Technical Information
Liquiline M CM42pH/ORP/conductivity/concentration/resistivity/digital sensorsTwo-wire transmitter for Ex and non-Ex areas
6 8 9 0 1
Application
Liquiline M CM42 is a modular two-wire transmitter for all areas of process engineering.Depending on the ordered version, Liquiline has one or two analog current outputs or it can be connected to field buses as per FOUNDATION Fieldbus, PROFIBUS PA and Hart protocol.Liquiline is developed acc. to the international safety standard IEC 61508.
The extremely robust, corrosion-resistant plastic version and the hygienic stainless steel version are designed for the following applications: Chemical processes Pharmaceuticals industry Foodstuff technology Ex applications
Your benefits
Cost-saving: Simple commissioning with Quick Setup and Navigator
(multifunction button) Thanks to Memosens technology, a calibration in the plant
is not required Predictive maintenance system detects when a sensor has
to be cleaned, calibrated or replaced Less storage thanks to modular design
Safe: Active display of cable interruption with Memosens
version Red alarm LED reports error immediately User-guided commissioning, graphic display and plain text
guidance SIL 2, ATEX, FM, CSA, NEPSI, TIIS, IP 67, NEMA 4X Code-protected commissioning and calibration Connection of all types of pH sensors (glass, ISFET, digital
sensors) and conductivity sensors Industry solutions:
Modular concept: sensor module replaceable Asset management (Fieldcare, W@M)
-
Liquiline M CM42
2 Endress+Hauser
General features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Modular design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Quick setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Navigator and plain text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Process check system (PCS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Software packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Special features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Output signal range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Ex specification current output 4/20 mA . . . . . . . . . . . . . . . . . . 10Ex specification PROFIBUS PA and FOUNDATION Fieldbus . . . . 10
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Housing grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Supply and signal circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Sensor connection: pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Sensor connection: conductivity . . . . . . . . . . . . . . . . . . . . . . . . . 17Sensor connection: digital sensors . . . . . . . . . . . . . . . . . . . . . . . . 18
Performance characteristics. . . . . . . . . . . . . . . . . . . . 19pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Installation in Ex area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 23Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 25Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 264 approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Digital sensors (additional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Operating Instructions CM42 . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
-
Liquiline M CM42
Endress+Hauser 3
General features
Modular design
Quick setup To the first measuring value within 1 minuteAfter setting up the few parameters in the Quick Setup menu, the measuring point is ready to measure. You can be sure, to have set up the most important settings.The first measuring value is reliably displayed.
Navigator and plain text The unique operating concept sets new standards: Fewer user errors thanks to very easy operation Quick configuration with the Navigator. Intuitive configuration and diagnosis due to plain text display
Sensor monitor Pressing the DIAG key takes you to the sensor monitor. All the important sensor data, incl. warning and alarm limits, are graphically displayed in one go.
Process check system (PCS) This function checks the measuring signal for stagnation. If the measuring signal does not change for some time (several measured values), an alarm is triggered. Soiling, blockage or similar could be the cause of such behaviour.
a0001296
Inside Liquiline (version with sensor module, without wiring)a0001979
CPU and sensor module
a0001984
Navigatora0001985-en
Plain text display
a0001771-en
Sensor monitor (example)
-
Liquiline M CM42
4 Endress+Hauser
Software packages The standard software contains the following features:
Safety Safety integration level acc. to IEC 61508 (SIL 2)
Liquiline is the first analysis transmitter to be developed to IEC 61508 and specified for use under SIL 2.The development to SIL 2 reduces the probability of hidden errors and stands for the highest degree of reliability.
Alarm LED
A bright, red LED draws your attention to an error (22 mA alarm).
Code protection
Liquiline is supplied without access protection.
You can specify passwords linked to relevant access rights for various user groups: Specialist Service Operator...
To guarantee the highest degree of safety and security in the pharmaceuticals sector, you can specify user roles with IDs and passwords as per 21CFR11.
Reordering validated software
You can order new devices with older, validated software so you do not have to constantly validate new software versions of new devices. This is possible as long as allowed by the hardware version.
Features
pH /ORP Conductivity Digital sensors
Offset and two-point calibration
Sample calibration Calibration with standard and
special buffers Manual buffer Temperature compensation Temperature adjustment Simulation of current output Self-diagnosis
Sample calibration Temperature calibration: one-point Simulation of current output Self-diagnosis Concentration measurement
Offset and two-point calibration
Sample calibration Calibration with standard and
special buffers Manual buffer Temperature compensation Temperature adjustment Simulation of current output Self-diagnosis
-
Liquiline M CM42
Endress+Hauser 5
Special features
pH / ORP Sensor Condition Check (SCC)
This function monitors the state of the electrodes or the degree of electrode ageing. The "Electrode OK", "Low wear" or "Replace electrode" messages inform you on the state of the electrode. The electrode state is updated after each calibration. When the "Replace electrode" message appears, an error message is displayed.
Sensor Check System (SCS)
The sensor check system alerts to deviations of the pH glass impedance or reference impedance from the normal range, thus indicating possible failure due to pH electrode blocking or damage.
Conductivity Polarisation monitoring
Polarisation effects in the boundary layer between the sensor and the solution to be measured limit the measuring range of conductive conductivity sensors.The transmitter can detect and indicate polarisation effects using an innovative, intelligent signal evaluation process.
United States Pharmacopeia (USP) and European Pharmacopoeia (EP)
The requirements on ultrapure water in the pharmaceutical industry are specified by the American USP and the European EP.
The transmitter meets the USP/EP requirements on conductivity measuring systems: Precise temperature measurement at point of conductivity measurement Simultaneous display of uncompensated conductivity values and temperature Display resolution 0.01 S/cm Exact adjustment of the transmitter in the factory with traceable precision resistances (optional) Exact adjustment of the sensors in the factory in accordance with ASTM D 1125-9 resp. ASTM D 5391-99
(optional) Temperature-dependent measured value monitoring acc. to USP and EP.
Digital sensors Memosens
The measuring point is safer with Memosens: No-contact, inductive signal transmission, thus optimum galvanic separation No contact corrosion Completely water-tight Automatic error message if data flow interrupted Sensor calibration possible in laboratory Look-ahead maintenance thanks to recording of sensor data, e.g.:
Total hours of operation Hours of operation for very high or very low pH values Hours of operation for high temperatures above 80 C (176 F) and 100 C (212 F) Number of steam sterilisations
-
Liquiline M CM42
6 Endress+Hauser
Measuring system
! Note! You can select from a wide variety of assemblies and sensors to set up your measuring point. You can find the corresponding information in the chapter "Accessories" resp. in the referenced documentations.
a0002012
Measuring system: Examples
pH / ORP CM42-P/R... Measuring cable CPK9 Assembly Cleanfit CPA471 Sensor Orbisint CPS11
Conductivity, inductive measurement CM42-I... Assembly Dipfit CLA111 Sensor Indumax CLS50
Conductivity, conductive measurement CM42-C... Measuring cable CPK9 Sensor Condumax CLS16
Digital sensor CM42-M... Measuring cable CYK10 Assembly Unifit CPA442 Sensor Oxymax COS22
-
Liquiline M CM42
Endress+Hauser 7
Input
pH / ORP Measured variable
pH value Oxidation-reduction potential (ORP) Temperature
Measuring range
Cable specification
Applicable temperature sensors
Pt100 Pt1000 NTC 30K
Ex specification
pH value 2 to 16
ORP 2000 to +2000 mV
Temperature 25 to + 150 C (13 to 302 F)
Without SCS 50 m (164.05 ft) max. cable length
With SCS 20 m (65.62 ft) max. cable length
0 Intrinsicallay safe sensor circuit with protection type: EEx ia IIC
Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co
Glass electrode10.08 V4.1 mA10.2 mW200 H150 nF
ISFET10.08 V50.7 mA128 mW200 H150 nF
-
Liquiline M CM42
8 Endress+Hauser
Conductivity Measured variable
Conductivity Resistivity Concentration
Measuring range
Cable specification
Temperature sensor
Pt100 Pt1000
Ex specification, conductive sensors
Conductivity, conductive measuredTwo-polar sensorFour-polar sensor
0.1 S.k1) to 20 mS.k0.1 S.k to 1.5 S.k
1) k = cell constant in cm-1 Example: 2-polar sensor, k=0.01 cm1, resulting measuring range is from 0.001 to 200 S/cm.
Conductivity, inductive measured 1.7 S.k2) to 1 S.k
2) k=cell constant in cm1 Example: sensor, k=2 cm1, resulting measuring range is from 3.4 S/cm to 2 S/cm.
Resistivity, conductive measured 10 M/k ... 50 /k
Concentration NaOH: 0...15% (0...100 C / 32...212 F)HNO3: 0...25% (0...80 C / 32...176 F)H2SO4: 0...30% (0...100 C / 32...212F)H3PO4: 0...15% (0...80 C / 32...176 F)HCl: 0...20% (0...65 C / 32...149 F)4 user tables
Conductivity/resistivity, conductive measured1)
Two-cell sensor10 S.k to 20 mS.k / 0.1 M/k to 50 /k5 S.k to 20 mS.k / 0.2 M/k to 50 /k0.1 S.k to 20 mS.k / 20 M/k to 50 /k
1) with CYK71 or CPK9 cables or sensor fixed cable
100 m (328.10 ft) max. cable length50 m (164.05 ft) max. cable length15 m (49.22 ft) max. cable length
Conductivity, conductive measuredFour-cell sensor10 S.k to 1.5 S.k0.1 S.k to 20 mS.k
100 m (328.10 ft) max. cable length15 m (49.2 ft) max. cable length
Conductivity, inductive measured2)
2) with CLK5 cable or sensor fixed cable
55 m (180.46 ft) max. cable length
0 Intrinsicallay safe sensor circuit with protection type: EEx ia IIC T6Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co
10.08 V23 mA57 mW335 H4 nF
-
Liquiline M CM42
Endress+Hauser 9
Ex specification, inductive sensors
Digital sensors Measured variable
pH value Oxidation-reduction potential Temperature
Measuring range
Cable specification
Ex specification
0 Intrinsicallay safe sensor circuit with protection type: EEx ia IICMax. output voltage UoMax. output current IoMax. output PoMax. outer capacity Co
10.08 V64 mA128 mW62 nF
For connection of the inductive sensor CLS50
pH value 2 to 16
ORP 2000 to +2000 mV
Temperature 25 to + 150 C (13 to 302 F)
With Memosens 100 m (328.10 ft) max. cable length
0 Intrinsically safe sensor circuit with protection type: EEx ia IIC T6Max. output voltage UoMax. output current IoMax. output PoMax. outer inductance LoMax. outer capacity Co
5.04 V70 mA128 mW447 nF105.6 H
-
Liquiline M CM42
10 Endress+Hauser
Output
Output signal 4 to 20 mA, potentially isolated against sensor circuit0.8 to 1.2 mA peak to peak1
PROFIBUS PA2
FOUNDATION Fieldbus3
Signal on alarm 21.5 mAdigital via field bus4
Load Max. load with an supply voltage of 24 V: 500 Max. load with an supply voltage of 30 V: 750
Output signal range
Ex specification current output 4/20 mA
Ex specification PROFIBUS PA and FOUNDATION Fieldbus
1) only for Hart
2) for version with PROFIBUS PA3) for version with FOUNDATION Fieldbus4) with Profibus PA or FOUNDATION Fieldbus only
pH adjustable, pH > 0.5
ORP adjustable, U > 5 mV
Conductivity, conductive measured none
Conductivity, inductive measured none
Temperature adjustable, T > 2 C
0 Intrinsically safe supply and signal circuit with protection type: EEx ia IIC T6 passiveMax. input voltage UiMax. input current IiMax. input PiMax. inner inductivity Li
Max. inner capacity Ci
30 V100 mA750 mW221 H (output 1)neglegiable (output 2)1.2 nF (output 1)neglegiable (output 2)
0 Suitable for use in a FISCO systemMax. input voltage UiMax. input current IiMax. input PiMax. inner inductivity LiMax. inner capacity Ci
17,5 V380 mA5.32 W38.4 Hneglegiable
-
Liquiline M CM42
Endress+Hauser 11
Wiring
Housing grounding Plastic housing
" Caution! You must connect the "EMC barrier" to the foundation earth with a separate 4 mm2 functional earth line.
Stainless steel housing
" Caution! You must connect the outer earth connection of the housing to the foundation earth with a separate 4 mm2 line (GN/YE).
Supply and signal circuit 4 ... 20 mA / Hart
! Note! Connect the transmitter via a two-wire cable screened on each end.
a0003617
Housing grounding12
EMC barrier4 mm2 functional earth
a0003616
Housing grounding12
Outer earth connection4 mm2 line (GN/YE)
1
2
a0001091
View in device (CPU module)
a0001092
Wiring diagram
+
133134
234
233
232
231
134
133
132
131
+
133
134
4...
20m
A
-
Liquiline M CM42
12 Endress+Hauser
PROFIBUS PA and FOUNDATION Fieldbus
! Note! Always use a fieldbus cable that is grounded on both sides (device and PCS).You can connect Profibus and FOUNDATION Fieldbus devices in three ways: Screened two-wire cable, "Hard grounding"
(generally to be preferred to capacitive grounding) Screened two-wire cable, "Capacitive grounding"
(if there is a risk of high equalizing currents) Using the fieldbus connection socket (accessories)
Version 1: "Hard grounding"
1. Place the cable screen on the "EMC barrier".
2. Connect the cable wires as per the assignment.
Version 2: "Capacitive grounding"
1. Strip back the screen braiding, push the extension wire onto the exposed screen and fasten the clip:
2. Place the extension wire on the "EMC barrier".
3. Connect the cable wires as per the assignment.
a0004060
View in device (CPU module)
a0001640
Wiring diagram
a0004071
View in device (CPU module)
a0004073
Wiring diagram
997
998
+
+
133
132
998
997997
998
GN/YE
-
Liquiline M CM42
Endress+Hauser 13
Version 3: Fieldbus connection socket (accessories):
1. Screw the fieldbus connection socket (accessories) into the housing bushing.
2. Trim the connection cores of the socket to approx. 15 cm (5.9 ").
3. Connect the cable cores as per the assignment. In doing so, you must place the cable screen (GN/YE) on the "EMC barrier".
Supply voltage
Sensor connection
Explanation of abbreviations in the following diagrams:
! Note! Cable colours indicated as per IEC 757 (see CD-ROM). You must connect shielded connections and terminals with functional earth ()) (there is no protective earth
(*) for plastic housings).
a0002378
View in device (CPU module)
a0004062
Wiring diagram
Current output 4 to 20 mA without Hart:Current output 4 to 20 mA with Hart:PROFIBUS / FOUNDATION Fieldbus: (max. 22 mA)
12.5 to 30 V DC12.5 to 30 V DC9 to 32 V DC (non-hazardous area) / 9 to 17.5 V DC (hazardous area)
Abbreviation Meaning
pH Signal from pH membrane glassRef Signal from reference electrodeSrc SourceDrn DrainPM Potential matchingU+ Digital sensor supplyUCom A
Digital sensor communicationCom B Temperature signald.n.c. Do not connect!
997
998
+
-
Liquiline M CM42
14 Endress+Hauser
Sensor connection: pH / ORP Glass electrodes with PML (symmetrical)
Glass electrodes without PML (asymmetrical)
a0001088
View in devicea0001079
Wiring diagram
a0001089
View in devicea0001080
Wiring diagram
PM
BN
317
318
320
316
315
111
113
112
319
GNWHYE
112113111
320318317
BKBN
BK
BN
GN
WH
YE
CPK9
318
320
112
113
111
317
CM42
319
316
315
J
PM
Ref
pH
Sensor
CPK1
d.n.c.
BN
317
318
320
316
315
111
113
112
319
BK
GNWHYE
112113111
317BN
320
BK
BN
GN
WH
YE
318
320
112
113
111
317
CM42
319
315
316
CPK9 / CPK1
J
d.n.c.
Ref
pH
Sensor
pH
CPK1
-
Liquiline M CM42
Endress+Hauser 15
ISFET sensors with PML (symmetrical)
ISFET sensors without PML (asymmetrical)
a0001090
View in devicea0001076
Wiring diagram
a0001084
View in devicea0001077
Wiring diagram
PM
BN
317
318
320
316
315
111
113
112
319
GNWHYE
112113111
320318
316
BKBNRD
315
BK
BN
GN
WH
YE
CPK12
317
318
112
113
111
319
CM42
315
316
320
RD
Src
J
Ref
PM
Drn
Sensor
BN
d.n.c.
317
318
320
316
315
111
113
112
319
GNWHYE
112113111
320316
BKRD
315
BN
BK
BN
GN
WH
YE
317
318
112
113
111
319
CM42
315
316
320
RD
Src
RefDrn
J
d.n.c.
CPK12 Sensor
-
Liquiline M CM42
16 Endress+Hauser
Overview, pH signal assignment
Version: CM42-P...
Version: CM42-R...As version CM42-P and also:
Terminal Glass/redox ISFET Difference electrodewith PML without PML with PML without PML
112 (GN) (GN) (GN) (GN) (GN)
113 (WH) (WH) (WH) (WH) (WH)
111 (YE) (YE) (YE) (YE) (YE)
315 Src Src
316 Drn Drn
320 PM Ref PM Ref PM
318 Ref Ref
317 pH pH pH
319 pH
Terminal pH+redox pH+pH
with PML without PML with PML without PML
112 (GN) (GN) (GN) (GN)
113 (WH) (WH) (WH) (WH)
111 (YE) (YE) (YE) (YE)
315
316
320 PM Ref (pH) PM Ref (pH-1)
318 Ref (pH+ORP) Ref (ORP) Ref (pH-1 + pH-2) Ref (pH-2)
317 pH pH pH pH
319 ORP ORP pH pH
-
Liquiline M CM42
Endress+Hauser 17
Sensor connection: conductivity
Conductive sensors, two-cell sensors
Conductive sensors, four-cell sensors
a0001086
View in devicea0001083
Wiring diagram
a0002363
View in device (sensor module)a0002371
Wiring diagram
221
222
220
219
111
113
112112
BK
GNWHYE
113111219220
BN
BK
BN
GN
WH
YE
CPK9 / CYK71
221
222
112
113
111
CM42
220
219
J
Sensor
d.n.c.22
122
222
021
911
111
311
2112
BK
GNWHYE
113111219220222221
BK
221
222
112
113
111
CM42
220
219
J
Sensor
-
Liquiline M CM42
18 Endress+Hauser
Inductive sensors
Sensor connection: digital sensors
a0001085
View in devicea0001082
Wiring diagram
217
218
111
113
112
216
215215
216112113111
WH
GNWHYE
RD
218217 WH
RD
GN
WH
YE
CLK5
217
218
215
216
112
CM42
111
113
J
Sensor
a0001087
View in devicea0001078
Wiring diagram
588
587
388
387
198
197
188
187 BN
WHGN
187188197198
YE
-
Liquiline M CM42
Endress+Hauser 19
Performance characteristics
pH / ORP Reference temperature
25 C (77 F)
Measured value resolution
Maximum measured error1
Conductivity Reference temperature
25 C (77 F)adjustable from 5 to 100 C (23 to 212 F)2
Measured value resolution
pH:
Glass electrodes:ISFET sensors:
0.01 pH0.01 pH
ORP: 1 mV
Temperature:
NTC30k:Pt100 / Pt1000:
0.1 C (0.1 F)0.1 C (0.1 F)
SCS glass electrodes:
Glass:Reference:
100 k100
Leak current, ISFET sensors: 100 nA
1) acc. to DIN IEC 746 part 1, under nominal operating conditions
pH:
Glass electrodes:ISFET sensors:
0.02 pH0.02 pH
ORP: 1 mV
Temperature:
NTC30k:Pt100 / Pt1000:
0.5 K0.25 K
SCS glass electrodes:
Glass:Reference:
200 M200
Leak current, ISFET sensors: 100 nA
2) with software package "Advanced" or "Advanced Plus"
Conductivity < 0.1% from the measured value (4 digits)
minimum, at:cell constant k 0.15 cm-1cell constant k 1.5 cm-1cell constant k > 1.5 cm-1
0.001 S/cm0.01 S/cm0.1 S/cm
Temperature (Pt100 / Pt1000) 0.01 C (0.01 F)
Resistivityminimum
< 0.1% from the measured value (4 digits)1 k.cm
-
Liquiline M CM42
20 Endress+Hauser
Maximum measured error1
Temperature compensation
Temperature adjustment
Digital sensors Measured value resolution
1) acc. to DIN IEC 746 part 1, under nominal operating conditions
Conductivity, conductive measured
Two-polar sensors 0.5 % from measured value 0.01 S.k
Four-polar sensors 0.5 % from measured value 0.01 S.k 1)
1.0 % from measured value 0.01 S.k 2)
2.0 % from measured value 0.01 S.k 3)
1) from 0.1 S.k to 20 mS.k (max. cable length 15 m (49.22 ft))
2) from 10 S.k to 1000 mS.k (max. cable length 100 m (328.1 ft))
3) from 10 S.k to 1500 mS.k (max. cable length 100 m (328.1 ft))
Conductivity, inductive measured 0.5 % from measured value 1.7 S.k
Resistivity, conductive measured(two-cell sensors)
0.5 % from measured value 0.01 S.k 4)
4) from 10 M/k to 50 /k (max. cable length 15 m (49.22 ft)) from 200 k/k to 50 /k (max. cable length 50 m (164.05 ft))
Type of compensationnonelinearNaCl acc. to IEC 746-3natural waters acc. to IEC 7888Ultra-pure water NaClUltra-pure water HCl
Range
= 0.00 to 20.00 % / K0 to 100 C (32 to 212 F)0 to 35 C (32 to 95 F)0 to 100 C (32 to 212 F)0 to 100 C (32 to 212 F)
Temperature offset 5 to +5 C (23 to 41 F)
pH
Glass electrodesISFET sensors
0.01 pH0.01 pH
ORP 1 mV
Temperature 0.1 C (0.1 F)
-
Liquiline M CM42
Endress+Hauser 21
Installation
Mounting plate
Weather protection cover Plastic housing
Stainless steel housing
a0001676-en
Weather protection cover
a0001075-en
Mounting plate
a0001671-en
Weather protection cover
320
/ 12.
60
270 / 10.63
mm / inch
300 / 11.81
-
Liquiline M CM42
22 Endress+Hauser
Installation in Ex area
Environment
Ambient temperature range Non-hazardous area version30 to 70 C (22 to 158 F)
Hazardous area versionT6: 20 to 50 C (4 to 122 F)T4: 20 to 55 C (4 to 131 F)
Ambient temperature limits -30 to +80 C (22 to 176 F)
Storage temperature 40 to 80 C (40 to 176 F)
Electromagnetic compatibility Interference emission and interference immunity as per EN 61326: 2004
Ingress protection IP 67
Relative humidity 10 to 95%, not condensing
a0001312-en
Installation in Ex area1234
Sensor in Ex versionIntrinsically safe sensor circuit EEx iaTransmitterSupply and signal circuit EEx ib (4...20 mA)
5678
Active barrier, e.g. Preline RN221Signal line Hart/PROFIBUS/FFFieldcare via PROFIBUS/FFHart handheld terminal
-
Liquiline M CM42
Endress+Hauser 23
Mechanical construction
Dimensions Plastic housing
Stainless steel housing
a0001074-en
Plastic housing
a0001073-en
Dimensions
145
/ 5.7
1
134
/ 5.2
811
0/ 4
.33
132.5 / 5.22136.5 / 5.37
160
/6.3
0
134.5 / 5.30136.5 / 5.37
26 /1.02
174 / 6.85
174
/6.8
510
/0.
39
3/ 0
.12
220 / 8.66
95/3
.74
mm / inch
-
Liquiline M CM42
24 Endress+Hauser
Weight Plastic housing
1.5 kg (3.3 lb)
Stainless steel housing
2.1 kg (4.6 lb)
Material Plastic housing
Stainless steel housing
Human interface
Operating elements
Housing:Housing seals:
PolycarbonateEPDM
Housing:Housing seals:
Stainless steel 1.4301 (AISI 304)Foamed silicone, EPDM
a0001301
Overview of operation
1 Display, current display: pH measuring mode2 Navigator3-6 Soft keys7 Soft key function (depends on menu)
-
Liquiline M CM42
Endress+Hauser 25
Ordering information
Product structure Sensor inputA No ModuleC Conductivity, conductive measurementI Conductivity, inductive measurementM Digital sensor, MemosensP pH (glass/ISFET) or ORPR pH (glass/ORP redundant) or ISFET
ApprovalA Non-hazardous area, CSA GPG Atex II (1)2G EEx ia/ib IIC T6 / II 3 DH Atex II 3G EEx na II T6N NEPSI Ex nA II T6P FM IS NI Cl. I, II, III, Div. 1&2, Groups A-GS CSA IS NI Cl. I, II, III, Div. 1&2, Groups A-GT TIIS
CertificateA not selectedB Test certificate acc. to EN 10204, 3.1C Test certificate acc. to EN 10204, 3.1, factory calibration certificate
Output0 1 current output 4 to 20 mA, Hart
1 2 current outputs 4 to 20 mA, 1x Hart
2 PROFIBUS PA3 FOUNDATION Fieldbus
Housing0 Plastic1 Stainless steel 1.4301 (AISI 304)
Cable entry0 M20 x 1.51 NPT "2 G
SoftwareEA Standard versionEB Advanced versionEC Advanced Plus version
Device languageC Zh/En (Chinese docs)D De/En (German docs)E En/De (English docs)F Fr/En (French docs)J Ja/En (Japanese docs)N Nl/En (Dutch docs)P Pl/En (Polish docs)S Es/En (Spanish docs)T It/En (Italian docs)
Documentation0 Installation and configuration1 Installation and configuration + safety manual
Additional equipment0 Basic version
CM42- complete order code
-
Liquiline M CM42
26 Endress+Hauser
Scope of delivery The scope of delivery comprises: A transmitter acc. to the ordered version A mounting plate including 4 screws A cable gland plate and 2 cable glands (in the packaging) A sticker sheet (embedded in the housing, containing nameplates and terminal assignment stickers) A test certificate acc. to EN 10204-3.1 (optional) A factory calibration certificate An Operating Instructions BA381C "Commissioning" An Operating Instructions BA382C "Operation" A Safety Manual acc. to SIL 2 (optional) A CD ROM with additional documentation
Certificates and approvals
4 approval Declaration of conformityThe product meets the legal requirements of the harmonized European standards. The manufacturer confirms compliance with the standards by affixing the 4 symbol.
Ex approval Depending on the version ordered: ATEX II (1)2G, EEx ia/ib IIC T6 / II 3D ATEX II 3G, EEx na II T6 NEPSI Ex nA II T6 CSA IS NI Cl.I, II, III, Div. 1&2, Grps. A-G FM IS NI Cl.I, II, III, Div. 1&2, Grps. A-G TIIS
-
Liquiline M CM42
Endress+Hauser 27
Accessories
pH / ORP Glass electrodes
Orbisint CPS11/CPS11DpH electrode for process applications, with PTFE diaphragm, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI028/C07/en)
Orbisint CPS12/CPS12DORP electrode for process applications, with PTFE diaphragm, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI367/C07/en)
Ceraliquid CPS41/CPS41DpH electrode with ceramics diaphragm and KCl liquid electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI079/C07/en)
Ceraliquid CPS42/CPS42DORP electrode with ceramics diaphragm and KCl liquid electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI079/C07/en)
Ceragel CPS71/CPS71DpH electrode with double chamber reference system and integrated bridge electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI245/C07/en)
Ceragel CPS72/CPS72DORP electrode with double chamber reference system and integrated bridge electrolyte, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI374/C07/en)
Orbipore CPS91/CPS91DpH electrode with open aperture for media with high dirt load, Memosens technology as option;Ordering acc. to product structure, see Technical Information (TI375C/07/en)
ISFET sensors
Tophit CPS471Sterilisable and autoclavable ISFET sensor for food and pharmaceuticals, process technology, water treatment and biotechnology;Ordering acc. to product structure, see Technical Information (TI283/C07/en)
Tophit CPS441Sterilisable ISFET sensor for media with low conductivity, with liquid KCl electrolyte;Ordering acc. to product structure, see Technical Information (TI352/C07/en)
Tophit CPS491ISFET sensor with open aperture for media with high dirt load;Ordering acc. to product structure, see Technical Information (TI377/C07/en)
-
Liquiline M CM42
28 Endress+Hauser
Assemblies
Cleanfit P CPA471Retractable assembly for tank and pipe installation,
Cleanfit P CPA473Pneumatically or manually driven retractable assembly with ball valve for reliable process termination, parts in contact with medium of 1.4404 (AISI 316L) stainless steel
Cleanfit P CPA474Pneumatically or manually driven retractable assembly with ball valve for reliable process termination, parts in contact with medium of PP, PEEK or PVDF
Cleanfit H CPA475Retractable assembly for tank and pipe installation under sterile conditions,
Dipfit W CPA111Immersion and installation assembly for open and closed tanks,
Unifit H CPA442Installation assembly for food industry, biotechnology and pharmaceutical industry, with EHEDG and 3A certificate,
! Note! All the assemblies can be ordered using the product structure. Please refer to the Technical Information of the desired assembly in question.
a0001305
Fig. 1: Cleanfit P CPA471a0001306
Fig. 2: Cleanfit P CPA473a0001307
Fig. 3: Cleanfit P CPA474
a0001308
Fig. 4: Cleanfit H CPA475a0001309
Fig. 5: Dipfit W CPA111a0001310
Fig. 6: Unifit H CPA442
-
Liquiline M CM42
Endress+Hauser 29
Measuring cables
CPK9 special measuring cableFor sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CPK12 special measuring cableFor pH/redox glass electrodes and ISFET sensors with TOP68 plug-in headOrdering acc. to product structure, see Technical Information (TI118C/07/en)
Buffer solutions
Technical buffer solutions, accuracy 0.02 pH, acc. to NIST/DIN pH 4.0 red, 100 ml (0.026 US gal.), order no. CPY 2-0 pH 4.0 red, 1000 ml (0.264 US gal.), order no. CPY 2-1 pH 7.0 green, 100 ml (0.026 US gal.), order no. CPY 2-2 pH 7.0 green, 1000 ml (0.264 US gal.), order no. CPY 2-3
Technical buffer solutions for single use, accuracy 0.02 pH, acc. to NIST/DIN pH 4.0 20 x 20 ml (0.005 US gal.), order no. CPY 2-D pH 7.0 20 x 20 ml (0.005 US gal.), order no. CPY 2-E
Technical buffer solutions for ORP electrodes +225 mV, pH 7, 100 ml (0.026 US gal.); order no. CPY3-0 +468 mV, pH 0, 100 ml (0.026 US gal.); order no. CPY3-1
Conductivity Conductive sensors
Condumax W CLS12Conductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to version, see Technical Information TI 082/C07/en
Condumax W CLS13Conductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to version, see Technical Information TI 083/C07/en
Condumax W CLS15Conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);Ordering acc. to version, see Technical Information TI 109/C07/en
Condumax H CLS16Hygienic conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);Ordering acc. to version, see Technical Information TI 227/C07/en
Condumax W CLS19Conductive conductivity sensor for pure and ultra-pure water applications;Ordering acc. to version, see Technical Information TI 110/C07/en
Condumax W CLS21Conductive conductivity sensor for applications with middle to high conductivity(incl. Ex); Ordering acc. to version, see Technical Information TI 085/C07/en
Inductive sensors
Indumax P CLS50Inductive conductivity sensor for standard, Ex and high temperature applications;Ordering acc. to the sensor version, see Technical Information (TI118C/07/en)
Indumax H CLS52Inductive conductivity sensor with short response time in hygienic design;Ordering acc. to the sensor version, see Technical Information (TI167C/07/en)
-
Liquiline M CM42
30 Endress+Hauser
Assemblies
Dipfit W CLA111Immersion assembly for open and closed tanks with flange DN 100;Order depending on version, see Technical Information (TI135C/07/en)
Dipfit P CLA140For the inductive sensor CLS50Immersion assembly with flange connection for processes with high requirements;Order depending on version, see Technical Information (TI196C/07/en)
Measuring cables
Special measuring cable / extension cable CYK71 for two-polar conductivity sensors with integrated temperature sensor, one low-noise coaxial line, four auxiliary cores at 0.75 mm2 each with a common screen,
outer diameter 7 mm (0.25") sold by the metre:
(minimum) length 5 m / 15 ft; order no. 50085333length 5 m / 15 ft; order no. 50088280length 10 m / 30 ft; order no. 50088281length 50 m / 150 ft; order no. 50088284length 100 m / 300 ft; order no. 50088285
Special measuring cable / extension cable CYK71-Ex for Ex applications, see CYK71, but with a blue sheath
sold by the metre, minimum length 5 m / 15 ft; order no. 50085673
Calibration solutions
Precision calibration solutions, acc. to SRM (Standard reference material) of NIST, error limit 0.5 %, reference temperature 25 C (77 F), with temperature table CLY11-A, 74.0 S/cm, 500 ml (0.132 Us.gal); order no. 50081902 CLY11-B, 149.6 S/cm, 500 ml (0.132 Us.gal); order no. 50081903 CLY11-C, 1.406 mS/cm, 500 ml (0.132 Us.gal); order no. 50081904 CLY11-D, 12.64 mS/cm, 500 ml (0.132 Us.gal); order no. 50081905 CLY11-E, 107.0 mS/cm, 500 ml (0.132 Us.gal); order no. 50081906
Digital sensors (additional) Measuring cables
CYK10 Memosens data cableFor digital pH sensors with Memosens technology (CPSxxD)Ordering according to product structure, see Technical Information (TI376C/07/en)
a0001303
Fig. 7: Dipfit W CLA111a0001304
Fig. 8: Dipfit P CLA140
-
Documentation
Operating Instructions CM42 Liquiline M CM42, Operating Instructions, KA381C/07/en and BA382C/07/enThe Operating Instructions correspond to the device version.Methods for (re-) ordering: Re-order the KA/BA by specifying the serial number of your device (via your local Sales Office) Order a specific KA/BA by specifying a valid Liquiline order code (via Internet).
International Head Quarters
Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland
Tel. +49 76 21 9 75 02
Fax +49 76 21 9 75 34 5
www.endress.com
TI381C/07/en/05.05Printed in Germany / FM+SGML 6.0 / POD/
51517610
-
ApplicationYour benefitsGeneral featuresModular designQuick setupNavigator and plain textSensor monitorProcess check system (PCS)Software packagesSafetySafety integration level acc. to IEC 61508 (SIL2)Alarm LEDCode protectionReordering validated software
Special featurespH / ORPSensor Condition Check (SCC)Sensor Check System (SCS)
ConductivityPolarisation monitoringUnited States Pharmacopeia (USP) and European Pharmacopoeia (EP)
Digital sensorsMemosens
Measuring systemInputpH / ORPMeasured variableMeasuring rangeCable specificationApplicable temperature sensorsEx specification
ConductivityMeasured variableMeasuring rangeCable specificationTemperature sensorEx specification, conductive sensorsEx specification, inductive sensors
Digital sensorsMeasured variableMeasuring rangeCable specificationEx specification
OutputOutput signalSignal on alarmLoadOutput signal rangeEx specification currentoutput 4/20mAEx specification PROFIBUSPA and FOUNDATION Fieldbus
WiringHousing groundingPlastic housingStainless steel housing
Supply and signal circuit4 ... 20 mA / HartPROFIBUS PA and FOUNDATION FieldbusSupply voltageSensor connection
Sensor connection: pH/ORPGlass electrodes with PML (symmetrical)Glass electrodes without PML (asymmetrical)ISFET sensors with PML (symmetrical)ISFET sensors without PML (asymmetrical)Overview, pH signal assignment
Sensor connection: conductivityConductive sensors, two-cell sensorsConductive sensors, four-cell sensorsInductive sensors
Sensor connection: digitalsensors
Performance characteristicspH / ORPReference temperatureMeasured value resolutionMaximum measured error
ConductivityReference temperatureMeasured value resolutionMaximum measured errorTemperature compensationTemperature adjustment
Digital sensorsMeasured value resolution
InstallationInstallationMounting plateWeather protection coverPlastic housingStainless steel housing
Installation in Ex area
EnvironmentEnvironmentAmbient temperature rangeAmbient temperature limitsStorage temperatureElectromagnetic compatibilityIngress protectionRelative humidity
Mechanical constructionDimensionsPlastic housingStainless steel housing
WeightPlastic housingStainless steel housing
MaterialPlastic housingStainless steel housing
Human interfaceOperating elements
Ordering informationProduct structureScope of delivery
Certificates and approvals4 approvalEx approval
AccessoriespH / ORPGlass electrodesISFET sensorsAssembliesMeasuring cablesBuffer solutions
ConductivityConductive sensorsInductive sensorsAssembliesMeasuring cablesCalibration solutions
Digital sensors (additional)Measuring cables
DocumentationOperating Instructions CM42