Please refer to the AimPlex assay user manual for details of the assay procedure.
Kit Contents Equipment Assay Procedure – Heat Lysis
Transcript of Kit Contents Equipment Assay Procedure – Heat Lysis
95°CLysis Block
WorkflowTray
TransferRack
Kit Contents Equipment
GASReaction
Tubes
C/GReaction
Tubes
DilutionBufferTube
LysisBufferTube
PCM305100EN00 (04/17)
Amplification and Detection
SolanaInstrument
Transfer 50 µL to each of the GAS and C/G
Reaction Tubes.
1
Rehydrate each Reaction Tube by pipetting up
and down 5 times.
2 43
Close lids tightlyand proceedto next step.
65 8
Close the lid and selectappropriate assay protocol.
Run complete in 25 minutes.
Review results.Lower the Reaction Tubesinto Solana.
Release the Reaction Tubesfrom the Transfer Rack bypulling the Transfer Rack
toward your body.
Procedure Card
Strep Complete AssayUsing ESwabTM Collection Device
Study the Package Insert thoroughly before using this Procedure Card. This is not a complete Package Insert.
7
Use the Transfer Rack to liftReaction Tubes from the
Workflow Tray.
ESwab
Assay Procedure – Heat Lysis
VORTEX5 seconds
VORTEX5 seconds
Heat for 5 minutesat 95°C.
Add 50 µL to Dilution
Buffer Tube.
1 3
Add 50 µL to
Lysis Tube.
42 5 6
Setup the Workflow Tray withpatient samples, Lysis BufferTubes, Dilution Buffer Tubes,
and Reaction Tubes.
7 8
VORTEX5 seconds