Line Isolation & Overload Monitor E05 Series Installation Instructions AL ARM MUTED OVERLOAD WAR NING PHC WARNING PR OSPE CTIVE HAZ ARD CUR R E NT LEVE L E0 5 LI NE ISOLATION AN D OVERLOAD MONITOR MU TE TE S T 1 2 3 4 5 6 8 7 9 mA POWER E. S .P.

Transcript of Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf ·...

Page 1: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

Line Isolation &Overload Monitor

E05 Series

Installation Instructions

ALARM MUTED

OVER LOAD WAR NING

PH C WARNING

PR OSPE CTIVE

HAZ ARD

CUR R E NT

LEVE L

E0 5

LI NE

IS OLATION

AN D

OV ER LOAD

MONITOR

MU TE

TE S T

12

34

56

87

9

mA

PO WE R

E. S.P .

Page 2: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

E05 Series Line Isolation & Overload Monitor Installation Instructions

© Clipsal Australia Pty Ltd

Contents

1.0 Description .................................................................................................. 3

2.0 Wiring Instructions ..................................................................................... 4

3.0 Operating Instructions ............................................................................... 5

4.0 Specifications ............................................................................................. 6

5.0 Two-Year Warranty ..................................................................................... 7

Copyright NoticeThe concepts, products and designs described in this document are the subject of international patents, and protected by international law.

Copyright 2010 Clipsal Australia Pty Ltd. All rights reserved.

TrademarksClipsal is a registered trademark of Schneider Electric.All other logos and trademarks are the property of their respective owners.

DisclaimerClipsal Australia Pty Ltd reserves the right to change specifi cations or designs described in this manual without notice and without obligation.

Page 3: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

3 of 8© Clipsal Australia Pty Ltd

E05 Series Line Isolator & Overload Monitor Installation Instructions

1.0

A

N

E MEN POINT

ISO L AT IO NT RANSFO RM E R

L1

L2

LIOM

L1

L2

RFAULT

POTENTIALFAULT

Description

The E05 Line Isolation and Overload Monitor in conjunction with the E14 SeriesIsolation Transformer, forms the basis of an isolated supply system whichprovides electrocution and overload protection as detailed in AS/NZS 3003‘Electrical Installations – Patient treatment areas of hospitals and medical anddental practices.’

An isolated supply is a continuous supply. The isolation transformer isolates thepower from earth therefore a fault on both lines of the secondary transformerwinding needs to occur before a fault current flows. A fault on only one line andearth will not cause a fault current to flow.

The E05 Line Isolation and Overload Monitor (LIOM) enables the integrity of anisolated power supply system to be continuously monitored. The LIOM is able todetect when one fault occurs therefore is able to warn of a potentially hazardoussituation before any fault actually flows. The first fault is known asProspective Hazard Current (PHC). The LIOM dislays the PHC on an LCDbargraph display and when 5 mA PHC is reached a mutable alarm sounds alongwith a flashing LED indicator labelled ‘PHC Warning.’

If a PHC alarm condition arises, the medical staff can, in light of theseriousness of the patient’s condition, choose to continue with the procedure. Inthis situation, the surgeon is relying on a second fault not occurring that wouldturn the PHC into a real flow of current. During this time measures can be takento rectify the fault condition by, for example, disconnecting all non-essentialequipment.

In addition to the LIOM’s monitoring isolation, it also warns of impendingoverload, thus averting problems which could result from a loss of supply. Whenthe overload level is reached a mutable alarm sounds along with a ‘OverloadWarning’ flashing LED indicator.

Page 4: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

4 of 8

E05 Series Line Isolation & Overload Monitor Installation Instructions

© Clipsal Australia Pty Ltd

2.0 Wiring Instructions

2 POL E CIRCUI TBREAKER

TO SUITHI G H IN RUS H

CURRENTES P

ISOL ATION TRANSF ORM E RE14 3.6kV A(15A )

ORE14 4.8kV A(20A )

LI NE 1

LI NE 2

ES P E05 L IO M ( 3.6k VA OR 4.8k VA)

LI NE 1

LI NE 2

E ARTH

A

N

E

LIN

E 1

IN

LIN

E 1

OU

T

LIN

E 2

IN

LIN

E 2

OU

T

N/O

COM

MO

N

N/C

EA

RTH

TE

RMIN

AT

ION

BL

OC

KRE

AR

OF

LIO

M

V OLT FRE EALARM

CONTACTS

The E05 must be installed by a qualified electrician in accordance with thedetailed instructions here and the requirements of AS/NZS 3003 &AS/NZS 3000.

Connect the E05 and associated electrical components as per the wiringdiagram, making sure to take into consideration the following installation guidelines.

• Long runs of cable between the transformer and the LIOM and/or long runs between the LIOM and the outlets should be avoided. Due to capacitive coupling between cables and between cables and conduit, runs longer than 50m can result in high standing PHC (ie, >2 mA).

• Parallel runs of cables from two or more isolated supplies should also be avoided. The ‘crosstalk’ between parallel cable may cause rhythmic fluctuations on the units’ displays. Where parallel wiring cannot be avoided, wiring should include an earth (eg, TPS). The earth conductor should be earthed only at one end of its run. While running cable in this manner will lessen crosstalk problems, it may increase standing PHC.

• Due to the LIOM’s monitoring the load current ‘on line’ the LIOM must be the first item connected to the secondary winding of the isolation transformer. All items to be powered from the isolated supply must connect back to the terminals of the LIOM labelled ‘Active 1 Out’ and ‘Active 2 Out.’ Failure to do this will result in incorrect measurements of the load current.

• In some installations there might be a high inrush current when power is first applied to the isolation transformer therefore a suitable circuit breaker needs to be chosen. It is recommended to use a Heinemann Electric CF series circuit breaker.

Page 5: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

5 of 8© Clipsal Australia Pty Ltd

E05 Series Line Isolator & Overload Monitor Installation Instructions

3.0

ALARM MUT E D

OVE R LOAD WAR NING

PH C WAR NING

PR OSPE C TIVEHAZ ARDCUR R E NTLEVE L

E0 5LI NE

IS OLAT IONAN D

OV ER LOADMONITOR

MU TE

TE S T

123456

87

9

mAPO W E R

E. S .P .

Operating Instructions

1. Check power is connected to the E05 LIOM by confirming the power indicator is on, which is located at the bottom of the LCD bargraph display.

2. Monitor the level of Prospective Hazard Current (PHC) using the LCD bargraph display. A correctly functioning system should display little or no PHC. When a PHC of at least 5 mA is detected the E05 will alarm both audibly and visually (PHC Warning Indicator) but the supply will not be interrupted. When this alarm condition arises the surgeon can, in light of the seriousness of the patient’s condition, choose to continue with the procedure. In this situation the surgeon is relying on a second fault not occurring that would turn the PHC into a real current flow. During this time measures can be taken to rectify the fault condition by, for example, disconnecting all non-essential equipment.

3. Monitor when the load current rating of the isolation transformer has been exceeded by monitoring the Overload Warning indicator. When the E05 detects that the load current rating has been exceeded the E05 will alarm both audibly and visually (Overload Warning indicator).

4. If required to mute the audible alarm then press the Mute button. The PHC Warning indicator or Overload Warning indicator will remain on until the fault is removed.

5. Test the PHC monitoring function of the unit by pressing the Test button. In the test mode the E05 should display approximately 7.5 mA PHC on the LCD bargraph display. The E05 should also go into Audible Alarm, and the PHC Warning indicator should turn on.

Page 6: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

6 of 8

E05 Series Line Isolation & Overload Monitor Installation Instructions

© Clipsal Australia Pty Ltd

4.0 Specifications

Parameter Description

Operational Voltage 240Vac 50 Hz Prospectiv e Ha zard Current Display LCD Bargraph (1 to 9mA) Prospectiv e Ha zard Current Alarm Point 5mA Ov erload Alarm Point 15A(3.6 kVA), 20A (4.8kVA ), or 30A (7.2 kVA)

Factory set to either of the above Audible Alarm 82dB typical at 1 metre Remote Alarm Contac ts Volt free, changeove r, 5A relay contacts Operating Temperature 10° C to 35° C

Surface Mount stainless steel Flush M ount stainless steel

Standard Finishes

Duct Mount powder coate d (Sand ) Surface Mount 220 x 275 x 90mm (H x W x D) Flush Mount 220 x 275 x 70 Duct Mount 221 x 200 x 70

Dimensions

Panel Mount 198 x 190 x 70

Page 7: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

7 of 8© Clipsal Australia Pty Ltd

E05 Series Line Isolator & Overload Monitor Installation Instructions

5.0 Two-Year Warranty

The benefi ts conferred herein are in addition to, and in no way shall be deemed to derogate, either expressly or by implication, any or all other rights and remedies in respect to the Clipsal product, which the consumer has under the Commonwealth Trade Practices Act or any other similar State or Territory Laws.

The warrantor is Clipsal Australia Pty Ltd. With registered offi ces in all Australian States.

This Clipsal product is guaranteed against faulty workmanship and materials for a period of two (2) years from the date of installation.

Clipsal Australia Pty Ltd reserves the right, at its discretion, to either repair (free of parts and labour charges), replace or offer a refund, in respect to any article found to be faulty due to materials, parts or workmanship.

This warranty is expressly subject to the Clipsal product being installed, wired, tested, operated and used in accordance with the manufacturer’s instructions.

All costs of a claim shall be met by Clipsal Australia Pty Ltd, however should the product that is the subject of the claim be found in good working order, all such costs shall be met by the claimant.

When making a claim the consumer shall forward the Clipsal product to the nearest offi ce of Clipsal Australia Pty Ltd with adequate particulars of the defect within 28 days of the fault occurring. The product should be returned securely packed, complete with details of the date and place of purchase, description of load, and circumstances of malfunction.

Warranty Statement

Page 8: Line Isolation & Overload Monitorupdates.clipsal.com/ClipsalOnline/Files/Brochures/W0001585.pdf · E MEN POINT ISO LATION TRANSFORM E R L1 L2 LIOM L1 L2 R FAULT POTENTIAL FAULT Description

January 2010 1035955

Clipsal Australia Pty Ltd reserves the right to change specifications, modify designs and dis con tin ue items without incurring obligation and whilst every effort is made to ensure that descriptions, specifications and other in for ma tion in this catalogue are correct, no warranty is given in respect thereof and the company shall not be liable for any error therein.

© Clipsal Australia Pty Ltd. The identifi ed trademarks and copyrights are the property of Clipsal Australia Pty Ltd unless otherwise noted.

Product of Clipsal Australia Pty LtdA member of Schneider Electric

Contact us: clipsal.com/feedback

National Customer Care Enquiries:

Tel 1300 2025 25 Fax 1300 2025 56

clipsal.com

Clipsal and Moduline - a winning combination!Moduline is the company behind the Medilec brand. Offering an extensive range of premium electrical products to suit hospital and other medical applications, the Medilec team includes some of Australia’s leading designers and manufacturers of hospital grade electrical safety equipment. The range includes high quality residual current devices that provide high levels of protection against electric shock and electrocution.

Clipsal and Moduline have now teamed up to provide the best hospital electrical solutions available on the Australian market. Backed by Clipsal, and with more than 80 years combined experience it is easy to see why a team like Moduline and Medilec are clear market leaders in providing innovative hospital electrical solutions.