![Page 1: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/1.jpg)
Live cell imaging and confocal microscopy
Laurent GelmanFacility for Advanced Imaging and Microscopy
Friedrich Miescher Institut, Basel
![Page 2: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/2.jpg)
Acknowledgments
Preparation of the lecture
- Arne Seitz (EPFL)
- Stefan Terjung (EMBL Heidelberg)
- Jens Rietdorf (FMI / Zeiss, Munich)
Data
- Rico Kunzmann (group A. Peters, FMI)
- Vincent Dion (group S. Gasser, FMI)
- Julia Kleylein-Sohn (Novartis)
- Flavio Donato (group P. Caroni, FMI)
- Gerald Moncayo (group B. Hemmigs)
- Jérôme Feige (UNIL / Novartis)
![Page 3: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/3.jpg)
Overview
Growing cells under a microscope…– Physical integrity– Attachment, immobilization of the sample– Incubation systems (temperature, gases, humidity, pH control)
Imaging living cells– Autofluorescence– Photodamage– Microscope settings
(speed, laser power, pinhole, voxel size…)– Microscope setups: spinning-disk versus single-beam LSM and wide-field microscope.
When performing live cell microscopy, cell viability must be at the forefront of any measurement to ensure that the biological processes that are under investigation are not altered in any way.
![Page 4: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/4.jpg)
Mounting: Physical Integrity
![Page 5: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/5.jpg)
Attachment, ImmobilizationCells usually adhere better on plastic (treated) but can be also plated on glass.
Coating the coverglass / SubstratePoly-L-LysineFibronectinConcanavalin AOther…
0.5 - 1% LMP agaroseCyGel1.5 - 3% methyl cellulose Silicon oil, Halocarbon oilTransparent films
Some dishes are made of a special plastic with low autofluorescence where cells adhere well and allowing the use of immersion objectives. (www.ibidi.de) (www.willcowells.com)
![Page 6: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/6.jpg)
Temperature
Small heating stages + Objective heater
![Page 7: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/7.jpg)
Temperature control (on microscope stage)Temperature controller for slides Multiwell stage-top incubator
(OKOlab, www.oko-lab.com)
+ Relatively cheap - System temperature not constant (focus drift
possible!)
(www.embl-em.de)
![Page 8: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/8.jpg)
Focus Drift
Constant temperature of the microscope is also important for time-lapse imaging:
Focus drift due to materialextension/shrinkage due to temperature changes
![Page 9: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/9.jpg)
Microscope incubators
Advantages• Constant conditions (10 – 40 °C)• Stabilized focus• Atmosphere controllable (CO2 …)
Disadvantages
• Expensive
• Microscope access impaired
(www.lis.ch)
(www.oko-lab.com)
(www.embl-em.de)
![Page 10: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/10.jpg)
Gases, pH, Humidity, Osmolarity
Physiological pH during imaging is required!
Without CO2 control (short-term):
• HEPES buffered medium instead of carbonate e.g. 30mM HEPES, 0.5g/l Carbonate instead of 2.2g/l Carbonate.
• Seal the sample chamber, e.g. with silicon (no gas exchange)
With CO2 control (long-term):
• Use perfusion chambers• Use incubators
![Page 11: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/11.jpg)
Gases, pH, Humidity, Osmolarity
Possible solutions:• Seal the sample chamber, e.g. silicon (gas exchange ?)• Use perfusion chambers• Use incubators with humidity control• Use silicon oil
Humidity control during long-term experiments: • Osmolarity of the medium stays constant • sample doesn´t run dry
![Page 12: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/12.jpg)
HEPES Buffered Media
Advantages
• Open system, easy to manipulate.
• Easy to handle and control.
Disadvantages• Usable for ~ 1 hour.• Toxicity of HEPES with some cells.• Evaporation.
![Page 13: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/13.jpg)
Sealed chamber
Advantages
• Easy to handle and control.
• No evaporation.
• Cheap (reusable).
Disadvantages• Usable for max 3 hours depending
on volume.• No manipulation.• Mounting of the chamber might be
tedious.
![Page 14: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/14.jpg)
Gas Permeable Chamber
Advantages• High optical quality (DIC)• Gas permeable• Water impermeable• Perfusable• Usable for days.
Disadvantages• No direct access to cells
(www.ibidi.de)
![Page 15: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/15.jpg)
Perfusion chambers
Advantages
• Constant conditions.
• Manipulation of media.
• Usable for days.
• No evaporation problems.
Disadvantages• Hard to assemble and control.• Shear stress possible.
![Page 16: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/16.jpg)
“scratch” or “would healing assay”Mammalian cells. Wide-field automated inverted microscope.2.5x objectiveTime-lapse: 50 points, 1 stack every 20min (16.5 hours total).
Courtesy of Gerald Moncayo (Hemmings lab, FMI)
![Page 17: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/17.jpg)
Overview
Growing cells under a microscope…– Physical integrity– Holders, attachment– Incubation systems
(Temperature, Gases, Humidity, pH…)
Imaging living cells– Autofluorescence– Photodamage– Microscope settings: speed, laser power, pinhole, voxel size…– Microscope setups: Spinning-disk versus single-beam LSM and wide-field
microscope.
![Page 18: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/18.jpg)
Choice of the imaging setup
Which microscope is the most appropriate for a live cell experiment?
Go more for sensitivity than resolution!
![Page 19: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/19.jpg)
Autofluorescence
Specific sources of autofluorescence– Aromatic amino acid residues (UV)– Reduced pyridine nucleotides (UV)– Flavins (UV, blue)– Chitin (broad)– Yolk– Chlorophyll (blue, green)– Phenol red indicator use ‘Imaging medium’ !
General sources of autofluorescence– Dead cells (broad)
Cures – Longer wavelengths– Avoid stress– Use imaging medium (no phenol red)– Choose excitation and detection windows properly
![Page 20: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/20.jpg)
Photodamage and photobleaching
With live-cell microscopy, there must be a compromise between acquiring beautiful images and collecting data that provide a high enough signal-to-noise ratio (S/N) to make meaningful quantitative measurements of a living specimen.
Triplett state:• chemical reactive• radicals• bleaching / phototoxicity
A lot of light, not absorbed by the fluorophore, can also potentially trigger chemical reactions in the sample by bringing directly energy to molecules or heating up locally the sample.
Fluorescent molecules used for the labeling of the sample can themselves cause deleterious effects.
Example: Be very careful when using DNA intercalating agents, such as DAPI or Hoechst!
![Page 21: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/21.jpg)
Recognize damaged cells
- Cells detach and round up- Bulges form (“Blebbing”)- Mitochondria swell and/or get isolated- Large vacuoles appear- Cells do not make it through mitosis- Any sign of necrosis or apoptosis
Monitoring of cellular metabolic activity can be monitored for example with AlamarBlue (Invitrogen).
![Page 22: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/22.jpg)
Avoid Photodamage
- Minimize illumination power, tolerate more noise and less S/N ratio
- Optimise detection:FiltersetsDetectorsResolution (x, y, z, t, intensity value, channels)
- Use decent dyes (high quantum efficiency).
- Add antioxidants (Trolox, ascorbic acid 2mg/ml)
![Page 23: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/23.jpg)
Is max resolution needed for my live cell experiment?
Relative Brightness of objective lenses NA4/M2
Magnification (M)
The brightness of an image varies directly as the fourth power of the NA but inversely as the square of the magnification.
Increasing the number of pixels or increasing magnification decreases the intensity in each pixel Resolution is increased only as long as the signal-to-noise ratio is good
40x/1.3
20x/0.8 63x/1.4
100x/1.35x/0.15
10x/0.45 25x/0.8
Detector
Sample
![Page 24: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/24.jpg)
Resolution, pixel size, binningIncreasing the number of pixels decreases the intensity in each pixel, so resolution is increased only as long as the signal-to-noise ratio is good (and optical resolution not limiting).
N.B.: Decreasing pixel size by two decreases signal-to-noise ratio by 2 also! (see also later section on noise).
Binning 3xNo binning
Binning pixels increases intensity but decreases resolution. Exposure time is also decreased (as well as file size) Interesting for live cell experiments.
![Page 25: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/25.jpg)
Binning, contrast and resolutionWhat are my object size and shape? I can’t be sure… Unless I do binning !NB: “binning” is achieved on a CLSM by reducing the number of pixels while keeping the field of view constant (“Frame size” 512X512 256x256).
No binning Binning 2x2
Zeiss Z1, 100x/1.4
![Page 26: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/26.jpg)
Binning, contrast, speed and photodamage
Mouse brain section, fixed section, Alexa488 staining, Zeiss Z1, 20x/0.8Binning 2x2, exposure time 10ms No binning, exposure time 320ms
Courtesy of Flavio Donato (group P. Caroni, FMI)
Playing on pixel size (binning) and contrast (do not use full dynamic range/well capacity) allowed here to reduce by 32 the exposure time more time for Z-stacks and less photodamage.
![Page 27: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/27.jpg)
Is max resolution needed for my live cell experiment?
Region bleached for FRAP
Ex: FRAP experimentGFP-labeled protein in the paternal pronucleus of a mouse zygote.LSM710 Zeiss microscope, 40x/1.3 objective.Time-lapse: 250 points, 1 frame every 134msec.Pinhole ~1.5 AU
Courtesy of Rico Kunzmann (Group A. Peters, FMI)
![Page 28: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/28.jpg)
Refraction around the coverslip
![Page 29: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/29.jpg)
NA and immersion medium
The higher the refractive index of the immersion medium, the higher the NA…
The higher the NA, the brighter the sample and the higher the resolution…
air, n = 1 oil, n = 1.51
![Page 30: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/30.jpg)
Immersion medium and mounting medium refractive indexesThe refractive index of the immersion medium must fit that of the mounting mediumwhen objects are located far away from the coverslip.
oil, n = 1.51
Thick tissue slicen < 1.51
![Page 31: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/31.jpg)
Objective numerical aperture - definition
Apochromat 63x/1.3 glycerol immersion Apochromat 63x/1.2 water immersion
Moss in water
(Courtesy of Nathalie Garin, Leica)
![Page 32: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/32.jpg)
Working distanceThe working distance is defined as the distance (in millimeters) from the front lens element of the objective to the closest surface of the coverslip when the specimen is in sharp focus.
In most instances, the working distance of an objective decreases as magnification increases.
![Page 33: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/33.jpg)
Principle of Confocal Laser Scanning Microscopy (CLSM)
PMT
Objective
Specimenfocal plane
Dichroic
Laser
Emission Filter
Pinhole
![Page 34: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/34.jpg)
Laser beam shape and photobleaching
confocal microscopy multi-photon microscopy
focal plane
coverslide
focal plane
Photo
bleac
hing
patte
rnLa
ser b
eam
shap
e
Scanned area
Reduced phototoxicity but heating problems and higher photobleaching may occur due to high laser power.
Scanned area
![Page 35: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/35.jpg)
Principle of Confocal Laser Scanning Microscopy (CLSM)
PMT
Objective
Specimenfocal plane
Dichroic
Laser
Emission Filter
Pinhole
![Page 36: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/36.jpg)
Principle of Confocal Laser Scanning Microscopy (CLSM)
The acquisition speed depends on - pixel dwell-time (time spent per pixel)- number of pixels
tone frame > number of pixel * pixel dwell time
Example 1: x*y format = 512 * 512pixel dwell time = 2.16 μst > 512 * 512 * 2.16 = 0.57 s
Example 2:with x*y format = 1024*1024pixel dwell time = 3.5 μsaverage = 4t > 1024 * 1024 * 305 * 4 = 14.7 s
Scanning is a slow process !! To go faster, parallel acquisition of several pixels is required (spinning-disk microscope)
PMT
Objective
Specimenfocal plane
Dichroic
Laser
Emission Filter
Pinhole
![Page 37: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/37.jpg)
Live Cell Confocal Settings
• Avoid averaging, but if it is necessary then use line average instead frame average .
• Use line-by-line sequential instead of frame-by-frame sequential
• Try to acquire channels in parallel and not sequentially
• Minimize laser power, compensate by:– Increased sensitivity (higher gain) more noise, process
the image after acquisition (see later)– Opening pinhole increased optical thickness– Less pixels larger voxel size (x, y, z) or scan every
second line.
Lineaverage
Frameaverage
Linesequential
Framesequential
![Page 38: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/38.jpg)
Parallel versus single point acquisition
Nipkow Disk
![Page 39: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/39.jpg)
Optical Path Configuration in CSU22
![Page 40: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/40.jpg)
Comparison CLSM, SD and wide-field systems
Camera chip
Hg Bulb
Pinhole
PMT
.…
Laser
Camera chip
Laser
![Page 41: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/41.jpg)
Higher sensitivity in SD/WF due to (EM-)CCD cameras
- NB: Confocal microscopes become more and more sensitive though, owing to a new generation of detectors: the GaAsP detectors (“BIG” for Zeiss, “HyD” for Leica).
- Read-out noise can be lowered when using lower read speeds of the camera.
![Page 42: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/42.jpg)
√ (read noise)2 + (dark noise)2 + (Shot noise)2
Signal to noise ratio (S/N)
S/N = Signal
Number of photons
S/N
Andor catalogue (2006)
noise dominated byphotonic component
noise dominated byread and dark components
![Page 43: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/43.jpg)
Duplication of supernumerary centrosomes
Multi-beam scanSingle-beam scan
mitotic DdCP224-GFP/GFP-histone2B cells
![Page 44: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/44.jpg)
Comparison of signal to noise ratio between LSM and SD
Single-beam LSM scan Multi-beam SD scan
![Page 45: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/45.jpg)
SD have weak illumination powers
![Page 46: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/46.jpg)
Imaging limitations: available signal
(Laser) excitation is not the only limiting factor, fluorophore saturation decreases overall signal
0
0.2
0.4
0.6
0.8
1
0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35
Excitation photons x1026
Satur
ation
Extended Jablonski diagram
![Page 47: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/47.jpg)
Saturation and triplet state in single- and multi-beam scanning
0.4
0.2
0.6
0.8
1.0
1E20 1E21 1E22 1E23 1E24 1E25Intensity (photons cm-2s-1)
Single-beam scanMulti-beam scan
1mW 488nm, 40x 1.25 NAFlu
orop
hore
satur
ation
Single-beam scanMulti-beam scan
0.4
0.2
0.6
0.8
1.0
1E20 1E21 1E22 1E23 1E24 1E25Intensity (photons cm-2s-1)
Tripl
et po
pulat
ion a
t stea
dy st
ate
![Page 48: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/48.jpg)
Intensity levels in single and multi-beam scanning
Beams 1Illumination 1 mWEmission rate 1.26 x 108 photons/secFluorophore saturation 63%Detection efficiency 0.25Overall efficiency 1
Beams > 1000Illumination 0.4-0.6 µWEmission rate 1.72 x 105 photons/secFluorophore saturation 0.09%Detection efficiency 0.9Overall efficiency 4
Model calculations based on scanning fluorescein at 1 mW 488 nm excitation with a 40x/1.25NA objective, from Live Cell Spinning Disk Microscopy, Ralph Gräf , Jens Rietdorf & Timo Zimmermann, Adv Biochem Engin/Biotechnol (2005) 95: 57–75
![Page 49: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/49.jpg)
Light-dependent photobleaching?
“Photochemical damage is one of the most important yet least understood aspects of the use of fluorescence in biology; in this discussion we can do little more than define our ignorance.”Roger Tsien (Nobel Prize for Chemistry 2008)
Triplett state:• chemical reactive• radicals• bleaching / phototoxicity
![Page 50: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/50.jpg)
Consequences of a not adjustable pinhole size in a SD
N.A. → ↘ → MAG → → ↘λ ↘ → →Resolution ↗ ↘ ~ ↘
![Page 51: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/51.jpg)
Comparison LSM vs. Spinning-Disk confocal
Property Single-beam scanning confocals
Multi-beam scanning confocals
Wide-field setup
Acquisition point by pointslow
Numerous pointsfast
All frame at a timevery fast
Detection Photomultiplier Tubelow sensitivity
CCD/EM-CCD Camerahigh sensitivity
CCD/EM-CCD Camerahigh sensitivity
Multichannel imaging
Parallel/Sequential= fast
Sequential/(Parallel)slow (time x nb of channels)
Sequential/(Parallel)slow (time x nb of channels)
Region bleaching Integrated Add-on Add-on
Optical sectioning Adjustable pinholesFlexible
a lot of light is lostNo crosstalk between
scan beams
Fixed pinhole sizeinflexible/suboptimala lot of light is lost
Crosstalk between scan beams
No sectioningall light collected
Possibility to do TIRF
![Page 52: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/52.jpg)
Image processing in live cell imaging
Very often, images are noisy, as low laser powers are used, and/or short exposure times, and/or fluorescent proteins with low quantum yields.
Deconvolution or denoising can often improves dramatically image quality.
![Page 53: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/53.jpg)
Image processing in live cell imaging: PureDenoise pluginPureDenoise is a plugin in ImageJ, Florian Luisier at the Biomedical Imaging Group (BIG), EPFL, Switzerland, http://bigwww.epfl.ch/algorithms/denoise/
Mammalian cells, GFP-labeled microtubule-organizing centre (MTOC)SD-microscope, 40x/1.3 objective, exposure time: 100ms per frame.Z-stack: 81 planes, dist=500nm.Time-lapse: 210 points, 1 stack every 2 min. After acquisition, images are denoised with PureDenoise over time and MIPs over Z-dimension are generated.
After denoising Before denoising
Courtesy of Julia Kleylein-Sohn (Novartis)
17’010 frames within 7 hours
![Page 54: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/54.jpg)
Image processing in live cell imaging: Deconvolution
YFP-labeled locus in Yeast nucleus.SD-microscope, 100x/1.45 objective, exposure time: 30ms per frame.Z-stack: 14 planes, dist=200nm.Time-lapse: 201 points, 1 stack every 1.5 sec.
After acquisition, images are deconvolved with Huygens professional (www.svi.nl) and MIPs over Z-dimension are generated.
Courtesy of Vincent Dion (Gasser lab, FMI)
After deconvolution Before deconvolution
2’814 frames within 5 minutes
![Page 55: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/55.jpg)
Working with fluorescent fusion proteins: be careful!- Check expression level sand degradation
- Check the functionality of the fusion protein… Adjusting expression levels to those of the endogenous protein!
ng of vector transfected 10 20 30 40 10 20 10 20A-ECFP B-EYFP EYFP-B
![Page 56: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/56.jpg)
Working with fluorescent fusion proteins: be careful!
- Look for artificial clustering of the over-expressed proteins
YFP-PPARα
![Page 57: Laurent Gelman Facility for Advanced Imaging and ... · Live cell imaging and confocal microscopy Laurent Gelman Facility for Advanced Imaging and Microscopy. Friedrich Miescher Institut,](https://reader030.fdocuments.in/reader030/viewer/2022040511/5e5ab02188bd643c1c0a118c/html5/thumbnails/57.jpg)
End