Preventive Maintenance Maintenance CourseCourse
• Inventory
• Menus
• Calibration
• Preventive Maintenance
• Miscellaneous Features
Course Overview
Power Supply Battery and Charger
Mode Module
Rear Panel
Sample and Calibration Traps
Inventory
Compressed AirCleaning Swabs
Dryer
MembranesDopant
Filters
O-Rings and Fuses
Inventory
Inventory
Customer Service• USA – 866-517-5580 (0830 to
1700 EST)
Technical Service• USA – 800-856-1716
Inventory
How many 90-minute batteries come with the VT2?
What power options are there for the VT2?
What is used to clean components on the VT2?
Operator mode 0 0 0
Administrator PIN 9 1 2
9 1 2
MENU TO QUIT
- - - ENTER CODE - - -
- ENTER - TO ACCEPT
Administrator Log-onMenus
MENU2 CAL ON SUBSTANCEINT l 8SMP l 0PLASMAGRAM
MENU3 ADJUST WINDOWSCAL FLOWSAVE PKS RAWCONNECT WIRE IR
Administrator MenusMenus
Menus
What is the Administrator log-on number?
What are some features of the Administrator mode?
How do you return to Operator mode?
Why calibrate?• Calibration introduces a
known substance into the detector for identification.
• After the VaporTracer2 identifies the calibrant, the software will adjust the positions of all substances.
• The VaporTracer2 must be calibrated for accurate detection.
Calibration
When do you calibrate?• After powering up• After changing modes• At the beginning of each
shift• When the unit is moved • At higher elevations• When atmospheric
conditions change (humidity, pressure)
Calibration
Press to begin calibration.Press to begin calibration.
CAL REQUIRED TNT _ _NIT _ _RDX _ _PET _ _HMX _ _AM _ _
MENU TO CONTINUE
Calibration
Use to highlight Use to highlight CAL EXT, , then press ..
MENU 1 SELECT SUBSTANCEBEEPER OFFLIGHT ONMETHOD VAP
INT
ONOFF
CAL EXT
PART
Calibration
INSERT CAL TRAP TIME STD LAST 6.169 6.134
-MENU- TO EXIT
-START- TO CAL
Insert the calibration trap into the desorber and then press .
Calibration
Press to accept the highlighted peak.
STD 6.16 LAST 6.13 HEIGHT TIME
-ENTER- TO ACCEPTGET EXPANDED LIST
5892 6.154
Calibration
• If the proper peak is not highlighted, use to select it.
• The correct peak is the peak closest to the STD peak (usually between 6 and 8 ms).
STD 6.16 LAST 6.13 HEIGHT TIME
-ENTER- TO ACCEPTGET EXPANDED LIST
6328 6.159 234 3.289 567 9.583
Calibration
CALIBRATION DONE
-ENTER- TO CLEAR
TO VERIFY THE CALIBRATION INSERT VERIFICATION TRAP THEN PRESS START
Verification
• When Calibration is done, insert a Verification trap (marked with “V”) into the desorber and press START.
Calibration
VERIFY SUCCESSFUL
REMOVE TRAP AND PRESS ENTER TO CLEARDOWN
VERIFY FAILURE
REMOVE TRAP AND PRESS ENTER TO CLEARDOWN
THEN CALIBRATE AGAIN
READY P+ 10 100 1k 10kTNT --NIT --RDX --PET --HMX --AM --
The READY screen displays after CLEARDOWN.
Calibration Verification Results
Calibration
Why is calibration necessary?
Which menu is used to access calibration?
How do you verify calibration?
If a calibration peak is not highlighted, which arrow do you use to select the proper peak?
When do you calibrate?
• Shift
• Weekly
• Monthly
• Annual
Preventive Maintenance
ITMS Detector Schematic
Preventive Maintenance
Detector Schematic
Dryer
Desorber
Display Screen
Dopant
Membrane
Preventive Maintenance
Power-down the unit. Wear white cotton gloves. Clean the desorber. Clean the membrane. Calibrate.
Shift MaintenancePreventive Maintenance
Power-down the unit. Wear white cotton gloves. Change the dryer. Inspect dryer tube o-rings. Clean the fan filter. Inspect the dopant. Calibrate.
Weekly MaintenancePreventive Maintenance
Power-down the unit. Wear white cotton gloves. Perform Weekly
Maintenance. Change the membrane. Power-on the unit. After 4 hours, calibrate.
Monthly MaintenancePreventive Maintenance
Wipe around the entire nozzle area.
Wipe around the entire seam between the top and bottom housing.
Annual Maintenance - Radiation Wipe TestPreventive Maintenance
When is preventive maintenance required?
How many hours do you wait to calibrate after monthly maintenance is completed?
What are the steps for weekly maintenance?
Are wearing cotton gloves optional?
Preventive Maintenance
• A substance must give off a sufficient amount of vapor to be detected.
• Heat increases the amount of vapor that is emitted.
• The vapor must be contained (vapor release point).
Vapor SamplingMiscellaneous Features
• After attaching the vapor plate, the unit will need 15 minutes to adjust temperatures.
• Perform calibration.• MENU 1 Highlight CAL INT.• Verify calibration using the
Check Sample.
Vapor Calibration and VerificationMiscellaneous Features
Power-down the unit. Remove the compartment
cover. Plug in and insert the module. Reinstall cover. Power-on the unit. Calibrate.
Changing ModesMiscellaneous Features
How is a mode module inserted into the VT2?
What is the vapor plate?
How do you verify vapor calibration?
Miscellaneous Features
• Inventory
• Menus
• Calibration
• Preventive Maintenance
• Miscellaneous Features
Course Review