![Page 1: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/1.jpg)
Kilovoltage X-ray dosimetry
in the clinic
Philip Mayles
The Clatterbridge Cancer Centre
![Page 2: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/2.jpg)
Kilovoltage Codes of Practice
Germany DIN 1988 and 1996
UK IPEMB (IPEM) 1996 and 2005
IAEA TRS 277 2nd edn 1997
Holland NCS 1997
USA AAPM TG61 2001
IAEA TRS 398 2004
![Page 3: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/3.jpg)
Formula for Medium Energy
• Measure at 2 cm deep
IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996
In phantom
![Page 4: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/4.jpg)
Formula for Low Energy
• Measure in air
In air
IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996
![Page 5: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/5.jpg)
Formula for Low Energy
• Measure in air
AAPM COP Ma et al Med Phys 28 868-893 2001
In air
![Page 6: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/6.jpg)
Formula for Low Energy
• Measure in air
In air
IPEMB recommends use of the in air protocol if
the prescription is to the max dose for medium
energy
IPEMB Addendum Aukett et al Phys Med Biol 50 2739-2748 2005
![Page 7: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/7.jpg)
IPEM code for Very Low Energy
• Measure at surface of phantom
• Note that kch is based on measurement
comparing to in air measurement
IPEMB COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996
IPEMB Addendum Aukett et al Phys Med Biol 50 2739-2748 2005
In phantom
![Page 8: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/8.jpg)
Grenz ray chamber
PTW23342
In phantom
![Page 9: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/9.jpg)
Grenz ray chamber
PTW23342
In phantom
![Page 10: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/10.jpg)
How are Low and Medium Energy
defined ?
• IPEMB based on HVL
– Medium energy 0.5 - 4 mm Cu (>160 kV)
– Low energy 1 – 8 mm Al (50–160 kV)
– Very low energy 0.035-1 mm Al (8–50 kV)
![Page 11: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/11.jpg)
How is Energy defined ?
• Penetrative power
– Depth dose
– First Half Value Layer
• Energy spectrum
– Generating potential
– Filtration
![Page 12: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/12.jpg)
Measurement of HVL
In air
AAPM COP Ma et al Med Phys 28 868-893 2001
![Page 13: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/13.jpg)
Specification of Beam Quality
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
0 50 100 150 200 250 300
kV
HV
L m
m A
l
Standards Lab
TECDOC 1455
Clatterbridge
0
0.5
1
1.5
2
2.5
3
0 20 40 60 80 100 120
kV
HV
L m
m A
l
Standards Lab
TECDOC 1455
Clatterbridge
Low Energy Medium Energy
![Page 14: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/14.jpg)
Adapted from TRS398
Does Beam Quality Matter ?
HVL mm Al
Nk
![Page 15: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/15.jpg)
Factors in Calculation: M
• Electrometer reading corrected for
temperature, pressure, ion recombination,
polarity effect and electrometer accuracy
In air In phantom
AAPM COP Ma et al Med Phys 28 868-893 2001
![Page 16: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/16.jpg)
Factors in Calculation: M
• Electrometer reading corrected for
temperature, pressure, ion recombination,
polarity effect and electrometer accuracy
In air In phantom
IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996
![Page 17: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/17.jpg)
Factors in Calculation: M
• Recombination correction
VH and VL are the high and low voltages
and MH and ML are the meter readings
Important for short SSDs
In air In phantom
AAPM COP Ma et al Med Phys 28 868-893 2001
![Page 18: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/18.jpg)
Factors in Calculation: μen/ρ air
• AAPM and IPEMB agree to better than
0.3% but the sources of data were not
independent
In air
![Page 19: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/19.jpg)
Dose is tissue dependent μen/ρ air
AAPM COP Ma et al Med Phys 28 868-893 2001
In air
![Page 20: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/20.jpg)
Factors in Calculation: μen/ρ 2cm
In phantom
AAPM COP Ma et al Med Phys 28 868-893 2001
1 10 100
![Page 21: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/21.jpg)
Factors in Calculation: Bw
• Since the backscatter factor is
fundamentally a water-kerma ratio, reliable
measurements are non-trivial. Therefore
for the application of this protocol
backscatter factors should not be
measured in the clinic
In air
AAPM COP Ma et al Med Phys 28 868-893 2001
![Page 22: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/22.jpg)
Factors in Calculation: Bw
• Since the backscatter factor is
fundamentally a water-kerma ratio, reliable
measurements are non-trivial. Therefore
for the application of this protocol
backscatter factors should not be
measured in the clinic
• Largest difference AAPM:IPEMB is 0.4%
In air
AAPM COP Ma et al Med Phys 28 868-893 2001
![Page 23: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/23.jpg)
Factors in Calculation: Pstem
• 1.0 if field size for calibration is same as
measurement
• Likely to be important only for plane
parallel chambers
• Need to know the value for the reference
chamber
In air
![Page 24: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/24.jpg)
Factors in Calculation: kch
• AAPM: PQchamPsheath
• IPEM: kch
In phantom
![Page 25: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/25.jpg)
Factors in Calculation: kch
• AAPM: PQchamPsheath
• IPEM: kch
• TRS277: kupu
In phantom
![Page 26: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/26.jpg)
Phantom Materials
• Compared Solid Water (RMI Gammex) and
Plastic Water (Computerized Imaging Reference Systems) to H2O
• Percentage Depth Dose
– Solid water: up to 2.4% difference (75kVp beam 40mm deep)
– Plastic water: up to 23.2% difference (75kVp beam 2mm deep)
In phantom
Hill et al Phys Med Biol 50 N331-N344 2005
![Page 27: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/27.jpg)
Phantom Materials
• Compared Solid Water (RMI Gammex) and
Plastic Water (Computerized Imaging Reference Systems) to H2O
• Percentage Depth Dose
– Solid water: up to 2.4% difference (75kVp beam 40mm deep)
– Plastic water: up to 23.2% difference (75kVp beam 2mm deep)
• Absolute dose measurement:
In phantom
Hill et al Phys Med Biol 50 N331-N344 2005
![Page 28: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/28.jpg)
Are In Air and In Water measurements
equivalent ?
In air In phantom
Ma et al Med Phys 25 2376-2384 1998
300kV 3.67mm Cu
![Page 29: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/29.jpg)
Are In Air and In Water measurements
equivalent ?
In air In phantom
Ma et al Med Phys 25 2376-2384 1998
300kV 3.67mm Cu
1.22
0.964
![Page 30: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/30.jpg)
Issues with Backscatter Method
Thanks to Rhydian Caines and Richard Clements
![Page 31: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/31.jpg)
Depth Dose measurement
• Measurement of doses close to the
surface is difficult
• Ma et al recommend using an NACP
chamber but found differences of 2.4%
• With a diode the differences were up to
10%
In phantom
![Page 32: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/32.jpg)
New Devices
New uses are being proposed
Zeiss Intrabeam Ariane Papillon 50
![Page 33: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/33.jpg)
Methodology for New Devices
• Ebert et al have carried out a thorough
evaluation of the dosimetry
• However, the uncertainties in dosimetry
probably place the priority on consistency
between centres
• An intercomparison jig has been proposed
by Armoogum and Watson Ebert et al Med Phys 30 2424-2431 2003
Armoogum and Watson Z Med Phys 18 120-127 2008
![Page 34: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/34.jpg)
Uncertainty Estimates
![Page 35: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/35.jpg)
Summary
• Protocols give reliable results at the point
of measurement
• Accurate measurement of percentage
depth doses is problematic
• A protocol should be chosen which
measures the dose close to the
prescription point
![Page 36: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/36.jpg)
Kilovoltage Codes of Practice
Germany DIN 1988 and 1996
UK IPEMB (IPEM) 1996 and 2005
IAEA TRS 277 2nd edn 1997
Holland NCS 1997
USA AAPM TG61 2001
IAEA TRS 398 2004
![Page 37: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/37.jpg)
TRS 398
• Calibrate chamber in absorbed dose to
water
• is a chamber specific factor
which corrects for differences between the
reference beam and the actual beam
In phantom
![Page 38: Kilovoltage X-ray dosimetry in the · PDF fileKilovoltage X-ray dosimetry in the clinic Philip Mayles ... Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements](https://reader031.fdocuments.in/reader031/viewer/2022030507/5ab6cb5d7f8b9ab7638e18f0/html5/thumbnails/38.jpg)
Is TRS 398 the answer ?
• Only PTB offers an absorbed dose to water
calibration
• Other standards labs may offer a calibration
based on one of the protocols
• Jurado et al have used it with calibrations from
PTB
• IAEA TECDOC 1455 shows agreement within
3% at low energies
In phantom
Jurado et al BJR 78 1-12 2005