14
S.No Discipline / Group Quantity Measured/ Instrument Range / Frequency * Calibration Measurement Capability(±) Remarks Permanent Facility 1 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Current @ 50 Hz to 1 kHz 1 mA to 10 mA 0.17 % to 0.25 % Using 6½ High Precision Digital Multimeter by Direct Method 2 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Current @ 50 Hz to 1 kHz 10 mA to 3 A 0.25 % to 0.25 % Using 6½ High Precision Digital Multimeter by Direct Method 3 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Current @ 50 Hz to 1 kHz 3 A to 10 A 0.25 % Using 6½ High Precision Digital Multimeter by Direct Method 4 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Voltage @ 50 Hz 1 kV to 5 kV 3.5% to 3.3 % Using High Voltage probe and 4½ Digit Multimeter by Direct Method 5 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Voltage @ 50 Hz to 1 kHz 1 V to 100 V 0.11 % to 0.10 % Using 6½ High Precision Digital Multimeter by Direct Method 6 ELECTRO- TECHNICAL- ALTERNATING CURRENT (< 1 GHZ) (Measure) AC Voltage @ 50 Hz to 1 kHz 100 mV to 1 V 0.12 % to 0.11 % Using 6½ High Precision Digital Multimeter by Direct Method This is annexure to 'Certificate of Accreditation' and does not require any signature. Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation Standard ISO/IEC 17025:2017 Certificate Number CC-2079 Page No. : 1 / 14 Validity 21/01/2019 to 20/01/2021 Last Amended on -

Transcript of Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION...

Page 1: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

Permanent Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50 Hz to1 kHz

1 mA to 10 mA 0.17 % to 0.25 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50 Hz to1 kHz

10 mA to 3 A 0.25 % to 0.25 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50 Hz to1 kHz

3 A to 10 A 0.25 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50 Hz 1 kV to 5 kV 3.5% to 3.3 % Using High Voltageprobe and 4½ DigitMultimeter by DirectMethod

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50 Hz to1 kHz

1 V to 100 V 0.11 % to 0.10 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50 Hz to1 kHz

100 mV to 1 V 0.12 % to 0.11 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 1 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 2: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50 Hz to1 kHz

100 V to 1000 V 0.10 % to 0.10 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current @ 45 Hz to1 kHz

1 A to 10 A 0.16 % to 0.4 % Using Multi ProductCalibrator by DirectMethod

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current @ 45 Hz to1 kHz

1 mA to 10 mA 0.15 % to 0.14 % Using Multi ProductCalibrator by DirectMethod

10 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current @ 45 Hz to1 kHz

10 mA to 1 A 0.14 % to 0.16 % Using Multi ProductCalibrator by DirectMethod

11 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current @ 50Hz 10 A to 500 A 0.4 % to 0.35 % Using Multi ProductCalibrator and Currentcoil 50 turns by DirectMethod

12 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Power SinglePhase 50 Hz @ U.P.F120 V to 240 V10 mAto 10 A

1.2 W to 2.4 kW 0.19 % to 0.17 % Using Multi ProductCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 2 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 3: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

13 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Voltage @ 45 Hz to1 kHz

100 V to 300 V 0.066 % to 0.065 % Using Multi ProductCalibrator by DirectMethod

14 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Voltage @ 45 Hz to1 kHz

300 V to 1000 V 0.065 % to 0.068 % Using Multi ProductCalibrator by DirectMethod

15 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Voltage @ 45 Hz to10 kHz

1 V to 100 V 0.04 % to 0.11 % Using Multi ProductCalibrator by DirectMethod

16 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Voltage @ 45 Hz to10 kHz

10 mV to 100 mV 0.41 % to 0.083 % Using Multi ProductCalibrator by DirectMethod

17 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Voltage @ 45 Hz to10 kHz

100 mV to 1 V 0.083 % to 0.04 % Using Multi ProductCalibrator by DirectMethod

18 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 1 kHz 1 nF to 10 nF 1.74 % to 0.7 % Using Multi ProductCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 3 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 4: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

19 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 1 kHz 10 nF to 100 nF 0.7 % to 0.42 % Using Multi ProductCalibrator by DirectMethod

20 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 1 kHz 100 nF to 1 µF 0.42 % to 0.41 % Using Multi ProductCalibrator by DirectMethod

21 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 100 Hz 1 µF to 10 µF 0.41 % to 0.53 % Using Multi ProductCalibrator by DirectMethod

22 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 100 Hz 10 µF to 100 µF 0.53 % to 0.70 % Using Multi ProductCalibrator by DirectMethod

23 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Capacitance @ 100 Hz 100 µF to 0.3 mF 0.70 % to 0.95 % Using Multi ProductCalibrator by DirectMethod

24 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 A to 3 A 0.082 % to 0.15 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

25 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 mA to 10 mA 0.064 % to 0.081 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 4 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 5: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

26 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10 mA to 100 mA 0.081 % to 0.064 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

27 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100 mA to 1 A 0.064 % to 0.082 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

28 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 3 A to 10 A 0.15 % to 0.20% Using 6½ HighPrecision DigitalMultimeter by DirectMethod

29 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 1 mV to 10 mV 0.41 % to 0.045 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

30 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 10 mV to 100 mV 0.045 % to 0.0084 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

31 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 10 V to 1000 V 0.0034 % to 0.006 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

32 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 100 mV to 10 V 0.0084 % to 0.0034 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

33 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 1 Mohm to 10 Mohm 0.013 % to 0.047 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 5 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 6: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

34 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 1 ohm to 10 ohm 0.36 % to 0.047 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

35 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 10 Mohm to 100 Mohm 0.047 % to 0.94 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

36 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 10 ohm to 100 ohm 0.047 % to 0.016 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

37 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 100 Mohm to 1000Mohm

0.94 % to 2.32 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

38 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

Resistance 100 ohm to 1 Mohm 0.016 % to 0.013 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

39 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 1 A to 10 A 0.04 % to 0.074% Using Multi ProductCalibrator by DirectMethod

40 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 1 mA to 10 mA 0.022 % to 0.019 % Using Multi ProductCalibrator by DirectMethod

41 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 A to 500 A 0.3% to 0.30 % Using Multi ProductCalibrator and Currentcoil 50 turns by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 6 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 7: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

42 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 mA to 1 A 0.019 % to 0.04 % Using Multi ProductCalibrator by DirectMethod

43 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 0.1 Mohm to 1 Mohm 2.31 % to 2.31 % Using High Stabilitydecade Megohm boxby Direct Method

44 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 0.1 ohm to 1 ohm 0.59 % to 0.06 % Using High precisiondecade resistance boxby Direct Method

45 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 1 kohm to 10 kohm 0.018 % to 0.018 % Using Multi ProductCalibrator by DirectMethod

46 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 1 Mohm to 1000 Mohm 2.31 % to 2.32 % Using High Stabilitydecade Megohm boxby Direct Method

47 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 1 ohm to 1 kohm 0.06 % to 0.058 % Using High precisiondecade resistance boxby Direct Method

48 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 1 ohmto 10 ohm

0.94 % to 0.107 %

Using Multi ProductCalibrator by DirectMethod

49 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 10 kohm to 10 Mohm 0.018 % to 0.076 % Using Multi ProductCalibrator by DirectMethod

50 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 10 Mohm to 100 Mohm 0.076 % to 0.58 % Using Multi ProductCalibrator by DirectMethod

51 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Resistance 10 ohm to 1 kohm 0.107 % to 0.018 % Using Multi ProductCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 7 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 8: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

52 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Voltage 1 mV to 10 mV 0.36 % to 0.04 % Using Multi ProductCalibrator by DirectMethod

53 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Voltage 1 V to 1000 V 0.007 % to 0.0067 % Using Multi ProductCalibrator by DirectMethod

54 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Voltage 10 mV to 1 V 0.04 % to 0.007 % Using Multi ProductCalibrator by DirectMethod

55 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

J-TYPE(Recorder/Indicator/Controller)

-200 °C to -30 °C 0.33 °C Using Multi ProductCalibrator by DirectMethod

56 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

J-TYPE(Recorder/Indicator/Controller)

-30 °C to 760 °C 0.22 °C Using Multi ProductCalibrator by DirectMethod

57 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

J-TYPE(Recorder/Indicator/Controller)

760 °C to 1200 °C 0.29 °C Using Multi ProductCalibrator by DirectMethod

58 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

K-TYPE(Recorder/Indicator/Controller)

120 °C to 1350 °C 0.48 °C Using Multi ProductCalibrator by DirectMethod

59 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

K-TYPE(Recorder/Indicator/Controller)

-200 °C to -25 °C 0.39 °C Using Multi ProductCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 8 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 9: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

60 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

K-TYPE(Recorder/Indicator/Controller)

-25 °C to 120 °C 0.23 °C Using Multi ProductCalibrator by DirectMethod

61 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

RTD (PT 100)(Recorder/Indicator/Controller)

-200 °C to 600 °C 0.15°C Using Multi ProductCalibrator by DirectMethod

62 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

R-TYPE(Recorder/Indicator/Controller)

0 °C to 1000 °C 0.7 °C Using Multi ProductCalibrator by DirectMethod

63 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

R-TYPE(Recorder/Indicator/Controller)

1000 °C to 1750 °C 0.75 °C Using Multi ProductCalibrator by DirectMethod

64 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

S-TYPE(Recorder/Indicator/Controller)

0 °C to 1000 °C 0.73 °C Using Multi ProductCalibrator by DirectMethod

65 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

S-TYPE(Recorder/Indicator/Controller)

1000 °C to 1750 °C 0.80°C Using Multi ProductCalibrator by DirectMethod

66 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

T-TYPE(Recorder/Indicator/Controller)

-200 °C to 400 °C 0.75 °C Using Multi ProductCalibrator by DirectMethod

67 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 100 Hz 0.08 % to 0.013 % Using 6½ HighPrecision DigitalMultimeter by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 9 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 10: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

68 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 100 Hz to 100 kHz 0.013% Using 6½ HighPrecision DigitalMultimeter by DirectMethod

69 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1.4 sec to 600 sec 0.13 sec to 0.41 sec Using Time Totaliser byComparison Method

70 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 10000 sec to 86400sec

5.91 sec to 50 sec Using Time Totaliser byComparison Method

71 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 600 sec to 10000 sec 0.41 sec to 5.91 sec Using Time Totaliser byComparison Method

72 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 10 Hz to 100 Hz 0.014 % to 0.004 % Using Multi ProductCalibrator by DirectMethod

73 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 100 Hz to 100 kHz 0.004 % to 0.003 % Using Multi ProductCalibrator by DirectMethod

74 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Bore DialGauge(Transmissionerror only)Span Dia. 6mm to 300 mm.L.C.: 1µm

Up to 1 mm 0.9 µm Dial Gauge Calibratorwith DRO Electronicprobe with Indicator(LVDT) by ComparisonMethod

75 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical measuringpins(Grade "1" andCoarser)

1 mm to 20 mm 1.0 µm Electronic probe withcomparator. Methodbased

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 10 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 11: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

76 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth MicrometerL.C.:10 µm

0 mm to 100 mm 3.0 µm Grade "0" GaugeBlocks & Surface plateby Comparison Method

77 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial / Digital PlungerDial GaugeL.C.: 1 µm

0 to 1 mm 0.88 µm Dial Gauge Calibratorwith DRO Electronicprobe with Indicator(LVDT) by ComparisonMethod

78 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ThicknessGaugeLC: 0.01 mm.

0 to 10 mm 6.4 µm Grade "0" GaugeBlocks by ComparisonMethod

79 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital Depth VernierL.C.: 10 µm

0 mm to 300 mm 9.30 µm Grade "0" GaugeBlocks, Grade "0" LongGauge Blocks byComparison Method

80 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital HeightGaugeL.C.: 10 µm

0 mm to 300 mm 9.5 µm Grade "0" GaugeBlocks,Grade "0" LongGauge Blocks byComparison Method

81 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital Micrometer L.C.:1 µm

> 25 mm to 150 mm 1.22 µm Grade "0" GaugeBlocks,Grade "0" LongGauge Blocks byComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 11 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 12: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

82 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital MicrometerL.C.:1 µm

0 to 25 mm 0.81 µm Grade "0" GaugeBlocks,Grade "0" LongGauge Blocks byComparison Method

83 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital Vernier CaliperL.C.: 10 µm

0 mm to 300 mm 9.5 µm Caliper Checker/Grade"0" Gauge Blocks byComparison Method

84 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Digital Vernier CaliperL.C.: 10 µm

300 mm to 600 mm 11.8 µm Caliper Checker/Grade"0" Gauge Blocks byComparison Method

85 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External MicrometerL.C.: 10 µm

> 150 mm to 300 mm 3.4 µm. Grade "0" GaugeBlocks,Grade "0" LongGauge Blocks byComparison Method

86 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Type Dial GaugeL.C.: 2 µm

Up to 0.2 mm 1.40 µm Using Dial GaugeCalibrator with DROElectronic probe withIndicator (LVDT) byComparison Method

87 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Type DialGaugeL.C.: 10 µm

0.8 mm 5.8 µm Using Dial GaugeCalibrator with DROElectronic probe withIndicator (LVDT) byComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 12 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 13: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

88 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Plug Gauge 5 mm to 90 mm 2.5 µm Floating Carriage DialMeasuring Machine byComparison Method

89 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plunger DialGaugeL.C.: 10 µm

1 mm to 12 mm 3.0 µm Dial Gauge Calibratorwith DRO Electronicprobe with Indicator(LVDT) by ComparisonMethod

90 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Test Sieves 10 mm to 100 mm 22.0 µm Digital Vernier Caliperby Comparison Methodbased

91 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thread Plug Gauge 5 mm to 100 mm 2.5 µm Floating CarriageDia.MeasuringMachine, SettingMaster & ThreadMeasuring Wires byComparison Method

92 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Vernier Height GaugeL.C.: 20 µm

300 mm to 600 mm 10.7 µm Grade "0" GaugeBlocksGrade "0" LongGauge Blocks byComparison Method

93 MECHANICAL-PRESSUREINDICATING DEVICES

Pressure ( Hydraulic )Pressure Gauges /Pressure Transducers

0 to 100 bar 0.24 % of rdg Pressure calibratorComparison Methodbased on DKD-R-6-1

94 MECHANICAL-PRESSUREINDICATING DEVICES

Pressure (Hydraulic)Pressure Gauges /Pressure Transducers.

0 to 700 bar 0.54 % of rdg Pressure calibrator byComparison Methodbased on DKD-R-6-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 13 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -

Page 14: Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59 ......Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM, COIMBATORE, TAMIL NADU, INDIA Accreditation

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

95 MECHANICAL-PRESSUREINDICATING DEVICES

Vacuum Gauges /Vacuum Transducers.

(-) 50 mmHg to (-) 600mmHg

0.81 % of rdg Digital pressure gaugeComparison Methodbased on DKD-R-6-2

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name BEST ENGINEERS CALIBRATION LAB, 59, THADAGAM ROAD, VELANDIPALAYAM,COIMBATORE, TAMIL NADU, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2079 Page No. : 14 / 14

Validity 21/01/2019 to 20/01/2021 Last Amended on -