Magic is in the air:Monitoring the respiratory system
C. PutensenDepartement of Anesthesiology and
Intensive Care Medicine
![Page 2: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/2.jpg)
![Page 3: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/3.jpg)
![Page 4: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/4.jpg)
Estimation of thetransvascular volumeflux
increased permeability ?
volume overload/hydrostatic
• ITBV ↓⊥ , EVLW ↑
• ITBV ↑, EVLW ↑
edema ?
![Page 5: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/5.jpg)
Am J Cardiol 1999;84:1158–1163
![Page 6: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/6.jpg)
4h
Fluid restriction in ARDS
PaO12h2/FiO2
Baseline74±6 85±6
92±518 ± 2 18 ± 2PEEP; cm H2O
18 ± 2Pei; cm H2O 28 ± 3 28 ± 3
28 ± 33,8 ± 0,7 3,6 ± 0,6CI, l/min/m2
2,8 ± 0,7101 ± 7HR; /min
110 ± 7 122 ± 818 ± 3
34 ± 3
19 ± 3
17 ± 4
31 ± 5
18 ± 3
CVP; mm Hg13 ± 3
PAP; mm Hg31 ± 4
PAOP; /min18 ± 3
![Page 7: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/7.jpg)
EVLW
I (m
l/kg)
ITBVI (ml/m2)200 400 600 800 1000 1200
28
26
24
22
20
18
16
14
12
10
Fluid restriction in ARDS
R = 0.75 R2 = 0.57
![Page 8: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/8.jpg)
What is evidence based?
![Page 9: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/9.jpg)
O.3
O.4
O.4
O.5
O.5
O.6
O.7
O7
O.7
O.8
O.9
O.9
O.9
1.0
1.0
1.0
1.0
5 5 8 8 10 10 10
12
14 14 14 16 18 18
20
22
24 cm H2O
FiO2
PEEP
What is evidence based?
ARDS Network. NEJM 2004
PEEP/FiO2titration
VTPei
RRpH
=≤
≤→
6 ml/kg pBW30 cm H2O
35 /min> 7,15
SaO2 > 90%
![Page 10: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/10.jpg)
Distension of the lungs
![Page 11: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/11.jpg)
Sres
s (m
bar)
The mechanical distension of thelungsCollagen and Elastin
≈100
50
40 80
Strain (%)
FRC
![Page 12: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/12.jpg)
Stress and strain of the lungs
Stress ≈ transpulmonal pressure (PTP)
Strain ≈ VT / FRC
The linkage is specific regional compliance
PTP = Espec *
Barotrauma
VT
FRC
Volotrauma
Chiumello D, Carlesso E, Cadringher P, Caironi P, Valenza F, Polli F, Tallarini F, Cozzi P, Cressoni M, Colombo A, Marini JJ, Gattinoni L.Lung stress and strain during mechanical ventilation for acute respiratory distress syndrome.Am J Respir Crit Care Med. 2008;178:346-55.
![Page 13: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/13.jpg)
FT
min maxMead J et al. J. Appl. Physiol. 28(5):596-6081970
Stress distribution - homogenous system
![Page 14: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/14.jpg)
min maxMead J et al. J. Appl. Physiol. 28(5):596-6081970
Spannungsverteilung - inhomogenes System
![Page 15: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/15.jpg)
25
EL EW
ETOT
5
EL EW
15 15
ETOT
PulmonalLung injury
Compliance Lungs(CL) ↓↓
Compliance Thoraxic wall(CW) ⊥
Extra-pulmonalLung injury
Compliance Lungs (CL) ⊥ ↓
Compliance Thoraxic wall(CW) ↓↓
Transpulmonal pressure (PTP) = EL/ETOT* PAW
Elastance (E) =1
Compliance (C)
PAW
![Page 16: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/16.jpg)
Che
st W
all E
last
ance
(cm
H2O
/L)
N=21R=0.83P<0.0001
3,6 7,2 10,8 14,4 21,618,0 25,2 28,8
Pulmonary und extrapulmonary induced ARDS –intraabdominal pressure
Intraabdominal Pressure (cmH2O)
Gattinoni L, Pelosi P, Suter PM, Pedoto A, Vercesi P, Lissoni A.Acute respiratory distress syndrome caused by pulmonary and extrapulmonarydisease. Differentsyndromes?Am J Respir Crit Care Med. 1998;158:3-11.
![Page 17: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/17.jpg)
Transpulmonary pressure
PTP = PAW - Ppl
Paw = 30 cmH2O
PTP = 18 cm H2O
PTP = 12 cm H2O
Low Ppl
High Ppl
EL/Etot = 0.6
EL/Etot = 0.4
PTP = EL/ETOT * PAW
![Page 18: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/18.jpg)
BouhuysA.Physiology and musical instruments. Nature 1969;221(187):1199–1204
![Page 19: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/19.jpg)
Tidal Volumex ETOT
Plateau Pressure
x EL/ ETOT
TranspulmonalPressure
VILI
?
![Page 20: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/20.jpg)
Lung
vol
ume
Putensen C., Baum M., Hörmann C.Selecting ventilator settings according to variables derived from the quasi-static pressure/volume relationship in patients with acute lung injury.Anesth Analg 1993; 77:436-447.
0
250
500
1750
1500
1250
1000
750
0 5 10 15 20 25 30 35 40 45 50
Paw
55 cm H2O
mL
PEEP
VT
PEI
Biotrauma BarotraumaAtelecttrauma
* p<0.05
VT ml/kg
PEI cm H2O
PEEP cm H2O
PaO2/FiO2
PaCO2 mm Hg
Adjustemen of ventilator settingsaccording V/P-curve
prior10,1±0,8
30±27±1
180±2534±4
after7,5±0,8 *28±112±1 *
265±19 *38±3
FRC
![Page 21: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/21.jpg)
N=24
-2.SD
+2.SD
MW
N=24
FRC in assisted mechanical ventilation N2-washouttechnique versus CT
Zinserling J, Wrigge H, Varelmann D, Hering R, Putensen C.Measurement of functional residual capacity by nitrogen washout during partial ventilatory support.Intensive Care Med. 2003; 29:720-6
R = 0.78
![Page 22: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/22.jpg)
Lung
stre
sstra
nspu
lmon
al p
ress
ure
(cm
H2O
)
PEEP 5 cmH2O
0
44 50 33 50 22 50 11 505
PEEP 15 cmH2O
P<0.001
P=0.001
patients with healthyPatients with ALI/ARDS
P<0.001
P<0.01
lungs
6 ml/kg 12 ml/kg6 ml/kg 12 ml/kg
Lung stresslow vs.high VTandlow vs. high PEEP
Chiumello D, Carlesso E, Cadringher P, Caironi P, Valenza F, Polli F, Tallarini F, Cozzi P, Cressoni M, Colombo A, Marini JJ, Gattinoni L.Lung stress and strain during mechanical ventilation for acute respiratory distress syndrome.Am J Respir Crit Care Med. 2008;178:346-55.
![Page 23: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/23.jpg)
Do we need to measure Ppl?
![Page 24: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/24.jpg)
Esophageal Pressure
• Technique:– Balloon catheter filled with 0.5 ml gas
– in the milde third of the esophagus – heart beat artifacts
• Limitations of Pes :– P/V characteristics und filling of the balloon
– In the supine position: assumption is questionable becauseof the weight of the lungs and the mediastinum
– No calibration – occlusion test ?
– Artifacts caused by swallowing, heart interactions,…– Displacement by swallowing,
Assumption: Pes= mean Ppl
Inspection of the Pes curve is essential
![Page 25: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/25.jpg)
ΔPes ≈ ΔPpl
Benditt, Respir Care 2005; 50:68
![Page 26: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/26.jpg)
Esophageal-Balloon-Catheter
![Page 27: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/27.jpg)
Postioning
Patient in supine position30 degree upright.
60cm
40cm
![Page 28: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/28.jpg)
positivepressureventilation
spontaneousbreathing
Inhalation
Exhalation
Chest wall compliance
Inhalation Exhalation
Inspiratory muscleeffort
Respir Physiol 1977;31:63Eur Respir J 1988;1:51
mm Hg18
10
mm Hg18
10
Br J Anaesth 1976;48:474Crit Care Med 1983;11:271Chest 2002; 21:533-538
![Page 29: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/29.jpg)
• PEEP levels were set to achieve atranspulmonary pressure of 0 to 10 cm ofwater at end expiration
• Keep transpulmonary pressure <25 cm ofwater at end inspiration.
![Page 30: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/30.jpg)
![Page 31: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/31.jpg)
![Page 32: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/32.jpg)
![Page 33: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/33.jpg)
Conclusion - PTP
• Determination of PTP is complex and requiresmeasurement of PES
• Despite of PTP varies regionally we only determin anaverage PES
• PTPvaries with equal PPLAT caused by changes in thethoraxic wall compliance and during spontaneousbreathing
• PTP is the major force contributing to VILI• Ventilatory setting targeting PTP may be favorable
• Easier monitoring would be required
![Page 34: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/34.jpg)
hyperinflated
poorly aerated
normally aerated
not aerated-900 -500
EI
EE
40
0-1000
-20
Protective mechanical ventilationdelta vol (ml)
120
60
-100
Hounsfield Units
Terragni PP, Rosboch G, Tealdi A, Corno E, Menaldo E, Davini O, Gandini G, Herrmann P, Mascia L,Quintel M, Slutsky AS, Gattinoni L, Ranieri VM.Tidal Hyperinflation during Low Tidal Volume Ventilation in Acute Respiratory Distress Syndrome.Am J Respir Crit Care Med 2007;175:160-6
![Page 35: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/35.jpg)
EI
EE
hyperinflated
poorly aerated
normally aerated
not aerated
-500
120
80
40
-100
Hounsfield Units
0-1000 -900
-20
Terragni PP, Rosboch G, Tealdi A, Corno E, Menaldo E, Davini O, Gandini G, Herrmann P, Mascia L,Quintel M, Slutsky AS, Gattinoni L, Ranieri VM.Tidal Hyperinflation during Low Tidal Volume Ventilation in Acute Respiratory Distress Syndrome.Am J Respir Crit Care Med 2007;175:160-6
Non protective – protective mechanical ventilationDelta vol (ml)
![Page 36: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/36.jpg)
Disadvantage of current measurementsof lung mechanics
• global measurements• no measurements of absolute FRC/EELV• does not give any information on specificlung regions
• recruitment and overdistension may occursimultaneously
![Page 37: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/37.jpg)
Regional ventilationimaging technology
![Page 38: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/38.jpg)
CTNot at the bedside
radiationIntermittently applicable
EITat the bedsidenon invasive
continously applicable
-
Regional GascontentComputertomography andElectroimpedance Tomography
Ventilation
+
![Page 39: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/39.jpg)
Electrical conductiv performance ofthe chest
J. Malmivuo and R. Plonsey, „Bioelectromagnetism“, Oxford Press, 1995 (modified)
![Page 40: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/40.jpg)
CHEST 1999; 116:1695–1702
![Page 41: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/41.jpg)
Electrical Impedance Tomoraphy- EIT -
EIT-device
![Page 42: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/42.jpg)
16
5
78910
11
13
15 3
4
21
I
U
U
U
6
U
U
UU
12
U
U
U
U
14
UU
Electrical Impedance Tomoraphy- EIT -
![Page 43: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/43.jpg)
16
78910
11
15 3
4
21
IU
U
U5
6
U
U
UUU
13
12
U
U
U
U
14
U
16 x 13 =208measurements
per„Frame“➔at 50Hz➔10400
measurements/s
Electrical Impedance Tomoraphy- EIT -
![Page 44: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/44.jpg)
Different EIT reconstruction algorithms
![Page 45: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/45.jpg)
Moerer O, Hahn G, Quintel M.Lung impedance measurements to monitor alveolar ventilation.Curr Opin Crit Care 2011; 17:260-7
Multiple plane EIT measurements
![Page 46: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/46.jpg)
Bikker et al. Critical Care 2011, 15:R193
Influence of PEEP on regionaldistribution of ventilation
![Page 47: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/47.jpg)
Contribution of impedance with increasingdistance from the cross-section
defined by the position of the electrode belt.
![Page 48: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/48.jpg)
5 10 15 250
5
10
30 cmDiaphragmatic cupolaApex
Regional distribution of PEEP-induced overinflation in 32Patients with Acute Lung Injury ( 6 « focal » and 26 « diffuse »)
Overinflated lung volume ( % of total pulmonary volume )
15
Nieszkowska et al., Critical Care Medicine, 32: 1496, 2004
![Page 49: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/49.jpg)
What does the image tell us?
![Page 50: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/50.jpg)
Image display
• Display of differences of impedancewith respect to end-expiratoryreference level
• Colour coded display
ΔZpositive
ΔZ = 0:reference level
new
![Page 51: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/51.jpg)
•
•
•
Image display
Tidal images or standard deviation of ventilation
Colour coding
Display of ventilation, not the lung itself!
![Page 52: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/52.jpg)
Expiration
Ventilation cycle duringspontaneous breathing
Inspiration
![Page 53: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/53.jpg)
Image display as movie
![Page 54: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/54.jpg)
Display of local time courses
Globalventilation
Localventilationat cursorposition
![Page 55: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/55.jpg)
Validation studies
![Page 56: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/56.jpg)
Local lung air content with EIT andelectron beam tomography
Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G.Detection of local lung air content by electrical impedance tomography compared with electron beamCT.J Appl Physiol. 2002;93(2):660-6.
![Page 57: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/57.jpg)
Local lung air content with EIT andelectron beam tomography
Change of lung density and impedance difference with tidal volume variation
Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G.Detection of local lung air content by electrical impedance tomography compared with electron beamCT.J Appl Physiol. 2002;93(2):660-6.
![Page 58: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/58.jpg)
Regional ventilationEIT vs. single photon emission tomography
Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, Burchardi H, Hedenstierna G.Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy inpigs.Chest. 2003 Jul;124(1):314-22.
![Page 59: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/59.jpg)
Gas
geha
lt C
T [%
]G
asge
halt
CT
[%]
Regional Gascontent Electroimpedance Tomographyvs. Computertomography
ventral
dorsal
right lung left lung
0 10 20 30 40 50
10
0
50
40
30
20
0 10 20 30 40 50
10
0
50
40
30
20
Gasgehalt EIT [%]0 10 20 30 40 50
10
0
30
20
50
40
0 10 20 30 40 50
10
0
30
20
50
40
Gasgehalt EIT [%]
Wrigge H, Zinserling J, Muders T, Varelmann D, Günther U, Groeben C, Magnusson A, Hedenstierna G, Putensen C.Electrical impedance tomography compared to thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury.Crit Care Med 2008;36:903-9
R=0,78
R=0,71R=0,70
R=0,79
![Page 60: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/60.jpg)
BA
A
C D
Regional Gascontent Electroimpedance Tomographyvs. Computertomography
B
ventral
dorsal
right lung left lung
Wrigge H, Zinserling J, Muders T, Varelmann D, Günther U, Groeben C, Magnusson A, Hedenstierna G, Putensen C.Electrical impedance tomography compared to thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury.Crit Care Med 2008;36:903-9
![Page 61: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/61.jpg)
Information provided by EIT
![Page 62: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/62.jpg)
30 %
35 % 15 %
20 %
Pleural effusion due to rupture of diaphragm
Information provided by EIT1. Continuous quantification of regional distribution
of tidal volumesmax.
min.
![Page 63: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/63.jpg)
Information provided by EIT2. Assess the impact of therapeutic interventions
before recruitment immediately after recruitment
max.
min.
Patient ventilated with same tidal volumes before and afterrecruitment (both images with same color scale)
![Page 64: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/64.jpg)
Information provided by EIT3. Quantification of changes of end-expiratory
lung volume
Tidal volume: 500 ml dEELV: + 350 ml
![Page 65: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/65.jpg)
Electric impedance tomography tracing duringPEEP optimization
Erlandsson K, Odenstedt H, Lundin S, Stenqvist O.Positive end-expiratory pressure optimization using electric impedance tomography in morbidly obesepatients during laparoscopic gastric bypass surgery.Acta Anaesthesiol Scand. 2006;50(7):833-9
![Page 66: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/66.jpg)
Two recruitment maneuvers in seriesFirst maneuver Second maneuver
Volume recruited no further volume recruited
![Page 67: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/67.jpg)
Information provided by EIT4. Localization of regional end-expiratory
lung volume - changes
![Page 68: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/68.jpg)
Present and futureapplications
![Page 69: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/69.jpg)
Meier T, Luepschen H, Karsten J, et al.Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med2008; 34: 543-550
ventilationgain
ventilationloss
ΔVT+9ml
TVG
TVL
PEEP 15 ➔ PEEP 10
![Page 70: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/70.jpg)
![Page 71: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/71.jpg)
Tidal immage 1
Tidal immage 2
GlobalImpedancecurve
Trend parameterof the ventilator
Differenceimmage
Trend
Information provided by EIT
![Page 72: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/72.jpg)
Differenceimmage
Trend
ΔEELI global
Global Impedancecurve
Regionale changein end-exspiratorylung impedance
Regionalimpedance curve
Trend parametersof thr ventilator
Information provided EIT
![Page 73: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/73.jpg)
Costa EL, Borges JB, Melo A, et al.Bedside estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomography.Intensive Care Med 2009; 35: 1132-1137
Collapse and hyperinflation- regional compliance -
![Page 74: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/74.jpg)
Collapse and hyperinflation- Regional compliance -
Costa EL, Borges JB, Melo A, et al.Bedside estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomography.Intensive Care Med 2009; 35: 1132-1137
![Page 75: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/75.jpg)
Frak
tion
of re
crui
tied
atel
ecta
ses
0 200 400 600 800 1000 1200 1400
EIT ventilation delay [ms]
0.5
0.4
0.3
0.2
0.1
0.0
r2=0.60
N=24
Regional VentilationRegional Recruitment
Recruitment maneuver withlow gas flow
VL0.6
100%
45%
25%
15%
15%
Wrigge H, Zinserling J, Muders T, Varelmann D, Günther U, Groeben C, Magnusson A, Hedenstierna G, Putensen C.Electrical impedance tomography compared to thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury.Crit Care Med 2008;36:903-9
![Page 76: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/76.jpg)
Homogeneity of regional ventilationRegionalVentilation Delay Index
Ventilation
Regionalventilation distribution
RVD pixel for pixel RVD Map
early late
Start of the global impedance change
![Page 77: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/77.jpg)
PEEP 0 PEEP 5 PEEP 10 PEEP 15 PEEP 20 PEEP 25
RVD in Quadrants
right ventral:r2=0.81
left ventral:r2=0.48
right dorsal:r2=0.32
left dorsal:r2=0.98
Cyclic alveolar collapse
![Page 78: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/78.jpg)
Vent
ilatio
nRV
D M
apHomogenity of ventilation – recruitment/cyclic alveolar collapse
early
PEEP 0 PEEP 5 PEEP 10 PEEP 15 PEEP 20 PEEP 25max
min
late
potential for alveolar recruitment
cyclic alveolar collapse (tidal recruitment)
![Page 79: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/79.jpg)
EIT, SDRVD maping, CT at different PEEP levels
Muders T, Luepschen H, Putensen C.Impedance tomography as a new monitoring technique.Curr Opin Crit Care. 2010 Jun;16(3):269-75.
![Page 80: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/80.jpg)
PEEP
20
cmH
2OPE
EP 2
5 cm
H2OEnd-expiratory
CTDelay map (EIT)
late
RVD
early
late
RVD
early
Is of SDRVDclinical relevance ?
![Page 81: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/81.jpg)
What is the goal of a EIT directedventilator setting?
regionalV/Q determination using SPECT
![Page 82: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/82.jpg)
ARDSnetPEEP
EITPEEP
OLPEEP ANOVA
VT[ml]VE[l/min
215(±18)7.3(±0.8)
210(±15)7.1(±0.9)
209(±21)7.4(±0.9)
nsns
PEEP[cmH2O]Ppeak[cmH2O]Pmean[H2O]
10.1(±2.7)§$34.7(±6.9)§$32.7(±6.4)§$
22.1(±2.1)#$39.0(±5.2)#$37.3(±5.6)#$
25.0(±3.8)#§42.8(±7.1)#§40.1(±7.3)#§
P<0.001P<0.001P<0.001
PaO2/FiO2[mmHg] 141(±40)§$ 417(±114)# 388(±120)# P<0.001
CO[l/min] 6.3(±1.2)§$ 4.8(±1.3)# 4.3(±1.0)# P<0.001
ITBV[ml]MAP[mmHg]
520(±66)97(±17)
515(±95)94(±21)
499(±107)92(±18)
nsns
HR[bpm] 120(±25)§$ 105(±20)# 109(±17)# P<0.05
Data are average values ± SD. p<0.05 (post hoc), # vs ARDSnet-PEEP, § vs. EIT-PEEP, $ vs. OL PEEP
ResultsVentilatory and cardio-vascular variables
![Page 83: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/83.jpg)
dorsal
ventral
dorsal
ventral
dorsal
ventral
ARDSnet PEEP
PEEP 10 cm H2O
Total lungPerfused lung
Resultsregional perfusion
EIT PEEP
PEEP 20 cm H2O
Total lungPerfused lung
Open Lung PEEP
PEEP 24 cm H2O
Total lungPerfused lung
![Page 84: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/84.jpg)
Pulm
onal
er B
lutfl
uss [
ml/m
in]
[ml/m
in]
Perf
usio
n
niedriges
ventral
dorsal
[%]
pulmonaler Blutfluss [ml/min] § $ &
ANOVA: p<0.05 for factor „V/Q “, „ventilatory modality“ and interactions. Post hoc (Newman-Keuls):p<0.05 # vs ARDSnet-PEEP, § vs. EIT-PEEP, $ vs. OL PEEP
#
#
# #
§$§$
# #
§$
Shunt
V/Q
normalesV/Q hohes
V/Q
Totraum
ARDSnet PEEP
Resultsregionale perfusion
EIT PEEP Open Lung PEEP
1.6
1.0
0.4
![Page 85: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/85.jpg)
dorsal
ventral
dorsal
ventral
dorsal
ventral
ARDSnet PEEP
PEEP 10 cm H2OTotal lungVentilated lung
Resultsregional ventilation
EIT PEEP
PEEP 20 cm H2O
Total lungVentilated lung
Open Lung PEEP
PEEP 24 cm H2OTotal lungVentilated lung
![Page 86: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/86.jpg)
Pulm
onal
er G
asflu
ss [m
l/min
]
[ml/m
in]
Vent
ilatio
n
niedriges
ANOVA: p<0.05 for factor „V/Q “, „ventilatory modalities“ and interactions.Post hoc (Newman-Keuls): p<0.05 # vs ARDSnet-PEEP, § vs. EIT-PEEP, $ vs. OL PEEP
# #
§$
# #
§$
Shunt
V/Q
normalesV/Q hohes
V/Q
Totraum
ventral
dorsal
[%]
pulmonaler Gasfluss [ml/min] $&
ARDSnet PEEP
Resultsregional ventilation
EIT PEEP Open Lung PEEP
2.5
1.5
0.5
![Page 87: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/87.jpg)
dorsal
ventral
d
v
d
v
>10
<0.1
1
ARDSnet PEEP
PEEP 10 cm H2O
Resultsregional V/Q
EIT PEEP
PEEP 20 cm H2O
Open Lung PEEP
PEEP 24 cm H2O
![Page 88: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/88.jpg)
The future
![Page 89: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/89.jpg)
Supine
Prone
Regionale Ventilation/Perfusion-Verteilung bestimmt mit 81mKr/99mTc-MAA SPECT ScansLamm W.J.E. et al. Am J Respir Crit Care Med 1994; 150:184-193.
Regional Ventilation/Perfusion-Distribution
EIT-Ventilation
EIT-Perfusion
EIT-Ventilation/Perfusion
The goal
![Page 90: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/90.jpg)
Visualizing cardiac relatedimpedance changes
• Separation of respiratory andcardiac related impedancechanges
• ATTENTION: Perfusion isdefined as flow in a certaindirection
• Interpretation of cardiacimpedance changes is stilldifficult
• Temporal and spatialinformation may help
• Contrast agents:– Saline
– Glucose
Cardiacimpedance
changesRespiratoryimpedance
changes
Curves knownfrom thermodilution
measurements
![Page 91: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/91.jpg)
Three components
Time
Right Heart (RH), Lung (L), Left Heart (LH)
![Page 92: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/92.jpg)
Validation
• SPECT/CT (in pigs)
Muders T, Luepschen H, Putensen C. Curr Opin Crit Care, 2010
![Page 93: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/93.jpg)
Costa EL, Lima RG, Amato MB.Electrical impedance tomography.Curr Opin Crit Care. 2009 Feb;15(1):18-24
![Page 94: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/94.jpg)
Advantages of EIT
• Noninvasive method
• Application at bedside
• Regional ventilation changes can bemonitored
– over time
– after maneuvers
– used to adjust mechanical ventilation (e.g. PEEP)
![Page 95: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/95.jpg)
Currently, no clinical data areavailable that advanced
respiratory monitoring (e.g. PPT, EIT)improve outcome in the overall or in a
selected critical care population
![Page 96: Magic is in the air: Monitoring the respiratory system](https://reader036.fdocuments.in/reader036/viewer/2022062501/5681613f550346895dd0ac85/html5/thumbnails/96.jpg)