Jennifer TronsonVVC RSPT 233
10/10/2015
Neurally Adjusted Ventilatory Assist
![Page 2: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/2.jpg)
• Diaphragm drops• Thoracic cavity
expands• Pressure drops • Negative Pressure
Gradient is formed from the mouth to the alveoli
• Air flows in and down the pressure gradient until equilibrium is reached
How Do We Inhale?
![Page 3: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/3.jpg)
How Do Ventilators Know When a Patient Wants to Take a Breath?
• Diaphragm drops• Thoracic cavity expands• Pressure drops • Negative Pressure Gradient
is formed from the mouth to the alveoli
• Air begins to flow down the gradient
• The ventilator is triggered to deliver a breath by settings predetermined by the therapist or a measurement taken of previous breaths
• Or….
![Page 4: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/4.jpg)
How Do Ventilators Know When a Patient Wants to Take a Breath?
• Water in the tube alters the pressure or flow
• Patient movement triggersa breath
• No breath is delivered because the patient is unable to alterflow or pressure enough to trigger the ventilator
• Even if everything works according to plan there is a significant delay between the patient’s brain deciding it should take a breath and the ventilator delivering the breath
![Page 5: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/5.jpg)
The respiratory center responds to chemo and pressure receptors in the body and sends a signal through the phrenic nerve to the respiratory muscles to contract and expand the thoracic cavity.
How Does Our Body Know We Want to Take a Breath?
![Page 6: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/6.jpg)
• What if the ventilator could be triggered by the same signal that triggers the respiratory muscles?– Increased patient synchrony– Increased patient comfort– Potentially lower levels of sedation
• What if that signal could also tell the ventilator how much support is needed for each breath?– Support will be calculated based on the signal the brain
sends to the diaphragm so there will not be a lack of support if the patient is requiring a deeper breathe nor too much support if the patient is trying to take a smaller breath.
Is There a Better Way?
![Page 7: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/7.jpg)
NAVA - Synchrony Redefined
![Page 8: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/8.jpg)
Triggering Comparison
![Page 9: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/9.jpg)
• Neurally Adjusted– Utilizes a catheter to detect
the electrical activity signaling in the diaphragm Edi
– Uses the strength of the Edi to determine how much assist to deliver to the patient
– Bases each breath on the electrical signal, not trending from previous breaths
NAVA
![Page 10: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/10.jpg)
NAVA• Ventilatory Assist
– The signal from the Edi catheter is used to determine the amount of support supplied and when the support is supplied.
The support is provided in the same fashion no matter the means of delivery to the patient (Invasive or Non-Invasive), and is unaffected by leaks as the breath is cycled on and off by neural signal not preset volume or pressure.
![Page 11: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/11.jpg)
NAVA
• Ventilatory Assist– The support is supplied in synchrony and
proportion with the Edi signal – The amplification applied to this support is the
NAVA level NAVA level X (Edi signal-Edi min)+PEEP
![Page 12: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/12.jpg)
Synchronized Ventilatory Assist
![Page 13: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/13.jpg)
![Page 14: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/14.jpg)
• NAVA is a system that belongs to MAQUET
• It is compatible with the SERVO- I ventilator
• Has been available in the US since late 2007
• The largest adopter of the technology in Southern California is Loma
Linda, but it is also used in Kaiser, Children’s Hospital of Orange County, and newly in the Huntington Hospital
Background
![Page 15: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/15.jpg)
• In the US it is primarily used in NICU/PICU, though in Europe it is widely used in adults.
• This is attributed to the obvious “ah-ha” moment that happens when you switch a baby over to NAVA and see them stop fighting the vent, relax and succeed on lower pressures and lower FiO2s
• Several factors have attributed to its slow adoption in adult markets in the US, Drs and therapists not wanting to relinquish control, learning curve, inertia…
Background
![Page 16: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/16.jpg)
Background
![Page 17: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/17.jpg)
• Edi catheter ‘s differ in iED making them uniquely suited to patients based on height.
• Only the 12 and 16 Fr sizes have a lumen for evacuation
Setup - Select the appropriate Edi Catheter
![Page 18: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/18.jpg)
• Insert the Edi Module into the SERVO-I
• Connect the Edi Cable to the Edi Module
• Perform Edi Module Function Test
Setup – Perform Edi Module Function Test
![Page 19: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/19.jpg)
Setup – Position the Edi Catheter
Measure NEX & Calculate Insertion Distance
![Page 20: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/20.jpg)
• Inspect the catheter
• Put the tip in sterile water – no other lubricant
• Allow the tip to remain in the water for 5 seconds before removing
Setup – Position the Edi Catheter
![Page 21: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/21.jpg)
• Insert the catheter to the Y value determined earlier• Connect the catheter to the Edi module
Setup – Position the Edi Catheter
![Page 22: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/22.jpg)
• Once you verify that the 2 middle rows are highlighted in blue note the insertion depth and secure the catheter in place.
Setup – Verify the Catheter Position
![Page 23: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/23.jpg)
• Edi signal can be seen in all modes including standby
• Edi Peak and Min are available for each breath cycle
• Signal trending for 24 hours are available
Setup – Monitor the Edi Signal
![Page 24: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/24.jpg)
• Open the NAVA preview window with the Neural Access key
• Set the initial NAVA level to achieve the same or slightly lower pressure level that is currently being delivered to the patient
• Press the NAVA level key then use the rotary dial to adjust the level
Setup – Set the Initial NAVA Level
![Page 25: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/25.jpg)
• Utilize the calculation to compare your pressure levels
Setup – Select and Set NAVA Mode
![Page 26: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/26.jpg)
• Set PEEP• Set FiO2• Set Trigger Edi
– This detects increases in Edi and should be set to filter out background noise so that it does not affect the patient’s ability to trigger a breath
Setup – Select and Set NAVA Mode
![Page 27: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/27.jpg)
• Set Pressure Support above PEEP– Must be set low
enough to not induce hyperinflation and interfere with the Edi signal
• Set Backup Ventilation– This is AC/PC
mode that is utilized if the patient becomes unable to signal for a breath
Setup – Select and Set NAVA Mode
![Page 28: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/28.jpg)
• Set Alarms that are adequate for the patient and in line with your facilities alarm limit policies.
Setup – Select Alarm Limits
![Page 30: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/30.jpg)
• Titrate NAVA level to obtain Edi Values that are 60% of the Edi max value during the SBT.
• Repeat at least once daily
• Extubate when SBT is successfully tolerated for 2 hours
Weaning
![Page 31: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/31.jpg)
• Any contraindications for naso-orogastric feeding tube such as upper airway surgery and skull base fractures
• NAVA would not be appropriate for a patient exhibiting an undesirable breathing pattern such as Biot’s breathing though the Edi information may remain a useful tool
• Known phrenic nerve lesions• Severe hiatal hernia• Need for MRIs as the Edi catheter is not approved
for use in MRI environments• NAVA may be contraindicated in a patient that is
experiencing seizures or frequent panic attacks that alter the normal breathing patterns
Contraindications
![Page 32: Jennifer Tronson VVC RSPT 233 10/10/2015 Neurally Adjusted Ventilatory Assist.](https://reader036.fdocuments.in/reader036/viewer/2022062422/56649f355503460f94c52f55/html5/thumbnails/32.jpg)
• http://ww2.maquet.com/uk/product/NAVA?tab=Features
• http://www.ncbi.nlm.nih.gov/pubmed/22015130
• http://rc.rcjournal.com/content/56/3/327.full
• http://www.respiratorytherapyfiles.net/uploads/2/6/3/1/26319264/servo_i.pdf
• http://www.researchgate.net/publication/41056512_Patient-ventilator_interaction_during_pressure_support_ventilation_and_neurally_adjusted_ventilatory_assist
• http://respiratory-care-sleep-medicine.advanceweb.com/Magazine/References/NAVA-Initial-Setup-Guide.aspx
• http://www.dicardiology.com/node/25122/3
• http://www.criticalcarenews.com/upload/pdf/CCN_case26_MX-1030_CCN_case26_SERVOI_NAVA_Weaning_Adults_EN_NONUS.pdf
• http://www.criticalcarenews.com/CcnTrainingPage.aspx?area=1&ContentTypeId=16&TrainingID=29
Works Cited