Lime Density Display Quick Set Up Notes for PPA
Calibration with Thermal Fisher Densometer:
In-line Densometer cannot be rigged in or Blender cannot have fluid in Densometer.
On lime Display Calibration Screen with no water in Densometer touch empty pipe button on Screen. Air voltage should read around 9.6 volts.
Fill detector to top with water and Press V water button on screen. Voltage should be around five volts on In-line and for blender denso around 2.5 volts on an eight inch. It really depends on what jumpers are connected on FSU in Densometer.
Calibration is finished.
Note: Can use real time simulation to change V empty and V water readings.
On front screen use Calculated Density not measured. Measured is good if water is showing 8.33 densities. Calculated will graph better than measured.
Percent Adjust on Cal Screen:
Percent adjust should be Zero
Average 3-5
Scales on Cal Screen:
Kg/m3 min 0 - Max 2650
PPA min 0 - Max 22.1
SGU min .8 – Max 2.8
KGA min 0 – Max 2650
LBS/GAL min 0 – Max 22.1
PPA Poly Screen: Touch on Top right corner of Lime Display.
PPA POLY
Value % out
2.207 9.987
4.420 20.000
6.627 29.987
8.840 40.000
11.047 49.987
13.260 60.000
15.467 69.987
17.680 80.000
19.887 89.987
22.100 100.00
On Accufrac computer or DAC software same screen as PPA Poly on right is STEP button. Step through 1 thru 10 and Accufrac if scaled right should show PPA for each step.
Make sure Analog One Output is used for PPA on front screen . The cable from densometer should be connected in Analog one in back of Line display.
Front Screen:
0.0 PPA
PPA – Analog Output one
SGU- Analog Output two
LBS/Gallon- US English Units (8.33 for Water)
How to set up IN-LINE Densometer in Accufrac:
Select Analog One on Accufrac computer
Select Scale Min-0 and Max to 22.1 using PPA
On Calibration Screen on Lime display Simulate 4ma and Set MIN DAC Counts on Accufrac. Turn off simulation.
Simulate 20ma on Lime display and Set Max DAC counts on Accufrac computer. Turn off simulation.
Now simulate 12 ma on Lime display and Accufrac density should display 11.05 PPA. Turn off Simulation.
Accufrac set up finished.
Locked Out from Calibration Screen On Lime Display:
If locked out from Cal Screen Touch on bottom left corner twice on display. Touch by lime Instruments version white letters. May take a few attempts.
E-Tech.
Sanjel USA
David Aakre