LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS...

34
LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance Inspection Procedure The Performance Inspection Procedure (PIP) is a set of manual and computer- assisted test procedures used by authorized Physio-Control service representatives. It will be used as an operational closed-case evaluation of the defibrillator. This section describes the test procedures you will perform to determine if the device is operating within the required specifications. Note: Use the Performance Inspection Procedure Checklist to record the results of each test to ensure compliance with each step of this process. Note: Perform the procedures in the order presented. TABLE OF CONTENTS PIP – Scope and Applicability PIP – Resource Requirements PIP – Test Equipment Requirements PIP – Instructions PIP – Checklist 3202983_J_PIP-TCP.pdf Page 1 of 34

Transcript of LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS...

Page 1: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents

Previous Page Table of Contents Back Index Next Page

Performance Inspection Procedure

The Performance Inspection Procedure (PIP) is a set of manual and computer-assisted test procedures used by authorized Physio-Control service representatives. It will be used as an operational closed-case evaluation of the defibrillator. This section describes the test procedures you will perform to determine if the device is operating within the required specifications.

Note: Use the Performance Inspection Procedure Checklist to record the results of each test to ensure compliance with each step of this process.

Note: Perform the procedures in the order presented.

TABLE OF CONTENTSPIP – Scope and ApplicabilityPIP – Resource RequirementsPIP – Test Equipment RequirementsPIP – InstructionsPIP – Checklist

3202983_J_PIP-TCP.pdf Page 1 of 34

Page 2: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PLUS/EXPRESS Defibrillator exclusively. m the combination of manual and computer structions section of this electronic service

ed from start to finish in the order

ements for a listing of the necessary equipment verification, workstation power,

equirements for a listing of test equired to complete the PIP.ur results.

32 Page 2 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Scope and Applicability

The PIP applies to the LIFEPAK CRTo complete the PIP, you must perforassisted tests outlined in the PIP – Inmanual. All PIP tests must be performpresented.Refer to the PIP – Resource Requirqualifications for PIP equipment, testand personnel. Refer to the PIP – Test Equipment Requipment, including specifications, rUse the PIP – Checklist to record yo

02983_J_PIP-TCP.pdf

Page 3: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

nts for PIP equipment, PIP test equipment nd PIP personnel qualifications.

PI equipment listed in the PIP – Test hough the table lists specific test equipment equivalent specifications may be

PIVe

e PIP must have a current calibration label. by a certified calibration facility.

PIPo

must be connected to a grounded power

32 Page 3 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Resource Requirements

This section describes the requiremeverification, PIP workstation power, a

P – Equipment To perform the PIP, you must use theEquipment Requirements table. Altby manufacturer, test equipment withsubstituted.

P – Test Equipment rification

All test equipment used to perform thThe calibration label must be issued

P – Workstation wer

The ac line power to the workstation source.

02983_J_PIP-TCP.pdf

Page 4: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

PEq ManufacturerDe

t be

Fluke - Impulse 7000DP

CaTe

3009139

De IET Model RS-200

Sa

t be

Fluke - ESA612

C 3011030-00El 3205979

TePI

3202957 revision G

32 Page 4 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Test Equipment Requirementsuipment Specificationsfibrillator analyzer Power range: 0-450 J

Load resistance: 50 Accuracy and/or guard banding mussufficient to ensure test limits.Waveforms: NSR and VF

ble, Quik-Combo to Snaprminationcade resistance box Range: 0-500 Resolution: 1

Accuracy: 1%fety analyzer 110 or 220 Vac line voltage

Current range: 0-1999 µAAccuracy and/or guard banding mussufficient to ensure test limits.

able Assembly, FAST-PATCHectrode Test Post Adapter (2 ea)

st Software, LIFEPAK CR Series P-TCP

02983_J_PIP-TCP.pdf

Page 5: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

Eq Manufacturer

IrD iFoundry IFSYS-8003ACHRe

3201616

32 Page 5 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Test Equipment Requirements

uipment Specifications

A Dongle IrDA FIR USB AdapterARGE-PAK/QUIK-PAK placement Kit

02983_J_PIP-TCP.pdf

Page 6: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIn

ions for performing the Performance

ented.inning of the procedure.

n Procedure Checklist to record your

PIIn

ion, described as follows.e device for the following:

(Continued on next page)32 Page 6 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – General structions

This section lists the general instructInspection Procedure (PIP).• Perform the PIP in the order pres• Always start the PIP from the beg• Use the Performance Inspectio

results.

P – Physical spection

Perform the exterior physical inspect1. Inspect all exterior surfaces of th

– Damage– Excessive wear– Improper mechanical function– Damaged connectors

02983_J_PIP-TCP.pdf

Page 7: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIn

listen for loose or rattling hardware. If loose te and correct or contact Physio-Control .

ntegrity of case seal).

or corroded connectors or pins. Clean or

l inspection on the PIP Checklist.

32 Page 7 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Physical spection (continued)

2. Lift and turn over the device and hardware is suspected, investigabefore performing this procedure

3. Inspect the case for separation (i

4. Check for debris or bent, broken,replace damaged components.

5. Record the results of the physica

02983_J_PIP-TCP.pdf

Page 8: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI f the device before you turn it on, and how re.e to proceed with inspection when the OK GE-PAK battery charger and QUIK-PAK the inspection. Make a note in the PIP and continue with inspection unless

INSTRUCTION

Proceed with PIP

Proceed with PIP

days ago. Delay PIP

days ago. Proceed with PIP

t. ERROR - do not continue

days ago and CHARGE-PAK Delay PIP

days ago and CHARGE-PAK ERROR - do not continue

t or CHARGE-PAK not replaced ERROR - do not continue

ERROR - do not continue

32 Page 8 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Initialization test This test confirms the overall status oto evaluate each step in the proceduNote: In some cases, it is acceptablsymbol is not visible since the CHARelectrodes are replaced at the end ofchecklist of which symbols are visibleotherwise indicated in the chart.

SYMBOL VISIBLE DEVICE HISTORY

OK n/a

BATTERY n/a

BATTERY & ATTENTION

Device used to treat patient < 3

Device used to treat patient > 3

Device NOT used to treat patien

ATTENTION

Device used to treat patient < 3 replaced

Device used to treat patient > 3 replaced

Device NOT used to treat patien

WRENCH n/a

02983_J_PIP-TCP.pdf

Page 9: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c

ot connected to the device.

in the Readiness Display.rs, the device must be serviced. Do not ction if the WRENCH symbol is illuminated.

les of symbols)

ization test, ensure that the device has seconds.

l is visible, the internal device battery is -PAK battery charger. Wait until the

ished before continuing with the PIP.

!

(Continued on next page)32 Page 9 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Initialization test ontinued)

Note: Ensure a patient load is n

• Verify the OK symbol appears• If the WRENCH symbol appea

attempt to perform this inspe

(Examp

• Before you perform the initialbeen turned off for at least 60

• If only the ATTENTION symborecharging from the CHARGEATTENTION symbol is extingu

OK

02983_J_PIP-TCP.pdf

Page 10: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c ry charger is installed in the device.

utton to open the lid and turn the device

NTION, and WRENCH symbols momentarily ay.ates briefly in semi-auto unit. (Move the

shows three black squares at the end of ymbols will be visible during this interval.y be visible. Make a note in the PIP

dure.

s flash red as the power cycles during

and that voice prompts are clearly audible,

32 Page 10 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Initialization test ontinued)

To perform the initialization test:1. Make sure a CHARGE-PAK batte2. Press the LID RELEASE/ON-OFF b

ON.3. Verify that the OK, BATTERY, ATTE

illuminate in the Readiness Displ4. Verify that the SHOCK LED illumin

electrode packet to view LED.)5. Verify that the Readiness Display

the startup sequence. No other sNote: The ATTENTION symbol machecklist but continue with proce

6. Verify that the electrode indicatorstartup.

7. Verify that an alert tone is heard which may include:

02983_J_PIP-TCP.pdf

Page 11: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c

T

TIC AND APPLY PADS TO EXPOSED CHEST

ATIENT’S CHEST

IC LY AS SHOWN IN THE PICTURE

STIC LY AS SHOWN IN THE PICTURE

vary depending on the device

32 Page 11 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Initialization test ontinued)

A- Original Voice Prompt Version:• CALL FOR HELP NOW• REMOVE CLOTHING FROM CHES• PULL RED HANDLE TO OPEN BAG• PEEL EACH PAD OFF BLUE PLAS

B- Current Voice Prompt Version:• CALL FOR HELP NOW• REMOVE ALL CLOTHING FROM P• PULL RED HANDLE TO OPEN BAG• LOOK AT PICTURES ON PADS• PEEL ONE PAD OFF BLUE PLAST• APPLY PAD TO BARE SKIN EXACT• PRESS PAD FIRMLY• PEEL OTHER PAD OFF BLUE PLA• APPLY PAD TO BARE SKIN EXACT• PRESS PAD FIRMLY

Note: Actual voice prompts mayconfiguration.

02983_J_PIP-TCP.pdf

Page 12: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c

t approximately every seven seconds if no

in the correct language.

s the device turns off.

lts on the PIP Checklist.

32 Page 12 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Initialization test ontinued)

Note: These prompts will repeafurther action is taken.

8. Verify that the voice prompts are

9. Close the lid.

10. Verify the voice prompts cease a

11. Record the Initialization test resu

02983_J_PIP-TCP.pdf

Page 13: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAs he defibrillator discharges up to 360 J of

py cable. e serial port communication link, verify the :00 auto test, perform defib verification, t the device time and date.er before you execute the test software.

RDA port is aligned with the service communication dongle.

vision G.

r left corner of the opening screen, is

ing screen and follow the instructions as

PIP passed.

32 Page 13 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Computer sisted tests

WARNING!Shock hazard. When discharged, telectrical energy through the theraPerform these steps to test the devicdevice serial number, perform the 03perform the defib charge time test, seNote: Record the device serial numb

1. Position the device such that its Ipersonal computer (PC) infrared

2. Execute software P/N 3202957 re

3. Verify the CRC value, in the lowe0x54A3.

4. Select the PIP button on the openprompted.

5. Verify the software P/N 3202957

PCOK

Device

02983_J_PIP-TCP.pdf

Page 14: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAs(c

ation fails, execute the TCP using software energy calibration. After the TCP passes ftware P/N 3202957 to obtain passing

st results on the PIP Checklist.

32 Page 14 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Computer sisted tests

ontinued)

Note: If the defibrillator energy verificP/N 3202957 to perform defibrillator you must re-execute the PIP from soresults for the PIP.

6. Record the Computer Assisted te

02983_J_PIP-TCP.pdf

Page 15: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIm

een electrode impedances and the

resistance box using the electrode test able.

50

utton to turn the device ON.

TING HEART RHYTHM.

TING HEART RHYTHM.

decade resistance box set to each of the ice prompts at each value.

32 Page 15 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Electrode pedance Test

This test verifies the relationship betwdevice’s responses.

1. Ensure the device is OFF.

2. Connect the device to the decadepost adapter with FAST-PATCH c

3. Set the decade resistance box to

4. Press the LID RELEASE/ON-OFF b

5. The voice prompt sounds:

A- Original Voice Prompt Version:• DO NOT TOUCH PATIENT/EVALUA

B- Current Voice Prompt Version:• DO NOT TOUCH PATIENT/EVALUA

6. Turn the device OFF.

7. Repeat steps 1 through 6 with thefollowing values and verify the vo

OK

FAST-PATCH cable

Decade Resistance

Box

ElectrodeTest post adapter

Device

02983_J_PIP-TCP.pdf

Page 16: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIm(C

OMPT VERSION:D CONTACT

- REMOVE CLOTHING FROM CHEST

OMPT VERSION: D CONTACT

- REMOVE ALL CLOTHING FROM

OMPT VERSION:T - EVALUATING HEART RHYTHM

OMPT VERSION: T - EVALUATING HEART RHYTHM

32 Page 16 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Electrode pedance Test ontinued)

Resistance Value (Ω) Voice Prompt

3 A- ORIGINAL VOICE PRCHECK PADS FOR GOOorCALL FOR HELP NOW

B-CURRENT VOICE PRCHECK PADS FOR GOOCHECK CONNECTORorCALL FOR HELP NOW PATIENT’S CHEST

11 A- ORIGINAL VOICE PRDO NOT TOUCH PATIEN

B-CURRENT VOICE PRDO NOT TOUCH PATIEN

02983_J_PIP-TCP.pdf

Page 17: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIm(C

OMPT VERSION:T - EVALUATING HEART RHYTHM

OMPT VERSION: T - EVALUATING HEART RHYTHM

OMPT VERSION:D CONTACT

- REMOVE CLOTHING FROM CHEST

OMPT VERSION: D CONTACT

- REMOVE ALL CLOTHING FROM

32 Page 17 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Electrode pedance Test ontinued)

Resistance Value (Ω) Voice Prompt

238 A- ORIGINAL VOICE PRDO NOT TOUCH PATIEN

B-CURRENT VOICE PRDO NOT TOUCH PATIEN

370 A- ORIGINAL VOICE PRCHECK PADS FOR GOOorCALL FOR HELP NOW

B-CURRENT VOICE PRCHECK PADS FOR GOOCHECK CONNECTORorCALL FOR HELP NOW PATIENT’S CHEST

02983_J_PIP-TCP.pdf

Page 18: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIIm(C

depending on the device configuration.

le and Electrode test post adapter from the

test results on the PIP Checklist.

OMPT VERSION:D CONTACT

- REMOVE CLOTHING FROM CHEST

OMPT VERSION: D CONTACT

- REMOVE ALL CLOTHING FROM

32 Page 18 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Electrode pedance Test ontinued)

Note: Actual voice prompts may vary

8. Disconnect the FAST-PATCH cabdecade resistance box.

9. Record the electrode impedance

Resistance Value (Ω) Voice Prompt

Open A- ORIGINAL VOICE PRCHECK PADS FOR GOOorCALL FOR HELP NOW

B-CURRENT VOICE PRCHECK PADS FOR GOOCHECK CONNECTORorCALL FOR HELP NOW PATIENT’S CHEST

02983_J_PIP-TCP.pdf

Page 19: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAuOp

isory system:et the Defibrillator Analyzer to output a

utton to turn the device ON.

s turn steady green.

clearly audible. They may include:

UATING HEART RHYTHM

CIRCULATION)

CIRCULATION), START CPRSEC, REPEAT “CONTINUE CARE” EVERY 15 SEC).

32 Page 19 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Defibrillator tomatic Advisory eration

To test the defibrillator automatic adv1. Establish the set-up as shown. S

Normal Sinus Rhythm waveform.

2. Press the LID RELEASE/ON-OFF b

3. Verify that the electrode indicator

4. Verify that the voice prompts are

A- ORIGINAL VOICE PROMPT VERSION:• CALL FOR HELP NOW• DO NOT TOUCH PATIENT — EVAL• NO SHOCK ADVISED• CHECK FOR PULSE (or SIGNS OF

• IF NO PULSE (or IF NO SIGNS OF• CONTINUE CARE (NOTE: AFTER 15

FAST-PATCH cable

Electrodetest post adapter

OK

Device

Defibrillator

Analyzer

02983_J_PIP-TCP.pdf

Page 20: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAuOp UATING HEART RHYTHM

CUE BREATHS AND CHEST COMPRESSIONSSEC, REPEAT “CONTINUE CARE” EVERY 15 SEC). depending on the device configuration.

he device discharges up to 360 J of py cable.utput a Ventricular Fibrillation waveform.

clearly audible. They may include:

32 Page 20 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Defibrillator tomatic Advisory eration (continued)

B- CURRENT VOICE PROMPT VERSION:• CALL FOR HELP NOW• DO NOT TOUCH PATIENT — EVAL• NO SHOCK ADVISED• CHECK FOR BREATHING• IF NOT BREATHING/PROVIDE RES• CONTINUE CARE (NOTE: AFTER 15

Note: Actual voice prompts may vary

WARNING!Shock hazard. When discharged, telectrical energy through the thera5. Set the Defibrillator Analyzer to o

6. Cycle power on the device.

7. Verify that the voice prompts are

02983_J_PIP-TCP.pdf

Page 21: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAuOp

RING SHOCK

depending on the device configuration.

fy that the voice prompts are clearly

32 Page 21 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Defibrillator tomatic Advisory eration (continued)

A- ORIGINAL VOICE PROMPT VERSION:• DO NOT TOUCH PATIENT DELIVE• EVALUATING HEART RHYTHM• STAND BY • PREPARING TO SHOCK

B- CURRENT VOICE PROMPT VERSION:• DO NOT TOUCH PATIENT • EVALUATING HEART RHYTHM• STAND BY • PREPARING TO SHOCK

Note: Actual voice prompts may vary

8. When the device is charged, veriaudible. They may include:

A- ORIGINAL VOICE PROMPT VERSION:• EVERYONE CLEAR• PRESS FLASHING BUTTON

B- CURRENT VOICE PROMPT VERSION:• EVERYONE CLEAR

02983_J_PIP-TCP.pdf

Page 22: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PIAuOp vary depending on the device

.

rge automatically.charge automatically; the SHOCK button

ON-OFF button to turn the device OFF.from the device and the Defibrillator

n of the automatic advisory operation test

32 Page 22 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Defibrillator tomatic Advisory eration (continued)

• DO NOT TOUCH PATIENT• DELIVERING SHOCK

Note: Actual voice prompts mayconfiguration.

9. Verify that a ready tone is audible

WARNING!Fully automatic models will dischaNote: Fully automatic models will diswill flash for semi-automatic models.

10. Press and hold the LID RELEASE/

11. Disconnect the Fast Patch cable Analyzer

12. Record the successful completioon the PIP Checklist.

02983_J_PIP-TCP.pdf

Page 23: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI meets the following industry standards:l Commission) 60601-1 and 60601-2-4

perform these tests could result in a current. Make sure you are familiar with t performance procedures.

pecific instructions for use of the Fluke Use of equivalent test equipment is

32 Page 23 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Leakage Current Check to ensure the leakage current• IEC (International ElectrotechnicaWARNING!Shock Hazard. Failure to properly failure to detect excessive leakageyour test equipment and these tes

Note: This test procedure provides sBiomedical ESA612 Safety Analyzer.acceptable.

02983_J_PIP-TCP.pdf

Page 24: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c

ermination cable to connect the device to e RA and LL snaps of the ESA612.

s the A button then the F3/Patient

measure IEC 60601 under the Setup

the ESA612 is properly configured to e.

akage Current from the RA snap to the LL e current limit, shown below.

Mode Leakage Current Limit

9 µA

32 Page 24 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Leakage Current ontinued)

1. Use the QUIK-COMBO to snap tthe ESA612 Safety Analyzer at th

2. Turn the device ON.

3. Turn the Fluke ESA612 ON, presAuxiliary button.

Note: The ESA612 must be setup tomenu.

4. Verify the test setup and confirm measure Patient Auxiliary leakag

5. Verify that the Patient Auxiliary Lesnap does not exceed the leakag

IEC 60601Parameter Current

Patient Auxiliary(QUIK-COMBO Leads, Lead-to-Lead)

AC + DC

02983_J_PIP-TCP.pdf

Page 25: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI(c

ex to the Sternum is identical to the flow ne leakage measurement is required.

, Polarity, Neutral, and Earth settings are battery power only.

esults on the PIP Checklist.

32 Page 25 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Leakage Current ontinued)

Note: As the current flow from the Apfrom the Sternum to the Apex, only o

Note: When using the Fluke ESA612irrelevant as the device operates via

6. Record the leakage current test r

7. Turn the device OFF.

02983_J_PIP-TCP.pdf

Page 26: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

ReQUanBa

HARGE-PAK battery charger have the be replaced at the end of this inspection to

QUInCHins

e the QUIK-PAK electrodes:hor pin from the slot in the device.

nchor pin into the slot on the device.

tor.

cket is centered on the device and is tucked lid.

ttery charger installation procedure on the

32 Page 26 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

plenishment of IK-PAK Electrodes

d CHARGE-PAK ttery Charger

The QUIK-PAK electrodes and the Csame Use By dates. They must bothensure the device is ready for use.

IK-PAK Electrodes stallation and

ARGE-PAK tallation

Perform the following steps to replac1. Remove the electrode packet anc

2. Slide the replacement electrode a

3. Plug in the new electrode connec

4. Ensure that the new electrode pabehind the lip before closing the

5. Proceed to the CHARGE-PAK banext page.

02983_J_PIP-TCP.pdf

Page 27: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

QUInCHins

y may be partially depleted (the ATTENTION inspection. This is considered normal placed back into active service until the OK roximately 10 days, depending on the state

e and replace the CHARGE-PAK battery

battery charger by pressing down on its charger will spring outward from the

tery charger into the device until you hear it pletely inserted, it will spring back out.

s on the Readiness Display.

battery charger using the discharger nd following the instructions provided.

QUIK-PAK electrodes and CHARGE-PAK PIP Checklist.

32 Page 27 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

IK-PAK Electrodes stallation and

ARGE-PAK tallation (continued)

Note: The defibrillator internal battersymbol will appear) at the end of the operation but the device must not be symbol is visible. (This may take appof the battery.)

Perform the following steps to removcharger used during this inspection.1. Remove the used CHARGE-PAK

latch. The CHARGE-PAK batterydevice.

2. Insert the new CHARGE-PAK bat“click” into position. If it is not com

3. Verify that the OK symbol appear

4. Deplete the used CHARGE-PAK provided in the replacement kit a

5. Record the installation of the newbattery charger test results on the

02983_J_PIP-TCP.pdf

Page 28: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF rformance Inspection Procedure

P

PI f the device after replacing the UIK-PAK electrodes.nnected to the device.

he self-test to finish.

the Readiness Display. r and the ATTENTION symbol is visible, the charging the internal battery and the n active service. he device must be serviced.

ples of symbols)

lt on the PIP Checklist.

!

32 Page 28 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator Pe

IP – Instructions

P – Final Status This test confirms the overall status oCHARGE-PAK battery charger and QNote: Ensure a patient load is not co

• Turn the device ON and wait for t• Turn the device OFF.• Verify the OK symbol appears in • If the OK symbol does not appea

CHARGE-PAK battery charger isdevice must not be placed back i

• If the WRENCH symbol appears, t

(Exam

• Record the Final Status test resu

OK

02983_J_PIP-TCP.pdf

Page 29: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents

Previous Page Table of Contents Back Index Next Page

Test and Calibration Procedures

This section covers the Test and Calibration Procedures (TCP). You should complete the relevant TCP to correct out-of-specification conditions detected during the Performance Inspection Procedure (PIP). Also follow TCP as necessary after replacement of device components. The listed procedures may be performed in any order.Note: The two PCBs in this device must be changed as a calibrated set. The analog and digital PCBs cannot be replaced individually.

Note: Any time the device is opened for repair, component replacement, or calibration, the device must successfully pass all portions of the closed-case PIP before being returned to active service. If a device exhibits a fault code, contact Physio-Control; do not place it back in service.

Scope and Applicability

Resource Requirements

Test Equipment Requirements

Defibrillator Energy Calibration

3202983_J_PIP-TCP.pdf Page 29 of 34

Page 30: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF Test and Calibration Procedures

S

ively. You may perform the procedures

for repair or component replacement, or cessfully pass the entire Performance turned to active service.r necessary equipment, test equipment

ents for a listing of test equipment, complete the TCP.

32 Page 30 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator

cope and Applicability

The TCP applies to the device exclusoutlined in this section in any order.Note: Any time the device is openedafter calibration, the device must sucInspection Procedure before it is reRefer to Resource Requirements foverification, and workstation power. Refer to Test Equipment Requirem(including specifications) required to

02983_J_PIP-TCP.pdf

Page 31: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF Test and Calibration Procedures

R

nts for TCP equipment, TCP test ion power, and TCP personnel.

TC e equipment listed in the Test Equipment ble lists specific test equipment by uivalent specifications may be substituted.

TCVe

e TCP must have a current calibration sued by a certified calibration facility.

TCPo

must be connected to a grounded power lectrostatic Discharge (ESD) protection.

32 Page 31 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator

esource Requirements

This section describes the requiremeequipment verification, TCP workstat

P – Equipment To perform the TCP, you must use thRequirements table. Although the tamanufacturer, test equipment with eq

P – Test Equipment rification

All test equipment used to perform thlabel. The calibration label must be is

P – Workstation wer

The ac line power to the workstation source. The workstation must have E

02983_J_PIP-TCP.pdf

Page 32: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF Test and Calibration Procedures

TeEq ManufacturerD

t

Fluke -Impulse 7000DP

C 3011030-00 El 3205979 TePI

3202957 revision G

32 Page 32 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator

st Equipment Requirementsuipment Specifications

efibrillator analyzer Power range: 0–450 JLoad resistance: 50 Accuracy and/or guard bandingmust be sufficient to ensure teslimits.

able Assembly, FAST-PATCH ectrode Test Post Adapter (2 ea)st Software, LIFEPAK CR Series P-TCP

02983_J_PIP-TCP.pdf

Page 33: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

P Back Index Next Page

LIF Test and Calibration Procedures

D Page 1 of 3

the energy meter. Dangerous voltages ectrode plates/posts. te all defibrillator-delivered energy levels:RDA port is aligned with the PC infrared

vision G. Select TCP button from the tructions as prompted.

fibrillator energy calibration.

rger must be replaced after defibrillator

32 Page 33 of 34

revious Page Table of Contents Section Contents

EPAK CR PLUS/EXPRESS Defibrillator

efibrillator Energy Calibration

WARNING!Shock hazard. Avoid contact with will be present on energy meter elPerform the following steps to calibra1. Position the device such that its I

communication dongle.

2. Execute program P/N 3202957 reopening screen and follow the ins

Note: Use a 50 patient load for de

Note: The CHARGE-PAK battery chaenergy calibration.

PC

Device

OK

IRDALink

02983_J_PIP-TCP.pdf

Page 34: LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents€¦ · LIFEPAK CR PLUS/EXPRESS Defibrillator Section Contents Previous Page Table of Contents Back Index Next Page Performance

LIFEPAK® CR PLUS/EXPRESS Defibrillator

Performance Inspection Procedure (PIP) Checklist

Model # __________________________________ Department/Location _______________________

Serial # __________________________________ Performed By ______________________________ Date _____________________________________ Service Report #____________________________

Pg 1 of 1

Inspection Pass Failed CommentsA Physical Inspection

Confirm the results of physical inspection. Testing1 Initalization Test

Confirm the results of Initalization test. 2 Computer Assisted Tests

a- Record the results of the Defibrillation Delivered Energy tests.

Shock # 1 Observed (J) ___________Shock # 2 Observed (J) ___________Shock # 3 Observed (J) ___________

b- Confirm the result of Computer assisted tests. 3 Electrode Impedance

Confirm the results of Electrode impedance test.

50 Do not touch patient3Check pads/Call for help/Check Connector11Do not touch patient238Do not touch patient370Check pads/Call for help/Check ConnectorOpen Check pads/Call for help/Check Connector

4 Automatic Advisory Operationa Record the results of Automatic Advisory Operation tests. 5 Leakage Current

Record the result of Leakage current test. (0 µA - 9 µA) Observed (µA) ___________6 Installation of New QUIK-PAK Electrodes and CHARGE-PAK battery charger

Confirm the result of Installation of new QUIK-PAK electrodes and CHARGE-PAK battery charger.

7 Final StatusConfirm the result of Final status test.

AAA

3202983_J_PIP-TCP.pdf Page 34 of 34