LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3)...

31
Setsco Services Pte Ltd Certificate No. : LA-1993-0051-C No. 18 Teban Gardens Crescent Singapore 608925 Issue No. : 20 Date : 26 April 2016 Page : 1 of 31 FIELD OF TESTING : Calibration and Measurement MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE CALIBRATED METHOD CALIBRATION & MEASUREMENT CAPABILITY (CMC*) A MECHANICAL 1. PRESSURE AND VACUUM MEASURING DEVICES (Lab/Site) - Compound Gauges - Digital Pressure Gauges - Pressure Gauges - Test Gauges - Vacuum Gauges - Transducer - Recorder - Pressure Modules - Pressure Instruments - Pressure Transmitters - Barometer Pressure Measuring Devices 1) 0 to 50 mbar ) MTD/CAL-05 : 2015 0.02 mbar 2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150 psi ) 0.02 psi 6) 150 to 300 psi ) 0.03 psi 7) 300 to 400 psi ) 0.04 psi 8) 400 to 600 psi ) 0.08 psi 9) 600 to 800 psi ) 0.1 psi 10) 800 to 1 000 psi ) 0.2 psi 11) 1 000 to 2 000 psi ) 0.4 psi 12) 2 000 to 3 000 psi ) 0.2 psi 13) 3 000 to 15 000 psi ) 2 psi

Transcript of LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3)...

Page 1: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Setsco Services Pte Ltd Certificate No. : LA-1993-0051-C No. 18 Teban Gardens Crescent Singapore 608925 Issue No. : 20 Date : 26 April 2016

Page : 1 of 31

FIELD OF TESTING : Calibration and Measurement

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

A MECHANICAL

1. PRESSURE AND VACUUM

MEASURING DEVICES (Lab/Site)

- Compound Gauges

- Digital Pressure Gauges

- Pressure Gauges

- Test Gauges

- Vacuum Gauges

- Transducer

- Recorder

- Pressure Modules

- Pressure Instruments

- Pressure Transmitters

- Barometer

Pressure Measuring Devices

1) 0 to 50 mbar ) MTD/CAL-05 : 2015 0.02 mbar

2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar

3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi

4) 30 to 100 psi ) 0.01 psi

5) 100 to 150 psi ) 0.02 psi

6) 150 to 300 psi ) 0.03 psi

7) 300 to 400 psi ) 0.04 psi

8) 400 to 600 psi ) 0.08 psi

9) 600 to 800 psi ) 0.1 psi

10) 800 to 1 000 psi ) 0.2 psi

11) 1 000 to 2 000 psi ) 0.4 psi

12) 2 000 to 3 000 psi ) 0.2 psi

13) 3 000 to 15 000 psi ) 2 psi

Page 2: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 2 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

14) 15 000 to 20 000 psi ) MTD/CAL-05 : 2015 15 psi

15) 20 000 to 30 000 psi ) BS EN 837-1 & 3 : 1998 52 psi

16) 30 000 to 60 000 psi ) DIN 16005 : 1987 125 psi

)

Vacuum Pressure Devices )

1) -1 to 0 bar ) 0.0002 bar

2) -50 to 0 mbar ) 0.02 mbar

)

Barometric Pressure Devices )

1) 960 to 1070 hpa ) 0.6 hpa

2. FORCE MEASURING DEVICES

(Lab/Site)

- Adhesion Tester

- Force Gauge

- Load Cells

- Hydraulic Load Cells

- Crane Scale

- Dynamometer

- Proving Rings

- Spring Load Tester

- Spring Scale

- Hydraulic Jacks

- Universal Testing Machine

- Load Testing Machine

- Compression Testing Machine

- Tension Testing Machine

In Compression Mode ) MTD/CAL-01 : 2015

1) Using dead weights ) MTD/CAL-06 : 2015

0 N to 5.0 N ) BS EN ISO 7500-1: 2004 0.001 N

5 N to 10 N ) 0.002 N

10 N to 20 N ) 0.001 N

20 N to 50 N ) 0.004 N

50 N to 200 N ) 0.007 N

200 N to 1 000 N ) 0.06 N

1 000 N to 3 000 N ) 1.0 N

3 000 N to 8 000 N ) 2.0 N

Page 3: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 3 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

2) Using load cells ) MTD/CAL-01 : 2015

0 N to 1.5 N ) MTD/CAL-06 : 2015 0.002 N

1.5 N to 2.5.N ) BS EN ISO 7500-1: 2004 0.007 N

0 kN to 2 kN ) 0.001 kN

2 kN to 5 kN ) 0.003 kN

5 kN to 10 kN ) 0.005 kN

10 kN to 50 kN ) 0.02 kN

50 kN to 100 kN ) 0.1 kN

100 kN to 300 kN ) 0.2 kN

300 kN to 400 kN ) 0.3 kN

400 kN to 500 kN ) 0.4 kN

300 kN to 1 500 kN ) 0.8 kN

1 200 kN to 2 000 kN ) 1.0 kN

1 800 kN to 3 000 kN ) 2.0 kN

)

In Tension Mode ) MTD/CAL-01 : 2015

1) Using dead weights ) MTD/CAL-06 : 2015

0 N to 20 N ) BS EN ISO 7500-1: 2004 0.001 N

20 N to 200 N ) 0.007 N

200 N to 1 000 N ) 0.1 N

1 000 N to 3 000 N ) 1.0 N

)

2) Using load cells )

0 kN to 1.5 N ) 0.003 N

1.5 N to 2.5 N ) 0.005 N

0 kN to 2.0 kN ) 0.0008 kN

2.0 kN to 5.0 kN ) 0.003 kN

5.0 kN to 10 kN ) 0.005 kN

10 kN to 100 kN ) 0.01 kN

100 kN to 200 kN ) 0.1 kN

200 kN to 450 kN ) 0.5 kN

)

3) Using universal testing machine )

0 kN to 950 kN ) 4.0 kN

Page 4: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 4 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

3. FORCE PROVING DEVICE (Lab)

- Proving Devices

(Classification 0.5, 1 and 2)

In Compression Mode ) MTD/CAL-39: 2015

1) Using dead weights ) BS EN ISO 376: 2011

0 N to 50 N ) 0.02 N

50 N to 200 N ) 0.06 N

200 N to 500 N ) 0.05 N

500 N to 3 000 N ) 0.8 N

2) Using load cells )

30 kN to 300 kN ) 0.05 kN

)

In Tension Mode )

1) Using dead weights )

0 N to 3 000 N ) 0.8 N

4. COMPRESSION TESTING

MACHINES (Site)

1) Platen Self-Alignment ) BS 1881 : Pt 115 : 1986

) SS 78 : Pt A15 : 1987

2) Frame Alignment Test ) MTD/CAL-26 : 2015

) BS EN 12390-4 : 2000

3) Offset Test )

0 kN to 10 kN ) 0.005 kN

10 kN to 50 kN ) 0.02 kN

50 kN to 100 kN ) 0.1 kN

100 kN to 300 kN ) 0.2 kN

300 kN to 500 kN ) 0.4 kN

500 kN to 2 000 kN ) 1.0 kN

2 000 kN to 3 000 kN ) 2.0 kN

Page 5: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 5 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

5. HARDNESS TESTING MACHINE

(Lab/Site)

1. Brinell Scale ) MTD/CAL-25 :2015

i) Indirect Verification ) ASTM E10-15

HBW ≤ 125 (1/10) ) BS EN ISO 6506-2 : 2005 0.004 mm

125 < HBW ≤ 225 (1/10) ) 0.002 mm

HBW > 225 (1/10) ) 0.002 mm

HBW ≤ 125 (1/30) ) 0.002 mm

125 < HBW ≤ 225 (1/30) ) 0.001 mm

HBW > 225 (1/30) ) 0.004 mm

HBW ≤ 125 (10/3000) ) 0.02 mm

125 < HBW ≤ 225 (10/3000) ) 0.02 mm

HBW > 225 (10/3000) ) 0.02 mm

)

ii) Direct Verification )

187.5 kgf ) 0.6 kgf

750 kgf ) 4.0 kgf

3 000 kgf ) 5.0 kgf

)

iii) Indenter Measurement )

(Ball Diameter) )

2.5 mm ) 0.0003 mm

5.0 mm ) 0.0003 mm

10.0 mm ) 0.0003 mm

)

iv) Measuring Scope )

1.0 mm to 6.0 mm ) 0.002 mm

)

2. Vickers Scale )

i) Indirect Verification )

a. Vickers (microindentation) )

>1 000 HV (load @ 100 gf) ) 0.2 μm

(load @ 200 gf) ) 0.2 μm

(load @ 300 gf) ) 0.2 μm

(load @ 500 gf) ) 0.3 μm

(load @ 1 000 gf) ) 0.4 μm

Page 6: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 6 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

b. Vickers ) MTD/CAL-25 : 2015

) ASTM E384: 2011e1

>1 000 HV (load @ 5 kgf) ) BS EN ISO 6506-2 : 2005 2.0 μm

(load @ 10 kgf) ) 2.0 μm

(load @ 30 kgf) ) 6.0 μm

)

ii) Direct Verification )

50 gf ) 0.08 gf

100 gf ) 0.1 gf

200 gf ) 0.5 gf

300 gf 0.4 gf

500 gf ) 0.3 gf

1 000 gf ) 0.4 gf

5 000 gf 2 gf

)

iii) Measuring Device )

Indenter Angle (deg) ) 0.03 deg

)

iv) Measuring Scope )

0 μm to 300 μm ) 0.1 μm

300 μm to 500 μm ) 0.2 μm

)

3. Rockwell Scale ) MTD/CAL-25 :2015

i) Indirect Verification ) ASTM E18-15

20 to 40 HRA ) BS EN ISO 6508-2 : 2005 0.4 HRA

70 to 75 HRA ) 0.4 HRA

80 to 84 HRA ) 0.2 HRA

40 to 50 HRBw ) 0.7 HRBw

60 to 79 HRBw ) 0.7 HRBw

85 to 100 HRBw ) 0.7 HRBw

20 to 30 HRC ) 0.2 HRC

35 to 51 HRC ) 0.3 HRC

60 to 65 HRC ) 0.2 HRC

70 to 77 HR15N ) 0.6 HR15N

78 to 88 HR15N ) 0.6 HR15N

90 to 92 HR15N ) 0.5 HR15N

42 to 52 HR30N ) 0.3 HR30N

55 to 73 HR30N ) 0.3 HR30N

77 to 82 HR30N ) 0.6 HR30N

Page 7: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 7 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

ii) Direct Verification )

15 kgf ) 0.03 kgf

60 kgf ) 0.08 kgf

100 kgf ) 0.10 kgf

150 kgf ) 0.14 kgf

)

iii) Indenter Measurement )

Indenter Angle (°) 120° ± 0.1° ) 0.03 deg

Tip Radium (mm) 0.2 ± 0.005 mm ) 0.006 mm

Ball Diameter (mm) 1.588 mm ) 0.001 mm

4. Direct Verification of Durometer (lab) ) MTD/CAL-24 : 2015

(Rubber Hardness Tester)

Shore A, B, C, D, DO, O, JA, JC, 0.1 Div

HAD & HDD

Indenter Length (mm) 0.004 mm

Indenter Diameter (mm) 0.004 mm

Indenter Tip Diameter (mm) 0.004 mm

Bore Diameter (mm) 0.004 mm

Indenter Angel (deg) 0.04 deg

6. BALANCES AND WEIGHING SCALES

(Lab/Site)

1) 0 to 5 000mg )MTD/CAL-07 :2015 0.02 mg

2) 5 000 mg to 20 000 mg 0.03 mg

3) 20 000 to 60 000 mg ) 0.07 mg

4) 60 000 to 200 000 mg ) 0.2 mg

5) 200 to 300 g ) 0.0008 g

6) 300 g to 400 g ) 0.001 g

7) 400 to 1 200 g ) 0.002 g

8) 1 200 to 6 000 g ) 0.02 g

9) 6 000 to 30000 g ) 0.1 g

10) 30 to 60 kg ) 0.0004 kg

11) 64 to 150 kg ) 0.002 kg

12) 150 to 300 kg ) 0.02 kg

13) 300 to 1 000 kg ) 0.2 kg

Page 8: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 8 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

7. WEIGHTS (Lab)

i) Standard Weights ) MTD/CAL-10 : 2015

1) 1 mg to 200 mg ) 0.00002 g

2) 500 mg to 1 g ) 0.00003 g

3) 2 g ) 0.00004 g

4) 5 g ) 0.00005 g

5) 10 g ) 0.00006 g

6) 20 g ) 0.00008 g

7) 50 g ) 0.00010 g

8) 100 g ) 0.0002 g

9) 200 g ) 0.0003 g

10) 500 g to 1 kg ) 0.002 g

11) 2 kg to 5 kg ) 0.02 g

12) 10 kg to 20 kg ) 0.2 g

13) 25 kg ) 0.2 g

)

ii) Weight Measurement )

1) 1 mg to 200 mg ) 0.00002 g

2) 500 mg to 1 g ) 0.00003 g

3) 2 g ) 0.00004 g

4) 5 g ) 0.00005 g

5) 10 g ) 0.00006 g

6) 20 g ) 0.00008 g

7) 50 g ) 0.00010 g

8) 100 g ) 0.0002 g

9) 200 g ) 0.0003 g

10) 500 g to 1 kg ) 0.002 g

11) 2 kg to 5 kg ) 0.02 g

12) 10 kg to 20 kg ) 0.2 g

13) 25 kg ) 0.2 g

14) 30 kg ) 0.2 g

8. TORQUE MEASURING DEVICES (Lab)

1) Torque Wrench Calibration ) MTD/CAL-22 :2015

0.15 Nm to 1.5 Nm ) BS EN ISO 6789: 2003 0.0015 Nm

1.0 Nm to 10 Nm ) 0.015 Nm

10 Nm to 100 Nm ) 0.13 Nm

100 Nm to 1 000 Nm ) 0.88 Nm

250 Nm to 2 500 Nm ) 1.2 Nm

Page 9: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 9 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

2) Torque Meter Tester ) MTD/CAL-37 : 2015

0.45 Nm to 4.5 Nm ) BS 7882: 2008 0.0011 Nm

2.5 Nm to 22.5 Nm ) 0.011 Nm

9. EXTENSOMETER (Lab/Site)

1) Strain displacement measurement ) MTD/CAL-23 :2015 0.002 mm

up to 25 mm or above ) BS EN ISO 9513: 2002

2) Gauge length 50 mm and above ) ASTM E83: 2010a 0.004 mm

Classification limit up to ASTM )

E83 : 2010a Class B1 for gauge

length 50 mm and above and

25 mm at near full scale

10. TACHOMETER (Lab/Site)

i) Non-contact Type (Lab)

10 rpm to 600 rpm ) MTD/CAL-201 : 2009 0.40 rpm

601 rpm to 6 000 rpm ) 1.8 rpm

6 001 rpm to 60 000 rpm ) 3.0 rpm 60 001 rpm to 100 000 rpm ) 4.0 rpm )

ii) Rotor Movement Apparatus (Site) )

30 rpm to 600 rpm ) 4.0 rpm

601 rpm to 6 000 rpm ) 7.0 rpm

6 001 rpm to 99 600 rpm ) 9.0 rpm

11. STROBOSCOPE (Lab)/Site)

i) Non-contact Type (Lab)

10 rpm to 600 rpm ) MTD/CAL-202 : 2009 0.40 rpm

601 rpm to 6 000 rpm ) 1.8 rpm

6 001 rpm to 60 000 rpm ) 2.6 rpm 60 001 rpm to 100 000 rpm ) 3.4 rpm

Page 10: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 10 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

12. SKID RESISTANCE (PENDULUM)

TESTER (Lab)

i) Height Adjustment ) MTD/CAL-138: 2012 See scope C38

ii) Distance of Center of Gravity of ) BS EN 13036-4: 2011 See scope C38

Pendulum/ Arm to Center of ) ASTM E 303: 1993(2008)

Rotation )

iii) Point Length ) See scope C38

iv) Slider to Center Rotation ) See scope C38

v) Mass of Pendulum/ Arm ) 1 g

vi) Pointer Weight ) 1 g

vii) Weight Calibration ) 1 g

viii) Measurement of Slide Force/ ) 0.1 N

Deflection Characteristics )

13. GAS FLOW METER (LAB)

- Flow Meter ) MTD/CAL-40:2015

- Digital Flow Meter ) ASTM D3195/D3195M -10

)

Flow Measuring Devices )

1) 0 to 5 l/min ) 0.03 l/min

2) 5 l/min to 10 l/min ) 0.07 l/min

B. TEMPERATURE MEASURING

DEVICES

1. LIQUID-IN-GLASS THERMOMETERS (Lab)

a) Total Immersion

i) For scale division: 0.01°C to 0.05°C ) MTD/CAL-12 : 2012

-80 °C to 0 °C ) ASTM E 1: 2007 0.028 °C

1 °C to 80 °C ) ASTM E 77: 2007 0.028 °C

81 °C to 280 °C ) ITS 90/ IPTS 68 0.028 °C

)

ii) For scale division: 0.1°C to 0.5°C )

-80 °C to 0 °C ) 0.03 °C

1 °C to 80 °C ) 0.03 °C

81 °C to 280 °C ) 0.03 °C

Page 11: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 11 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

iii) For scale division: 1°C to 2°C ) MTD/CAL-12 : 2012

-80 °C to 0 °C ) ASTM E 1: 2007 0.1°C

1 °C to 80 °C ) ASTM E 77: 2007 0.1°C

81 °C to 280 °C ) ITS 90/ IPTS 68 0.1°C

b) Partial Immersion

i) For scale division: 0.01°C to 0.5°C ) MTD/CAL-12 : 2012

-80 °C to 0 °C ) ASTM E 1: 2007 0.03 °C

1 °C to 80 °C ) ASTM E 77: 2007 0.03 °C

81 °C to 280 °C ) ITS 90/ IPTS 68 0.03 °C

)

ii) For scale division: 1°C to 2°C )

-80 °C to 0 °C ) 0.1°C

1 °C to 80 °C ) 0.1°C

81 °C to 280 °C ) 0.1°C

2. TEMPERATURE SENSOR (Lab)

a) Resistance Temperature Detector ) MTD/CAL-13 : 2012

-80 °C to 280 °C ) ASTM E220: 2007a 0.03°C

281 °C to 400 °C ) ASTM E230/ E230M: 2012 0.13°C

) ASTM E644: 2011

) ASTM E1137/ E1137M: 2008

) JIS C1602: 1995

b) Thermocouple ) JIS C1604: 1997

) IEC 60751: 2008 (Ed.2)

i) Type E ) IEC 60584-1: 1995 (Ed.2)

-80 °C to 280 °C ) IEC 60584-2: 1982 (Ed.1) 0.4°C

281 °C to 400 °C + Amd 1: 1982 (Ed.1) 1.5°C

)

ii) Type J )

-80 °C to 280 °C ) 0.4°C

281 °C to 400 °C ) 1.6°C

)

iii) Type K )

-80 °C to 280 °C ) 0.4°C

281 °C to 400 °C ) 1.6°C

Page 12: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 12 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

iv) Type N )

-80 °C to 280 °C ) 0.5°C

281 °C to 400 °C ) 1.7°C

v) Type R )

0 °C to 280 °C ) 1.0°C

281 °C to 400 °C ) 3.0°C

)

vi) Type S )

0 °C to 280 °C ) 1.0°C

281 °C to 400 °C ) 3.0°C

)

vii) Type T )

-80 °C to 280 °C ) 0.4°C

281 °C to 400 °C ) 0.4°C

3. TEMPERATURE DISPLAY

WITH TEMPERATURE SENSOR (Lab)

a) Temperature Display ) MTD/CAL-14 : 2012

with RTD Sensor ) ITS 90/ IPTS 68

-80 °C to 0 °C ) 0.03°C

1 °C to 80 °C ) 0.03°C

81 °C to 280 °C ) 0.03°C

281 °C to 400 °C ) 0.11°C )

b) Temperature Display )

with Thermocouple Sensor )

-80 °C to 0 °C ) 0.2°C

1 °C to 80 °C ) 0.2°C

81 °C to 280 °C ) 0.2°C

281 °C to 400 °C ) 0.4°C )

c) Temperature Gauge )

-80 °C to 0 °C ) 0.3°C

1 °C to 80 °C ) 0.3°C

81 °C to 280 °C ) 0.3°C

281 °C to 400 °C ) 0.6°C

Page 13: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 13 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

4. TEMPERATURE DISPLAY

a) Thermocouple Display ) MTD/CAL-15 : 2012

(using voltage standard) (Lab) ) ASTM E220: 2007a

) ASTM E230/ E230M: 2012

i) Type E ) ASTM E644: 2011

-270 °C to 600 °C ) ASTM E1137/ E1137M: 2008 0.3°C

600 °C to 1 000 °C ) JIS C1602: 1995 0.3°C

) JIS C1604: 1997

ii) Type J ) IEC 60751: 2008 (Ed.2)

-210 °C to 800 °C ) IEC 60584-1: 1995 (Ed.2) 0.3°C

800 °C to 1 200 °C ) IEC 60584-2: 1982 (Ed.1) 0.3°C

+ Amd 1: 1982 (Ed.1)

iii) Type K )

-270 °C to 400 °C ) 0.4°C

400 °C to 1 372 °C ) 0.4°C

)

iv) Type N )

-270 °C to 900 °C ) 0.4°C

900 °C to 1 300 °C ) 0.5°C

)

v) Type R )

-50 °C to 100 °C ) 1.8°C

100 °C to 1 767 °C ) 1.4°C

)

vi) Type S )

0 °C to 200 °C ) 1.7°C

200 °C to 1 400 °C ) 1.4°C

1 400 °C to 1 767 °C ) 1.4°C

)

vii) Type T )

-200 °C to 0 °C ) 0.4°C

0 °C to 400 °C ) 0.4°C

)

b) Thermocouple Display )

(using calibrator) )

)

i) Type E )

-250 °C to 1 000 °C ) 0.5°C

Page 14: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 14 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

ii) Type J ) MTD/CAL-15 : 2012

-210 °C to 1 200 °C ) ASTM E220: 2007a 0.5°C

) ASTM E230/ E230M: 2012

iii) Type K ) ASTM E644: 2011

-200 °C to 1 372 °C ) ASTM E1137/ E1137M: 2008 0.5°C

) JIS C1602: 1995

iv) Type N ) JIS C1604: 1997

-200 °C to 1 300 °C ) IEC 60751: 2008 (Ed.2) 0.5°C

) IEC 60584-1: 1995 (Ed.2)

v) Type R ) IEC 60584-2: 1982 (Ed.1)

0 °C to 1 767 °C ) + Amd 1: 1982 (Ed.1) 2.8°C

)

vi) Type S )

0 °C to 1 767 °C ) 2.8°C

)

vii) Type T )

-250 °C to 400 °C ) 0.9°C

)

c) RTD Display )

(using decade resistance box) )

-200 °C to 850 °C ) 0.02°C

5. TEMPERATURE ENCLOSURE) (Lab/Site)

) MTD/CAL-20 : 2012

-80 °C to 0 °C ) ASTM E145: 1994(2011) 0.6°C

1 °C to 280 °C ) SAE AMS 2750E: 2012 0.6°C

281 °C to 400 °C ) AS 2853: 1986 0.7°C

401 °C to 1 200 °C ) ITS 90/ IPTS 68 1.8°C

6. TEMPERATURE BLOCK CALIBRATION ) MTD/CAL-27 : 2012

(Lab/ Site) ) EURAMET/CG-13/v.01

-45 °C to 280 °C ) ITS 90/ IPTS 68 0.09°C

280 °C to 1 200 °C ) 0.9°C

Page 15: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 15 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

7. THERMO-HYGROMETER / ) MTD/CAL-29 : 2012

THERMO-HYGROGRAPH / ) BS 1339-1: 2002

(Lab) ) BS 1339-2: 2004

) BS 1339-3: 2004

)

a) Relative Humidity @ 10°C )

20%r.h. @ 10°C ) 0.9% r.h. & 0.3°C

35%r.h. @ 10°C ) 1.2% r.h. & 0.3°C

55%r.h. @ 10°C ) 1.2% r.h. & 0.3°C

75%r.h. @ 10°C ) 2.2% r.h. & 0.3°C

85%r.h. @ 10°C ) 2.5% r.h. & 0.3°C

b) Relative Humidity @ 25°C )

12%r.h. @ 25°C ) 0.7% r.h. & 0.3°C

35%r.h. @ 25°C ) 1.1% r.h. & 0.3°C

55%r.h. @ 25°C ) 1.5% r.h. & 0.3°C

75%r.h. @ 25°C ) 1.9% r.h. & 0.3°C

95%r.h. @ 25°C ) 2.3% r.h. & 0.3°C

c) Relative Humidity @ 55°C )

12%r.h. @ 55°C ) 0.7% r.h. & 0.3°C

35%r.h. @ 55°C ) 1.0% r.h. & 0.3°C

55%r.h. @ 55°C ) 1.2% r.h. & 0.3°C

75%r.h. @ 55°C ) 1.7% r.h. & 0.3°C

95%r.h. @ 55°C ) 2.0% r.h. & 0.3°C

d) Relative Humidity @ 70°C )

12%r.h. @ 70°C ) 0.6% r.h. & 0.3°C

35%r.h. @ 70°C ) 0.9% r.h. & 0.3°C

55%r.h. @ 70°C ) 1.0% r.h. & 0.3°C

75%r.h. @ 70°C ) 1.6% r.h. & 0.3°C

90%r.h. @ 70°C ) 1.9% r.h. & 0.3°C

)

e) Relative Humidity @ 85°C )

12%r.h. @ 85°C 0.6% r.h. & 0.3°C

35%r.h. @ 85°C 0.8% r.h. & 0.3°C

55%r.h. @ 85°C 1.0% r.h. & 0.3°C

75%r.h. @ 85°C 1.3% r.h. & 0.3°C

85%r.h. @ 85°C 1.5% r.h. & 0.3°C

Page 16: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 16 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

8. TEMPERATURE & RELATIVE ) MTD/CAL-36 : 2012

HUMIDITY CHAMBER (Site) ) Guideline – The Society of

Environmental Engineers

a) Relative Humidity @ 10°C )

20%r.h. @ 10°C ) 1.9% r.h. & 0.5°C

55%r.h. @ 10°C ) 2.4% r.h. & 0.5°C

85%r.h. @ 10°C ) 3.4% r.h. & 0.5°C

)

b) Relative Humidity @ 25°C )

12%r.h. @ 25°C ) 1.7% r.h. & 0.5°C

55%r.h. @ 25°C ) 2.3% r.h. & 0.5°C

90%r.h. @ 25°C ) 3.7% r.h. & 0.5°C

)

c) Relative Humidity @ 55°C )

12%r.h. @ 55°C ) 1.8% r.h. & 0.5°C

55%r.h. @ 55°C ) 2.6% r.h. & 0.6°C

95%r.h. @ 55°C ) 3.8% r.h. & 0.5°C

d) Relative Humidity @ 70°C )

12%r.h. @ 70°C ) 1.6% r.h. & 0.6°C

55%r.h. @ 70°C ) 2.3% r.h. & 0.6°C

90%r.h. @ 70°C ) 3.5% r.h. & 0.5°C

)

e) Relative Humidity @ 85°C )

12%r.h. @ 85°C ) 1.8% r.h. & 1.2°C

55%r.h. @ 85°C ) 3.3% r.h. & 1.2°C

85%r.h. @ 85°C ) 3.2% r.h. & 0.9°C

Page 17: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 17 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

C. DIMENSIONAL

Measuring Instruments / Tools

(Metric and Imperial)

1) External Micrometers (Lab) ) MTD/CAL-103 : 2012

Micrometer Screw Accuracy ) BS 870 : 2008

Up to 125 mm ) 1 m

above 125 to 300 mm ) 2 m

above 300 to 425 mm ) 3 m

above 425 to 600 mm ) 4 m

above 600 to 750 mm ) 5 m

above 750 to 900 mm ) 6 m

above 900 to 1000 mm ) 7 m

2) ) Micrometer Head (Lab) ) MTD/CAL-114 : 2012

i) Accuracy ) BS 1734 : 1951

Up to 1.27 mm ) 0.3 m

above 1.27 - 25 mm ) 1 m

above 25 - 50 mm ) 2 m

ii) Flatness ) 0.41 m )

3) Holtest (Micrometer Head Only) - ) (Lab) )

Accuracy (Transverse Distance): )

Up to 13 mm ) 1 m

above 13 - 25 mm ) 2 m

4) Internal Micrometer (Lab) ) MTD/CAL-137 : 2009

Up to 1 500 mm ) BS 959 : 2008

i) Micrometer Screw Accuracy ) 2m ii) Extension Rods: )

up to 13 mm ) 2m above 13 mm to 25 mm ) 2m above 25 mm to 50 mm ) 2m above 50 mm to 75 mm ) 2m above 75 mm to 100 mm ) 2m above 100 mm to 150 mm ) 3m above 150 mm to 200 mm ) 3m above 200 mm to 300 mm ) 3m above 300 mm to 400 mm ) 4m

Page 18: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 18 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

ii) Extension Rods: )

above 400 to 500 mm ) 7 m

above 500 to 600 mm ) 8 m

above 600 to 700 mm ) 10m

above 700 to 800 mm ) 11m

above 800 to 900 mm ) 12m

5) Depth Micrometer (Lab) ) MTD/CAL-107 : 2012

i) ) Micrometer Screw Accuracy ) BS 6468 : 2008

0 mm to 150 mm ) 1 m

150 mm to 300 mm ) 2 m

ii) ) Rod Accuracy: )

0 mm to 150 mm ) 1 m 150 mm to 300 mm ) 2 m

iii) ) Parallelism Measurement ) 2 m

iv) ) Flatness Measurement ) 0.41 m

6) Dial Gauges (Lab) ) MTD/CAL-110 : 2012

Up to 5 mm ) BS 907 : 2008 1.1m Above 5 to 20 mm 1.6m Above 20 to 25 mm 1.7m

7) Linear / Indicating Instrument (Lab) ) MTD/CAL-109 : 2012

(Digital and Dial Type) ) ASME/ANSI B89.1.10M: 2001

Up to 10 mm 0.2m Above 10 to 25 mm 0.3m Above 25 to 50 mm 0.5m Above 50 to 100 mm 2m

8) Dial Test Indicators (Lab) ) MTD/CAL-111 : 2012

Up to 0.6 mm ) BS 2795 : 1981 1.1 m

Above 0.6 to 1.5 mm 1.5 m

9) Dial Depth Gauge (Lab) ) MTD/CAL-113 : 2012

Up to 25 mm ) ASME/ANSI B89.1.10M: 2001 2 m

Flatness Measurement ) 0.41 m

10) Dial Thickness Gauge (Lab) ) MTD/CAL-115 : 2012

Up to 12 mm ) ASME/ANSI B89.1.10M: 2001 1 m

above 12 to 20 mm ) 2 m

Page 19: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 19 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED

METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

11) Caliper (Vernier, Dial & Digital) ) MTD/CAL-104 : 2012

(Lab) ) BS 887 : 2008

Up to 1 200 mm ) JIS B 7507 : 1993 10 m

above 1 200 to 2 000 mm ) 20 m

12) Caliper Depth Gauge (Lab) ) MTD/CAL-106 : 2012

Up to 600 mm ) BS 6365 : 2008 10 m

13) Height Gauge (Lab/Site) ) MTD/CAL-105 : 2012

Up to 300 mm ) BS 1643 : 2008 2 m

above 300 to 400 mm ) JIS Z 8103: 2000 3 m

above 400 to 600 mm ) 4 m

above 600 to 1 000 mm ) 10 m

14) Measuring Profile Projector ) MTD/CAL-112 : 2010

(Lab/Site) ) JIS B 7184 : 1999

i) Linear Measurement )

X-Axis : Up to 250 mm ) 0.008 mm

Y-Axis : Up to 140 mm ) 0.007 mm

ii) Magnification Measurement ) 0.01 %

iii) Angular Measurement ) 0.0167°

15) End Standard Length ) MTD/CAL-108 : 2012

Measurement (Lab/Site) )

i) Indirect Measurement )

up to 25 mm ) 0.4 m

above 25 to 50 mm ) 0.4 m

above 50 to 75 mm ) 0.4 m

above 75 to 100 mm ) 0.5 m

above 100 to 150 mm ) 1.1 m

above 150 to 200 mm ) 1.4 m

above 200 to 250 mm ) 1.8 m

above 250 to 300 mm ) 2.1 m

above 300 to 400 mm ) 2.8 m

above 400 to 500 mm ) 3.5 m

Page 20: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 20 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

ii) Direct Measurement )

(For Resolution: 0.1 m) )

up to 10 mm ) 0.8 m

above 10 mm to 25 mm ) 0.9 m

above 25 mm to 50 mm ) 0.9 m

above 50 mm to 75 mm ) 1.6 m

above 75 mm to 100 mm ) 1.7 m

(For Resolution: 1 m) )

above 500 mm to 600 mm ) 12.0 m

above 600 mm to 700 mm ) 13.0 m

above 700 mm to 800 mm ) 13.0 m

above 800 mm to 900 mm ) 13.0 m

above 900 mm to 975 mm ) 13.0 m

16) Feeler Gauge (Lab) ) MTD/CAL-108 : 2012

up to 0.5 mm ) BS 957 : 2008 0.9 m

above 0.5 mm to 5 mm ) 1.1 m

17) Gauge Block (Lab) ) MTD/CAL-116 : 2012

Grade ‘0’ & below ) BS 4311 : 2007

(Steel & Tungsten Carbide Material) ) up to 10 mm ) 0.12 m

above 10 mm to 25 mm ) 0.13 m

above 25 mm to 50 mm ) MTD/CAL-116 : 2012 0.13 m

above 50 mm to 75 mm ) BS 4311 : 2007 0.14 m

above 75 mm to 100 mm ) 0.16 m

18) Height Setting Micrometer & ) MTD/CAL-120 : 2004 Raiser Block (Lab) ) ISO 7863: 1984 i) Flatness Measurement ) 0.4 m

ii) Parallelism Measurement ) 0.3 m

iii) Column Accuracy (up to 300 mm) ) 2.4 m

iv) Leadscrew Accuracy ) 1.2 m

v) Riser Block Accuracy (up to 300 mm) ) 2.4 m

Page 21: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 21 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

19) Universal Length Machine ) MTD/CAL-121 : 2012

(Lab/Site) )

)

Accuracy Measurement: )

i) For Resolution: 0.01 μm )

up to 10 mm ) 0.06 m

above 10 to 20 mm ) 0.13 m

above 20 to 50 mm ) 0.26 m

above 50 to 80 mm ) 0.39 m

above 80 to 100 mm ) 0.51 m

ii) For Resolution: 1 μm )

Up to 100 mm ) 1 m

Above 100 up to 200 mm ) 1 m

Above 200 up to 300 mm ) 2 m

Above 300 up to 400 mm ) 2 m

Above 400 up to 500 mm ) 3 m

Above 500 up to 800 mm ) 4 m

Above 800 up to 1 000 mm ) 5 m

20) Steel Rule (Lab) ) MTD/CAL-123 : 2010

i) Graduation Accuracy ) BS 4372 : 1968 0.006 mm

ii) Length Accuracy ) JIS B 7516 : 2005

0 to 600 mm ) DIN 868 : 1983 0.1 mm

0 to 1 000 mm 0.2 mm

21) Coating Thickness Gauge (Lab) ) MTD/CAL-124 : 2004

i) Foil Measurement (up to 7.8 mm) ) 0.4 m

ii) Linear Measurement (up to 1 000 m) ) 1.1 m

22) Welding Inspection Gauge (Lab) ) MTD/CAL-126:2004

i) Length Accuracy (Gauge Block) ) 116 m

(up to 25 mm)

ii) Length Accuracy (Profile Projector) ) 6 m

(up to 60 mm)

iii) Angular Accuracy ) 0.027 °

Page 22: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 22 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

23) Measuring Tape (Lab) ) MTD/CAL-131: 2010

Accuracy Measurement: )

0 - 1 m ) 0.2 mm

0 - 5 m ) 0.3 mm

0 - 10 m ) 0.5 mm

0 - 15 m ) 0.5 mm

0 - 20 m ) 0.6 mm

0 - 25 m ) 0.7 mm

0 - 30m ) 0.8 mm

0 - 35 m ) 0.8 mm

0 - 40 m ) 0.9 mm

0 - 45 m ) 0.9 mm

0 - 50 m ) 1.0 mm

24) Bore Gauge (Lab) ) MTD/CAL-132: 2010

Accuracy Measurement: )

20 - 150 mm ) 4 m

150 - 250 mm ) 5 m

250 - 300 mm ) 6 m

300 - 600 mm ) 7 m

25) Surface Plate ) MTD/CAL-128 : 2010

(up to 1 200 x 900 mm) ) BS 817 : 2008

i) Flatness ) DIN 876 Pt 1 & 2 1984 3.3 m

ii) Variation ) GGG-P- 463C 1973 1.3 m

26) Parallel Screw Thread (Lab) ) MTD/CAL-130: 2012

Generally as stated in FED )

Std H 28 : 1978 )

Thread Plug : M2 – M100 )

Pitch : 0.5 to 6 mm )

i) Pitch & Major Dia. : 1.2 m ) 1.0 m

ii) Pitch Measurement : 3.7 m ) 6 m

iii) Angular : 0.038° ) 0.027 °

)

Thread Ring : M3 – M200 )

Pitch : 0.5 to 6 mm )

i) Pitch Diameter : 3.9 m ) 6.0 m

ii) Minor Diameter : 3.7 m ) 6 m

Page 23: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 23 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

27) Plain Plug or Pin Gauges (Lab) ) MTD/CAL-117: 2012

up to 15 mm ) 1.5 m

above 15 mm to 70 mm ) 1.6 m

28) Plain Ring Gauges (Lab) ) MTD/CAL-117: 2012

up to 25 mm ) 0.9 m

above 25 mm to 30 mm ) 1.0 m

above 30 mm to 40 mm ) 1.4 m

above 40 mm to 100 mm ) 1.6 m

29) Dimensional Measureable (Lab) ) MTD/CAL-134: 2014

Circle & Radius : ≤ 50 mm ) Using Profile Projector

Angular : ≤ 360° ) Using Scope and Small

Surface & Line Distance : ≤ 180 mm ) Tools

i) Length Accuracy ) 6 m

ii) Angular Accuracy ) 0.027 °

30) Flatness Measurement (Lab) ) MTD/CAL-101: 2012 0.41 m

31) Parallelism Measurement (Lab) ) MTD/CAL-102: 2012 0.48 m

32) Gauge Block Comparator (Lab) ) MTD/CAL-118 : 2008

0.5 mm to 100 mm 0.05 m

33) Measurement by Coordinate ) MTD/CAL-135 : 2008

Measuring Machine (CMM) (Lab)

)

i) Accuracy: )

100 mm ) 2.8m

300 mm ) 3.7m

600 mm ) 16.2m

ii) Angular ) 0.0020 °

34) Cube Mould (Lab) ) MTD/CAL-122 : 2012

Up to 150 mm ) BS 1881 Part 108 : 1983

i) Accuracy (Depth) ) BS EN 12390-1 2000 4m

ii) Accuracy (Distance) ) SS 78 Part A8 : 1987 9m

35) Test Sieve (Lab) ) MTD/CAL-127 : 2010 6m

) BS 410 : 2000

) ISO 3310-1: 2000

) ISO 3310-2 : 1999

) ASTM E11: 2009

Page 24: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 24 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

36) NDT Instrumentation (Lab) ) MTD/CAL-125 : 2012

Ultrasonic Pulse Echo Instrument ) ASTM E317 : 2006a

Calibration Block ) AWS D1.1/D1.1M : 2010

i) Length Accuracy (up to 300 mm) ) BS EN 12223 : 2000 0.016 mm

ii) Angular Accuracy ) BS EN 27963 : 1992 0.040 °

37) Coordinate Measuring Machine ) MTD/CAL-136 : 2009

(CMM) Calibration )

i) Accuracy : 0 - 25 mm ) 0.57 m

0 - 100 mm ) 0.39 m

0 - 200 mm ) 0.37 m

0 - 300 mm ) 0.37 m

0 - 500 mm ) 0.63 m

0 - 700 mm 0.34 m

ii) Angular ) 0° 1’22”

38) Skid Resistance (Pendulum)

Tester (Lab)

i) Height Adjustment ) MTD/CAL-138: 2012 1 mm

ii) Distance of Center of Gravity of ) BS EN 13036-4: 2011 1 mm

Pendulum/ Arm to Center of ) ASTM E 303: 1993(2008)

Rotation )

iii) Point Length ) 1 mm

iv) Slider to Center Rotation ) 1 mm

v) Mass of Pendulum/ Arm ) See scope A12

vi) Pointer Weight ) See scope A12

vii) Weight Calibration ) See scope A12

viii) Measurement of Slide Force/ ) See scope A12

Deflection Characteristics )

39) Calibration using laser (Lab/Site) ) MTD/CAL-140: 2014

i) Up to 100 mm ) ASME B89.1.10M:2001 0.15µm

ii) Above 100 to 200 mm ) 0.7 µm

iii) Above 200 to 300 mm ) 1.4µm

iv) Above 300 to 500 mm ) 1.8µm

v) Above 500 to 800 mm ) 1.9µm

vi) Above 800 to 1 000 mm ) 1.5µm

Page 25: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 25 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

D. ELECTRICAL CALIBRATION (Lab)

1. DC Voltage Measuring Instrument ) MTD/Cal-204 : 2014

0 - 330 mV ) MTD/Cal-205 : 2014 0.008 mV

0.33 - 3.3 V ) MTD/Cal-206 : 2014 0.038 mV

3.3 - 33 V ) MTD/Cal-214 : 2012 0.42 mV

33 - 330 V ) 0.0061 V

330 - 1000 V ) 0.020 V

2. AC Voltage Measuring Instrument ) MTD/Cal-204 : 2014

) MTD/Cal-205 : 2014

) MTD/Cal-206 : 2014

) MTD/Cal-214 : 2012

1.0 mV - 33 mV 10 Hz - 45 Hz 0.032 mV

45 Hz - 10 kHz 0.011 mV

10 kHz - 20 kHz 0.013 mV

20 kHz - 50 kHz 0.039 mV

50 kHz - 100 kHz 0.13 mV

100 kHz - 500 kHz 0.31 mV

33 mV - 330 mV 10 Hz - 50 kHz 0.12 mV

50 kHz - 100 kHz 0.30 mV

100 kHz - 500 kHz 0.73 mV

0.33 V - 3.3 V 10 Hz - 100 kHz 0.0024 V

100 kHz - 500 kHz 0.0085 V

3.3 V - 33 V 10 Hz - 50 kHz 0.012 V

50 kHz - 100 kHz 0.031 V

33 V - 330 V 45 Hz - 50 kHz 0.11 V

50 kHz - 100 kHz 0.71 V

330 V - 1000 V 45 Hz - 10 kHz 0.4 V

3. DC Current Measuring Instrument ) MTD/Cal-204 : 2014

) MTD/Cal-205 : 2014

0 - 330 A ) MTD/Cal-206 : 2014 0.070 A

0 - 3.3 mA 0.00038 mA

3.3 - 33 mA 0.0036 mA

33 - 330 mA 0.036 mA

Page 26: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 26 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

0.033 - 1.1 A 0.00026 A

1.1 - 3 A 0.002 A

3 - 11 A 0.007 A

11 - 20.0 A 0.021 A

Clamp Meter Coil

Up to 800 A 6.1 A

4. AC Current Measuring Instrument ) MTD/Cal-204 : 2014

) MTD/Cal-205 : 2014

) MTD/Cal-206 : 2014

29 μA - 330 μA 10 Hz - 20 Hz 0.76 μA

20 Hz - 45 Hz 0.60 μA

45 Hz - 1 kHz 0.51 μA

1 kHz - 5 kHz 1.1 μA

5 kHz - 10 kHz 2.8 μA

10 kHz - 30 kHz 5.7 μA

0.33 mA - 3.3 mA 10 Hz - 20 Hz 0.0068 mA

20 Hz - 45 Hz 0.0043 mA

45 Hz - 1 kHz 0.0035 mA

1 kHz - 5 kHz 0.0068 mA

5 kHz - 10 kHz 0.017 mA

10 kHz - 30 kHz 0.034 mA

3.3 mA - 33 mA 10 Hz - 20 Hz 0.061 mA

20 Hz - 45 Hz 0.032 mA

45 Hz - 1 kHz 0.015 mA

1 kHz - 5 kHz 0.028 mA

5 kHz - 10 kHz 0.069 mA

10 kHz - 30 kHz 0.14 mA

33 mA - 330 mA 10 Hz - 20 Hz 0.61 mA

20 Hz - 45 Hz 0.32 mA

45 Hz - 1 kHz 0.15 mA

1 kHz - 5 kHz 0.38 mA

5 kHz - 10 kHz 0.76 mA

10 kHz - 30 kHz 1.5 mA

0.33 A - 1.1 A 10 Hz - 45 Hz 0.0021 A

45 Hz - 1 kHz 0.00065 A

1 kHz - 5 kHz 0.0076 A

5 kHz - 10 kHz 0.033 A

Page 27: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 27 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

1.1 A - 3 A 10 Hz - 45 Hz 0.006 A

45 Hz - 1 kHz 0.003 A

1 kHz - 5 k Hz 0.019 A

5 kHz - 10 kHz 0.080 A

3 A - 11 A 45 Hz - 100 Hz 0.009 A

100 Hz - 1 kHz 0.013 A

1 kHz - 5 kHz 0.33 A

11 A - 20 A 45 Hz - 100 Hz 0.030 A

100 Hz - 1 kHz 0.036 A

1 kHz - 5 kHz 0.62 A

Clamp Meter Coil

Up to 800 A 45 Hz - 65 Hz 6.5 A

65 Hz - 200 Hz 9.3 A

5. Resistance Measuring Instrument ) MTD/Cal-204 : 2014

0 Ω - 11 Ω ) MTD/Cal-205 : 2014 0.0014 Ω

11 Ω - 33 Ω ) MTD/Cal-206 : 2014 0.0025 Ω

33 Ω - 110 Ω ) 0.005 Ω

110 Ω - 330 Ω ) 0.011 Ω

330 Ω - 1.1 kΩ ) 0.000033 kΩ

1.1 kΩ - 3.3 kΩ ) 0.00011 kΩ

3.3 kΩ - 11 kΩ ) 0.00033 kΩ

11 kΩ - 33 kΩ ) 0.0011 kΩ

33 kΩ - 110 kΩ ) 0.004 kΩ

110 kΩ - 330 kΩ ) 0.013 kΩ

330 kΩ - 1.1 MΩ ) 0.00004 MΩ

1.1 MΩ - 3.3 MΩ ) 0.00023 MΩ

3.3 MΩ - 11 MΩ ) 0.0015 MΩ

11 MΩ - 33 MΩ ) 0.011 MΩ

33 MΩ - 110 MΩ ) 0.06 MΩ

110 MΩ - 330 MΩ ) 1.1 MΩ

330 MΩ - 1100 MΩ ) 17 MΩ

Page 28: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 28 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

6. Capacitance Measuring Instrument ) MTD/Cal-205 : 2014

) MTD/Cal-206 : 2014

0.19 nF - 0.40 nF ) 10 Hz to 10 kHz 0.012 nF

0.40 nF - 1.1 nF ) 10 Hz to 10 kHz 0.02 nF

1.1 nF - 3.3 nF ) 10 Hz to 3 kHz 0.03 nF

3.3 nF - 11 nF ) 10 Hz to 1 kHz 0.1 nF

11 nF - 33 nF ) 10 Hz to 1 kHz 0.2 nF

33 nF - 110 nF ) 10 Hz to 1 kHz 1.0 nF

110 nF - 330 nF ) 10 Hz to 1 kHz 2.0 nF

0.33 μF - 1.1 μF ) 10 Hz to 600 Hz 0.01 μF

1.1 μF - 3.3 μF ) 10 Hz to 300 Hz 0.02 μF

3.3 μF - 11 μF ) 10 Hz to 150 Hz 0.1 μF

11 μF - 33 μF ) 10 Hz to 120 Hz 0.2 μF

33 μF - 110 μF ) 10 Hz to 80 Hz 1.0 μF

110 μF - 330 μF ) 0 to 50 Hz 2.0 μF

0.33 mF - 1.1 mF ) 0 to 20 Hz 0.01 mF

1.1 mF - 3.3 mF ) 0 to 6 Hz 0.02 mF

3.3 mF - 11 mF ) 0 to 2 Hz 0.06 mF

11 mF - 33 mF ) 0 to 0.6 Hz 0.28 mF

33 mF - 50 mF ) 0 to 0.2 Hz 1.3 mF

7. Frequency Measuring Instrument ) MTD/Cal-205 : 2014

) MTD/Cal-206 : 2014

0.01 Hz - 120 Hz ) 0.001 Hz

120.0 Hz - 1200 Hz ) 0.01 Hz

1.200 kHz - 12 kHz ) 0.0001 kHz

12.00 kHz - 120 kHz ) 0.001 kHz

120.0 kHz - 1200 kHz ) 0.01 kHz

1.200 MHz - 2.000 MHz ) 0.00001 MHz

8. 8 Stopwatch Calibration ) MTD/Cal-211 : 2009

1 minute – 10 minutes ) 1 s

10 minutes – 24 hours ) 4 s

9. 8 Insulation Tester Calibration ) MTD/Cal-214 : 2012

Decade Resistance (Sourcing) )

0.1 Ω to 10 Ω ) 0.01 Ω

10 Ω to 100 Ω ) 0.1 Ω

100 Ω to 1 000 Ω ) 1 Ω

1.0 kΩ to 10 kΩ ) 0.01 kΩ

10 kΩ to 100 kΩ ) 0.1 kΩ

100 kΩ to 1 000 kΩ ) 1 kΩ

1.0 MΩ to 10 MΩ ) 0.01 MΩ

Page 29: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 29 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

10 MΩ to 100 MΩ ) 0.2 MΩ

100 MΩ to 1 000 MΩ ) 2 MΩ

1.0 GΩ to 10 GΩ ) 0.02 GΩ

10. 8 Sourcing Instrument ) MTD/Cal-215 : 2014

i) DC Voltage ) MTD/Cal-214 : 2012

0 mV to 120 mV 0.010 mV

0.12 mV to 1.2 V 0.00006 V

1.2 V to 12 V 0.0005 V

12 V to 120 V 0.006 V

119 V to 1 000 V 0.06 V

ii) AC Voltage ) MTD/Cal-215 : 2014

1 mV to 120 mV 5 Hz – 10 Hz 0.47 mV

10 Hz – 20 kHz 0.12 mV

20 kHz – 50 kHz 0.20 mV

50 kHz – 100 kHz 0.82 mV

100 kHz – 300 kHz 5.4 mV

0.11 V to 1.2 V 5 Hz – 10 Hz 0.0046 V

10 Hz – 20 kHz 0.0011 V

20 kHz – 50 kHz 0.002 V

50 kHz – 100 kHz 0.0082 V

100 kHz – 300 kHz 0.054 V

1.2 V to 12 V 5 Hz – 10 Hz 0.046 V

10 Hz – 20 kHz 0.011 V

20 kHz – 50 kHz 0.020 V

50 kHz – 100 kHz 0.082 V

100 kHz – 300 kHz 0.54 V

12 V to 120 V 10 Hz – 20 kHz 0.11 V

20 kHz – 50 kHz 0.20 V

50 kHz – 100 kHz 0.82 V

100 kHz – 300 kHz 5.4 V

120 V to 750 V 10 Hz – 20 kHz 0.68 V

20 kHz – 50 kHz 1.3 V

50 kHz – 100 kHz 5.1 V

Page 30: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 30 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

iii) DC Current ) MTD/Cal-215 : 2014

0 mA to 12 mA 0.0084 A

12 mA to 120 mA 0.066 A

0.12 A to 1.2 A 0.0013 A

1.2 A to 3A 0.0042 A

3 A to 20 A 0.12 A

iv) AC Current ) MTD/Cal-215 : 2014

0.004 A – 1.2 A 5 Hz – 10 Hz 0.0041 A

10 Hz – 5 kHz 0.0017 A

1.2 A – 3 A 5 Hz – 10 Hz 0.012 A

10 Hz – 5 kHz 0.007 A

3 A – 20 A 45 Hz – 66 Hz 0.060 A

v) Frequency ) MTD/Cal-215 : 2014

3 Hz to 10 Hz 0.005 Hz

10 Hz to 40 Hz 0.012 Hz

40 Hz to 300 kHz 0.03 kHz

vi) Resistance ) MTD/Cal-215 : 2014

0 Ω to 100 Ω 0.014 Ω 100 Ω to 1 kΩ 0.00011 kΩ 1 kΩ to 10 kΩ 0.0011 kΩ 10 kΩ to 100 kΩ 0.011 kΩ 0.1 MΩ to 1 MΩ 0.00011 MΩ 1.0 MΩ to 10 MΩ 0.0041 MΩ 10 MΩ to 100 MΩ 0.81 MΩ

vii) High Voltage Sourcing ) MTD/Cal-215 : 2014

DC Voltage 0.5 kV to 5 kV 0.03 kV 5 kV to 30 kV 0.18 kV

AC Voltage

0.5 kV to 5 kV 50 Hz 0.04 kV

5 kV to 28 kV 50 Hz 0.18 kV

DC Current

1 mA to 3 mA 0.01 mA

3 mA to 10 mA 0.1 mA

Page 31: LA-1993-0051-C - Setsco Services Pte Ltd...2) 50 to 1 000 mbar ) BS EN 837-1 & 3 : 1998 0.1 mbar 3) 3 to 30 psi ) DIN 16005 : 1987 0.003 psi 4) 30 to 100 psi ) 0.01 psi 5) 100 to 150

Certificate No. : LA-1993-0051-C Issue No. : 20 Date : 26 April 2016 Page : 31 of 31

MEASURED QUANTITIES / INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

AC Current

1 mA to 3 mA 50 Hz 0.01 mA

3 mA to 30 mA 50 Hz 1 mA

30 mA to 100 mA 50 Hz 1 mA

* CMC is expressed as an expanded uncertainty estimated at a level of confidence of approximately 95%.

Approved signatories :

Mr Juraimi Bin Sabit - Category B, C and D8 only.

Mr Chong Pin Siong - All except items C22-C26, C32-C33, C37, C38 and D.

Mr Neo Tien Sing - Category C except C32, C33, C37 and C38.

Mr Raj Kumar - Category A1 to A9, A12- A13, C29 and C38 only.

Mr Andy Yew Geok Ann - Category D1-D10 and A10-A11 only.

Ms Rosana R. Tolentino - Category C1-C3, C5-C20, C23-C24, C26-C31 and C35-C36 only.

Mr Teu Cihua - Category C1-C15, C18-C19, C22, C25, C29, C33, C35 to 37 only.

Mr Edwin Leong Heng Fatt - Category D1-D7 and D9-D10 only.

Mr James Song Lei Guang - Category B only.

Note :

This laboratory is accredited in accordance with the recognised International Standard ISO/IEC 17025:2005. A laboratory's fulfilment of the requirements of ISO/IEC 17025:2005 means the laboratory meets both the technical competence requirements and management system requirements that are necessary for it to consistently deliver technically valid calibration results. The management system requirements In ISO/IEC 17025:2005 (Section 4) are written in language relevant to laboratory operations and meet the principles of ISO 9001:2008 Quality Management Systems — Requirements and are aligned with its pertinent requirements.