TI381C/07/en/05.05 51517610 Technical Information Liquiline M CM42 pH/ORP/conductivity/concentration/resistivity/digital sensors Two-wire transmitter for Ex and non-Ex areas 6 8 9 0 1 Application Liquiline M CM42 is a modular two-wire transmitter for all areas of process engineering. Depending on the ordered version, Liquiline has one or two analog current outputs or it can be connected to field buses as per FOUNDATION Fieldbus, PROFIBUS PA and Hart protocol. Liquiline is developed acc. to the international safety standard IEC 61508. The extremely robust, corrosion-resistant plastic version and the hygienic stainless steel version are designed for the following applications: Chemical processes Pharmaceuticals industry Foodstuff technology Ex applications Your benefits • Cost-saving: Simple commissioning with Quick Setup and Navigator (multifunction button) Thanks to Memosens technology, a calibration in the plant is not required Predictive maintenance system detects when a sensor has to be cleaned, calibrated or replaced Less storage thanks to modular design • Safe: Active display of cable interruption with Memosens version Red alarm LED reports error immediately User-guided commissioning, graphic display and plain text guidance SIL 2, ATEX, FM, CSA, NEPSI, TIIS, IP 67, NEMA 4X Code-protected commissioning and calibration Connection of all types of pH sensors (glass, ISFET, digital sensors) and conductivity sensors Industry solutions: Modular concept: sensor module replaceable Asset management (Fieldcare, W@M)

Transcript of Liquiline_CM42_LIGAÇÕES_TI381CEN.pdf

  • TI381C/07/en/05.0551517610

    Technical Information

    Liquiline M CM42pH/ORP/conductivity/concentration/resistivity/digital sensorsTwo-wire transmitter for Ex and non-Ex areas

    6 8 9 0 1

    Application

    Liquiline M CM42 is a modular two-wire transmitter for all areas of process engineering.Depending on the ordered version, Liquiline has one or two analog current outputs or it can be connected to field buses as per FOUNDATION Fieldbus, PROFIBUS PA and Hart protocol.Liquiline is developed acc. to the international safety standard IEC 61508.

    The extremely robust, corrosion-resistant plastic version and the hygienic stainless steel version are designed for the following applications: Chemical processes Pharmaceuticals industry Foodstuff technology Ex applications

    Your benefits

    Cost-saving: Simple commissioning with Quick Setup and Navigator

    (multifunction button) Thanks to Memosens technology, a calibration in the plant

    is not required Predictive maintenance system detects when a sensor has

    to be cleaned, calibrated or replaced Less storage thanks to modular design

    Safe: Active display of cable interruption with Memosens

    version Red alarm LED reports error immediately User-guided commissioning, graphic display and plain text

    guidance SIL 2, ATEX, FM, CSA, NEPSI, TIIS, IP 67, NEMA 4X Code-protected commissioning and calibration Connection of all types of pH sensors (glass, ISFET, digital

    sensors) and conductivity sensors Industry solutions:

    Modular concept: sensor module replaceable Asset management (Fieldcare, W@M)

  • Liquiline M CM42

    2 Endress+Hauser

    General features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Modular design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Quick setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Navigator and plain text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Process check system (PCS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Software packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    Special features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Output signal range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Ex specification current output 4/20 mA . . . . . . . . . . . . . . . . . . 10Ex specification PROFIBUS PA and FOUNDATION Fieldbus . . . . 10

    Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Housing grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Supply and signal circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Sensor connection: pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Sensor connection: conductivity . . . . . . . . . . . . . . . . . . . . . . . . . 17Sensor connection: digital sensors . . . . . . . . . . . . . . . . . . . . . . . . 18

    Performance characteristics. . . . . . . . . . . . . . . . . . . . 19pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

    Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Installation in Ex area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 23Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 25Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 264 approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Digital sensors (additional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

    Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Operating Instructions CM42 . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

  • Liquiline M CM42

    Endress+Hauser 3

    General features

    Modular design

    Quick setup To the first measuring value within 1 minuteAfter setting up the few parameters in the Quick Setup menu, the measuring point is ready to measure. You can be sure, to have set up the most important settings.The first measuring value is reliably displayed.

    Navigator and plain text The unique operating concept sets new standards: Fewer user errors thanks to very easy operation Quick configuration with the Navigator. Intuitive configuration and diagnosis due to plain text display

    Sensor monitor Pressing the DIAG key takes you to the sensor monitor. All the important sensor data, incl. warning and alarm limits, are graphically displayed in one go.

    Process check system (PCS) This function checks the measuring signal for stagnation. If the measuring signal does not change for some time (several measured values), an alarm is triggered. Soiling, blockage or similar could be the cause of such behaviour.

    a0001296

    Inside Liquiline (version with sensor module, without wiring)a0001979

    CPU and sensor module

    a0001984

    Navigatora0001985-en

    Plain text display

    a0001771-en

    Sensor monitor (example)

  • Liquiline M CM42

    4 Endress+Hauser

    Software packages The standard software contains the following features:

    Safety Safety integration level acc. to IEC 61508 (SIL 2)

    Liquiline is the first analysis transmitter to be developed to IEC 61508 and specified for use under SIL 2.The development to SIL 2 reduces the probability of hidden errors and stands for the highest degree of reliability.

    Alarm LED

    A bright, red LED draws your attention to an error (22 mA alarm).

    Code protection

    Liquiline is supplied without access protection.

    You can specify passwords linked to relevant access rights for various user groups: Specialist Service Operator...

    To guarantee the highest degree of safety and security in the pharmaceuticals sector, you can specify user roles with IDs and passwords as per 21CFR11.

    Reordering validated software

    You can order new devices with older, validated software so you do not have to constantly validate new software versions of new devices. This is possible as long as allowed by the hardware version.

    Features

    pH /ORP Conductivity Digital sensors

    Offset and two-point calibration

    Sample calibration Calibration with standard and

    special buffers Manual buffer Temperature compensation Temperature adjustment Simulation of current output Self-diagnosis

    Sample calibration Temperature calibration: one-point Simulation of current output Self-diagnosis Concentration measurement

    Offset and two-point calibration

    Sample calibration Calibration with standard and

    special buffers Manual buffer Temperature compensation Temperature adjustment Simulation of current output Self-diagnosis

  • Liquiline M CM42

    Endress+Hauser 5

    Special features

    pH / ORP Sensor Condition Check (SCC)

    This function monitors the state of the electrodes or the degree of electrode ageing. The "Electrode OK", "Low wear" or "Replace electrode" messages inform you on the state of the electrode. The electrode state is updated after each calibration. When the "Replace electrode" message appears, an error message is displayed.

    Sensor Check System (SCS)

    The sensor check system alerts to deviations of the pH glass impedance or reference impedance from the normal range, thus indicating possible failure due to pH electrode blocking or damage.

    Conductivity Polarisation monitoring

    Polarisation effects in the boundary layer between the sensor and the solution to be measured limit the measuring range of conductive conductivity sensors.The transmitter can detect and indicate polarisation effects using an innovative, intelligent signal evaluation process.

    United States Pharmacopeia (USP) and European Pharmacopoeia (EP)

    The requirements on ultrapure water in the pharmaceutical industry are specified by the American USP and the European EP.

    The transmitter meets the USP/EP requirements on conductivity measuring systems: Precise temperature measurement at point of conductivity measurement Simultaneous display of uncompensated conductivity values and temperature Display resolution 0.01 S/cm Exact adjustment of the transmitter in the factory with traceable precision resistances (optional) Exact adjustment of the sensors in the factory in accordance with ASTM D 1125-9 resp. ASTM D 5391-99

    (optional) Temperature-dependent measured value monitoring acc. to USP and EP.

    Digital sensors Memosens

    The measuring point is safer with Memosens: No-contact, inductive signal transmission, thus optimum galvanic separation No contact corrosion Completely water-tight Automatic error message if data flow interrupted Sensor calibration possible in laboratory Look-ahead maintenance thanks to recording of sensor data, e.g.:

    Total hours of operation Hours of operation for very high or very low pH values Hours of operation for high temperatures above 80 C (176 F) and 100 C (212 F) Number of steam sterilisations

  • Liquiline M CM42

    6 Endress+Hauser

    Measuring system

    ! Note! You can select from a wide variety of assemblies and sensors to set up your measuring point. You can find the corresponding information in the chapter "Accessories" resp. in the referenced documentations.

    a0002012

    Measuring system: Examples

    pH / ORP CM42-P/R... Measuring cable CPK9 Assembly Cleanfit CPA471 Sensor Orbisint CPS11

    Conductivity, inductive measurement CM42-I... Assembly Dipfit CLA111 Sensor Indumax CLS50

    Conductivity, conductive measurement CM42-C... Measuring cable CPK9 Sensor Condumax CLS16

    Digital sensor CM42-M... Measuring cable CYK10 Assembly Unifit CPA442 Sensor Oxymax COS22

  • Liquiline M CM42

    Endress+Hauser 7

    Input

    pH / ORP Measured variable

    pH value Oxidation-reduction potential (ORP) Temperature

    Measuring range

    Cable specification

    Applicable temperature sensors

    Pt100 Pt1000 NTC 30K

    Ex specification

    pH value 2 to 16

    ORP 2000 to +2000 mV

    Temperature 25 to + 150 C (13 to 302 F)

    Without SCS 50 m (164.05 ft) max. cable length

    With SCS 20 m (65.62 ft) max. cable length

    0 Intrinsicallay safe sensor circuit with protection type: EEx ia IIC

    Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co

    Glass electrode10.08 V4.1 mA10.2 mW200 H150 nF

    ISFET10.08 V50.7 mA128 mW200 H150 nF

  • Liquiline M CM42

    8 Endress+Hauser

    Conductivity Measured variable

    Conductivity Resistivity Concentration

    Measuring range

    Cable specification

    Temperature sensor

    Pt100 Pt1000

    Ex specification, conductive sensors

    Conductivity, conductive measuredTwo-polar sensorFour-polar sensor

    0.1 S.k1) to 20 mS.k0.1 S.k to 1.5 S.k

    1) k = cell constant in cm-1 Example: 2-polar sensor, k=0.01 cm1, resulting measuring range is from 0.001 to 200 S/cm.

    Conductivity, inductive measured 1.7 S.k2) to 1 S.k

    2) k=cell constant in cm1 Example: sensor, k=2 cm1, resulting measuring range is from 3.4 S/cm to 2 S/cm.

    Resistivity, conductive measured 10 M/k ... 50 /k

    Concentration NaOH: 0...15% (0...100 C / 32...212 F)HNO3: 0...25% (0...80 C / 32...176 F)H2SO4: 0...30% (0...100 C / 32...212F)H3PO4: 0...15% (0...80 C / 32...176 F)HCl: 0...20% (0...65 C / 32...149 F)4 user tables

    Conductivity/resistivity, conductive measured1)

    Two-cell sensor10 S.k to 20 mS.k / 0.1 M/k to 50 /k5 S.k to 20 mS.k / 0.2 M/k to 50 /k0.1 S.k to 20 mS.k / 20 M/k to 50 /k

    1) with CYK71 or CPK9 cables or sensor fixed cable

    100 m (328.10 ft) max. cable length50 m (164.05 ft) max. cable length15 m (49.22 ft) max. cable length

    Conductivity, conductive measuredFour-cell sensor10 S.k to 1.5 S.k0.1 S.k to 20 mS.k

    100 m (328.10 ft) max. cable length15 m (49.2 ft) max. cable length

    Conductivity, inductive measured2)

    2) with CLK5 cable or sensor fixed cable

    55 m (180.46 ft) max. cable length

    0 Intrinsicallay safe sensor circuit with protection type: EEx ia IIC T6Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co

    10.08 V23 mA57 mW335 H4 nF

  • Liquiline M CM42

    Endress+Hauser 9

    Ex specification, inductive sensors

    Digital sensors Measured variable

    pH value Oxidation-reduction potential Temperature

    Measuring range

    Cable specification

    Ex specification

    0 Intrinsicallay safe sensor circuit with protection type: EEx ia IICMax. output voltage UoMax. output current IoMax. output PoMax. outer capacity Co

    10.08 V64 mA128 mW62 nF

    For connection of the inductive sensor CLS50

    pH value 2 to 16

    ORP 2000 to +2000 mV

    Temperature 25 to + 150 C (13 to 302 F)

    With Memosens 100 m (328.10 ft) max. cable length

    0 Intrinsically safe sensor circuit with protection type: EEx ia IIC T6Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co

    5.04 V70 mA128 mW447 nF105.6 H

  • Liquiline M CM42

    10 Endress+Hauser

    Output

    Output signal 4 to 20 mA, potentially isolated against sensor circuit0.8 to 1.2 mA peak to peak1

    PROFIBUS PA2

    FOUNDATION Fieldbus3

    Signal on alarm 21.5 mAdigital via field bus4

    Load Max. load with an supply voltage of 24 V: 500 Max. load with an supply voltage of 30 V: 750

    Output signal range

    Ex specification current output 4/20 mA

    Ex specification PROFIBUS PA and FOUNDATION Fieldbus

    1) only for Hart

    2) for version with PROFIBUS PA3) for version with FOUNDATION Fieldbus4) with Profibus PA or FOUNDATION Fieldbus only

    pH adjustable, pH > 0.5

    ORP adjustable, U > 5 mV

    Conductivity, conductive measured none

    Conductivity, inductive measured none

    Temperature adjustable, T > 2 C

    0 Intrinsically safe supply and signal circuit with protection type: EEx ia IIC T6 passiveMax. input voltage UiMax. input current IiMax. input PiMax. inner inductivity Li

    Max. inner capacity Ci

    30 V100 mA750 mW221 H (output 1)neglegiable (output 2)1.2 nF (output 1)neglegiable (output 2)

    0 Suitable for use in a FISCO systemMax. input voltage UiMax. input current IiMax. input PiMax. inner inductivity LiMax. inner capacity Ci

    17,5 V380 mA5.32 W38.4 Hneglegiable

  • Liquiline M CM42

    Endress+Hauser 11

    Wiring

    Housing grounding Plastic housing

    " Caution! You must connect the "EMC barrier" to the foundation earth with a separate 4 mm2 functional earth line.

    Stainless steel housing

    " Caution! You must connect the outer earth connection of the housing to the foundation earth with a separate 4 mm2 line (GN/YE).

    Supply and signal circuit 4 ... 20 mA / Hart

    ! Note! Connect the transmitter via a two-wire cable screened on each end.

    a0003617

    Housing grounding12

    EMC barrier4 mm2 functional earth

    a0003616

    Housing grounding12

    Outer earth connection4 mm2 line (GN/YE)

    1

    2

    a0001091

    View in device (CPU module)

    a0001092

    Wiring diagram

    +

    133134

    234

    233

    232

    231

    134

    133

    132

    131

    +

    133

    134

    4...

    20m

    A

  • Liquiline M CM42

    12 Endress+Hauser

    PROFIBUS PA and FOUNDATION Fieldbus

    ! Note! Always use a fieldbus cable that is grounded on both sides (device and PCS).You can connect Profibus and FOUNDATION Fieldbus devices in three ways: Screened two-wire cable, "Hard grounding"

    (generally to be preferred to capacitive grounding) Screened two-wire cable, "Capacitive grounding"

    (if there is a risk of high equalizing currents) Using the fieldbus connection socket (accessories)

    Version 1: "Hard grounding"

    1. Place the cable screen on the "EMC barrier".

    2. Connect the cable wires as per the assignment.

    Version 2: "Capacitive grounding"

    1. Strip back the screen braiding, push the extension wire onto the exposed screen and fasten the clip:

    2. Place the extension wire on the "EMC barrier".

    3. Connect the cable wires as per the assignment.

    a0004060

    View in device (CPU module)

    a0001640

    Wiring diagram

    a0004071

    View in device (CPU module)

    a0004073

    Wiring diagram

    997

    998

    +

    +

    133

    132

    998

    997997

    998

    GN/YE

  • Liquiline M CM42

    Endress+Hauser 13

    Version 3: Fieldbus connection socket (accessories):

    1. Screw the fieldbus connection socket (accessories) into the housing bushing.

    2. Trim the connection cores of the socket to approx. 15 cm (5.9 ").

    3. Connect the cable cores as per the assignment. In doing so, you must place the cable screen (GN/YE) on the "EMC barrier".

    Supply voltage

    Sensor connection

    Explanation of abbreviations in the following diagrams:

    ! Note! Cable colours indicated as per IEC 757 (see CD-ROM). You must connect shielded connections and terminals with functional earth ()) (there is no protective earth

    (*) for plastic housings).

    a0002378

    View in device (CPU module)

    a0004062

    Wiring diagram

    Current output 4 to 20 mA without Hart:Current output 4 to 20 mA with Hart:PROFIBUS / FOUNDATION Fieldbus: (max. 22 mA)

    12.5 to 30 V DC12.5 to 30 V DC9 to 32 V DC (non-hazardous area) / 9 to 17.5 V DC (hazardous area)

    Abbreviation Meaning

    pH Signal from pH membrane glassRef Signal from reference electrodeSrc SourceDrn DrainPM Potential matchingU+ Digital sensor supplyUCom A

    Digital sensor communicationCom B Temperature signald.n.c. Do not connect!

    997

    998

    +

  • Liquiline M CM42

    14 Endress+Hauser

    Sensor connection: pH / ORP Glass electrodes with PML (symmetrical)

    Glass electrodes without PML (asymmetrical)

    a0001088

    View in devicea0001079

    Wiring diagram

    a0001089

    View in devicea0001080

    Wiring diagram

    PM

    BN

    317

    318

    320

    316

    315

    111

    113

    112

    319

    GNWHYE

    112113111

    320318317

    BKBN

    BK

    BN

    GN

    WH

    YE

    CPK9

    318

    320

    112

    113

    111

    317

    CM42

    319

    316

    315

    J

    PM

    Ref

    pH

    Sensor

    CPK1

    d.n.c.

    BN

    317

    318

    320

    316

    315

    111

    113

    112

    319

    BK

    GNWHYE

    112113111

    317BN

    320

    BK

    BN

    GN

    WH

    YE

    318

    320

    112

    113

    111

    317

    CM42

    319

    315

    316

    CPK9 / CPK1

    J

    d.n.c.

    Ref

    pH

    Sensor

    pH

    CPK1

  • Liquiline M CM42

    Endress+Hauser 15

    ISFET sensors with PML (symmetrical)

    ISFET sensors without PML (asymmetrical)

    a0001090

    View in devicea0001076

    Wiring diagram

    a0001084

    View in devicea0001077

    Wiring diagram

    PM

    BN

    317

    318

    320

    316

    315

    111

    113

    112

    319

    GNWHYE

    112113111

    320318

    316

    BKBNRD

    315

    BK

    BN

    GN

    WH

    YE

    CPK12

    317

    318

    112

    113

    111

    319

    CM42

    315

    316

    320

    RD

    Src

    J

    Ref

    PM

    Drn

    Sensor

    BN

    d.n.c.

    317

    318

    320

    316

    315

    111

    113

    112

    319

    GNWHYE

    112113111

    320316

    BKRD

    315

    BN

    BK

    BN

    GN

    WH

    YE

    317

    318

    112

    113

    111

    319

    CM42

    315

    316

    320

    RD

    Src

    RefDrn

    J

    d.n.c.

    CPK12 Sensor

  • Liquiline M CM42

    16 Endress+Hauser

    Overview, pH signal assignment

    Version: CM42-P...

    Version: CM42-R...As version CM42-P and also:

    Terminal Glass/redox ISFET Difference electrodewith PML without PML with PML without PML

    112 (GN) (GN) (GN) (GN) (GN)

    113 (WH) (WH) (WH) (WH) (WH)

    111 (YE) (YE) (YE) (YE) (YE)

    315 Src Src

    316 Drn Drn

    320 PM Ref PM Ref PM

    318 Ref Ref

    317 pH pH pH

    319 pH

    Terminal pH+redox pH+pH

    with PML without PML with PML without PML

    112 (GN) (GN) (GN) (GN)

    113 (WH) (WH) (WH) (WH)

    111 (YE) (YE) (YE) (YE)

    315

    316

    320 PM Ref (pH) PM Ref (pH-1)

    318 Ref (pH+ORP) Ref (ORP) Ref (pH-1 + pH-2) Ref (pH-2)

    317 pH pH pH pH

    319 ORP ORP pH pH

  • Liquiline M CM42

    Endress+Hauser 17

    Sensor connection: conductivity

    Conductive sensors, two-cell sensors

    Conductive sensors, four-cell sensors

    a0001086

    View in devicea0001083

    Wiring diagram

    a0002363

    View in device (sensor module)a0002371

    Wiring diagram

    221

    222

    220

    219

    111

    113

    112112

    BK

    GNWHYE

    113111219220

    BN

    BK

    BN

    GN

    WH

    YE

    CPK9 / CYK71

    221

    222

    112

    113

    111

    CM42

    220

    219

    J

    Sensor

    d.n.c.22

    122

    222

    021

    911

    111

    311

    2112

    BK

    GNWHYE

    113111219220222221

    BK

    221

    222

    112

    113

    111

    CM42

    220

    219

    J

    Sensor

  • Liquiline M CM42

    18 Endress+Hauser

    Inductive sensors

    Sensor connection: digital sensors

    a0001085

    View in devicea0001082

    Wiring diagram

    217

    218

    111

    113

    112

    216

    215215

    216112113111

    WH

    GNWHYE

    RD

    218217 WH

    RD

    GN

    WH

    YE

    CLK5

    217

    218

    215

    216

    112

    CM42

    111

    113

    J

    Sensor

    a0001087

    View in devicea0001078

    Wiring diagram

    588

    587

    388

    387

    198

    197

    188

    187 BN

    WHGN

    187188197198

    YE

  • Liquiline M CM42

    Endress+Hauser 19

    Performance characteristics

    pH / ORP Reference temperature

    25 C (77 F)

    Measured value resolution

    Maximum measured error1

    Conductivity Reference temperature

    25 C (77 F)adjustable from 5 to 100 C (23 to 212 F)2

    Measured value resolution

    pH:

    Glass electrodes:ISFET sensors:

    0.01 pH0.01 pH

    ORP: 1 mV

    Temperature:

    NTC30k:Pt100 / Pt1000:

    0.1 C (0.1 F)0.1 C (0.1 F)

    SCS glass electrodes:

    Glass:Reference:

    100 k100

    Leak current, ISFET sensors: 100 nA

    1) acc. to DIN IEC 746 part 1, under nominal operating conditions

    pH:

    Glass electrodes:ISFET sensors:

    0.02 pH0.02 pH

    ORP: 1 mV

    Temperature:

    NTC30k:Pt100 / Pt1000:

    0.5 K0.25 K

    SCS glass electrodes:

    Glass:Reference:

    200 M200

    Leak current, ISFET sensors: 100 nA

    2) with software package "Advanced" or "Advanced Plus"

    Conductivity < 0.1% from the measured value (4 digits)

    minimum, at:cell constant k 0.15 cm-1cell constant k 1.5 cm-1cell constant k > 1.5 cm-1

    0.001 S/cm0.01 S/cm0.1 S/cm

    Temperature (Pt100 / Pt1000) 0.01 C (0.01 F)

    Resistivityminimum

    < 0.1% from the measured value (4 digits)1 k.cm

  • Liquiline M CM42

    20 Endress+Hauser

    Maximum measured error1

    Temperature compensation

    Temperature adjustment

    Digital sensors Measured value resolution

    1) acc. to DIN IEC 746 part 1, under nominal operating conditions

    Conductivity, conductive measured

    Two-polar sensors 0.5 % from measured value 0.01 S.k

    Four-polar sensors 0.5 % from measured value 0.01 S.k 1)

    1.0 % from measured value 0.01 S.k 2)

    2.0 % from measured value 0.01 S.k 3)

    1) from 0.1 S.k to 20 mS.k (max. cable length 15 m (49.22 ft))

    2) from 10 S.k to 1000 mS.k (max. cable length 100 m (328.1 ft))

    3) from 10 S.k to 1500 mS.k (max. cable length 100 m (328.1 ft))

    Conductivity, inductive measured 0.5 % from measured value 1.7 S.k

    Resistivity, conductive measured(two-cell sensors)

    0.5 % from measured value 0.01 S.k 4)

    4) from 10 M/k to 50 /k (max. cable length 15 m (49.22 ft)) from 200 k/k to 50 /k (max. cable length 50 m (164.05 ft))

    Type of compensationnonelinearNaCl acc. to IEC 746-3natural waters acc. to IEC 7888Ultra-pure water NaClUltra-pure water HCl

    Range

    = 0.00 to 20.00 % / K0 to 100 C (32 to 212 F)0 to 35 C (32 to 95 F)0 to 100 C (32 to 212 F)0 to 100 C (32 to 212 F)

    Temperature offset 5 to +5 C (23 to 41 F)

    pH

    Glass electrodesISFET sensors

    0.01 pH0.01 pH

    ORP 1 mV

    Temperature 0.1 C (0.1 F)

  • Liquiline M CM42

    Endress+Hauser 21

    Installation

    Mounting plate

    Weather protection cover Plastic housing

    Stainless steel housing

    a0001676-en

    Weather protection cover

    a0001075-en

    Mounting plate

    a0001671-en

    Weather protection cover

    320

    / 12.

    60

    270 / 10.63

    mm / inch

    300 / 11.81

  • Liquiline M CM42

    22 Endress+Hauser

    Installation in Ex area

    Environment

    Ambient temperature range Non-hazardous area version30 to 70 C (22 to 158 F)

    Hazardous area versionT6: 20 to 50 C (4 to 122 F)T4: 20 to 55 C (4 to 131 F)

    Ambient temperature limits -30 to +80 C (22 to 176 F)

    Storage temperature 40 to 80 C (40 to 176 F)

    Electromagnetic compatibility Interference emission and interference immunity as per EN 61326: 2004

    Ingress protection IP 67

    Relative humidity 10 to 95%, not condensing

    a0001312-en

    Installation in Ex area1234

    Sensor in Ex versionIntrinsically safe sensor circuit EEx iaTransmitterSupply and signal circuit EEx ib (4...20 mA)

    5678

    Active barrier, e.g. Preline RN221Signal line Hart/PROFIBUS/FFFieldcare via PROFIBUS/FFHart handheld terminal

  • Liquiline M CM42

    Endress+Hauser 23

    Mechanical construction

    Dimensions Plastic housing

    Stainless steel housing

    a0001074-en

    Plastic housing

    a0001073-en

    Dimensions

    145

    / 5.7

    1

    134

    / 5.2

    811

    0/ 4

    .33

    132.5 / 5.22136.5 / 5.37

    160

    /6.3

    0

    134.5 / 5.30136.5 / 5.37

    26 /1.02

    174 / 6.85

    174

    /6.8

    510

    /0.

    39

    3/ 0

    .12

    220 / 8.66

    95/3

    .74

    mm / inch

  • Liquiline M CM42

    24 Endress+Hauser

    Weight Plastic housing

    1.5 kg (3.3 lb)

    Stainless steel housing

    2.1 kg (4.6 lb)

    Material Plastic housing

    Stainless steel housing

    Human interface

    Operating elements

    Housing:Housing seals:

    PolycarbonateEPDM

    Housing:Housing seals:

    Stainless steel 1.4301 (AISI 304)Foamed silicone, EPDM

    a0001301

    Overview of operation

    1 Display, current display: pH measuring mode2 Navigator3-6 Soft keys7 Soft key function (depends on menu)

  • Liquiline M CM42

    Endress+Hauser 25

    Ordering information

    Product structure Sensor inputA No ModuleC Conductivity, conductive measurementI Conductivity, inductive measurementM Digital sensor, MemosensP pH (glass/ISFET) or ORPR pH (glass/ORP redundant) or ISFET

    ApprovalA Non-hazardous area, CSA GPG Atex II (1)2G EEx ia/ib IIC T6 / II 3 DH Atex II 3G EEx na II T6N NEPSI Ex nA II T6P FM IS NI Cl. I, II, III, Div. 1&2, Groups A-GS CSA IS NI Cl. I, II, III, Div. 1&2, Groups A-GT TIIS

    CertificateA not selectedB Test certificate acc. to EN 10204, 3.1C Test certificate acc. to EN 10204, 3.1, factory calibration certificate

    Output0 1 current output 4 to 20 mA, Hart

    1 2 current outputs 4 to 20 mA, 1x Hart

    2 PROFIBUS PA3 FOUNDATION Fieldbus

    Housing0 Plastic1 Stainless steel 1.4301 (AISI 304)

    Cable entry0 M20 x 1.51 NPT "2 G

    SoftwareEA Standard versionEB Advanced versionEC Advanced Plus version

    Device languageC Zh/En (Chinese docs)D De/En (German docs)E En/De (English docs)F Fr/En (French docs)J Ja/En (Japanese docs)N Nl/En (Dutch docs)P Pl/En (Polish docs)S Es/En (Spanish docs)T It/En (Italian docs)

    Documentation0 Installation and configuration1 Installation and configuration + safety manual

    Additional equipment0 Basic version

    CM42- complete order code

  • Liquiline M CM42

    26 Endress+Hauser

    Scope of delivery The scope of delivery comprises: A transmitter acc. to the ordered version A mounting plate including 4 screws A cable gland plate and 2 cable glands (in the packaging) A sticker sheet (embedded in the housing, containing nameplates and terminal assignment stickers) A test certificate acc. to EN 10204-3.1 (optional) A factory calibration certificate An Operating Instructions BA381C "Commissioning" An Operating Instructions BA382C "Operation" A Safety Manual acc. to SIL 2 (optional) A CD ROM with additional documentation

    Certificates and approvals

    4 approval Declaration of conformityThe product meets the legal requirements of the harmonized European standards. The manufacturer confirms compliance with the standards by affixing the 4 symbol.

    Ex approval Depending on the version ordered: ATEX II (1)2G, EEx ia/ib IIC T6 / II 3D ATEX II 3G, EEx na II T6 NEPSI Ex nA II T6 CSA IS NI Cl.I, II, III, Div. 1&2, Grps. A-G FM IS NI Cl.I, II, III, Div. 1&2, Grps. A-G TIIS

  • Liquiline M CM42

    Endress+Hauser 27

    Accessories

    pH / ORP Glass electrodes

    Orbisint CPS11/CPS11DpH electrode for process applications, with PTFE diaphragm, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI028/C07/en)

    Orbisint CPS12/CPS12DORP electrode for process applications, with PTFE diaphragm, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI367/C07/en)

    Ceraliquid CPS41/CPS41DpH electrode with ceramics diaphragm and KCl liquid electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI079/C07/en)

    Ceraliquid CPS42/CPS42DORP electrode with ceramics diaphragm and KCl liquid electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI079/C07/en)

    Ceragel CPS71/CPS71DpH electrode with double chamber reference system and integrated bridge electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI245/C07/en)

    Ceragel CPS72/CPS72DORP electrode with double chamber reference system and integrated bridge electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI374/C07/en)

    Orbipore CPS91/CPS91DpH electrode with open aperture for media with high dirt load, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI375C/07/en)

    ISFET sensors

    Tophit CPS471Sterilisable and autoclavable ISFET sensor for food and pharmaceuticals, process technology, water treatment and biotechnology;Ordering acc. to product structure, see Technical Information (TI283/C07/en)

    Tophit CPS441Sterilisable ISFET sensor for media with low conductivity, with liquid KCl electrolyte;Ordering acc. to product structure, see Technical Information (TI352/C07/en)

    Tophit CPS491ISFET sensor with open aperture for media with high dirt load;Ordering acc. to product structure, see Technical Information (TI377/C07/en)

  • Liquiline M CM42

    28 Endress+Hauser

    Assemblies

    Cleanfit P CPA471Retractable assembly for tank and pipe installation,

    Cleanfit P CPA473Pneumatically or manually driven retractable assembly with ball valve for reliable process termination, parts in contact with medium of 1.4404 (AISI 316L) stainless steel

    Cleanfit P CPA474Pneumatically or manually driven retractable assembly with ball valve for reliable process termination, parts in contact with medium of PP, PEEK or PVDF

    Cleanfit H CPA475Retractable assembly for tank and pipe installation under sterile conditions,

    Dipfit W CPA111Immersion and installation assembly for open and closed tanks,

    Unifit H CPA442Installation assembly for food industry, biotechnology and pharmaceutical industry, with EHEDG and 3A certificate,

    ! Note! All the assemblies can be ordered using the product structure. Please refer to the Technical Information of the desired assembly in question.

    a0001305

    Fig. 1: Cleanfit P CPA471a0001306

    Fig. 2: Cleanfit P CPA473a0001307

    Fig. 3: Cleanfit P CPA474

    a0001308

    Fig. 4: Cleanfit H CPA475a0001309

    Fig. 5: Dipfit W CPA111a0001310

    Fig. 6: Unifit H CPA442

  • Liquiline M CM42

    Endress+Hauser 29

    Measuring cables

    CPK9 special measuring cableFor sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68Ordering acc. to product structure, see Technical Information (TI118C/07/en)

    CPK12 special measuring cableFor pH/redox glass electrodes and ISFET sensors with TOP68 plug-in headOrdering acc. to product structure, see Technical Information (TI118C/07/en)

    Buffer solutions

    Technical buffer solutions, accuracy 0.02 pH, acc. to NIST/DIN pH 4.0 red, 100 ml (0.026 US gal.), order no. CPY 2-0 pH 4.0 red, 1000 ml (0.264 US gal.), order no. CPY 2-1 pH 7.0 green, 100 ml (0.026 US gal.), order no. CPY 2-2 pH 7.0 green, 1000 ml (0.264 US gal.), order no. CPY 2-3

    Technical buffer solutions for single use, accuracy 0.02 pH, acc. to NIST/DIN pH 4.0 20 x 20 ml (0.005 US gal.), order no. CPY 2-D pH 7.0 20 x 20 ml (0.005 US gal.), order no. CPY 2-E

    Technical buffer solutions for ORP electrodes +225 mV, pH 7, 100 ml (0.026 US gal.); order no. CPY3-0 +468 mV, pH 0, 100 ml (0.026 US gal.); order no. CPY3-1

    Conductivity Conductive sensors

    Condumax W CLS12Conductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to version, see Technical Information TI 082/C07/en

    Condumax W CLS13Conductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to version, see Technical Information TI 083/C07/en

    Condumax W CLS15Conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);Ordering acc. to version, see Technical Information TI 109/C07/en

    Condumax H CLS16Hygienic conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);Ordering acc. to version, see Technical Information TI 227/C07/en

    Condumax W CLS19Conductive conductivity sensor for pure and ultra-pure water applications;Ordering acc. to version, see Technical Information TI 110/C07/en

    Condumax W CLS21Conductive conductivity sensor for applications with middle to high conductivity(incl. Ex); Ordering acc. to version, see Technical Information TI 085/C07/en

    Inductive sensors

    Indumax P CLS50Inductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to the sensor version, see Technical Information (TI118C/07/en)

    Indumax H CLS52Inductive conductivity sensor with short response time in hygienic design;Ordering acc. to the sensor version, see Technical Information (TI167C/07/en)

  • Liquiline M CM42

    30 Endress+Hauser

    Assemblies

    Dipfit W CLA111Immersion assembly for open and closed tanks with flange DN 100;Order depending on version, see Technical Information (TI135C/07/en)

    Dipfit P CLA140For the inductive sensor CLS50Immersion assembly with flange connection for processes with high requirements;Order depending on version, see Technical Information (TI196C/07/en)

    Measuring cables

    Special measuring cable / extension cable CYK71 for two-polar conductivity sensors with integrated temperature sensor, one low-noise coaxial line, four auxiliary cores at 0.75 mm2 each with a common screen,

    outer diameter 7 mm (0.25") sold by the metre:

    (minimum) length 5 m / 15 ft; order no. 50085333length 5 m / 15 ft; order no. 50088280length 10 m / 30 ft; order no. 50088281length 50 m / 150 ft; order no. 50088284length 100 m / 300 ft; order no. 50088285

    Special measuring cable / extension cable CYK71-Ex for Ex applications, see CYK71, but with a blue sheath

    sold by the metre, minimum length 5 m / 15 ft; order no. 50085673

    Calibration solutions

    Precision calibration solutions, acc. to SRM (Standard reference material) of NIST, error limit 0.5 %, reference temperature 25 C (77 F), with temperature table CLY11-A, 74.0 S/cm, 500 ml (0.132 Us.gal); order no. 50081902 CLY11-B, 149.6 S/cm, 500 ml (0.132 Us.gal); order no. 50081903 CLY11-C, 1.406 mS/cm, 500 ml (0.132 Us.gal); order no. 50081904 CLY11-D, 12.64 mS/cm, 500 ml (0.132 Us.gal); order no. 50081905 CLY11-E, 107.0 mS/cm, 500 ml (0.132 Us.gal); order no. 50081906

    Digital sensors (additional) Measuring cables

    CYK10 Memosens data cableFor digital pH sensors with Memosens technology (CPSxxD)Ordering according to product structure, see Technical Information (TI376C/07/en)

    a0001303

    Fig. 7: Dipfit W CLA111a0001304

    Fig. 8: Dipfit P CLA140

  • Documentation

    Operating Instructions CM42 Liquiline M CM42, Operating Instructions, KA381C/07/en and BA382C/07/enThe Operating Instructions correspond to the device version.Methods for (re-) ordering: Re-order the KA/BA by specifying the serial number of your device (via your local Sales Office) Order a specific KA/BA by specifying a valid Liquiline order code (via Internet).

    International Head Quarters

    Endress+Hauser

    GmbH+Co. KG

    Instruments International

    Colmarer Str. 6

    79576 Weil am Rhein

    Deutschland

    Tel. +49 76 21 9 75 02

    Fax +49 76 21 9 75 34 5

    www.endress.com

    [email protected]

    TI381C/07/en/05.05Printed in Germany / FM+SGML 6.0 / POD/

    51517610

  • ApplicationYour benefitsGeneral featuresModular designQuick setupNavigator and plain textSensor monitorProcess check system (PCS)Software packagesSafetySafety integration level acc. to IEC 61508 (SIL2)Alarm LEDCode protectionReordering validated software

    Special featurespH / ORPSensor Condition Check (SCC)Sensor Check System (SCS)

    ConductivityPolarisation monitoringUnited States Pharmacopeia (USP) and European Pharmacopoeia (EP)

    Digital sensorsMemosens

    Measuring systemInputpH / ORPMeasured variableMeasuring rangeCable specificationApplicable temperature sensorsEx specification

    ConductivityMeasured variableMeasuring rangeCable specificationTemperature sensorEx specification, conductive sensorsEx specification, inductive sensors

    Digital sensorsMeasured variableMeasuring rangeCable specificationEx specification

    OutputOutput signalSignal on alarmLoadOutput signal rangeEx specification currentoutput 4/20mAEx specification PROFIBUSPA and FOUNDATION Fieldbus

    WiringHousing groundingPlastic housingStainless steel housing

    Supply and signal circuit4 ... 20 mA / HartPROFIBUS PA and FOUNDATION FieldbusSupply voltageSensor connection

    Sensor connection: pH/ORPGlass electrodes with PML (symmetrical)Glass electrodes without PML (asymmetrical)ISFET sensors with PML (symmetrical)ISFET sensors without PML (asymmetrical)Overview, pH signal assignment

    Sensor connection: conductivityConductive sensors, two-cell sensorsConductive sensors, four-cell sensorsInductive sensors

    Sensor connection: digitalsensors

    Performance characteristicspH / ORPReference temperatureMeasured value resolutionMaximum measured error

    ConductivityReference temperatureMeasured value resolutionMaximum measured errorTemperature compensationTemperature adjustment

    Digital sensorsMeasured value resolution

    InstallationInstallationMounting plateWeather protection coverPlastic housingStainless steel housing

    Installation in Ex area

    EnvironmentEnvironmentAmbient temperature rangeAmbient temperature limitsStorage temperatureElectromagnetic compatibilityIngress protectionRelative humidity

    Mechanical constructionDimensionsPlastic housingStainless steel housing

    WeightPlastic housingStainless steel housing

    MaterialPlastic housingStainless steel housing

    Human interfaceOperating elements

    Ordering informationProduct structureScope of delivery

    Certificates and approvals4 approvalEx approval

    AccessoriespH / ORPGlass electrodesISFET sensorsAssembliesMeasuring cablesBuffer solutions

    ConductivityConductive sensorsInductive sensorsAssembliesMeasuring cablesCalibration solutions

    Digital sensors (additional)Measuring cables

    DocumentationOperating Instructions CM42