LCG Meeting, May 14th 20 03 V. Daniel Elvira 1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab

Transcript of LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel...

Page 1: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 1

G4 (OSCAR_1_4_0) Validation of CMS HCal

V. Daniel Elvira

Fermilab

Page 2: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 2

General Description

• Simulate HCal test beam 2002 setup using OSCAR_1_4_0, physics list 1.8

•Includes beam line (trigger scint. tiles, wire chambers), ECal box, HB, HO, aluminium slab (representing magnet), iron (muon) layer

• Mimic the data analysis, including calibration from electrons (ECal) and pions (HB): study resolution, response (linearity), transverse shower profiles

•Compare with data measurements for G4 validation

Page 3: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 3

ECal 7x7 Crystal Matrix (movable)

2 HCal HB Modules

HO Panels

Aluminum Slab

HCal 2002 Test Beam

beam

Moving table

Small scale experiment to demonstrate that HCal works: 49 ECal crystals, 144 HB channels, 16 HO channels.

Over 100 Million

Events!

-: 225 GeV

e-: 20,30,50,100 GeV

-: 20,30,50,100,

300 GeV

Page 4: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 4

Based on 1-5,000 events onto the ()=(6,2) tower of the HB and crystal 25 of the ECal matrix

TB02 Simulation (OSCAR/G4 )

Crystal 25

Page 5: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 5

Angle view of the full TB02 detector

10 GeV electron

HB

ECalHO

Beam Line

HO

HB

ECal

Beam Line

TB02 Simulation (OSCAR/G4 )

100 GeV pion

Page 6: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 6

Ecorr= 122*EHB

5x5+1.05*EECal3x3

EHBscint EHB

scint+ 0.1* EHBscint MeV * Rand

EHBtower EHB

tower+ 262 MeV * RandLong. Non-uniformity

Electronic Noise

Calibration factors (already there in previous presentation)

EECaltower EECal

tower+ match* Rand (To match the measured electron resolution-much worse than M.C.)

•Add more energy points at 10, 25, 200, 250, 300 GeV (in addition to 20, 30, 50, 100, 150 GeV)• 5 times more statistics 10-30 GeV

Improved M.C. Simulation

Page 7: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 7

/E=30.9% (similar to the fit to ¾ of second peak by Jordan)

Data analysis: sources systematic errors:• muon (pion decay) & electrons (from scrapping?) backgrounds• calibration

Data Analysis

Page 8: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 8

From J.Damgov

Electron background

Muon background Background

subtracted muon peak

It should be possible to estimate background subtraction error by defining a band limited by an “aggressive” (bckgnd rejection) and a “conservative cut” (signal efficiency)

The resolution comes down from 31% to 26.7% !

Pion Signal

TB02: E/E and e/

Page 9: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 9

Pion peak

Muon background

Background subtracted pion peak

E/E = 20.3%

TB02: E/E and e/

50 GeV pions

Pions showering in the crystals

Data energy resolution for 50 GeV pions (almost negligible background close to the pion peak)

Page 10: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 10

TB02: E/E and e/

The agreement is excellent in all the energy range

Shape difference: e/h (e.m. & nuclear x-sec), leakage……?

Data systematic error analysis in progress

Validate GEANT4 physics models

Resolution Linearity

Page 11: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 11

Transverse Shower Profile

Future: remove ECal, add integrated profiles, longitud. profile studies, HO fraction & compare with OSCAR CMS & CMSIM (Salavat), and TB data (Jordan)

G4/OSCAR TB02

crystals

HB towers

Fill ( D, E )

With D: distance of crystal/tower to the pion direction E: energy deposition

Page 12: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 12

•TB02 OSCAR_1_4_0 fully functional. Validation studies in progress, for week data uncertainties are essential.

•Need to port the simulation to OSCAR_2, hopefully for next CMS week in June

fighting technical problems related to translation to XML, among other things

•Document on TB02 M.C. simulation by July

Summary & Plans

Page 13: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 13

Data & MC Electron Resolution

2.6% at 100 GeV for 3x3 tower matrix

2.8% at 100 GeV for 3x3 tower matrix

EECaltower EECal

tower+ 115 MeV * Rand

Electronic noise

M.C resol much lower than measurement

Page 14: LCG Meeting, May 14th 2003 V. Daniel Elvira1 G4 (OSCAR_1_4_0) Validation of CMS HCal V. Daniel Elvira Fermilab.

LCG Meeting, May 14th 2003

V. Daniel Elvira 14

• EECaltower EECal

tower+ 115 MeV * Rand (noise)• 2.3%/ sqrt(Eele) (photo-statistics)• 0.3% (longitudinal non-uniformity)• 0.4% (calibration)

Other Contributions

No significant change

EECaltower EECal

tower+ match* Rand (To match the measured electron resolution)

•Temperature Resp=2%/oC

(crystals) 0.5%/oC (tubes)• Longitudinal crystal uniformity• Radiation damage

Causes (?):