1 ISRRT Chesney Research Grant 2016 Project: An investigation into the effectiveness of thyroid shielding as a dose optimisation tool when carrying out radiographic imaging prior to, or during, orthodontic treatment of children By Rebekah Goulston Radiographer Manchester University NHS Foundation Trust Dental Radiology Department University Dental Hospital of Manchester Higher Cambridge Street Manchester UK The following is a general summary of the research study that was funded by the ISRRT Chesney Research Fund following a call for research proposals on the theme ‘A novel approach on optimization and/ or justification of medical exposures by radiographers for radiation protection and safety.’ Complete data will not be presented here but will be published in due course in a peer reviewed academic journal. Approximately one fifth of all X-ray examinations are performed by dentists and a high proportion of those images are carried out on children and young people. Dental radiology predominantly involves low dose techniques; 1,2 however, some radiosensitive organs and tissues are in close proximity to the radiation beam and/or scatter, including the thyroid gland. 3 Globally, guidelines on the use of lead collars to protect the thyroid differ widely and the evidence-base is limited. For example, US guidelines specify that thyroid shielding must always be used for dental radiography, 4,5 while UK guidelines only suggest that it is used if the thyroid gland is in line of the primary beam. 6 The development of Cone Beam CT (CBCT) for dental examinations, which has radiation doses typically an order of magnitude greater than those of conventional radiography and sometimes higher, has raised the potential importance of shielding. It has also highlighted the impact that high atomic number materials used in thyroid shields can have on image quality. 7 Limiting X-ray beam size to the smallest required is an obvious means of dose limitation. Rectangular collimation of the X-ray beam for intraoral radiography has been recommended for at least two decades and is now available from manufacturers as a standard equipment specification. A recent UK survey found that 67% of intraoral X-ray sets had rectangular collimation. 8 It is possible that a reduced beam size might itself lower thyroid dose by excluding it from the primary beam and reducing scatter. If so, thyroid shielding might have less impact on dose or none at all. This research project was therefore designed to provide evidence as to whether putting lead shielding on young patients to protect the thyroid significantly reduces the radiation dose to the thyroid during a variety of dental radiographic examinations. It also aimed to identify if wearing a thyroid collar had an impact on the image quality, and therefore the diagnostic value of extraoral images taken, especially in the case of CBCT. A final objective was to identify if rectangular collimation can reduce thyroid dose when taking intraoral dental radiographs.

Transcript of ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are...

Page 1: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

1

ISRRT Chesney Research Grant – 2016 Project: An investigation into the effectiveness of thyroid shielding as a dose

optimisation tool when carrying out radiographic imaging prior to, or during, orthodontic treatment of children

By Rebekah Goulston Radiographer Manchester University NHS Foundation Trust Dental Radiology Department University Dental Hospital of Manchester Higher Cambridge Street Manchester UK

The following is a general summary of the research study that was funded by the ISRRT Chesney Research Fund following a call for research proposals on the theme ‘A novel approach on optimization and/ or justification of medical exposures by radiographers for radiation protection and safety.’ Complete data will not be presented here but will be published in due course in a peer reviewed academic journal. Approximately one fifth of all X-ray examinations are performed by dentists and a high proportion of those images are carried out on children and young people. Dental radiology predominantly involves low dose techniques;1,2 however, some radiosensitive organs and tissues are in close proximity to the radiation beam and/or scatter, including the thyroid gland.3 Globally, guidelines on the use of lead collars to protect the thyroid differ widely and the evidence-base is limited. For example, US guidelines specify that thyroid shielding must always be used for dental radiography,4,5 while UK guidelines only suggest that it is used if the thyroid gland is in line of the primary beam.6 The development of Cone Beam CT (CBCT) for dental examinations, which has radiation doses typically an order of magnitude greater than those of conventional radiography and sometimes higher, has raised the potential importance of shielding. It has also highlighted the impact that high atomic number materials used in thyroid shields can have on image quality.7 Limiting X-ray beam size to the smallest required is an obvious means of dose limitation. Rectangular collimation of the X-ray beam for intraoral radiography has been recommended for at least two decades and is now available from manufacturers as a standard equipment specification. A recent UK survey found that 67% of intraoral X-ray sets had rectangular collimation.8 It is possible that a reduced beam size might itself lower thyroid dose by excluding it from the primary beam and reducing scatter. If so, thyroid shielding might have less impact on dose or none at all. This research project was therefore designed to provide evidence as to whether putting lead shielding on young patients to protect the thyroid significantly reduces the radiation dose to the thyroid during a variety of dental radiographic examinations. It also aimed to identify if wearing a thyroid collar had an impact on the image quality, and therefore the diagnostic value of extraoral images taken, especially in the case of CBCT. A final objective was to identify if rectangular collimation can reduce thyroid dose when taking intraoral dental radiographs.

Page 2: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

2

To be able to ascertain whether thyroid shielding and/or rectangular collimation are an effective way to reduce the dose to the thyroid received by children and young adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry scenarios in a manner similar to the method previously used by Hidalgo et al.9 An anthropomorphic phantom (Figure 1b) filled with thermoluminescent dosemeters (TLDs) at the level of the thyroid (Figure 1c), was positioned and exposed to radiation as if it were a patient undergoing the examination (including scout projections for CBCT).

Page 3: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

3

Figure 1: Images of Cone Beam Computed Tomography study investigations being

undertaken

Figure 1

a. The 3D Accuitomo CBCT equipment made by the J. Morita Company, Japan; installed at the University Dental Hospital of Manchester.

b. The ATOM® 706-C anthropomorphic phantom (Computerized Imaging Reference Systems Inc., Norfolk, VA) representing a 10-year-old child positioned in the CBCT scanner

c. One of the slices of the ATOM® 706-C with 4 TLDs loaded in (circled in red)

d. The small adult female skull from a historic collection held in the Division of Dentistry at the University of Manchester in a head-shaped water bath to simulate soft tissues being imaged in the 3D Accuitomo F170 CBCT scanner (J. Morita, Kyoto, Japan)

e. A screen shots of a lower third molar image bank scan in the Image J software (version 1.52g, National Institutes of Health, Bethesda, MD) with contrast-to-noise measurements being taken.

All imaging machines used are in the clinical department at the University Dental Hospital of Manchester.

a b

d

e

c

Page 4: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

4

The scenarios and types of thyroid shielding tested can be seen in Table 1. Two basic designs of thyroid shield were used:

RothbandTM 0.25mm Pb (Rothband & Co. Ltd, Rossendale, UK.) – a flexible collar design fixed using Velcro at the back of the neck

Protectoray 0.5 mm lead lined thyroid shield (Hager & Werken, Duisburg, Germany) – a rigid flat design normally held in place by the patient in clinical use

Many alternative thyroid shields are available, with designs varying in size and shape, but these two were chosen to represent those most typically used by dentists. The rigid design of the Protectoray shield results in it being impractical for CBCT and Panoramic radiography examinations as it impedes correct patient positioning and in some cases the rotation of the machine. The Protectoray shield was therefore only used for the intraoral scenarios being considered.

Table 1 – Summary table of dosimetry scenarios in this study and how they were investigated

Examination Types of thyroid collar With and without

rectangular collimation

RothbandTM

0.25mm Pb (Rothband & Co. Ltd, Rossendale,

UK.)

Protectoray 0.5 mm lead lined thyroid shield

(Hager & Werken, Duisburg, Germany)

Intraoral anterior periapical

Intraoral upper maxillary occlusal

Panoramic radiograph N/A

Small FoV* CBCT scan of the anterior mandible N/A

Small FoV* CBCT scan of the posterior mandible (impacted wisdom teeth)

N/A

Small FoV* CBCT scan of the anterior maxilla

N/A

*FoV stands for field of view

The results of this first section of the study demonstrated that rectangular collimation can statistically reduce the radiation dose to the thyroid for periapical radiography (Figures 2b, 2c & 2d); but that a shielding device such as the Protectoray (Figures 2a & 2c) is required to reduce radiation dose to the thyroid significantly for the anterior maxillary occlusal projection. The results also suggested that a thyroid collar reduces the radiation dose received by the thyroid in patients undergoing dental CBCT examinations in many cases. However, the results demonstrated that the use of lead shielding does not reduce the dose to the thyroid when a panoramic radiograph is being taken (Figure 2e).

Page 5: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

5

Figure 2: Images of two dimensional imaging study investigations being undertaken

a b

c d

Figure 2

The ATOM® 706-C anthropomorphic phantom (Computerized Imaging Reference Systems Inc., Norfolk, VA) representing a 10-year-old child positioned for two-dimensional dosimetry studies

a. Occlusal imaging using the Planmeca Intra (Planmeca

Oy, Helsinki, Finland) with the Protectoray 0.5 mm lead

lined thyroid shield (Hager & Werken, Duisburg,

Germany) & no rectangular collimation

b to d show periapical imaging using the Planmeca Intra (Planmeca Oy, Helsinki, Finland) and a holder for paralleling technique with:

b. No thyroid collar or rectangular collimation

c. A Protectoray 0.5 mm lead lined thyroid shield (Hager

& Werken, Duisburg, Germany) & rectangular

collimation

d. A RothbandTM 0.25mm Pb (Rothband & Co. Ltd,

Rossendale, UK.) thyroid collar & rectangular

collimation

e. Panoramic radiographic examination set up with a RothbandTM 0.25mm Pb (Rothband & Co. Ltd, Rossendale, UK.) thyroid collar in the Planmeca Promax (Planmeca Oy, Helsinki, Finland)

All X-ray equipment used are in the clinical department at the University Dental Hospital of Manchester.

e

Page 6: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

6

The second phase of the study investigated the impact that wearing a thyroid collar has on image quality when undertaking small field of view CBCT examinations. To do this a bank of images was created of the two small field of view mandible scenarios, using a small adult female skull from a historical collection held in the School of Dentistry of the University of Manchester positioned in a head-shaped water bath to simulate soft tissues following the method of Shelley et al.10 (Figures 1d & 1e). 50% of the scans were taken with the thyroid collar in-situ and 50% without. Five specialist dental and maxillofacial consultant radiologists, who were blinded to thyroid collar use, then gave their opinions on the image quality of each scan via a questionnaire. In addition, the image quality of each of the CBCT scans was measured quantitatively using contrast-to-noise ratio on multiple scan slices. The results of both the subjective and quantitative image quality studies carried out in this phase of the study demonstrated that CBCT image quality was not reduced, either quantitatively or subjectively, by thyroid shielding. The results of the study are helpful in expanding the evidence base in this area of radiation protection in dental radiography. The key points for practitioners to consider include:

Rectangular collimation is the most effective way of reducing radiation dose to the thyroid when performing intraoral periapical imaging, but the RothbandTM 0.25mm Pb thyroid collar also resulted in a significant dose reduction to the thyroid if rectangular collimation is not available.

When maxillary occlusal projections are undertaken, radiation dose to the thyroid can be reduced by the use of a Protectoray thyroid shield. However, the RothbandTM 0.25mm Pb thyroid collar does not reduce the thyroid dose.

The use of a RothbandTM 0.25mm Pb thyroid collar can reduce the dose to the thyroid of young patients undergoing CBCT imaging without reducing the diagnostic quality of the resultant images.

However, it also needs to be noted that there was a difference in effectiveness of the two thyroid shielding designs, so a general recommendation to use thyroid shields may be too simplistic. The two thyroid shield designs in this study are not the only ones available, so caution is needed before extrapolating the results of this study to other designs.

Page 7: ISRRT Chesney Research Grant 2016 Project: An investigation into …€¦ · adults when they are undergoing dental radiographic imaging, the research team set up a series of dosimetry

7

References 1. UNSCEAR 2000 Report to the General Assembly, with scientific annexes.

Volume 1: Sources. Annex D: Medical Radiation Exposures. United Nations Scientific Committee on the Effects of Atomic Radiation, Vienna, 2000.

2. Horner K. Dental radiology: the forgotten problem? In: Justification of Medical Exposure in Diagnostic Imaging. Proceedings of an International Workshop Held in Brussels, Belgium, 2–4 September 2009. International Atomic Energy Agency, Vienna 2012.

3. International Commission on Radiological Protection. The 2007 recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007; 37:1–332.

4. American Thyroid Association. Policy statement on thyroid shielding during diagnostic medical and dental radiology, 2013. Available from: http://www.thyroid.org/wp-content/uploads/statements/ABS1223_policy_statement.pdf (Accessed: 24th March 2016).

5. American Academy of Oral and Maxillofacial Radiology. Clinical recommendations regarding use of cone beam computed tomography in orthodontics. Oral Surg Oral Med Oral Pathol Oral Radiol 2013; 116: 238–57.

6. Isaacson KG, Thorn AR, Atack NE, Horner K, Whaites E. Guidelines for the use of radiographs in clinical Orthodontics. Fourth edition. London, British Orthodontic Society, 2015.

7. Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, Schoemer E. Artefacts in CBCT: a review. Dentomaxillofac Radiol 2011;40: 265-273.

8. Holroyd JR, Smith JRH, Edyvean S. PHE-CRCE-51: Dose to patients from dental radiographic X-ray imaging procedures in the UK – 2017 review. Public Health England, London, 2019.

9. Hidalgo A, Davies J, Horner K, Theodorakou C. Effectiveness of thyroid gland shielding in dental CBCT using a paediatric anthropomorphic phantom. Dentomaxillofac Radiol 2015; 44: 20140285.

10. Shelley AM, Brunton P, Horner K. Subjective image quality assessment of cross sectional imaging methods for the symphyseal region of the mandible prior to dental implant placement. J Dent. 2011 Nov;39(11):764-70.