on instruments for nominal mass accuracy.

• Calibrations can be done with any compound with a known elemental formula. The best calibration compounds are similar type of compounds to the type you are going to measure accurately.

• It is best to have calibration compounds above and below the compound of interest

Page 11: Maldi

Mass Accuracy

• Higher resolution allows for higher mass accuracy in two ways– Resolves stardards and sample from background– Better centroiding of peaks

• Use of internal standards with a high resolution (>10,000) is the “Gold standard” for mass accuracy.

• Mass Accuracy needed is dependent on use and available information.

Page 12: Maldi

R Voltage across resistor

Ion Detection

Page 13: Maldi

WHY MALDI?

• Less Sensitive to Salts• Lower PRACTICAL detection limits• Easier to interpret spectra (less multiple

charges)• Quick and easy• Higher mass detection • Higher Throughput (>1000 samples per

hour)

Page 14: Maldi

Matrix

• Needs to be involatile (most are solids at room temperature)

• Needs to absorb the laser wavelength that you are using. (most cases 337 nm)

• Preferably dissolves in same solvent as the sample

• Typically, the matrices are acidic.

Page 15: Maldi

HOT and Cold Matrices

• DHB is a cold matrix, the samples are not as likely to be fragmented, may not ionize some molecules.

• Alpha-cyano dihydroxybenzoic acid is considered a hot matrix. More likely to fragment the molecules. Can produce multiply charged proteins.

Page 16: Maldi

Sample preparation

• Dried Droplet- mix sample with matrix and drop on plate at the same time. –Easy to do.

• Layered method- put matrix on plate then dry before adding sample. Good for low concentrations, more difficult.

• Want high Matrix-Analyte ratio.

Page 17: Maldi

The MALDI Instrument

S ou rceD e layed E xtrac tion

A n a lyzerR eflec tron

P S D

D etec to rW h y d e flec t th e ion s?

Tim e-o f F lig h tM ass S p ec trom ete r

Page 18: Maldi

Basic Physics

• Kinetic Energy= ½ mv2

• Potential Energy of charged particle=qV

• Potential Energy +Kinetic Energy=Constant

Page 19: Maldi

Delayed Extraction

• Kinetic Energy of ions leaving the surface is not constant. Also, their could be a time delay, in addition.

• Delayed Extraction is a method used to compensate for the spread of kinetic energy of the ions leaving the surface.

• Delayed Extraction is mass dependent.• Delayed Extraction will also compensate for some

time delay of ions coming off the surface.

Page 20: Maldi
Page 21: Maldi

M1 kE < M2 kE

Page 22: Maldi
Page 23: Maldi
Page 24: Maldi

Pote

ntia

l Volta

ge

Sam

ple

pla

te

M1M2

T=t4

E=kE +PEM1 kE< M2 kE

M1 kE > M2 kE

Dete

cto

r

Page 25: Maldi

Reflectron

• The reflectron is another method used to compensate for the spread in Kinetic energies in the time of flight.

• Reflectron does not compensate for any time delay coming off the surface.

• Three types of reflectrons ( single stage, two stage, and curved field)

Page 26: Maldi

M1M2

KE M1 > KE M2

Page 27: Maldi

M1

M2

KE M1 > KE M2T=t2

Page 28: Maldi

M1

M2KE M1 > KE M2

T=t3

KE M2=KE M1=0

Page 29: Maldi
Page 30: Maldi

T=5

Page 31: Maldi

Why Deflect the ions?

• Detector is made up of multiple detectors similar to the electron multiplier called “channels”.

• When an ion strikes the detector the “channel” is dead for milliseconds.

• When to many matrix ions hit the detector, it kills the signal.

• Linear detector is specially design to compensate for larger number of ions, but is not perfect.

Page 32: Maldi

Why Electrospray Ionization?

• Electrospray Ionization can be easily interfaced to LC.

• Absolute signals from Electrospray are more easily reproduced, therefore, better quantitation.

• Mass Accuracy is considered better.• Multiple charging is more common then

MALDI.

Page 33: Maldi

Detection of Peptides

MALDIDetectedPeptides

Page 34: Maldi

Advantages of Multiple Charging

• Can use instruments with lower maximum m/z (i.e., Quadrupoles, ion traps, FTMS)

• For FTMS, the resolution is better at lower m/z values, therefore, ESI helps one obtain better resolution at higher m/z values.

• Multiply charge ions tend to fragment easier then singly charge ions.

Page 35: Maldi

To Pumping

- 3kV

Basic ESI Source

Page 36: Maldi

Needle options

• High flow needles (for connection to regular LC columns(up to 1ml/min))

• Nanospray needles (high sensitivity work)• Microspray needles (connected to micro LC

columns)• Atmospheric Pressure Chemical Ionization

(for non-polar molecules)• Multiple needles setups

Page 37: Maldi

Why Fragment the ions?

• Molecular weight is only one piece of data the instrument can provide

• Data base searches on tryptic peptides do not always reveal the protein.

• Sequence and modification information can be obtained by fragmentation

Page 38: Maldi

Ways to Induce Fragmentation

• Post Source Decay-MALDI

• In source Dissociation-ESI

• Low Energy –High Energy Collision induced dissociation

• Inferred multi-photon dissociation (in a trap)

• Electron Capture dissociation (in a trap)

Page 39: Maldi

Tandem Mass Spectrometers

• Traps

• Methods using single MS’s

• Triple Quadrupoles

• Q-Tofs

• TOF-TOF

Page 40: Maldi

PSD

• Easily done on simple MALDI with reflectron

• Might not produce enough fragments

• Relatively low precursor ion selection

Page 41: Maldi

Physics Reminder

• Conservation of Energy assures that the kinetic energy of fragments is determined by the relative masses of the products.

• KE=1/2mv2

• KEprecursor=KEproduct ion+KEproduct neutral

• vprecursor=vproduct ion

• KEproduct ion=KEprecursor*(mproduct ion/mprecursor)

Page 42: Maldi

M > M 2> M 2

K > 2> K 21

Page 43: Maldi

M > M 2> M 2

K > 2> K 21

Page 44: Maldi

M > M 2> M 2

K > 2> K 21

Velocity in Field FreeRegion on the samefor both fragments

Page 45: Maldi

Types of Reflectrons• Two stage- Give best resolution on single MS

mode, requires doing PSD in Stages.• Single Stage-Large Reflector takes a significant

amount of space. PSD can be done in single stage, however, low mass product ions have poor resolution.

• Curved field-Another Large Reflector that takes up space. Allows PSD to be obtained in single stage with good resolution throughout PSD, but single MS mode does not have as good a resolution.

Page 46: Maldi

In Source Dissociation

• This is done by increasing the voltage between the entrance to the sample orifice and the first skimmer.

• Can be done on simple electrospray instrument.• Produces results similar to low energy CID• No mass selection, therefore, requires sample to

be extremely pure.• Can be used to produce multiple MS on triple and

hybrid instruments.

Page 47: Maldi

CID

• CID is divided into low energy (<100 eV) and high energy(>1000 eV) based on the collision energy.

• High energy CID produces more fragments, but is more complicated to interpret.

• Low energy CID has a limit on the m/z it can dissociate of approximately 1000.

• High energy CID produces charge remote fragmentation that can be used to distinguish leucine and isoleucine.

Page 48: Maldi

IRMPD and ECD

• Complimentary techniques

• Can only be performed in trapping instruments-Mostly FTMS.

• ECD requires a multiply charged ion

• Used for dissociating high molecular weight fragments

Page 49: Maldi

Interpretation of Tandem Mass Spectra of Peptides

• Known Sequence- Calculate expected fragments and compare to tandem spectra to see match

• Modified Sequence-Calculate unmodified sequence compare to tandem spectra to see difference where modification occurs.

• Unknown Sequence- Check Database to see if it is a match

• Unknown Sequence not in Database- Manual Interpretation (Practice!Practice!Practice!)

Page 50: Maldi

Peptide fragmentation

-RCH-C-NH-CHR-O

x y z

cba

Nterm Cterm

Page 51: Maldi

Manual Interpretation

• Goal-Assign as many abundant fragments as possible to a spectra

• Remember Cysteine modifications• Know type of fragments that are typically

observed by dissociation method.– Low Energy (b and y, loss of neutrals from these

fragments)– High Energy (x,y,z, a,b,c, v, d,w)Reference: Ioannis A Papayannopoulos, Mass Spec.

Rev.,1995, vol. 14, 49-73.

Page 52: Maldi

Steps in Manual Interpretation• Choose a large peak• Look for ions that are different in mass from the

fragment by a specific amino acid mass. If there is look for an other fragment that differs by a specific amino acid mass to build a partial sequence.

• Decide the type of fragments that corresponds to the above sequence by either mass difference to the precursor or complimentary fragments.

• Look for other complimentary fragments and low mass ions to confirm the partial sequence.

• Choose another unassigned peak and repeat procedures until sequence is determined.

Page 53: Maldi

Peptide modifications to helpInterpretation

• O18 water digestion- all C terminal fragments will have unique Isotope pattern

• N-terminal or C-terminal modifications-results in distinct fragmentation patterns.

Page 54: Maldi

LC Interfaces

• Direct LC with Electrospray.– Direct connection to mass spectrometer– Real Time monitoring

• Vacuum deposition for MALDI– Indirect conection to mass spectrometer– Not Real Time, allows one to make decisions

later. Multiple MS/MS on single LC peaks.

Page 55: Maldi

LC detection methods

• Simple acquiring of spectra.

• Precursor and product ion scans

• Data Dependent scans

• Selective Ion Monitoring

• Selective Reaction monitoring

Page 56: Maldi

Applications

• Protein Identifications

• Protein Modifications

• Bacteria Identifications

• SNP’s

Page 57: Maldi

Protein Identification

• Identify Protein mass• Digest Protein• Analyze by MALDI• Data Base Search• If negative results, go to Tandem Mass

Spectrometry• Data Base Search on Tandem Mass

Spectrum

Page 58: Maldi

Protein Modifications

• Tandem Mass Spectrometry is very important• Can obtain full mass of proteins to help determine

mass and number of modifications• Digest proteins and run mass spectrometry.• Use Tandem to determine the exact position• Use of unmodified protein or peptide can be useful

in the interpretation

Page 59: Maldi

Bacteria Identification

• Uses MALDI of Bacteria to Identify Proteins produced by bacteria

• Data base search of MALDI spectra identifies bacteria

• At this point, no large consistent Data Base is available.