L 7 ct physics

23
INTRODUCTION TO CT PHYSICS Shahid Younas Lecture 7

Transcript of L 7 ct physics

INTRODUCTION TO CT PHYSICS

Shahid Younas Lecture 7

Slice Thickness: Multiple Detector Array Scanners

Shahid Younas Lecture 7

The slice thickness of multiple detector array CT scanners is determined not by the collimation but by the width of the detectors in the slice thickness dimension.

Slice Thickness: Multiple Detector Array Scanners

Shahid Younas Lecture 7

Slice Thickness: Multiple Detector Array Scanners

Shahid Younas Lecture 7

The width of the detectors is changed by binning different

numbers of individual detector elements together.

The electronic signals generated by adjacent detector

elements are electronically summed.

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

Pitch is a parameter that comes to play when helical scan protocols are used.

In a helical CT scanner with one detector array, the pitch is determined by the collimator (collimator pitch).

Single and Multiple detector arrays scanner have different definitions.

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

In a helical CT scanner with one detector array, the pitch is

determined by the collimator (collimator pitch),

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

Pitch fundamentally influences radiation dose to the patient,

image quality, and scan time.

For single detector array scanners, a pitch of 1.0 implies that

the number of CT views acquired, when averaged over the

long axis of the patient, is comparable to the number acquired

with adjoining axial CT.

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

A pitch of less than 1.0 involves over-scanning may result in

some slight improvement in image quality and a higher

radiation dose to the patient.

pitches greater than 1.0, and pitches up to 1.5 are commonly

used.

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

Collimator Pitch = T / C

Detector width = T / D

For a multiple detector array CT scanner with four detector arrays, a collimator pitch of 1.0 is equal to a detector pitch of 4.0.

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

Scanners that have multiple detector arrays require a different

definition of pitch.

N = Number of detectors used

Detector Pitch & Collimator Pitch

Shahid Younas Lecture 7

For scanners with four detector arrays, detector pitches running

from 3 to 6 are used. A detector pitch of 3 for a four-detector

array scanner is equivalent to a collimator pitch of 0.75 (3/4), and

a detector pitch of 6 corresponds to a collimator pitch of 1.5

(6/4).

CT Number or Hounsfield Units

Shahid Younas Lecture 7

Each pixel in the image is represented by a high-precision

floating point number that is useful for computation but less

useful for display.

Consequently, after CT reconstruction, but before storing and

displaying, CT images are normalized and truncated to integer

values.

CT Number or Hounsfield Units

Shahid Younas Lecture 7

where µ(x,y) is the floating point number of the (x,y) pixel before conversion

CT numbers are quantitative, and this property leads to more accurate diagnosis in some clinical settings.

CT Dose Measurement

Shahid Younas Lecture 7

Compton scattering is the principal interaction mechanism

in CT, so the radiation dose attributable to scattered

radiation is considerable, and it can be higher than the

radiation dose from the primary beam.

The multiple scan average dose (MSAD) is the standard

for determining radiation dose in CT.

CT Dose Measurement Methods

Shahid Younas Lecture 7

CT Dose Measurement

Shahid Younas Lecture 7

The MSAD is the dose to tissue that includes the dose

attributable to scattered radiation emanating from all

adjacent slices.

The MSAD is defined as the average dose, at a particular

depth from the surface, resulting from a large series of CT

slices.

CT Dose Measurement

Shahid Younas Lecture 7

Anther way is the CT Dose Index (CTDI) to estimate MSAD.

The CTDI is defined by the U.S. Food and Drug Agency as

the radiation dose to any point in the patient including the

scattered radiation contribution from 7 CT slices in both

directions, for a total of 14 slices.

CT Dose Measurement Methods

Shahid Younas Lecture 7

TLD Dosimeters

Pencil Ionization

Chambers

CT Dose Measurement Methods

Same as previous but a correction factor is needed,

CT Artifacts

Shahid Younas Lecture 7

1. Beam Hardening 2: Motion Artifacts

CT Spatial Resolution Affecting Factors

1. Detector pitch

2. Detector aperture

3. Number of views

4. Number of rays

5. Focal spot size

6. Object magnification:

7. Slice thickness

8. Slice sensitivity Profile

9. Helical Pitch

10. Reconstruction kernel

11. Pixel matrix

12. Field of View

Practical Sessions for Medical Physics

Dose Calibrator and sealed sources used in Nuclear Medicine

March, 10, 20144 to 5 pm

Hot Lab of Nuclear Medicine

Future Books