Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of...

70
Low Vacuum Scanning Electron Microscopy and Microanalysis Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes, John Wiley&Sons 2008 Several graphs courtesy of prof. Bradley Thiel

Transcript of Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of...

Page 1: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Low Vacuum Scanning Electron Microscopy and Microanalysis

Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes, John Wiley&Sons 2008Several graphs courtesy of prof. Bradley Thiel

Page 2: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

• Conventional scanning electron microscopy (SEM) is a widely used as an analytical tool. However, there are several limitations on the type of samples, which may be observed.

• When a specimen consists of a non-conductive material, or not properly grounded to the specimen stub, charging occurs. Charging is the build-up of an excess of electrons on the surface of the specimen, which causes many undesirable artefacts. Both secondary and backscattered electrons utilised for SEM observations are significantly affected.

• The elimination of specimen charging is accomplished in the LowVac/ESEM scanning electron microscope by the introduction of a gas, water, nitrogen or air into the specimen chamber.

Page 3: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Outline

Introduction to Variable Pressure (Environmental) Scanning Electron Microscopy

Image formation in VP-SEM

X-ray Microanalysis in VP-SEM

Variable pressure method

EBSD in VP-SEM

Conclusions

Page 4: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

LV-SEM – Low Vacuum Scanning Electron MicroscopyVacuum from 0 to 1 Torr (0 mbar – 1.33 mbar)

VP-SEM or Environmental Scanning Electron MicroscopyVacuum from 0 to 20 Torr (0 mbar – 26.6 mbar)

15

10

5

0

-10℃ 0℃ 10℃ 20℃

ESEM

Low

Vac

uu

m S

EM

Solid

phase

Liquid phase

Gaseous phase

Pre

ssu

re [

Torr

]

Page 5: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Specimen Chamber1-20 Torr

Gun

Pressure limiting aperture 1

Pressure limiting aperture 2

Water vapor/gas

Vent

(Not to scale)

RP DP

IP

RP DP

RP

10-8 Torr

10-5 Torr

10-1 Torr

10-3 Torr

Variable Pressure Scanning Electron Microscope

Page 6: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Pumping system of ESEM

PLA1

PLA2

Gas flow from chamber max. 20 Torr

Major gas flow(to rotary pump)

Minor gas flow(to ODP or TMP)

Emission area: min. 10-5 Torr

Page 7: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

E

Grounded Stub

Dielectric Specimen

Biased Anode

+ +

+ ++

+

+

+++

+++

+

+

+

Cascade Amplification

Page 8: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Cascade Amplification

E

Grounded Stub

Dielectric Specimen

Biased Anode

The SEs collide with the gas molecules in the chamber and create further electrons which are also accelerated in the electric field: a cascade or avalanche promotes amplification

Page 9: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

SE Amplification

Specimen

Detector

Pop

ula

tio

n

EnergyV0

E

Second Generation

First Generation

Third Generation

Second Generation

First Generation

Fourth Generation

Third Generation

Second Generation

First Generation

Swarm Energy

Page 10: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

• The primary electron beam impinges on the sample, leading to the production of backscattered and secondary electrons.

• The electric field between the positively biased anode and the grounded specimen stage accelerates secondary electrons towards the anode.

• Secondary electrons collide with and ionise gas molecules in their path and amplify the secondary signal.

• Positive ions drift towards the sample surface to compensate negative charge build up.

A simplified schematic diagram of gaseous signal amplification via ionisation

Page 11: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

0 1 2 3 4 5 6 7 8 9 10Pressure (Torr)

Ca

scad

e A

mp

lific

atio

n (

a.u

.)

General Shape of Amplification Curve

If the pressure is too high (concentration of gas molecules is large), MFP is short and significant fraction of secondary

electrons is scattered before they acquire sufficient energy to cause ionisation

If the pressure is too small (concentration of gas molecules is

small too), MFP is large and amplification events are few and the

secondary electron signal is weak

The most efficientamplification

Page 12: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Signal Components

Sign

alWater

0%

20%

40%

60%

80%

100%

0 1 2 3 4 5 6 7 8 9 10Pressure (Torr)

S.E.

B.S.E.

P.E.

Helium

0%

20%

40%

60%

80%

100%

0 1 3 5 7 9 11Pressure (Torr)

Sign

al

S.E.

B.S.E.

P.E.

PE – signal due to ionisation from primary electronsBSE – signal due to ionisation from backscatter electronsSE – signal due to ionisation from secondary electrons (most desired!)

Page 13: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Alternative Gases

0

50

100

150

200

250

300

0 2 4 6 8 10 12Pressure (Torr)

Re

lati

ve A

mp

lific

atio

n

H2O

CO 2

N2O

N2

He

Page 14: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

H2O

4 torr 8 torr6 torr

He

2 torr 5 torr 12 torr

Copper grid on Carbon

Page 15: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

deG

0VBPd

APe

P – pressure, V0 – anode bias, A & B - gas specific constants

d

S

d

SexpII BSEPEd

oc

Ic – amplified gaseous electron current Io – primary beam current SPE – ionisation efficiency of primary electronsSBSE – ionisation efficiency of backscatter electronsδ – secondary electron coefficientη - backscatter electron coefficientα – ionisation efficiency of the gasd – sample-anode gap

The factor α is called Townsend’s first ionisation coefficient

Signal gain G is a function of the ionisation coefficient α and the specimen-anode distance d

Page 16: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

0 2 4 6 8 10Pressure (torr)

Re

lati

veC

on

trib

uti

on

to T

ota

l Sig

nal

Total

S.E.

PE

BSE

Graph of signal contributions as a function of chamber gas pressure (for water vapour). The maximum total amplification occurs at slightly higher pressure than the secondary electron maximum. At high pressures, the signal is dominated by

background signals such as ionisation of gas molecules by primary electrons

Page 17: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Cascade Signal Components

The cascade signal is then composed of:

Amplified SE

BSE derived signal

PE derived signal

All of the above from the “skirt region”

Detector noise

Of these, only the first is desired!

Page 18: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

High vacuum versus low vacuum mode

SE image of Au coated Trich made in low vacuum mode (without charging)

SE images of Au coated Trich:despite the coating, the image still shows some charging effects

Page 19: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Microstructure

E-SEM with W hair-pin filamentE-SEM with Shottky FEG

Natural HAp/bioglass compositeDiatomite(a powdery mineral composed of the fossilised

skeletal remains of microscopic single-celled aquatic plants called diatoms)

Page 20: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

The minimum pressure that can sustain water in the liquid phase is about 4.6 Torr at 0oC.Higher temperatures require higher pressures

To maintain a sample in its natural hydrated state it is necessary to drop the temperature below room temperature,

Peltier cooling stage -25°C to +50°C

Page 21: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Hydrated sample orchid petal

1.4 Torr H2Ovapor at 6oC

dehydration is obvious

7 Torr H2O vapor at 6oC

H2O environment becomes saturated

and dehydration does not occur

before 2 min. exposure after 2 min. exposure

Page 22: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

ESEM ‘wet’ mode

Fresh bacteria

2 µm

Oil-in-water emulsion

Page 23: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Matrix ice

Interfacial ice T = -120oC P = 0.7 Torr N2

x520

Stokes, Mugnier & Clarke (2004) J Microscopy 213(Pt 3)

Page 24: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Microanalysisunder low vacuum conditions

Low vacuum SEM: charge is eliminated by a gas (water, air or N2)

Two major problems :

- beam damage (heating of sample)

- beam spread (skirt effect)

Beam spread

Electrons are scattered due to collisions with gas atoms

X-ray is generated outside the probe

X-ray information comes from up to 500 micron from central spot (skirt area)

Page 25: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Pressure Limiting Aperture

High pressure region

Low pressure region

H2O= H2O+ + 1 e-

Non-conductive sample

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

+ + + + + + + + + + + + + + + + +

Elastic scattering – Skirt formation

Beam spread :

• X-rays from the probe spot (actualinformation)

• X-rays due to atmospheric effect (gas)

• X-rays information up to 500 micron from central spot (skirt)

Page 26: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,
Page 27: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,
Page 28: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Fraction of electrons remaining in central probe:

where:

σ – scattering cross section

P – pressure

R – gas constant

d – gas path length

T – temperature

Electrons in the Skirt

RTPdef /σ-=

Page 29: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Elastic Cross Sections

0.01

0.1

1

1 10 100

Energy (keV)

Cro

ss S

ect

ion

2) Nitrogen

Water

Helium

Page 30: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Scattering out of the Probe

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 500 1000 1500 2000 2500

Pressure Distance (GPL) (Pa-mm)

frac

tio

n in

pro

be

Air

2 keV

30 keV

10 keV5 keV

Use as low a pressure and as short a Gas Path Length as possible!

Page 31: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Influence of atomic number on primary electron mean free path λ as a function of primary beam

energy

Influence of atomic number on skirt radius rs as a function of primary beam energy

Which gas?

Page 32: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Influence of atomic number on the percentage of primary electrons remaining in the focused probe as a function of primary beam energy

Helium –theoretically the best gas.However it is the most difficult gas to ionize and the small size of helium atoms makes this gas notoriously difficult to handle with a typical vacuum pump. Prolonged use of helium is definitely not recommended.

Page 33: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Helium

Nitrogen

Argon

Exploring the Gas Path Length

The general trends can be seen in all cases: there is a sharp rise in the radius of the skirt as the primary beam energy decreases below about 5 keV and almost two orders of magnitude difference between radii for GPL = 1 mm compared to 15 = mm

Page 34: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Exploring the Gas Path LengthWith the exception of helium, the percentage of electrons remaining in the focused probe falls sharply as the GPL increases. That brings with it a reduction in electron beam current.For argon, it would be advisable to maintain gas path length of a millimeter or two.For nitrogen and oxygen the GPL of the order a few millimeters is advisable.

Page 35: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Influence of energy on elastic mean free path λ as a function of pressure for nitrogen

How much gas?

Page 36: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

How much gas?

Influence of atomic number on skirt radius rs as a function of pressure

Page 37: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Influence of pressure on skirt radius rs for several GPLs for nitrogen

p = 100 PaRs = 2.4 μm for GPL = 1 mmrs = 26 μm for GPL = 5 mm

p = 2.8 kPars = 12.5 μm for GPL = 1 mmrs = 139 μm for GPL = 5 mm

How much gas?

Page 38: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Percentage of electron remaining in the focused probe as a function of pressure and atomic

number

GPL = 2 mm, E0 = 20 keV

Percentage of electron remaining in the focused probe as a function of pressure and a range of

primary beam energy

GPL = 1 mm, nitrogen

How much gas?

Page 39: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Microanalysisunder low vacuum conditions

• Low vacuum SEM: charge is eliminated by a gas (water, air or N2)

• Two major problems :

- beam damage (heating of sample)

- beam spread (skirt effect)

Beam spread

• Electrons are scattered due to collisions with gas atoms

• X-ray is generated outside the probe

• X-ray information comes from up to 500 micron from central spot (skirt area)

Page 40: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,
Page 41: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

X-ray Skirt

Electron beam

Precipitates

Matrix

„Skirt”

Page 42: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Sample : Jadeite NaAlSi2O6

Matrix 1 : polymer

Matrix 2 : Cu/Zn block

Distance between the spot position (centre of mineral)and the Cu/Zn block is more than 500 micron

Large sample in homogeneous matrix

Page 43: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

High-vacuum spectrum - Jadeite 25kV

Cu and Zn peaks are not visible

Page 44: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Microanalysisunder low vacuum conditions

To reduce the skirt effect:• use short gas path (a special cone attached

to pole piece)• use high acceleration voltage (25kV)• use low pressure (0.1 - 0.3 mbar)

VP gas compensated technique(with 2 spectra measured at different pressures)

developed by Wernisch and Dijkstra

Page 45: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

A pressure variation methodinvolves collecting spectra at two different pressures (with other parameters constant)

and uses the difference spectrum to extrapolate back to zero-scattering case.

This method assumes that the composition does not change with pressure!!!

It is only X-ray intensity that changes!

Page 46: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Pressure variation method

More pressure: more X-rays from B, less from A

AB

pressure

Intensity

Page 47: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Pressure variation methodTake 2 measurementsat 2 different pressures

Extrapolate the intensitiesto pressure 0

Apply matrix correction to extrapolated intensities

to get compositionpressure

Intensity

Charging

Page 48: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Low-vacuum spectrum0.4 mbar 25kV

Contribution of the matrix (brass holder) is clearly visible

Low-vacuum spectrum0.8 mbar 25kV

Page 49: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Results

High Vac

Low Vac results (gas compensated) are close to those of the HVanalysis.

Low Vac

(0.2 - 0.4 mbar)

Page 50: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

X-ray Mapping: Curing cement

Dry Wet

Page 51: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Summary - Skirt Effect

Imaging: No problem; skirt adds a nearly uniform backgroundResolution is defined by central probe

X-ray analysis:quantification is difficult but possiblemapping stillz possible

Trade-offs between variables: Pressure

Working distance

Type of gas

Energy of primary electrons

Resolution

Charge control

To reduce skirt effect:use short gas path (a special

cone attached to pole piece)

use high acceleration voltage

use low pressure (0.3 - 0.4 torr)

Page 52: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,
Page 53: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,
Page 54: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Hi-vac

Lo-vac

ESEM

Imaging: No problem; skirt adds a nearly uniform backgroundResolution is defined by central probe

Page 55: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Charge control by Duane –Hunt limit

Electron beam energy 3 keV

Page 56: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

EBSD from non-conductive samples

Page 57: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Orientation maps acquired from non-conductive smaple (cubic ZrO2)

50 μm 10 μm

a) C-SEM – electric charge non compensated

All Euler map

b) VP-SEM – electric charge compensated

All Euler map

Page 58: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Rule of thumb: Use as low pressure and small polepiece-specimen

distance as possible!

Trade-offs between several variables:

Pressure

Working distance

Type of gas

Energy of primary electrons

Resolution

Charge control

Elastic Cross Section

Scattering out of the Probe

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 500 1000 1500 2000 2500

Pressure Distance (Pa-mm)

frac

tio

n in

pro

be

Air

2 keV

30 keV

10 keV5 keV

0.01

0.1

1

1 10 100

Energy (keV)

Cro

ss S

ect

ion

(Å2

)

Nitrogen

Water

Helium

Page 59: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

PLZT as a ferroelectric ceramics reveals spontaneous polarization within the grains.

The polarization direction differs from grain to grain and depends on crystallographic orientation.

Applied electric field changes the polarization direction leading to mechanical stresses within the grains.

If the applied electric field is too strong, it can produce such large mechanical stress and corresponding strain within the grains that they can be pulled out and cracks are formed.

What happens when the charge is not compensated

Page 60: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Effect of coating on diffraction pattern from NiO monocrystala) no coating, b) carbon coating, c) gold coating

Page 61: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Gas inlet

Variable Pressure SEM seems to be a good solution to overcome

charging problems!

Courtesy of EDAX

Effect of coating on diffraction patter qualitya) no coating – VP-SEM, b) carbon coating, c) gold coating

Low backscatter yield of the ceramics due to low average atomic number !!!

Page 62: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Effect of vacuum in SEM chamber on diffraction pattern from Pta) 0.05 Torr, b) 0.5 Torr, c) 1.0 Torr

Page 63: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

EBSD pattern recorded from the PLZT ceramics in FEGSEM at different pressures:a) 50 Pa, b) 70 Pa, c) 90, d) 130 Pa

FEGSEM• a huge electrical charging

due high beam current• mechanical and electron

beam drift

Page 64: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

In order to achieve the best electron backscatterdiffraction quality in low vacuum conditions, there aresome parameters in the SEM which need to beadjusted, mainly:• beam energy and current,• gas path length,• gas pressure,• dwell time.

Page 65: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Inverse pole figure shows randomly oriented grains

91% of diffraction patterns solved!!! All Euler map

PLZT ceramicsPb1-3x/2LaxZr0.65Ti0.35O3 for x = 0.08

(denoted as PLZT 8/65/35)

Page 66: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Misorientation Distribution Function (Al2O3/WC )

32

32

O]2231[]1000[

O)0101()0101(

AlWC

AlWC

32

32

O]2594[]1000[

O)0101()0101(

AlWC

AlWC

32

32

O]0110[]1000[

O)0112()0112(

AlWC

AlWC

32

32

O]1891[]12110[

O)0213()0213(

AlWC

AlWC

Page 67: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Misorientation Distribution Function (Al2O3/ WC )

The crystallographic relationships correspond respectively to: 12%, 5%, 2%, 7% and 4% of the total Al2O3/WCinterphase boundary length.

0

2

4

6

8

10

12

14

1 2 3 4 5

Serie1

32

32

O]2231[]1000[

O)0101()0101(

AlWC

AlWC

32

32

O]2594[]1000[

O)0101()0101(

AlWC

AlWC

32

32

O]0110[]1000[

O)0112()0112(

AlWC

AlWC

32

32

O]1891[]12110[

O)0213()0213(

AlWC

AlWC

32

32

O]0101[]0211[

O)1000()1000(

AlWC

AlWC

% o

f th

e to

tal in

terp

ha

se b

oun

dary

len

gth

Page 68: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Al2O3 (H2O pressure - 0.4 mbar) Al2O3 (pressure H2O - 1.33 mbar)

Page 69: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Coarse grained Al2O3 Fine grained Al2O3

When the pressure is too low…

Page 70: Low Vacuum Scanning Electron Microscopy and … to Low Vacuum SEM...Principles and Practice of Variable Pressure/Environmental Scanning Electron Microscopy (VP-ESEM), Debbie J Stokes,

Thank you for attention