Micro-sized IVF
labDr. Yasmin Magdi
Abd-Elkreem
![Page 2: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/2.jpg)
IVF Unit• The IVF Unit may consist of a number of Functional Zones:ENTRY/ CONSULT ZONE: Entry/ Reception and waiting areas Administration/ Records Interview Room/s Consult/ Examination/ Treatment Room/s Ultrasound room/s Collection Room/s with shower and Toilet Public ToiletsPATIENT PROCEDURAL ZONE: Operating Room/s Recovery areas Change areas and toilets for staff and patients
LABORATORY ZONE: Laboratories (Embryology, Andrology, Genetics) Cryopreservation facilities Gas Bottle StoreSTAFF AND SUPPORT ZONE: Clean-up and Disposal room Store rooms and Sterile store Offices, meeting rooms, staff room Sterilising area:
![Page 3: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/3.jpg)
Services provided by the IVF Unitinclude: Patient consultation and interview on an outpatient basis Pre treatment assessment Blood collection Semen collection Artificial insemination Ovarian stimulation therapy Ultrasound examination Oocyte (egg) collection Embryo culture In vitro / ICSI fertilisation Cryopreservation Embryo transfer
![Page 4: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/4.jpg)
In vivo is not in vitro• in vitro environment is a very stressful place to be: temperature
and pH fluctuate; light exposure and high levels to oxygen. In addition, nutrients and energy sources may not be correct during all developmental stages.
• A prerequisite for implantation and successful pregnancy, in vitro induced stress factors must be reduced to a minimum.
• The ART laboratory has been designed to provide an environment that is as close to optimum as possible for the growth of human embryos and to provide the best resulting pregnancy rates for patients undergoing IVF cycles.
• The layout of the laboratory must be sensibly planned and logically designed to allow smooth running of routine procedures.
![Page 5: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/5.jpg)
ART laboratory set-upFocus on three keys area1. Place• Design• Equipment
Working equipment Air quality equipment QC and monitoring equipment Stainless steel furniture
• Consumables• Paints and adhesives 2. Stuff• Manager• Clinical embryologists• Dress
3. Procedures• Standard operation protocols• Quality control • Double witness
![Page 6: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/6.jpg)
Incubators• Embryos spend the majority of their time within incubator
• Function: Provides a stable and appropriate culture environment required for optimizing embryo development and clinical outcomes.
• Selection: Type of gas sensors, temperature regulation methods and size of the incubator are different variables and important for selecting the working incubator.
• Management : (Air quality, temperature, humidity, gas monitoring and recovery, sterilization)
![Page 7: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/7.jpg)
Time Lapse incubator
Bench top incubator
Mini-incubator
Middle sized incubator
![Page 8: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/8.jpg)
Gas cylinders• Co2 (5-7%) required for:
1. maintain pH of bicarbonate buffered media2. incorporate into protein and nucleic acids by the embryos.
• In vivo, oocytes and embryos are exposed to a maximum of ~5-8% O2 in the reproductive system.
• Atmospheric O2 may lead to supraphysiological ROS levels, potentially causing oxidative stress (damage to cell organelles, lipids, membranes, DNA, gene expression), and ultimately poor embryo development.
• Reduced O2 concentration results in enhanced development in vitro.• Management: (in a separated gas room, alarm system, regulators,
responsible personnel, spare line and cylinders, clean the room and cylinders, color of cylinders, time and date of change)
![Page 9: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/9.jpg)
Laminar flow• Types & mechanism
![Page 10: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/10.jpg)
Laminar flow• Function: 1.Clean work station for gametes and embryos
manipulation. 2. Media preparation.
• Management: (Air quality, temperature, decontamination, UV, humidity and gases if close chamber)
![Page 11: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/11.jpg)
Heating surfaces• Separated unit or included in the laminar flow. • Function: are used for maintaining the temperature at
37°C.• Management: (routinely and sent for recalibration to
ensure that the temperature level is accurate, calibration if not accurate).
![Page 12: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/12.jpg)
Micromanipulation system• Inverted microscope: The light source is at the top (above the stage)
and the objectives (lenses) are below the stage (upside down as compared to conventional microscopes) equipped with 4x, 10x, and 20x objectives.
• The microscope stage: (where the dish containing gametes or embryos will be placed) is heated to 37 degrees C (body temperature).
• Enhanced with two identical micro-manipulators to can handle embryos and gametes (capable of three dimensional movements).
• The microinjector consists of a syringe and silicone tubing connected to the microtool at the end of the tubing. (holding the oocyte and sperm)
• Function: 1. Intracytoplasmic sperm injection. 2. Fertilization, cleavage, blastocyst assessment. 3. Laser assisted hatching.4. Blastomer biopsy • Management: (temperature , cleaning).
![Page 13: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/13.jpg)
Pipette• The capacity of a micropipette can range from less than 1 µl to
1000 µl (1 ml), while macropipettes can measure volumes greater.• Function: 1. accurately measure and dispense small volumes of
media to prepare dishes. 2. Embryo manipulation. 3. Ovum pack up.
• Types used in IVF: 1. Volumetric with disposable tips. 2-Pasteur
![Page 14: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/14.jpg)
QC and monitoring equipment
pH meter
Monitor your lab conditions at all times
Co2, o2, temperature analyzer for incubators
Thermometer
![Page 15: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/15.jpg)
Air quality equipment • Importance of Air quality in IVF High levels of VOC in unfiltered IVF lab air Efficient particle and VOC removal by air filtration Embryo development inversely correlated with VOC Better outcome (PR, IR, miscarriage) in IVF cycles performed
laboratories with air particle and VOC filtration
• Types of filters: 1. HEPA (High efficiency particulate air) filters 2. ULPA (ultra-low penetration air) filters 3. VOC (volatile organic compound) Filters
HEPA ULPA
![Page 16: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/16.jpg)
Positive Pressure system• The laboratory should be pressurized to prevent air from adjacent
rooms entering the laboratory.• Function: 1. Reduce microbial flora to minimum
2. Reduce VOC 3. exclude contamination coming from surrounding areas
with the provision of fresh air intake to replace C02 Gas released in the IVF Lab.
4. protect the embryos but also the embryologist.• 4 stages filtration • Management: (change filters, maintenance, choice the time and
frequancy of switch on).
![Page 17: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/17.jpg)
Air purification system• Function: Clean air pollutants, dust particles, mould spores,
dander, pollen, dust mites, cleaning chemicals, volatile organic compounds (VOCs), chemically active compounds (CACs), aldehydes, carcinogenic materials, carbon monoxide, viruses and bacteria.• Mostly portable.• Air filters + UV lamp• Management: (Daily check, change filters and UV lamp, increase
velocity according to the level of pollution)
![Page 18: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/18.jpg)
Liquid nitrogen tanks • Function: store gametes and embryos at -196°C• Liquid nitrogen, when returned to the gaseous state, can displace
oxygen from the air under the right conditions.
• It may be prudent to install oxygen monitors in areas where liquid nitrogen is stored and ventilation is minimal.
• Management: (Daily measure it’s level, refilling tank always ready, keep closed, always be stored in the upright position, always wear safety gloves, high concentrations of escaping gas should not be allowed to collect in an enclosed area, avoid prolonged breathing of cryogenic liquid vapors ).
![Page 19: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/19.jpg)
Centrifuge• Function: 1. required to prepare sperm (Fresh or frozen Semen,
TESE, FNA) for IVF insemination or intracytoplasmic spermatozoa injection (ICSI).
2. Sperm Cryopreservation.
• Management: regularly cleaned and maintained.
![Page 20: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/20.jpg)
Refrigerator• Function: keep the ( medias , oil, Kits )• Should be adapting with its requirement ( big or small ).• Management: (temperature check twice and adjusted according
to manufacture instructions, never use for something else, one for stock and one working, decontamination)
![Page 21: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/21.jpg)
Equipment…….. Which one???• Proven or established models that have been used successfully in
other successful centers should be bought. • highest quality and reliability• safe to use.• Regular servicing and maintenance of equipment is mandatory.
For this• reason, equipment that has maintenance support and reliable
breakdown service should be used. • It is important that the suppliers have spare equipment as backup
in the event of any malfunction of equipment.
![Page 22: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/22.jpg)
Paints and adhesives • Walls of the laboratory should be painted with non odor emitting,
nontoxic, glossy, non-vaporizing paint. Ceiling should be solid. Flooring should be slip proof preferably vinyl which is easy to clean.
• Lighting should be adjustable by dimmer. There should be enough electrical outlets.
• There should be a wall mounted data logger displaying the laboratory ambient temperature and humidity levels.
• The entire walls should be washed with sterile distilled water and alcohol weekly, and a yearly maintenance by the suppliers is recommended.
![Page 23: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/23.jpg)
Stuff• There should be adequate number of staff to do all the procedures. • The laboratory should be directed by a qualified, experienced, and
responsible person with expertise in the field of embryology.
• Clinical embryologists should have hands-on experience in all facets of clinical embryology which is an absolute requirement when starting a new setup.
• Embryologists should be given opportunities to increase their knowledge and competence through attending conferences, workshops, and continuing educational programs.
• Scrub • Proficiency Testing
![Page 24: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/24.jpg)
Procedures• All activities performed within the ART laboratory need to be clearly
written with sufficient details on the techniques (Called slandered operation protocols; SOPs).
• All SOP’s should be signed and dated and should be made available to all staff.
• There should be strict adherence to the SOP.
• Non adherence to SOP’s should be dealt with seriously and should not be excused.
• Witness: When doing procedures there should be at least two identifiers for patients’ specimen. The name and an additional identifier such as an identity number should be used.
![Page 25: Ivf lab microsized](https://reader035.fdocuments.in/reader035/viewer/2022062503/588722bd1a28abfb0b8b5ee3/html5/thumbnails/25.jpg)
Conclusions…………..• The ART laboratory plays a crucial role in the treatment of infertile couples and
setting up a state of the art laboratory is no easy task.
• To setup a quality ART laboratory, the combination of the three key areas will provide the main framework.
• The main goal of the laboratory should be to provide a safe and secure environment while maintaining optimal parameters for embryonic development.
• The embryologists selected should be experienced and highly skilled and be able to deliver consistent results.
• Quality standards are needed to ensure consistency and reproducibility of all methods.
• Procedures and processes should be clearly written and available. For a successful ART program, the ART laboratory should establish and maintain strict quality controls.