LCMSMS Solutions For Steroid Analysis - SCIEX notes/LCMSMS_Solutions...LC-MS/MS, including: ‒...
Transcript of LCMSMS Solutions For Steroid Analysis - SCIEX notes/LCMSMS_Solutions...LC-MS/MS, including: ‒...
RUO-MKT-18-1166-A
LCMSMS Solutions For Steroid Analysis
Dan Blake
Manager, Clinical Applications
For Research Use Only. Not for use in diagnostic procedures.
2 © 2016 SCIEX RUO-MKT-18-1166 A
Agenda
• Special considerations for steroid analysis
• Methods of Steroid analysis 1: Simple & Fast ‒ Testosterone iMethod™
‒ Online Extraction with 2DLC and simple sample processing
‒ Low Level Estrogen iMethod™
• Improving Selectivity with Ion Mobility
• Methods of Steroid Analysis 2: Extending the coverage ‒ Steroid Panel iMethod™
‒ Commercial Kits
• Summary of and conclusions from the different approaches
Special considerations for steroid analysis
4 © 2016 SCIEX RUO-MKT-18-1166 A
• Traditional immunoassays, while convenient, are unreliable for
reasons previously discussed
• Varying sample volumes available ‒ Paediatric/neonatal samples
• Pressures of sample number and turnaround time demand a
fast and easy to setup method
• Again, similar pressures require the use of rapid, automatable
extraction procedures ‒ Ideally some form of protein precipitation
‒ Liquid Liquid extraction is an option but not appropriate for all laboratories
‒ SPE is an option but can be costly
‒ Derivitisation, for example with dansyl chloride, is an option for challenging
steroids but should be considered as a last resort.
5 © 2016 SCIEX RUO-MKT-18-1166 A
• What Steroids to analyse?
The LC/MS/MS method should give us as much information as possible
However, increasing the number of analytes could potentially increase the complexity of the assay
6 © 2016 SCIEX RUO-MKT-18-1166 A
The “ideal” LC/MS/MS Steroid Application
• The convenience of an immunoassay style kit approach
• Sensitive enough to assay the lowest level steroids ‒ The “which instrument” part of the question
• Able to cope with sample volume and throughput demands
• The highest control ‒ Wide dynamic range standard curve
‒ Deuterated internal standards, for each analyte if possible
‒ Ability to utilise traceable Quality Control material
Testosterone iMethod™
8 © 2016 SCIEX RUO-MKT-18-1166 A
• Testosterone is one of the most widely measured steroids
• Measurement of testosterone is used to research a variety of
disorders, such as congenital adrenal hyperplasia and
hypogonadism.
• Measurement of the typical levels in men is straightforward;
however, measurement of the low levels present in women and
children can be especially challenging.
• SCIEX provides solutions for routine analysis for both
circumstances. ‒ Simple male testosterone assay on the API3200™ platform using protein
precipitation
‒ Low level (female) assay on the API5000™ or Triple Quad 5500™ using
more thorough sample preparation
9 © 2016 SCIEX RUO-MKT-18-1166 A
Testosterone iMethod™, API3200™
Data from a 3.4 nMol/L testosterone sample analyzed using an SCIEX API
3200™ LC/MS/MS System.
10 © 2016 SCIEX RUO-MKT-18-1166 A
Testosterone iMethod™, API5000™
Data from a 0.069 nMol/L testosterone sample analyzed using an SCIEX
API 5000™ LC/MS/MS System.
Online Extraction with 2DLC and simple sample
processing
Testosterone iMethod™
12 © 2016 SCIEX RUO-MKT-18-1166 A
• A simple method for analysis of a small number of routinely
measured steroids has been proposed
• The method employs a simple protein precipitation with direct
injection onto a 2DLC system (no need for offline sample
concentration)
• The remainder of the sample cleanup process is performed on a
trapping column, followed by a reversed flow gradient onto a
common C18 HPLC column, with a total runtime of 5 minutes
• As a consequence, sample preparation time is kept to a
minimum and sample throughput is maximised
• The trap column is reusable, with a lifetime in excess of 1000
injections
13 © 2016 SCIEX RUO-MKT-18-1166 A
Switching Valve
TRAPPING COLUMN Analytical Column
Valve position A:
(Analyte Loading onto Trapping Column)
10
6
9
8
7 5
1 2
3
4
GRADIENT PUMPS A/B
AUTOSAMPLER
VALVE ISOCRATIC LOAD PUMP C
WASTE WASTE
6 5
Valve position B:
(Analyte Elution from SPE Column onto
Analytical Column)
MS System
Proposed Flow and Plumbing diagram for
2DLC Steroids method
Column Oven
Switching Valve:
Time(min) Valve
Position
0.0 A
1.0 B
5.5 A
14 © 2016 SCIEX RUO-MKT-18-1166 A
Extracted Steroid sample, approx 0.15nmol/L, Analysed by 2DLC approach
Testosterone
17-OH-P
11-Deoxycortisol
Androstenedione
Low Level Estrogen iMethod™
16 © 2016 SCIEX RUO-MKT-18-1166 A
The Importance of Estrogens
• Estrogen is an entire class of related female sex hormones: estrone (E1) estradiol (E2), and estriol (E3)
• Estrone (E1) is widespread throughout the body. It is the only one of estrogens that’s present in any amount in women after menopause.
• Estradiol (E2) is the primary sex hormone of childbearing women. It is formed from developing ovarian follicles. Estradiol is responsible for female characteristics and sexual functioning. Also, estradiol is important to women’s bone health. Estradiol contributes to most gynecologic problems and even female cancers.
• Estriol (E3) is made from the placenta. It’s important only during pregnancy.
17 © 2016 SCIEX RUO-MKT-18-1166 A
Estrone and Estradiol:
Conditions:
• Instrumentation ‒ API 5000™ and SCIEX Triple Quad™ 5500 Mass Spectrometers
‒ Shimadzu Prominence 20A and Agilent 1200 integrated systems
• Method Information ‒ 500 µL sample volume required
‒ Liquid-liquid extraction without derivatization
‒ Clinically relevant dynamic range of 5-500 pg/mL
‒ 6 minute total run time
18 © 2016 SCIEX RUO-MKT-18-1166 A
Estrone and Estradiol: Lower Limit of Quantitation
Estrone 5 pg/mL LLOQ, S/N=22 Estradiol 5 pg/mL LLOQ, S/N=19
19 © 2016 SCIEX RUO-MKT-18-1166 A
Estrone and Estradiol: Precision and Accuracy
Sample (pg/mL) %CV % Accuracy
Day 1-3 Day 1-3
LLOQ (5) 9 104
Low QC (15) 3 104
High QC (100) 3 101
ULOQ (500) 3 100
•Estrone (inter-day)
•LLOQ CV’s ≤ 9% ,QC’s CV’s ≤ 5%
•Accuracy deviation ≤ 4%
Sample (pg/mL) %CV % Accuracy
Day 1-3 Day 1-3
LLOQ (5) 10 101
Low QC (15) 4 96
High QC (100) 5 102
ULOQ (500) 4 101
Estrone (E1) Estradiol (E2)
•Estradiol (inter-day)
•LLOQ CV’s ≤ 10% ,QC’s CV’s ≤ 5%
•Accuracy deviation ≤ 4%
20 © 2016 SCIEX RUO-MKT-18-1166 A
Estrone and Estradiol: Linearity
Clinically relevant dynamic range 5-500 pg/ml, R≤ 0.999
Estrone
Estradiol
21 © 2016 SCIEX RUO-MKT-18-1166 A
Estrone and Estradiol: Summary
• The method was verified internally across laboratories for
analytical analysis
• LLOQ with suitable signal to noise (20)
• Inter-day Precision and Accuracy ‒ LLOQ CVs ≤ 10% ,QCs CV’s ≤ 5%
‒ Accuracy Deviation ±4%
• Control samples at clinically relevant low and high
concentrations (5-500 pg/mL) with linear response r ≤ 0.999
• Liquid-Liquid extraction without derivatization
• 6 minutes run time
Improving Selectivity with Ion Mobility
23 © 2016 SCIEX RUO-MKT-18-1166 A
Ion Mobility to improve selectivity
• There are many different ways to achieve improved selectivity in applications of
LC-MS/MS, including:
‒ Improved chromatography
‒ Tandem mass spectrometry (MS/MS, or even higher MS3)
‒ High-resolution mass spectrometry
‒ More up-front sample cleanup
• In many applications of LC-MS/MS, mass measurements alone (even high
resolution) cannot provide sufficient selectivity
‒ Isobaric interferences may have the same “exact mass” as analyte ions
‒ Interferences may have common fragment ions
‒ Chemical background may pose a challenge
• The SelexION™ ion mobility technology provides an orthogonal means of
separating / filtering ions prior to, and independent of, MS/MS analysis.
24 © 2016 SCIEX RUO-MKT-18-1166 A
SelexION™ Ion Mobility Device Components:
New hardware
Existing ion path
25 © 2016 SCIEX RUO-MKT-18-1166 A
SelexION™ Ion Mobility Device Components:
Robust, easy-to-install, hardware components:
No tools required
No cables
No need to break vacuum
Installation in about 2 minutes
1. Orifice Plate 2. Ion Mobility Cell 3. Curtain Plate
26 © 2016 SCIEX RUO-MKT-18-1166 A
0
0
=0
t1
t2
)0(
)0()()(
K
KEKE
Tuning the SelexION™ Ion Mobility Device
Ions migrate towards one of the planar electrodes
– If an ion’s high-field mobility K(E) is larger than its low-field mobility K(0), >0
– If an ion’s low-field mobility K(0) is larger than it high-field mobility K(E), <0
Any ion can be steered back onto the center-line, by application of a compound-specific DC compensation voltage (CV).
Compensation
Voltage (COV)
Separation
Voltage (SV)Compensation
Voltage (COV)
Separation
Voltage (SV)
ions in
transport gas to mass
spectrometer
27 © 2016 SCIEX RUO-MKT-18-1166 A
Tuning the SelexION™ Ion Mobility Device
In infusion mode, the compensation voltage (CV) parameter is ramped in order to determine the optimized value.
Note that this is analogous to tuning any other compound-specific parameter, e.g. the collision energy (CE)
-14 -12 -10 -8 -6 -4 -2 0 2 4 6 8 10 12 14 16 180%
50%
100%
Rel. I
nt.
(%
)
Compensation Voltage (CV)
-14 -12 -10 -8 -6 -4 -2 0 2 4 6 8 10 12 14 16 180%
50%
100%
Rel. I
nt.
(%
)
Compensation Voltage (CV)
The ion residence time, and
hence the resolution of the ion
mobility device, can be controlled.
CV peak widths range from
approximately 0.5 – 2.5 V FWHM.
28 © 2016 SCIEX RUO-MKT-18-1166 A
XIC of +MRM (6 pairs): Exp 1, 287.2/97.0 Da ID: ASD from Sample 15 (Androgens mix 5) of Quant-RestekC18-150msec.wiff (Turbo Spr... Max. 8.7e4 cps.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5Time, min
0.00
5000.00
1.00e4
1.50e4
2.00e4
2.50e4
3.00e4
3.50e4
4.00e4
4.50e4
5.00e4
5.50e4
6.00e4
6.50e4
7.00e4
7.50e4
8.00e4
8.50e4
9.00e4
9.50e4
1.00e5
1.05e5
1.10e5
1.15e5
1.20e5
1.25e5
1.29e5
Inte
ns
ity
, c
ps
3.17
ASD
T DHT
AND
DHEA 5-ASD
75% B 80% B
98% B
55% B
(~400pg/mL)
(~500pg/mL)
(~30ng/mL) (~10ng/mL)
(~4ng/mL)
(~10ng/mL)
Mixture of 6 androgens in 60:40 Methanol:Water
Sample chromatogram from QTRAP® 5500 LC/MS/MS system
29 © 2016 SCIEX RUO-MKT-18-1166 A
Structures of two androgens, DHEA and T
• Testosterone, C19H28O2
O
CH3
CH3OH
H
HH
O
CH3
CH3
OH
H
HH
Dehydroepiandrosterone (DHEA), C19H28O2
Low ionization efficiencies
Common precursor ion masses (m/z 289)
Similar fragmentation patterns
Similar chromatographic retention properties
30 © 2016 SCIEX RUO-MKT-18-1166 A
MS/MS of DHEA and Testosterone (m/z 289)
(A) MS/MS fragmentation pattern (Collision Energy ramp) for DHEA
MS/MS fragmentation pattern (Collision Energy ramp) for Testosterone (B)
O
CH3
CH3OH
H
HH
O
CH3
CH3
OH
H
HH
Used for MRM
transition
Used for MRM
transition
31 © 2016 SCIEX RUO-MKT-18-1166 A
Resolving DHEA and Testosterone using the SelexION™ Ion Mobility Device
a) There is a significant interference in the MRM channel for DHEA, caused by
the presence of testosterone
DHEA (289.0/271.1)
DHEA (289.0/253)
Interference caused by the
presence of Testosterone (a) No interference caused by the
presence of Testosterone
DHEA (289.0/271.1)
DHEA (289.0/253)
(b)
b) When the SelexION™ ion mobility device is employed the interference is completely removed. Thus there is no longer a need for chromatographic separation, and run-times can be reduced dramatically.
SelexION™ device “off” SelexION™ device “on”
32 © 2016 SCIEX RUO-MKT-18-1166 A
Liquid-Liquid Extraction (LLE) – LC-MS/MS
289/109 289/97
Testosterone
peak in
female serum
Testosterone
peak in
female serum Interference
Interference
Area ratio ~ 23
Expected ratio
=1.23
x
Liquid-Liquid Extraction (LLE) – LC-MS/MS with SelexION™ Technology
289/109 289/97
Area ratio =
1.28
Expected ratio
=1.23
33 © 2016 SCIEX RUO-MKT-18-1166 A
Protein Precipitation (PPT) – LC-MS/MS
289/109 289/97 Testosterone
peak in
female serum
Testosterone
peak in
female serum
Protein Precipitation (PPT) – LC-MS/MS with SelexION™ Technology
289/109 289/97
34 © 2016 SCIEX RUO-MKT-18-1166 A
STD Curve in Serum Sample using the SelexION™Ion Mobility Device (sample preparation = PPT)
4 female
subjects
Sample Neat (n=28) Patient PPT (n=25) Patient LLE (n=25)
Average Ion Ratio
(97/109) 1.23 1.28 1.24
STDEV 0.04 0.07 0.04
%CV 3.44 5.47 3.12
Accuracy 96- 105%
Sample Neat (N=28) Serum PPT (N=25) Serum LLE (N=25)
4 serum samples
from females
35 © 2016 SCIEX RUO-MKT-18-1166 A
SelexIon performance for Estrogen Analysis Serum (40 uL) by PPT and no SelexIon™
XIC of -MRM (3 pairs): 275.200/147.100 Da ID: d4-E2 from Sample 1 (PlasmaNew ppt 15) of 111220-NoDms-PlasNew-15.wiff (Turbo Spray) Max. 6.8e4 cps.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5Time, min
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
In
te
ns
it
y,
c
ps
XIC of -MRM (3 pairs): 271.200/145.100 Da ID: E2 from Sa... offset by -2.05 min
Max. 3710.6 cps.
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0Time, min
0
200
400
600
800
1000
In
te
ns
ity
, c
ps
8.72
7.827.98
XIC of -MRM (3 pairs): 269.200/145.100 Da ID: E1 from Sa... offset by -2.05 min
Max. 1.4e4 cps.
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0Time, min
1000
2000
3000
4000
5000
6000
7000
8000
9000
In
te
ns
ity
, c
ps 8.68
7.76
E2
E1
36 © 2016 SCIEX RUO-MKT-18-1166 A
XIC of -MRM (3 pairs): 269.200/145.100 Da ID: E1 from Sample 6 (plasma 15ppt) of 111222-Plasma15-demifast.wiff (Turbo Spray) Max. 2313.0 cps.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5Time, min
0
500
1000
1500
2000
2313
In
te
ns
it
y,
c
ps
8.66
XIC of -MRM (3 pairs): 271.200/145.100 Da ID: E2 from Sa... Max. 289.3 cps.
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0Time, min
0
50
100
150
200
250
289
In
te
ns
ity
, c
ps
8.64
XIC of -MRM (3 pairs): 269.200/145.100 Da ID: E1 from Sa... Max. 2313.0 cps.
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0Time, min
0
500
1000
1500
2000
2313In
te
ns
ity
, c
ps
8.66E2
E1
SelexIon performance for Estrogen Analysis Serum (40 uL) by PPT with SelexIon™
37 © 2016 SCIEX RUO-MKT-18-1166 A
Spiked plasma sample analyzed without SelexION™ (A) and with SelexION™ Technology (B). The light blue marked regions defined where signal and noise were collected.
E1: S/N (1 sigma): 77
E2: S/N (1 sigma): 27
E1: S/N (1 sigma): 462
E2: S/N (1 sigma): 162
Steroid Panel iMethod™
39 © 2016 SCIEX RUO-MKT-18-1166 A
• Increasing the number of compounds analysed in an assay is a
perfect way of maximising potential return on investment
• LC-MS/MS is ideally suited to this approach
• The SCIEX Steroid Panel iMethod™ allows simultaneous
analysis of 10 commonly measured steroids and also 25-
Hydroxyvitamin D3
• Sample preparation is simple protein precipitation with dilution
• LC separation is required in order to keep the sample
preparation to a minimum, run time is 12.5 minutes
• In order to achieve the required sensitivity for all compounds
assayed, this method is designed for use on the API5000™and
Triple Quad 5500™instruments
40 © 2016 SCIEX RUO-MKT-18-1166 A
Steroid Panel iMethod™, Extracted sample at ULOQ
41 © 2016 SCIEX RUO-MKT-18-1166 A
Performance Data, Steroid Panel iMethod™
42 © 2016 SCIEX RUO-MKT-18-1166 A
• SCIEX offers several solutions for steroid analysis
• Various forces will determine which approach is needed ‒ Analysis of less challenging steroids
‒ Requirements for analysis of multiple compounds
‒ Sensitivity requirements on challenging steroids
‒ Desire for the convenience of a kit-based assay
‒ Multiplexing with additional applications
• From this, the appropriate method and instrument can be
matched to the analytical requirements
43 © 2016 SCIEX RUO-MKT-18-1166 A
Trademarks/Licensing
SCIEX is doing business as SCIEX.
For Research Use Only. Not for use in diagnostic procedures.
© 2016 SCIEX. The trademarks mentioned herein are
the property of SCIEX Pte. Ltd. or their respective owners.
SCIEX™ is being used under license.