Knobology 01 2010 (2)

41
Ultrasound machine Ultrasound machine knobology knobology Copyright © 2010 Academy of Veterinary Imaging

Transcript of Knobology 01 2010 (2)

Page 1: Knobology 01 2010 (2)

Ultrasound machine Ultrasound machine knobologyknobology

Copyright © 2010 Academy of Veterinary Imaging

Page 2: Knobology 01 2010 (2)

Ultrasound machine knobology…Ultrasound machine knobology…

In other words, what the heck do all of those knobs and buttons do, and how can adjusting

them affect image quality?!?!

Page 3: Knobology 01 2010 (2)

• Image optimization– Presets, depth, frequency, auto-optimize, focal

points, gain/TGC

• Features– Archiving, transducers, SRI, compounding,

harmonics, 3-D, LOGIQView, Doppler

GE ultrasoundGE ultrasound

Page 4: Knobology 01 2010 (2)

GE ultrasoundGE ultrasound

LOGIQ 7

LOGIQ Book

Go to:Image optimization:

Presets

Depth

Frequency

Auto-optimize

Focal points

Gain/TGC

LOGIQ e

LOGIQ 5

Archiving

Transducers

SRI

CrossXBeam

Harmonics

3-D

LOGIQView

% affected

GE features:

LOGIQ P5

In this presentation, click on the Home button as pictured here to return to this page

Page 5: Knobology 01 2010 (2)

Image optimizationImage optimization

• Image quality is comprised of many things• During an ultrasound exam, the ultrasound machine settings should be changed as needed to optimize image quality• Understanding just a few of the knobs can accomplish this for most scans

L7

L5

LB

Le

P5

Page 6: Knobology 01 2010 (2)

PresetsPresets

• Settings can be saved as presets• Viewed in Application or Preset

screen• Different presets for different

transducers/structures/patient species/sizes

• The more ‘complicated’ machine settings are saved in presets, so these do not need to be adjusted during every scan

L7

L5

LB

P5

Le

Page 7: Knobology 01 2010 (2)

DepthDepth

• Adjust as needed throughout scan

Optimal

L7

L5

LB

P5

Le

Page 8: Knobology 01 2010 (2)

FrequencyFrequency

• High frequency• Better resolution• Less depth

• Lower frequency• Lower resolution • Better for depth

L7

L5

LB

P5

Le

Page 9: Knobology 01 2010 (2)

Auto optimizeAuto optimize

• ‘Personal preference’ button.• Adds more contrast…Whites are

whiter, blacks are blacker• Try on or off. Chose the setting

that looks better to you. • Auto optimize must be deselected

then reselected for every new structure imaged.

L7

L5

LB

P5

Le

Page 10: Knobology 01 2010 (2)

Focal pointsFocal points

• Improved resolution is seen where the focal points are.• Place focal points at the region of interest.

L7

L5

LB

P5

Le

Page 11: Knobology 01 2010 (2)

GainGain

• Overall gain– Amplification of received signal

• TGC– Time Gain Compensation– Compensates for beam

attenuation (loss of energy as it passes through tissue)

L7

L5

LB

P5

Le

Page 12: Knobology 01 2010 (2)

LOGIQ 7LOGIQ 7

TGC

Auto optimize

Gain

Depth

Frequency

Focal points

(touch screen)

L5

LB

Le

P5

Page 13: Knobology 01 2010 (2)

LOGIQ 5LOGIQ 5

TGCAuto

optimize

Gain

Depth

Focal points

Frequency

L7

LB

Le

P5

Page 14: Knobology 01 2010 (2)

LOGIQ P5LOGIQ P5

Auto optimizeGain

DepthTGC

Focal points

Frequency

L7

L5

LB

Le

Page 15: Knobology 01 2010 (2)

LOGIQ eLOGIQ e

Auto optimize

Gain

Depth

TGC

Focal points

Frequency

L7

L5

LB

P5

Page 16: Knobology 01 2010 (2)

LOGIQ BookLOGIQ Book

Auto optimizeGain

Depth

TGC

Focal points

Frequency

(toggle)

L7

L5

Le

P5

Page 17: Knobology 01 2010 (2)

Archiving Archiving

• ‘P’ buttons can be programmed

• Default settings:– P1: Saves as raw dicom clip or

still image (image appearance, M-mode, 3-D can be done on a raw dicom clip, but it is a very large file, thus too large to send)

– P2: Print on the thermal printer

– P3: Save as dicom

Page 18: Knobology 01 2010 (2)

Archiving Archiving

• Images can be transferred to a CD or USB flash drive.

• Depending on GE model, media must be formatted first– Press Utilities button – Select Connectivity, then removable media

• To eject media after saving images, press F3– Select the media to eject, then finalize

Page 19: Knobology 01 2010 (2)

Some GE transducersSome GE transducers

8C 12L

4C

3S

Rectal

i739

Page 20: Knobology 01 2010 (2)

SRISRI

• Speckle Reduction Imaging:– Improved contrast resolution– Improved overall clarity– Smoothes regions where no feature or edges

appear– Maintains or enhances edges and borders– Does not eliminate any information– Increased Signal-to-Noise ratio

Page 21: Knobology 01 2010 (2)

SRISRI

• Always is on in the LOGIQ e• Can adjust to have different levels in the

LOGIQ 7 and LOGIQ P5– Level is set within the preset (Application)

settings

Page 22: Knobology 01 2010 (2)

Spatial compoundingSpatial compounding

• CrossXBeam – = GE’s spatial compounding– Available on the LOGIQ e and LOGIQ P5

• Interrogates tissue at multiple angles• Introduces new tissue info• Averages out speckle• SRI and CrossXBeam together can have

complimentary and additive effects

Page 23: Knobology 01 2010 (2)

CrossXBeamCrossXBeam

• Perpendicular angle of insonation generates higher signal

• Images generated with compounding have more angles that are perpendicular to structures

CrossXBeam OFF CrossXBeam ON

Page 24: Knobology 01 2010 (2)

CrossXBeamCrossXBeam

• Improved border definition• Improved contrast resolution• Reduced angle dependence • Reduced speckle and noise• Increased visualization of biopsy needles• Can reduce helpful artifact • Can reduce frame rate

LOGIQ eLOGIQ P5

Page 25: Knobology 01 2010 (2)

CrossXBeam: P5CrossXBeam: P5

Press L and R buttons simultaneously to see side-by-side display of CrossXBeam on and CrossXBeam off

Press X Beam to turn on

Page 26: Knobology 01 2010 (2)

CrossXBeam: LeCrossXBeam: Le

F9 or ‘Comp’ to turn on

Press L and R buttons simultaneously to see side-by-side display of CrossXBeam on and CrossXBeam off

Page 27: Knobology 01 2010 (2)

Harmonics (THI)Harmonics (THI)

Display is from echos of a frequency different from the transmitted frequency (i.e. 2x transmitted frequency)

•Reduced artifact

•Improved axial and lateral resolution

•Better margin definition and increased contrast

Page 28: Knobology 01 2010 (2)

Harmonics (THI)Harmonics (THI)

Press L and R buttons simultaneously to see side-by-side display of harmonics on and harmonics off

Press THI or Harmonics button to turn on

Page 29: Knobology 01 2010 (2)

3-D3-D

• Volume imaging– Volume vs. area affected can be assessed

• Appears more ‘life-like’• Infinite viewing planes• May approximate CT or MRI imaging• Scan parallel* or sweep (fan)• Angle of imaging less operator dependent

LOGIQ eLOGIQ P5

Page 30: Knobology 01 2010 (2)

3-D: P53-D: P5

Press Left button to start. Scan anatomy. Press Right button to stop.

Press 3D to turn on

Click arrow to go to images

Page 31: Knobology 01 2010 (2)

3-D: Le3-D: Le

Press F7 to turn on

Press Left button to start. Scan anatomy. Press Right button to stop.

Click arrow to go to images

Page 32: Knobology 01 2010 (2)

3-D3-DUse set button to alternate

hand and rotate tools

Drag image to corner

Page 33: Knobology 01 2010 (2)

3-D3-D

Rotate image

Use set button to alternate hand and rotate tools

Page 34: Knobology 01 2010 (2)

3-D3-D

Slice through

Use set button to alternate hand and rotate tools

Page 35: Knobology 01 2010 (2)

3-D3-D

Rotate again

Use set button to alternate hand and rotate tools

Page 36: Knobology 01 2010 (2)

3-D3-D

Slice through another plane

Use set button to alternate hand and rotate tools

Page 37: Knobology 01 2010 (2)

LOGIQViewLOGIQView

• Extended field of view• Better demonstration of anatomical

relationships• Ability to measure large structures

LOGIQ e

LOGIQ P5

Page 38: Knobology 01 2010 (2)

LOGIQView: P5LOGIQView: P5

Press LOGIQ View to turn on

Press Left button to start, scan anatomy. Press Right button to stop.

Page 39: Knobology 01 2010 (2)

LOGIQView: LeLOGIQView: Le

Press F8 View to turn on

Press Left button to start, scan anatomy. Press Right button to stop.

Page 40: Knobology 01 2010 (2)

LOGIQViewLOGIQView

Scan in a straight plane or image will ‘bend’

Page 41: Knobology 01 2010 (2)

% affected% affected

• Measure % of tendon affected by the lesion– Press Measure button– Select % stenosis folder– Select % sten (area)– Trace around entire tendon– Press set– Trace around lesion– Press set– % affected = 100 - %sten