kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V...

9
PosterREPRINT Introduction LC/MS applications are filling an analysis gap and making valuable contributions to the clinical and toxicology laboratory. We will demonstrate how LC/MS has become crucial for our routine operations. As an example, an LC/MS assay that monitors compliance of drug rehabilitation patients using a transdermal patch, administering Fentanyl, will be shown. Clinical/toxicological regulatory guidelines have started to address the appearance of the recent introduction of LC/MS and LC/MS/MS instrumentation into a routine analysis environment. Compound identification criteria for single quadrupoles based on ion ratios must make use of at least two ratios or one ratio of two characteristic ions if few ions exist. The unknowns must have ion ratios within ±20% of the extracted standards. The internal standard identification must be monitored by at least one ion ratio. In addition, selected ion monitoring (SIM) spectra identification must be based on retention time. These and other acceptance parameters require specialized quantitation and reporting features to be available within the mass spectrometry processing software. We will present an LC/MS method that allows the use of multiple fragment ions for confident confirmation (as many as five ions). Analytical results for Fentanyl and its metabolite, Norfentanyl, are presented. The identification of each analyte is based on the routine assessment of one precursor ion and two product ions formed by in-source collision-induced dissociation (CID). Methods and Instrumentation Urine samples are from patients being tested for mandatory drug compliance. Standards and two quality control samples were prepared by dilution of stock standards stored in methanol into pooled, drug-free urine. Five standard levels, 10, 50, 100, 200, 500 and 1000 ng/ml were used to generate a calibration curve. Method development included the use of Norfentanyl-d5, Normeperidine-d4 as internal standards. External standard quantitation was also performed. The drugs of interest were extracted by solid phase extraction (SPE), using an Oasis ® 1cc/30mg MCX cartridge, prior to analysis. Two LC/MS methods were developed. An initial HPLC method, using an existing amphetamine analysis template, was developed for Fentanyl/Norfentanyl. This 25-minute method used a pH tolerant, hybrid Waters ® XTerra ® MS C 18 , 2.1x100mm, 3.5μ column. Mobile phases were 20mM ammonium bicarbonate buffer, pH 9.6(A) and methanol(B). The second method was developed to allow a desirable analyte selectivity change, shortened analysis time ( 8 minutes) and additional opiate family members to be incorporated into the analysis. A Waters Atlantis dC 18 IS (Intelligent Speed ), 2.1x20mm, 3μ shows promise. The mobile phases for the 20mm long column were 10mM ammonium formate buffer, pH 3.7(A) and methanol(B). Eight ions were monitored for each MS method. See Figures A and B for details. Detection was accomplished with a single quadrupole MS detector. The ionization mode was positive ESI. Data collection used SIM for quantitation and full scans to monitor for other compounds of interest. MassLynx 4.0 Software controlled the LC/MS system. TargetLynx , a MassLynx Application Manager, accomplished multi-ion quantitative confirmation. Robert Franey 1 , Kim Lilley 2 , Kate Yu 2 1 Franey Medical Labs, 8 Jan Sebastion Way, Sandwich, MA 02563, 2 Waters Corporation, 34 Maple Street, Milford, MA 01757 Presented at ASMS, Montréal, 8th-12th June 2003 QUANTIFICATION OF FENTANYL AND NORFENTANYL IN URINE BY LC/MS-ESI WITH MULTI-ION CONFIRMATION

Transcript of kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V...

Page 1: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

Introduction

LC/MS applications are filling an analysis gap andmaking valuable contributions to the clinical andtoxicology laboratory. We will demonstrate howLC/MS has become crucial for our routine operations.As an example, an LC/MS assay that monitorscompliance of drug rehabilitation patients using atransdermal patch, administering Fentanyl, will beshown.

Clinical/toxicological regulatory guidelines havestarted to address the appearance of the recentintroduction of LC/MS and LC/MS/MS instrumentationinto a routine analysis environment. Compoundidentification criteria for single quadrupoles based onion ratios must make use of at least two ratios or oneratio of two characteristic ions if few ions exist. Theunknowns must have ion ratios within ±20% of theextracted standards. The internal standardidentification must be monitored by at least one ionratio. In addition, selected ion monitoring (SIM) spectraidentification must be based on retention time. Theseand other acceptance parameters require specializedquantitation and reporting features to be availablewithin the mass spectrometry processing software.

We will present an LC/MS method that allows the useof multiple fragment ions for confident confirmation (asmany as five ions). Analytical results for Fentanyl andits metabolite, Norfentanyl, are presented. Theidentification of each analyte is based on the routineassessment of one precursor ion and two product ionsformed by in-source collision-induced dissociation (CID).

Methods and Instrumentation

Urine samples are from patients being tested formandatory drug compliance.

Standards and two quality control samples wereprepared by dilution of stock standards stored inmethanol into pooled, drug-free urine. Five standardlevels, 10, 50, 100, 200, 500 and 1000 ng/ml wereused to generate a calibration curve.

Method development included the use ofNorfentanyl-d5, Normeperidine-d4 as internalstandards. External standard quantitation was alsoperformed.

The drugs of interest were extracted by solid phaseextraction (SPE), using an Oasis® 1cc/30mg MCXcartridge, prior to analysis.

Two LC/MS methods were developed. An initialHPLC method, using an existing amphetamineanalysis template, was developed forFentanyl/Norfentanyl. This 25-minute method useda pH tolerant, hybrid Waters® XTerra® MS C18,2.1x100mm, 3.5µ column. Mobile phases were20mM ammonium bicarbonate buffer, pH 9.6(A)and methanol(B). The second method wasdeveloped to allow a desirable analyte selectivitychange, shortened analysis time ( 8 minutes) andadditional opiate family members to beincorporated into the analysis. A Waters Atlantis™

dC18 IS (Intelligent Speed™), 2.1x20mm, 3µ showspromise. The mobile phases for the 20mm longcolumn were 10mM ammonium formate buffer, pH3.7(A) and methanol(B). Eight ions were monitoredfor each MS method. See Figures A and B fordetails.

Detection was accomplished with a singlequadrupole MS detector. The ionization mode waspositive ESI. Data collection used SIM forquantitation and full scans to monitor for othercompounds of interest.

MassLynx™ 4.0 Software controlled the LC/MSsystem. TargetLynx™, a MassLynx ApplicationManager, accomplished multi-ion quantitativeconfirmation.

Robert Franey1, Kim Lilley2, Kate Yu2

1Franey Medical Labs, 8 Jan Sebastion Way, Sandwich, MA 02563, 2Waters Corporation, 34 Maple Street, Milford, MA 01757

Presented at ASMS, Montréal, 8th-12th June 2003

QUANTIFICATION OF FENTANYL ANDNORFENTANYL IN URINE BY LC/MS-ESI

WITH MULTI-ION CONFIRMATION

Page 2: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

Flow Injection Analysis (FIA): Fentanyl CID Fragments

FIA: Norfentanyl CID Fragments

College of American Pathologists (CAP)Guidelines: Chemistry/Toxicology – March 2003

• Are the identification criteria for single stagemass spectrometry (i.e., GC/MS, LC/MS) incompliance with recommendations?

NOTE: An acceptable criterion for compoundidentification using ion ratios is that the unknownresult must have ion ratios within ± 20% of theextracted calibrator(s). Identification using ionratios typically requires the use of at least 2 ionratios. However, one ion ratio of 2 characteristicions may be acceptable if there are only a fewcharacteristic ratios. The internal standard'sidentification should be monitored with at leastone ion ratio. An acceptable criterion forcompound identification using total spectra is thatthe unknown result must have a "spectral match"quality or fit that is within the defined limits thatthe laboratory has set and validated. Ion ratiosdetermined from total spectra analysis are anacceptable identification method, and should fulfillthe same criteria as given above for ion ratioidentifications.

COMMENTARY:

Mass spectral identification criteria must complywith recommendations.

• When applicable, are TIC identification criteriabased on retention time and MS Library?

COMMENTARY:

The mass spectral identification criteria must beconsistent with retention times and library.

Are SIM spectra identification criteria based onretention time?

COMMENTARY:

SIM spectra identification criteria must be basedon retention time.

80 100 120 140 160 180 200 220 240 260 280 300 320 340 360m/z0

100

%

0

100

%

0

100

%

0

100

%

337.4

338.4

337.4

188.2338.4

188.2

104.8 134.1

337.4

216.3189.5 338.4

188.2

104.9

83.9 134.1 146.2 337.4216.3189.5

Cone=20V

Cone=30V

Cone=40V

Cone=50V

Precursor and product ions were determined byFIA of the stock standards. The MS method wascreated using this information.

FIA: Norfentanyl CID fragments

80 100 120 140 160 180 200 220 240 260m/z0

100%

0

100%

0

100%

0

100%

0

100%

233.4

83.8 234.5

83.8

233.4177.3150.2

83.7

233.3150.2 177.3

83.8

94.0 150.2

94.083.8

93.3150.1132.2120.0104.8

Cone=20V

Cone=30V

Cone=40V

Cone=60V

Cone=70V

Two CID fragments, 150.2 and 177.3 m/z, areeasily detected at a cone voltage that would beappropriate for other drug candidates that wehave an interest in monitoring. 94 m/z is alsoavailable for use in the method. While 83.8 m/zappears to be an attractive choice to monitor, theinternal standard, Norfentanyl-d5, shares this CIDfragment as can be seen in the following slide.

Page 3: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

CAP Guidelines: Chemistry/Toxicology –March 2003

• For LC/MS, does the laboratory's assayprocedure include an evaluation for possibleion-suppression?

COMMENTARY:Ion suppression is a recognized analyticalanomaly in LC/MS experiments. Such suppressioncan lead to false negative results or poorquantitative analyses. While difficult to predictand observe from specimen to specimen, certainprecautions should be used to try to recognizewhen ion suppression occurs. As an example, forisotopically-labeled internal standards, if there ispoor recovery of the internal standard, a signal tonoise ratio greater than 3:1 should still suffice foracceptance of the specimen in question. Ifrecovery of the isotopically-labeled internalstandard is considered poor, then an alternateanalysis should be considered, e.g., the method ofstandard addition. For analogue-type internalstandards, internal standard recovery may be usedas a guide for identification of ion suppression,although the method of standard addition wouldbe better. For unique matrices, the method ofstandard addition is recommended.

FIA: Norfentanyl-d5 CID Fragments (Initial IS)

Fentanyl LC/MS Method

The LC/MS method was successfully implementedin the laboratory. Interference from other matrixcomponents was not noted during the period thatthe method was developed.

MS: Waters Micromass® ZQ™ MassSpectrometer (Initial Method)

Ionization: ESI+

Source (C°): 125°C

Desolvation (C°): 300°C

Capillary Voltage: 3.0kV

Data Acquisition: SIM(Centroid Scan 80 to 475m/z: Sample Review)

80 100 120 140 160 180 200 220 240 260m/z0

100

%

0

100

%

0

100

%

0

100

%

238.483.8

182.3239.4

83.7

155.2

83.8

98.8

155.2125.1

98.8

83.7

84.3 109.5 137.3113.3

Cone=25V

Cone=45V

Cone=65V

Cone=85V

The identification of the internal standard ismonitored with one CID fragment from thoseavailable.

y

1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00Time0

100

%

0

100

%

0

100

%

0

100

%

0

100

%

0

100

%

9.29

11.00

9.29

11.00

9.29

11.00

10.92

9.29

9.29 10.84

9.29

10.92

NorfentanylFentanyl

Patient JC

Patient MO

Patient JQ

100 ng/ml Std

50 ng/ml Std

10 ng/ml Std

Page 4: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

Initial Method

HPLC Waters Alliance® 2695 Separations Module

Column: Waters XTerra MS C18, 2.1x100mm, 3.5µ

Column (C°): 40 °C

Mobile Phase: A. 20mM NH4HCO3 in Water, pH 9.6

B. Methanol

Flow Rate: 0.25 ml/min

Inject volume: 15µl

Gradient:

Time A% B% Curve

0.0 85 15 1

3.0 25 75 6

14.0 25 75 6

15.0 10 90 6

18.0 10 90 6

20.0 85 15 6

25.0 85 15 6

Sample Preparation: SPE

Oasis MCX (cation-exchange and RP sorbent)

TargetLynx Overview

Mass spectrometry is an analytical tool that hasproven utility in the identification and quantificationof compounds. The spectral information suppliedduring an acquisition can indicate the identity of ananalyte. In addition to the precursor m/z obtained,other information such as product ion m/z, production relative abundance and chromatographicretention time provide invaluable criteria to enablethe confident identification of a compound.

The importance of developing a robust analyticalmethod has a special emphasis placed upon itwhen the results are used in areas such as humanhealth, environmental monitoring, and foodanalysis. Complex matrices necessitate samplepreparation procedures. For complex cleanupstrategies, recovery may vary. During analysis othersample components may be present that disturbchromatographic retention time. Compounds mayexist in greater abundance and suppress detectionof the analyte of interest. Monitoring theperformance and cleanliness of the analyticalsystem for each assay is a concern. In addition tothese issues, there are other sources of variation thatrequire ongoing review. TargetLynx is a specializedMassLynx Application Manager that provides toolsthat aid in the identification and quantificationprocess. An assortment of routine tasks isautomatically executed within the quantitationmethod.

• A product ion and up to four precursor ions canbe monitored for identification and quantificationof an analyte.

• If targeted ions are determined to be present, theyare integrated. The abundance ratios between thetarget ions must fall within the method’s definedlimits or the sample is flagged.

Load: 1 ml urine on 1cc/30mg cartdridge

Wash1: 1 ml 0.1N HCL

Wash2: 1 ml Methanol

Wash3: 1 ml 60/40 Methanol:2%Acetic

Extract: 1.0 ml 90/8/2 Methanol:H2O:NH4OH

Inject 15ul

Precondition cartridge: 1 ml MEOH / 1ml Water

Page 5: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

• Retention time, blank response, S/N ratio andcoefficient of determination are additionalacceptance criteria. Parameters outside thedesignated limits will be flagged in the report.

• LOD and LOQ requirements for target analytesmust be met or the results will be flagged.

• QC samples are monitored.

TargetLynx Report: Ion Ratio Flags

MS: Waters Micromass ZQ Mass Spectrometer(Shortened Analysis)

Ionization: ESI+

Source (C°): 125 °C

Desolvation (C°): 300 °C

Capillary Voltage: 3.0kV

Data Acquisition: SIM (Selected Ion Monitoring)Centroid Scan: Sample Review)

Shortened Analysis Method

HPLC: Waters Alliance 2695 Separations Module

Column: Waters Atlantis dC18 IS, 2.1x20mm, 3m

Column (C°): 45 °C

Mobile Phase: A. 10mM NH4Formate in Water, pH 3.7

B. Methanol

Flow Rate: 1.0 mL/min, 1.25 mL/min at 6-8 minutes

Inject Volume: 12 µL

Gradient:

Time A% B% Curve

.0 100 0 1

1.0 100 0 6

3.0 75 25 6

4.0 60 40 6

5.0 0 100 6

6.0 0 100 6

6.5 100 0 6

8.0 100 0 6

Two Analysis Goals: Confirmation andQuantitation

• Detect illegal drug use

• Monitor renewed episodes of drug use

• Confirm drug therapy regimens followed

• Useful indicator of effective metabolism withmultiple drug regimens

• Metabolites monitored as test of compliance

Page 6: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

TargetLynx Method: Targeting Properties

• Possible to "target" four ions

• Ratio and Tolerance monitored

• Reporting method uses Acceptance Parameters(S/N, retention time, ion ratio) to detect andreport compound to fulfill regulatory guidelines

TargetLynx Report: Reviewing Results Aided byPop-Up Summary

Patient X: Fentanyl/Norfentanyl Levels over the course of six weeks therapy.

Date Fent (ng/ml) Norfent (ng/ml) Comments Date Fent (ng/ml) Norfent (ng/ml) Comments

3/20 42 293 100 ug/h 4/16 261 548 100 ug/h

3/21 114 279 100 ug/h 4/17 129 499 100 ug/h

3/24 27 336 100 ug/h 4/18 130 362 100 ug/h

3/28 77 491 100 ug/h 4/19 41 516 100 ug/h

3/29 31 412 100 ug/h 4/21 76 596 100 ug/h

3/31 52 410 100 ug/h 4/22 267 538 100 ug/h

4/1 32 73 100 ug/h 4/23 36 320 100 ug/h

4/3 40 44 clarithromycin 4/24 11 113 100 ug/h

4/4 151 60 (Biaxin) 4/26 34 816 100 ug/h

4/5 13 31 therapy 4/28 59 479 100 ug/h

4/8 164 55 “””” 4/29 151 260 100 ug/h

4/9 224 69 “””” 4/30 68 740 100 ug/h

4/10 118 63 “””” 5/1 4 210 100 ug/h

4/11 318 163 “””” 5/2 38 587 100 ug/h

4/12 70 243 100 ug/h 5/3 164 393 100 ug/h

4/14 211 824 100 ug/h 5/6 178 532 100 ug/h

Expected initial serum response: Level off 12-24 hours. Relatively constant 24-72 hours. Several sequential applications: Individually determined steady state.

Page 7: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

Investigation: Ratios Fail Acceptance Parameters

Library Search: Trimethoprim Metab

Opiate Method Optimization: ShortenedAnalysis and Selectivity Change

1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00Time0

100

%

0

100

%

0

100

%

9.29

10.92

11.00

7.91

5.474.08

12.22

11.25

9.21?

FentanylNorfentanyl

Norfentanyl

Fentanyl

Patient "TMP": TIC

Patient "TMP": 337.2+233.2 m/z

200 ng/ml Std

Interference?

Investigation of the full scan data from the patient that failed the TargetLynx acceptance

Investigation of the full scan data, from the patientthat failed the TargetLynx acceptance parameters,showed an unknown compound eluting in theretention time window for norfentanyl.

Library Search: Trimethoprim metab

Library file

Interfering Cmpd

A user built library quickly identifies the unknown as a metabolite of trimethoprim.

A user-built library quickly identifies the unknownas a metabolite of trimethoprim.

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00Time0

100

%

0

100

%

0

100

%

0

100

%

0

100

%

3.224.45

3.21

4.33

4.32

5.054.954.34

3.643.21 3.89

4.885.73

5.37 5.84

TMP patient:234.2m/z

TMP patient:177.2m/z

TMP patient:233.2m/z

FentanylNorF

TIC: TMP patient

Standard

NorF

QC: External Standard MethodFentanyl/Norfentanyl Quality Control

Date Fent Norfent Date Fent Norfent (ng/ml) (ng/ml) (ng/ml) (ng/ml)

Level 1 Level 1 Level 2 Level 2

3/24 118 300 3/24 16 102

3/24 94 253 3/24 14 93

3/25 115 322 3/25 16 121

3/28 106 289 3/28 23 97

3/28 106 395 3/28 11 115

3/31 119 376 3/31 17 121

4/1 104 338 4/1 16 113

4/1 110 311 4/1 16 102

4/1 94 266 4/1 15 89

4/5 110 384 4/5 16 122

4/4 104 357 4/4 18 145

4/7 115 364 4/7 16 118

4/7 116 372 4/7 16 121

4/12 106 322 4/12 13 101

4/15 126 337 4/15 11 110

4/16 121 302 4/16 13 114

4/21 115 299 4/21 13 100

4/22 125 333 4/22 14 107

4/23 126 312 4/23 10 101

4/25 120 300 4/25 13 92

MEAN 112.1 328.0 14.9 110.1

SD 9.6084 39.5221 2.9340 13.2157

CV 8.6 12.0 19.6 12.0

• Atlantis dC18 IS Column, 3.5m, 2.1x20 mm• 10mM Ammonium Formate, pH 3.7/MEOH,

Opiate resolution• 0 to 40% methanol from 0 to 4 minutes, 100%

4 6 minutesThis modified method has shortened the analysistime, as well as, provided a selectivity change.The trimethoprim and metabolite elute well outsidethe retention time window for norfentanyl.

Page 8: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

MS Method: Spectrum from Scan. Retentiontime 11.0 minutes.

Interference: Modify MS Method (Dwell on 84 m/z)

Opiate Standard: LC Separation ...Importanceof Retention Time

Conclusions

• Confident ID is possible with single quadrupoleMS and automated TargetLynx monitoring ofRetention Time, CID fragments, Ion Ratios (up tofour) and S/N

• LC separation with MS detection providesneeded specificity to detect and confirm drugsin a clinical setting

• Sample preparation is reduced with LC/MS

• LC/MS is robust and requires little dailymaintenance

� Atlantis dC18 IS Column, 3.5µ, 2.1x20 mm

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00Time0

100

%

0

100

%

0

100

%

0

100

%

0

100

%

3.224.45

3.21

4.33

4.32

5.054.954.34

3.643.21 3.89

4.885.73

5.37 5.84

TMP patient:234.2m/z

TMP patient:177.2m/z

TMP patient:233.2m/z

FentanylNorF

TIC: TMP patient

Standard

NorF

In addition to fentanyl, several other opiates are ofimportance in this patient group. This new LC/MSanalysis, in the ammonium formate buffer, allowstheir simultaneous detection. Among the acceptanceparameters monitored by TargetLynx, the use of theretention time window flag allows positiveidentification of the drugs that have the samemolecular weight within this closely related family.

???

80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460 480m/z0

100

%

277.2

260.2

233.2

177.2100.0 151.1

278.3

279.3 392.3

NorFentanyl Ions233.2m/z177.2m/z

234.2151.1 177.2

The spectrum of the unidentified compound showsseveral ions that are also used for the analysis ofnorfentanyl.

y y g

9.50 10.00 10.50 11.00 11.50 12.00 12.50 13.00 13.50 14.00 14.50 15.00 15.50 16.00 16.50Time0

100

%

0

100

%

0

100

%

0

100

%

10.80

10.88

11.14

11.14

MS Scan: 260m/z

MS Scan: 177m/z

MS Scan: 233m/z

MS Scan: 84m/z

Norfentanyl

Norfentanyl

Tmp+Metab

Tmp+Metab

• SIM function: 84 m/z replaces 177.1 m/z• Choose New IS: Norfentanyl-d5 shares 84m/z• External standard method substitution when

dealing with this interference• Modify HPLC method: Goals include selectivity

change and shorten time of analysisA quick modification of the analysis methodresolved the problem within 45 minutes. The 84m/z ion was substituted for the 177 m/z. The ionratio had failed because of this 177 m/zinterference. The norfentanyl-d5 was removed asthe internal standard since the 84 m/z is also aproduct of the IS. An external standard methodwas substituted while the fentanyl assay wasmodified.

Page 9: kimlilly reprint edits - Waters Corporation · 93.3 104.8 120.0 132.2 150.1 Cone=20V Cone=30V Cone=40V Cone=60V Cone=70V Two CID fragments, 150.2 and 177.3 m/z, are easily detected

Post

erRE

PRIN

T

Sales OfficesAUSTRIA AND EXPORT (CENTRAL EUROPE, CIS, MIDDLE EAST,INDIA AND INDIA SUBCONTINENT) 43 1 8771807

AUSTRALIA 61 2 9933 1777

BELGIUM AND LUXEMBOURG 02 726 1000

BRAZIL 55 11 5543 7788

CANADA 800 252 4752 X2205

CIS 7 095 931 9193

CZECH REPUBLIC 42 02 617 11384

DENMARK 46 59 8080

FINLAND 358 09 506 4140

FRANCE 1 3048 7200

GERMANY 06196 40 06 00

HONG KONG 852 29 64 1800

HUNGARY 36 1 350 5086

INDIA 91 80 837 1900

IRELAND 1 4481500

ITALY 02 274211

JAPAN 81 3 3471 7191

KOREA 82 2 3284 1300

MEXICO 52 55 5524 7636

THE NETHERLANDS 76 508 7200

NORWAY 6 384 6050

PEOPLES REPUBLIC OF CHINA 86 10 8451 8918

POLAND 48 22 833 4400

PUERTO RICO 787 747 8445

SINGAPORE 65 6278 7997

SPAIN 93 600 9300

SWEDEN 555 11 500

SWITZERLAND 62 889 2030

TAIWAN 886 2 2543 1898

UK 44 208 238 6100

U.S.A. AND ALL OTHER COUNTRIES: WATERS CORPORATION34 Maple St.Milford, MA 01757 U.S.A.T: 508 478 2000F: 508 872 1990www.waters.com

Waters, Micromass, Oasis, XTerra, Atlantis, Intelligent Speed, MassLynx, TargetLynx, ZQ and Alliance are trademarks of Waters Corporation.All other trademarks are the property of their respective owners.©2003 Waters Corporation Produced in the U.S.A. June 2003 720000666EN SD-PDF

WATERS CORPORATION34 Maple St.Milford, MA 01757 U.S.A.T: 508 478 2000F: 508 872 1990www.waters.com